APA 6000 Manual

100
Catalog Number 62005-18 APA 6000™ Process Analyzer Installation and Maintenance Manual © Hach Company, 2001, 2002, 2004. All rights reserved. Printed in the U.S.A. ce/dk 04 February 2005 Edition 4

Transcript of APA 6000 Manual

Catalog Number 62005-18

APA 6000 Process AnalyzerInstallation and Maintenance Manual

Hach Company, 2001, 2002, 2004. All rights reserved. Printed in the U.S.A. ce/dk 04 February 2005 Edition 4

Table of ContentsSafety Precautions .................................................................................................................................................... 4 Specifications ............................................................................................................................................................ 5

Section 1 Introduction ........................................................................................................................................ 71.1 How to Use the Manuals ..................................................................................................................................... 7 1.2 How the Analyzer Works ..................................................................................................................................... 7 1.2.1 Components .............................................................................................................................................. 8 1.2.2 Flow Through the Instrument..................................................................................................................... 8

Installation ........................................................................................................................................................... 11 Section 2 Installation......................................................................................................................................... 13 2.1 Selecting a Location for the APA 6000 Analyzer ............................................................................................... 13 2.2 Mounting ........................................................................................................................................................... 13 2.2.1 Bench Mount ........................................................................................................................................... 14 2.2.2 Wall Mount............................................................................................................................................... 14 2.3 Electrical Connections ...................................................................................................................................... 15 2.3.1 Power Connections at the Analyzer......................................................................................................... 16 2.3.2 Alarm Relay Connections ........................................................................................................................ 20 2.3.3 Recorder Output Connections ................................................................................................................. 21 2.4 Plumbing Installation ......................................................................................................................................... 22 2.4.1 Sample Line Installation .......................................................................................................................... 23 2.4.1.1 Sample Conditioning................................................................................................................... 24 2.4.2 Drain Line Connection ............................................................................................................................. 25 2.4.2.1 Waste Disposal ........................................................................................................................... 26 2.5 Air Purge ........................................................................................................................................................... 26 Startup .................................................................................................................................................................. 27 Section 3 Startup ............................................................................................................................................... 29 3.1 Introduction ....................................................................................................................................................... 29 3.2 Preparing Sample Conditioning and Starting Sample Flow .............................................................................. 29 3.3 Installing Reagents and Standards ................................................................................................................... 29 3.3.1 Supplying Power ...................................................................................................................................... 30 3.3.2 Priming the Analyzer ............................................................................................................................... 31 Section 4 Instrument Menus ........................................................................................................................... 33 4.1 The AquaTrend Interface................................................................................................................................... 33 4.1.1 The AquaTrend Interface Keypad ............................................................................................................ 33 4.1.2 The AquaTrend Interface Display ............................................................................................................ 34 4.2 Menu Functions................................................................................................................................................. 37 4.2.1 Alarm Log ................................................................................................................................................ 37 4.2.2 AquaTrend Menu ..................................................................................................................................... 37 4.2.2.1 Display Setup.............................................................................................................................. 37 4.2.2.2 Language .................................................................................................................................... 40 4.2.2.3 Edit Name ................................................................................................................................... 40 4.2.2.4 Set Time/Date ............................................................................................................................. 41 4.2.2.5 Security Setup ............................................................................................................................ 42 4.2.3 Sensor Menu ........................................................................................................................................... 44 4.2.3.1 Calibration................................................................................................................................... 44 4.2.3.2 Alarm Setup ................................................................................................................................ 47 4.2.3.3 Recorder Setup........................................................................................................................... 48 4.2.3.4 Measurement Options ................................................................................................................ 50Page 2 Table of Contents

6200518TOC.fm

Table of Contents4.2.3.5 Enabling Solution Level Monitoring ............................................................................................ 51 4.2.3.6 Setting Solution Levels ............................................................................................................... 51 4.2.3.7 Solutions & Cleaning Options ..................................................................................................... 52 4.2.3.8 Sensor Diagnostics ..................................................................................................................... 52 4.3 Other Device Menu and Network Menu ............................................................................................................ 55 4.4 Data Display...................................................................................................................................................... 55 4.4.1 Numerical Display.................................................................................................................................... 55

Maintenance ........................................................................................................................................................ 57 Section 5 Maintenance and Troubleshooting ............................................................................................. 59 5.1 Maintenance ..................................................................................................................................................... 59 5.1.1 Shutting the Analyzer Down (Off or Standby).......................................................................................... 59 5.1.2 Scheduled Maintenance .......................................................................................................................... 59 5.1.2.1 Cleaning the Instrument.............................................................................................................. 60 5.1.2.2 Sample Conditioning................................................................................................................... 60 5.1.2.3 Changing Modules ...................................................................................................................... 62 5.1.2.4 Changing Tubing ........................................................................................................................ 63 5.1.2.5 Rotary Valve Module .................................................................................................................. 65 5.1.2.6 Maintaining the Autoburette Module ........................................................................................... 68 5.1.2.7 Greasing the Lead Screw in the Autoburette Module ................................................................. 70 5.1.3 Unscheduled Maintenance ...................................................................................................................... 71 5.1.3.1 Fuse Replacement...................................................................................................................... 71 5.1.3.2 Cleaning the Mixer Module Chamber ......................................................................................... 74 5.2 Troubleshooting................................................................................................................................................. 75 5.2.1 General Troubleshooting Procedure ........................................................................................................ 75 5.2.1.1 Check Alarm Log ........................................................................................................................ 75 5.2.1.2 Check Solutions .......................................................................................................................... 75 5.2.2 Examine the Rotary Valve ....................................................................................................................... 76 5.2.3 Examine the Autoburette ......................................................................................................................... 76 5.2.4 Examine the Tubing and Fittings ............................................................................................................. 76 5.2.5 Back Pressure Gauge ............................................................................................................................. 76 5.3 Specific Alarm/Warning Messages ................................................................................................................... 77 Appendix A Network Connections....................................................................................................................... 81Network Connections in the APA 6000 Analyzer.......................................................................................... 81

Appendix B Printer and Computer Formats ........................................................................................................ 85APA 6000 Printer and Computer Formats........................................................................................................ 85

Appendix C Second Sample Options .................................................................................................................. 89Sample Sequencing ......................................................................................................................................... 89 Quality Control.................................................................................................................................................. 89 Grab Sample .................................................................................................................................................... 89 Setting Up the Sample Sequence .................................................................................................................... 90

General Information........................................................................................................................................... 93Replacement Parts and Accessories ...................................................................................................................... 94 How to Order........................................................................................................................................................... 96 Repair Service ........................................................................................................................................................ 97 Limited Warranty ..................................................................................................................................................... 98 Certification ............................................................................................................................................................. 99

6200518TOC.fm

Page 3 Table of Contents

Safety Precautions

Section 2.3 Electrical Connections on page 15 Section 2.3.1 Power Connections at the Analyzer on page 16 Section 2.3.2 Alarm Relay Connections on page 20 Section 2.3.3 Recorder Output Connections on page 21 Section 2.4 Plumbing Installation on page 22 Section 3.3 Installing Reagents and Standards on page 29 Section 5.1.2.3 Changing Modules on page 62 Section 5.1.2.4 Changing Tubing on page 63 Section 5.1.2.6 Maintaining the Autoburette Module on page 68 Section 5.1.3.1 Fuse Replacement on page 71

Page 4 Safety Precautions

6200518 Safety.fm

SpecificationsSpecifications are subject to change without notice. Specifications for range, accuracy, repeatability, response time, cycle time, detection limit, and other parameter specific items are listed in the parameter-specific manual.

PerformanceRecorder Outputs: Two isolated, 420 mA outputs standard. Output span programmable over any portion of the range (500 V isolation from earth ground). Alarms: 14 programmable internal alarms Relays: Two SPDT relays with a maximum contact rating of 5 A resistive load at 240 V ac. Additional relays available through optional Signal Output Module. Power Requirements: 95240 V ac, 50 or 60 2 Hz, 150 VA Installation Category: II Sample Inputs: Two sample streams (with optional hardware) Maximum Operating Temperature Range: 5 to 50 C (41 to 122 F) Relative Humidity: 5 to 95% relative humidity, non-condensing Network Interface: LonWorks FTT-10 twisted-pair free topology Maximum Node-to-node Distance: 400 meters (1320 feet)

FeaturesInstrument Enclosure: NEMA 4X (Indoor) and IEC 529 (IP 66) with provision for air purge Instrument Installation Category: II Reagent Enclosure: Fully enclosed latch cases with protected ducting for instrument drain; additional plumbing connections for instrument case and reagent enclosure overflow drains.

Sample RequirementsSample Filtration: 0.5 m or smaller. Sample Pressure: 0.5 to 30 psig (0.035 to 2 bar) maximum at sample inlet block Sample Temperature: 5 to 50 C (41 to 122 F) Sample Flow: 1000 to 2000 mL/min.

