— ABB MEASUREMENT & ANALYTICS | DATA SHEET
System components for sample gas sampling
2 SYSTEM COMPONENTS FOR SAMPLE GAS SAMPLING | DS/SCS-EN REV. A
— Measurement made easy System components for sample gas sampling
— Sample gas sampling best suited to the application ensures perfect measurement results in the gas analyzer • In most cases, the sample gas taken from the process
cannot be processed by the gas analyzer without further treatment. Excessive dust content, temperature and dew point as well as excessively high or low pressure and, last but not least, interfering components in the sample gas can impair the operability of the gas analyzer and falsify the measurement result.
• System components such as the sampling probes, filter units and sample gas lines ensure that regardless of the process and local conditions, the gas inlet conditions of the connected sample gas coolers, pumps and gas analyzers are met and perfect measurement results are obtained.
— System components for sample gas sampling • FE2 Filter unit • Probe H (cement flue gas probe) • Heating sleeve • Probe 2 • PFE2 Filter unit • Probe 11 • PFE3 Filter unit • Ceramic filter • Type 40, Type 40 W
Probe tubes • Process pressure
reducers • Type 42 Probe tube
(heated) • Electronic temperature controller
and switching relay • Type 60 S Probe tube
(cement flue gas probe) • CGWB 13, TBL 01-S, TBL 01-C
Heated sample gas lines
SYSTEM COMPONENTS FOR SAMPLE GAS SAMPLING | DS/SCS-EN REV. A 3
— Modular sample gas probe system
• Easily adaptable to different sampling conditions thanks to modular construction
• Components: – FE2, PFE2, PFE3 filter units – Heating sleeve – Probe tube types 40, 42, 60 S
• Probe tubes and filter unit heatable
• Connections for test gas and back purging
• Filter with high retention rate and long service life
• Easy filter replacement
• Usable in emission measuring systems conforming to European Directive 2010/75/EU (IED)
Sales Service
4 SYSTEM COMPONENTS FOR SAMPLE GAS SAMPLING | DS/SCS-EN REV. A
— FE2 Filter unit
Use The FE2 filter unit is a component of the modular gas sampling probe system. It is used for filtering gases containing dust. It can be heated with the regulated heating sleeve (see page 5).
Process gas sampling conditions Pressure pabs = 50 to 600 kPa
Temperature Max. 200 °C (coated max. 180 °C)
Flow rate 30 to 500 l/h, relative to 100 kPa and 0 °C
Pressure drop Approx. 0.6 hPa at 100 l/h
Dust content Max. 3 g/m3 (application recommendation); max. 40 g/m3 with probe tube type 40 (see page 19) with purge-back device; max. 2000 g/m3 with probe tube type 60 S (see page 21)
Filter characteristics Filter element Surface filter, ceramic
Filter fineness 0.3 µm
Retention rate 99.98 %
Seal integrity 10−4 hPa l/s
Dead volume 280 ml
Materials of gas-conducting parts Casing and connections Stainless steel 1.4571 (AISI 316Ti)
Flange Stainless steel 1.4541 (AISI 321)
Filter Aluminum oxide surface filter
Coating Thermoset plastic, resistant up to 180 °C
Installation Flange DN 65, PN 6, facing type A to DIN EN 1092-1; positioning of holes: 45°, 90°, 135°, 180°, 225°, 270°, 315°, 360°
Mounting Flange connect or bolt to the probe tube
Weight Approx. 8.0 kg
Connection ports Sample gas, test gas G¼ internal thread
Compressed air for back-purging (2 ports, optional, see page 6) G½ internal thread
Probe tube G¾ internal thread
Dimensions (in mm)
Ordering information
Catalog No. Housing and flange SS, filter 0.3 µm ceramic, gaskets standard, without purge connectors
23042-4-0730743
Housing and flange SS, filter 0.3 µm ceramic, gaskets standard, with purge connectors
23042-4-0730745
Housing and flange SS, coated, filter 0.3 µm ceramic, gaskets standard, with purge connectors
23042-5-0730746
Housing and flange SS, filter 0.3 µm ceramic, gaskets for high corrosive gases, without purge connectors
23042-4-0730747
Housing and flange SS, filter 0.3 µm ceramic, gaskets for high corrosive gases, with purge connectors
23042-4-0730749
141
30
16
0
65
280 40...651
05
SYSTEM COMPONENTS FOR SAMPLE GAS SAMPLING | DS/SCS-EN REV. A 5
— Heating sleeve
Regulated heating sleeve for FE2 filter unit Heater Micanite
Holding temperature Max. 350 °C
Temperature control Max. 200 °C (Pt 100)
Power supply 120 V AC or 230 V AC
Power consumption 250 VA
Connection Terminal box with M16 cable gland, degree of protection IP54
Supply cable 3 x 1 mm2 with silicone sheath, length 1 m
Ambient temperature −20 to +45 °C
Relative humidity ≤ 75 % yearly mean; infrequent slight condensation permissible
Weight Approx. 1 kg
Essential accessories Insulating mat (see ordering information) Temperature controller (see page 27)
Ordering information
Catalog No. Heating sleeve, 230 V, 50/60 Hz, 250 VA 23043-5-8018410 Heating sleeve, 120 V, 50/60 Hz, 250 VA 23043-5-8018411 Insulating mat 23043-5-8018412
6 SYSTEM COMPONENTS FOR SAMPLE GAS SAMPLING | DS/SCS-EN REV. A
— PFE2 Filter unit
Use and design The PFE2 filter unit consists of the FE2 filter unit (see page 4) with a regulated heating sleeve (see page 5, installed in a stainless-steel protective case. It is used to filter dust-containing gases and is a component of the modular gas sampling probe system.
When used for gases with high dust content, the PFE2 can option-ally be fitted with valves for automatic cleaning of the filter ele-ment in the FE2 filter unit. Depending on the dust load, in addition to 1-stage or 2-stage back-purging the type of back-purging can be selected. For this purpose, a pneumatically controlled valve cir-cuitry with diaphragm back-purging valves and a direct acting version with coaxial sliding valves are available. The coaxial sliding valves are particularly suitable for frequent back-purging cycles.
