Calibration Copy Working - TIAERtiaer.tarleton.edu/transfer/Shepherd/QAM-RI-107rev... · QAM-RI-107...
-
Upload
trinhduong -
Category
Documents
-
view
216 -
download
2
Transcript of Calibration Copy Working - TIAERtiaer.tarleton.edu/transfer/Shepherd/QAM-RI-107rev... · QAM-RI-107...
![Page 1: Calibration Copy Working - TIAERtiaer.tarleton.edu/transfer/Shepherd/QAM-RI-107rev... · QAM-RI-107 Operation and Calibration of the Canberra Alpha Analyst Spectrometer Rev.0 Page](https://reader031.fdocuments.us/reader031/viewer/2022022512/5ae81c347f8b9a9e5d900ecd/html5/thumbnails/1.jpg)
QAM-RI-107
Operation and Calibration of the
Canberra Alpha Analyst Spectrometer
Revision: 0
{1ra /ULaborato ry Manager/LQAO/RSO
Renewal Date:
t' -lo ,i L
Date
Effective Date: f / r^ ,
Date
lnitials:
Texas institute for Applied Environmental Research
Work
ing C
opy
![Page 2: Calibration Copy Working - TIAERtiaer.tarleton.edu/transfer/Shepherd/QAM-RI-107rev... · QAM-RI-107 Operation and Calibration of the Canberra Alpha Analyst Spectrometer Rev.0 Page](https://reader031.fdocuments.us/reader031/viewer/2022022512/5ae81c347f8b9a9e5d900ecd/html5/thumbnails/2.jpg)
QAM-RI-107 Operation and Calibration of the Canberra Alpha Analyst Spectrometer
Rev.0 Page 2 of 12 TIEAR
1.0 Applicability and Purpose
A. This procedure applies to the operation and calibration of the
Canberra Alpha Analyst Spectroscopy system using Apex
Alpha and Genie 2000 software in the TIAER Laboratory at
Tarleton State University, Stephenville, Texas. The procedure
outlined in this text ensures that there is consistency between
trained analysts to produce accurate and precise data of high
quality. In performing these calibration and confirmation steps
the analyst reduces anomalies caused by instrument sensitivity,
interference and fluctuations.
B. Principals of alpha spectroscopy: Alpha emitting sample
sources are placed on the sample holders arranged in certain
geometries in the alpha spectrometer and a vacuum is pulled
on the spectrometer chambers. When alpha particles are
emitted by the sample, the alpha detectors produce electronic
signals that are analyzed by the MCA (multi-channel analyzer)
to generate a digital spectrum equivalent to the energy of the
alpha particles. The spectrum is then analyzed by computer
software that identifies and quantifies the alpha energies.
C. Operation summary: The alpha spectrometer is set-up and
ready for operation. The pulser and background checks are
performed, and the instrument is calibrated on a continuing
basis. For operation of the instrument performance
measurements of the continuing quality control standards,
spikes, duplicates and blanks are analyzed with each sample
batch.
D. Alpha-emitting radionuclides (or their short-lived decay
progeny) on the sample test sources may contaminate analyzer
chambers. The contaminants may emit alpha rays with peaks
at energies that cannot be adequately resolved from the tracers
or analytes of following sample test sources. The contaminants
are checked and removed before sample analysis with proper
instrument maintenance and sample preparation steps.
2.0 Definitions
Work
ing C
opy
![Page 3: Calibration Copy Working - TIAERtiaer.tarleton.edu/transfer/Shepherd/QAM-RI-107rev... · QAM-RI-107 Operation and Calibration of the Canberra Alpha Analyst Spectrometer Rev.0 Page](https://reader031.fdocuments.us/reader031/viewer/2022022512/5ae81c347f8b9a9e5d900ecd/html5/thumbnails/3.jpg)
QAM-RI-107 Operation and Calibration of the Canberra Alpha Analyst Spectrometer
Rev.0 Page 3 of 12 TIEAR
A. Refer to QAM-R-100, “TIAER Laboratory Radiochemistry
Program” for general radiological terms
B. FWHM- Full Width Half Maximum- the point on a spectral peak
where the surrounding area is most effectively calculated
C. ROI- Range of Interest- viewing area for energy peaks
D. MDC- Minimum Detectable Concentration
3.0 Equipment and Reagents
A. Computer with Apex Alpha and Genie 2000 software B. Alpha Analyst Spectroscopy system with multiple chambers C. Vacuum pump D. NIST traceable standard sources, Echert & Ziegler 7400-SRC
containing 234U, 238U, 239Pu, and 241Am, or equivalent
4.0 Procedure
A. The analyst is trained in accordance with QAM-Q-107 and approved by the RSO and LM to operate the instrument. All other aspects of the Quality Assurance Manual are followed.
