Application of UV/Vis Spectroscopy Common Problems: a) Mixtures: - blank samples often contain...

38
Application of UV/Vis Spectroscopy Common Problems: a) Mixtures: - blank samples often contain multiple absorbing species. - the absorbance is the sum of all the individual absorbencies A= A 1 + A 2 +A 3 + … = 1 bc + 2 bc + 3 bc … - substances in both the blank and sample which absorb can be “blanked out” in both double and single beam spectrometers.

Transcript of Application of UV/Vis Spectroscopy Common Problems: a) Mixtures: - blank samples often contain...

Page 1: Application of UV/Vis Spectroscopy Common Problems: a) Mixtures: - blank samples often contain multiple absorbing species. - the absorbance is the sum.

Application of UV/Vis Spectroscopy

Common Problems:a) Mixtures:

- blank samples often contain multiple absorbing species.- the absorbance is the sum of all the individual absorbencies

A= A1 + A2 +A3 + … = 1bc + 2bc + 3bc …

- substances in both the blank and sample which absorb can be “blanked out” in both double and single beam spectrometers.

Page 2: Application of UV/Vis Spectroscopy Common Problems: a) Mixtures: - blank samples often contain multiple absorbing species. - the absorbance is the sum.

If blank absorbance too high:- dilute the sample- use a different where the analyte absorbs more relative to the interference.- use a different method of separation

But, If the blank absorbance is high, Po will decrease too much, the response will be slow and the results inaccurate

scan of substance

Large blank

absorbance

Page 3: Application of UV/Vis Spectroscopy Common Problems: a) Mixtures: - blank samples often contain multiple absorbing species. - the absorbance is the sum.

b) Instrumental Deviations from Beer’s Law:- stray light (already discussed).- polychromatic light (more then a single )

‚ since all instruments have a finite band pass, a range of ’s are sent through the sample.

‚ may be different for each

Deviation’s from Beer’s law at high concentration

Page 4: Application of UV/Vis Spectroscopy Common Problems: a) Mixtures: - blank samples often contain multiple absorbing species. - the absorbance is the sum.

Illustration of Deviation from Beer’s law:

Let us say that exactly 2 wavelengths of light were entering the sample = 254 nm 254 =10,000

= 255 nm 255 = 5,000

let Po = 1 at both ’s

What happens to the Beer’s law plot as c increases?

A =A254 + A255

A = log Po/P (total) = log (Po254 + Po

255)/(P254 + P255)

At the individual ’s:

A254 = 254bc = log Po254/P254

10bc = Po254/P254

P254 = Po/10bc = Po10-bc

For both together:

A = log (Po254 + Po

255)/(Po25410-254bc + Po

25510-255bc)

Page 5: Application of UV/Vis Spectroscopy Common Problems: a) Mixtures: - blank samples often contain multiple absorbing species. - the absorbance is the sum.

Since Po254 = Po

255 =1:

A = 2.0/(Po25410-254bc + Po

25510-255bc)

A = 2.0/(10-10,000x1.0xc + 10-5000x1.0xc)

C A (actual) A(expected)

10-6M 0.0075 0.0075

10-5M 0.074 0.075

10-4M 0.068 0.75

10-3M 5.3 7.50

1

2

3

4

5

6

7

8

2 10-4 4 10-4 6 10-4 8 10-4 1 10-3

Concentration (M)

Negative Deviation

A = bc

The results are the same for more ’s of light. The situation is worse for greater differences in ’s (side of absorption peak, broad bandpass)

Always need to do calibration curve! Can not assume linearity outside the range of linearity curve!

Page 6: Application of UV/Vis Spectroscopy Common Problems: a) Mixtures: - blank samples often contain multiple absorbing species. - the absorbance is the sum.

HInKa

H+ + In-

Red, =600nm colorless

phenolphthalein:

If solution is buffered, then pH is constant and [HIn] is related to absorbance.

b) Chemical Deviations from Beer’s Law:- Molar absorptivity change in solutions more concentrated than 0.01M

‚ due to molecular interactions

‚ Beer’s law assumes species are independent

‚ electrolytes may also cause this problem

‚ is also affected by the index of refraction

- association, dissociation, precipitation or reaction of analyte

‚ c in Beer’s law is the concentration of the absorbing species

‚ commonly use the analytical concentration – concentration of all forms of the species.