Protective case Dimensions Standard version: 450 x 450 x 400 mm (H x W x D) Basic version: 270 x 203 x 330 mm (H x W x D)
Material Stainless steel 1.4301 (AISI 304)
Degree of protection of housing IP54
Weight Standard version: approx. 25 kg incl. FE2 filter unit; Basic version: approx. 17 kg incl. FE2 filter unit
Ambient temperature −20 to +45 °C
Option Heating sleeve (see page 5)
Back purging Compressed air quality Instrument air, free of dust, water and oil
Pneumatic valves
Type 2/2-way straight-way diaphragm valve
Operating pressure/control pressure 600 kPa (6 bar)
Nominal diameter DN 12
Connections G½ internal thread, control air: G⅛ internal thread
Material Brass
Mode of operation Closed in normal position
Ambient temperature 0 to +100 °C
Control valves
Type 3/2-way solenoid valve
Pressure range 600 kPa (6 bar)
Nominal diameter DN 2.0
Connections G⅛ internal thread
Material Brass
Mode of operation Closed in normal position
Ambient temperature 0 to +100 °C
Electrical connections Standard connector
Power supply 120 V AC or 230 V AC, 50 to 60 Hz
Power consumption 8 VA
Coaxial valves
Type 2/2-way direct acting valve
Operating pressure 0 to 64 bar
Nominal diameter DN 10
Connections 10/12 x 1 mm
Material Body: brass, valve seat: synthetic resin on metal, seal: NBR
Mode of operation Closed in normal position
Ambient temperature −10 to +80 °C
Power supply, current consumption 230 V AC, 50 to 60 Hz, 0.13 A
SYSTEM COMPONENTS FOR SAMPLE GAS SAMPLING | DS/SCS-EN REV. A 7
Probe protective case, standard version: Dimensions (in mm)
40
0
450
45
0
155
440
30
016
0
17
0
170
210
75
75
11
0
65
80
90
180
80
50
15
5
M4
M4
165
ø6
,5 ø11
,5
ø1
9,5
M4
M4
45
08
0
155
550 300
Te
st
ga
sco
nn
ectio
nB
ulk
he
ad
fitt
ing
4/6
x1
mm
Co
mp
resse
da
irP
urg
ing
Bu
lkh
ea
dfitt
ing
10
/12
x1
mm
Sa
mp
leg
as
co
nn
ectio
n
Co
ntr
ola
irB
ulk
he
ad
fitt
ing
4/6
x1
mm
Fla
ng
ep
late
with
pip
ecla
mp
21
/2"
RS
T3
7-2
ho
tg
alv
an
ize
dD
IN3
56
8fo
rm
ou
ntin
ga
tp
rob
em
ou
ntin
ga
da
pte
r
Elo
ng
ate
dh
ole
sfo
rM
8sa
fety
co
llar
bo
lts
Ele
ctr
ica
lco
nn
ectio
ns
Co
ve
rsw
ive
lsu
pw
ard
s
2x
M3
23
xM
25
2x
M2
0
Ta
ke
ho
les
for
ca
ble
gla
nd
so
nth
ele
ftin
toa
cco
un
t
Te
rmin
alb
ox
with
ca
ble
gla
nd
sU
se
as
req
uire
d
8 SYSTEM COMPONENTS FOR SAMPLE GAS SAMPLING | DS/SCS-EN REV. A
— … PFE2 Filter unit
Probe protective case, standard version: Mounting
Minimum distance xmin of the mounting flange from the wall depending on mounting angle α: α 10° 15° 20° 25° 30° 35° xmin / mm 229 248 268 287 307 324
Probe protective case, basic version: Dimensions (in mm)
1 Terminal box -X1 IP66 2 Sample gas line 3 Sample gas connection 6 mm 4 Test gas connection bulkhead fitting 6 mm
X
3 xM20x1.5
2 xM12x1.5
2 xM20x1.5
1234
270
330
130
58
124
80
266
200
SYSTEM COMPONENTS FOR SAMPLE GAS SAMPLING | DS/SCS-EN REV. A 9
Pneumatic connections for 2-stage back-purging, standard version
C
D
B
A
A:
B:
C:
D:
Back
Pre
ssure
Purg
ing
Filt
er
Inle
tG
1/2
"->
12
mm
Tube
Fitting
Back
Pre
ssure
Purg
ing
Filt
er
Surf
ace
and
Pro
be
Tube
Inle
tG
1/2
"->
12
mm
Tube
Fitting
Sam
ple
Outlet
G1/4
"->
6m
mor
8m
mT
ube
Fitting
Calib
ration
Gas
Inle
tG
1/4
"->
6m
mT
ube
Fitting
Instr
um
entA
irIn
let
Bulk
Head
Unio
n12
mm
Calib
ration
Gas
Inle
tB
ulk
Head
Unio
n6
mm
Contr
olA
irIn
let
Bulk
Head
Unio
n6
mm
Pilo
tO
pera
tion
Valv
eC
leanin
gF
ilter
-Y2.1
Dia
phra
gm
Valv
eC
leanin
gF
ilter
-Y2.2
Pilo
tO
pera
tion
Valv
eP
uls
-Instr
um
entA
ir-Y
1.1
Dia
phra
gm
Valv
eP
uls
-Instr
um
ente
nlu
ft-Y
1.2
Pilo
tO
pera
tion
Valv
eC
leanin
gF
ilter
Surf
ace
and
Pro
be
Tube
-Y3.1
Dia
phra
gm
Valv
eC
leanin
gF
ilter
Surf
ace
and
Pro
be
Tube
-Y3.2
Sole
noid
Valv
eA
era
tion
-Y4
Heate
dS
am
ple
Lin
e-E
2
Junction
Box
Pro
tection
Box
Check
Valv
e
Filt
er
Unit
Heate
dC
heck
Valv
e-Y
5
Pt1
00
Connection
Pow
er
Supply
10 SYSTEM COMPONENTS FOR SAMPLE GAS SAMPLING | DS/SCS-EN REV. A
— … PFE2 Filter unit
Pneumatic connections for 2-stage back-purging, standard version with coaxial valves
0
Junction b
ox
Pro
tection b
ox
Check v
alv
e
Filt
er
unit
Pow
er
supply
Connection for
back-p
urg
ing filt
er
G 1
/2 in. to
12 m
m p
ipe c
ouplin
g
Test gas c
onnection G
1/4
in.
to 6
mm
pip
e c
ouplin
g
Sam
ple
gas o
utlet G
1/4
in.
to 6
mm
pip
e c
ouplin
g
Sole
noid
valv
e for
cle
anin
g filt
er
-Y2
Connection for
instr
um
ent
air (
max. 6 b
ar)
bulk
head fitting 1
2 m
m
Connection for
test gas
bulk
head fitting 6
mm
Connection for
Pt1
00
Heate
d s
am
ple
gas lin
e -
E2
Connection for
instr
um
ent
air (
max. 6 b
ar)
bulk
head fitting 1
2 m
m
Heate
d c
uto
ff v
alv
e -
Y5 (
option)
Sole
noid
valv
e for
cle
anin
g filt
er
surf
ace
and p
robe tube -
Y1
Connection for
back-p
urg
ing
filter
surf
ace a
nd p
robe tube
G 1
/2 in. to
12 m
m p
ipe c
ouplin
g
SYSTEM COMPONENTS FOR SAMPLE GAS SAMPLING | DS/SCS-EN REV. A 11
Probe protective case, standard version: Circuit and terminal diagram
-E1
-E2
LN
PE
-X1
12
��
��
-X1
56
PE
38
-E3
.2
��
-X1
LN
PE
-X1
-E5
.2-E
5.1
LN
PE
17
-T0
1
11
5V
~
24
V=
1,3
A
-X2
12
LN
PE
45
LN
PE
78
LN
PE
LN
PE
-Y3.1
-Y1.1
-Y2.1
-Y4
LN
PE
15
36
A1
A2
-Y5 -X
2
......
...
-X3
12
34
56
78
910
11
12
13
14
15
16
18
17
14
-E4
-B01
set
@30°C
3...1
8x
0,1
4m
m²
18
19
32
1
3x1,5
mm
²3x2,5
mm
²
3x1,5
mm
²
3x1,5
mm
²
3x1,5
mm
²4x1,5
mm
²
20
21
15
16
13
14
12
10
11
913
7
N
4
L
3x1,5
mm
²2x0,1
4m
m²2
x0,1
4m
m²
2x0,1
4m
m²
2x0,1
4m
m²
2x
0,1
4m
m²
2x
0,1
4m
m²
-K0
13x1,5
mm
²
L
11
10
9
PE
N
12
-E3
.1
External Protection Box Junction Box Analyzer Cabinet
Power Supply Heating Sleave Filter Unit,230 or 115 VAC, 50...60 Hz, 250 W,required fusing 6 A,controller required !
Heatin
gS
leeve
Filte
rU
nit
Power Supply Heater Probe Tube,230 VAC, 50...60 Hz, max. 800 W,required fusing 6 A,in case of 115 VAC power supply wire theheating elements -E3.1 and -E3.2 parallel.controller required !
Power Supply Sample Line,230 VAC: < 35 m -> 1~, 35...60 m -> 3~,115 VAC: < 15 m -> 1~, 15...40 m -> 3~,50...60 Hz, 90 W/m,in case of a 3~ sample line connectL2 -> Termination 5, L3 -> Termination 6,required fusing 16 A,controller required !
>>
Heate
dP
robe
Tube
Power Supply Frost Protection,230 or 115 VAC, 50...60 Hz, 300 W,required fusing 6 A
Power Supply PTC-Elements andPowerpack -G02 / Heated Solenoid Valvein Sample Outlet,230 or 115 VAC, 50...60 Hz, 250 W,required fusing 6 A
Power Supply Pilot Operation Valve -Y3.1,Cleaning Filter Surface and Probe Tube,230 or 115 VAC, 50...60 Hz, 8 Wrequired fusing 1 A
Power Supply Pilot Operation Valve -Y1.1,Puls-Instrument Air,230 or 115 VAC, 50...60 Hz, 8 Wrequired fusing 1 A
Power Supply Pilot Operation Valve -Y2.1,Cleaning Filter,230 or 115 VAC, 50...60 Hz, 8 Wrequired fusing 1 A
Power Supply Pilot Operation Valve -Y4,Aeration,230 or 115 VAC, 50...60 Hz, 8 Wrequired fusing 1 A
Power Supply Relais -K05/Check Valve -Y5,230 or 115 VAC, 50...60 Hz,required fusing 1 A
Resistance Thermometer PT100,Filter Unit Sample Outlet,3-Wire-Connection
Resistance Thermometer PT100,Sample Line,3-Wire-Connection
Resistance Thermometer PT100,Probe Tube,3-Wire-Connection
Resistance Thermometer PT100,Filter Unit Sample Outlet (Safety Wiring), 3-Wire-Connection
Resistance Thermometer PT100,Sample Line (Safety Wiring),3-Wire-Connection
Resistance Thermometer PT100,Probe Tube (Safety Wiring),3-Wire-Connection
Fro
st
Pro
tectio
n
Heate
dC
heck
Valv
e
21.1
0.2
005
Heate
dS
am
ple
Lin
e
12 SYSTEM COMPONENTS FOR SAMPLE GAS SAMPLING | DS/SCS-EN REV. A
— … PFE2 Filter unit
Probe protective case, standard version with coaxial valves: Circuit and terminal diagram
-E1
-E2
LN
PE
-X1
12
-X1
56
L2
L3
PE
38
-E3
.1-E
3.2
-X1
LN
PE
-X1
-E5.2
-E5
.1
LN
PE
17-X
21
24
5
LN
PE
LN
PE
-Y1
-Y2
LN
PE
93
6
A1
A2
-Y5 -X
2
......