B. Instrument start
1. Power on the instrument and computer.
2. Start Apex software and log on.
3. Make sure vacuum pump is on and working properly before
starting operation of the instrument.
C. Pulser check and detector background check:
1. Pulser check
a. Click on the “Sample Assigner” tab. Select detectors for
pulser check.
b. Click on the “Pulser Check” button.
c. Click on the “Load Samples” button. A screen with the list
of detectors will appear. Click on “OK”. The pulser count
will start for each available detector. The count time is 5
minutes. To view the pulsers, click on the “Main” tab. After
the pulsers have finished counting, click on the “Reports”
tab.
d. In the “Reports” window, click on the “QA Summary
Report” tab.
Work
ing C
opy
![Page 4: Calibration Copy Working - TIAERtiaer.tarleton.edu/transfer/Shepherd/QAM-RI-107rev... · QAM-RI-107 Operation and Calibration of the Canberra Alpha Analyst Spectrometer Rev.0 Page](https://reader031.fdocuments.us/reader031/viewer/2022022512/5ae81c347f8b9a9e5d900ecd/html5/thumbnails/4.jpg)
QAM-RI-107 Operation and Calibration of the Canberra Alpha Analyst Spectrometer
Rev.0 Page 4 of 12 TIEAR
e. Select the Pulser button then click on the “Check All”
button.
f. Click on the “View” button to get the pulser report on the
screen.
g. If needed, print the pulser report, or export to an electronic
file and obtain the required approval.
2. Detector background check
a. Click “Sample Assigner” tab, select detectors for
background check.
b. Click on the “Background Check” button.
c. Click on the “Load Samples” button. A screen with the list
of detectors will appear. Click on “OK”. The background
check count will start for each available detector. The
count time is 30 minutes.
d. After the background check has finished counting, click on
the “Quality Assurance View” tab and click on "Charts and
Reports". Select chamber and geometry, select
“Background Check”, and select the report of “Last
Measurement” to see the last check, or “Control Chart” to
see the chart graph. If the data are out of control limits,
investigate the reason, take correction action and
documented it.
e. Print out or save the check results and file.
f. Record results on the control chart for the detector and
geometry used.
g. Every control chart has a warning limit (2σ) and an action
limit (3σ). The LM is notified when the data are out of
warning limit, and a Corrective Action Report generated
when the data are out of control limit. The control chart is
established based on all data accumulated from an
assigned date to present (See QAM-Q-100).
D. Initial calibration 1. Calibration is performed upon initial set-up, after major
repair or service, or when QC results indicate the need to
Work
ing C
opy
![Page 5: Calibration Copy Working - TIAERtiaer.tarleton.edu/transfer/Shepherd/QAM-RI-107rev... · QAM-RI-107 Operation and Calibration of the Canberra Alpha Analyst Spectrometer Rev.0 Page](https://reader031.fdocuments.us/reader031/viewer/2022022512/5ae81c347f8b9a9e5d900ecd/html5/thumbnails/5.jpg)
QAM-RI-107 Operation and Calibration of the Canberra Alpha Analyst Spectrometer
Rev.0 Page 5 of 12 TIEAR
reestablish operating parameters for each chamber of the instrument. The calibration is performed on each shelf of each detector chamber that is used. These geometries are recorded in the Alpha Spec Log (attachment 1). The energy range of the spectrometry system is set to at least include the region between 3 and 8 MeV. Before initial calibration, the analyst performs a pulser check and detector background check to ensure the instrument is operating normally.
2. Use the NIST traceable source, Echert & Ziegler 7400-SRC to calibrate.
3. Click “Sample Assigner” tab. Select Calibration and choose combined calibration. Drag source 7400-SRC into the assigned chamber. Choose the appropriate geometry shelf. Place the standard source in the chamber and double check to ensure the computer file assignment matches the geometry. Start counting. The count time is 60 minutes. Calibration can also be performed from “Calibration” table.
4. Upon completing the calibration a calibration report is printed out and stored electronically. For geometry at shelf 2 the centroid error for energy calibration should be less than 2 (6 KeV) channel, the FWHM should be less than 33 channels (100 KeV), and the efficiency should be between 17% and 25%.
E. Continuing calibration check
1. After calibration the instrument is routinely checked before and after each batch using standard sources to ensure the instrument operating conditions are (1) acceptable for analysis of sample test sources and (2) equivalent to those to be established during calibration.