Page 7: Application of UV/Vis Spectroscopy Common Problems: a) Mixtures: - blank samples often contain multiple absorbing species. - the absorbance is the sum.

CHIn [HIn] [In-] [HIn]/[In-]

10-5 8.5x10-7 9.2x10-6 0.0924

10-4 3.8x10-5 6.2x10-5 0.613

10-3 7.3x10-4 2.7x10-4 2.70

But, if unbuffered solution, equilibrium will shift depending on total analyte concentration

example: if Ka = 10-4 HInKa

H+ + In-

CHIn

Expected

Actual

“Apparent” deviation since can be accounted for by chemical equilibrium

Page 8: Application of UV/Vis Spectroscopy Common Problems: a) Mixtures: - blank samples often contain multiple absorbing species. - the absorbance is the sum.

Isosbestic point At the isosbestic point in spectra:

A = b([HIn] + [In-])

But, if unbuffered solution, equilibrium will shift depending on total analyte concentration

example: if Ka = 10-4 HInKa

H+ + In-

Page 9: Application of UV/Vis Spectroscopy Common Problems: a) Mixtures: - blank samples often contain multiple absorbing species. - the absorbance is the sum.

c) Non-constant b:- worse for round cuvettes- use parallel cuvettes to help

P0

A = log10 Po/P = bc

P1

B1

B2

P2

Page 10: Application of UV/Vis Spectroscopy Common Problems: a) Mixtures: - blank samples often contain multiple absorbing species. - the absorbance is the sum.

d) Instrument noise:

noise - short term baseline fluctuations, which decrease the precision of the analysis

‚ can not measure A precisely

‚ the various sources of noise each cause some uncertainty

in the absorbance measurement and can be treated as individual standard deviations.

1) 0%T noise:- noise when light beam is blocked- seldom important

- typically 0.01%T

2) Readout Precision:- especially with a meter- typically 0.5%T 1-3% error in concentration

3) Shot Noise:- occurs when e- transfers across a junction (like the space between cathode & anode in PMT).- causes random fluctuations in current since individual e- arrive at random times- increases with increase current (%T). Especially bad above 95%T.

Page 11: Application of UV/Vis Spectroscopy Common Problems: a) Mixtures: - blank samples often contain multiple absorbing species. - the absorbance is the sum.

4) Flicker Noise:- noise from the lamp due to intensity changes- important at high transmittances.

5) Cell positioning uncertainty:- not really noise, but affects precision- minor imperfections, scratches or dirt change %T- may be the major cause of imprecision

Category Characterized by Typical Sources Likely to be Important in

Case IST = k1

Limited Readout resolution Inexpensive photometers and spectrophotometers having small meter scales

Heat detector Johnson noise IR and near-IR spectrophotometers and photometers

Dark current and amplifier noise Regions where source intensity and detector are low

Case II Photon detector shot noise High-quality UV-visible spectrophotometers

Case IIIST = k3T

Cell positioning uncertainties High-quality UV-visible and IR spectrophotometers

Source flicker Inexpensive photometers and spectrophotometers

22TS k T T

Page 12: Application of UV/Vis Spectroscopy Common Problems: a) Mixtures: - blank samples often contain multiple absorbing species. - the absorbance is the sum.

Taken together, these noise sources indicate that the intermediate absorbance and transmittance ranges should be used.

- at low %T, 0%T, noise and readout precision are important - at high %T, shot and flicker noise are large.

Keep A in range of 0.1 – 1.5 absorbance units (80 -3%T)

Page 13: Application of UV/Vis Spectroscopy Common Problems: a) Mixtures: - blank samples often contain multiple absorbing species. - the absorbance is the sum.

Applications:A) Molar Absorptivities () in UV-Vis Range:

= 8.7 x 1019 PA

P – transition probability (ranges from 0.1 to 1, for likely transitions)A – cross-section area of target molecule (cm2)

- ~10-15 cm2 for typical organics

- max = 104 to 105 L/mol-cm

- < 103 – low intensity (P 0.01)

Name Structure max

but-1-en-3-yne 219 7,600

cyclohex-2-enone 225 10,300

toluene 206 7,000

3,4-dimethylpent-3-en-2-one 246 5,300

O

CH

C CHH2C

CH3

OCH3

H3C

CH3

Page 14: Application of UV/Vis Spectroscopy Common Problems: a) Mixtures: - blank samples often contain multiple absorbing species. - the absorbance is the sum.