...
-X3
12
34
56
78
910
111
213
141
51
61
81
78
-E4
-B01
set @
30°C
18
19
32
1
3x1,5 mm
²5x2,5 mm
²3
x1,5 mm
²
3x1,5
mm
²4x1,5 m
m²
20
151
613
14
77
N
4
L1
3x1,5
mm
²
2x0,14 m
m²
2x0,14 m
m²
2x0,14 mm
²
2x0,14 mm
²
2x0,14m
m²
2x0,14
mm
²
-K05
3x1,5 mm
²
L
111
09
PE
N
12
LN
-T01 2
30
V~
24
V=
1,3
A
+-
External Protection Box Junction Box Analyzer Cabinet
Power Supply Heating Sleave Filter Unit,230 VAC, 50...60 Hz, 250 W,required fusing 6 A,controller required!
Hea
ting
Sle
eve
Filte
r Un
it
Power Supply Heater Probe Tube,230 VAC, 50...60 Hz, max. 800 W,required fusing 6 A,controller required!
Power Supply Sample Line,230 VAC, 50...60 Hz,90 W/m, max. 11 kW / 17 kW, required fusing 16 A / 25 A,controller required!
Hea
ted
Pro
be Tu
be
Power Supply Frost Protection,230 VAC, 50...60 Hz, 300 W,required fusing 6 A
Power Supply PTC Elements and Powerpack -G02 / Heated Solenoid Valve in Sample Outlet,230 VAC, 50...60 Hz, 250 W,required fusing 6 A
Power Supply Solenoid Valve -Y1Cleaning Filter Surface and Probe Tube,230 VAC, 50...60 Hz, 30 VArequired fusing 1 A
Power Supply Relay -K05 for Cutoff Valve -Y5,230 VAC, 50...60 Hz,required fusing 1 A
Resistance Thermometer Pt100,Filter Unit Sample Gas Outlet,3-Wire ConnectionResistance Thermometer Pt100,Sample Gas Line,3-Wire ConnectionResistance Thermometer Pt100,Probe Tube,3-Wire ConnectionResistance Thermometer Pt100,Filter Unit Sample Gas Outlet (Safety Wiring),3-Wire ConnectionResistance Thermometer Pt100,Sample Gas Line (Safety Wiring),3-Wire ConnectionResistance Thermometer Pt100,Probe Tube (Safety Wiring),3-Wire Connection
Fro
st Prote
ction
He
ated C
heck V
alve
Hea
ted
Sam
ple Lin
ePower Supply Solenoid Valve -Y2Cleaning Filter,230 VAC, 50...60 Hz, 30 VArequired fusing 1 A
SYSTEM COMPONENTS FOR SAMPLE GAS SAMPLING | DS/SCS-EN REV. A 13
Ordering information
Catalog No. 23028-0- 0 0 Design with protection box, Standard, IP 54 (450 x 450 x 400 mm) 2 with protection box, Standard, IP 54, anti-freeze heating 3 with protection box, Standard, IP 54, ambient temp. max. 65 °C 4 with protection box, Basic, IP 54 (270 x 203 x 330 mm) 5 Filter housing and flange (equal FE2) Material SS, filter ceramic 0.3 µm, w/o connectors for back purging A Material SS, coated, filter ceramic 0.3 µm, w/o connectors for back purging B Material SS, filter stainless steel 1.0 µm, w/o connectors for back purging C Heating with heating sleeve and insulation (max. 190 °C, Pt 100 required) 2 Controlling and monitoring without 1 with 1 x Pt 100 2 with 2 x Pt 100 3 Back purging unit with test/emergency purge valve (w/o compressed air unit) without 0 with back purging, 1-stage and SS connection adapter 1 with back purging, 2-stage and SS connection adapter 2 without back purging, with test/emergency purge valve, unheated, without SS connection adapter 3 with back purging, 1-stage, with test/emergency purge valve, heated shut-off valve, SS connection adapter 5 with back purging, 2-stage, with test/emergency purge valve, heated shut-off valve, SS connection adapter 6 with back purging, 2-stage, heated shut-off valve, SS connection adapter 7 with back purging, 1-stage via Coax valve, SS connection adapter A with back purging, 2-stage via Coax valves, SS connection adapter B with back purging, 1-stage via Coax valve, with test/emerg. purge valve, heated shut-off-valve, SS connection ad. D with back purging, 2-stage via Coax valves, with test/emerg. purge valve, heated shut-off valve, SS connection ad. E with back purging, 2-stage via Coax valves, heated shut-off valve, SS connection adapter F Power supply 230 V, 50 to 60 Hz 1 120 V, 50 to 60 Hz 2 120 V, 50 to 60 Hz CSA, with AWG wiring 3 Approvals without 0 GOST 1 Gaskets Standard 1 for high corrosive gases 2 Connectors for back-purging at the filter housing (FE2) without 0 with 1 Bulkhead unions (at the protection box) metric 1 imperial (sample gas outlet in metric only) 2
14 SYSTEM COMPONENTS FOR SAMPLE GAS SAMPLING | DS/SCS-EN REV. A
— PFE3 Filter unit
Design The PFE3 filter unit consists of a modified FE2 filter unit with a self-regulating heater. It is installed in a stainless steel protective case.
Process gas sampling conditions Pressure pabs = 50 to 600 kPa
Temperature Max. 200 °C
Flow rate 30 to 500 l/h, relative to 100 kPa and 0 °C
Pressure drop Approx. 0.6 hPa at 100 l/h
Dust content Max. 3 g/m3 (application recommendation)
Filter characteristics Filter element Surface filter
Filter fineness 0.3 µm
Seal integrity 10−4 hPa l/s
Dead volume 280 ml
Dimensions 50/20 x 135 mm
Heater Type Self-regulating heater (400 W)
Temperature 180 °C (ambient temperature 20 °C)
Temperature alarm Approx. 150 °C
Power supply 120 V AC or 230 V AC, 50 to 60 Hz
Options Heating sleeve (see page 5)
Materials of gas-conducting parts Casing, flange and connections Stainless steel 1.4571 (AISI 316Ti)
Seals FPM
Installation Flange DN 65, PN 6, facing type A to DIN EN 1092-1; positioning of holes: 45°, 135°, 225°, 315°
Mounting Flange connect or bolt to the probe tube
Connection ports Sample gas Pipe fitting 6 mm
Test gas Bulkhead fitting 6 mm (option)
Purge gas Bulkhead fittings 18 mm (option)
Probe tube G¾ internal thread
Protective case Dimensions 270 x 203 x 330 mm (H x W x D)
Material Stainless steel 1.4301 (AISI 304)
Degree of protection of case IP54
Weight Approx. 17 kg incl. FE2 filter unit
Ambient temperature −20 to +45 °C
SYSTEM COMPONENTS FOR SAMPLE GAS SAMPLING | DS/SCS-EN REV. A 15
Dimensions, gas connections (dimensions in mm)
58
270
266
130
124
80
330
200
3x
M2
0x1
,52
xM
12
x1
,5
M2
0x1
,5
Ju
nctio
nB
ox
-X1
IP6
6
Tu
be
,S
S3
16
Ti,
6x1
mm
Ca
libra
tio
nG
as
Bu
lkH
ea
dU
nio
n6
mm
Cle
an
ing
Filt
er
Su
rfa
ce
an
dP
rob
eT
ub
e,
Bu
lkH
ea
dU
nio
n1
8m
m
Cle
an
ing
Filt
er ,
Bu
lkH
ea
dU
nio
n1
8m
m
Tu
be
,C
U1
5x1
mm
Tu
be
,C
U1
5x1
mm
16 SYSTEM COMPONENTS FOR SAMPLE GAS SAMPLING | DS/SCS-EN REV. A
— … PFE3 Filter unit
Circuit and terminal diagram
-E2
LN
PE
-X1
12
�
-X1
34
LN
PE
PE
PE
-X1
56
78
910
3x2,5
mm
²
2x0,1
4m
m²
-E1.1
-E1.2
-E1
.3
-E1
.4
set@
140°C
�
11
12
13
2x0,1
4m
m²
2x0,2
mm
²
3x1,5
mm
²
External Protection Box Junction Box Analyser Cabinet
Power SupplyHeated Aluminium Form Filter Unit,230 or 115 VAC, 50...60 Hz, 250 W,required fusing 6 A
Power Supply Sample Line,230 or 115 VAC, 50...60 Hz,90 W/m, max. 3600 W / 5700 W,required fusing 16 A / 25 A,controller required !
