Avalon FM20/FM30 Fetal/Maternal Monitor - · PDF fileAvalon FM20/FM30 Fetal/Maternal Monitor...

20
Avalon FM20/FM30 Fetal/Maternal Monitor Philips M2702A/M2703A Technical Data Sheet - Release J.3 The Philips Avalon FM20/FM30 fetal monitor offers a reliable, cost-effective, high-performance solution for fetal monitoring needs in clinics and hospitals, doctors’ practices and private households. This ranges from prenatal check-ups to antepartum monitoring of high-risk pregnancies, and during labor and delivery. With a cutting-edge feature set, including advanced transducers, triplets monitoring, high-quality signal processing, and a touchscreen interface, the Avalon FM20/FM30 sets new standards for performance, flexibility, convenience, and ease of use. The unique Smart Pulse measurement integrated in the Toco transducer simplifies monitoring. Without an additional sensor, the maternal pulse rate is monitored and continuously compared with the heart rate from the fetal sensor (cross-channel verification). In combination with an Avalon CL Transducer System all fetal and maternal measurements can be monitored cablelessly, including twins, triplets, and Smart Pulse. The battery option gives monitoring flexibility for in-house transport or bridging a mains power outage.

Transcript of Avalon FM20/FM30 Fetal/Maternal Monitor - · PDF fileAvalon FM20/FM30 Fetal/Maternal Monitor...

Page 1: Avalon FM20/FM30 Fetal/Maternal Monitor - · PDF fileAvalon FM20/FM30 Fetal/Maternal Monitor ... The dual MIB, Medical Information ... • Connection of the external device to the

Avalon FM20/FM30 Fetal/Maternal Monitor

Philips M2702A/M2703A Technical Data Sheet - Release J.3

ted

e

ith

nel

be

nd

for

The print quality of

The Philips Avalon FM20/FM30 fetal monitor offers a

reliable, cost-effective, high-performance solution

for fetal monitoring needs in clinics and hospitals,

doctors’ practices and private households. This

ranges from prenatal check-ups to antepartum

monitoring of high-risk pregnancies, and during labor

and delivery.

With a cutting-edge feature set, including advanced

transducers, triplets monitoring, high-quality signal

processing, and a touchscreen interface, the

Avalon FM20/FM30 sets new standards for

performance, flexibility, convenience, and ease of

use.

The unique Smart Pulse measurement integra

in the Toco transducer simplifies monitoring.

Without an additional sensor, the maternal puls

rate is monitored and continuously compared w

the heart rate from the fetal sensor (cross-chan

verification).

In combination with an Avalon CL Transducer

System all fetal and maternal measurements can

monitored cablelessly, including twins, triplets, a

Smart Pulse.

The battery option gives monitoring flexibility

in-house transport or bridging a mains power

outage.

this copy is not an accurate representation of the original.

Page 2: Avalon FM20/FM30 Fetal/Maternal Monitor - · PDF fileAvalon FM20/FM30 Fetal/Maternal Monitor ... The dual MIB, Medical Information ... • Connection of the external device to the

trols

ic

the

tion,

tting

n of

ers,

y of

ith

least

er

and

ery

ith

M30

ours

a

ew

Features

Transducers and Measurements• Smart transducers are comfortable to wear, waterproof, and easy to

clean

• Ultrasound (US) fetal heart rate monitoring of singletons, twins, and

even triplets

• Fetal Movement Profile (FMP) automatically detects fetal body

movements for antepartum applications using an ultrasound

transducer

• External uterine activity (Toco) monitoring

• Smart Pulse integrated in the Toco MP transducer detects the

maternal pulse rate without the need for an additional sensor (SpO2

or MECG), confirming correct fetal heart rate monitoring

• Cross-Channel Verification (CCV) detects the coincidence

between all fetal and maternal heart and pulse rates

• NST timer for antepartum applications

• Optional Non-Stress Test (NST) report software allows

interpretation of fetal heart rate and Toco traces. It generates a

printed report automatically for an NST, using NICHD guidelines

from 2008 for electronic fetal heart rate monitoring

• Optional maternal noninvasive blood pressure (NBP) including

pulse rate

• Heart rate with maternal ECG (MECG)

• Cableless monitoring with an Avalon CL or Avalon CTS Fetal

Transducer System

• Optional monitoring of maternal temperature with the Tympanic

Temperature Thermometer

• Entering of manually taken maternal temperature

• Direct fetal heart rate monitoring with fetal ECG (DECG) including

waveform display (FM30 only)

• Intrauterine pressure monitoring (IUP) (FM30 only)

• Optional maternal pulse oximetry (SpO2) and continuous maternal

pulse rate (FM30 only)

Monitor• High-resolution color TFT touchscreen display with a wide

viewing angle. Fetal and maternal measurements are displayed as

large, easy-to-read numerics

• Intuitive user functionality via the touchscreen

• Quick and simple cleaning due to the absence of mechanical con

• Optimized size and presentation of measurements with automat

screen layouts

• Displaying information such as: Patient identification (including

bed label when connected to a surveillance system such as

OB TraceVue/IntelliSpace Perinatal), date and time, alarm informa

recorder status, and prompt and error messages

• Fast access to functions, for example starting a recording or rese

the Toco baseline via Configurable SmartKeys

• Adjustment of the heart rate and alarm volume with an indicatio

the current setting

• With the automatic channel management all wired fetal transduc

patient module, or remote event marker can be connected to an

the fetal sensor sockets at the left side of the monitor

• Identification of the relation between transducer and numeric

through a blue transducer finder LED, while the position

indication on the screen shows to which fetal sensor socket a

transducer is connected

• Comprehensive patient data presentation, including at-a-glance

patient identification, and the choice to add user-configurable

notes

• Optional battery operation based on a built-in Smart Battery

(M4605A) and on an external AC/DC power supply (M8023A w

option E25); the monitor can operate using battery power for at

four hours

Recorder• Built-in six-inch high-resolution thermal array recorder with pap

type detection and paper-out warning

• Trace output of fetal and maternal heart rates, fetal movements,

uterine activity, and annotation of time, date, trace identification

symbols, patient information, and software versions

• Maternal heart and pulse rate are printed every five minutes

• SpO2 saturation is printed every five minutes (configurable to ev

1 minute) (FM30 only)

• Maternal noninvasive blood pressure and temperature values are

annotated at the end of the measurement

• Recording of six-second MECG and/or DECG wave snapshots w

25 mm/s scaling in parallel to, or separate from the fetal traces (F

only)

• Internal backup memory for at least 7 hours of trace data (>3.5 h

with previous hardware revision <A 00.18):

– Stored data recording to print a specific patient episode from

selected patient at high speed, even multiple times

– Bridge paper-out for printing the missing piece of trace after n

paper has been loaded

2

The print quality of this copy is not an accurate representation of the original.

