APA 6905 INSTRUMENTATION FOR RESEARCH IN MOVEMENT SCIENCE
SCHOOL OF HUMAN KINETICSMario Lamontagne PhD
Vicon Calibration Vicon Calibration Vicon Calibration Vicon Calibration
David GrohUniversity of Nevada –
Las VegasDepartment of Kinesiology
APA 6905 INSTRUMENTATION FOR RESEARCH IN MOVEMENT SCIENCE
SCHOOL OF HUMAN KINETICSMario Lamontagne PhD
Purpose of CalibrationPurpose of CalibrationPurpose of CalibrationPurpose of Calibration
To assist in the reconstruction 3D positions of captured markers from 2D camera images
To establish an absolute reference of origin and orientation
APA 6905 INSTRUMENTATION FOR RESEARCH IN MOVEMENT SCIENCE
SCHOOL OF HUMAN KINETICSMario Lamontagne PhD
Earlier Calibration MethodsEarlier Calibration MethodsEarlier Calibration MethodsEarlier Calibration Methods
Direct Linear Transformation (DLT)• Abdel-Aziz YI and Karara HM (1971)
Non-Linear Transformation (NLT)• Dapena J, Harman EA, and Miller JA (1982)
APA 6905 INSTRUMENTATION FOR RESEARCH IN MOVEMENT SCIENCE
SCHOOL OF HUMAN KINETICSMario Lamontagne PhD
Direct Linear TransformationDirect Linear TransformationDirect Linear TransformationDirect Linear Transformation
Two or more 2D views (cameras) of a control object (cube or cage) to solve for camera parameters relating the image space (external) to the object space (internal)
Both reference frame and the units of measurement are defined by the known 3-D point coordinates on the control object (Hinrichs).
Time consuming Costly
APA 6905 INSTRUMENTATION FOR RESEARCH IN MOVEMENT SCIENCE
SCHOOL OF HUMAN KINETICSMario Lamontagne PhD
Non-Linear TransformationNon-Linear TransformationNon-Linear TransformationNon-Linear Transformation
Proposed by Dapena, et al. Also uses control object (pole) Specific point coordinates on the object are not necessary
Internal parameters calculated separately
APA 6905 INSTRUMENTATION FOR RESEARCH IN MOVEMENT SCIENCE
SCHOOL OF HUMAN KINETICSMario Lamontagne PhD
APA 6905 INSTRUMENTATION FOR RESEARCH IN MOVEMENT SCIENCE
SCHOOL OF HUMAN KINETICSMario Lamontagne PhD
DLT Calibration AccuracyDLT Calibration AccuracyDLT Calibration AccuracyDLT Calibration Accuracy
Chen L, et al. (1994)• Five groups of control (known) points (8,12,16,20,24)> Six configurations each
• Results: > accuracy increased as control points increased
> Best accuracy achieved with even distribution
> Accuracy decreased as distance from control region increased
APA 6905 INSTRUMENTATION FOR RESEARCH IN MOVEMENT SCIENCE
SCHOOL OF HUMAN KINETICSMario Lamontagne PhD
APA 6905 INSTRUMENTATION FOR RESEARCH IN MOVEMENT SCIENCE
SCHOOL OF HUMAN KINETICSMario Lamontagne PhD
NLT vs. DLT NLT vs. DLT Calibration Accuracy Calibration Accuracy
NLT vs. DLT NLT vs. DLT Calibration Accuracy Calibration Accuracy
Hinrichs RN and McLean SP (1995) Purpose:
• to compare the accuracy of the NLT vs. DLT methods
• To compare accuracy with and without extrapolation
APA 6905 INSTRUMENTATION FOR RESEARCH IN MOVEMENT SCIENCE
SCHOOL OF HUMAN KINETICSMario Lamontagne PhD
MethodsMethodsMethodsMethods
DLT control object• Cuboid, 27 strings hanging from ceiling
• Two defined control volumes– Entire volume (185cm x 490 cm x 213 cm)
– ~25% of entire volume
• Four groups of control points (16, 24, 40, 60)
APA 6905 INSTRUMENTATION FOR RESEARCH IN MOVEMENT SCIENCE
SCHOOL OF HUMAN KINETICSMario Lamontagne PhD
APA 6905 INSTRUMENTATION FOR RESEARCH IN MOVEMENT SCIENCE
SCHOOL OF HUMAN KINETICSMario Lamontagne PhD
MethodsMethodsMethodsMethods
