AMI Hydrazine

download AMI Hydrazine

of 84

Transcript of AMI Hydrazine

  • 8/18/2019 AMI Hydrazine

    1/84

     AMI Hydrazine Version 4.00 and later 

     A-96.250.501 / 070606

       O  p  e  r  a   t  o  r   ’  s   M

      a  n  u  a   l

    SWAN sets the standard 

  • 8/18/2019 AMI Hydrazine

    2/84

    © 2007, SWAN ANALYTISCHE INSTRUMENTE AG, Switzerland, all rights reserved

    subject to change without notice

    Customer Support

    SWAN and its representatives maintain a fully trained staff of technical specialists

    around the world. For any technical question, contact your nearest

    SWAN representative, or the manufacturer:

    SWAN ANALYTISCHE INSTRUMENTE AG

    Turicaphonstrasse 29

    8616 Riedikon

    Switzerland

    Internet: www.swan.ch

    E-mail: [email protected]

    Document Status

    Title: Monitor AMI Hydrazine Operator’s Manual

    ID:  A.96.250.501

    Revision Issue

    01 June 2007

  • 8/18/2019 AMI Hydrazine

    3/84

    AMI HydrazineTable of Contents

     A-96.250.501 / 070606 1

    Table of Contents1. Safety Instructions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4

    1.1 General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4

    1.2 Warning Notices. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5

    1.3 General Safety Regulations . . . . . . . . . . . . . . . . . . . . . . . 6

    1.3.1 Legal Requirements. . . . . . . . . . . . . . . . . . . . . . . . . . . 6

    1.3.2 General Inspection and Maintenance Duties . . . . . . . . 6

    1.3.3 Spare Parts and Disposables . . . . . . . . . . . . . . . . . . . 6

    1.3.4 Modifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61.3.5 Electrical shock hazard . . . . . . . . . . . . . . . . . . . . . . . . 6

    1.4 Restrictions for use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7

    2. Product Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . 8

    2.1 Description of the system. . . . . . . . . . . . . . . . . . . . . . . . . 8

    2.2 Instrument Specification. . . . . . . . . . . . . . . . . . . . . . . . . . 10

    2.3 Instrument Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13

    2.4 Display, Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14

    2.5 Software Structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16

    2.6 Changing parameters and values . . . . . . . . . . . . . . . . . . 17

    3. Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18

    3.1 Installation Checklist . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18

    3.2 Mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19

    3.3 Connecting sample and waste. . . . . . . . . . . . . . . . . . . . . 19

    3.4 Electrical Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . 19

    3.4.1 Electrical connection scheme . . . . . . . . . . . . . . . . . . . 21

    3.4.2 Power Supply. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22

    3.5 Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 223.6 Relay Contacts. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23

    3.6.1 Alarm Relay . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23

    3.6.2 Relay Contacts 1 and 2 . . . . . . . . . . . . . . . . . . . . . . . . 23

    3.7 Signal Outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25

    3.7.1 Signal Outputs 1 and 2 (Current Outputs) . . . . . . . . . . 25

    3.7.2 Signal Output 3 (Optional) . . . . . . . . . . . . . . . . . . . . . . 25

    3.8 Interface. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26

    3.8.1 Interface RS232 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26

    3.8.2 Interface RS485 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26

    3.9 Installation of Filter Vessel, Electrode andDiisopropylamine Bottle . . . . . . . . . . . . . . . . . . . . . . . . . . 27

  • 8/18/2019 AMI Hydrazine

    4/84

    2  A-96.250.501 / 070606

    AMI HydrazineTable of Contents

    4. Instrument Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29

    5. Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30

    5.1 Maintenance Schedule. . . . . . . . . . . . . . . . . . . . . . . . . . . 30

    5.2 Stop Operation before Maintenance . . . . . . . . . . . . . . . . 30

    5.3 Cleaning of Sample Supply/Filters. . . . . . . . . . . . . . . . . . 30

    5.4 Cleaning of Protective Filter . . . . . . . . . . . . . . . . . . . . . . . 30

    5.5 Maintenance of Hydrazine Sensor. . . . . . . . . . . . . . . . . . 31

    5.6 Maintenance of Reference Electrode. . . . . . . . . . . . . . . . 32

    5.7 Changing Diisopropylamine Bottle. . . . . . . . . . . . . . . . . . 32

    5.8 Changing the Diffusion Tube . . . . . . . . . . . . . . . . . . . . . . 335.9 Maintenance of Flow Cell. . . . . . . . . . . . . . . . . . . . . . . . . 34

    5.9.1 Longer Stop of Operation. . . . . . . . . . . . . . . . . . . . . . . 36

    6. Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37

    6.1 Calibration Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . 37

    6.2 Hydrazine Sensor. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37

    7. Trouble Shooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40

    7.1 Error messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 407.2 Error List. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41

    A. Programm Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . 43

    B. Profibus (Option) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46

    B.1. Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46

    B.2 Cyclic Data Telegram. . . . . . . . . . . . . . . . . . . . . . . . . . . . 47

    B.3 Customizing the Cyclic Data Telegram . . . . . . . . . . . . . . 48

    B.4 Data Formats . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49

    B.5. Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53B.5.1 Device Master Files (GSD). . . . . . . . . . . . . . . . . . . . . . 53

    B.5.2 GSD File Types . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53

    B.6. Acyclic Data Communication . . . . . . . . . . . . . . . . . . . . . . 54

    C. Modbus (Option) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55

    C.1. Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55

    C.2. Data and Control Functions . . . . . . . . . . . . . . . . . . . . . . . 57

    C.3. Function Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59

  • 8/18/2019 AMI Hydrazine

    5/84

    AMI HydrazineTable of Contents

     A-96.250.501 / 070606 3

    D. Programm List and Explanations. . . . . . . . . . . . . . . . . 63

    1 Messages. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63

    2 Diagnostics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63

    3 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64

    4 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65

    5 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69

    E. Default Values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77

    F. Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79

    http://ami_sodiump_en02.pdf/

  • 8/18/2019 AMI Hydrazine

    6/84

    4  A-96.250.501 / 070606

    AMI Hydrazine AMI Hydrazine - Operator’s Manual

     AMI Hydrazine - Operator’s ManualThis document describes the main steps for instrument setup,operation and maintenance.

    1. Safety Instructions

    This chapter contains general information assuring safe operation

    of the instrument. More safety instructions are given throughout thismanual, at the respective locations where observation is most

    important.

    Strictly follow all safety instructions in this publication.

    1.1 General

    The instructions included in this section explain the potential risksassociated with instrument operation and provide important safetypractices designed to minimize these risks.

    If you carefully follow the information contained in this section, youcan protect yourself from hazards and create a safer workenvironment.

    Target audience

    Operator: Qualified person who uses the equipmentfor its intended purpose.

    Instrument operation requires thorough knowledge of applications,

    instrument functions and software program as well as all applicablesafety rules and regulations.

    Qualification, Training

    To be qualified for instrument operation, you must read andunderstand the written instructions in this operator’s manual, haveappropriate training (by SWAN Service Technician), and know thesafety rules and regulations.

    Operator’s Manual Location

    The AMI Powercon Operator’s Manual shall be kept in proximity ofthe instrument.

  • 8/18/2019 AMI Hydrazine

    7/84

    AMI HydrazineSafety Instructions

     A-96.250.501 / 070606 5

    1.2 Warning NoticesThe symbols used for safety-related notices have the following

    significance:

    WARNINGSymbols

    Warning

    Generally, the triangular warning symbol indicates the possibilityof personal injury or even loss of life if instructions are not followed,e.g.

    electrical shock hazard

    harmful to health

    inflammable

    AttentionSymbols

    Attention

    The attention symbol indicates the possibility of equipmentdamage, malfunctions or incorrect process results if instructionsare not followed.

    WarningFor safe instrument installation and operation you must read andunderstand the instructions in this manual, as well as the MaterialSafety Data Sheets (MSDS) of diisopropylamine.

    Only SWAN trained and authorized personnel shall perform thetasks described in this document.

  • 8/18/2019 AMI Hydrazine

    8/84

    6  A-96.250.501 / 070606

    AMI HydrazineSafety Instructions

    1.3 General Safety Regulations1.3.1 Legal Requirements

    The user is responsible for proper system operation.

     All precautions must be followed to ensure safe operation of the

    instrument.

    1.3.2 General Inspection and Maintenance Duties

    The SWAN instrument shall only be operated in optimal running

    condition. The preventive maintenance procedures listed in thismanual must be performed.

    In case of safety relevant deficiencies, SWAN must immediately beinformed.

    1.3.3 Spare Parts and Disposables

    Use only official SWAN spare parts and disposables. If other partsare used during the normal warranty period, the manufacturer’swarranty is voided.

    1.3.4 Modifications

    Modifications and instrument upgrades shall only be carried out byan authorized Service Technician. SWAN will not acceptresponsibility for any claim resulting from unauthorized modificationor alteration.

    1.3.5 Electrical shock hazard

    WARNING

    If proper operation is no longer possible, the instrument must bedisconnected from all power lines, and measures must be taken toprevent inadvertent operation.

    To prevent from electrical shock, always make sure thatthe ground wire is connected.

    Service shall be performed by authorized personnel only.

    Whenever electronic service is required, disconnectinstrument power and power of devices connected to

    • relay 1,

    • relay 2,

    • alarm relay.

  • 8/18/2019 AMI Hydrazine

    9/84

    AMI HydrazineSafety Instructions

     A-96.250.501 / 070606 7

    1.4 Restrictions for useAlcalizing

    ReagentPlease correct the pH of the sample only with Diisopropylamine,p.a. However, Diisopropylamine is not included in delivery. Pleasebuy it from your local supplier: E.g. VWR Merck (6x 1l: 803646),Fluka (1x 1l: 38300) or Riedel-de Haen (1x 1l: 62580).

    SampleRequire-

    ments

    No sand (or other polishing material) or oil are allowed in thesample.

    The instrument is specified for a max. pressure of 2 bar (28 psi). Ifthe sample pressure exceeds 2 bar, please mount a pressure

    reduction in front of the instrument inlet.Phosphates may interfere with the hydrazine measurement.

    The pH value of the sample must be equal or higher than pH 7.

    FlowInterruption

    High concentrations of Diisopropylamine reagent may attack theacrylic material of the flow cell. This can cause cracks and renderthe flow cell completely unusable. This will not happen duringnormal operation. However, if the sample flow is stopped, undiluteddiisoproylamine will diffuse through the diffusion tube and cancause damage. Therefore, the following rules should be observed if

    the sample flow is stopped for more than one day: Remove diisopropylamine bottle and close it. Keep it in a

    safe place for further use.Note: Diisopropylamine is highly inflammable and toxic.Wear googles and protective clothing! Do not breathevapours! Avoid contact with skin! 