- For Compounds with Multiple Chromophores:‚ If greater then one single bond apart

- are additive- constant

CH3CH2CH2CH=CH2 max= 184 max = ~10,000

CH2=CHCH2CH2CH=CH2 max=185 max = ~20,000

‚ If conjugated- shifts to higher ’s (red shift)

H2C=CHCH=CH2 max=217 max = ~21,000

Page 15: Application of UV/Vis Spectroscopy Common Problems: a) Mixtures: - blank samples often contain multiple absorbing species. - the absorbance is the sum.

Example 6: The equilibrium constant for the conjugate acid-base pair

HIn + H2O H3O+ + In-

Calculate the absorbance at 430 nm for an indicator concentration of 3.00x10 -4 M

K = 8.00x10-5

= 8.04x103 = 0.755x103

Page 16: Application of UV/Vis Spectroscopy Common Problems: a) Mixtures: - blank samples often contain multiple absorbing species. - the absorbance is the sum.

Applications:

B) Absorbing Species in UV/Vis:

1) Electronic transitions involving organic compounds, inorganic compounds, complexes, etc.

Basic process:

M + h M*

10-8 – 10-9s

M* M + heat (or fluorescence, light, or phosphorescence)

or

10-8 – 10-9s

M* N (new species, photochemical reaction)

Note: excited state (M*) is generally short and heat produced not generally measurable. Thus, get minimal disturbance of systems (assuming no photochemical reaction)

Page 17: Application of UV/Vis Spectroscopy Common Problems: a) Mixtures: - blank samples often contain multiple absorbing species. - the absorbance is the sum.

2) Absorption occurs with bonding electrons.

- E() required differs with type of bonding electron.

- UV-Vis absorption gives some information on bonding electrons (functional groups in a compound.

- Most organic spectra are complex

‚ electronic and vibration transitions superimposed

‚ absorption bands usually broad

‚ detailed theoretical analysis not possible, but semi-quantitative or qualitative analysis of types of bonds is possible.

‚ effects of solvent & molecular details complicate comparison

Page 18: Application of UV/Vis Spectroscopy Common Problems: a) Mixtures: - blank samples often contain multiple absorbing species. - the absorbance is the sum.

- Single bonds usually too high excitation energy for most instruments (185 nm)

‚ vacuum UV

‚ most compounds of atmosphere absorb in this range, so difficult to work with.

‚ usually concerned with functional groups with relatively low excitation energies (190 850 nm).

- Types of electron transitions:

i) , , n electrons

Sigma () – single bond electron

Low energy bonding orbital High energy anti-bonding orbital

Page 19: Application of UV/Vis Spectroscopy Common Problems: a) Mixtures: - blank samples often contain multiple absorbing species. - the absorbance is the sum.

Pi () – double bond electron

Low energy bonding orbital High energy anti-bonding orbital

Non-bonding electrons (n): don’t take part in any bonds, neutral energy level.

Example: Formaldehyde

Page 20: Application of UV/Vis Spectroscopy Common Problems: a) Mixtures: - blank samples often contain multiple absorbing species. - the absorbance is the sum.

‚ * transition in vacuum UV‚ n * saturated compounds with non-bonding electrons

~ 150-250 nm ~ 100-3000 ( not strong)

‚ n *, * requires unsaturated functional groups (eq. double bonds)most commonly used, energy good range for UV/Vis ~ 200 - 700 nm n * : ~ 10-100 *: ~ 1000 – 10,000

Page 21: Application of UV/Vis Spectroscopy Common Problems: a) Mixtures: - blank samples often contain multiple absorbing species. - the absorbance is the sum.