Page 3: Avalon FM20/FM30 Fetal/Maternal Monitor - · PDF fileAvalon FM20/FM30 Fetal/Maternal Monitor ... The dual MIB, Medical Information ... • Connection of the external device to the

for to r to

itor,

de

ed

ing

efer

m

– Data recovery after an interruption of the online data transmission

to an OB TraceVue/IntelliSpace Perinatal system over a LAN

connection.

• Paper Save Mode for maternal measurements

AlarmsWhen an alarm condition is fulfilled, it is signaled by the monitor in the

following ways:

• An alarm tone sounds

• An alarm message is shown on the screen

• The numeric of the alarming measurement flashes on the screen

Alarm categories• Physiological Alarms for fetal and maternal parameters (to

indicate vital sign alarm limit violations, for example)

• Technical Alarms (INOPs) are triggered by signal quality

problems, equipment malfunction, or equipment disconnection

Alarm modes• INOP only: INOPs are enabled, with audible and visual indications

active (default setting)

• All: patient alarms and INOPs are enabled, with audible and visual

indications active

Alarm suspension• Alarms Off/Pause Alarms function

• Alarm pause period: indefinite, or one, two, or three minutes

Alarm latching/non-latching• Visual

• Audible

• Visual and audible

Standard Interfaces

Cableless monitoring using the Avalon CL Transducer System, or the

Avalon CTS Cordless Transducer System with connection at the fetal

sensor socket

Optional Interfaces

LAN/RS-232 System InterfaceThe LAN/RS-232 interface has two fully isolated ports:

1. The LAN port is for connecting the monitor to an OB TraceVue/

IntelliSpace Perinatal obstetrical information and surveillance system

(revision E.0 or later) on a network. OB TraceVue/IntelliSpace

Perinatal controls all patient admit/discharge events.

2. The RS232 port can be used for connecting the monitor to an

obstetrical information and surveillance system, such as

OB TraceVue/IntelliSpace Perinatal.

Dual system interface support: If an obstetrical information and

surveillance system is connected via LAN, another information system

(EMR) can be independently connected via RS232 in read-only mode.

Dual MIB/RS-232 Interface

The dual MIB, Medical Information Bus (IEEE P1073), is a standardinterfacing medical devices, used for the Data Export functionalityexport measurement values from the monitor to other devices, oconnect the Tympanic Temperature Thermometer

Flexible Nurse Call Interface• Means for alarms generated on the monitor to be signaled on an

external device such as a nurse call system, a beeper, or a light

• Providing three general alarm relays

• Connection of the external device to the alarm relay of the mon

and configurable alarm criteria

• Active open and closed contacts and a user-definable delay time

Device Interface (USB Interfaces)• Allowing connection of USB devices (mouse, keyboard, or bar co

reader) to the monitor

Service Features

• The Configuration Mode is password-protected and allows train

users to customize the monitor configuration

• The Demo Mode is password-protected and is intended for train

and educational purposes

• The Service Mode is password-protected and ensures that only

trained staff can access service tests and tasks

• The Support Tool uses the LAN interface of the fetal monitor

• The Support Tool helps technical personnel to

– carry out configuration, upgrades, and troubleshooting on an

individual monitor

– share configuration settings between monitors

– back up the monitor settings

Related Products

Avalon CL and Avalon CTS Transducer SystemThe monitor is compatible with the Avalon CL and CTS, allowing

expectant mothers to be continuously monitored without wires,

whether in the bath or shower, in the bed, or while ambulating. R

to the Avalon CL and the Avalon CTS1 Data Sheets for further

information.

1 The Avalon CTS Transducer System is no longer for sale.

Avalon CL Transducer System Avalon CTS Transducer Syste

3

The print quality of this copy is not an accurate representation of the original.

Page 4: Avalon FM20/FM30 Fetal/Maternal Monitor - · PDF fileAvalon FM20/FM30 Fetal/Maternal Monitor ... The dual MIB, Medical Information ... • Connection of the external device to the

cept

est

f it is

°F)

)

Tympanic Temperature Thermometer (866149)The monitor is compatible with the Tympanic Temperature

Thermometer allowing the measurement of maternal temperature.

The temperature is taken in the ear. The value of this measurement can

be automatically adjusted to correspond to a different body reference

site. The measured value is shown on the display of the thermometer

and transmitted to the monitor. Refer to the Tympanic Temperature

Thermometer Data Sheet for further information.

Regulatory and Standards Compliance

The monitor complies with the requirements of the European Medical

Devices Directive 93/42/EEC and the following major international

standards:

Safety Specifications• IEC 60601-1:2005+A1:2012 / EN 60601-1:2006+AC:2010

• IEC 60601-1-6:2010 / EN 60601-1-6:2010

• IEC 60601-1-8:2006+A1:2012 / EN 60601-1-8:2007+AC:2010

• IEC 60601-2-49:2011 / EN 60601-2-49:2001

• ANSI/AAMI ES60601-1:2005+A1:2012+C1:2009+A2:2010

• CAN/CSA C22.2#60601-1-08

• JIS T 1303 2005

• AS/NZS 3200.1.0-1998

The possibility of hazards arising from hardware and software errors

was minimized in compliance with ISO 14971:2012,

IEC 60601-1:2005+A1:2012/EN 60601-1:2006+AC:2010.

Alarm sounds are compliant with Standard IEC 60601-1-8:2006+A1:2012 /

EN 60601-1-8:2007+AC:2010.

Electromagnetic Compatibility• IEC 60601-1-2:2007/EN 60601-1-2:2007+AC:2010

• ICES-003:2012

The monitor is classified as electromagnetic emissions Class B, ex

when used with the IUP/ECG patient module M2738A, then it is

classified as Class A.

This ISM device complies with Canadian ICES-003. Cet appareil ISM

conforme à la norme NMB-001 du Canada.

Environmental Specifications

The monitor may not meet the given performance specifications, i

stored and used outside the specified temperature and humidity

ranges.

Tympanic Temperature Thermometer

FM20/FM30 Monitor

Temperature Range

Operating without battery: 0-45°C (32-113°F)

with fully charged battery: 0-40°C (32-104°F)

with battery charging: 0-35°C (32-95°F)

Storage -20-60°C (-4-140°F) (except recorder paper)

Humidity Range

Operating <95% relative humidity @ 40°C (104°F)

Storage <90% relative humidity @ 60°C (140°F)

(except recorder paper)

Altitude Range

Operating -500-3000 m (-1640-9840 ft)

Storage -500-13100 m (-1640-43000 ft)

Transducers (M2734B/M2735A/M2736A/M2738A)

Temperature Range

Operating 0-40°C (32-104°F)

Storage -20-60°C (-4-140°F)

Humidity Range

Operating <95% relative humidity @ 40°C (104°F)

Storage <90% relative humidity @ 60°C (140°F)

Altitude Range

Operating -500-3000 m (-1640-9840 ft)

Storage -500-13100 m (-1640-43000 ft)

Philips SpO2 Sensors (FM30 only)

Operating Temperature Range

Reusable sensors (M119xX) 10-37°C (50-98.6

Disposable sensors (M113xA) 0-55°C (32-131°F

4

The print quality of this copy is not an accurate representation of the original.