    Read the safety data sheet carefully before use

    Screw an empty reagent bottle on the holder.

    Let the sample flow for another 30 minutes to rinse out theDiisopropylamine dissolved in the diffusion tubing material.

    Stop sample flow.

  • 8/18/2019 AMI Hydrazine

    10/84

    8  A-96.250.501 / 070606

    AMI HydrazineProduct Description

    2. Product Description

    2.1 Description of the system

    Application Hydrazine is used in power plants as an oxygen scavenger. It helpsto remove oxygen, which may cause corrosion in the power plant.The reaction of hydrazine with oxygen produces nitrogen andwater.

    Measuringprinciple

    3-electrode amperometry:

    The sensor consists of two platinum electrodes and a referenceelectrode. A voltage is applied to the platinum electrodes and kepton a optimum level by the reference electrode. The hydrazine in thesample generates a small current on the platine pin electrode,which is proportional to the hydrazine concentration.

    For an optimal sensitivity and stability, a rotor continuously cleansthe surfaces of the platinum electrodes (hydrodynamic cleaning).The flow sensor measures the rotations of the rotor to detectsufficient flow.

    The signal of amperometric systems depends on flow. The constanthead guarantees a constant flow if the sample always overflowsinto the longer constant head tube.

    Temperature compensation is done automatically.

    On-lineoperation

    The measurement of hydrazine requires a high pH of the sample.The AMI Hydrazine buffers the pH to pH>10.5 by addingdiisopropylamine. The diisopropylamine is dissolved in the waterwhile flowing through a diffusion tube.

    The sample enters at the sample inlet , passes the filter  and theneedle valve, where sample flow is adjusted and flows through the

    diffusion tube in the diisopropylamine bottle, picking updiisopropylamine and reaching pH>10.5. Afterwards, the samplefills the constant head . Sample must always overflow into thelonger constant head tube (waste) to ensure constant flow andpressure at the hydrazine sensor.

  • 8/18/2019 AMI Hydrazine

    11/84

    AMI HydrazineProduct Description

     A-96.250.501 / 070606 9

     A large part of the sample flows into the shorter overflow tube tothe hydrazine sensor , turns on the rotor  and leaves it into waste.

     A small part of the sample leaves thje instrument by the longerconstant head tube.

    The rotation of the rotor is monitored with a flow detector  to assuresufficient sample flow. An inconsistent sample flow causes the rotorto turn slowly (or stop) and produces a system error.

    3

    1

    2

    4

    8

    7

    6

    5

    9

    10

    2

    11

    1 sample inlet

    2 filter 

    3 needle valve

    4 sample flow tohydrazine sensor 

    5 hydrazine sensor

    6 flow measurement

    7 temperaturemeasurement

    8 constant head overflow

    9 sample outlet/waste

    10 grab sample

    11 Diisopropylaminebottle with diffusiontube

  • 8/18/2019 AMI Hydrazine

    12/84

    10  A-96.250.501 / 070606

    AMI HydrazineProduct Description

    2.2 Instrument SpecificationSample

    Require-ments

    Flow rate: approx. 15 l/h

    Sample pressure inlet: 0.15 - 2 bar (3 - 28 PSI)

    Sample outlet: pressure-free

    Temperature range: 15 - 45 °C (41 - 113 °F)

    No oil, no grease, no sand.

    pH value of the sample must be equal or higher than pH 7.

    Reagent consumption: < 1l Diisopropylamine per month (25 °C).

    On-siteRequire-

    ments

    Sample inlet: Tube 4x6 mm.Drain: 14x20 mm (1/2”) hose nozzle which must end

    in convenient atmospheric waste of sufficientcapacity.

    Power: 85-265 VAC, 47-63 Hz or 24 VDC, isolated,power consumption max. 20 VA.

    MeasuringRange

    Hydrazine

    Range: 0.1 - 600ppb

     Accuracy: ±5% of reading up to 200 ppb,

    ±15% up to 600ppb, or ±2ppb

    Stability: ±5% of reading per month, or ±2ppb per month

    Response time: 90% of change in 60 sec after the sample entered

    the flow cell.

    MeasuringRange

    Temperature

    Measuring range: up to 60°C

    Resolution: 0.1°C

    Sample FlowMeasurement

    With Swan sample flow meter and alarm in cases of insufficient

    sample flow

    SafetyFeatures

    No data loss after power failure, all data is saved in non-volatile

    memory. Overvoltage protection of in- and outputs. Galvanic

    separation of measuring inputs and signal outputs.

    AlarmRelay

    One potential free contact for summary alarm indication for

    programmable alarm values and instrument faults.

    Maximum load: 1A / 250 VAC

    Input One input for potential-free contact. Programmable hold or remote

    off function.

  • 8/18/2019 AMI Hydrazine

    13/84

    AMI HydrazineProduct Description

     A-96.250.501 / 070606 11

    Relays Two potential-free contacts programmable as

    • limit switches for measuring values

    • controllers

    • timer for system cleaning with automatic hold function

    Rated load: 1A / 250VAC

    SignalOutputs

    Two programmable signal outputs for measured values (freelyscaleable, linear, bilinear or logarithmic) or as continuous controloutput (control parameters programmable).

    Current loop: 0/4 - 20mA

    Maximum burden: 510ΩThird signal output with the same specifications as option available.

    Communi-cation Inter-

    face (option)

    RS232 interface for logger download with HyperTerminal or RS485

    interface with Fieldbus protocol Modbus or Profibus DP V1.

    ControlFunction

    Relays or current outputs programmable for 1 or 2 pulse dosing

    pumps, solenoid valves or for one motor valve.

    Programmable P, PI, PID or PD control parameters

  • 8/18/2019 AMI Hydrazine

    14/84

    12  A-96.250.501 / 070606

    AMI HydrazineProduct Description

    Dimensions

    Mounting: Panel: Distance between mounting holes: 254x824 mm

    Screws: 6 mm in diameter 

    Enter 

    ppb

    RUN

    Ø 10

    Exit

    *

    *

    *

    13

     AMI Hydrazine

    24.8°C

    14:36:15

    13

    254280

    15.3

    850

    824

  • 8/18/2019 AMI Hydrazine

    15/84

    AMI HydrazineProduct Description

     A-96.250.501 / 070606 13

    2.3 Instrument Overview

    Enter 

    ppb

    14:36:15

     AMIHydrazine

    RUN

    Exit

    *

    *

    24.8°C

    15.3

    *

    Mounting panel

    Bottle holder 

    Diisopropylamine

    bottle with diffusion

    tube

    Diffusion tube

    Reference

    electrode

    Grab sample

    valve

    Grab sample

    outlet

    Needle

    valve

    Hydrazine

    sensor with rotor 

    Transmitter 

    Temperature

    sensor 

    Constant head

    cover 

    Constant head

    Constant headtube long

    Constant head

    tube short

    Flow cell block

    Flow sensor (behind inlet)

    Sample inlet

    Filter vessel and

    filter 

    Waste

  • 8/18/2019 AMI Hydrazine

    16/84

    14  A-96.250.501 / 070606

    AMI HydrazineProduct Description

    2.4 Display, KeysDisplay of Measured

    Value backlit LCD

    Menu navigation keys

    upper / lower limit not yet reachedupper / lower limit reached

    control upw. / dnw.: no action

    control upw. / dnw.: active, dark bar indicates control intensity

    timer 

    timer: timing active (hand rotating)

    Relay Status

    motor valve: valve closed

    rpm

    motor valve: open, dark bar indicates approx. position

    °C

    Status Time

    R1 Process

    ValuesR2

    blinking: fatal error error 

    HOLD input closed or cal delay: Instrument on hold (shows status of signal

    outputs). Add. info see Programm List and Explanations, p. 63.

    OFF input closed: Control/limit is interrupted (shows status of signal

    outputs). Add. info see Programm List and Explanations, p. 63.

    RUN normal operation

    sample

    temperature

    sample

    flow

    Hydrazine xx.x ppb

  • 8/18/2019 AMI Hydrazine

    17/84

    AMI HydrazineProduct Description

     A-96.250.501 / 070606 15

    Keys

    ProgramAccess, Exit

    to exit a menu or command (without accepting)

    to move back to the previous menu level

    to move DOWN in a menu list and to decrease digits

    to move UP in a menu list and to increase digits

    to open a selected sub-menu

    to accept an entry

    Exit

    Enter 

    25.4°C

    RUN

    305 rpm

    14:10:45

    R1

    15.3 ppbR2

    1

    InstallationOperation

    DiagnosticsMessages

    Maintenance

    Main MenuEnter 

    Exit

  • 8/18/2019 AMI Hydrazine

    18/84

    16  A-96.250.501 / 070606

    AMI HydrazineProduct Description

    2.5 Software Structure1

    Messages

    OperationMaintenanceDiagnostics

    Main Menu

    Menu 4: Operation

    Subset of menu 5 -

    installation, but process-

    related. User relevant

    parameters that might need to

    be modified during dailyroutine. Normally password

    protected and used by the

    process-operator.

    Menu 1: Messages

    Reveals pending errors as

    well as an event history

    (time and state of events that

    have occurred at an earlier

    point of time).

    It contains user relevant data.

    Menu 2: Diagnostics

    Provides user relevant

    instrument and sample data.

    Menu 5: Installation

    For initial instrument set up by

    SWAN authorized persons, to

    set all instrument parameters.

    Can be protected by means ofa password.

    Menu 3: Maintenance

    For instrument calibration,

    relay and signal output

    simulation, to set theinstrument time. It is used by

    the service personnel.

    Installation

    1.1

    Pending ErrorsMessage List

    Messages

    2.1

    InterfaceI/O StateSample

    IdentificationSensors

    Diagnostics

    3.1

    CalibrationSimulation

    Maintenance

    Set Time 01.01.05 16:30:00

    4.1

    Logger Relay ContactsSensors

    Operation

    5.1

    Interface

    MiscellaneousRelay Contacts

    SensorsSignal Outputs

    Installation

  • 8/18/2019 AMI Hydrazine

    19/84

    AMI HydrazineProduct Description

     A-96.250.501 / 070606 17

    2.6 Changing parameters and valuesChanging

    para-metersThe following example shows how to change the current loop from

    4-20 mA to 0-20 mA (see Programm Overview, p. 43 Installation):

    Changingparameter 

    values

    Highlight the menu item

    indicating the parameter you

    want to change.

    Press .

    Press or to highlight

    the required parameter.

    Press to confirm the

    selection (or [Exit] to keep the

    previous parameter).

    The selected parameter isindicated (but not saved yet).