2. The continuing calibration check results are compared to control charts and tolerance limits. Data out of control limits are investigated, documented and corrective action may be needed. The continuing calibration checks include calibration check and background subtraction count. The background subtraction count is performed daily. Before continuing instrument quality check, perform pulser check and detector background check to ensure the instrument is operating normally.
Work
ing C
opy
![Page 6: Calibration Copy Working - TIAERtiaer.tarleton.edu/transfer/Shepherd/QAM-RI-107rev... · QAM-RI-107 Operation and Calibration of the Canberra Alpha Analyst Spectrometer Rev.0 Page](https://reader031.fdocuments.us/reader031/viewer/2022022512/5ae81c347f8b9a9e5d900ecd/html5/thumbnails/6.jpg)
QAM-RI-107 Operation and Calibration of the Canberra Alpha Analyst Spectrometer
Rev.0 Page 6 of 12 TIEAR
3. Calibration check
a. Click “Sample Assigner” tab, select detectors to be checked for calibration.
b. Click on the “Calibration Check” button. Drag the 7400-SRC source to the appropriate chamber and geometry described on the screen.
c. Click on the “Load Samples” button. A screen with the list of detectors will appear. Place the standard source in the corresponding chamber and shelf. Click on “OK”. The calibration check count will start for the loaded detectors.
d. After the calibration check has finished counting, click on the “Quality Assurance View” tab and click on "Charts and Reports". Select chamber and geometry, select “Calibration Check”, and then select the report of “Last Measurement” to see the last check, or “Control Chart” to see the chart graph.
e. Data that are out of warning limits are investigated and the LM is notified. A CAR is generated when data are out of control limits.
f. Special attention is paid to instrument efficiency data
(CPM, the element activity) which directly influences the analytical data accuracy. Any corrective actions are documented with a CAR.
g. Print out and electronically store the check results and file.
F. Background subtraction count 1. The background subtraction count, or the system
background count, is used for sample analysis data correction.
2. Use a blank planchet of the same type that is used for making samples or test sources for counting. The geometry of a background subtraction count is the same as that used for counting samples or sources.
3. Before starting, ensure the system background procedure is set to the "ROI directed" count mode and the ROI should be the same as that of the target analytes.
4. Click “Sample Assigner” tab, select detectors to be checked for background subtraction count.
Work
ing C
opy
![Page 7: Calibration Copy Working - TIAERtiaer.tarleton.edu/transfer/Shepherd/QAM-RI-107rev... · QAM-RI-107 Operation and Calibration of the Canberra Alpha Analyst Spectrometer Rev.0 Page](https://reader031.fdocuments.us/reader031/viewer/2022022512/5ae81c347f8b9a9e5d900ecd/html5/thumbnails/7.jpg)
QAM-RI-107 Operation and Calibration of the Canberra Alpha Analyst Spectrometer
Rev.0 Page 7 of 12 TIEAR
5. Click on the “System Background Count” button. 6. Click on the “Load Samples” button. Choose the
appropriate geometry (i.e. “Shelf 2”, “Detector 1”), Place the blank planchet in the chamber at the required shelf.
7. A screen with the list of detectors to be counted will appear. Click on “OK”. The system background check count will start. The count time is 1000 minutes.
8. After the background subtraction count has finished counting, click on the “Quality Assurance View” tab and click on "Charts and Reports". Select chamber and geometry. Select “System background count”, and then
select the report of “Last Measurement” to see the last background subtraction count results, or “Control Chart” to see the chart graph. If the data are out of control limits, investigate the reason, take correction action and document and/or notify the LM.
9. Print out the check results and store electronically. The data are
used for sample count data correction.
G. Creating an analytical batch
1. Click on the “Batch” tab, enter Batch ID. The format of batch
ID is month-date-year-number. For example, the batch ID of
the first batch created on January 14, 2017 should be Jan-14-
17-001. Input batch description, select sample matrix, input
sample number (unknowns), select QC samples included,
select procedure (for Uranium, select Uranium 60 or Uranium
120; for Americium, select Americium 60 or Americium 120,
etc.), input sample amount, tracer amount, and QC spike
amount etc. An existing Batch may be copied, then assigned a
new ID for the batch and edited.
2. Click “Next”, to next page. Input information of each sample
including QC samples to create sample list in the Batch.
When inputting duplicate and spike samples, right click the
mouse to assign the sample to be spiked or duplicated. After
input each sample, click “Update Sample” to save it. After all
samples are inputted, click “Save” to save the Batch.
