BITC 1402 General, Organic and Biochemistry or GOC, Chapters 3, 4, 5 Review of General Chem 1...
-
Upload
vernon-stone -
Category
Documents
-
view
216 -
download
3
Transcript of BITC 1402 General, Organic and Biochemistry or GOC, Chapters 3, 4, 5 Review of General Chem 1...
BITC 1402
General, Organic and Biochemistry or GOC, Chapters 3, 4, 5
Review of General Chem 1
Laboratory Methods or LM, 24, 14, 15
Chapter 3What is the octet rule?Define electronegativityCompare and contrast ionic and covalent bonding, which is stronger etc.Compare and contrast nonpolar and polar covalent bondingWhat is a Lewis structure and what does it show for a molecule?Compare and contrast Lewis structure and VSEPR modelsWhat is a dipole?How are binary molecular compounds named?
Chapter 4Be able to calculate formula, molecular weightsBe able to write chemical equations including a net ionic equationBe able to determine limiting reagent, percent yieldBe able to determine spectator ionsBe able to determine what is oxidized and what is reduced in a redox rxnDefine heat of reaction, exothermic and endothermic, and heat of combustion
Chapter 5
Define intermolecular forces of attraction and be able to list them in order of increasing strength
Define surface tension, vapor pressure, boiling point, melting point, and crystallization
Define phase, phase change, sublimation, heat of fusion, heat of vaporization, triple point, and a phase diagram
Chapter 7pg 201
Le Chatelier’s Principle:If an external stress is applied to a system in equilibrium, the system reacts in such a way
as to partially relieve that stress
Activities Chapters 3, 4, and 5
Ch3: 53, 58, 69, 72, 77, 102, 105
Ch4: 29, 41, 46, 50, 53, 73, 90, 92
Ch5: 51, 56, 57, 88
LM: Chapter 24
Contaminants of water (table 24.1)Dissolved inorganicsDissolved organicsSuspended particlesDissolved gasesMicroorganismsPyrogens/endotoxins Measure total organic carbon (TOC) Measure total solids (TS)
Types of Water (Table 24.3)
I (highest purity) used for analytical procedures, tissue culture, some HPLC, electrophoresis buffers, immunology assays
II suitable for microbiology procedures
III used for initial glassware rinses
Biologically pure- tissue culture
Organically pure- HPLC, GC/MS
WFI- water for injection
Methods of Water PurificationReview table 24.4What method(s) will remove pyrogens, dissolved ionized gases?DistillationIon ExchangeDeionization (remove all ionic contaminants)Carbon Adsorption –remove dissolved organic compoundsFiltration Depth Microfiltration Ultrafiltration Reverse osmosis
Other Methods
UV oxidation – removes organic contaminants and may kill bacteria
Ozone – kills bacteria
Systems in the Lab
RO
Deionization distillation
What type(s) of water are possible and how do we determine what type(s) of water are being produced?
Handling of Reagent Water
Highly purified water is an extremely aggressive solventReadily leach contaminants from any
vessel and will also dissolve carbon dioxide from the air (Is this why the pH value continually changed?)
Type 1 cannot be stored for any length of time –Type II can be stored for short periods of time
Monitoring
Resistivity – The theoretical max ionic purity for water is 18.3 megohm-cm and 17 is acceptable
Bacterial counts
Pyrogens
Organic carbon (potassium permanganate)
Maintenance
Distilled water: frequent cleaning prepared daily Ion exchange/ deionization: regenerated and sanitizedFilters: tested, flushed and sanitizedActivated carbon: washed/ rechargedUV: replaced annuallyDOCUMENTATION
Characteristics & Cleaning of Glass and Plastic Labware
Review tables 24.5 and 24.6Cleaning Prerinse Contaminant removal (review table 24.7) Rinse Final rinse Drying
(I used chromic acid to clean pipettes in graduate school—why is this a problematic process in today’s lab?)
Sterilization and Storage of Solutions
Review Tables 24.9 and 24.10
Activity: problem 1 and 3
Chapter 14: Introduction to Instrumental Methods and Electricity
Mechanical measurement instrument has these components: Interface Transducer (Sensor) Signal Processor Display (Readout)
Identify these components in a balance
Certain Requirements Must be Met:
The instrument’s response must have a consistent and predictable relationship with the property being measured
The instrument’s response must be related to internationally accepted units of measurement
Calibration is the process by which the response of an instrument is related to internationally accepted measurement units
Calibration of Common Bioinstrumentation
pH meter
Balance
Centrifuge
(How is calibration different from validation?)
Basic Terminology and Concepts of Electricity
Define the following:CurrentAmperes or amps (6.25 x 1018
electrons/sec = 1amp)DC (delivered by batteries)AC (USA- frequency is 60 times/sec) 1
Hz=1cycle/sec (why is AC delivered by power companies and NOT DC?)
Rectification- change AC to DC
Voltage
What is the definition of energy? Potential energy?
Electrical potential is also called electromotive force (EMF) or voltage (V).
The voltage supplied by the power company is either in the range of 110 to 120 or 220.
Resistance
Impedance to electron flow
The units are ohms (1 ohm is the value of resistance through which 1 V maintains a current of 1 A)
What is a conductor? What is a semiconductor? An insulator?
Ohm’s LawV=IRApplication of Ohm’s Law Gel electrophoresis: The sample mixture is placed
in a gel matrix. The gel is placed in a box, buffer is poured over it, and a current (I) is applied. The positive and negative ions in the gel and the buffer conduct the current. A power supply provides the voltage (V), source of the I. The gel provides R, which increases with time as the ions are run out of the gel. Therefore, to maintain I, V must be increased BUT this also increases the heat in the gel-thus, it is better to maintain constant V instead of constant I TO prevent excessive heating of the gel.