NLT control volume• Similar dimensions to large DLT volume• Single 223cm pole with six marks
> Eleven vertical positions, four horizontal positions
> Encompassed the control volume
Two cameras Mean absolute error – calculation of the resultant difference between the predicted and surveyed locations of selected points from the DLT control object (Hinrichs)
APA 6905 INSTRUMENTATION FOR RESEARCH IN MOVEMENT SCIENCE
SCHOOL OF HUMAN KINETICSMario Lamontagne PhD
ResultsResultsResultsResults
Smallest resultant errors recorded with standard DLT re-predicting control points
Errors in re-predicting control points increased as control points increased
Errors in predicting non-control points decreased as control points increased
APA 6905 INSTRUMENTATION FOR RESEARCH IN MOVEMENT SCIENCE
SCHOOL OF HUMAN KINETICSMario Lamontagne PhD
APA 6905 INSTRUMENTATION FOR RESEARCH IN MOVEMENT SCIENCE
SCHOOL OF HUMAN KINETICSMario Lamontagne PhD
DynaCal3DynaCal3DynaCal3DynaCal3
Vicon Workstation’s current calibration method
Replaced DC2 in early 2001 Implements reconstruction algorithm Provides System information:
• Positions and orientations of cameras wrt each other
• Focal length of lenses• Lens distortion• Absolute reference origin of lab
APA 6905 INSTRUMENTATION FOR RESEARCH IN MOVEMENT SCIENCE
SCHOOL OF HUMAN KINETICSMario Lamontagne PhD
Calibration ProcedureCalibration ProcedureCalibration ProcedureCalibration Procedure
Dynamic calibration • Wand dance• Determines camera and lens variables
• Processed first
APA 6905 INSTRUMENTATION FOR RESEARCH IN MOVEMENT SCIENCE
SCHOOL OF HUMAN KINETICSMario Lamontagne PhD
Calibration ProcedureCalibration ProcedureCalibration ProcedureCalibration Procedure
Static calibration • L-frame• Determines origin of lab space (absolute reference point), only
• Depends on successful dynamic calibration
• Processed last
APA 6905 INSTRUMENTATION FOR RESEARCH IN MOVEMENT SCIENCE
SCHOOL OF HUMAN KINETICSMario Lamontagne PhD
DefinitionsDefinitionsDefinitionsDefinitions
Residuals – numbers that indicate the quality of calibration
Linearization – the process of correcting for lens distortion
Bundle Adjustment – an algorithm that performs the linearization process by optimizing the camera parameters to give the best re-projection error
APA 6905 INSTRUMENTATION FOR RESEARCH IN MOVEMENT SCIENCE
SCHOOL OF HUMAN KINETICSMario Lamontagne PhD
Dynamic CalibrationDynamic CalibrationDynamic CalibrationDynamic Calibration
Performed first Algorithm presumptions
• Known distance between wand markers
• No additional markers moving within the Volume during calibration
• 2D marker images seen by different cameras originate from the same 3D positions
APA 6905 INSTRUMENTATION FOR RESEARCH IN MOVEMENT SCIENCE
SCHOOL OF HUMAN KINETICSMario Lamontagne PhD
Dynamic CalibrationDynamic CalibrationDynamic CalibrationDynamic Calibration
Algorithm first looks at camera pair with the greatest overlap.
Accumulation of additional cameras initiate optimization algorithm to determine best fit.
Bundle Adjustment performs linearization process to correct for lens distortion.
APA 6905 INSTRUMENTATION FOR RESEARCH IN MOVEMENT SCIENCE
SCHOOL OF HUMAN KINETICSMario Lamontagne PhD
Dynamic CalibrationDynamic CalibrationDynamic CalibrationDynamic Calibration
Dynamic calibration algorithm uses known distance between markers to establish the scale of the volume.
This distance constant is specified in the Calibration Reference Object (CRO) file.