Chromophore Example Solvent max (nm) max Type of transition

Alkene n-Heptane 177 13,000 *

Alkyne n-Heptane 178

196

225

10,000

2,000

160

*

_

_

Carbonyl n-Hexane

n-Hexane

186

280

180

293

1,000

16

Large

12

n*

n*

n*

n*

Carboxyl Ethanol 204 41 n*

Amido Water 214 60 n*

Azo Ethanol 339 5 n*

Nitro CH3NO2Isooctane 280 22 n*

Nitroso C4H9NO Ethyl ether 300

665

100

20

_

n*

Nitrate C2H5ONO2Dioxane 270 12 n*

C6H13HC CH2

C5H11C C CH3

CH3CCH3

O

CH3CH

O

CH3COH

O

CH3CNH2

O

H3CN NCH3

Absorption Characteristics of Some Common Chromophores

Page 22: Application of UV/Vis Spectroscopy Common Problems: a) Mixtures: - blank samples often contain multiple absorbing species. - the absorbance is the sum.

Other Examples of Some Common Chromophores

Page 23: Application of UV/Vis Spectroscopy Common Problems: a) Mixtures: - blank samples often contain multiple absorbing species. - the absorbance is the sum.

ii) d/f electrons (transition metal ions)‚ Lanthanide and actinide series

- electronic transition of 4f & 5f electrons - generally sharp, well-defined bands not affected by associated ligands

‚ 1st and 2nd transition metal series- electronic transition of 3d & 4d electrons- broad peaks

Crystal-Field Theory

- In absence of external field d-orbitals are identical - Energies of d-orbitals in solution are not identical- Absorption involves e- transition between d- orbitals- In complex, all orbitals increase in energy where orbitals along bonding axis are destabilized

Page 24: Application of UV/Vis Spectroscopy Common Problems: a) Mixtures: - blank samples often contain multiple absorbing species. - the absorbance is the sum.

Magnitude of depends on:- charge on metal ion- position in periodic table- ligand field strength :

I- < Br- < Cl- < F- < OH- < C2O42- ~ H2O < SCN- < NH3

< ethylenediamine < o-phenanthroline < NO2- < CN-

increases with increasing field strength, so wavelength decreases

Page 25: Application of UV/Vis Spectroscopy Common Problems: a) Mixtures: - blank samples often contain multiple absorbing species. - the absorbance is the sum.

iii) Charge Transfer Complexes‚ Important analytically because of large (> 10,000)‚ Absorption of radiation involves transfer of e- from the donor to orbital

associated with acceptor- excited state is product of pseudo oxidation/reduction process

‚ Many inorganic complexes of electron donor (usually organic) & electron acceptor (usually metal)

- examples: Iron III thiocyanate

Iron II phenanthroline

(deep red color)(colorless)

Page 26: Application of UV/Vis Spectroscopy Common Problems: a) Mixtures: - blank samples often contain multiple absorbing species. - the absorbance is the sum.

C) Qualitative Analysis:

1) Limited since few resolved peaks- unambiguous identification not usually possible.

2) Solvent can affect position and shape of curve.- polar solvents broaden out peaks, eliminates fine structure.

Loss of fine structure for acetaldehyde when transfer to solvent from gas phase

Also need to consider absorbance of solvent.

Page 27: Application of UV/Vis Spectroscopy Common Problems: a) Mixtures: - blank samples often contain multiple absorbing species. - the absorbance is the sum.

Loss of fine structure for 1,2,4,5-tetrazine as solvent polarity increases

(a) Vapor

(b) Hexane solution

(c) Aqueous

2) Solvent can affect position and shape of curve.- polar solvents broaden out peaks, eliminates fine structure.

Page 28: Application of UV/Vis Spectroscopy Common Problems: a) Mixtures: - blank samples often contain multiple absorbing species. - the absorbance is the sum.

3) Solvent can also absorb in UV-vis spectrum.

Page 29: Application of UV/Vis Spectroscopy Common Problems: a) Mixtures: - blank samples often contain multiple absorbing species. - the absorbance is the sum.

3) Can obtain some functional group information for certain types of compounds..- weak band at 280-290 nm that is shifted to shorter ’s with an increase in polarity (solvent) implies a carbonyl group.

‚acetone:in hexane, max = 279 nm ( = 15)

in water, max = 264.5 nm

- solvent effects due to stabilization or destabilization of ground or excited states, changing the energy gap.

‚ since most transitions result in an excited state that is more polar than the ground state- 260 nm with some fine structure implies an aromatic ring.