    Press .

    Yes highlighted, press to

    save the new parameter. The

    system reboots, the new

    parameter is set.

    Enter 

    Enter 

    Exit

    Enter 

    5.2.1Signal output 1

    Current loop.Parameter x Sec

    Scaling 1 MinFunction Free chlorine

    Curr. loop.

    4-20 mA

    5.2.1Signal output 1

    Current loop 0-20 mAParameter Hydrazine

    Function linNo

    Save ?

    Yes

    5.2.1Signal output 1

    Function lin

    Parameter Hydrazine

    Current loop 4-20 mA

    5.2.1Signal output 1

    Function lin

    Parameter HydrazineCurrent loop 0-20 mA

    Scaling

    0-20 mA

    Scaling

    Scaling

    5.3.1.1

     Alarm High 1000 ppb

    Alarm Dis

     Alarm Low 0.00 ppbHysteresis 1.0 ppbDelay 5 Sec

    5.3.1.1Alarm Dis

     Alarm Low 0.00 ppbHysteresis 1.0 ppbDelay 5 Sec

     Alarm High 30.0 ppb

    Select parameter and press .

      to set required value.

     to confirm new value.

     to exit the menu.

    Enter 

    Enter 

    Exit

  • 8/18/2019 AMI Hydrazine

    20/84

    18  A-96.250.501 / 070606

    AMI HydrazineInstallation

    3. InstallationThe first part of this chapter describes the preparing and placing ofthe system in position for use.Restrictions on use

    If exposed to weather, install a cover to protect the instrument from

    insolation and precipitation. Please see also Restrictions for use, p.

    7. See Instrument Specification, p. 10 for temperature limitations.

    3.1 Installation Checklist

    Check Instrument’s specification must conform to your AC power ratings.

    Do not turn on power until instructed to do so.

    On site

    requirements

    85-265 VAC, 47-63 Hz or 24 VDC power outlet with ground

    connection and 20 VA.

    Sample line with 15 l/h and 0.15 - 2 bar (3 - 28 psi).

    Waste line with atmospheric drain.

    See Connecting sample and waste, p. 19 and Electrical Connec-

    tions, p. 19

    Installation Mount the instrument in vertical position. Display should be at eye

    level.

    Mount the filter vessel. Connect the sample and waste line.

    Electrical

    Wiring

    Connect all external devices like limit switches, current loops and

    pumps. Connect power cord; do NOT switch on power yet!

    Electrodes Install reference electrode. Connect to cable TR.

    Insert temperature sensor into small opening of cover.

    Diiso-proplyamine

    Wear safety gloves and safety glasses. Add 20 ml high puritywater to full Diisopropylamine bottle. Slide bottle over diffusion

    tubes and fix on holder.

    Power-up Turn on the sample flow and wait until the rotor on the Hydrazine

    sensor starts turning.

    Switch on power.

    Instrument

    set-up

    Program all parameters for external devices (interface, etc.).

    Program all parameters for instrument operation (limits, alarms).

    Run-in period Let the instrument run-in for approximately 30 min continuousoperation (flow on, power on). The Diisopropylamine needs

    roughly 30 min to penetrate the walls of new diffusion tubes.

  • 8/18/2019 AMI Hydrazine

    21/84

    AMI HydrazineInstallation

     A-96.250.501 / 070606 19

    3.2 MountingFor ease of operation mount the instrument so that the display is oneye level.

    For the panel size, see Dimensions, p. 11

    3.3 Connecting sample and waste

    Sample inlet:Use plastic tube (4x6 mm) toconnect the sample line.

    Waste:

    Connect the 1/2” tube to the wastenozzle and place it into theatmospheric drain.

    3.4 Electrical Connections

    Warning

     Always turn off AC power before manipulating electric parts.Grounding requirements: Only operate the instrument from an

    power outlet which has a ground connection.Make sure the power specification of the instrument corresponds tothe power on site.

    inlet

    outlet/waste

  • 8/18/2019 AMI Hydrazine

    22/84

    20  A-96.250.501 / 070606

    AMI HydrazineInstallation

    Cablethick-

    nesses

    In order to comply with IP66, use the following cable thicknesses:

    Wire Power, Relays

    Use max. 1.5 mm2 / AWG 14 stranded wire with end sleeves.

    Signal Outputs, InputUse 0.25 mm2 / AWG 23 stranded wire with end sleeves.

    WARNING

    Make sure that the devices to be connected to

    relay 1

    relay 2

    alarm relay

    are disconnected from the power before resuming installation.

    WARNING

    To prevent from electrical shock, do not connect the instrument tothe power unless the ground wire (PE) is connected.

    Do not connect unless specifically instructed to do so.

    PG 9 fittings: Cable Øouter  4 - 8 mm

    PG 11 fitting: Cable Øouter  5 - 10 mm

    PG 7 fittings: Cable Øouter  3 - 6.5 mm

    Protect

    unused

    glands.

  • 8/18/2019 AMI Hydrazine

    23/84

    AMI HydrazineInstallation

     A-96.250.501 / 070606 21

    3.4.1 Electrical connection scheme

    Use only the terminals shown in this diagram, and only for the men-

    tioned purpose. Use of any other terminals will cause short circuits with

    possible damage to material and personnel.

  • 8/18/2019 AMI Hydrazine

    24/84

    22  A-96.250.501 / 070606

    AMI HydrazineInstallation

    3.4.2 Power Supply

    Terminal 1 Phase L

    Terminal 2 Neutral N

    Attention:

    It is not allowed to power external loads out of the AMI transmitter.

    3.5 Input

    Note: Use only potential-free (dry) contacts.

    Terminals 16/42

    For programming see Programm Overview, p. 43 

    Menu Installation

    power supply connector 

    Neutral

    Power 

    The protective earth cable(ground) MUST be

    connected to the provided

    grounding screw.

  • 8/18/2019 AMI Hydrazine

    25/84

    AMI HydrazineInstallation

     A-96.250.501 / 070606 23

    3.6 Relay Contacts3.6.1 Alarm Relay

    Note: Max. load 1 A / 250 VAC  

     Alarm output for system errors.

    Error codes see Maintenance, p. 30.

    Open during normal operation.

    Closed on error and loss of power.

    Terminals 10/11.

    3.6.2 Relay Contacts 1 and 2

    Note: Rated load 1 A / 250 VAC  

    Relay 1: Terminals 6/7

    Relay 2: Terminals 8/9

    For programming see Programm Overview, p. 43, Menu Installation

    AttentionHeavy inductive or directly controlled loads (solenoid valves,

    dosing pumps) must be switched by external power relays. In this

    case, use the AMI relaybox which is available as an option.

    Small inductive loads (max 0.1 A) as for example the coil of a

    power relay can be switched directly. To avoid noise voltage in the

     AMI it is mandatory to connect an snubber circuit in parallel to the

    load (this procedure is not necessary if a AMI Relaybox is used)

  • 8/18/2019 AMI Hydrazine

    26/84

    24  A-96.250.501 / 070606

    AMI HydrazineInstallation

    Resistive loads (max. 1 A) and control signals for PLC, impulse

    pumps and so on can be connected without further measures..

     Actuators are using both relays: One relay contact is used for

    opening (relay 1), the other for closing (relay 2) the valve, i.e. with

    the 2 relay contacts available, only one motor valve can be

    controlled.

    Motors with loads bigger than 0.1 A must be controlled via external

    power relais or an AMI Relaybox.

  • 8/18/2019 AMI Hydrazine

    27/84

    AMI HydrazineInstallation

     A-96.250.501 / 070606 25

    3.7 Signal Outputs3.7.1 Signal Outputs 1 and 2 (Current Outputs)

    Note: Max. burden 510 Ω .

    If signals are sent to two different receivers, use signal

    isolator (loop isolator).

    Signal output 1: Terminals 14 (+) and 13 (-)

    Signal output 2: Terminals 15 (+) and 13 (-)

    For programming see Programm Overview, p. 43, Menu Installation

    3.7.2 Signal Output 3 (Optional)

    Requires the additional board for the third signal output 0/4 - 20 mA

    PCB Max. burden: 510 Ω.

    The third signal output is installed in the holder on the main board.

    You can operate either 3. signal output OR communication

    interface, not both!

    Terminal 38 (+) and 37 (-).

    For programming see Installation Signal output 3 in Programm

    Overview, p. 43.

    Third signal output

    0/4 - 20 mA PCB

    installed on main board 

  • 8/18/2019 AMI Hydrazine

    28/84

    26  A-96.250.501 / 070606

    AMI HydrazineInstallation

    3.8 Interface3.8.1 Interface RS232

    Terminal 50, 52, 53

    The AMI Interface RS232 Print Circuit Board (=PCB) is used for

    Logger download and Firmware upload. For detailed information

    see the corresponding manual “AMI RS232 Interface”.

    RS232 Interface PCB

    3.8.2 Interface RS485

    Terminal 37 PB, Terminal 38 PA

    To connect several instruments by means of a network or to

    configure a PROFIBUS DP connection, consult the PROFIBUS

    section in this manual (Profibus (Option), p. 46). Use appropriatenetwork cable.

    Note: Bus termination only by means of ON/OFF switch on the

    RS485 PCB. If only one instrument is installed, or on the last

    instrument in a bus, the switch must be ON.

    RS485 PCB

    on/off switch

  • 8/18/2019 AMI Hydrazine

    29/84

    AMI HydrazineInstallation

     A-96.250.501 / 070606 27

    3.9 Installation of Filter Vessel, Electrode andDiisopropylamine Bottle

    Position of Sensors in

    the Flow Cell

    ConstantHead

    Mount the filter vessel.

    Enter 

    ppb

    14:36:15

     AMIHydrazine

    RUN

    Exit

    *

    *

    24.8°C

    15.3

    *

    Temperature

    sensor 

    Reference

    electrode, cable

    marked TR

    Hydrazine sensor,

    cable with BNC plug

    marked H

    Filter andfilter vessel

    Diisoproplyamine

    bottle

    Holder for Diiso-

    proplyamine bottle

    Diffusion tube

    Grab sample valve

    and outlet

    Screw cap

    Fixing screw

  • 8/18/2019 AMI Hydrazine

    30/84

    28  A-96.250.501 / 070606

    AMI HydrazineInstallation

    ReferenceElectrode

    1 Cautionally pull off the rubber cap from the reference electrodetip.

    2 Loosen the fixing screw and insert the electrode until it stops.

    3 Tighten fixing screw hand tight, take off plug cap, and connectto cable marked TR.

    4 Store the plug cap and the rubber cap for interruption ofoperation.

    TemperatureSensor 

    Fragile! Handle with care!