Page 5: Avalon FM20/FM30 Fetal/Maternal Monitor - · PDF fileAvalon FM20/FM30 Fetal/Maternal Monitor ... The dual MIB, Medical Information ... • Connection of the external device to the

)

Physical Specifications

Performance Specifications

Complies with IEC 60601-2-37:2007/EN 60601-2-37:2008. ECG measurement follows IEC 60601-2-27:2011+AC:2012/EN 60601-2-

27:2006+AC:2006 except clauses listed below:

• 201.6.2, 201.8.5.5

• 201.12.1.101

• 202.6.2.101

FM20/FM30 Monitor

Power (without battery option)

Supply Voltages 100-240 VAC ± 10%

Supply Frequency Range 50-60 Hz

Power consumption (current) 0.7-0.4 A

Dimensions and Weight

Size: W x H x D (without options) 286 x 134 x 335 mm (11.3 x 5.3 x 13.2 in

Weight without battery 5.1 kg (11.2 lbs)

Weight with battery 5.3 kg (11.7 lbs)

Classification

Degree of Protection Against Electrical Shock Type CF

Electrical Class Class II equipment

Mode of Operation Continuous operation

Startup Time (Time taken from switching on the monitor to seeing the first parameter labels) <30 seconds

Ingress Protection

IPX1 (provided recorder drawer is shut)

Transducers (M2734B/M2735A/M2736A/M2738A)

Shock Resistance

Withstands a 1 m drop to concrete surface with possible cosmetic damage only

Water Ingress Protection Code

M2734B/M2735A/M2736A IP68 (immersion up to 1 m water depth for 5 hours)

M2738A IP67 (immersion up to 0.5 m water depth for 30 minutes)

Dimensions and Weight

M2734B/M2735A/M2736A Size (diameter) 83 mm (3.27 in)

Weight (incl. cable) 0.2 kg (0.5 lb)

M2738A W x H x D 42 x 30 x 123 mm (1.7 x 1.2 x 4.8 in)

Cable length 2.5 m (8.2 ft)

Weight 0.2 kg (0.5 lb)

Degree of Protection Against Electrical Shock

Type CF

Fetal/Maternal Performance Specifications

Ultrasound (M2736A/AA)

Measurement Method Ultrasound Pulsed Doppler

Measurement Range US 50-240 bpm

Resolution Display 1 bpm

Printer 1/4 bpm

Jitter @ 200 bpm ≤3 bpm

US Intensity Average output power P = (7.4 ± 0.4) mW

Peak rarefactional pressure pr = (40.4 ± 4.3) kPa

5

The print quality of this copy is not an accurate representation of the original.

Page 6: Avalon FM20/FM30 Fetal/Maternal Monitor - · PDF fileAvalon FM20/FM30 Fetal/Maternal Monitor ... The dual MIB, Medical Information ... • Connection of the external device to the

Output beam intensity (Iob) (= spatial average

- temporal average intensity)

Iob= Isata = (2.38 ± 0.59) mW/cm2

Spatial-peak temporal average intensity Ispta = (15.0 ± 3.2) mW/cm2

Effective radiating area @ -12 dB (3.11 ± 0.74) cm2

US Frequency 1 MHz ± 100 Hz

US Signal range 3.5 μVpp-350 μVpp @ 200 Hz

US Burst Repetition Rate 3.0 kHz

Duration ≤100 μs

Signal Quality Indication Good Full

Acceptable Half-full

Poor Empty

Beat to Beat Change (max.) for Ultrasound 28 bpm

Fetal Heart Rate (DECG) (FM30 only)

Measurement Range 30-240 bpm

Resolution Display 1 bpm

Recorder 1/4 bpm

Accuracy ±1 bpm or 1%, whichever is greater

Filter Bandwidth 0.8-80 Hz

INOP Auxiliary Current (Leads Off Detection) <100 μA

Input Signal Range DECG 20 μVpp-6 mVpp

Defibrillator Protection None

ESU Protection None

Toco

Measurement Method Strain Gauge Sensor Element

Sensitivity 1 unit = 2.5 g

Resolution Display 1 unit

Printer 1/4 unit

Measurement Range 400 units

Signal Range 0-127 units

Maximum Offset Range -300 units

Baseline Setting 20 units

Auto Offset Correction 3 seconds after connecting the transducer, the Toco value is set to 20 units

Auto Zero Adjust Toco value is set to zero following a negative measurement value for 5 seconds

Pulse from Toco MP

Measurement Range 40-240 bpm

Accuracy ±2% or 1 bpm, whichever is greater

Resolution 1 bpm

Wavelength Range 780-1100 nm

Emitted light energy ≤15 mW

Maternal Heart Rate (MECG)

Measurement Range 30-240 bpm

Resolution Display 1 bpm

Recorder 1/4 bpm

Accuracy ±1 bpm or 1%, whichever is greater

Filter Bandwidth 0.8-80 Hz

INOP Auxiliary Current (Leads Off Detection) <100 μA

Input Signal Range MECG 150 μVpp to 6 mVpp

Fetal/Maternal Performance Specifications

6

The print quality of this copy is not an accurate representation of the original.

Page 7: Avalon FM20/FM30 Fetal/Maternal Monitor - · PDF fileAvalon FM20/FM30 Fetal/Maternal Monitor ... The dual MIB, Medical Information ... • Connection of the external device to the

Alarm Specifications

Defibrillator Protection None

ESU Protection None

Fetal/Maternal Performance Specifications

IUP (FM30 only)

Measurement Method Passive Resistive Strain Gauge Elements

Measurement Range -100-300 mmHg

Signal Range -99-127 mmHg

Resolution Display 1 mmHg

Printer 1/4 mmHg

Sensitivity 5 μV/V/mmHg

Offset Compensation +100 to -200 mmHg

Baseline Setting 0 mmHg

Accuracy (not including sensor accuracy) ±0.5% per 100 mmHg

Auto Offset Correction 3 seconds after connecting the transducer, the IUP value is set to 0 mmHg

Fetal Heart Rate (Ultrasound Alarm Specifications)

FHR/DECG Alarm Limits Range Adjustment Default

Bradycardia (low limit) 60-200 bpm 10 bpm steps 110 bpm

Tachycardia (high limit) 70-210 bpm 10 bpm steps 170 bpm

FHR/DECG Alarm Delay

Bradycardia (low limit) 10-300 seconds 10 seconds 240 seconds

Tachycardia (high limit) 10-300 seconds 10 seconds 300 seconds

Signal Loss Delay 10-300 seconds 10 seconds

MHR/MECG Alarm Specifications

Range Adjustment Default

High Range 31-240 bpm 1 bpm steps (31-40 bpm), and

5 bpm steps (40-240 bpm)

120 bpm

Low Range 30-235 bpm 1 bpm steps (30-40 bpm), and

5 bpm steps (40-235 bpm)

50 bpm

Tachycardia Difference to high limit:

0-50 bpm

Adjustable in 5 bpm steps 20 bpm

Clamping at: 150-240 bpm Adjustable in 5 bpm steps 200 bpm

Bradycardia Difference to low limit:

0-50 bpm

Adjustable in 5 bpm steps 20 bpm

Clamping at: 30-100 bpm adjustable in 5 bpm steps 40 bpm

7

The print quality of this copy is not an accurate representation of the original.