6200518 Specifications.fm

Page 5 Specifications

Visit http: //www.hach.com

Section 1

Introduction

1.1 How to Use the ManualsThis instrument requires two manuals for complete operating instructions; the APA 6000 Installation and Maintenance Manual and a parameter-specific operation manual. This APA 6000 Installation and Maintenance Manual (Cat. No. 62005-18) provides instructions for instrument installation, menu setup, operation, maintenance, and troubleshooting. The parameter-specific manual contains information related to parameter specific instrument components and software functions. Descriptions of reagents and standards and their consumption rates, method of analysis, operational sequence, and bench comparison testing are all addressed in the parameter-specific manual.

1.2 How the Analyzer WorksThe APA 6000 Process Analyzer is a microprocessor-controlled process analyzer designed to continuously monitor a specific substance in a sample stream (see Figure 1). The analyzer consists of five basic components: Figure 1 Module Components AquaTrend Interface Autoburette Module Detector Module and Holding Coil Valve Module Mixer Module

6200518 Introduction.fm

Page 7 Introduction

Section 11.2.1 ComponentsThe AquaTrend Interface The APA 6000 Process Analyzer has an AquaTrend Interface built into the instrument door. It provides a numerical or graphical display and a keypad for user input. The AquaTrend Interface displays, graphs, and logs sensor readings, and is used for operator control and sensor setup. Autoburette Module The analyzer uses a positive-displacement autoburette to measure fluids through the various analytical components in the instrument. The autoburette module consists of a clear acrylic block which houses a ceramic piston, a double seal for the piston, a linear actuator, a two-way solenoid valve, and control circuitry. Detector Module Samples are measured in a temperature controlled flow cell detector. All detector modules contain a holding coil. The holding coil acts as a storage area for fluids as they are drawn into the instrument before being directed to the appropriate module for mixing or sensor readings. The detector module for some instruments also contains a reaction coil. The reaction coil optimizes the dispersion utilized in some analysis methods. Valve Module The valve module directs fluids to various components in the instrument. Mixer Module The mixer module mixes reagents with the sample or standards. The mixing action also removes excess air from the sample.

1.2.2 Flow Through the InstrumentThe following is a simplified version of how fluids are cycled through the analyzer: 1. In a normal measuring cycle, the instrument uses the autoburette module to draw sample in through the sample conditioning block into the holding coil. 2. The pump then draws reagent(s) into the holding coil. 3. Next, the pump moves the reagent(s) and sample from the holding coil into the mixing module with the sample. 4. When the sample and reagent are adequately mixed, the instrument draws the mixture into the holding coil. 5. The next pump actuation moves the mixture from the holding coil into the reaction coil (not used in all chemistries). As the mixture flows past the reaction coil and through the detector, a measurement is taken and the fluid continues to flow into the drain. The cycle then starts over again at Step 1.

Page 8 How the Analyzer Works

6200518 Introduction.fm

Section 1Step One The burette draws sample into the Holding Coil. 10-Port Rotary Valve Reaction Coil

DetectorHolding Coil

Mixer Drain

Filtered Sample IN Reagent IN 10-Port Rotary Valve Mixing Chamber Reaction Coil Waste

Step Two The burette draws reagent into the Holding Coil.

DetectorHolding Coil

Mixer Drain

Filtered Sample IN Reagent IN Step Three The burette moves sample and reagent into the Mixing Chamber. 10-Port Rotary Valve Mixing Chamber Reaction Coil Waste

DetectorHolding Coil

Mixer Drain

Filtered Sample IN Reagent IN Step Four The burette draws the sample/reagent mixture into the Holding Coil. 10-Port Rotary Valve Mixing Chamber Reaction Coil Waste

DetectorHolding Coil

Mixer Drain

Filtered Sample IN Reagent IN Step Five The burette moves the sample/reagent mixture through the Reaction Coil and the Detector and measures the response of the Detector. Mixing Chamber Waste

10-Port Rotary Valve

Reaction Coil

DetectorHolding Coil

Mixer Drain

Filtered Sample IN Reagent IN Mixing Chamber Waste

6200518 Introduction.fm

Page 9 How the Analyzer Works

Visit http://www.hach.com

Installation

DANGERSome of the following manual sections contain information in the form of warnings, cautions and notes that require special attention. Read and follow these instructions carefully to avoid personal injury and damage to the instrument. Only personnel qualified to do so, should conduct the installation/maintenance tasks described in this portion of the manual.

DANGERCertains des chapitres suivants de ce mode demploi contiennent des informations sous la forme davertissements, messages de prudence et notes qui demandent une attention particulire. Lire et suivre ces instructions attentivement pour viter les risques de blessures des personnes et de dtrioration de lappareil. Les tches dinstallation et dentretien dcrites dans cette partie du mode demploi doivent tre seulement effectues par le personnel qualifi pour le faire.

PELIGROAlgunos de los captulos del manual que presentamos contienen informacin muy importante en forma de alertas, notas y precauciones a tomar. Lea y siga cuidadosamente estas instrucciones a fin de evitar accidentes personales y daos al instrumento. Las tareas de instalacin y mantenimiento descritas en la presente seccin debern ser efectuadas nicamente por personas debidamente cualificadas.

GEFAHREinige der folgenden Abschnitte dieses Handbuchs enthalten Informationen in Form von Warnungen, Vorsichtsmanahmen oder Anmerkungen, die besonders beachtet werden mssen. Lesen und befolgen Sie diese Instruktionen aufmerksam, um Verletzungen von Personen oder Schden am Gert zu vermeiden. In diesem Abschnitt beschriebene Installations- und Wartungsaufgaben drfen nur von qualifiziertem Personal durchgefhrt werden.

PERICOLOAlcune parti di questo manuale contengono informazioni sotto forma davvertimenti, di precauzioni e di osservazioni le quali richiedono una particolare attenzione. La preghiamo di leggere attentivamente e di rispettare quelle istruzioni per evitare ogni ferita corporale e danneggiamento della macchina. Solo gli operatori qualificati per luso di questa macchina sono autorizzati ad effettuare le operazioni di manutenzione descritte in questa parte del manuale.

6200518 Installation Stopper.fm

Page 11 Installation

Visit http: //www.hach.com

Section 2

Installation

2.1 Selecting a Location for the APA 6000 AnalyzerLocate the analyzer as close to the sampling point as possible. Shorter distances between the process stream and the analyzer mean faster analyzer response to changes in analyte concentration. Long sample lines are also more vulnerable to sediment accumulation. Do not place the sampling point near a point of chemical addition. The process stream at that location may be inadequately mixed, yielding erroneous readings. The instrument enclosure is designed for general-duty, indoor installation. Do not mount the analyzer in direct sunlight. Ambient temperatures may range from 550 C (41122 F), but best performance will result if ambient temperature is between 5 and 35 C (41 and 95 F). Use Figure 2 to determine the space needed to install the instrument. To control the analyzer from a location other than the instrument location, connect a remote AquaTrend Interface. It can be mounted on a bench, pole, or panel as far away as 400 m (1320 feet). Figure 2 Overall Dimensions of the APA 6000 Analyzer

2.2 MountingAn installation kit (Cat. No. 51040-00) is supplied with the instrument. The kit includes: Grab sample kit (Cat. No. 62037-00) Four 0.5 micron filters (Cat. No. 62011-00) 20 ft. of -in. poly tubing (Cat. No. 47438-00)Page 13 Installation

6200518 Installation.fm

Section 22.2.1 Bench MountImportant Note: The analyzer must be bolted to the bench.

Secure the rear of the instrument to the bench or table with two -20 UNC screws through the holes in the metal frame on the back of the instrument (see Figure 3). This will prevent the instrument from tipping while the module panel is open.

2.2.2 Wall MountThe back plate of the instrument has four key-shaped holes for wall mounting. The analyzer can be mounted directly to a wall or to rails capable of holding 157.5 kg (350 lb.). Leave at least 4656 cm (1822 in.) of space above the top of the bolt pattern (see Figure 3) to allow access to the reagent enclosures. 1. Make sure no power is supplied to the analyzer. 2. Install the top two 3/8-16 UNC hex bolts into the wall or fixture according to the dimensions in Figure 3, leaving a -inch gap under each bolt head. 3. Slide the bracket keyholes onto the bolts and lower the frame so the bolts settle in the narrow part of the keyholes. 4. Install two bolts through the lower holes in the frame, into the wall or fixture. 5. Tighten all four bolts securely. Figure 3 Mounting Dimensions

Page 14 Mounting

6200518 Installation.fm

Section 22.3 Electrical ConnectionsDANGER All electrical connections should be made by a qualified technician to assure compliance to all applicable electrical codes. DANGER Tous les branchements lectriques doivent tre effectus par un technicien qualifi pour assurer la conformit toutes les normes lectriques applicables. PERIGO Todas las conexiones elctricas deben ser ubstaladas por personal tcbico cualificado a fin de asegurar su conformidad con las normas elctricas vigentes. GEFAHR Alle elektischen Anschlsse mssen van qualifizierten Techniekem durchgefhrt werden, damit die bereinstimmung mit allen elektrischen Anschlussdaten gewrhrleistet ist. PERICOLO Tutte le connessioni elettriche devono essere fatte da un tecnico qualificato.