    Insert the temperature sensor into the small opening in the cover of 

    the flow cell.

    Check all sensor connections between AMI Hydrazine and sensors.

    Diisopropyl-amine Bottle

    Only install the diisopropylamine bottle if you intend to start

    operation immediately. Do NOT install if no sample in

    available!

    Attention!

    Diisopropylamine is a highly inflammable and toxic

    compound. Read the safety data sheet carefully before use.

    Avoid contact with skin! Do not inhale!You can fill the Diisopropylamine into the glass bottle (G45 thread),

    delivered with the instrument.

    If you want to connect a Merck bottle directly, you need a thread

    adapter (to be ordered at Merck).

     Add 20 ml high purity water into the full Diisopropylamine bottle to

    limit the consumption of Diisopropylamine at the beginning. Slide

    the Diisopropylamine bottle over the diffusion tubes and fix it to the

    holder by the screw cap.

  • 8/18/2019 AMI Hydrazine

    31/84

    AMI HydrazineInstrument Setup

     A-96.250.501 / 070606 29

    4. Instrument Setup After the analyzer is installed according to the previous instructions,connect the power cord. Do not switch on power, yet!

    BeforeStarting

    Sample Flow

    Ensure the grab sample valve is closed.Ensure the sample outlet is pressure-free.

    Open SampleFlow

    Open the flow regulating valve to allow sample flow into theconstant head and into the waste. Sample must always overflow

    into the longer constant head tubeWatch the rotor of the Hydrazine sensor. As soon as it startsturning, switch on power.First, the analyzer performs a self test, displays the firmwareversion and then starts normal operation.

    Programming Program all parameters for external devices (interface, etc.). Set allparameters for instrument operation (limits, alarms).

    Run-in Period Let the instrument run-in for approximately 30 min continuousoperation (flow on, power on). The Diisopropylamine needs roughly

    30 min to penetrate the walls of new diffusion tubes.

    Correction of Hydrazine

    Sensor 

    If necessary, correct the instrument after at least 30 min running-in. A calibration (zero and slope) is NOT necessary.

    To correct the instrument the concentration of hydrazine in thesample has to be known. Hydrazine reacts withdimethylaminobenzaldehyde in an acid solution and forms a yellowcolour. The colour intensity can be measured by a photometer.Refer to an appropriate standard method (e.g. DIN 38413).

    The manual sample must be taken from the grab sample outlet of

    the flow cell. Let the sample flow out for roughly 1 minute beforegrabbing the sample for the manual analysis.

    Do not calibrate.

  • 8/18/2019 AMI Hydrazine

    32/84

    30  A-96.250.501 / 070606

    AMI HydrazineMaintenance

    5. MaintenanceMaintenance frequency depends strongly on the water quality.

    5.1 Maintenance Schedule

    5.2 Stop Operation before Maintenance

    Put on safety gloves and wear safety glasses!

    Take off diisopropylamine bottle and close well. Screw an emptybottle on the holder. Let the sample run for 30 min.

    Stop sample flow. Wait until rotor stops and hydrazine reading is0 ppb.

    Shut off power of the instrument.

    Empty the constant head by opening the grab sample tap.

    5.3 Cleaning of Sample Supply/Filters

    Depending on the water quality, the sample supply and its filter mayneed cleaning from time to time.

    5.4 Cleaning of Protective Filter 

    If the protection filter shows deposits, proceed as follows:

    Close flow regulating valve.

    Stop the sample supply before the filter.

    Screw out the filter vessel.

    Screw out the filter (must be held at the top).

    Backwash the filter under pressure of tap water. Clean theoutside of the filter.

    Mount the filter and filter vessel again.

    Open sample supply and flow regulating valve again.

    Weekly Check sample supply for dirt.

    Check sample flow.

    Make manual measurement. If necessary, correct.

    Monthly Exchange Diisopropylamine bottle if there is lessthan 150 ml (half of the diffusion tube has to be

    covered with Diisopropylamine).

     Yearly Exchange diffusion tubes

  • 8/18/2019 AMI Hydrazine

    33/84

    AMI HydrazineMaintenance

     A-96.250.501 / 070606 31

    5.5 Maintenance of Hydrazine Sensor Important

    Run-in Period After cleaning, the sensor is very sensitive. Do not correct thevalue! Let the instrument run in for at least 24 h! 

    Cleaning Shut down the instrument as

    described in chapter Stop Operation beforeMaintenance, p. 30.

    Take off the sensor cable and putit to the side to prevent

    wetting it. Unscrew the two white knurled

    fastening screws.

    Attention:

    The sensor will glide easily out of the cell. Do not let it drop to thefloor. Any mechanical damage will make it unfit for further use. Donot touch the platinum parts with your fingers or metallic objects.

    Take the hydrazine sensor out of the flow cell.

    There is always some water left in the sensor chamber. Whentaking off the hydrazine sensor, this water will pour down. Preventthe connector on the cable and on the sensor from getting wet.

    Take off the rotor.

    Clean the two orifices with a pipe cleaner or a toothpick.

    Clean the rotor with a pipe cleaner. Wipe the sensor with a softtissue, mainly the platinum parts and the whole area which is incontact with water.

    Rinse all parts thoroughly with clean water.

    Put rotor on the sensor.

    Mount sensor. Fasten screws hand-tight.

    Connect sensor cable.

    Open sample flow.

     As soon as the rotor is turning, switch on power.

  • 8/18/2019 AMI Hydrazine

    34/84

    32  A-96.250.501 / 070606

    AMI HydrazineMaintenance

    5.6 Maintenance of Reference ElectrodeCleaning Shut down the instrument as described in chapter Stop Operation

    before Maintenance, p. 30.

    Disconnect the electrode from the cable

    Loosen the fixing screw and pull out the electrode of the samplechamber.

    Wipe sensor tip carefully with a soft tissue.

    Grease and oily films only may cause problems: Clean the tipcautiously with alcohol using a soft tissue. Rinse with water.

    Do not use acid! Remount the electrode into the sample chamber, push it down

    till it stops and connect the electrode to the cable.

    5.7 Changing Diisopropylamine Bottle

    Shut down the instrument as described in Stop Operation beforeMaintenance, p. 30

    Attention! Wear safety gloves and safety glasses! Please note

    the warnings in the safety data sheet of Diisopropylamine! Donot inhale!

    Put on safety gloves and safety glasses!

    Fill roughly 750 ml high purity water into a 1 l beaker.

    Screw off the almost empty Diisopropylamine bottle. Take outdiffusion tube. Close bottle immediately.

    Dip the diffusion tube into the beaker (filled with high puritywater). Rinse some seconds. Take out.

    Open new Diisopropylamine bottle, add 20 ml of high puritywater and insert diffusion tube. Fasten bottle to holder.

    (If you fill up Diisopropylamine into the almost empty bottle, nohigh purity water needs to be added).

    The liquid in the almost empty Diisopropylamine bottle must bedischarged as chemical waste.

  • 8/18/2019 AMI Hydrazine

    35/84

    AMI HydrazineMaintenance

     A-96.250.501 / 070606 33

    5.8 Changing the Diffusion TubeShut down the instrument as described in chapter Stop Operationbefore Maintenance, p. 30.

    Attention! Wear safety gloves and safety glasses! Please notethe warnings in the safety data sheet of Diisopropylamine! Donot inhale!

    Put on safety gloves and wear safety glasses!

    Fill roughly 750 ml high purity water into a 1 l beaker.

    Screw off Diisopropylamine bottle. Take out diffusion tube.

    Close bottle immediately. Dip the diffusion tube into the beaker filled with high purity

    water. Rinse some seconds. Take out.

    Take off the old diffusion tube.

    Slide on the new diffusion tube.

    Open Diisopropylamine bottle, insert diffusion tube, and fastenbottle to the holder.

    The correct measuring value will appear after 30 min of continuousoperation (flow and power on). Diisopropylamine needs roughly30 min to penetrate the walls of the new diffusion tube.

  • 8/18/2019 AMI Hydrazine

    36/84

    34  A-96.250.501 / 070606

    AMI HydrazineMaintenance

    5.9 Maintenance of Flow CellFor technical details see the component drawing on the next page.

    Attention!

    Never use organic solvents or scrubbing materials to clean acrylicglass parts.Use soft detergent and rinse well.

    Cleaning   Shut down the instrument as described in section Stopoperation before maintenance.

    Remove all sensors. See also drawing on the following page.

    Put the rubber cap on the tip of the reference electrode andplug cap on sensor plug. For longer interruptions see LongerStop of Operation, p. 36.

    Take off the protection filter, the tabs and the fittings. Do notallow water on the sensor plugs!

    Loosen fastening screws, take off the cover of the constanthead, and take out all tubes. Remove also the the tubes,leading to the Diisopropylamine bottle.

    The flow cell block must not be taken from the panel during

    cleaning.  All acrylic parts are cleaned with a soft brush (bottle cleaner)

    using soapy water.

     All gasket have to be replaced before reassembling the flowcell. A film of teflon paste (e.g. Fomblin from Solvay Solexis) onthe gaskets improves tightness and life time. Wrap teflon tapearound the threads of the sample inlet and outlet.

    The tube, leading to the

    diisoproplyamine bottle, is

    pushed into the hole before the

    small overflow tube. The tube

    coming from the

    diisopropylamine bottle is put

    into the small overflow tube.

    Bottle

    holder 

    fromflow cell

    toflow cell

    into smaller over-

    flow tube (closest

    to stainless steel

    panel

  • 8/18/2019 AMI Hydrazine

    37/84

    AMI HydrazineMaintenance

     A-96.250.501 / 070606 35

    Component Drawing

    cover 

    outer tube

    smaller overflow tube

    longer overflow tube

    hydrazine sensor chamber 

    tap / outlet for 

    tap of sample inlet

    sample

    inlet

    hydrazine sensor 

    rotor 

    filter 

    filter vessel

    sampleoutlet

    to diisoprop. bottle

    temp. sensor 

    from diisoprop. bottle

    flowsensor 

    fastening ref. electrode

    grab sample

    flow cell block

  • 8/18/2019 AMI Hydrazine

    38/84

    36  A-96.250.501 / 070606

    AMI HydrazineMaintenance

    5.9.1 Longer Stop of Operation

    Do not switch the instrument off if your operation is suspended forless than a week. Power consumption is very low, and the sensorsremain ready for use.

    Do not spill water on connectors! Ensure they remain dry!

    Shut-down the instrument as decribed in chapter Stop Operationbefore Maintenance, p. 30.

    Switch off power of all connected devices.

    Remove filter vessel, empty, and remount.