Resistance Thermometer PT100,Filter Unit Sample Outlet,3-Wire-Connection
Resistance Thermometer PT100,Sample Line,3-Wire-Connection
Heate
dS
am
ple
Lin
e
He
ate
dA
lum
iniu
mF
orm
sF
ilter
Un
it
Temperature Alarm,Surface heated Aluminium Forms
SYSTEM COMPONENTS FOR SAMPLE GAS SAMPLING | DS/SCS-EN REV. A 17
Ordering information
Catalog No. 23046-0- 0 0 0 0 0 Design with protection box IP54 (330 x 203 x 270 mm) 1 Filter housing and flange in stainless steel Filter unit < 0.3 µm (filtration fineness), without purge connectors 4 Filter unit < 0.3 µm, coated, without purge connectors 5 Filter unit FE2, < 0.3 µm, with purge connectors, 2-stage 6 Filter unit FE2, coated, < 0.3 µm, with purge connectors, 2-stage 7 Heating with self-regulated heating and temperature alarm (120/230 V AC) 1 with heated sleeve (max. 190 °C, Pt 100 required) 3 Controlling and monitoring without 3 with 1 x Pt 100 (only for heated sleeve) 4 with 2 x Pt 100 (only for heated sleeve) 5 Connectors for back purging (without compressed air unit) without 1 with bulkhead unions for back purging, 1-stage 2 with bulkhead unions for back purging, 2-stage 3 Power supply 230 V AC, 50 to 60 Hz 1 120 V AC, 50 to 60 Hz 2 Test gas / emergency purging valve without 1 with, unheated, bulkhead union for connecting 2
18 SYSTEM COMPONENTS FOR SAMPLE GAS SAMPLING | DS/SCS-EN REV. A
— Selection table for probe tube types and probes
Prob
e tu
be t
ype 40
SS
1.457
1 (A
ISI 3
16Ti
)
Prob
e tu
be t
ype 40
W
SS 1
.457
1 (A
ISI 3
16Ti
)
Prob
e tu
be t
ype 40
SS
1.457
1 (A
ISI 3
16Ti
) co
ated
Prob
e tu
be t
ype 40
SS
1.4
893
Prob
e tu
be t
ype 40
si
licon
car
bide
Prob
e tu
be t
ype 42
SS
hea
ted
Prob
e tu
be t
ype 60
S
wat
er c
oole
d 3)
Prob
e H
w
ater
coo
led 3
)
Prob
e 2
pa
rtly
hea
ted
Prob
e 2
he
ated
Temperature < 130 °C X < 180 °C X X 2) < 300 °C X
180 to 450 °C X X 180 to 900 °C X
< 1300 °C X X X < 1400 °C X X
Dust content < 50 mg/m3 X X X X X X < 1000 mg/m3 X X X X X X
< 3 g/m3 X X X X X X X X > 3 g/m3 X 1) X 1) X 1) X 1) X 1) X X
40 to 100 g/m3 X 1) X 1) X 1) X 1) < 2000 g/m3 X 1) X 1)
Water vapor dew point < 65 °C X X X X X X X > 65 °C X X X X X X X
Acidic aerosols (acid dew point 130 °C) X X X X
Pressure < 2 bar X X X < 5 bar X X X X X
Technical data see … Page 19 Page 19 Page 19 Page 19 Page 19 Page 20 Page 21 Page 22 Page 23 Page 23
1) Back purging required 2) With regulated heating sleeve (see page 5) 3) See data sheet ‘SCK Sampling system for dry gas sampling’
— Lead-through tube installation
Brickwork with metal sheeting Isolated sheet-metal channel Brickwork
Flow direction of process gas
1 Lead-through tube 2 Mounting flange DN 65, PN 6,
facing type A to DIN EN 1092-1 3 Gasket 4 Welded-on rectangular block 5 Sampling probe flange
Dimensions in mm
Minimum distance xmin of the mounting flange from the wall depending on mounting angle α: α 10° 15° 20° 25° 30° 35° xmin/mm 133 138 143 147 151 153
1 2 5 3 1 2 5 3
1 2 5 34
SYSTEM COMPONENTS FOR SAMPLE GAS SAMPLING | DS/SCS-EN REV. A 19
— Type 40, Type 40 W Probe tubes
Use For sampling hot dust-containing gases in conjunction with the FE2/PFE2 filter unit (see page 4 or page 6)
Process gas sampling conditions Pressure pabs = 50 to 500 kPa (0.5 to 5 bar)
Temperature Stainless steel 1.4571 (AISI 316Ti) max. 450 °C Stainless steel 1.4571 (AISI 316Ti) coated max. 180 °C Stainless steel 1.4893 max. 900 °C Stainless steel 1.4571 (AISI 316Ti) Type 40 W max. 300 °C Silicon carbide (max. length 1000 mm) max. 1300 °C
Flow rate Max. 300 l/h
Pressure drop Approx. 10 hPa at 30 to 90 l/h
Dust content Max. 3 g/m3 (application recommendation); max. 40 g/m3 with purge-back device in conjunction with FE2/PFE2 filter unit (not for coated probe tube)
Installation Connection G¾ external thread
Mounting angle Any; 8° to 90° downwards from the horizontal is recommended. It is imperative to install the silicon carbide probe tube vertically.
Dimensions (in mm)
Ordering information
Catalog No. 23036-0- 0 0 0 0 0 0 0 0 0 0 Version Stainless steel (1.4571), up to 500 °C 1 Stainless steel (1.4571), coated, up to 180 °C 2 Silicon carbide up to 1300 °C, max. 1000 mm 4 Stainless steel (1.4893), up to 900 °C 5 Stainless steel (1.4571), up to 300 °C, heated (Type 40 W for cement) 6 Nominal length 500 mm 1 1000 mm 2 1500 mm 3 2000 mm 4 2500 mm 5 3500 mm (only for heated version) 6 4000 mm (only for heated version) 7 4500 mm (only for heated version) 8
G3/4
G3/4
G3/4
L1
L1
L1
16
21,4
ø
12,6
ø16
,8 ø
15 ø
23 ø
34 ø
16
16105
17,2
ø
Silicon-Carbide (max. 1300 °C)
Special Steel Material No. 1.4571 (max. 450 °C)
Special Steel Material No. 1.4893 (max. 900 °C)
SiC Protective Tube
20 SYSTEM COMPONENTS FOR SAMPLE GAS SAMPLING | DS/SCS-EN REV. A
— Type 42 Probe tube (heated)
Use Low gas temperature
High moisture content in gas
Corrosive components in the process gas
Typical places of use are:
• Sampling points in wet desulphurization plants (crude gas, clean gas, part-cleaned clean gas)
• Waste incineration plants
• Auxiliary boilers and stack emission measurements
Process gas sampling conditions Pressure pabs = 200 kPa (2 bar)
Temperature Max. 200 °C
Flow rate Max. 250 l/h
Flow velocity Max. 12 m/s (> 1000 mm max. 7 m/s)
Dust content Max. 3 g/m3 (application recommendation); max. 40 g/m3 with purge-back device in conjunction with FE2/PFE2 filter unit (see page 4 or page 6)
T90 time Approx. 0.4 s at 250 l/h
Design Materials of gas-conducting parts Stainless steel 1.4571 (AISI 316Ti)
Flange DN 65, PN 6, facing type A to DIN EN 1092-1; positioning of holes: 45°, 135°, 225°, 315°
Weight Nominal length 600 mm 1000 mm 1500 mm 2000 mm Weight 7 kg 8 kg 10 kg 12 kg
Probe tube heater Heat output Nominal length 600 mm 1000 mm 1500 mm 2000 mm Heating power 600 W 800 W 1200 W 1500 W
Nominal voltage 120 V AC or 230 V AC, 50 to 60 Hz
Control Control temperature max. 200 °C
Temperature sensor Pt 100
Ordering information
Catalog No. Length 600 mm 3KXG839050U0100 Length 1000 mm 23041-4-8018392 Length 1500 mm 23041-4-8018393 Length 2000 mm 23041-4-8018394
Dimensions and connections (dimensions in mm [inch])
SYSTEM COMPONENTS FOR SAMPLE GAS SAMPLING | DS/SCS-EN REV. A 21
— Type 60 S Probe tube (cement flue gas probe)
Use and design The water-cooled probe tube type 60 S is used in the sampling system SCK for sampling hot cement flue gas with high dust con-tent for CO, CO2, NO and SO2 analysis in ‘dry’ cement production process (e.g. recuperator furnaces, furnaces with grate coolers), at rotary kiln inlet and/or after precalcination.