Benzene in heptane

More complex ring systems shift to higher ’s (red shift) similar to conjugated alkenes

Page 30: Application of UV/Vis Spectroscopy Common Problems: a) Mixtures: - blank samples often contain multiple absorbing species. - the absorbance is the sum.

UV Spectral Nomenclature

Page 31: Application of UV/Vis Spectroscopy Common Problems: a) Mixtures: - blank samples often contain multiple absorbing species. - the absorbance is the sum.

C) Quantitative Analysis (Beer’s Law):1) Widely used for Quantitative Analysis Characterization

- wide range of applications (organic & inorganic)- limit of detection 10-4 to 10-5 M (10-6 to 10-7M; current)- moderate to high selectivity- typical accuracy of 1-3% ( can be ~0.1%)- easy to perform, cheap

2) Strategiesa) absorbing species- detect both organic and inorganic compounds containing any of these species (all the previous examples)

Page 32: Application of UV/Vis Spectroscopy Common Problems: a) Mixtures: - blank samples often contain multiple absorbing species. - the absorbance is the sum.

b) non- absorbing species- react with reagent that forms colored product- can also use for absorbing species to lower limit of detection- items to consider:

, pH, temperature, ionic strength- prepare standard curve (match standards and samples as much as possible)

Complex(red)

As Fe3+ continues to bind proteinred color and absorbance increases.

When all the protein is bound to Fe3+,no further increase in absorbance.

Page 33: Application of UV/Vis Spectroscopy Common Problems: a) Mixtures: - blank samples often contain multiple absorbing species. - the absorbance is the sum.

Standard Addition Method (spiking the sample)

- used for analytes in a complex matrix where interferences in the UV/Vis for the analyte will occur: i.e. blood, sediment, human serum, etc..

- Method:(1) Prepare several identical aliquots, Vx, of the unknown sample.

(2) Add a variable volume, Vs, of a standard solution of known concentration, cs, to each unknown aliquot.

Note: This method assumes a linear relationship between instrument response and sample concentration.

Page 34: Application of UV/Vis Spectroscopy Common Problems: a) Mixtures: - blank samples often contain multiple absorbing species. - the absorbance is the sum.

(3) Dilute each solution to an equal volume, Vt.

(4) Make instrumental measurements of each sample to get an instrument response, IR.

Page 35: Application of UV/Vis Spectroscopy Common Problems: a) Mixtures: - blank samples often contain multiple absorbing species. - the absorbance is the sum.

(5) Calculate unknown concentration, cx, from the following equation.

Note: assumes a linear relationship between instrument response and sample concentration.

Where:S = signal or instrument responsek = proportionality constantVs = volume of standard added

cs = concentration of the standard

Vx = volume of the sample aliquot

cx = concentration of the sample

Vt = total volume of diluted solutions

V s

Inst

rum

ent

Res

pon

se (

S) m = y/ x

b = y-intercept

(V s ) 0

t

xx

t

ss

V

ckV

V

ckVS

bmVS s

x

sx mV

bcc

1 s sx2 1 x

S V cC

( S S )V

Page 36: Application of UV/Vis Spectroscopy Common Problems: a) Mixtures: - blank samples often contain multiple absorbing species. - the absorbance is the sum.

c) Analysis of Mixtures- use two different ’s with different ’s

A1 = 1MbcM + 1

NbcN (1)

A2 = 2MbcM + 2

NbcN (2)

Note: need to solve simultaneous equations

Page 37: Application of UV/Vis Spectroscopy Common Problems: a) Mixtures: - blank samples often contain multiple absorbing species. - the absorbance is the sum.

d) Photometric titration- can measure titration with UV-vis spectroscopy.- requires the analyte (A), titrant (T) or titration product (P) absorbs radiation

Page 38: Application of UV/Vis Spectroscopy Common Problems: a) Mixtures: - blank samples often contain multiple absorbing species. - the absorbance is the sum.

Example 7: Given:

Calculate the concentrations of A and B in solutions that yielded an absorbance of 0.439 at 475 nm and 1.025 at 700 nm in a 2.50-cm cell.

Absorbance (1.00 cm cell)

Species 475 nm 700 nm

A (7.50x10-5 M) 0.155 0.755

B (4.25x10-5 M) 0.702 0.091