To access electrical connections in the APA 6000 instrument, remove the two hold-down fasteners in the upper corners of the module panel. Slowly pull the panel forward until it is fully supported by the steel restraining cable (see Figure 4 and Figure 5). Four access holes for electrical connections are provided along the back of the instrument enclosure (see Figure 6). Remove the cover seals by loosening the wing nuts from inside the instrument. This product is designed to meet NEMA 4X (INDOOR) and IEC 529 IP 66 enclosure requirement when the front door is securely latched. Achieving these protection levels, however, depends on the use of proper hardware in the installation of this product. Standard off-the-shelf NEMA-approved conduit hardware is required for proper installation of all electrical connections. For process or industrial applications, the national electrical codes of most countries require that ac service feeds be hard-wired and contained in conduit systems. The APA 6000 Analyzer has been designed to conform to this requirement. Hach recommends conduit for two reasons: 1. It is generally required by most local electrical codes, and 2. use of metal conduit can improve immunity to lightning surges and ac power transients. In hard-wired electrical applications, the power and safety ground service drops for the analyzer should be no longer than 6 meters (20 feet) unless metal conduit is used to shield the ac power wiring. In applications where power cords are allowed by local electrical codes and power surges and transients are not a great concern, an 18-gauge, 3-conductor power cord (including a safety ground wire) can be used, but its length must not exceed 3 meters (10 feet). The power cord or power cable is not supplied with the APA 6000 Analyzer and must be selected to meet applicable electrical codes for the installation site.

6200518 Installation.fm

Page 15 Electrical Connections

Section 2Hach recommends that care be taken with hook-up configurations other than shown in these illustrations. First, be sure your configuration satisfies local electrical codes and second, ensure that your configuration does not cause susceptibility to lightning, power surges, or transients upsets.

2.3.1

Power Connections at the AnalyzerIf power cords are allowed by local electrical code, a 125 V UL/CSA-approved power cord with an approved NEMA-style strain relief and a standard 115 V North American-style plug (Hach Cat. No. 46306-00) or a 230 V VDE-approved power cord with an approved NEMA-style strain relief and a Continental European-style plug (Hach Cat. No. 46308-00) can be ordered. The APA 6000 Analyzer operates on any ac line voltage from 95240 V ac, 50 or 60 Hz. A 12- to 18-gauge power lead supply wire (power cord), with insulation rated at 300 V (600 V peak), is required. The analyzer is designed for hard-wire connection to ensure conformation to local electrical codes.

Note: Be sure the analyzer power switch is turned off. Dry operation may damage the analyzer.

Route the power leads through the access hole on the back of the enclosure. Line voltage connections are wired directly to the terminal strip J10 (see Figure 6 and Figure 7). Make the ground connection to the adjacent ground stud using an appropriate closed loop connector (see Table 1). Table 1 Wiring Color Codes

Connector hot line conductor 01 neutral conductor 02 ground conductor (#8 stud)

North American black white green/green w/yellow stripe

IEC brown blue green with yellow stripe

Page 16 Electrical Connections

6200518 Installation.fm

Section 2Figure 4 Looking Inside the APA 6000 Analyzer, with Hinged Module Panel Down

Figure 5

Looking Inside the APA 6000 Analyzer, Removing the Customer Access Cover

6200518 Installation.fm

Page 17 Electrical Connections

Section 2Figure 6 Looking Inside the APA 6000 Analyzer, Locating and Identifying the Customer Connections

Back View of Instrument

Air Purge

Power Connections Alarm Connections Drain

Network Connections Recorder Connections

Page 18 Electrical Connections

6200518 Installation.fm

Section 2Figure 7 Wiring Options

6200518 Installation.fm

Page 19 Electrical Connections

Section 22.3.2 Alarm Relay ConnectionsDANGER La zone de raccordement des relais est conue pour des connexions soit tout en basse tension (30 V). Un risque dlectrocution existe si des connexions basse et haute tension sont faites dans la zone de raccordement des relais. PELIGRO El rea de conexin de los rels esta diseada para conexiones que sean todas de baja tensin (inferior a 30V o bien todas de alta tensin (superior a 30V). Existe riesgo de descargas si se combinan conexiones de alto y bajo voltaje en el rea de conexin de los rels. GEFAHR Die Ralaisanschlsse sind entweder fr Niederspannung (< 30V) oder fr Hochspannung (>30 V) vorgesehen. Wenn sowohl Nieder-als auch Hochspannung an die Relais angeschlossen wird, kann es zu einem Stromschlag kommen. PERICOLO La zona riservata ai contatti di allarme disegnata per bassi voltaggi ( 30V). Assolutamente non confondersi nell'utilizzo dei bassi e degli alti voltaggi. Note: Limit current to the relay contacts to 5 amps. This can be accomplished with an external switch and a 5-amp fuse or with a switched 5-amp circuit breaker. You must be able to remove power from the relays locally in case of an emergency or for servicing of the product.

DANGER The relay connection area is designed for either all low voltage (30 V) connections. A shock hazard exists if low and high voltage connections occur simultaneously in the relay connection area.

The analyzer contains two alarm relays designed for use with either high voltage (greater than 30 V-RMS and 42.2 V-PEAK or 60 V dc) or low voltage (less than 30 V-RMS and 42.2 V-PEAK, or less than 60 V dc), but not a combination of both high and low voltage. Wire alarm relays to the connectors labeled Alarm A or Alarm B on the back electrical panel (see Figure 8). For terminal strip and strain relief locations see Figure 6. SPDT relay contacts are unpowered and rated to carry up to 5 amperes at 240 V ac with a resistive load. If high-current or inductive-load devices are to be controlled by analyzer alarms, use a heavy-duty slave relay actuated by the alarm relay. Both normally open and normally closed contacts are available. The connector will accept 22- to 12-gauge wire, insulated as appropriate for the voltage. Size the wires according to the load and local electrical code regulations. If more relays are desired, supplement your system with the AquaTrend Signal Output Module (SOM) (Cat. No. 51250-00). Each SOM provides two additional relays.

Page 20 Electrical Connections

6200518 Installation.fm

Section 2Figure 8 Alarm Relay Connections

2.3.3

Recorder Output ConnectionsDANGER Les installations lectriques dans les usines dveloppent des tensions de masse si de mauvaises techniques de mise la terre sont utilises. Pour rduire les risques dlectrocution, prendre des prcautions pour le branchement dquipements raccords la terre des emplacements diffrents. Suivre attentivement les instructions donnes dans ce chapitre. PELIGRO Si no se emplean tcnicas adecuadas de conexin a tierra, los sistemas elctricos de la instalacin pueden generar tensiones de masa. Con el fin de disminuir el riesgo de descarga elctrica, haga la conexin a tierra en lugares diferentes. Respecte las instrucciones que se indican en esta seccin. GEFAHR Spannungsversorgungsanlagen fhren evlt. Erdpotential, wenn eine schlechte Systemerdung vorgenommen wurde. Vosicht ist geboten, wenn das angeschlossene Gert an verschiedenen Orten geerdet ist. Die Anweisungen in diesem Abschnitt sind genau zu befolgen. PERICOLO Assicurarsi che la messa a terra sia attiva, altrimenti esiste la possibilit di tensioni latenti nella carcassa dello strumento.

DANGER Facility power systems may have ground potentials if poor grounding techniques have been used. To reduce the chance of shock hazards, care must be taken when connecting equipment grounded at different locations (potentials). Carefully follow the instructions given in this section.

The analyzer contains two isolated 420 mA analog outputs which can drive a recorder or control a process (see Figure 9). Recommended output impedance is 3.6 to 500 ohms. Route 22-to 12-gauge recorder wires through the access hole below the terminal strip and connect them to the terminalPage 21 Electrical Connections

6200518 Installation.fm

Section 2strip J8, labelled Recorder (see Figure 6). Use shielded, twisted-pair cable to connect the analyzer to the recorder. Tie the shield to earth ground only at the analyzer end to avoid ground loops. To ensure the analyzer meets NEMA 4X and IP 66 enclosure ratings, use only NEMA-style enclosure strain relief fittings. If more analog outputs are desired, you can supplement your system with the AquaTrend Signal Output Module (SOM). Each SOM provides two additional analog outputs. Figure 9 APA to Recorder Connections

2.4

Plumbing InstallationThe nature of tasks described in this section requires individuals to understand the associated dangers. Please read all caution statements in the left-hand column of this section before installing the instrument. Burns, shock, eye injury, fire, and chemical exposure may occur. Hach Company assumes individuals performing these tasks are qualified and aware of proper safety procedures.

CAUTION This instrument is not intended for use with flammable samples or those containing hydrocarbons.

PRUDENCE Cet instrument nest pas conu pour une utilisation avec des chantillons inflammables ou des chantillons contenant des hydrocarbures. CUIDADO Este instrumento no est destinado para uso con muestras inflamables o que contengan hidrocarburos. VORSICHT Dieses Gert darf nicht fr Tests mit brennbaren Proben oder Proben, die Kohlenwasserstoffe enthalten, benutzt werden. ATTENZIONE Questo strumento non utilizzabile con campioni infiammabili o contenenti idrocarburi.