    Dismount the reference electrode, fill water in the rubber capand put the cap on the electrode tip to preserve it from dryingout! Disconnect from cable and put plug cap on the electrodeplug. Store in dry and frost protected place with tip pointingdownwards.

    Remove cable from hydrazine sensor. Dismount sensor, dry itwith a soft, clean tissue, and store dry.

  • 8/18/2019 AMI Hydrazine

    39/84

    AMI HydrazineCalibration

     A-96.250.501 / 070606 37

    6. Calibration

    6.1 Calibration Procedure

    ManualMeasurement

    Hydrazine reacts with dimethylaminobenzaldehyde in an acidsolution and forms a yellow colour. The colour intensity isproportional to the concentration and can be determined by aphotometer. Refer to an appropriate standard method.

    Correction The adjustment of the slope is called correction. The zero remains

    unchanged. A calibration (changing slope and zero) is notnecessary. The zero point is very stable and no change is

    necessary.

    6.2 Hydrazine Sensor 

    ProcessCalibration

    The sample must always overflow into the longer overflow tube ofthe constant head. Only perform process calibration if thedifference is significant.

    Grab Sample Grab sample directly on the flow cell (grab sample valve and tap).Open the grab sample valve and let the water flow out for a minutebefore taking the sample.

    Note the displayed value of the AMI Hydrazine while taking the grabsample. The reading has to be stable .

    Determine the hydrazine concentration of the sample by manualanalysis.

    Compare the result to the reading of the AMI.If the reading of the AMI has changed while performing the manual

    analysis, enter the difference between the reading of the AMI.

    For example:Displayed measurement value at AMI during grab sample: 10 ppbDisplayed measurement value at AMI after concentrationdetermination: 15 ppb.Determined hydrazine concentration: 8 ppbThere is a change of the measurement value of 50% during theconcentration determination. This means, you have to correct thedetermined hydrazine concentration also with a factor of 50%.This will result in a value of 8ppm+ (50% of 8 ppb) = 12ppb

    Finally, you have to enter 12ppb as a process value at the AMItransmitter 

  • 8/18/2019 AMI Hydrazine

    40/84

    38  A-96.250.501 / 070606

    AMI HydrazineCalibration

    During calibration, control is interrupted. The signal outputs arefrozen if hold after calibration has been programmed (menu4.3.4.2). Otherwise the outputs track the measuring value. Hold

    after calibration is indicated by Hold  in the display.

    Enter Process Hydrazine

    Current Value 18.8 ppbRaw value x n A

    Save Process Value 24.0 ppb

    3.1.2.5

    3.1.2.5Process Hydrazine

    Current Value 15.0 ppbRaw value x n A

    Calibration successfulpress Enter 

    3.1.2.5Process Hydrazine

    Current Value 18.8 ppbRaw value x n A

    Process Value 24.0 ppb

    Save

    Enter 

    Process Hydrazine

    Raw value x n A

    Save Process Value 18.8 ppb

    3.1.2.5Enter 

    Current Value 18.8 ppb

    3.1

    CalibrationSimulation

    Maintenance

    Set Time 01.01.05 16:30:00

    3.1.2CalibrationEnter  Zero Hydrazine

    Process HydrazineCalibration

    Enter correct value with

    arrow keys.

  • 8/18/2019 AMI Hydrazine

    41/84

    AMI HydrazineCalibration

     A-96.250.501 / 070606 39

    PossibleError 

    Message

    Current value too low or no sample flow. Possible reasons:

    For additional explanations see Trouble Shooting, p. 40 andError List, p. 41

    ZeroHydrazine

     A zero point calibration is NOT necessary.

    If your quality procedure specifies a zero point calibration, pleaseproceed as described below:

    During a zero calibration, the sample flow is turned off and theresidual hydrazine in the small water volume around the sensor isconsumed within 15-20min. 0 ppb remains and after 30min., a zerocalibration is done.

    Let the instrument run continuously for AT LEAST 5 DAYS innormal operation before doing a zero point calibration!

    Choose Menu 3.1.3. The instrument guides you through thecomplete calibration process. If you have completed the requiredaction press to proceed.

    Close tap of the water inlet. Wait 30 min. (Timer counts down).

    Current value / offset (Progress of zero cal. is shown). Wait untilfinished.

    Open tap of the water inlet and regulate flow.

    Sample

    flow low

    Low sample flow

    resulting in no signal

    difference to zero.

    Check rpm on main display.

    Process

    value too

    low

    ppb-value is too low to

    provide enough

    current difference from

    the calibrated zero

    current.

    Process value used for

    calibration should be higher.

    Check the diagnostic value

    of the zero point.

    Contami-

    nated

    sensor 

    Sensor does not pro-

    vide enough current

    for the value measured

    with the manual analy-

    sis.

    Clean sensor. In case of

    repeated sensor

    contamination, check for

    water treatment chemicals

    such as phosphates.

  • 8/18/2019 AMI Hydrazine

    42/84

    40  A-96.250.501 / 070606

    AMI HydrazineTrouble Shooting

    7. Trouble ShootingThis chapter provides some hints to make trouble shooting easier.For any detailed information how to handle/clean parts please seeMaintenance, p. 30.For any detailed information how to program the instrument pleasesee Display, Keys, p. 14.

    Attention: The sample for the manual analysis must be takendirectly from the flow cell!

    If you need further help please contact your dealer. Note serial

    number of instrument and all diagnostic values before.

    7.1 Error messages

    The menu 1 - Messages - reveals pending errors as well as themessage list with text, time and state of errors that have occuredbefore.

    Error

    Nonfatal error. The system continues to operate normally, except

    for the defective part.

    Fatal error   (blinking symbol)

    If used for control relay 1, 2 as well as the Signal Outputs aredeactivated, i. e. any control is interrupted. The indicated measuredvalues are incorrect. Depending on the error the system may stopoperating.

    To solve the problem, please consult Error List, p. 41

    If an error is corrected, by intervention or by the system itself, it islisted in the the Message List (Menu 1.2). The symboldisappears.

    The last 64 messages are stored for viewing. They are deleted ifSET Default Values (Menu 5.4.2) is carried out.

    1.1.5

    Pending ErrorsError Code E010

    Sample Flow low

    to Acknowledge

    Errors remain in the Pending

    Errors window until corrected,even if acknowledged.

     Acknowledging the error onlydeactivates the Alarm Relay, i.e.it switches off an alarm device ifavailable.

  • 8/18/2019 AMI Hydrazine

    43/84

    AMI HydrazineTrouble Shooting

     A-96.250.501 / 070606 41

    7.2 Error ListE0xx (bold and red) are fatal errors.

    Error Error text Corrective action

    E001 Hydrazine Alarm high check process

    E002 Hydrazine Alarm low check process

    E007 Sample Temp. high check sample temperature

    E008 Sample Temp. low check sample temperature

    E009 Sample Flow high check sample inlet pressure,

    check sample flow, check alarm

    value in Installation/Relays/

     Alarm relay/Sample

    E010 Sample Flow low establish sample flow, check

    sample inlet pressure, check

    alarm value in Installation/

    Relays/Alarm relay/Sample,clean instrument

    E011 Temp. shorted check wiring of temperature sen-

    sor, check temp. sensor 

    E012 Temp. disconnected check wiring of temperature sen-

    sor, check temp. sensor 

    E013 Case Temp. high check environment temperature

    E014 Case Temp. low check environment temperature

    E015 Reference electrode Check conductivity value of

    sample (must be > 5 μS/cm)

    Check gain of hydrazine sensor.

    If too small, clean hydrazine

    sensor.

    Exchange reference electrode.

    E017 Control time-out check control device or program-

    ming in Installation, Relay con-

    tact, Relay 1/2

  • 8/18/2019 AMI Hydrazine

    44/84

    42  A-96.250.501 / 070606

    AMI HydrazineTrouble Shooting

    E024 Input active Information that the Input is

    active (see programming Instal-

    lation, Input, Fault “Yes”)

    E025 IC MK41T56 call service

    E026 IC LM75 call service

    E027 IC PCF8574 call service

    E028 EEProm Microcon call service

    E029 EEProm Motherboard call service

    E030 EEprom Front-End call service

    E031 Calibration RecOut call service

    E032 Wrong Front-End call service

    E033 Power-on none, status message

    E034 Power-down none, status message

    Error Error text Corrective action

  • 8/18/2019 AMI Hydrazine

    45/84

    AMI HydrazineProgramm Overview

     A-96.250.501 / 070606 43

    A. Programm Overview

    Menu Messages may be protected by a password. Menu Diagnostics is always

    accessible for everybody. No password protection. In both menus, no settings can be

    modified.

    Pending Errors 1.1*Messages Message List 1.2*(Main Menu 1)

    Designation AMI Hydrazine 2.1.1*Version 3.81-04/07 2.1.2*

    Instrument 2.1.3.1*Identification Factory Test Motherboard 2.1.3.2*2.1*  2.1.3*  Frontend 2.1.3.3*

    Years 2.1.4.1*Operating Time Days 2.1.4.2*2.1.4*  Hours 2.1.4.3*

    Minutes 2.1.4.4*Seconds 2.1.4.5*

    Current Value 2.2.1.1*Raw Value 2.2.1.2*

    Hydrazine Sensor Ref. Voltage 2.2.1.3*

    Sensors 2.2.1*  Cal. History 2.2.1.5*2.2*

    Miscellaneous Case Temp. 2.2.2.1*Diagnostics 2.2.2*(Main Menu 2)

    Sample ID  2.3.1*Temperature (°C) 2.3.2*

    Sample Nt5k (Ohm) 2.3.3*2.3*  Sample Flow (rpm) 2.3.4*

    Raw Value (Hz) 2.3.5*

     Alarm Relay 2.4.1*

    Relay 1 2.4.2*I/O State Relay 2 2.4.3*2.4*  Input 2.4.4*

    Signal Output 1/2 2.4.5/2.4.6*

    Interface Protocol 2.5.1*2.5* Device address/Baud rate 2.5.2* (depends on

      Baud rate/ID No. 2.5.3* chosen protocol)  Local operation/Parity 2.5.4*

    * Menu numbers

  • 8/18/2019 AMI Hydrazine

    46/84

    44  A-96.250.501 / 070606

    AMI HydrazineProgramm Overview

    Menu Maintenance is for service: Calibration, simulation of outputs and set time/

    date. Please protect with password.

    Menu Operation is for the user , allowing to set limits, alarm values, etc. The

    presetting is done in the menu Installation (only for the System engineer). Please

    protect with password.