Page 8: Avalon FM20/FM30 Fetal/Maternal Monitor - · PDF fileAvalon FM20/FM30 Fetal/Maternal Monitor ... The dual MIB, Medical Information ... • Connection of the external device to the

onal en

01-2-

se

1-2-

a

nce

ax.

Noninvasive Blood Pressure

Measurement Validation: In adult mode, the blood pressure measurements determined with this device comply with the American NatiStandard for Electronic or Automated Sphygmomanometers (ANSI/AAMI SP10 - 1992) in relation to mean error and standard deviation, whcompared to intra-arterial or auscultatory measurements (depending on the configuration) in a representative patient population. For the auscultatory reference, the 5th Korotkoff sound was used to determine the diastolic pressure. Complies with IEC 80601-2-30:2010/EN 80630:2010.

Alarm Specifications NBP

SpO2 (FM30 only)

Maternal pulse oximetry provides SpO2, pulse numerics, and a pulse trace for oxygen saturation (SpO2) of maternal aterial hemoglobin. Pulrate is derived from SpO2 if ECG is not connected.Complies with ISO 80601-2-61:2011/EN 80601-2-61:2011 (except alarm system; alarm system complies with IEC 60601-2-49:2011/EN 606049:2001).Measurement Validation: The SpO2 accuracy has been validated in human studies against arterial blood sample reference measured withCO-oximeter. Pulse oximeter measurements are statistically distributed, only about two-thirds of the measurements can be expected to fallwithin the specified accuracy compared to CO-oximeter measurements. Display Update Period: Typical: 2 seconds, Maximum: 30 seconds. Maximum with noninvasive blood pressure INOP suppression on: 60 seconds. The specified accuracy is the root-mean-square (RMS) differebetween the measured values and the reference values.

Noninvasive Blood Pressure Performance Specifications

Measurement Ranges Systolic 30-270 mmHg (4-36 kPa)

Diastolic 10-245 mmHg (1.5-32 kPa)

Mean 20-255 mmHg (2.5-34 kPa)

Accuracy Max. Std. Deviation: 8 mmHg (1.1 kPa) M

Mean Error: ±5 mmHg (±0.7 kPa)

Pulse Rate Range 40-300 bpm

Accuracy: (average over noninvasive blood

pressure measurement cycle)

40-100 bpm: ±5 bpm

101-200 bpm: ±5% of reading

201-300 bpm: ±10% of reading

Measurement Time Typical at HR >60 bpm Auto/manual: 30 seconds Maximum time: 180 seconds

Cuff Inflation Time Typical for normal cuff: less than 10 seconds

Initial Cuff Inflation Pressure 165 ± 15 mmHg

Auto Mode Repetition Times 1, 2, 2.5, 3, 5, 10, 15, 20, 30, 45, 60, or

120 minutes

Venipuncture Mode Inflation

Inflation Pressure 20-120 mmHg (3-16 kPa)

Automatic deflation after 170 seconds

Alarm Specifications Range Adjustment

Systolic 30-270 mmHg (4-36 kPa)10-30 mmHg: 2 mmHg (0.5 kPa) >30 mmHg: 5 mmHg (1 kPa)

Diastolic 10-245 mmHg (1.5-2 kPa)

Mean 20-255 mmHg (2.5-34 kPa)

Overpressure Settings Adjustment

>300 mmHg (40 kPa) >2 sec not user adjustable

SpO2 Performance Specifications

Range 0-100%

Accuracy Philips Reusable Sensors:

M1191A/B, M1191AL/BL, M1191ANL, M1192A, M1192AN = 2% (70% to 100%) M1191T,

M1192T, M1194A, M1194AN, M1196A, M1196T = 3% (70% to 100%)

8

The print quality of this copy is not an accurate representation of the original.

Page 9: Avalon FM20/FM30 Fetal/Maternal Monitor - · PDF fileAvalon FM20/FM30 Fetal/Maternal Monitor ... The dual MIB, Medical Information ... • Connection of the external device to the

)

%)

Alarm Specifications SpO2

External Power Supply Specifications

Philips Disposable Sensors with M1943A(L):

M1131A, M1901B, M1903B, M1904B = 3% (70% – 100%) M1133A, M1134A = ±2% (70%

100%)

NellcorPB® Sensors with M1943A(L):

MAX-A, MAX-AL, MAX-P, MAX-N, D-25, D-20, N-25, OxiCliq A, P, N = 3% (70% – 100%

Accuracy Masimo Reusable Sensors® with LNOP MP12 or LNC MP10:

LNOP DC-I, LNOP DC-IP, LNOP YI, LNCS DC-I, LNCS DC-IP: 2% (70% – 100%) LNOP

TC-I, LNCS TC-I: 3.5% (70% – 100%)Masimo Disposable Sensors® with LNOP MP12 or LNC MP10:

LNOP Adt, LNOP Adtx, LNOP Pdt, LNOP Pdtx, LNCS Adtx, LNCS Pdtx: 2% (70% – 100

LNOP Neo-L, LNCS Neo-L: 3% (70% to 100%)

Resolution 1%

Pulse

Range 30-300 bpm

Accuracy ±2% or 1 bpm, whichever is greater

Resolution 1 bpm

Sensors

Wavelength range 500-1000 nm information about the wavelength range can be especially useful to clinicians

(for instance, when photodynamic therapy is performed)

Emitted Light Energy ≤15 mW

Pulse Oximeter Calibration Range 70%-100%

SpO2 Performance Specifications

SpO2 Alarm Range Adjustment Delay

SpO2 50-100% 1% steps (0, 1, 2, 3,... 30) + 4 sec

Desat 50 - low alarm limit 1% steps

Pulse 30-300 bpm 1 bpm steps (30-40 bpm) 5 bpm

steps (40-300 bpm)

max. 14 sec

Tachycardia Difference to high limit 0-50 bpm 5 bpm steps max. 14 sec

Clamping at 150-300 bpm 5 bpm steps

Bradycardia Difference to low limit 0-50 bpm 5 bpm steps max. 14 sec

Clamping at 30-100 bpm 5 bpm steps

M8023A External Power Supply Specifications

Dimensions and Weight

Size: W x H x D (without options) 208 x 105 x 135 mm (8.2 x 4.1 x 5.3 in)

Weight 0.6 kg (1. lbs)

Power Specifications

Line Voltage 100-240 V approximately

Current 1.3-0.7 A

Frequency 50-60 Hz approximately

Indicators

AC Power green LED

Ingress Protection for External Power Supply (M8023A)

IP31 (protected against the ingress of solid foreign objects 2.5 mm in diameter or larger, and the ingress of water when water is dripping

vertically) when rested on its rubber feet on a flat, level surface

9

The print quality of this copy is not an accurate representation of the original.