Page 22 Plumbing Installation

6200518 Installation.fm

Section 2DANGER This instrument should be installed by qualified technical personnel to ensure adherence to all applicable electrical and plumbing codes. DANGER Cet appareil doit tre install par du personnel technique qualifi, afin dassurer le respect de toutes les normes applicables dlectricit et de plomberie. PELIGRO Este instrumento debe ser instalado por personal tcnico capacitado para assegurar el cumplimiento con todos los cdigos elctricos y de plomera aplicables. GEFAHR Um zu gewhrleisten, da alle elektrischen und sanitrinstallationstechnischen VDE-Vorschriften und gegebenefalls die Zusatzvorschriften der zustndigen Elektrizitts- und Wasserwerke erfllt werden, darf dieses Gert nur von geschultem Fachpersonal installiert werden. PERICOLO La strumentazione deve essere installata da personale qualificato, sia per le connessioni elettriche sia per quelle idrauliche.

2.4.1 Sample Line InstallationSelecting a good, representative sampling point is important for optimum performance from the analyzer. The sample analyzed by the analyzer must be representative of the condition of the entire system. If a sample is drawn from a location too close to points of chemical additions to the process stream, inadequate mixing or an incomplete reaction can result in erratic readings from the analyzer.Note: Plumbing connections should be made only by qualified personnel to assure conformance to all applicable plumbing codes.

Run sample lines with small diameter tubing to the analyzer to minimize lag time (the time required for sample to flow from sample point to the analyzer). Route the tubing as directly as possible between the sampling point and the analyzer. Long sample lines and low flow rates result in significant lag time between actual process line conditions and the sampling point of the instrument. Install sample line taps into larger process pipes to minimize the chances of ingesting sediment from the pipe line bottom or air bubbles from the top. A tap projecting into the center of the pipe is ideal (see Figure 10).

Figure 10 Sample Line Location in Process Stream

Air (Typical) Sediment (Typical)

Poor

Poor

Good

Best

6200518 Installation.fm

Page 23 Plumbing Installation

Section 2New tubing, valves and other sample conditioning equipment may be contaminated with substances (oils, dust, etc.) which may contribute to low or high readings on initial startup. Flush the sample line with sample for 12 hours before calibration. For faster cleaning, plumb the sample line to a container of 14 liters of a dilute caustic solution such as 1 N (5%) sodium hydroxide solution and cycle it through to clean sample system components. When the solution is depleted, connect the sample line to the sample source and run the analyzer for an hour to purge the sodium hydroxide solution.

2.4.1.1 Sample ConditioningThe analyzer is equipped with a sample conditioning block that includes a 0.5 m filter. The unit is located on the left side of the instrument. See Figure 11. Use -in OD poly tubing from the sample source to the sample conditioning block. The tubing connects to the block through the lower push-in fitting. From the upper push-in fitting of the sample conditioning block, excess sample flows to drain. A vent assembly is required to avoid excess vacuum at the sample conditioning block. The vent should be located above the sample conditioning block. For wastewater applications or other samples containing high solids, an APA 6000 Micro Filter System is recommended (Cat. No. 51339-00, 115 V model or Cat No. 51339-01, 230 V model). A 40-mesh y-strainer (Cat. No. 46618-00) can be used to remove large solids alone or with the Micro Filter System. When testing highly aerated samples, the manufacturer recommends installing a Process Sample Heater such as Cat. No. 48685-00, prior to the sample conditioning unit. Warming the samples removes some of the excess air and stabilizes the analyzer readings.Important Note: Sample conditioning requires regular maintenance for proper operation. Important Note: If the sample conditioning system supplied with the instrument is removed, it must be replaced with an alternative filtration system of 0.5 m or smaller. Failure to provide the required filtration will void the instrument warranty.

Page 24 Plumbing Installation

6200518 Installation.fm

Section 2Figure 11 Location of the Sample Conditioning Block

Sample Conditioning Block

Vent

Unfiltered Sample Out

Drain

Unfiltered Sample In Remove these two screws to take off the Sample Conditioning Block. Filtered sample into instrument.

2.4.2 Drain Line ConnectionNote: It is imperative to mount the drain point below the instrument drain block. Failure to do so may result in improper functioning of the instrument or back flow of waste into the instrument cabinet.

The analyzer is connected to a drain system through a -inch hose barb fitting on a drain block on the right side of the analyzer case (see Figure 12). After analysis, the waste exits the instrument through the manifold at the bottom of the instrument and into the drain block. This block connects to a -inch ID hose and provides a vented vacuum break at this point. A vent is located on the top of the drain block. Also see section 2.4.2.1. Route instrument waste through a -in. ID hose to a PVC pipe assembly, which must be located one foot below the level of the instrument drain block. The catastrophic drain is located in the middle of the tubing manifold and is sized to accommodate a 1-in. ID hose. This drain is used if an upset inside the instrument results in a significant leak. The supplied hose from this drain port must be connected and run to a drain which is one foot below the bottom of the instrument. The reagent enclosures each have an overflow fitting in the bottom of the case. Plumb these to the drain by punching out the knock-out in the fitting, attaching a 3/8-inch ID hose to each fitting and running it to a drain which is one foot lower in elevation than the bottom of the instrument.

6200518 Installation.fm

Page 25 Plumbing Installation

Section 2Figure 12 Drain Line Connection

Drain Block

All drains must be at least one foot below the bottom of the instrument.

Drains from Reagent Enclosures

Drain from Instrument Drain Block (Customer supplied 3/4 inch tubing.) From Catastrophic Drain

2.4.2.1 Waste DisposalThe effluent from the APA analyzer is non-hazardous per federal RCRA regulations and may be safely discharged to a sanitary sewer based solely on federal rules. Other state and local regulations or permitting requirements may apply and it is the responsibility of the end-user to know and apply state or local requirements to their specific applications. Hach Company does not advocate the pouring, dumping, or discharging of any chemicals, hazardous or nonhazardous, to the ground or to septic systems.

2.5 Air PurgeAir purge may be necessary if the analyzer is located in an environment with high humidity and/or caustic vapors. The goal is to maintain a slight positive pressure in the instrument with dry instrument air. The air purge connection is located in the middle of the back of the instrument. To connect an air supply, remove the plug in the quick connect fitting. Connect in. poly tubing by pushing the tubing into the fitting. You should feel two distinct stops as you push. If the tubing is not seated, it will pop out when air pressure is applied. Use only dry instrument air at 15 scfh (7 L/min.)

Page 26 Air Purge

6200518 Installation.fm

Startup

DANGERHandling chemical samples, standards, and reagents can be dangerous. Review the necessary Material Safety Data Sheets and become familiar with all safety procedures before handling any chemicals.

DANGERLa manipulation des chantillons chimiques, talons et ractifs peut tre dangereuse. Lire les Fiches de Donnes de Scurit des Produits (FDSP) et se familiariser avec toutes les procdures de scurit avant de manipuler tous les produits chimiques.

PELIGROLa manipulacin de muestras qumicas, estndares y reactivos puede ser peligrosa. Revise las fichas de seguridad de materiales y familiarcese con los procedimientos de seguridad antes de manipular productos qumicos.

GEFAHRDas Arbeiten mit chemischen Proben, Standards und Reagenzien ist mit Gefahren verbunden. Es wird dem Benutzer dieser Produkte empfohlen, sich vor der Arbeit mit sicheren Verfahrensweisen und dem richtigen Gebrauch der Chemikalien vertraut zu machen und alle entsprechenden Materialsicherheitsdatenbltter aufmerksam zu lesen.

PERICOLOLa manipolazione di campioni, standard e reattivi chimici pu essere pericolosa. La preghiamo di prendere conoscenza delle Schede Techniche necessarie legate alla Sicurezza dei Materiali e di abituarsi con tutte le procedure di sicurezza prima di manipolare ogni prodotto chimico.

6200518 Startup Stopper.fm

Page 27 Startup

Visit http: //www.hach.com

Section 33.1 Introduction

Startup

It is assumed that the analyzer has already been plumbed to the sample source and a drain, as well as wired to a power source and any other electrical wiring associated with 420 mA connections, relay connections, and network connections by the installation technician as outlined in Section 2. This section presents the steps, in sequence, an operator should take to begin operation. This section will: describe how to prepare the sample conditioning describe the steps to install the reagents and standards list the steps required to prime the sample lines and reagents.

3.2 Preparing Sample Conditioning and Starting Sample FlowThe analyzer is supplied with a 0.5 m ceramic filter (already installed in the sample conditioning block) and four replacement filters. 1. Set the sample flow rate through the sample conditioning block in the following range: 10002000 mL/min. 2. Check the flow at the air break of the drain (where the tubing from the sample inlet block meets the drain tee). If desired, measure the flow rate by directing the sample into a large measuring container (flask or cylinder) over a set amount of time. Keep the end of the tube at the same height as the push-in fitting. 3. Ensure that the filter is tightly threaded into the sample conditioning block to prevent unfiltered sample from entering the analyzer.