    Calibration Zero Hydrazine 3.1.1*3.1*  Process Hydrazine 3.1.2*

     Alarm Relay 3.2.1*Maintenance Simulation Relay 1/2 3.2.2/3.2.3*(Main Menu 3) 3.2*  Signal Output 1/2 3.2.4/3.2.5*

    Set Time3.3*

      Sensors Filter Time Const. 4.1.1*4.1* Hold after Cal. 4.1.2*

     Alarm Relay Alarm Hydrazine 4.2.1.1*4.2.1*

    Operation Relay Contacts Relay 1/2 Limit value depends on pre-(Main Menu 4) 4.2* 4.2.2/4.2.3* or  setpoint setting in Menu 5:

    or  timer  Installation

    Input 4.2.4* depends on presetting in Menu 5: Installation

    Logger Log Interval 4.3.1*4.3* Clear Logger 4.3.2*

    * Menu numbers 

  • 8/18/2019 AMI Hydrazine

    47/84

    AMI HydrazineProgramm Overview

     A-96.250.501 / 070606 45

    Menu Installation: Defining assignment of all inputs and outputs, measuring param-

    eters, interface, passwords, etc. Menu for the system engineer. Password stronlgy

    recommended

    Sensors Temp. Compensation 5.1.1*5.1*

    HydrazineParameter Temperature

    Signal Output 1 5.2.X.1* Sample FlowSignal Outputs Signal Output 2 5.2* Signal Output 3 Current Loop (0/4 – 20 mA) 5.2.X.2*

    (only with optional

    3rd

     Signal Output)  Function 5.2.X.3*5.2.1-5.2.3*

      Scal./Contr.par. (depends on chosen Function)5.2.X.4*

     Alarm high

     Alarm Hydrazine  Alarm low  5.3.1.1* Hysteresis

    Delay

    Flow alarm

     Alarm Relay Sample Flow Flow alarm high5.3.1* 5.3.1.2* Flow alarm low

    Sample temp 5.3.1.3* Temp. highTemp. low

    Case Temp. high 5.3.1.4*Installation  Case Temp. low 5.3.1.5*(Main Menu 5) 

    Function 5.3.2.1*Relay Contacts Relay 1/2 Parameter 5.3.2.2*5.3* 5.3.2/5.3.3* Limit/Set contr./Timer (depends on chosen Function)

    5.3.2.3*

     Active 5.3.4.1*Input Signal Outputs 5.3.4.2*5.3.4*  Output/Control 5.3.4.3*

    Fault 5.3.4.4*Delay 5.3.4.5*

    Language 5.4.1*Miscellaneous Set defaults 5.4.2*5.4*  Load Firmware 5.4.3*

    Password 5.4.4* (for Messages, Maint., Operation, Installation)Sample ID 5.4.5*

    Interface Protocol 5.5.1* (only with optional additional board)

    5.5*  Device Address/Baud Rate 5.5.2*Baud Rate/ID No. 5.5.3* (depends on chosen protocol)

      Local operation/Parity 5.5.4*

    * Menu numbers 

  • 8/18/2019 AMI Hydrazine

    48/84

    46  A-96.250.501 / 070606

    AMI HydrazineProfibus (Option)

    B. Profibus (Option)

    B.1. Introduction

    The AMI family of transmitters provides a common base for anumber of analytical instruments with application specific sensorinterfaces. Each instrument has a basic structure with identicalfunctionality.

     AMI instruments can (optionally) be equipped with a PROFIBUSDP-V1 interface. The interface supports the PROFIBUS-PA Profile3.0. This part of the manual describes the functionality of theinterface and the integration of the AMI family of transmitters intoan automation system.

    Hardware Wiring, bus terminals and connectors are not part of this manual. It

    should, however, be noted that the observation of the

    corresponding standards is of paramount importance to proper

    operation. Setting the transmission rate to reasonable values will

    help avoid communication errors.

    GSD file for AMI

    Hydrazine

    The PROFIBUS system requires a description of the device

    parameters, e. g. output data, input data, data format, data volume

    and supported transmission rate, so that it can integrate the field

    devices into the bus system. These data are contained in a

    Device Master File (GSD file) which is placed at the disposal of

    the PROFIBUS master while communication system is

    being commissioned.

    Device bitmaps can also be integrated. These appear as icons inthe network tree.

    The manufacturer-specific GSD file SW17C6.gsd is available for

    download from http://www.swan.ch

    http://www.swan.ch/http://www.swan.ch/http://www.swan.ch/

  • 8/18/2019 AMI Hydrazine

    49/84

    AMI HydrazineProfibus (Option)

     A-96.250.501 / 070606 47

    B.2 Cyclic Data Telegram AMI Hydrazine makes the following modules available as input datafor the cyclic data exchange:

    1 Main Process Value

    2 Main Temperature

    3 Main Flow

    4 Control AMI

    5 Diagnostic Values (5 values)

    a Raw value in nA

    b Counter Electrode Voltage in mV

    c Offset in nA

    d Slope in nA

    e Case Temperature in °C

    The input data is transferred from AMI Hydrazine in the followingstructure :

    Table 1: Input data structure of AMI Hydrazine

    Indexinputdata

    Data AccessData format /comments

    Configuration data

    0 ... 4 Analog Input block 1Main Process Value

    read Measured value (32-bitfloating point number)Status byte

    0x42, 0x84, 0x08, 0x05 or0x42, 0x84, 0x81, 0x81 or0x94

    5 ... 9 Analog Input block 2Main Temperature

    read Measured value (32-bitfloating point number)Status byte

    0x42, 0x84, 0x08, 0x05 or0x42, 0x84, 0x81, 0x81 or0x94

    10 ... 14 Analog Input block 3

    Main Flow

    read Measured value (32-bit

    floating point number)Status byte

    0x42, 0x84, 0x08, 0x05 or

    0x42, 0x84, 0x81, 0x81 or0x94

    15 ... 34 Diagnostic Values read 5 Measured values (32-bit floating pointnumbers)

    0x45, 0x93, 0x08, 0x08, 0x08,0x08, 0x08, 0xE9

  • 8/18/2019 AMI Hydrazine

    50/84

    48  A-96.250.501 / 070606

    AMI HydrazineProfibus (Option)

    The output data is transferred to AMI Hydrazine in the following structure:

    B.3 Customizing the Cyclic Data Telegram

    You can customize the cyclic telegram to better meet the

    requirements of a process. The tables above represent the

    maximum contents of the cyclic data telegram. If you do not want to

    use all the cyclic data of AMI Hydrazine, you can use the device

    configuration (Chk_Cfg) to eliminate individual data blocks from the

    cyclic telegram via the PLC software. Shortening the telegram

    improves the data throughput rate of a PROFIBUS system. You

    should only keep those blocks active which you process further inthe system. You can do this by means of a "negative" selection in

    the configuration tool.

    To achieve the correct structure of the cyclic data telegram, the

    PROFIBUS master must send the identification FREE_PLACE

    (00h) for the non-active blocks.

    Indexoutputdata

    Data AccessData format /

    commentsConfiguration data

    0 ... 1 Control AMI (SP_D) write Byte 0

    Bit Description

    0 Relay 10 = open1 = close

    1 Relay 20 = open1 = close

    2..7 Not used

    Byte 1

    Bit Description

    0 Hold

    1 Control Off

    2..7 Not used

    0x82, 0x81, 0x05, 0x05 or0x82, 0x81, 0x84, 0x82 or0xA1

  • 8/18/2019 AMI Hydrazine

    51/84

    AMI HydrazineProfibus (Option)

     A-96.250.501 / 070606 49

    B.4 Data FormatsEach measured value has the same format: one four-byte (32-bit)

    floating-point value in IEEE 754 Short Real Number format and a

    status code byte referring to the measured value. There are four

    quality states, 16 sub-status values (not all defined) for each quality

    state, and four limit states for the measured value indicated in its

    associated status byte (See Figures 2 ,3, 4 and 5).

    Figure 2: Out Value + Status

    Figure 3: IEEE 754 floating point number:

    Formula: Value = (-1) sign * 2(exponent - 127) * (1 + mantissa)Example: 40 F0 00 00 h = 0100 0000 1111 0000 0000 0000 0000 0000 b

    Value = (-1)0 * 2(129 - 127) * (1 + 2-1 + 2-2 + 2-3)= 1 * 22* (1 + 0.5 + 0.25 + 0.125)= 1 * 4 * 1.875= 7.5

    Figure 4: Status Code Byte

    Byte 1 Byte 2 Byte 3 Byte 4 Byte 5 Analog values in IEEE 754 Format Status

    Byte n Byte n+1 Byte n+2 Byte n+3Bit 7 Bit 6 Bit 0 Bit 7 Bit 6 Bit 0 Bit 7 Bit 0 Bit 7 Bit 0

    Sign 27 2

    6 2

    5 2

    4 2

    3 2

    2 2

    12

    02

     –1 2

     –2 2

     –3 2

     –4 2

     –5 2

     –6 2

     –72

     –8 2

     –9 2

     –10 2

     –11 2

     –12 2

     –13 2

     –14 2

     –152

     –16 2

     –17 2

     –18 2

     –19 2

     –20 2

     –21 2

     –22 2

     –23

    Exponent Mantissa Mantissa Mantissa

    7 6 5 4 3 2 1 0

    Quality Sub-status Limits Meaning

    0 0 bad

    0 1 uncertain

    1 0 good

    0 0 ok

    0 1 low limited

    1 0 high limited

    The sub-status hasa different quality

    depending on the coding

    1 1 constant

  • 8/18/2019 AMI Hydrazine

    52/84

    50  A-96.250.501 / 070606

    AMI HydrazineProfibus (Option)

    Figure 5: Status codes of the measured value

    Figure 6: DP Diagnostics

    Statuscode

    Device status Meaning Limits

    0x08 BAD Not connected (Temp Sensor notconnected)

    0x0C BAD device failure

    0x10 BAD sensor failure (Temp Sensor ShortCircuit)

    0x47 UNCERTAIN last usable value (HOLD) CONST

    0x50 UNCERTAIN sensor conversion not accurate (No

    Flow)

    OK

    0x640x67

    UNCERTAIN sensor calibrationsensor calibration, hold

    OKCONST

    0x80 GOOD ok OK

    0x890x8A

    GOOD Warning LowWarning High

    LOW_LIMHIGH_LIM

    0x8D0x8E

    GOOD Alarm Low Alarm High

    LOW_LIMHIGH_LIM

    Byte Description

    0 Status 11 Status 22 Status 33 Master address4 Ident number (high byte)5 Ident number (low byte)

    Standard DP Diagnostics

    6 Header (block length incl. Header byte) = 147 Status Type (0x81)8 Slot Number (0x00)9 Status Specifier

    Status coding accordingto DP/V1

    10 ..13 Diagnosis14 ..19 Diagnosis Extension

  • 8/18/2019 AMI Hydrazine

    53/84

    AMI HydrazineProfibus (Option)

     A-96.250.501 / 070606 51

    The diagnosis object consists of 4 Bytes

    Byte Bit Meaning when “1” Abbreviation

    0 0 Hardware failure electronics DIA_HW_ELECTR

    1 Hardware failure mechanics DIA_HW_MECH

    2 -

    3 Electronic temperature too high DIA_TEMP_ELECTR

    4 -

    5 Measurement failure DIA_MEASUREMENT

    6 -

    7 -

    1 0 -

    1 -

    2 -

    3 -

    4 -

    5 Maintenance required DIA_MAINTENANCE

    6 -

    7 -

    2 0 -

    1 -

    2 -

    3 -4 -

    5 -

    6 -

    7 -

    3 0 -

    1 -

    2 -

    3 -

    4 -

    5 -

    6 -

    7 Extension Available EXTENSION_AVAILABLE

  • 8/18/2019 AMI Hydrazine

    54/84

    52  A-96.250.501 / 070606

    AMI HydrazineProfibus (Option)

    The diagnosis extension object consists of 6 Bytes.