Page 10: Avalon FM20/FM30 Fetal/Maternal Monitor - · PDF fileAvalon FM20/FM30 Fetal/Maternal Monitor ... The dual MIB, Medical Information ... • Connection of the external device to the

nds

ad

m

m

Battery Specifications

Recorder Specifications

IP32 when mounted with the connectors facing downwards

M8023A External Power Supply Specifications

M4605A Battery Specifications

Performance Specifications Avalon FM20/30 Battery Option E25

Operating Time (with new, fully charged

battery)

Basic monitoring configuration: >2 hours (Display Brightness: 70%, Recorder: "On" at 3 cm/min, NBP: Auto Mode at 15 min,

2 US Transducers, 1 Toco+ with MECG, 1 Patient Module with DECG)

Charge Time When monitor is off, approx. 6 hours, when monitor is in use, more than 10 hours

(depending on monitor configuration)

Built-in Thermal Array Fetal Trace Recorder

Paper and Printing

Type Standard Z-fold thermal paper

Standard Speeds (real-time

traces)

3 cm/min, 2 cm/min, 1 cm/min

ECG Wave Recording (FM30

only. Not real-time.)

Emulated 25 mm/s print speed is variable up to 20 mm/s and depe

on the print load

Stored Data Recording/Trace

Recovery Printout

Print speed is variable up to 20 mm/s and depends on the print lo

Paper Advance Up to 20 mm/s

Sensing Optical Reflex Sensor for black page marks

Accuracy @ 3 cm/min, 2 cm/

min, 1 cm/min

±5 mm/page

Usable Print Width 128 mm

Resolution 8 dots/mm (200 dpi)

Time delay to see trace on paper <30s @ 1 cm/min

Trace Separation Offset for FHR

(Ultrasound/DECG)

Twin Standard FHR2 +20 bpm

Classic FHR1 +20 bpm

Triplet Standard FHR2 +20 bpm/FHR3 -20 bp

Classic FHR1 +20 bpm/FHR3 -20 bp

10

The print quality of this copy is not an accurate representation of the original.

Page 11: Avalon FM20/FM30 Fetal/Maternal Monitor - · PDF fileAvalon FM20/FM30 Fetal/Maternal Monitor ... The dual MIB, Medical Information ... • Connection of the external device to the

lable.

with

your

d

cers.

sary.

d

Ordering Information and Accessory Options

You can order the Avalon FM20/FM30 fetal monitor under the product number M2702A/M2703A. There are several ordering options avai

Fetal Monitor Ordering Options

The following tables give an overview of the monitor’s standard capabilities, and its options. To order, precede the required option number

M2702A or M2703A (for example, M2702A with option B71 adds the noninvasive blood pressure measurement). “L” options let you modify

order. Features listed as options can also be added at a later time (see “Upgrade Options” on page 19).

The fetal monitor can be operated with the Avalon CL Transducer System, or with cabled transducers. The maternal measurements NBP an

SpO2 can also be measured cablelessly by ordering the IntelliVue Cableless Pods that are supported by the Avalon CL Transducer System.

If you want to operate the fetal monitor only cablelessly, order the L00 option, so that the fetal monitor is delivered without cabled transdu

The fetal monitor can also be ordered with both, cabled and cableless transducers, to switch during monitoring between the options if neces

Refer to “The following accessories are delivered with any fetal monitor order.” on page 17 for a list of items shipped as part of the standar

configuration. Start-up quantities of fetal supplies are available in five convenient kits.

Cableless Option Avalon CL Transducer System

Optional Interfaces

Paper Scaling Options (Factory configuration)

Transport Options (Battery and External Power Supply)

Description Comments Options

FM20/30 Fetal Monitor (M2702A/M2703A) The monitor is delivered without cabled transducers. The

Avalon CL Transducer System is required for monitoring.

L00

CL Base Station (866074)

See the Avalon CL Technical Data Sheet for ordering

details, options, and accessories.

CL Toco+ MP Transducer (866075)

CL Ultrasound Transducer (866076)

CL ECG/IUP Transducer (866077)

IntelliVue CL SpO2 Pod (865215) See the IntelliVue Cableless Measurements Technical Data

Sheet for ordering details, options, and accessories.IntelliVue CL NBP Pod (865216)

Description Comments Option1

System Interface (RS232 port and LAN port) For interfacing to Philips IntelliSpace Perinatal and service

(software upgrade).

J70

Dual MIB/RS232 interface board For data export or interface to the Tympanic Temperature

Thermometer.

J13

Dual PS/2 interface board For keyboard and mouse connection (M8024A #A01) J22

Quad. USB interface board For keyboard, mouse, or bar code reader) J25

Flexible Nurse Call Relay interface board J30

Telemetry interface cable (Avalon CTS Cable M2731-

60001) 453564203941

For interfacing with an Avalon CTS transducer system (no

longer for sale, but still supported).

K30

1 Only two interfaces from the Jxx options can be fitted in the fetal monitor at the same time.

Description Option

50-210 bpm paper scaling P01

30-240 bpm paper scaling P02

Description Option

Battery (M4605A), external power supply (M8023A with option E25), and MSL cable (1 m) E25

11

The print quality of this copy is not an accurate representation of the original.

Page 12: Avalon FM20/FM30 Fetal/Maternal Monitor - · PDF fileAvalon FM20/FM30 Fetal/Maternal Monitor ... The dual MIB, Medical Information ... • Connection of the external device to the

51

71

31

41

Standard Measurements Cabled Transducers

Fetal Heart Rate (FHR)

Direct ECG (DECG) FM30 only

Remote Event Marker

Uterine Measurements

Intrauterine Pressure

Measurement Description Comments Option

Ultrasound FHR singletons One M2736A (M2736AA for US)

transducer M2736A includes Fetal Movement

Profile (FMP)

L01

Ultrasound FHR twins Two M2736A (M2736AA for US)

transducers

L02

Ultrasound FHR triplets Three M2736A (M2736AA for US)

transducers

Option C73 is required. US

transducers may be borrowed from a

second Avalon fetal monitor.