3.3

Installing Reagents and StandardsThe solution bottles are labeled (i.e., Reagent 1, Standard 1, Cleaning Solution). The tubing into each of these bottles is also labeled with the solution name to prevent reagent line mix-ups. Some analyzer solutions require special accessories and handling. Please see the parameter specific manual for more information. To change solutions, follow these instructions: 1. Use a permanent marker to put a mark on the tubing where it exits the bottle. 2. Remove the bottle vent cap assembly and tubing from the top of the existing solution bottle. 3. Rinse the straw surrounding the tubing with deionized water and shake off any residual water before inserting it into the new solution. This avoids possible contamination. Be sure the tubing extends to the bottom of the straw. If not, adjust it and re-mark the tubing. 4. Insert the straw into the new bottle and screw the vent cap on. Be sure the mark on the tubing is still located where the tubing exits the bottle. This ensures the inner tube has remained at the bottom of the bottle.

6200518 Startup.fm

Page 29 Startup

Section 35. Place the new bottle in the reagent compartment. Hach recommends changing all the solutions at the same time and does not recommend adding fresh reagent to fill a partially full bottle.

3.3.1 Supplying PowerAfter all appropriate reagents and standards are installed and sample conditioning is prepared, the APA 6000 is ready to start up. Turn on the analyzer by setting the power switch to ON (I). The power switch is located at the center top of the bulkhead that contains the instrument modules (see Figure 13). Close the main door of the analyzer and fasten the latches. The main door should remain closed during operation to maintain correct temperature. Figure 13 Location of the Instrument Power Switch

Page 30 Installing Reagents and Standards

6200518 Startup.fm

Section 33.3.2 Priming the AnalyzerNote: Do not allow the instrument to run before installing reagents and standards and providing sample.

After internal checks, the analyzer will begin operation. It is important to prime the analyzer because operating without fluid may damage components. Priming the analyzer ensures that the standards and/or reagents fill the tubing going to the valve module. To prime the analyzer, complete the following steps: 1. Press the MENU key to start from the MAIN MENU. 2. Select SENSOR MENU and press ENTER. 3. Select the sensor by name (default is #1 APA6000) and press ENTER. 4. Select SOLUTIONS & CLEANING and press ENTER. 5. Select PRIMING and press ENTER. 6. Select PRIME ENTIRE SYSTEM NOW. This will prime all the lines from the solution bottles and sample lines. The system prime will take approximately 715 minutes. The analyzer will display various messages about the progress of the cycle. 7. A pop-up window will indicate that priming will begin after the current operation cycle. When priming starts, a pop-up window will indicate the time remaining to complete the prime cycle. 8. When priming is complete, the analyzer will automatically begin measuring sample. The analyzer will require 36 hours for readings to stabilize. During this time, the analyzer will establish temperature control. After readings stabilize, a calibration is required to obtain accurate values. See section 4.2.3.1 on page 44.

6200518 Startup.fm

Page 31 Installing Reagents and Standards

Visit http: //www.hach.com

Section 4

Instrument Menus

4.1 The AquaTrend InterfaceThe AquaTrend Interface integrated into the door of the APA 6000 Analyzer provides the interface necessary to communicate with the analyzer (sensor) and the Hach network, if present. The interface provides a numerical or graphical display and a keypad for user input. A remote AquaTrend can be connected for access to the analyzer from another location. The AquaTrend interface can be used with a number of Hach sensors and therefore, not all of the functions available are used when interfacing with the APA 6000. Functions that are not appropriate for the APA 6000 will not be explained fully in the following text. Data can be viewed in numeric or graphical format and reviewed on trend graphs for details of measurement history. With the addition of the Hach Serial I/O Module (see Replacement Parts and Accessories on page 94), data can be downloaded to a computer or printer.

4.1.1 The AquaTrend Interface KeypadThe AquaTrend Interface keypad is designed for simple navigation through the AquaTrend Interface menus (see Figure 14). The keypad contains four arrow keys, a DISPLAY key, a MENU key, an EXIT key, and an ENTER key. Figure 14 and Table 2 describe how these keys work. Figure 14 AquaTrend Interface Keypad1 2 4DISPLAY EXIT

3

5

MENU

ENTER

Table 2 AquaTrend Interface Key FunctionsNumber 1 Key Function of Keys Allows navigation through the various menus, to change settings, and to move the cursor when observing graph data. Table 3 summarizes the functions of the arrow keys in each menu view. Displays the data display screen. Once the data is displayed, pressing the DISPLAY key toggles between data presented in numerical form and graph form. Displays the Main Menu from any menu level or from the data display. From the Main Menu, press the EXIT key to access the data display screen. From any other menu level, press the EXIT key to display the previous menu screen. The EXIT key can also be used to abandon any alphanumeric entry or selection process. Press the ENTER key while in any menu to select menu options and confirm alphanumeric entries.

2

3

4

5

6200518 Instrument Menus.fm

Page 33 Instrument Menus

Section 4The arrow icons and their positions in the arrow field in the top right corner of the display (item 7 on Figure 16) indicates the active arrow keys. For example, if three arrow icons are present in the arrow field, pointing up, down, and right, then the up, down, and right arrow keys can be used while viewing that screen. Figure 15 shows examples of the screen categories and Table 3 contains summaries of the arrow key functions for each screen category. Figure 15 AquaTrend Interface Screen Categories

1. Menu Screen

2. Numeric Data Display Screen

3. Graph Display Screen

4. Pop-up Entry Screen

Table 3 Arrow Key FunctionsScreen Category Function Use the arrow keys to move the pointer to the next (down) or previous (up) item. For menus no longer than one page, the pointer will loop around from the bottom to the top or from the top to the bottom. For multiple-page menus, the pointer will move to the next (down) or previous (up) page. Left and Right Arrow: Changes the number of channels shown. Press the Right Arrow to increase the number of channels or press the Left Arrow to decrease the number of channels. Up and Down Arrow: Changes the active channel designation. Up Arrow: Displays preceding channel graph. Down Arrow: Displays next channel graph. Left and Right Arrow: Moves curser if it is enabled; if not enabled, changes the graph time base. Up and Down Arrows: Increment (up) and decrement (down) currently highlighted character. Left and Right Arrows: Moves highlight to next (right) or preceding (left) character. Loops around at either end.

1. Menu Screen

2. Numeric Data Display Screen (1 to 8 numeric readings) 3. Graph Display Screen

4. Pop-up Entry Screen

4.1.2 The AquaTrend Interface DisplayAquaTrend Interface screens appear in several different formats (see Figure 15). See Figure 16 and Figure 17 for descriptions of each field. When the Main Menu is displayed on the AquaTrend Interface screen, press the EXIT or DISPLAY key to return to the data display mode. Consecutive presses of the EXIT key from any Menu takes you back, one screen at a time, to the data display screen. Pressing DISPLAY from any menu immediately returns you to the measurement display screen. When changing settings or performing diagnostics on a device on the network, temporarily exit the data display mode and enter the Main Menu by pressing the MENU key. The AquaTrend Interface will display data from the analyzer except when the operator uses it for menu setup. When viewing data, press the DISPLAY key to alternate between displaying the data in numeric and graphical mode. When viewing data in numeric mode, you can display a single measurement or up to sixteen measurement channels at the same time. Pressing the right arrow key causes more measurement channels to be displayed; pressing the left arrow key causes fewer measurement channels to be displayed.Page 34 The AquaTrend Interface

6200518 Instrument Menus.fm

Section 4When viewing data in graph mode, you can see a history of readings over the last 2 hours, 8 hours, 30 hours, 7 days, or 30 days. The AquaTrend Interface, however, displays data for only one measurement channel at a time in the graphic mode. For details on selecting the preferred graph time and scale, see section 4.2.2.1 on page 37. Figure 16 AquaTrend Interface Menu and Numeric Data Display Fields

Table 4 Display FieldsNumber 1 2 3 4 Field Status Bar Text/Data Date Alarms Security Icon 5 6 7 Hourglass Icon Time Navigation Arrow Information Shown in Display Field Numeric Data display: shows DISPLAY DATA, Main menu: shows MAIN MENU Lower menus: shows top level menu (main menu item) Numeric data (116 channels), graphical data, menus, and pop up boxes appear here. Current date Numbers that appear here, 1 through 8, correspond to the sensor with active alarms. If N appears, a Network alarm is active. If no indicators appear in this field, no system alarm is active. See section 5.2.1.1 on page 75 for information on reviewing alarms. When displayed, indicates a security level of Partial or Total When displayed, indicates the datalog is completing a change. The keyboard buffer is cleared when the icon disappears. Current time displayed as military time (24 hour clock). Direction of these arrow icons correspond to active arrow keys for each screen.

6200518 Instrument Menus.fm

Page 35 The AquaTrend Interface

Section 4Figure 17 AquaTrend Interface Graph Fields1 16

2

AquaTrend

15 14 13

3

4

5 12 6 11 7Alarms

8

9

10

Table 5 Graph FieldsNumber Field 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 Status Bar Channel assignment Graph Data Y-axis upper limit Cursor Y-data Units Species Graph data Y-axis lower limit Graph time X-axis lower limit Cursor Time/Date Graph time X-axis upper limit Graph Cursor Graph Area Sensor name Units and chemical form Navigation Arrows Current Data Information Shown in Graph Field Shows REVIEW MEASUREMENTS Channel number of the measurement shown in the graph Upper numeric limit of trend graph data. (Ymax) Y value of data at cursor position Units of the measurement on the displayed channel Chemical form of the measurement on the displayed channel. Lower numeric limit of trend graph data (Y-min) Lower time limit (oldest data). If data is being reviewed in the 24 hour or 30 day mode, this field is replaced with the date. If data is being reviewed in the 24 hour or 30 day mode, this field is replaced with the date.

