BMC Medical Co., Ltd.

30
BMC Medical Co., Ltd. Nov. 2011 Training Course

description

BMC Medical Co., Ltd. Training Course. Nov. 2011. Table of Contents CATALOGUE INSTALL& DISASSEMBLE SET UP. 1. CATAOLGUE. Resmart CPAP. 2. INSTALL & DISASSEMBLE. Install Water tank. Component : Water tank, Humidifier, Device. Clearing Water tank. - PowerPoint PPT Presentation

Transcript of BMC Medical Co., Ltd.

Page 1: BMC Medical Co., Ltd.

BMC Medical Co., Ltd.

Nov. 2011

Training Course

Page 2: BMC Medical Co., Ltd.

2

Table of Contents

1.CATALOGUE

2.INSTALL& DISASSEMBLE

3.SET UP

Page 3: BMC Medical Co., Ltd.

3

1. CATAOLGUE

Page 4: BMC Medical Co., Ltd.

4

Resmart CPAP

Page 5: BMC Medical Co., Ltd.

5

2. INSTALL & DISASSEMBLE

Page 6: BMC Medical Co., Ltd.

Install Water tank

Heater Plate

Indicator Light

Outlet Port

Water Chamber

Fill Line

Page 7: BMC Medical Co., Ltd.

Component : Water tank, Humidifier, Device.

1

2

3

Page 8: BMC Medical Co., Ltd.

Disconnect the tubing from the water chamber. Press down

on the water chamber and slide it out of the humidifier platform.

Open the silica gel cover of water chamber and discard any remaining water.

Fill a solution of warm water and a mild dishwashing detergent into the chamber, cover the silica gel cover, rock the chamber a few minutes, then outflow the solution. Rinse the chamber several times with clean water and allow to air dry.

Fill the water chamber and close the silica gel cover. Inspect the water chamber for any leaks or damage.

Replace the water chamber if any damage is present.

Clearing Water tank

Page 9: BMC Medical Co., Ltd.

DISASSEMBLE

Open the cover of the filter carefully. And then take it out.

Page 10: BMC Medical Co., Ltd.

DISASSEMBLE

Press the button at the bottom of the main machine, then pull the heated humidifier.

Page 11: BMC Medical Co., Ltd.

SET UP

Control Button

Page 12: BMC Medical Co., Ltd.

CPAP

CPAP USER SET UP

ENTER USER MENU

Page 13: BMC Medical Co., Ltd.

CPAP SET UP

Verify/Change the RESmart USER settings. To start the setup menu, press and hold the Ramp button until the setup menu appear (about 5 seconds) when system standby.

IMPORTANT! Press the Ramp Button to enter change status and press +/- User Buttons to change the setting. Press Ramp Button again to confirm the setting.

Press +/- User Buttons to access the next setting.

Page 14: BMC Medical Co., Ltd.

CPAP

Setting the Humidifier When the optional InH2™ Heated Humidifier is used, this setting can change the temperature and humidity of airflow between level 0 to 5. Default setting is 3.

Page 15: BMC Medical Co., Ltd.

CPAP

Setting the Altitude

The altitude setting can be changed between level 0 to 2. Default setting is 0.

0 = less than 2,460 ft. (<750 m)

1 = 2,460 to 4,921 ft. (750 m to 1500 m)

2 = 4,924 to 8,202 ft. (1501 m to 2500 m)

*over 8,202 ft. = The airflow pressure may not be accurate. Contact your home care provider to have your pressure setting adjusted.

Page 16: BMC Medical Co., Ltd.

CPAP

Setting Auto ON

When the RESmart CPAP is standby and

wearing mask, your deep breath will start

airflow automatically. Default setting is Enable.

Page 17: BMC Medical Co., Ltd.

CPAP

Setting Auto Off

When the mask is removed, the RESmart CPAP will stop airflow automatically. Default setting is Disable.