    Bytes 4 and 5 are reserved for future extensions.

    Byte Bit Meaning when “1”

    0 0 Hyd. Alarm high -

    1 Hyd. Alarm low -

    2 - -

    3 - -

    4 - -

    5 - -

    6 Sample Temp. high DIA_MEASUREMENT

    7 Sample Temp. low DIA_MEASUREMENT

    1 0 Sample Flow high DIA_MEASUREMENT

    1 Sample Flow low DIA_MEASUREMENT

    2 Temp. shorted DIA_HW_ELECTR

    3 Temp. disconnected DIA_HW_ELECTR

    4 Case Temp. high DIA_TEMP_ELECTR

    5 Case Temp. low DIA_TEMP_ELECTR

    6 Counter Electrode Reference DIA_MEASUREMENT

    7 -

    2 0 Control Timeout -

    1 - -

    2 - -

    3 - -

    4 - -

    5 - -

    6 Cleaning Reagent DIA_MAINTENANCE

    7 Input Active DIA_HW_ELECTR

    3 0 IC MK41T56 DIA_HW_ELECTR

    1 IC LM75 DIA_HW_ELECTR

    2 IC PCF8574 DIA_HW_ELECTR

    3 EEprom Microcon DIA_HW_ELECTR

    4 EEprom Motherboard DIA_HW_ELECTR

    5 EEprom Front-end DIA_HW_ELECTR

    6 Cal. Recout DIA_HW_ELECTR

    7 Wrong Front-end DIA_HW_ELECTR

  • 8/18/2019 AMI Hydrazine

    55/84

    AMI HydrazineProfibus (Option)

     A-96.250.501 / 070606 53

    B.5. ConfigurationB.5.1 Device Master Files (GSD)

    The device is ready for system integration once commissioning has

    been effected via the local display or the Class 2 master. The

    PROFIBUS system requires a description of the device

    parameters, e.g. output data, input data, data format, data volume

    and supported transmission rate, so that it can integrate the field

    devices into the bus system. These data are contained in a Device

    Master File (GSD file) which is placed at the disposal of thePROFIBUS master while the communication system is being

    commissioned. Device bitmaps can also be integrated. These

    appear as icons in the network tree.

    The Profile 3.0 Device Master File (GSD) allows field devices from

    various manufacturers to be exchanged without having to

    reconfigure.

    B.5.2 GSD File Types

    Prior to configuration, decide which GSD you want to use tooperate the system.

    You can change the setting by means of a Class 2 master (under

    Physical Block-Parameter Ident_Number_Selector).

    The following two Device Master Files with different functionalities

    are available:

      Manufacturer-specific GSD with Profile 3.0 functionality

    (default setting):

    This GSD guarantees the unlimited functionality of the fielddevice. Device-specific process parameters and functions are

    therefore available.

    Profile GSD's:

    If a system is configured with profile GSD's, it is possible to

    exchange devices that are supplied by various manufacturers. It

    is, however, essential that the cyclic process values follow the

    same sequence. The transmitter supports the profiles Analyzer

    and Multivariable (IEC 61158-2). These GSD's comprise Analog

    Input blocks (AI).

  • 8/18/2019 AMI Hydrazine

    56/84

    54  A-96.250.501 / 070606

    AMI HydrazineProfibus (Option)

    GSD files for AMI Hydrazine:

    Working with GSD files 

    The GSDs must be integrated into the automation system.

    Depending on the software that is being used, the GSD files can becopied to the program-specific directory or can be read into thedatabase using the import function within the configurationsoftware.

    B.6. Acyclic Data Communication

    The AMI Series of analytical instruments is available for a widerange of parameters. Instrument configuration depends onconnected sensors and application. Therefore, configuration mustbe performed by trained personnel with a background inanalytical methods.

    It should be noted that there is a clear distinction between

    instruments that are configured and calibrated in an application(flow, level, pressure) and instruments that are calibrated with aknown standard (pH, chlorine). AMI instruments are of thesecond type.

    To ensure reliability and consistency of the measurement, remoteconfiguration has not been enabled at this time.

    Device name Ident_number_

    Selector

    ID number GSD Bitmaps

    Manufacturer-specific GSD with Profile 3.0 functionality:

     AMI Hydrazine 1 17C6Hex SW17C6.gsd SW17C6_D.bmpSW17C6_N.bmpSW17C6_D.bmp

    Profile 3.0 GSD Analyzer:

     AMI Hydrazine 0 9750 Hex PA139750.gsd PA_9750n.bmp

    Profile 3.0 GSD Multivariable:

     AMI Hydrazine 3 9760 Hex PA139760.gsd PA_9760n.bmp

  • 8/18/2019 AMI Hydrazine

    57/84

    AMI HydrazineModbus (Option)

     A-96.250.501 / 070606 55

    C. Modbus (Option)

    C.1. Introduction

    The AMI Hydrazine is a Microprocessor controlled electronic

    transmitter & controller for the conductivity measurement in power

    cycles. It can be operated as an intelligent Slave device on the

    Modbus/JBUS using twisted pair RS485 lines.

    Modbus/Jbus The Modbus / Jbus Protocol communicates using a master-slave

    technique, in which only one device (the master) can initiatetransactions (queries). The other devices (the slaves) respond by

    supplying the requested data to the master, or by taking the action

    requested in the query. Typical master devices include host

    processors, PLCs etc. Typical slaves can be PLCs or intelligent

    transmitter/controller like the AMI family of the instruments.

    The Query The function code in the query tells the addressed slave device

    what kind of action to perform. The data bytes contain any

    additional information the slave will need to function. For example,

    function code 03 will query the slave to read holding registers andrespond with their contents. The data field must contain the

    information telling the slave which register to start at and how many

    registers to read.

    TheResponse

    If the slave makes a normal response, the function code in the

    response is an echo of the function code in the query.

    The data bytes contain the data collected by the slave.

    Serial

    TransmissionModes

    Modbus supports two transmission modes: ASCII or RTU.

    The AMI Instruments support Modbus RTU (Remote Terminal Unit)mode only.

    RTU Mode In RTU mode, each 8-bit byte in a message contains two 4-bit

    hexadecimal characters. The main advantage of this mode is that

    its greater character density allows better data throughput than

     ASCII for the same baud rate. Each message must be transmitted

    in a continuous stream. Users select the desired serial port

    communication parameters (baud rate, parity mode, etc). The serial

    parameters must be the same for all devices on a Modbus network.

  • 8/18/2019 AMI Hydrazine

    58/84

    56  A-96.250.501 / 070606

    AMI HydrazineModbus (Option)

    Coding System

    Eight bit binary

    1 start bit

    8 data bits, least significant bit sent first

    1 bit for even / odd parity, no bit for no parity

    1 stop bit if parity is used, 2 bits if no parity

    Cyclic Redundancy Check (CRC)

    RTU framing: messages start with a silent interval of at least 3 ½character times. A similar interval of at least 3 ½ character timesmarks the end of the message.

     A typical message frame is shown below

    StartSlave

     AddressFunction Data CRC END

    T1-T2-T3-T4 8 Bits 8 Bits n x 8 Bits 16 Bits T1-T2-T3-T4

  • 8/18/2019 AMI Hydrazine

    59/84

    AMI HydrazineModbus (Option)

     A-96.250.501 / 070606 57

    C.2. Data and Control FunctionsData Address Data Addresses in Modbus Messages

     All data addresses in Modbus Messages are referenced to zero.The first occurrence of a data item is addressed as item numberzero. For example:

    Holding Register 40001 is addressed as register 0000 in thedata address field of the message. The function code fieldalready specifies a holding register operation. Therefore the 4xreference is implicit.

    FunctionCodes

    Modbus Function CodesThe list below shows the function codes supported by the AMIInstruments. Codes are listed in decimal. Function codes notsupported return an exception response 02 to the master.

    Table 1: Modbus function codes

    Data Re-presentation

    Data Representation

    The process- and parameter data of the AMI Instruments arerepresented by float (IEEE 754), signed- or unsigned integervalues. Depending on the range of the value, float, integer8,unsigned8, integer16, unsigned16, integer32 or unsigned32representation is used.

    Code Name Description02 Read Input Status Read Alarm & Status Bits

    03 Read Holding Registers Read Parameter Data

    04 Read Input Registers Read Process Data

    05 Force Single Coil Set Relays, fake Keystrokes etc.16 Preset Multiple Registers Preset Parameter Data (SWAN Use only)

  • 8/18/2019 AMI Hydrazine

    60/84

    58  A-96.250.501 / 070606

    AMI HydrazineModbus (Option)

    Discrete Inputs

     Alarm &Status bits are packed as one input per bit of the data field.

    Definition: big-endian and little-endian

    Big-endian and little-endian are terms that describe the order inwhich a sequence of bytes are stored in computer memory. Big-endian is an order in which the "big end" (most significant value inthe sequence) is stored first (at the lowest storage address). Little-endian is an order in which the "little end" (least significant value inthe sequence) is stored first. For example, in a big-endian

    computer, the two bytes required for the hexadecimal number 4F52would be stored as 4F52 in storage. In a little-endian system, itwould be stored as 524F.

    Attention: big-endian format is used !!!