The probe tube type 60 S is installed in kilns where less hard en-crustations are to be expected.
By using the probe type 60S with two sample gas openings at the side of the probe tip, an effective pre-dedusting of the sam-ple gas is achieved at the tip thus reducing the frequency of the automatic back purging actions.
Back-purging of the probe tube and the filter with compressed air as well as the probe retraction out of and insertion into the oven are controlled by the sampling system SCK (see data sheet ‘Sampling System SCK’).
Process gas sampling conditions Pressure pabs = max. 200 kPa (2 bar)
Temperature Installation length/mm 3.500 3.000 2.500 2.500 1.500 Max. temperature/°C 1.200 1.250 1.300 1.350 1.400
Flow rate Max. 250 l/h
Dust content Max. 2000 g/m3 with periodical compressed air cleaning in con-junction with FE2 filter unit (see page 4)
Dead time (Tt) for length = 3500 mm Approx. 40 s at 60 l/h, approx. 9 s at 250 l/h
90% time (T90) for length = 3500 mm Approx. 82 s at 60 l/h, approx. 17 s at 250 l/h
Pressure drop Approx. 10 hPa
Design Materials of gas-conducting parts Sampling pipe: Stainless steel 1.4571 (AISI 316Ti); Cooling tube: Stainless steel 1.4541 (AISI 321)
Flange DN 150, PN 10
Weight 45 to 70 kg
Mounting A flanged probe connector (internal diameter 200 mm) should be installed by the customer. The mounting flange is essential for installing the probe tube. The connections should be aligned according to the local situation.
Cooling water supply Quality Drinking water quality with approx. 30 % anti-freeze if required
Outlet temperature Min. 50 °C (or > dew point of the cement flue gas), max. 85 °C
Pressure pabs = max. 400 kPa (4 bar)
Quantity circulated Approx. 3 m3/h (closed system)
Compressed air supply Quality Instrument air, free of dirt, oil and water droplets, 600 kPa (6 bar) positive pressure, dew point −40 °C
Consumption Approx. 3 m3/h at standard conditions
Ordering information
Catalog No. Length 1500 mm 23038-4-0730642 Length 2000 mm 23038-4-0730643 Length 2500 mm 23038-4-0730644 Length 3000 mm 23038-4-0730645 Length 3500 mm 23038-4-0730646
Dimensions (in mm)
1 Sample gas inlet ports 2 Thickened shank of probe tube 3 Cooling water connections L1 Probe tube length L2 Length of thickened shank:
800 mm for L1 ≤ 2000 mm 1300 mm for L1 ≥ 2500 mm
L3 200 mm with reinforced protection tube
L1
L2
90
35
A
A
~60
3
3
1 2
~70
ø60
,3
ø89
L3
22 SYSTEM COMPONENTS FOR SAMPLE GAS SAMPLING | DS/SCS-EN REV. A
— Probe H (cement flue gas probe)
Use and design The water-cooled gas sampling probe type H is used in the sam-pling system SCK for sampling hot cement flue gas with high dust content for CO, CO2, NO and SO2 analysis in ‘dry’ cement production process (e.g. recuperator furnaces, furnaces with grate coolers), at rotary kiln inlet and/or after precalcination.
The probe type H is installed in kilns, where extremely hard en-crustations are to be expected.
By using the probe type H, long measuring periods without inter-ruptions are possible: The pneumatically driven plunger which is welded on the filter tip mechanically breaks the encrustations at the probe gas entry without interruption of the measurement.
Back-purging of the probe with compressed air as well as the plunger movement and the probe retraction out of and insertion into the oven are controlled by the sampling system SCK (see data sheet ‘Sampling System SCK’).
Process gas sampling conditions Pressure pabs = max. 200 kPa (2 bar)
Temperature Installation length/mm 3.500 3.000 2.500 2.500 1.500 max. temperature/°C 1.200 1.250 1.300 1.350 1.400
Flow rate Max. 250 l/h
Dust content Max. 2000 g/m3 with periodical compressed air cleaning
Dead time (Tt) for length = 3500 mm Approx. 40 s at 60 l/h, approx. 9 s at 250 l/h
90% time (T90) for length = 3500 mm Approx. 82 s at 60 l/h, approx. 17 s at 250 l/h
Pressure drop Approx. 10 hPa
Design Materials of gas-conducting parts Sampling pipe: Stainless steel 1.4436 (AISI 316); Cooling tube: Stainless steel 1.4541 (AISI 321)
Flange DN 150, PN 10
Weight 60 to 90 kg
Mounting A flanged probe connector (internal diameter 200 mm) should be installed by the customer. The mounting flange is essential for installing the probe tube. The connections should be aligned according to the local situation.
Cooling water supply Quality Drinking water quality with approx. 30 % anti-freeze if required
Outlet temperature Min. 50 °C (or > dew point of the cement flue gas), max. 85 °C
Pressure pabs = max. 400 kPa (4 bar)
Quantity circulated Approx. 3 m3/h (closed system)
Compressed air supply Quality Instrument air, free of dirt, oil and water droplets, 600 kPa (6 bar) positive pressure, dew point −40 °C
Consumption Approx. 3 m3/h at standard conditions
Ordering information
Catalog No. Length 1500 mm 23020-4-0730656 Length 2000 mm 23020-4-0730647 Length 2500 mm 23020-4-0730648 Length 3000 mm 23020-4-0730649 Length 3500 mm 23020-4-0730650
Dimensions (in mm)
1 Control air conn. G¼ 2 Sample gas conn. G¼
3 Compressed air connection for pipe nozzle with 18 mm outer diameter
4 Cooling water connections G 1 inch 5 Bleeding for cooling water circulation
6 Flange DN 150, PN 10 L Installation length
18
L350535202
1105
L+1105
1
2
34
14
5
6
SYSTEM COMPONENTS FOR SAMPLE GAS SAMPLING | DS/SCS-EN REV. A 23
— Probe 2
Use Wherever large volumes of gas are sampled, gas pressure is raised and there is a high moisture or dust content in the gas.