H. Sample analysis
Work
ing C
opy
![Page 8: Calibration Copy Working - TIAERtiaer.tarleton.edu/transfer/Shepherd/QAM-RI-107rev... · QAM-RI-107 Operation and Calibration of the Canberra Alpha Analyst Spectrometer Rev.0 Page](https://reader031.fdocuments.us/reader031/viewer/2022022512/5ae81c347f8b9a9e5d900ecd/html5/thumbnails/8.jpg)
QAM-RI-107 Operation and Calibration of the Canberra Alpha Analyst Spectrometer
Rev.0 Page 8 of 12 TIEAR
1. Load samples (or background planchets) for counting in
calibrated detectors.
2. Click on the “Sample Assigner” tab, click “Queued Batch”, and
select the Batch to run. Assign each sample in the batch to a
chamber.
3. Click “Load Sample”, vent chambers, and load samples into
the corresponding chambers with calibrated geometries.
4. Click “OK”, a sample run log table will show up. Print the
sample run log table for filing. Carefully review the sample run
log to ensure each sample is placed in correct chamber and
geometry. Press “OK” to start counting.
5. Double check the sample ID and the chamber ID to ensure
they agree with the run log table.
I. Data review
1. Click on the “Data Review” tab, Click “Batch ID” on the search
type box. Type the batch number for samples to be reviewed,
select procedure used by the batch, click "search now" to get
the batch. You can use procedure to find batch ID from
procedure too. To do so, select procedure, assign the date
filter (date start and date end), click "search now" to list the
batches associated with this procedure, select the one you
want.
2. Check the batch number in the search results window to list
the samples to be reviewed, click the “Next” button located on
the bottom right of the data review page. A spectrum of the
highlighted data file (on the up-right box) is shown.
3. Review the regions of interest (ROI) to determine if they are
within specification.
4. The ROIs associated for each isotope may be adjusted at one
time using the “Shift All ROIs” slide bar or the left marker may
be adjusted using the “Shift only Left Markers” slide bar. If the
ROIs are acceptable, click on the “Update ROI”, and
“Reanalyze”.
Work
ing C
opy
![Page 9: Calibration Copy Working - TIAERtiaer.tarleton.edu/transfer/Shepherd/QAM-RI-107rev... · QAM-RI-107 Operation and Calibration of the Canberra Alpha Analyst Spectrometer Rev.0 Page](https://reader031.fdocuments.us/reader031/viewer/2022022512/5ae81c347f8b9a9e5d900ecd/html5/thumbnails/9.jpg)
QAM-RI-107 Operation and Calibration of the Canberra Alpha Analyst Spectrometer
Rev.0 Page 9 of 12 TIEAR
5. If the ROIs created are not acceptable, they may be adjusted
manually in the review box. If this approach is taken, each ROI
that requires adjustment must first be deleted by clicking on the
“delete ROI” button. Then the new ROI is set by adjusting the
left and right markers and clicking on the “Add ROI” button.
The ROIs are updated by clicking on the “Update ROI” button.
Once the ROIs have been updated, click on the “Reanalyze”
button.
6. From "View" click on the "report" button to review the report on
screen. Check the chemical recovery factor. It should be >
75% and <125%. If the data are acceptable, add any
comments on this sample by click on "Enter comments" button
when needed, then click on the “Approved” button. The
following important items are reported in the report: sample
and count information, chemical recovery factor, sample
activities of tracer and analytes and their relative errors
(uncertainty), MDCs and their errors of analytes, FWHM of
each peak. The percent recoveries of lab control spike and
matrix spike are also included in the reports of QC samples.
7. Click on the “Print Report” button. A print option screen will
appear. Select "Print Report" option, click "OK". A hard copy
of the report will be printed. Save a copy electronically in the
assigned folder.
8. Repeat the above steps to review the next sample. Repeat
this process until the review of all the samples in the batch is
completed.
9. Review the QC results: click on the “Quality Assurance View”
tab and select "chemistry". Select the type of QC (spike,
duplicate, unknown etc.), then select sample matrix and
element. Select the report of the “Last Measurement” to see
the "RPD" (relative percent difference) of last duplicate, or
"RPR" (relative percent recovery) of last spike, or chemical
recovery of last unknowns. Select "control chart" to see the
QC data chart. If the data are out of control limits, investigate
Work
ing C
opy
![Page 10: Calibration Copy Working - TIAERtiaer.tarleton.edu/transfer/Shepherd/QAM-RI-107rev... · QAM-RI-107 Operation and Calibration of the Canberra Alpha Analyst Spectrometer Rev.0 Page](https://reader031.fdocuments.us/reader031/viewer/2022022512/5ae81c347f8b9a9e5d900ecd/html5/thumbnails/10.jpg)
QAM-RI-107 Operation and Calibration of the Canberra Alpha Analyst Spectrometer
Rev.0 Page 10 of 12 TIEAR
the reason, take correction action and document with a CAR.