Power, Work, and Circuits
Power = voltage x current or
W = (V)(I)
A good equation to know if you’re trying to figure out how many instruments you can run in a lab at one time
Equipment in the Laboratory
Power supplies convert AC to DCTransducers or detectors generate an electrical signal in response to a physical or chemical property of a sample (review table 14.1)Detection limit, sensitivity or range of the
equipment is the minimum level of the material or property of interest that causes a detectable signal
Electrical noise is an important factor in determining rangeNoise Short-term is defined as random, rapid “spikes” Long-term or drift is a relatively long-term increase
or decrease in readings due to changes in the instrument
Expressed as signal-to-noise ratio: the higher the ratio, the better the performance (root mean square noise- RMS)
The detection limit of a detector is therefore often defined in practice as the minimum level of sample that generates a signal at least twice the average noise level
The dynamic range of a detector is the range of sample concentrations that can be accurately measured by the detector
Signal Processing Units
Amplifier boosts the voltage or current from a detector in proportion to the size of the original signalGain is the degree to which a signal can be increased or decreasedFor example, if input is 1 mV and the output is 100 mV then Gain = 100/1=100
Attenuator reduces a signal in order to best display it by a readout deviceReadout devices Meters, strip chart recorders, computer screens
Quality and Safety Issues Preventive maintenance
Performance verification Instrument validation
Environmental Factors that Affect the Performance of Instruments (table 14.4)Electrical Safety (table 14.5)Problems: 1, 2, 8
BITC 1402The Measurement of Weight
Chapter 15
Chapter 15: Weight
The force of gravity on an objectBalances are instruments used to measure this forceMass The amount of matter in an object expressed in units
of grams Why are these instruments named balances? How is a mechanical balance different from an
electronic balance? Which one would you use to balance tubes for
centrifugation?
Characteristics and Types of Balances
Range (highest to lowest)Capacity (the heaviest sample that can be weighed)Sensitivity (or readability)- the smallest value of weight that will cause a change in the response of the balance (determines the no. of places right of the decimal point)Which is more sensitive, an analytical balance or a pan balance?(review figure 15.4)How about the balances in the lab?
How Does an Electronic Balance Work?
The weighing pan is depressed by a small amount when an object is placed on it.The balance has a detector that senses the depression of the panAn electromagnetic force is generated to restore the pan to its original (“null”) position. This force is measured as an electrical signal that is in turn converted to a digital display of weight value (i.e. the balance compares the electrical signal of the unknown sample to the signal of standard(s) of known weight)
General Procedure for Weighing a Sample with an Electronic Analytical
BalanceMake sure the balance is levelAdjust the balance to zero (the pan should be clean and doors shut to avoid air currents)Tare the weighing container or weigh the empty containerPlace the sample on the weighing pan & read the value for the measurementRemove the sample; clean the balance and area around it (I will fail you if you do not!)
Factors That Affect the Quality of Weight Measurements
Temperature (samples are heavier when they are cold)
“Warming up” period for the balance?
Static charge (ionizing blower, antistatic brush) Ground yourself (hold the sample on the spatula
and touch the balance)
Review table 15.1
Calibration and Maintenance of a Balance
Must be checked periodically
Must be checked when a balance is moved
For microbalances (5 places pass the decimal to the right), they must be calibrated when the weather changes!
Mechanical Balance
Must be calibrated by a trained technician (ASTM Standard E 319-85)
Electronic Balance: Two-point
Calibration The balance is set to zero using the appropriate knob, when the weighing pan is clean and emptyThe first weight is added and the balance is set to the value of the standardThe second weight (usually heavier than the first) is added and the balance is set to the value of the standardA quick check is to add exactly 1mL of water to a tared weigh boat-it should weigh exactly 1g (of course your micropipettor needs to be measuring exactly 1 mL)
StandardsPurchased with a certificate showing their traceability to NIST The appropriate weights for a balance have been established by ASTM Standard E 617There are different classes: 1, 2, 3 and 4 with 1 being the most rigorous (4 is recommended for student use)Handle with tongs because they are damaged by skin oils and by cleaners that remove such oils(Keep two sets: one for use and one for storage; every 6 months measure the one for use against the one for storageReview the Appendix
Quality Programs and Balances
Laboratories that meet the requirements of a quality system, such as ISO 9000 or cGMP, have procedures that detail how to operate each balance and how to maintain, calibrate, and check their performance.
Verifying Balance Accuracy, Precision, and Linearity
Accuracy is tested by weighing one or more mass standards
Precision is measured by weighing a sample multiple times and calculating the standard deviation
Linearity is tested by weighing subsets of weights and comparing the sum of the subsets to the weight of the objects all together
Checking the Linearity of a Balance
Select 4 items whose weight is about the capacity of the balance and label them A, B, C, DWeigh all 4 items separately and added up the valuesWeigh all 4 items together and compare that value to the summed valueDo this for a low weight (25%), the midpoint weight (50%) and for a weight close to capacity of the balance (75%)
Writing SOPs for the Use of Balances
Examples are found on page 292
Mass versus WeightThe major force measured in weighing is the force of gravity. However there is also a slight buoyant force from air. Therefore objects weighed in a vacuum are heavier than weighed in air. This is the priniciple of buoyancy. (Note 1 g of metal weighs differently than 1 g of water)The discrepancy between mass and weight is called the buoyancy error-usually this error is ignored!Exercises: 4, 5, 6