L03

Description Comments Option

M2735A Toco+ transducer (includes ECG/IUP capability)

Includes DECG wave display. All transducers require a K51

DECG adapter cable.

L11

1 M2738A ECG/IUP patient module L21

2 M2738A ECG/IUP patient modules L22

3 M2738A ECG/IUP patient modules L23

DECG Accessories: Philips DECG Solution Part No.

DECG reusable leg plate adapter cable (with flushing port) (K51) 9898031376

DECG leg attachment electrode for DECG leg plate adapter cable, box of 25 9898031397

DECG fetal scalp electrode: single spiral, worldwide availability, box of 25 9898031376

DECG fetal scalp electrode: double spiral, Europe only. Not for USA, box of 25 9898031376

Description Comment Option

Remote Event Marker (989803143411) Can be connected to the fetal sockets at the fetal monitor,

and is operated manually to record an event on the trace.

L30

Measurement Description Comments Option

Toco (external)

M2734B Toco MP transducer

(maternal Smart Pulse integrated)

Toco and maternal pulse are measured

in parallel.

L10

M2735A Toco+ transducer (includes

ECG/IUP capability)

L11

Measurement Description Comments Option

Intrauterine pressure

M2735A Toco+ transducer (includes

ECG/IUP capability)All transducers require a K52 IUP

adapter cable.

L11

1 M2738A ECG/IUP patient module L21

2 M2738A ECG/IUP patient modules L22

3 M2738A ECG/IUP patient modules L23

12

The print quality of this copy is not an accurate representation of the original.

Page 13: Avalon FM20/FM30 Fetal/Maternal Monitor - · PDF fileAvalon FM20/FM30 Fetal/Maternal Monitor ... The dual MIB, Medical Information ... • Connection of the external device to the

31

Optional Fetal Measurements

Maternal Heart Rate (MHR)

MECG Accessories

Optional Maternal Measurements

Noninvasive Blood Pressure (NBP)

Noninvasive Blood Pressure Accessories

Multi-Patient Comfort Cuffs and Disposable Cuffs with M1598B (1.5 m) or M1599B (3.0 m) Tubing

Koala IUP Accessories Part No.

Disposable Koala IUP catheter, box of 10 M1333A

Reusable IUP adapter cable (corresponds to option K52) 9898031439

Description Comments Option

NST Report Capability and NST Trace Interpretation Analyzes the data according to the NICHD guidelines. C71

Triple FHR via ultrasound Requires the option L03. FM30 only: Triplets monitoring is also possible with 2 x US

and 1 x DECG (membranes ruptured).

C73

Measurement Description Comments Option

Maternal pulse (MP) rate monitoringM2734B Toco MP transducer

(maternal Smart Pulse integrated)

Toco and maternal pulse are measured

in parallel

L10

Maternal ECG (MECG) heart rate

monitoring

M2735A Toco+ transducer (includes

ECG/IUP capability) All transducers require a K53 MECG

adapter cable. MECG wave available

with FM30 only.

L11

1 M2738A ECG/IUP patient module L21

2 M2738A ECG/IUP patient modules L22

3 M2738A ECG/IUP patient modules L23

Maternal Heart Rate Accessories Part No.

MECG adapter cable (corresponds to option K53) M1363A

Foam MECG electrodes, snap-fit, for MECG adapter cable, 5 per pack, 300 per case 40493D

Foam MECG electrodes, snap-fit, for MECG adapter cable, 30 per pack, 300 per case 40493E

Measurement Description Option

Noninvasive Blood Pressure (NBP) with pulse rate B71

NBP cuff assortment and tubing Accessory option K31

Maternal Patient

CategoryLimb Circumference Bladder Width

Disposable Cuff Part No.

Reusable Cuff Part No.

Adult (thigh) 42.0-54.0 cm 20.0 cm M1879A M1576A

Large adult 34.0-43.0 cm 16.0 cm M1878A M1575A

Adult 27.0-35.0 cm 13.0 cm M1877A M1574A

Small adult 20.5-28.0 cm 10.5 cm M1876A M1573A

Cuff kit of 4 adult sizes M1578A

13

The print quality of this copy is not an accurate representation of the original.

Page 14: Avalon FM20/FM30 Fetal/Maternal Monitor - · PDF fileAvalon FM20/FM30 Fetal/Maternal Monitor ... The dual MIB, Medical Information ... • Connection of the external device to the

Antimicrobial Coated Reusable Cuffs with M1598B (1.5 m) or M1599B (3.0 m) Tubing

Soft Single Patient Single-Hose Disposable Cuffs with M1598B (1.5 m) or M1599B (3.0 m) Tubing

NBP Cuff Assortment Kit

Pulse Oximetry

Maternal Patient Category

(color)Limb Circumference Bladder Width Part No.

Adult thigh (gray) 45.0-56.5 cm 21.0 cm M4559B

Adult thigh (gray), pack of 5 cuffs 45.0-56.5 cm 21.0 cm M4559B5

Large adult X-long (burgundy) 35.5-46.0 cm 17.0 cm M4558B

Large adult X-long (burgundy), pack of

5 cuffs

35.5-46.0 cm 17.0 cm M4558B5

Large adult (burgundy) 35.5-46.0 cm 17.0 cm M4557B

Large adult (burgundy), pack of 5 cuffs 35.5-46.0 cm 17.0 cm M4557B5

Adult X-long (navy blue) 27.5-36.5 cm 13.5 cm M4556B

Adult X-long (navy blue), pack of 5

cuffs

27.5-36.5 cm 13.5 cm M4556B5

Adult (navy blue) 27.5-36.5 cm 13.5 cm M4555B

Adult (navy blue), pack of 5 cuffs 27.5-36.5 cm 13.5 cm M4555B5

Small adult (royal blue) 20-28.5 cm 10.6 cm M4554B

Small adult (royal blue), pack of 5 cuffs 20.5-28.5 cm 10.6 cm M4554B5

Cuff kits containing one small adult, one adult, one large adult, and one thigh cuff 864288

Cuff kits containing one small adult, one adult, one adult X-long, one large adult, one large adult X-long, and one thigh cuff 864291

Patient Category Limb Circumference Bladder Width Part No.

Adult (thigh) 45.0-56.5 cm 20.4 cm M4579B

Large adult X-long 35.5-46.0 cm 16.4 cm M4578B

Large adult 35.5-46.0 cm 16.4 cm M4577B

Adult X-long 27.5-36.5 cm 13.1 cm M4576B

Adult 27.5-36.5 cm 13.1 cm M4575B

Small adult 20.5-28.5 cm 10.4 cm M4574B

Description Part No.