Note: We advice you can set Auto Off Disable. Most machine may appear this kind of problem, because apnea and mask off  is not to be judged at pressure lower than 7 hpa

Page 18: BMC Medical Co., Ltd.

CPAP

Setting the LCD Light

This setting allows you to adjust the LCD light among Auto, On and Off. Default setting is Auto.

Auto: LCD light will be turned on when any button pressing and turned off after few seconds.

On: LCD light will be turned ON always.

Off: LCD light will be turned OFF always.

Page 19: BMC Medical Co., Ltd.

CPAP

Setting the Alert

This setting allows you to turn on/off the patient disconnect alert. When a large, continuous air leak (such as mask removal) has been detected in the circuit, this setting enables/disables the audible alert (a beeping sound). This setting is only effective when the Auto Off setting is disabled. (This function is NOT available on certain model.) Default setting is Disable.

Page 20: BMC Medical Co., Ltd.

CPAP

Setting the Delay Off

When the optional InH2™ Heated Humidifier is used, this setting allows the air flow continue for about 15 minutes at very low pressure (about 2hPa) after RESmart CPAP is turned off. This will blow off the left vapor and protect the RESmart CPAP. Default setting is Enable.

Page 21: BMC Medical Co., Ltd.

CPAP

Use Days

The number of nights the RESmart CPAP was used for therapy for more than 4 hours will appear. This screen is only for reference. Your home care provider may periodically ask you for this information.

Page 22: BMC Medical Co., Ltd.

CPAP

Exit the Settings

The settings are complete. Press the +/- user button to access the SAVE setting and Ramp Button to save all changes and exit the settings menu.

Page 23: BMC Medical Co., Ltd.

CPAP

CPAP MAINTENANCE SET UP

ENTER MAINTENANCE MENU

Page 24: BMC Medical Co., Ltd.

CPAP

Entering the Menu: Press and hold the Pressure Start/Stop Button and Ramp Button while connecting the power cord into the AC inlet. After several seconds the Maintenance Menu will appear on display screen as below.

IMPORTANT! Press the Ramp Button to enter change status and press +/- User Buttons to change the setting. Press Ramp Button again to confirm the setting.

Press +/- User Buttons to access the next setting as below.

Page 25: BMC Medical Co., Ltd.

CPAP

Ramp Starting Pressure: The starting pressure after blower is turned on. Setting range is 4hPa to Treatment Pressure. Adjustment step is 0.5cm H2O. Default setting is 4cm H2O.

Init P 4.0

Ramp Starting Pressure

Page 26: BMC Medical Co., Ltd.

CPAP

Treatment Pressure: The prescription pressure after ramp time. Setting range is 4-20cm H2O. Adjustment step is 0.5cm H2O. Default setting is 8cm H2O.

Page 27: BMC Medical Co., Ltd.

CPAP

Ramp Timer: The ramp time from starting pressure to prescription pressure. Setting range is 0-60mins. Adjustment step is 5mins. Default setting is 15mins.

Page 28: BMC Medical Co., Ltd.

CPAP

Manual Pressure: This setting allows user to adjust Treatment Pressure by oneself. Setting range is 0-2cm H2O. Adjustment step is 0.5cm H2O. Default setting is 0.

For example: if this setting is 1cm H2O and Treatment Pressure setting is 8cm H2O, user may change the treatment pressure between 7-9cm H2O by pressing +/- User Button.

Page 29: BMC Medical Co., Ltd.

CPAP

Use Time: Total hours this RESmart has been used will appear. This number is only for reference.

Page 30: BMC Medical Co., Ltd.

CPAP

Save and Exit: The settings are complete. Press the +/- user button to access the SAVE setting and Ramp Button to save all changes and exit the Maintenance Menu.

IMPORTANT! To make all changes active, press the +/- User Button to access the SAVE setting and Ramp Button to save all changes and exit the Maintenance Menu.