To see total QC results and control chart in a period of time,
select "Full Report" and define the start and the end dates.
10. If no QC information was entered in the batch, e.g. the QC
samples are treated as normal samples, the QC results may
be manually calculated.
5.0 Quality Control and Safety Aspects
A. All aspects of this procedure comply with QAM-Q-101,
“Laboratory Quality Control”, QAM-S-101, “Laboratory Safety”,
and QAM-R-100, “TIAER Laboratory Radiochemistry Program”.
Only personnel properly trained and authorized by the LM and
RSO are allowed to perform this procedure. Area swipes for
loose surface contamination are performed per S-101 and R-100.
B. All data are documented and maintained in accordance with,
QAM-A-102, “Document and Data Control”. All raw data are
recorded in the appropriate logbook or E-log.
C. Waste is handled and disposed of in accordance with QAM-W-
101, “Disposal of Laboratory Waste”. All waste from this
procedure should initially be considered radioactive.
D. Any sample or standard test source is removed from detector
chamber after counting to prevent contamination.
E. Any part in the detector chamber, such as test source supporter
and shelf regulator, should not be left outside the chamber.
F. To prevent damaging or contaminating detectors, the door of
each detector is kept closed at all times except when loading or
unloading the test sources or samples.
G. Current activities for source checks are found in the Calibration
Source Log, Q-102-4. Document source ID and source dates.
H. Ensure that the instrument has been calibrated prior to use.
I. Always take into account background radiation levels and record
the most recent with all data. Measure and record background
radiation levels in the appropriate logbook or E-log. Background
values must be within +/- 20% of the initial background average
Work
ing C
opy
![Page 11: Calibration Copy Working - TIAERtiaer.tarleton.edu/transfer/Shepherd/QAM-RI-107rev... · QAM-RI-107 Operation and Calibration of the Canberra Alpha Analyst Spectrometer Rev.0 Page](https://reader031.fdocuments.us/reader031/viewer/2022022512/5ae81c347f8b9a9e5d900ecd/html5/thumbnails/11.jpg)
QAM-RI-107 Operation and Calibration of the Canberra Alpha Analyst Spectrometer
Rev.0 Page 11 of 12 TIEAR
used for calibration. Measurements must fall within +/- 20% of the
initial average used for Quality Control.
J. During operation internal components contain live voltage. Turn
the instrument OFF, unplug coax cable and allow it to sit for one
minute before accessing internal components.
K. The connector contains voltage exceeding 1kV. Contacting it
during operation may result in electric shock.
L. Follow all safety precautions outlined in QAM-S-101 when
working with radioactive samples.
M. System background checks for each tray and geometry used are
performed at least weekly and recorded in the log.
N. Perform alpha and beta source Calibration Source check
measurements daily using a consistent location and geometry.
O. Standard test sources and sample test sources are radioactive
and must be handled with care. After counting the test sources
must be placed in their containers.
P. Safety glasses, gloves, and lab coats are recommended personal
protective equipment when handling samples.
Q. Staff members wear radiation dosimeters and a Geiger counter is
available to monitor the workplace.
R. Annual HazCom and safety training is performed in accordance
S-101 and R-100.
6.0 References
A. User's manual of Canberra Apex-Alpha Spectroscopy Software Suite.
B. Operation manual of Canberra Genie 2000.
C. ASTM D7282 “Standard Practice for Set-up, Calibration, and Quality Control of Instruments Used for Radioactivity Measurements,” ASTM Book of Standards 11.02, current version, ASTM International, West Conshohocken, PA
7.0 Attachments
A. Alpha Spectrometer Log, RI-107-1 (example)
Work
ing C
opy
![Page 12: Calibration Copy Working - TIAERtiaer.tarleton.edu/transfer/Shepherd/QAM-RI-107rev... · QAM-RI-107 Operation and Calibration of the Canberra Alpha Analyst Spectrometer Rev.0 Page](https://reader031.fdocuments.us/reader031/viewer/2022022512/5ae81c347f8b9a9e5d900ecd/html5/thumbnails/12.jpg)
QAM-RI-107 Operation and Calibration of the Canberra Alpha Analyst Spectrometer
Rev.0 Page 12 of 12 TIEAR
Attachment A
Alpha Spectrometer Log (example)
Instrument ID
Analysis Date
Analyst Source ID Background Date
Background Results
Tracer ID Prep Date Batch ID Comments
Attach printed copies of QC Analytical Batches and QC results
Work
ing C
opy