Time or date value of data at cursor position Current time (latest data).

Retrieves value of data at cursor position Graph of 60 minutes, 8 hours, 24 hours, 7 days, or 30 days of data. These settings are selectable in the DISPLAY SETUP menu (see section 4.2.2.1). Depending on the sensor type and configuration, the display shows measurement name or the sensor location, or a combination of both. Units and chemical form of the measurement on the channel Direction of these arrow icons correspond to enabled arrow keys for each screens. Most recent measurement (same as numeric display)

Page 36 The AquaTrend Interface

6200518 Instrument Menus.fm

Section 44.2 Menu FunctionsThe first level of menu functions is comprised of Alarm Log, Aquatrend Menu, Sensor Menu, Other Device Menu, and Network Menu. Each level of menus will be discussed in detail below.

4.2.1 Alarm LogIf an alarm warning is active, a number will appear at the bottom of the display in the alarms area and if the FLASH ON ALARM option has been enabled, the entire display will flash (to disable the Flash on Alarm see section 4.2.2.1). The alarms may suggest an over- or under-range condition, or a Sensor Warning. Proceed to the Alarm Log to find out why the alarm was triggered: 1. With the cursor at Alarm Log, press ENTER. 2. Select device for review of current alarms. 3. Select either Warning/Alarm Log or Setpoint Alarm Log. The Warning/Alarm Log provides information on the operation of the analyzer. The SET POINT ALARM LOG provides information about sample measurements that exceed the setpoint values.

The most recent alarm/warning is listed first. A bell icon to the left of the warning indicates the alarm/warning is active. Selecting a specific alarm/warning and pressing ENTER provides more detailed information about that alarm/warning (i.e., time, date, value).

4.2.2 AquaTrend MenuThe following options are provided in the AquaTrend Menu: Display Setup: Controls specific screen functions. Language: Changes the language of the displayed information. Edit Name: Allows the user to change the name of the AquaTrend. Set Time/Date: Sets the time and date for the AquaTrend. Security Setup: Controls access to the analyzer functions. Print Data Log: If a com-link is set up on the network, data can be output to a printer or computer. AT Diagnostics: Allows the user to test various AquaTrend functions and clear the AquaTrend alarm log.

4.2.2.1 Display SetupThe following options are controlled in the Display Setup Menu: Measurement To Channel: Allows assignment of the analyzer measurement (using the user-selected chemical form and units) to a channel. Channel Graph Setting: Controls the range displayed in the graph.

6200518 Instrument Menus.fm

Page 37 Menu Functions

Section 4Flash On Alarm: Controls whether or not the display will flash when an alarm has been detected. Contrast Level: Controls the level of contrast in the display.

Measurement to ChannelAPA 6000 Analyzers can measure from 1 or 2 sample streams, requiring assignment of a measurement channel for each parameter on a stream. For example, parameter options could be called 1Hardness and 2Hardness. In addition, each sample stream may report up to five different parameters. To assign a measurement to a channel, use the MEASUREMENT TO CHANNEL function.

Attaching a Measurement to a Channel1. While in the Display Set Up Menu, move the cursor to MEASUREMENT TO CHANNEL and press ENTER. 2. Move the cursor to ATTACH and press ENTER. 3. Move the cursor to the sensor whose measurement is to be attached and press ENTER. 4. Select chemistry measurement and press ENTER. 5. Move the cursor to the channel where the measurement is to be displayed and press ENTER. 6. Move the cursor to ATTACH and press ENTER. The message OPERATION COMPLETE will appear. The screen will return to the display setup menu upon completion of the measurement to channel setup.

Detaching a Measurement From a Channel1. While in the Display Set Up Menu, move the cursor to MEASUREMENT TO CHANNEL and press ENTER. 2. Move the cursor to DETACH and press ENTER. 3. Move the cursor to the sensor to be detached and press ENTER. 4. Move the cursor to the measurement to be detached and press ENTER. The message OPERATION COMPLETE will appear. The screen will return to the display setup menu upon completion of the measurement to channel setup.

Page 38 Menu Functions

6200518 Instrument Menus.fm

Section 4Channel Graph SettingsTo optimize the graphical display, establish the minimum and maximum values to be displayed. For example, if the expected range is 07 ppm, assign 0 as the measurement minimum and 7 as the measurement maximum. 1. Press the MENU key to access the Main Menu. 2. Select AquaTrend Menu and press ENTER. 3. Select DISPLAY SETUP and press ENTER. 4. Select CHANNEL GRAPH SETTINGS from the Display Setup Menu and press ENTER. A list of available channels will appear. 5. Select the channel of interest and press ENTER. 6. Select YMAX and press ENTER. 7. Use the arrow keys to change the displayed values. Press ENTER to accept the value or EXIT to cancel. 8. Repeat step 6 and step 7 to change the YMIN: displayed value. Enter negative numbers by changing the first character to a negative sign with the arrow keys. Setting the Graph Time (x-axis): 1. Press the DISPLAY key until a graph is displayed. 2. Press the up or down arrow keys to select the channel of interest. 3. Press the right or left arrow key to change the time base. The time base will cycle through 2 hours, 8 hours, 30 hours, 7 days, and 30 days. The time scale is shown below the X-axis.

Flash on AlarmWhen an alarm is active, the display can flash to bring attention to the alarm condition. Enable/Disable the Flash on Alarm as follows: 1. Press the MENU key to access the Main Menu. 2. Select AquaTrend Menu and press ENTER. 3. Select Display Setup and press ENTER. 4. Move the pointer to Flash On Alarms. YES or NO will be displayed to the right of that line. If the displayed option is correct, press the EXIT key to return to the previous menu level. If the displayed option is not correct, press ENTER. A pop-up window will appear. Use the arrow keys to move the pointer to the desired option and press the ENTER key to accept it. Press the EXIT key to return to the previous menu level.

6200518 Instrument Menus.fm

Page 39 Menu Functions

Section 4Adjusting the Display ContrastAdjusting the contrast level setting of the AquaTrend Interface allows the operator to choose the most readable display for the lighting conditions. 1. Press MENU to start from the Main Menu. 2. Select AquaTrend Menu and press ENTER. 3. Select Display Setup and press ENTER. 4. Select Contrast Level. 5. A pop-up window provides a choice of settings between 0 and 24 (0 being the lowest setting). Use the up and down arrow keys to adjust the contrast. Press ENTER to accept the displayed contrast level.

4.2.2.2 LanguageEnglish is the default language of the APA 6000 Analyzer. To display another language follow the instructions below: 1. Press the MENU key to start from the Main Menu. 2. Select AquaTrend Menu and press ENTER. 3. Select Language and press ENTER. 4. Use the up or down arrow key to select the appropriate language and press ENTER. All screens will be displayed in this language.

4.2.2.3 Edit NameThe EDIT NAME function may be used to assign a unique name to a device connected to the AquaTrend network. (For example: FAB 3, Boiler_Room, etc.) Refer to the AquaTrend Interface Manual for more information. 1. Press the MENU key to start from the Main Menu. 2. Select AquaTrend Menu and press ENTER. 3. Use the up or down arrow key to select Edit Name and press ENTER. 4. A pop-up window will appear instructing the user to Edit AquaTrend Name. 5. Use the left and right arrow keys to choose the character to change and use the up and down arrow keys to cycle through letters, numbers, and symbols until you reach the character you want.Note: In order to distinguish between devices on the network, use a unique and descriptive name. The name can be up to 12 characters long in any combination of alphanumeric characters. The first character must be a letter or the # sign.

6. Press ENTER to accept the displayed name.

Page 40 Menu Functions

6200518 Instrument Menus.fm

Section 44.2.2.4 Set Time/DateThe current time and date have been factory set on the AquaTrend Interface, but the time may need to be adjusted to your time zone. If it is necessary to reset the time and date, proceed as follows:

Setting the Time1. Press the MENU key to start from the Main Menu. 2. Use the up/down arrow keys to select AquaTrend Menu and press ENTER. 3. Select Set Time/Date and press ENTER. 4. Select Set Time and press ENTER. 5. A pop-up window will prompt for the new time. The numbers representing the hour will be highlighted. Press the up arrow key to increase the hour setting and press the down arrow key to decrease the setting. When the proper hour is displayed, press the right arrow key and set the minutes in the same manner. 6. The AquaTrend Interface uses a 24-hour clock (military time), so there is no selection for a.m. or p.m. If the current time is 4:00 p.m., for example, you would enter the 24-hour format equivalent of 16:00. 7. Press ENTER to save the change or EXIT to cancel.

Setting the DateUse this procedure to change the current date: 1. Press the MENU key to start from the Main Menu. 2. Use the up/down arrow keys to select AquaTrend Menu and press ENTER. 3. Select Set Time/Date and press ENTER. 4. Select Set Date and press ENTER. 5. A pop-up window will prompt for the new date. The numbers representing the month will be highlighted. Press the up arrow key to increase the month setting and press the down arrow key to decrease the setting. When the proper month is displayed, press the right arrow key and set the day. Repeat to set the last two digits of the current year. 6. Press ENTER to save the change or press EXIT to cancel.