Process gas sampling conditions Pressure pabs = max. 400 kPa (4 bar)
Temperature Max. 450 °C (probe tube partially heated), max. 130 °C (probe tube heated
Flow rate Max. 2000 l/h
Dust content Max. 20 g/m3 (application recommendation), max. 100 g/m3 with purging
Purging pressure Max. 600 kPa (6 bar)
Heating temperature of inlet filter ≥ 200 °C (at process gas temperature of 20 °C)
90% time (T90) < 2 s at 250 l/h
Filter fineness 0.3 µm
Material of gas-conducting parts Support tube, heating rod, connecting pipe Stainless steel 1.4571 (AISI 316Ti)
Inlet filter Aluminum oxide
Seals Ceramic fiber
Flange Stainless steel 1.4581
Mounting Flange DN 65, PN 6, facing type A to DIN EN 1092-1; positioning of holes: 45°, 135°, 225°, 315°
Mounting angle Any; 10° to 35° downwards from the horizontal recommended (take direction of flow into account), see ‚Dimensions‘
Sample gas outlet G¼ internal thread
Test gas/purge air inlet G⅜ internal thread
Probe tube heating Heat output (unregulated) Nominal length 1200 mm: 650 W, partially heated (for heating of the wall bushing): 255 W
Nominal voltage 230 V AC, 50 to 60 Hz (120 V AC on inquiry)
Electrical connection Silicone cable, 3 x 1.5 mm2, length 1000 mm
Protective case (option) Dimensions 270 x 203 x 330 mm (H x W x D)
Material Stainless steel 1.4301 (AISI 304)
Degree of protection of case IP54
Weight Approx. 7 kg
Ambient temperature −20 to +45 °C
Dimensions and connections (dimensions in mm)
1 Mounting for the inlet filter 2 Inlet filter (inner filter) with internal heating rod 3 Gas sampling tube 4 Wall tube with flange 5 Sample gas outlet or test gas/purge air inlet Minimum distance xmin of the mounting flange from the wall depending on mounting angle α: α 10° 15° 20° 25° 30° 35° xmin/mm 133 138 143 147 151 153
Ordering information
Catalog No. 1200 mm, 230 V, 50/60 Hz, 650 W 23022-4-0768301 1200 mm, 230 V, 50/60 Hz, 255 W, partially heated 23022-4-0768305 Protection box, insulated, IP54 (330 x 203 x 270 mm) 23022-4-0730719
180
650
12
24 SYSTEM COMPONENTS FOR SAMPLE GAS SAMPLING | DS/SCS-EN REV. A
— Probe 11
Use Use in interior air, garage and tunnel monitoring, including corro-sive atmospheres. The materials used ensure that with the gases normally occurring no adsorption occurs.
Technical data Sampling pressure Atmospheric pressure
Gas temperature −25 to +50 °C, dry
Sample gas flow rate Max. 70 l/h
Dust content Max. 1 mg/m3
Pressure drop with new filter ≤ 10 hPa (mbar) at 70 l/h
Service life of filter (estimated)
d (days) = filter constant (2500) flow rate (l/h) x dust concentration (mg/m3)
Sample gas connection port Bolted connection for flexible pipe with internal diameter 4 mm and external diameter 7 mm
Materials Polycarbonate (probe head), aluminum oxide (filter)
Transport and storage temperature −25 to +70 °C
Mounting On wall or suspended from ceiling
Dimensions (in mm)
Ordering information
Catalog No. Room air probe 11 23021-4-0859585
SYSTEM COMPONENTS FOR SAMPLE GAS SAMPLING | DS/SCS-EN REV. A 25
— Ceramic filter
Use The ceramic filter is used for filtering dust from gases in the gas conditioning systems of analyzer systems when the gas sampling probe does not have a filter. The ceramic filter is positioned up-stream the other devices in the sample gas conditioning chain.
Technical data Flow rate (air) Max. 300 l/h
Gas pressure pabs = 20 kPa to 1.7 MPa (0.2 to 17 bar) at 20 °C filter temperature, max. 200 kPa (2 bar) at 180 °C
Pressure drop 0.1 to 0.85 hPa at 30 to 250 l/h (air)
Dynamic characteristics at Flow rate (air) 30 l/h 60 l/h 90 l/h Dead time Tt 3.7 s 2.3 s 1.7 s 90% time T90 11.5 s 5.5 s 3.9 s
Filter fineness 0.3 µm
Construction, mounting Materials of gas-conducting parts Housing: steel, stainless steel 1.4571 (AISI 316Ti); filter: aluminum oxide; seals: FPM, Cu, Al2O3 fiber
Gas connection ports G¼ internal thread or pipe connectors for pipe of internal diame-ter 4 mm
Filter surface Approx. 212 cm2
Weight Approx. 2.1 kg
Mounting On wall with angle bracket
Dimensions (not to scale, dimensions in mm)
Ordering information
Catalog No. Ceramic filter 23031-5-8018178
incl
. T-h
and
le (
not
show
n)
26 SYSTEM COMPONENTS FOR SAMPLE GAS SAMPLING | DS/SCS-EN REV. A
— Process pressure reducers
Use If the pressure of the process gas is too high for the analyzer, a pressure reducer is required to lower the pressure. The type of pressure reducer used is determined by the properties of the process gas and the required sample gas pressure.
Technical data and ordering information
Single-stage with one manometer (outlet)
Single-stage with two manometers (inlet/outlet)
Single-stage with two manometers (inlet/outlet)
Inlet pressure Max. 1.5 MPa (15 bar) Max. 3.5 MPa (35 bar) Max. 3.5 MPa (35 bar) Outlet pressure (adjustable) 0.8 to 10 kPa (8 to 100 mbar) 10 to 40 kPa (0.1 to 0.4 bar) 10 to 70 kPa (0.1 to 0.7 bar) Safety valve set to – 50 kPa (0.5 bar) 80 kPa (0.8 bar) Manometer display inlet pressure – 0 to 4 MPa (0 to 40 bar) 0 to 4 MPa (0 to 40 bar) Manometer display outlet pressure 0 to 250 hPa (0 to 250 mbar) 0 to 60 kPa (0 to 0.6 bar) −100 to +150 kPa (−1 to +1.5 bar) Flow rate 25 to 250 l/h 25 to 250 l/h 25 to 250 l/h Material diaphragm Buna rubber Buna rubber Stainless steel Material gasket Buna rubber Buna rubber EPDM Material housing Brass Stainless steel; gas-conducting parts: brass Stainless steel Weight Approx. 1.25 kg Approx. 2.65 kg Approx. 2.35 kg Gas and ambient temperature Max. 50 °C Max. 50 °C Max. 50 °C Scope of supply Pressure reducer with angle bracket Pressure reducer with angle bracket Pressure reducer with angle bracket
Catalog No. 23035-5-5864958 23035-5-5864951 23422-4-0730700
Dimensions (in mm) Pressure reducer single-stage with one manometer (outlet) (Catalog No. 23035-5-5864958)
Pressure reducer single-stage with two manometers (inlet /outlet) (Catalog No. 23035-5-5864951)
Pressure reducer single-stage with two manometers (inlet /outlet) (Catalog No. 23422-4-0730700)
118
Ø4
GasTech148
approx. 94(depending on handwheel position)
approx. 30° (variable)
¼
NPT
F
¼
NPT
F
¼
NPT
F
¼
NPT
F
¼ NPTF
SYSTEM COMPONENTS FOR SAMPLE GAS SAMPLING | DS/SCS-EN REV. A 27
— Electronic temperature controller and switching relay
Temperature controller Control range −200 to +850 °C
Measurement input Pt 100
Outputs 2 x relay 3 A, 250 V AC
Power supply 110 to 240 V AC +10/−15 %, 48 to 63 Hz
Dimensions 48 x 48 mm, C-rail housing
Weight 140 g
Switching relay Load voltage Urms = 24 to 265 V
Max. load current Irms = 30 A (ambient temperature = 25 °C)
Weight 200 g
Ordering information
Catalog No. Temperature controller Parameters not set 23345-5-8328907 Parameters set for self-regulating heater for ACMP, ACX
23345-5-8328907C
Parameters set for sample gas line 23345-5-8328907L Parameters set for filter unit 23345-5-8328907P Parameters set for probe 42 23345-5-8328907R Switching relay 23345-5-8328731
Electrical connections
L1 N 342 343 341 142 143
83242 243 11280 111 11381
GN
D
++
0/12V230V/3A
/3A230V
110.
..240
VA
C-1
5/+
10%
48...
63 H
z
4+ 3-
1 2
M2A
max.max.