APA 6905 INSTRUMENTATION FOR RESEARCH IN MOVEMENT SCIENCE
SCHOOL OF HUMAN KINETICSMario Lamontagne PhD
Dynamic CalibrationDynamic CalibrationDynamic CalibrationDynamic Calibration
Wand wave (“wand dance”) tips:• Avoid breaking cameras’ views with one’s body.
• Position body so the wand can be seen by as many cameras as possible.
• Move around the entire volume to give all cameras an equal opportunity to see the wand wave.
• No need to collect more than 1000 samples
APA 6905 INSTRUMENTATION FOR RESEARCH IN MOVEMENT SCIENCE
SCHOOL OF HUMAN KINETICSMario Lamontagne PhD
Static CalibrationStatic CalibrationStatic CalibrationStatic Calibration
Establishes absolute spatial reference (origin)
Utilizes L-frame with two axes• Tripleton – three markers• Singleton – one marker
APA 6905 INSTRUMENTATION FOR RESEARCH IN MOVEMENT SCIENCE
SCHOOL OF HUMAN KINETICSMario Lamontagne PhD
Static CalibrationStatic CalibrationStatic CalibrationStatic Calibration
Tripleton• Establishes direction of first (X) axis
Singleton• Establishes position of origin• Combines with the first axis to establish the second axis
• Resultant cross product is the third axis
• Origin specified in CRO file
APA 6905 INSTRUMENTATION FOR RESEARCH IN MOVEMENT SCIENCE
SCHOOL OF HUMAN KINETICSMario Lamontagne PhD
ResidualsResidualsResidualsResiduals
Individual cameras• A measure the accuracy of one camera with respect to all other cameras
• Established by reconstructing wand marker positions using all cameras except the one for which the residual is being calculated> Camera in question is non-contributory to its own 3D reconstruction residual
> New with DC3
APA 6905 INSTRUMENTATION FOR RESEARCH IN MOVEMENT SCIENCE
SCHOOL OF HUMAN KINETICSMario Lamontagne PhD
ResidualsResidualsResidualsResiduals
Individual cameras (cont.)• The distance between the reconstructed markers image and the camera’s own image averaged across all available samples
• High camera residual = 2D contribution that tends to be less accurate than the other cameras
APA 6905 INSTRUMENTATION FOR RESEARCH IN MOVEMENT SCIENCE
SCHOOL OF HUMAN KINETICSMario Lamontagne PhD
Residuals (cont.)Residuals (cont.)Residuals (cont.)Residuals (cont.)
Mean Residual – average of individual residuals
Residual Range – the highest and lowest residuals
Wand Visibility – the average percentage of the wand wave that contributed to each camera’s calibration
Static Reproducibility – a measure of the accuracy of the reconstructed positions of the L-frame markers as compared to the CRO file
APA 6905 INSTRUMENTATION FOR RESEARCH IN MOVEMENT SCIENCE
SCHOOL OF HUMAN KINETICSMario Lamontagne PhD
Common Reasons for Failed Common Reasons for Failed CalibrationsCalibrations
Common Reasons for Failed Common Reasons for Failed CalibrationsCalibrations
Inappropriate wand wave• Too fast (low frequency)• Solution: slow wand wave
Poor camera positioning• Inadequate overlap• Solution:
> adjust cameras> May need to customize wand wave
Background static noise• Reposition cameras• Adjust camera sensitivities (can be turned up after calibration
APA 6905 INSTRUMENTATION FOR RESEARCH IN MOVEMENT SCIENCE
SCHOOL OF HUMAN KINETICSMario Lamontagne PhD
ReferencesReferencesReferencesReferences
Dapena J, Harman EA, and Miller JA. (1982) Three-dimensional cinematography with control object of unknown shape. J. Biomechanics 15:11-19
Chen L, Armstrong CW, and Raftopoulos DD. (1994) An investigation on the accuracy of three-dimensional space reconstruction using the direct linear transformation technique. J. Biomechanics 27:493-500
APA 6905 INSTRUMENTATION FOR RESEARCH IN MOVEMENT SCIENCE
SCHOOL OF HUMAN KINETICSMario Lamontagne PhD
ReferencesReferencesReferencesReferences
Hinrichs RN, McLean SP. (1995) NLT and extrapolated DLT: 3-D cinematography alternatives for enlarging the volume of calibration. J. Biomechanics 28:1219-1223