  • 8/18/2019 AMI Hydrazine

    61/84

    AMI HydrazineModbus (Option)

     A-96.250.501 / 070606 59

    C.3. Function DescriptionInput Status Read Input Status (Alarm &Status Bits), Function 02

    Up to 32 bits of Alarm are available as summarised in the tablebelow:

    Table 2: Alarm and Status addresses

    Addr. Type Description0  Alarm Hyd. Alarm high

    1  Alarm Hyd. Alarm low

    2  Alarm -

    3  Alarm -4  Alarm -

    5  Alarm -

    6  Alarm Sample Temp. high

    7  Alarm Sample Temp. low

    8  Alarm Sample Flow high

    9  Alarm Sample Flow low

    10  Alarm Temp. shorted

    11  Alarm Temp. disconnected

    12  Alarm Case Temp. high

    13  Alarm Case Temp. low

    14  Alarm Reference15  Alarm -

    16  Alarm Control Timeout

    17  Alarm -

    18  Alarm -

    19  Alarm -

    20  Alarm -

    21  Alarm -

    22  Alarm Cleaning Reagent

    23  Alarm Input Active

    24  Alarm IC MK41T56

    25  Alarm IC LM7526  Alarm IC PCF8574

    27  Alarm EEprom Microcon

    28  Alarm EEprom Motherboard

    29  Alarm EEprom Front-end

    30  Alarm Cal. Recout

    31  Alarm Wrong Front-end

  • 8/18/2019 AMI Hydrazine

    62/84

    60  A-96.250.501 / 070606

    AMI HydrazineModbus (Option)

    InputRegister 

    Read Input Registers (Process data): Function 04

    The process data are made up of the measuring values of the AMIHydrazine.

    Table 3: Input registers

    Single CoilFunction

    Force Single Coil: Function 05This function is used to set relays and reset the instrument. Relayscan only be forced if set to fieldbus function.

    Table 4: single coil function

    Read Holding Registers (Parameter data): Function 03

    Read Parameter data (see Parameter Table). All objects of typeRecord are defined in the Manufacturer specific tables following theparameter table.

    Note: The Address is used in a proprietary way to select thedesignated record.

    see parameter table on the next page.

    Byte-Offset

    Data type Description Range ofphysical value

    0 Float Hydrazine 0.. 1000 ppb

    4 Float Temperature -30 … 120 ºC

    8 Float Flow 0.. 600 rpm

    12 Float Counter Electrode Voltage mV

    16 Float Raw Hydrazine 0.. 100’000 nA20 Float Raw Temp. (NT5k) 150 …66’000

    Ω 

    24 Float Raw Flow 0.. 100 Hz

    28 Float Case Temperature -55... 125 ºC

    32 Float Signal Output 1 0.. 20 mA

    36 Float Signal Output 2 0.. 20 mA

    40 INSTRUMENT_ STATUS

    Instrument Display Status

    Addr. Description OFF (0) ON (1)

    0.. 1 Relay Contacts 1..2 open close

    2 Hold no yes

    3 Remote Off no yes

    4 Channel Select Channel 0 Channel 1

    15 Reset Instrument (like hardware reset) reset

  • 8/18/2019 AMI Hydrazine

    63/84

    AMI HydrazineModbus (Option)

     A-96.250.501 / 070606 61

    Table 5: parameter table

    Specific Data Manufacturer-specific Data Types & Records

    DATE_STAMP Format

    Address Modbus Addr Parameter Object Type Data Type Description Acc.

    2300 1151 Version Record FIRMWARE_REC Firmware Version r

    2400 1201 Num_Evt Simple Unsigned16 Number of Event Records r

    2402 1202 Num_Log Simple Unsigned16 Number of Log Records r

    3000 1501 Cal_History Array Array of CAL_HIST_1[15] Cal. History 1 r

    3001 1502 Cal_History Array Array of CAL_HIST_2[15] Cal. History 2 r

    4000.. 4126 2001.. 2064 Event_Rec Record EVENT_REC Get Event Record64 Records max.

    r

    5000.. 8070 2501.. 4036 Log_Rec Record LOG_REC Get Logger Record

    1536 Records max.

    r

    31.. 26 25..22 21..17 16..12 11.. 6 5.. 0

    YYYYYY MMMM DDDDD HHHHH MMMMMM SSSSSS

    Year0... 63

    Month1... 12

    Date1... 31

    Hour0... 23

    Minutes0... 59

    Seconds0... 59

  • 8/18/2019 AMI Hydrazine

    64/84

    62  A-96.250.501 / 070606

    AMI HydrazineModbus (Option)

      Parameter Size Data Type Description Range

    INSTRUMENT_STATUS 32 Record Instrument Display Status

    Language 1 Unsigned8 Language Selection 0 = German1 = English2 = French3 = Spanish

    RunMode 1 Unsigned8 Run Mode 0 = Run1 = Hold2 = Off3 = Grab

     AlaMode 1 Unsigned8 Alarm Mode 0 = Off1 = on2 = Blinking

    MaintMode 1 Unsigned8 Maintenance Mode 0 = Off1 = on2 = Blinking

    RelMode1 1 Unsigned8 Relais 1 Display Mode 0 = None1 = Limit Hi On2 = Limit Hi Off3 = Limit Lo On4 = Limit Lo Off5 = Timer Off6 = Timer On7 = Percent Control8 = Valve9 = Bus

    RelMode2 1 Unsigned8 Relais 2 Display Mode 0 = None

    1 = Limit Hi On2 = Limit Hi Off3 = Limit Lo On4 = Limit Lo Off5 = Timer Off6 = Timer On7 = Percent Control8 = Valve9 = Bus

    PercDos1 2 Unsigned16 Percent Dosage Relais 1 0.. 100 %

    PercDos2 2 Unsigned16 Percent Dosage Relais 2 0.. 100 %

    MaintValue 2 Unsigned16 Maintenance Value 0.. 100 %

     AlarmsActive 4 Unsigned32 Alarms Pending AlarmsActive

     AlarmsLatched 4 Unsigned32 Alarms no Acknowledged AlarmsLatched

     AlarmsActiveE 4 Unsigned32 Extended Alarms Pending AlarmsActiveE

     AlarmsLatchedE 4 Unsigned32 Extended Alarms not Acknowledged AlarmsLatchedE

    TimeStamp 4 DATE_STAMP Current Time

    end Record

  • 8/18/2019 AMI Hydrazine

    65/84

    AMI HydrazineProgramm List and Explanations

     A-96.250.501 / 070606 63

    D. Programm List and Explanations

    1 Messages

    1.1 Pending ErrorsProvides the list of active errors with their status (active,acknowledged). If an active error is acknowledged, the alarm relayopens again. Cleared errors are moved to the Message list.

    1.2 Message ListShows the error history: Error code, date and time of issue, andstatus (active, acknowledged, cleared).65 errors are memorized. Then the oldest error is cleared to savethe newest error (circular buffer).

    2 DiagnosticsIn diagnostics mode, the values can only be viewed, not modified.

    2.1 Identification

    2.1.1 View the Designation of instrument: Always AMI Hydrazine

    2.1.2 Version: Firmware of instrument, e.g. V4.00-05/07

    2.1.4 Factory test of the instrument; date of mother board and front-endQC test

    2.1.5 Operating time: years, days, hours, minutes, seconds

    2.2 Sensors2.2.1 Hydrazine sensor

    Current value: in ppbRaw value: Uncompensated current in nA.Ref. voltage: Applied voltage at the counter electrode (CE) in mV

    2.2.1.5 Cal. history:

    Review the diagnostic values of the last calibrations of thehydrazine sensor. Only for diagnostic purpose.Max. 65 data records are memorized.

    2.2.5 MiscellaneousCase Temp.: Actual temperature in °C of electronics housing.

  • 8/18/2019 AMI Hydrazine

    66/84

    64  A-96.250.501 / 070606

    AMI HydrazineProgramm List and Explanations

    2.3 Sample2.3.1 Sample ID

    Sample ID: Review the programmed code. The code is defined bythe user to identify the sample point in the plant.

    Sample TemperatureReview the actual temperature in °C and in Ohm.

    Sample flowReview if sample flows (detection: rpm and raw value in Hz)

    2.4 I/O State2.4.1- 2.4.2 Reveals the actual status of all in and outputs:

     Alarm relay: Open or closedRelay 1 and Relay 2: Open or closedInput: Open or closedSignal output 1, 2 (and 3, if option is installed): Actual current in mA.

    2.5 InterfaceOnly available if optional interface is installed.

    Review programmed communication settings.

    3 Maintenance

    3.1 Calibration

    3.1.1 Zero Hydrazine:

    Zero calibration (0 ppb). See Zero Hydrazine, p. 39 for moredetails.

    3.1.2 Process Hydrazine:

    For a process calibration of the hydrazine sensor. Follow thecommands on the screen. Save the value with the key.

  • 8/18/2019 AMI Hydrazine

    67/84

    AMI HydrazineProgramm List and Explanations

     A-96.250.501 / 070606 65

    3.2 Simulation3.2.1-3.2.6 Select relay or signal output to be tested with the key and

    press the up/down key to change the value. After confirming thesetting with the Enter key, the value is simulated by the relay/signaloutput.

     At the absence of any key activities, the instrument will switch backto normal mode after 20 min.

    If you quit the menu, all simulated values will be reseted.

    3.3 Set Time

    To adjust date and time.

    4 Operation

    4.1 Sensors

    4.1.1 Filter time constant:

    Used to damp noisy signals. Constant definition: 99% ofmeasurement value change in a given time (t99)Range: 5-300 sec

    4.1.2 Hold after cal: To allow the instrument to stabilize again after calibration. Duringcal. and hold time the signal outputs are frozen, alarms and limitsare not active.Range: 0 - 6000 sec

  • 8/18/2019 AMI Hydrazine

    68/84

    66  A-96.250.501 / 070606

    AMI HydrazineProgramm List and Explanations

    4.2 Relay Contacts4.2.1 Alarm Relay

    The alarm relay is used as cumulative error indicator.Under normal operating conditions, the contact is open.The contact closes upon:

    power loss

    detection of system faults like defective sensors or electronicparts

    high case temperature

    process values out of programmed ranges

    See chapter 5 for a complete list of parameters.

    Program process value alarm levels here or in menu 5.3.1

    4.2.1.1 Alarm Hydrazine

    4.2.1.1.1 Alarm high: If the measured value rises above the alarm highvalue, the alarm relay closes and E001 is displayedin the message list. Range: 0-1000 ppb

    4.2.1.1.28 Alarm low : If the measured value falls below the alarm lowvalue, the alarm relay closes and E002 is displayed

    in the message list. Range: 0-1000 ppb

    4.2.1.1.36 Hysteresis: Within the hysteresis range, the relay does not switch.This prevents damage of relais contacts when the measuredvalue fluctuates around the limit value.Range: 0-1000 ppb

    4.2.1.1.46 Delay  : During run time plus the delay time the signal and controloutputs are held in the operating mode.Range: 0-28800 sec

  • 8/1