L
L
N
Last
Logi
cO
utp
uts
Bin
ary
Out
put
2
Bin
ary
Out
put
1
not
used
Alarm 1Temperature
use coupling relayFor load current > 2 A
Alarm 2Limiter
28 SYSTEM COMPONENTS FOR SAMPLE GAS SAMPLING | DS/SCS-EN REV. A
— Heated sample gas lines
CGWB 13 Heated sample gas line • Parallel strip heater (holding temperature max. 200 °C)
• Pt 100 resistance thermometer for regulation
• Self-regulating anti-frost heater (max. 30 °C)
• Self-regulating heater (option, holding temp. max. 100 °C)
TBL 01-S Heated sample gas line • Fixed resistor heat. conductor (holding temp. max. 200 °C)
• Pt 100 resistance thermometer for regulation
• Self-regulating heater (option, holding temp. max. 100 °C)
TBL 01-C Heated sample gas line • Self-regulating heater (holding temp. max. 100/120 °C)
• Integral pipe-end connectors for cement analysis systems
Selection table for heated sample gas lines
CGWB 13 TBL 01-S TBL 01-C Heater Self-regulating strip heater (30 °C) x – – Self-regulating strip heater (100 °C) x x x Self-regulating strip heater (120 °C) – – x Fixed resistor heat. cond. (200 °C) – x – Parallel strip heater (200 °C) x – –
Outer covering Polyvinyl chloride (PVC) Polyamide (PA) Metal or PVC
Options Line can be shortened x with self-regulating strip heater: yes
with fixed resistor heating conductor: no x
Small bending radius – x x Gas connections Various options Various options Various options Electrical connections Heater and Pt 100 fixed Heater and Pt 100 fixed Heater and Pt 100 fixed Maximum line length 130 m 100 m 70 m
Applications CO, O2, CO2 Anti-frost heater – – CO, O2, CO2 Holding temperature 100 °C Holding temperature 100 °C – SO2, NOx, CnHm Holding temperature 200 °C Holding temperature 200 °C – Cement analysis systems – – Holding temperature ≤ 120 °C
Technical data see … Page 30 Page 32 Page 35
SYSTEM COMPONENTS FOR SAMPLE GAS SAMPLING | DS/SCS-EN REV. A 29
Integral pipe-end connectors Flange Silicone cap
Metal cap POM cap and shrunk-on hose
Hard cap
30 SYSTEM COMPONENTS FOR SAMPLE GAS SAMPLING | DS/SCS-EN REV. A
— CGWB 13 Heated sample gas line
Technical data
Frost protected line Self-regulating strip heater Regulated parallel strip heater Outer covering PVC PVC PVC Medium-carrying pipe PTFE PTFE PTFE Integral pipe-end connectors Various options (see ordering info.) Various options (see ordering info.) Various options (see ordering info.) Heater self-regulating strip heater self-regulating strip heater CPD heating element, regulated Power supply 230 V or 120 V, 50 to 60 Hz 230 V or 120 V, 50 to 60 Hz 230 V or 120 V, 50 to 60 Hz Heat output 16 W/m 30 W/m 60 W/m Current consumption Making/holding current = 65/25 mA 0.15 A/m 0.3 A/m Holding temperature see diagram see diagram 200 °C Ambient temperature −30 to +60 °C −30 to +60 °C −30 to +60 °C Outer diameter 30 mm 43 mm 40 mm Flexible sample gas pipe 6 x 4 x 1 mm or 8 x 6 x 1 mm 6 x 4 x 1 mm or 8 x 6 x 1 mm 6 x 4 x 1 mm or 8 x 6 x 1 mm Maximum line length 130 m 90 m 60 m Line duct M42 M42 M42 Minimum bending radius 300 mm 300 mm 300 mm Weight Approx. 1 kg/m Approx. 1 kg/m Approx. 1 kg/m Connect. lead for heater (fixed) 3 x 2.5 mm2; length 5 m 3 x 2.5 mm2; length 5 m 3 x 2.5 mm2; length 5 m Connect. lead for Pt 100 (fixed) – – LIYCI 2 x 0.14 mm2; length 5 m
Holding temperature diagrams
Ordering information
Catalog No. 23281-0- 0 0 0 0 0 0 0 Version Parallel strip heater 60 W/m, line can be shortened, regulated (Pt 100), 200 °C 2 Self-regulating 16 W/m, 30 °C (antifreeze) 3 Self-regulating 30 W/m, 100 °C 7 Integral pipe-end connectors Probe: metal cap for PFE2/3 – analysis: plastic cap with shrunk-on hose 2 Probe: metal cap for PFE2/3 – analysis: flange 3 Both sides with POM plastic cap (for self-regulated versions only) 4 Probe: PTFE cap for PFE2/3 in conf. set – analysis: POM plastic cap 5 Probe: metal cap for PFE2/3 in conf. set – analysis: plastic cap with shrunk-on hose 6 Gas conducting line PTFE tube 6x4x1 mm 1 PTFE tube 8x6x1 mm 2 Power supply 230 V, 50 to 60 Hz 1 120 V, 50 to 60 Hz 2 Controlling and monitoring without Pt 100 (for self-regulated versions only) 0 with 1 x Pt 100 (fixed built-in) 1
50
40
30
20
- 30 - 20 - 10 + 10 + 20 + 30 + 400
Frost protected line, 6 x 4 x 1 mm PTFE, 16 W/m, self-regulating
Me
diu
mp
ipe
tem
pe
ratu
re°C
Ambient temperature °C
100
90
80
70
- 30 - 20 - 10 + 10 + 20 + 30 + 400
Ambient temperature °C
Me
diu
mp
ipe
tem
pe
ratu
re°C
Self-regulating strip heater, 6 x 4 x 1 mm PTFE, 30 W/m
SYSTEM COMPONENTS FOR SAMPLE GAS SAMPLING | DS/SCS-EN REV. A 31
Dimensions (in mm) CGWB 13 Self-regulating strip heater CGWB 13 Regulated parallel strip heater
42
300
PO
Mcap
Shrinkin
ghose
Sam
ple
take-o
ffsid
eA
naly
zer
sid
e
Pow
er
Supply
Cable
3x
1.5
...2
.5m
m²
46
46
300
Length
by
custo
mer’
sre
quest
max.le
ngth
dependin
gon
MC
B
Pro
cess
tube
overh
ang,
SS
-nip
ple
6x1
or
8x1
mm
46
42
63
49
25
54
30
0
Le
ng
thb
ycu
sto
me
r’s
req
ue
st
ma
x.
len
gth
de
pe
nd
ing
on
MC
Bo
rsw
itch
ing
loa
do
fth
eco
ntr
olle
rre
sp
ective
ly
PO
Mca
p
Sh
rin
kin
gh
ose
Sa
mp
leta
ke
-off
sid
eA
na
lyze
rsid
e
Re
pla
ca
ble
tub
en
ipp
le,
31
6ss,
6x1
or
8x1
mm
Pt1
00
Ca
ble
2x
0.1
4m
m²
Po
we
rS
up
ply
Ca
ble
3x
1.5
...2
.5m
m²
32 SYSTEM COMPONENTS FOR SAMPLE GAS SAMPLING | DS/SCS-EN REV. A
— TBL 01-S Heated sample gas line
Technical data TBL 01-S Self-regulating strip heater TBL 01-S Fixed resistor heating conductor Outer covering Polyamide corrugated tubing Polyamide corrugated tubing Medium-carrying pipe PTFE pipe 6 x 4 x 1 mm or 8 x 6 x 1 mm PTFE pipe 6 x 4 x 1 mm or 8 x 6 x 1 mm Test gas pipe (unheated, option) PTFE pipe 6 x 4 x 1 mm (500 mm protruding) PTFE pipe 6 x 4 x 1 mm (500 mm protruding) Integral pipe end connectors Probe: metal cap, analyzer: flange or hard caps on both sides (pipe protruding 300 mm) or silicone caps
on both sides (probe: 6/4 stainless steel connections, analyzer: pipe protruding 300 mm) or PTFE caps with 6/4 stainless steel connections on both sides, test gas pipe protruding 500 mm on both sides
Power supply 230 V or 120 V, 50 to 60 Hz 230 V or 120 V, 50 to 60 Hz Heat output 38 W/m 90 W/m Holding temperature see diagram on page 30 200 °C Ambient temperature −30 to +60 °C −30 to +60 °C Outer diameter 43 mm 43 mm Maximum line length 100 m 40 m Line duct M48 M48 Minimum bending radius 200 mm 200 mm Weight Approx. 1 kg/m Approx. 1 kg/m Electrical connection for heater and Pt 100 Fixed, 3 x 2.5 mm2, length 5 m Fixed, 3 x 2.5 mm2 or 2 x 0.14 mm2, length 5 m
Ordering information Catalog No. 23283-0- 0 0 0 0 0 0