Antimicrobial NBP cuff assortment kit with 3 m (10 ft) connection tubing (corresponds to option K31) M1599B

M1573A

M1574A

M1575A

Measurement Description Option

Pulse Oximetry (maternal SpO2) B73 includes NBP monitoring (FM30 only) B73

Reusable maternal SpO2 Sensor M1191BL (3 m cable) K91

14

The print quality of this copy is not an accurate representation of the original.

Page 15: Avalon FM20/FM30 Fetal/Maternal Monitor - · PDF fileAvalon FM20/FM30 Fetal/Maternal Monitor ... The dual MIB, Medical Information ... • Connection of the external device to the

SpO2 AccessoriesAll listed sensors operate without the risk of exceeding 41°C on the skin if the initial skin temperature is below 37°C (FM30 only).

Philips Reusable Sensors

Philips Disposable Sensors

Description Comments Part No.

Adult sensor (2.0 m cable), for patients over 50 kg. Any

finger except thumb.

No adapter cable required.

M1191B

M1191B with longer cable (3.0 m) M1191BL

Small adult, pediatric sensor (1.5 m cable) for patients

between 15 kg and 50 kg. Any finger except thumb. Use

only on adult patients with FM30/40/50.

M1192A

Ear sensor (1.5 m cable) for patients more than 40 kg. Use

only on adult patients with FM30/40/50.

M1194A

Adult clip sensor (3 m cable) for patients over 40 kg. Any

finger except thumb.

M1196A

Adult clip sensor (2 m cable) for patients over 40 kg. Any

finger except thumb.

M1196S

Adult sensor (0.45 m cable), for patients over 50 kg. Any

finger except thumb.

Requires M1943A (1.0 m) or M1943AL (3.0 m) adapter

cable.

M1191T

Small adult or pediatric sensor (0.45 m cable) for patients

between 15 kg and 50 kg. Any finger except thumb. Use

only on adult patients with FM30/40/50.

M1192T

Adult clip sensor (0.9 m cable) for patients over 40 kg. Any

finger except thumb.

M1196T

Special Edition (SE): Adult sensor (3 m cable), for patients

over 50 kg. Any finger except thumb.

No adapter cable required. SE sensors work with FM30/40/

50, as well as with OxiMax-compatible SpO2 versions of

other Philips monitors.

M1191ANL

Special Edition (SE): Small adult or pediatric sensor (1.5 m

cable) for patients between 15 kg and 50 kg. Any finger

except thumb. Use only on adult patients with FM30/40/50.

M1192AN

Special Edition (SE): Ear sensor (1.5 m cable) for patients

more than 40 kg.

M1194AN

Description Comments Part No.

Identical to OxiMax MAX-ARequires M1943A (1.0 m) or M1943AL (3.0 m) adapter

cable

M1904B1

Identical to OxiMax MAX-P M1903B2

Identical to OxiMax MAX-N M1901B3

Adult/pediatric finger sensor (0.45 m cable). Use only on

adult patients with FM30/40/50.Requires M1943A (1.0 m) or M1943AL (3.0 m) adapter

cable

M1131A

Adult/pediatric finger sensor (0.9 m cable) for patients

>40 kg. Any finger except thumb. Use only on adult

patients with FM30/40/50.

M1133A

1 Not available in the USA.2 Not available in the USA.3 Not available in the USA.

15

The print quality of this copy is not an accurate representation of the original.

Page 16: Avalon FM20/FM30 Fetal/Maternal Monitor - · PDF fileAvalon FM20/FM30 Fetal/Maternal Monitor ... The dual MIB, Medical Information ... • Connection of the external device to the

r

r

r

21

31

41

81

91

01

31

61

91

Nellcor SensorsNellcor sensors must be ordered from Medtronic.

Masimo LNOP Reusable Sensors

These sensors require the adapter cable Masimo MP12 Cable (M1020-61100).

Masimo LNCS Reusable Sensors

These sensors require the adapter cable LNC MP10 (989803148221).

Masimo LNOP Disposable Adhesive Sensors

These sensors require the adapter cable Masimo MP12 Cable (M1020-61100).

Product Name Description Comments

OxiMax MAX-A Adult finger sensor (patient size >30 kg)

Requires M1943A (1.0 m) or M1943AL (3.0 m) adapte

cable.

OxiMax MAX-AL OxiMax MAX-A with long cable

OxiMax MAX-P Pediatric or small adult foot/hand sensor (patient size

10-50 kg). Use only on adult patients with FM30/40/50

OxiMax MAX-N Adult finger or neonatal foot/hand sensor (patient size

>40 kg or <3 kg). Use only on adult patients with FM30/

40/50.

Oxisensor II D-25 Adult sensor (patient size >30 kg)

Requires M1943A (1.0 m) or M1943AL (3.0 m) adapte

cable

Oxisensor II D-20 Pediatric sensor or small adult (patient size 10-50 kg). Use

only on adult patients with FM30/40/50.

Oxisensor II N-25 Neonatal/adult sensor (patient size <3 kg or >40 kg). Use

only on adult patients with FM30/40/50.

OxiCliq A See OxiMax MAX-A

Requires M1943A (1.0 m) or M1943AL (3.0 m) adapte

cable together with OC-3 adapter cable

OxiCliq P See OxiMax MAX-P Use only on adult patients with

FM30/40/50

OxiCliq N See OxiMax MAX-N Use only on adult patients with

FM30/40/50

Product Name Description Part No.

LNOP DC-I Adult finger sensor (>30 kg) 9898031403

LNOP DC-IP Pediatric finger sensor (10-50 kg). Use only on adult patients with FM30/40/50. 9898031403

LNOP YI Multi-site sensor (>1 kg). Use only on adult patients with FM30/40/50. n/a

LNOP TC-I Ear sensor (>30 kg). Use only on adult patients with FM30/40/50. 9898031403

Product Name Description Part No.

LNCS DC-I Adult Finger Sensor (>30 kg) 9898031482

LNCS DC-IP Pediatric Finger Sensor (10-50 kg). Use only on adult patients with FM30/40/50 9898031482

LNCS TC-I Ear Sensor (>30 kg). Use only on adult patients with FM30/40/50 9898031483

Product Name Description Part No.

LNOP Adt Adult Sensor (>30 kg) 9898031402

LNOP Adtx Adult Sensor (>30 kg) n/a

LNOP Pdt Pediatric Sensor (10-50 kg). Use only on adult patients with FM30/40/50 9898031402

LNOP Pdtx Pediatric Sensor (10-50 kg). Use only on adult patients with FM30/40/50 n/a

LNOP Neo-L Neonatal (<3 kg) or Adult adhesive Sensor (>40 kg). Use only on adult patients with FM30 9898031402

16

The print quality of this copy is not an accurate representation of the original.

