ANL MCP #145 Phosphor Tests 1/24/2011

3
ANL MCP #145 Phosphor Tests 1/24/2011 Incom substrate, 33mm diameter, ANL MCP#145 20µm pores, 8° bias, 60:1 L/D, 65% OAR Chemistry 2 resistive layer, MgO SEE layer Electrode deposited on top of ALD layers Annealing conditions unknown (Anil?) 25MΩ in air with DMM. 25MΩ in vacuum initially with DMM, then drifts up to 70MΩ. 25MΩ in vacuum at voltage (20µA at 500V) 25 January 2011 J. McPhate – LAPP Main Telecon 1

description

ANL MCP #145 Phosphor Tests 1/24/2011. Incom substrate, 33mm diameter, ANL MCP#145 20µm pores, 8° bias, 60:1 L/D, 65% OAR Chemistry 2 resistive layer, MgO SEE layer Electrode deposited on top of ALD layers Annealing conditions unknown (Anil?) 25MΩ in air with DMM. - PowerPoint PPT Presentation

Transcript of ANL MCP #145 Phosphor Tests 1/24/2011

Page 1: ANL MCP #145 Phosphor Tests 1/24/2011

J. McPhate – LAPP Main Telecon 1

ANL MCP #145 Phosphor Tests1/24/2011

• Incom substrate, 33mm diameter, ANL MCP#145• 20µm pores, 8° bias, 60:1 L/D, 65% OAR• Chemistry 2 resistive layer, MgO SEE layer• Electrode deposited on top of ALD layers• Annealing conditions unknown (Anil?)• 25MΩ in air with DMM.• 25MΩ in vacuum initially with DMM, then drifts up

to 70MΩ.• 25MΩ in vacuum at voltage (20µA at 500V)

25 January 2011

Page 2: ANL MCP #145 Phosphor Tests 1/24/2011

J. McPhate – LAPP Main Telecon 2

MCP #145, Phosphor Imaging

25 January 2011

1000V900V

900V

Screen 3000V for all

Page 3: ANL MCP #145 Phosphor Tests 1/24/2011

J. McPhate – LAPP Main Telecon 3

Output vs. Voltage

25 January 2011

200 300 400 500 600 700 800 900 1000 1100 12000.01

0.1

1

10

100

1000

10000

"Gain" vs. Voltage, ALD MCP 145

MCP 145 (1 lamp)MCP 103 (2 lamps)

MCP HV (volts)

Corr

ecte

d Ph

otoc

urre

nt (n

A)

Hard to make direct comparison to MCP #103 (Ar anneal discolored MCP), but looks to be significantly higher gain.

Note use of one penray lamp on MCP #145, versus two lamps for MCP #103, yet output is about twice.