Version Fixed resistor heating strip, 90 W/m, holding temperature 200 °C (for ACX) 1 Self-regulated heating strip, 38 W/m, holding temperature 100 °C 2 Fixed resistor heating strip 90 W/m, holding temperature 200 °C (for ACF) 4 Self-regulated heating strip, 38 W/m, holding temperature 100 °C (for SCK) 6 Gas conducting line with PTFE tube 6x4x1 mm 1 with PTFE tube 8x6x1 mm (ACF) 3 with PTFE-tube 6x4x1 mm incl. an additional unheated tube for test gas 6x4x1 mm (500 mm protruding) 4 with PTFE-tube 8x6x1 mm incl. an additional unheated tube for test gas 6x4x1 mm (500 mm protruding, ACF) 5 Material of sheathing with plastic hose of PA (outside and indoors installation) 1 Connections Metal caps w. 6/4 SS connectors on both sides (ACF) A Probe: metal cap – analyzer: flange, 6/4 SS connectors on both sides B Plastic hard caps on both sides (tube 300 mm protruding, self-regulated only) C Silicone caps both sides – probe 6/4 SS conn. – analyzer: tube 300 mm protruding (SCK) D PTFE caps with 6/4 SS connectors on both sides (ACF) E PTFE caps with 6/4 SS connectors on both sides for sample gas and tube for test gas 500 mm protruding on both sides F Metal caps with 6/4 SS connectors on both sides for sample gas and tube for test gas 500 mm protruding on both sides G Controlling and monitoring without Pt 100 (self-regulated version only) 0 with 1 x Pt 100 1 with 2 x Pt 100 2 Power supply 230 V, 50 to 60 Hz, for self-regulated heating strip (for SCK, not for ACF) 1 120 V, 50 to 60 Hz, for self-regulated heating strip (for SCK, not for ACF) 2 230 V, 50 to 60 Hz, for fixed resistor heating conductor 90 W/m 3 120 V, 50 to 60 Hz, for fixed resistor heating conductor 90 W/m 4 230 V, 50 to 60 Hz, for fixed resistor heating conductor 90 W/m (for ACX, ACF, length ≤ 35 m) 5 120 V, 50 to 60 Hz, for fixed resistor heating conductor 90 W/m (for ACX, ACF, length ≤ 15 m) 6 400 V/230 V, 50 to 60 Hz, for fixed resistor heating conductor 90 W/m (for ACX, ACF, length = 35 to 60 m) 7 208 V/120 V, 50 to 60 Hz, for fixed resistor heating conductor 90 W/m (for ACX, ACF, length = 15 to 40 m) 8 120 V, 50 to 60 Hz, self-regulated heating strip (for SCK with AWG wiring) 9
Additional ordering information BA No. Fixed resistor heating strip 90 W/m, 6x4x1 mm, length … m 301 Fixed resistor heating strip 90 W/m, 8x6x1 mm, length … m 303 Self-regulated heating strip 38 W/m, 6x4x1 mm, length … m 304 Self-regulated heating strip 38 W/m, 6x4x1 mm, length … m 307 Self-regulated heating strip 38 W/m, 8x6x1 mm, length … m 309 Fixed resistor heating strip, PTFE tube 6x4x1 mm incl. add. unheated PFTE tube for test gas 6x4x1 mm (500 mm protruding), length … m 311 Fixed resistor heating strip, PTFE tube 8x6x1 mm incl. add. unheated PFTE tube for test gas 6x4x1 mm (500 mm protruding), length … m 313 Fixed resistor heating strip 90W/m, 400V/230V, 50 to 60 Hz, (ACF, line length 35 to 60m) with PTFE tube 8x6x1 mm, length … m 315 Fixed resistor heating strip 90W/m, 208V/120V, 50 to 60 Hz, (ACF, line length 15 to 40m) with PTFE tube 8x6x1 mm, length … m 317
SYSTEM COMPONENTS FOR SAMPLE GAS SAMPLING | DS/SCS-EN REV. A 33
Dimensions (in mm) TBL 01-S, pipe end connectors with hard caps TBL 01-S, pipe end connectors with flange and metal cap
34 SYSTEM COMPONENTS FOR SAMPLE GAS SAMPLING | DS/SCS-EN REV. A
— … TBL 01-S Heated sample gas line
Dimensions (in mm) TBL 01-S, pipe end connectors with metal caps for ACF
SYSTEM COMPONENTS FOR SAMPLE GAS SAMPLING | DS/SCS-EN REV. A 35
— TBL 01-C Heated sample gas line
Technical data
TBL 01-C Self-regulating strip heater Outer covering Metal corrugated tubing (for ACK) or PVC outer covering (for ACMP, ACX) Medium-carrying pipe PTFE pipe 6 x 4 x 1 mm Integral pipe end connectors Probe: metal cap – analyzer: silicone cap (mandatory for ACK) or POM plastic caps on both sides (man-
datory for ACMP, ACX) or probe: PTFE cap – analyzer: POM plastic cap (mandatory for ACMP, ACX 120 °C) Power supply 230 V or 120 V, 50 to 60 Hz Heat output 30 W/m or 47 W/m Holding temperature 100 °C or 120 °C Ambient temperature −30 to +60 °C Outer diameter 43 mm Maximum line length 70 m Line duct M48 Minimum bending radius 200 mm Weight Approx. 1 kg/m Electrical connection for heater and Pt 100 Fixed, 3 x 2.5 mm2 or 2 x 0.14 mm2, length 5 m
Ordering information
Catalog No. 23285-0- 0 0 0 0 0 0 Design Self-regulating strip heater, 30 W/m, 100 °C (ACK) 2 Self-regulating strip heater, 30 W/m, 100 °C (ACMP, ACX coal bin) 3 Self-regulating strip heater, 47 W/m, 120 °C (ACMP, ACX pre-heater) 4 Gas conducting line with PTFE tube 6x4x1 mm 1 Material of sheathing with metal corrugated tubing for ACK (outdoor and indoor installation) 1 with PVC sheathing for ACMP, ACX (outdoor and indoor installation) 2 Connections Probe: metal cap, analyzer: silicone cap (mandatory for ACK) + shortening kit 1 POM plastic caps on both sides (mandatory for ACMP, ACX) + shortening kit 2 Probe: PTFE caps, analyzer: POM caps (mandatory for ACMP, ACX 120 °C) + shortening kit 3 Control/monitoring with 1 x Pt 100 1 Power supply 230 V, 50 to 60 Hz 1 120 V, 50 to 60 Hz 2
36 SYSTEM COMPONENTS FOR SAMPLE GAS SAMPLING | DS/SCS-EN REV. A
— … TBL 01-C Heated sample gas line
Dimensions (in mm) TBL 01-C, pipe end connectors for ACK TBL 01-C, pipe end connectors for ACMP and ACX
300
51
45
49
81
63
54
25
Sam
ple
take-o
ffsid
eA
naly
zer
sid
e
Meta
llcorr
ugate
dhose
Length
by
custo
mer’
sre
quest
max.le
ngth
dependin
gon
MC
Bor
sw
itchin
glo
ad
ofth
econtr
olle
rre
spectively
Pro
cess
tube
overh
ang
Sili
cone
cap
Pt1
00
Cable
2x
0,1
4m
m²
Pow
er
supply
Cable
3x
1,5
...2
,5m
m²
Repla
cable
tube
nip
ple
,316ss,6x1
mm
42
300
46
46
300
PV
CJacket
PO
Mcap
Shrinkin
ghose
Sam
ple
take-o
ffsid
eA
naly
zer
sid
e
Pow
er
Supply
Cable
3x
1.5
...2
.5m
m²
Length
by
custo
mer’
sre
quest
max.le
ngth
dependin
gon
MC
B
Pro
cess
tube
overh
ang
Pt1
00
Cable
2x
0.1
4m
m²
SYSTEM COMPONENTS FOR SAMPLE GAS SAMPLING | DS/SCS-EN REV. A 37
— Notes
38 SYSTEM COMPONENTS FOR SAMPLE GAS SAMPLING | DS/SCS-EN REV. A
— … Notes
— ABB Automation GmbH Measurement & Analytics Stierstädter Str. 5 60488 Frankfurt am Main Germany E-mail: [email protected]
abb.com/analytical
DS
/SC
S-E
N R
ev. A
11.
2018
— We reserve the right to make technical changes or modify the contents of this document without prior notice. With regard to purchase orders, the agreed particulars shall prevail. ABB does not accept any responsibility whatsoever for potential errors or possible lack of information in this document. We reserve all rights in this document and in the subject matter and illustrations contained therein. Any reproduction, disclosure to third parties or utilization of its contents – in whole or in parts – is forbidden without prior written consent of ABB.
© Copyright 2018 ABB 3KXG171100R1001