Page 17: Avalon FM20/FM30 Fetal/Maternal Monitor - · PDF fileAvalon FM20/FM30 Fetal/Maternal Monitor ... The dual MIB, Medical Information ... • Connection of the external device to the

31

41

71

1

2

r

01

Masimo LNCS Disposable Adhesive Sensors

These sensors require the adapter cable LNC MP10 (989803148221).

Tympanic Temperature Thermometer

Additional Accessories

The following accessories are delivered with any fetal monitor order.

Fetal Supplies Starter Kits

Additional Fetal Accessories

Product Name Description Part No.

LNCS Adtx Adult Sensor (>30 kg) 9898031482

LNCS Pdtx Pediatric Finger Sensor (10-50 kg). Use only on adult patients with FM30/40/50 9898031482

LNCS Neo-L Neonatal Foot Sensor (<3 kg) or Adult Finger Sensor (>40 kg). Use only on adult patients with

FM30/40/50

9898031482

Tympanic Temperature Thermometer Part No.

Tympanic Temperature Thermometer (requires J13), including:

• a base station for mounting on a table, wall, or a roll stand

• A starter pack of 96 probe covers

866149

Adding a 1.5 m, RS232 connection cable 866149 #A0

Adding a 3.0 m, RS232 connection cable 866149 #A0

Description Quantity

Fetal paper pack (country-specific, installed) 1

Power cable 1

Documentation DVD (Instructions for Use, and all other fetal monitor documentation) 1

Description Contents Kit Numbe

Standard Antepartum Kit (North America) Paper (30-240), belts, gel, belts buttons, belt clips 862214

Standard Antepartum Kit (Europe, Latin America, Asia) Paper (50-210), belts, gel, belts buttons, belt clips 862215

Standard Antepartum Kit (Japan) Paper (50-210), belts, gel, belts buttons, belt clips 862216

FM30 only

Intrapartum DECG Kit (single spiral) Fetal scalp electrodes, fetal adhesive pad electrodes 862217

Intrapartum DECG Kit (double spiral) Europe only Fetal scalp electrodes, fetal adhesive pad electrodes 862218

Fetal Accessories Part No.

Belt (reusable, gray, water resistant)

32 mm wide, 15 m roll (32 mm belts for use with M2738A Patient Module only) M4601A

60 mm wide, 5 belts M4602A

60 mm wide, 15 m roll M4603A

50 mm wide, 5 belts M1562B

Belt (disposable, yellow, water resistant)

60 mm wide, pack of 100 M2208A

Ultrasound Gel

12 Bottles 40483A

5 liter refill (with dispenser) for 40483A Shelf life: 24 months maximum 40483B

Belt Buttons and Belt Clip

Belt buttons (pack of 10) M1569A

Butterfly belt clip (pack of 6) 9898031434

17

The print quality of this copy is not an accurate representation of the original.

Page 18: Avalon FM20/FM30 Fetal/Maternal Monitor - · PDF fileAvalon FM20/FM30 Fetal/Maternal Monitor ... The dual MIB, Medical Information ... • Connection of the external device to the

41

01

01

01

02

03

d

Recorder PaperSupplied in cases of 40 packs. Each pack has 150 numbered pages

Input Devices

Mounting Hardware

Telemetry Interface Cable

Telemetry interface cable Avalon CTS Cable M2731-60001 4535642039

Fetal Accessories Part No.

Geography FHR Scale Grid Color Scale UnitsHighlighted 3 cm

Lines?Part No.

USA/Canada and Asia 30-240 Red/Orange mmHg Yes M1910A

Europe/Japan 50-210 Green mmHg and kPa No M1911A

Japan 50-210 Green mmHg Yes M1913A

Japan 50-210 Green1 mmHg Yes M1913J1 Bradycardia and tachycardia alarm ranges are shaded.

Description Option

Slimline keyboard with integrated trackball (includes spill cover) M8024A #A

Optical mouse M8024A #B

Trackball M8024A #C

Wireless trackball M8024A #C

Hand-Track M8024A #C

Mounts, Carts, and Rollstands

M2740A #A01 M2740A #A05 M2740A #B01 M2740A #C02

Flush Wall Mount for a flat

wall mounting

Wall Mount (Arm) Bed Hanger Cart with fixed angle mount an

two drawers

M2740A #R01 M2740A #U02 M2740A #W01 M2740A #C02 Option C11

Rollstand with tray Mounting Kit for Avalon CL for

use with M2740A #C02

Wall Channel required for wall

mounts A01 and A05

Additional drawer for the cart

18

The print quality of this copy is not an accurate representation of the original.

Page 19: Avalon FM20/FM30 Fetal/Maternal Monitor - · PDF fileAvalon FM20/FM30 Fetal/Maternal Monitor ... The dual MIB, Medical Information ... • Connection of the external device to the

Upgrade Options

Upgrade options are prefixed with M2702AU/ M2703AU. For example, to add a system interface upgrade, order M2702AU option J70.

Only two interfaces from the Jxx options can be fitted in the fetal monitor at the same time.

Description Option

Noninvasive blood pressure, no supplies included B71

Pulse oximetry (SpO2), no supplies included (FM30 only) B72

NBP and SpO2 measurement (FM30 only) B73

NST report capability C71

Triplets monitoring capability C73

Support for cableless maternal measurements (IntelliVue CL NBP and CL SpO2). Not available with option E25 CL2

Battery upgrade for monitors with Rev. G software or higher E25

Dual MIB/RS232 Interface for Data Export functionality, and for maternal infrared temperature measurement J13

Dual PS/2 Interface for connecting a keyboard and mouse J22

Quadruple USB Interface (Keyboard, Mouse, Bar Code Reader) J25

Flexible Nurse Call Relay Interface J30

System Interface: RS232 port and LAN port J70

Avalon CTS Interface Cable M2731-60001 K30

Rel. J.3 Software Upgrade SJ3

19

The print quality of this copy is not an accurate representation of the original.

Page 20: Avalon FM20/FM30 Fetal/Maternal Monitor - · PDF fileAvalon FM20/FM30 Fetal/Maternal Monitor ... The dual MIB, Medical Information ... • Connection of the external device to the

© 2016 Koninklijke Philips N.V. All rights are reserved.

Philips Healthcare reserves the right to make changes in specifications and/or to discontinue any product at any time without notice or obligation and will not be liable for any consequences resulting from the use of this publication.

Printed in The Netherlands. 452299111931* March 2016

Philips Healthcare is part of Royal Philips

How to reach us

www.philips.com/healthcare [email protected] fax: +31 40 27 64 887

Asia +852 2821 5888

Europe, Middle East, Africa +49 7031 463 2254

Latin America +55 11 2125 0744

North America +1 425 487 7000 800 285 5585 (toll free, US only)

Avalon FM20/M2702A and FM30/

M2703A comply with the

requirements of the Council Directive

93/42/EEC of 14 June 1993 (Medical

Device Directive).

0123

Please visit www.philips.com/