6200518 Instrument Menus.fm

Page 41 Menu Functions

Section 4Setting the Correct Day of the WeekUse this procedure to change the current day of the week: 1. Press the MENU key to start from the Main Menu. 2. Select AquaTrend Menu and press ENTER. 3. Select Set Time/Date and press ENTER. 4. Select Day and press ENTER. 5. A pop-up window will prompt for the correct day. Use the up or down arrow key to highlight the desired day. 6. Press ENTER to save the change or press EXIT to cancel. The day will change automatically. Check to be sure the clock is set at the correct time.

Changing the Date FormatUse the following procedure to change the format of the date displayed by the AquaTrend Interface: 1. Press the MENU key to start from the Main Menu. 2. Select AquaTrend Menu and press ENTER. 3. Select Set Time/Date and press ENTER. 4. Select Date Format: and press ENTER. 5. A pop-up window will appear with SELECT DATE FORMAT. Use the up and down arrow keys to select a date format: MM/DD/YY is month/day/year, and DD/MM/YY is day/month/year. 6. Press ENTER to save the change or press EXIT to cancel.

4.2.2.5 Security SetupTo limit access and avoid unauthorized changes to the AquaTrend Interface Network setup features, set security options by: setting a password; setting the security level; or accessing a temporary unlock option. This temporarily suspends the full or partial security levels, allowing access to the system.

The default password is HACH. The default security level is NONE. This security level allows access to all functions. If the password is changed and then is lost or forgotten, contact your Hach representative for further instructions.

Page 42 Menu Functions

6200518 Instrument Menus.fm

Section 4Changing the System Password1. Press MENU to start from the Main Menu. 2. Select AquaTrend Menu and press ENTER. 3. Select Security Setup and press ENTER. 4. Select Set Password and press ENTER. 5. Enter the current password for access to the AquaTrend Interface menus. Pressing the right and left arrow keys moves the cursor right and left; pressing the up and down arrow keys scrolls through the available characters. The default password is HACH. 6. Press ENTER to confirm or press EXIT to cancel. 7. Enter the new password by using the arrow keys as in step 5. 8. Re-enter the new password.

Setting the Security LevelThe analyzers security level determines which menu parameters can be modified by the operator. The security levels are NONE, PARTIAL, and TOTAL. TOTAL lockout means the operator is unable to change any of the analyzers settings or cause the analyzer to perform any operations (priming, cleaning, setting fluid levels, calibrating, etc.) from the keyboard. Any user will be able to review all settings, but will be unable to modify them if they are locked out. PARTIAL lockout allows user access to the following menus and functions: display setup language alarm log execute calibration (see section 4.2.3.1) all priming operations setting all fluid levels clean system now

Setting the security level to NONE allows access, without a password, to any of the menus except SECURITY SETUP.

6200518 Instrument Menus.fm

Page 43 Menu Functions

Section 4Set the security level as follows:1. Press the MENU key to start from the Main Menu. 2. Select AquaTrend Menu and press ENTER. 3. Select Security Setup and press ENTER. 4. Select Security Level: and press ENTER. 5. Enter the password (default is HACH). Use the left and right arrow keys to choose the character to change and use the up and down arrow keys to cycle through letters, numbers and symbols until you reach the character you want. 6. Use the arrow keys to select NONE, PARTIAL, or TOTAL. 7. Press ENTER to confirm or press EXIT to cancel.

Using Temporary Unlock and Unlock DurationSelecting temporary unlock allows the operator to temporarily override the security level setting and have full access to all menu parameters. Unlock duration allows the operator to select the length of time that the temporary unlock is in effect. 1. Press the MENU key to start from the Main Menu. 2. Select AquaTrend Menu and press ENTER. 3. Select Security Setup and press ENTER. 4. Select Temporary Unlock and press ENTER. 5. Enter the password (this is necessary only if the security level is set to Partial or Total). 6. Choose the duration of time you want the keypad to remain unlocked: 30, 60, 120, or 360 minutes. 7. Press ENTER to confirm or press EXIT to cancel.

4.2.3 Sensor Menu4.2.3.1 CalibrationCalibration prepares the APA analyzer to calculate concentrations in unknown samples based on a calibration generated using two known standards. These two standards are the high and low standard solutions. For immediate trend analysis, an electronic factory default calibration is already established and allows the analyzer to provide approximate measurements.The once-a-day (at 4:00 a.m.)* auto-calibration interval may be changed to every 12 hours, a specific day of the week, or may be turned off. It can also be calibrated on demand as necessary.

* APA6000 Alkalinity and HR Hardness instruments defaults are each Monday at 4:00 a.m. APA6000 HR Hardness instrument default is each Monday at 4:00 a.m. Page 44 Menu Functions

6200518 Instrument Menus.fm

Section 4Reducing the frequency of the auto-calibration will reduce standard usage, but it may also reduce measurement accuracy. Hach recommends that the analyzer be calibrated when any reagent or standard is changed or as often as necessary to achieve acceptable results. Calibration should only be performed after the analyzer has been turned on for more than five hours. This allows the temperature inside the analyzer to stabilize.

Setting Up the Auto-Calibration ScheduleThe analyzer accuracy specifications are based on the factory-set default interval. The APA 6000 Analyzer can calibrate at any time. To change the calibration time, use the auto-cal setup procedure: 1. Press the MENU key to start from the Main Menu. 2. Select Sensor Menu and press ENTER. 3. Select the sensor by name and press ENTER. 4. Select Calibration and press ENTER. 5. Select Auto-Cal Setup and press ENTER. 6. The analyzer calibration is a factory set default interval. In order to change the settings, select one of the following: Select AUTO-CAL to turn the automatic calibration option off, or to set the day or hourly interval you want the instrument to calibrate. Choose any day of the week, every day of the week, or every 12 hours (as shown below). Select Auto-Cal Time to specify the clock hour you want the instrument to calibrate (use a 24-hour clock). Use the up and down arrow keys to enter an hour between 1 and 24. For example, if you want the instrument to calibrate at 2:00 p.m., enter 14.

Note: When auto-calibrate is set to 12 hours, the analyzer calibrates at the set auto-cal time and again 12 hours later.

7. Press ENTER to confirm or press EXIT to cancel.

Execute CalibrationTo start a calibration manually, proceed as follows. The instrument will finish its current cycle, then perform the calibration. 1. Press the MENU key to start from the Main Menu. 2. Select Sensor Menu and press ENTER. 3. Select the sensor by name and press ENTER. 4. Select Calibration and press ENTER. 5. Select Execute Calibration and press ENTER. 6. The CALIBRATION PENDING message will appear.

6200518 Instrument Menus.fm

Page 45 Menu Functions

Section 4The analyzer will wait until any current measurement or priming cycle is complete before calibrating. Calibration typically takes 1 hours to complete.

Viewing the Calibration HistoryThis option allows you to review calibration data of past calibrations. 1. Press the MENU key to start from the Main Menu. 2. Select Sensor Menu and press ENTER. 3. Select the sensor by name and press ENTER. 4. Select Calibration and press ENTER. 5. Select Cal History and press ENTER. 6. A pop-up window will appear with the calibration date and time of the seven most recent calibrations (see the parameter specific operation manual for specific example). Select Review Next Cal to step through the pages of the calibration history. After the last calibration, press ENTER to return to the previous menu. To exit the Cal History Menu, press EXIT. Data displayed for the seven most recent calibrations will include the standard concentrations, and the date and time of the calibration. Please see the parameter-specific operation manual for specific information. Executing the Default Calibration If the analyzer is giving incorrect readings, it can be reset to use the default calibration. The default calibration allows the analyzer to give good approximations of the measurement value. To ensure that measurements meet analyzer accuracy specifications, follow the default calibration with a manual calibration as described above. To set the analyzer to its default calibration, follow these steps: 1. Press the MENU key to start from the Main Menu. 2. Select Sensor Menu and press ENTER. 3. Select the sensor by name and press ENTER. 4. Select Calibration and press ENTER. 5. Select Default Calibration and press ENTER. 6. Select Yes and press ENTER to set the analyzer to its default calibration.Note: To exit without restoring the default calibration, select No and press ENTER or press EXIT.

Page 46 Menu Functions

6200518 Instrument Menus.fm

Section 44.2.3.2 Alarm SetupEach APA 6000 Analyzer provides 14 alarm signals for your system. Two of these, ALA and ALB, are permanently attached to the analyzers internal alarm outputs, labeled Relay A and Relay B. The remaining alarm signals, AL1 through AL12, can be attached to relays in optional Signal Output Modules (SOMs). SOMs can be purchased separately, and provide two relays and two analog outputs each. For information see the SOM Manual (Cat. No. 51250-18). Each of the alarm setup options provides a setting for measurement, enabling/disabling alarms, alarm type, and alarm set point. Alarms can be configured as one of six types: high, low, rate (in mg/L/hour units), sensor warning, sensor alarm, or unconfigured. Table 6 describes the parameters available on each alarm type. Parameters in bold print indicate default settings. Table 6 Parameters for Setting AlarmsOptions TYPE MEASUREMENT SET POINT ENABLED MEAS FAILURE Set point Alarms High, Low, Rate, Unconfigured [Sensor measurement] [Legal value] Yes, No Off, On, Hold Sensor Warnings and Alarms Sensor Warning, Sensor Alarm, Unconfigured ----Yes, No ---

To set up the alarms:1. Press the MENU key to start from the Main Menu. 2. Select Sensor Menu and press ENTER. 3. Select the sensor by name and press ENTER. 4. Select Alarm Setup and press ENTER. 5. Select the alarm to set up and press ENTER. ALA and ALB are internal analog outputs connected to relays in the APA analyzer. AL1 through AL12 may be connected to relays in an SOM 6. Select Type and press ENTER. 7. Select one of the following to choose the type of alarm: To disable an alarm, select Unconfigured and press ENTER. Select High, Low, or Rate to configure a set point alarm within those ranges and press ENTER. Go to step 8. Select Sensor Warning or Sensor Alarm to configure an alarm that will be triggered by a sensor alarm or warning. Go to step 9.

8. If High, Low or Rate was selected in step 7, the following setup options are available:

6200518 Instrument Menus.fm

Page 47 Menu Functions

Section 4 Select Measurement to change the sample measurement that is monitored for alarm generation. Select Set Point to specify a numeric value for the alarm. Select Enabled to turn the alarm on or off. Select Meas Failure to set the alarms response when sample is not flowing to the analyzer or the analyzer experiences a fault that prevents it from providing the selected measurement. On and Off turn the alarm on or off; Hold leaves the alarm in the state it was in prior to the failure.

9. If you selected Sensor Warning or Sensor Alarm in step 7, select Enabled. Select Yes or No to turn the alarm on or off, respectively. 10. Press ENTER to save the alarm setting, or press EXIT to cancel.

4.2.3.3 Recorder SetupEach APA 6000 analyzer provides up to 14 analog output signals. Two of these signals, RCA and RCB, are permanently attached to the analyzers internal analog outputs, labeled Output A and Output B. RCA and RCB may also be configured as recorder outputs. The remaining analog signals, numbered RC1 through RC12, can be attached to analog outputs in SOMs. SOMs can be purchased separately and provide two relays and two analog outputs each (see the AquaTrend Interface Manual for networking information). Table 7 describes the parameters available for recorders and controllers. Parameters in bold print indicate default settings. Table 7 Parameters for Setting RecordersOptions TYPE MEAS MIN MAX ENABLED MEAS FAILURE OUTPUT SETUP Recorder Recorder, Unconfigured [Sensor measurement] [value] (As a recorder output, this value is always shown in the default units of mg/L regardless of what is selected and displayed on the AquaTrend) [value] (As a recorder output, this value is always shown in the default units of mg/L regardless of what is selected and displayed on the AquaTrend) Yes, No Go Max, Go Min., Hold Only RCA and RCB

Configuring Analog Outputs as Recorder SignalsTo set up an analog output to a recorder, perform the following steps: 1. Press the MENU key to start from the Main Menu. 2. Select Sensor Menu and press ENTER. 3. Select the sensor by name and press ENTER. 4. Select Recorder Setup and press ENTER. 5. Select the appropriate analog output to use and press ENTER. Select RC [1 -12] to configure a SOM output to a recorder. Select RCA or RCB to configure one of the APA analog outputs as a recorder output.Page 48 Menu Functions

6200518 Instrument Menus.fm

Section 46. To change the type of output, select TYPE and press ENTER. 7. Select Recorder to send the analog output to a recorder 8. Press ENTER to confirm the setting or press EXIT to cancel. 9. To change the measurement that the recorder output will track, select Measurement: XXXX and press ENTER. 10. Select the desired measurement and press ENTER. 11. To change the minimum measurement value that corresponds to the minimum recorder output, select Min: XXXX and press ENTER. 12. Modify the minimum value by using the arrow keys and press ENTER when the value entry is complete. 13. To change the maximum measurement value that corresponds to the maximum recorder output, select Max: XXXX and press ENTER. 14. Modify the maximum value by using the arrow keys and press ENTER when the value entry is complete. 15. To enable or disable the recorder output, select Enabled and press ENTER. 16. Select Yes or No to indicate that the recorder output is enabled or disabled, respectively, and press ENTER. 17. To change the recorder output action when the analyzer is unable to perform measurements in the selected measurement range, select MEAS Failure: XXXX and press ENTER. 18. Select Hold, Go Max, or GO Min to indicate what the output action should be when the analyzer cannot perform a selected measurement. Go Max sets the output to 20 mA, Go Min sets the output to 4 mA, and HOLD keeps the output at the current value. Press ENTER. 19. To complete the recorder setup for RCA or RCB see the following section. To complete the recorder setup for RC112 see Analog Output Settings in the SOM Manual.

Adjusting Output for RCA and RCB (Trim)Use the trim adjustments to set zero or full scale output levels. The trim adjustments allow correction for recorder response without the need to adjust the recorder itself. The range of adjustment is 0.011 mA for zero scale and 13.222.5 mA for full scale. The current values are nominal values and are accurate to 1.5%. The recorder full scale (20 mA) may not exactly match the APA full scale (20 mA). You can adjust the outputs so that full scale on the APA is equal to full scale on the recorder. The same holds true for the zero value.

6200518 Instrument Menus.fm

Page 49 Menu Functions

Section 4To adjust (trim) the milliamp output range for the internal recorder/controller outputs (RCA or RCB) or to change the range: 1. Select Output A Setup or Output B Setup. Press ENTER. 2. To adjust the mA output at the bottom of the scale: a. Select Zero Scale Trim and press ENTER. b. Use the up/down arrow to select Fine or Coarse Adjustment. Press ENTER. The new trim mA value will appear. c. Select Accept New Value when done and press ENTER. 3. To adjust the mA output at the top of the scale: a. Select Full Scale Trim and press ENTER. b. Use the up/down arrow to select Fine or Coarse Adjustment. Press ENTER. The new trim mA value will appear. c. Select Accept New Value when done and press ENTER 4. To select either 020 mA or 420 mA as the output range, select Output Range and press ENTER.

4.2.3.4 Measurement OptionsSelecting Measurement UnitsThe selection of units that each sensor can display is parameter specific. To view the selection and change the display, follow the directions below.Important Note: Changing units causes the data log for the measurement to be erased.

1. Press the MENU key to start from the Main Menu. 2. Select Sensor Menu and press ENTER. 3. Select the sensor by name and press ENTER. 4. Select Measurement Options and press ENTER. a. Select Chemical Form and press ENTER. Select the desired chemical form and press ENTER. b. Select Sample Sequence and press ENTER. Assign a sample sequence and press ENTER. c. Select Display and assign the reading to the appropriate DDM if desired. Press ENTER. d. Select Units and press ENTER. Select Hold Outputs and press ENTER. Select the units you want to use and press ENTER, or press EXIT to cancel.Note: When units are changed, the display shows new units with the next measurement.

Page 50 Menu Functions

6200518 Instrument Menus.fm

Section 44.2.3.5 Enabling Solution Level MonitoringSolution level monitoring by default is set to Off. The display will show Level Monitor: On/Off. When On is chosen (solution monitoring enabled), solution levels will be displayed as 0 to 100% full. To turn Solution Level Monitoring on or off proceed as follows: 1. Press the MENU key to start from the Main Menu. 2. Select Sensor Menu and press ENTER. 3. Select the sensor by name and press ENTER. 4. Select Solutions and Cleaning and press ENTER. 5. Select Level Monitor: On/Off and press ENTER. a. If monitoring is off the display will show: Turn on Level Monitor Levels Will Start @ 100% b. Select Yes and press ENTER to enable monitoring or select No to exit. c. If monitoring is on, the display will show: Turn off Level Monitor. Select Yes and press ENTER to disable or select No to exit.

4.2.3.6 Setting Solution Levels1. Press the MENU key to start from the Main Menu. 2. Select Sensor Menu and press ENTER. 3. Select the sensor by name and press ENTER. 4. Select Solutions and Cleaning and press ENTER. 5. Select Solution Levels and press ENTER. 6. The current % Full is displayed for each reagent. 7. If all reagents have been cleaned and refilled, select Reset All to 100% and select ENTER. 8. If all reagents have not been cleaned and refilled to 100%, select the solution(s) value to be changed and press ENTER. 9. A pop-up box will show the default setting of 100%. Use the up and down ARROW keys to change the setting to reflect the amount of solution in the bottle. 10. Press the ENTER key to accept the value or press Exit to Cancel.

6200518 Instrument Menus.fm

Page 51 Menu Functions

Section 44.2.3.7 Solutions & Cleaning OptionsThere are two cleaning options, Clean System Now and Auto-Clean. Clean System Now will start the clean cycle as soon as the current operation finishes. The Auto-Clean option allows the user to change the frequency of the cleaning cycle. 1. Press the MENU key to start from the Main Menu. 2. Select Sensor Menu and press ENTER. 3. Select the sensor by name and press ENTER. 4. Select Solutions & Cleaning and press ENTER. 5. Select Instrument Clean and press ENTER. 6. Select Auto-Clean and press ENTER. Select the Automati