A Systematic Approach to Optimize Real-Time Quantitative RT-qPCR Experiments with the Agilent 2200 TapeStation System Application Note Author Arunkumar Padmanaban Agilent Technologies, Inc. Bangalore, India Nucleic Acid Analysis Abstract This Application Note describes and demonstrates the efficiency of the Agilent 2200 TapeStation system and the RNA ScreenTape assay in RT-qPCR workflows for assessing the integrity of the initial total RNA template. This template is crucial for the successful amplification of the target genes. Moreover, the D1000 ScreenTape assay helps assess the amplified products for sizing and eliminate false positive results. The data presented demonstrate that the 2200 TapeStation system is an ideal quality control platform for RT-qPCR workflow solutions. The Application Note further describes the advantages of presenting the qPCR results in accordance with the MIQE guidelines.

Transcript of A Systematic Approach to Optimize Real-Time Quantitative ... · TapeStation System Application Note...

Page 1: A Systematic Approach to Optimize Real-Time Quantitative ... · TapeStation System Application Note Author ... by designing and reporting experimental ... peak with a T m of 80.3

A Systematic Approach to Optimize Real-Time Quantitative RT-qPCR Experiments with the Agilent 2200 TapeStation System

Application Note

AuthorArunkumar Padmanaban Agilent Technologies, Inc. Bangalore, India

Nucleic Acid Analysis

AbstractThis Application Note describes and demonstrates the efficiency of the Agilent 2200 TapeStation system and the RNA ScreenTape assay in RT-qPCR workflows for assessing the integrity of the initial total RNA template. This template is crucial for the successful amplification of the target genes. Moreover, the D1000 ScreenTape assay helps assess the amplified products for sizing and eliminate false positive results. The data presented demonstrate that the 2200 TapeStation system is an ideal quality control platform for RT-qPCR workflow solutions. The Application Note further describes the advantages of presenting the qPCR results in accordance with the MIQE guidelines.

Page 2: A Systematic Approach to Optimize Real-Time Quantitative ... · TapeStation System Application Note Author ... by designing and reporting experimental ... peak with a T m of 80.3

2

RNA degradationTwo vials of high integrity total RNA samples containing 1 µg each were incubated at 90 °C for different amounts of time (5 and 10 minutes) to generate RNA with different qualities. Following the degradation, the samples were immediately incubated on ice to stop further degradation. A 1 μL amount was taken for analysis using the 2200 TapeStation and 2100 Bioanalyzer systems. The samples were then diluted 1:5, and 2 µL were used in subsequent RT reactions.

Reverse transcriptionA five-point standard curve was generated using total RNA at 0.001, 0.01, 0.1, 1, and 10 ng as final load template amounts. The total RNA was reverse transcribed using the Agilent AffinityScript QPCR cDNA Synthesis Kit (p/n 600559) following the manufacturer guidelines in a 20 µL reaction volume (25 °C for 5 minutes, 42 °C for 15 minutes, 95 °C for 5 minutes) using an Agilent SureCycler 8800 Thermal Cycler. Each standard was run in triplicate. A 4 µL amount of the cDNA was used directly in subsequent qPCR reactions.

importance of the initial quality control of total RNA. Studies have shown that the integrity of the starting material is a crucial factor determining the success of the qPCR assay in amplifying the desired product3. The RNA integrity number equivalent (RINe) provides an instant and objective evaluation of total RNA degradation for the 2200 TapeStation system. RINe is directly comparable in assessing RNA integrity to the widely accepted and highly cited RNA Integrity Number (RIN) of the Agilent 2100 Bioanalyzer system4.

Materials and MethodsRNA purification and analysisHEK293 cells (ATCC) were grown in MEM media with 10 % FBS, 4 mM L-glutamine, and 1 % strep/penicillin antibiotics, and left to reach 90 % confluency. The cells were then trypsinized and collected as pellets containing 1 × 106 cells, and immediately frozen at –80 °C until use. Total RNA was extracted using the Agilent Total RNA Isolation Minikit (p/n 5185-6000) following the manufacturer guidelines. The extracted samples were then analyzed for integrity using the RNA ScreenTape and RNA 6000 Nano assays with 2200 TapeStation and 2100 Bioanalyzer systems respectively following manufacturer guidelines. The samples were also parallel analyzed using NanoDrop spectrophotometer.

IntroductionRT-qPCR is a complex assay and must be carefully optimized for specificity, sensitivity, and reproducibility. In addition to RNA quality, several experimental variables affect the outcome of RT-qPCR assay performance and may lead to false positive results. The Minimum Information for Publication of Quantitative Real-Time PCR Experiments (MIQE)2 guidelines were put forth in 2009 to develop robust experimental design and evaluate the results. The MIQE guidelines help achieve high quality data by designing and reporting experimental conditions that yield reproducible results. This Application Note demonstrates the successful application of the 2200 TapeStation system to optimize real-time quantitative PCR experiments by recapitulating the experimental outline of a previous study with an Agilent 2100 Bioanalyzer1.

The experimental design for this study followed the guidelines and considerations of the MIQE guidelines. The sample extraction and handling procedure, quality control of total RNA, details of commercial kits and reagents, primer sequences, location of the target genes, and qPCR specificity with standard curve and melt analysis are reported at the appropriate sections of the RT-qPCR experimental workflow. The MIQE guidelines direct special attention to the

Page 3: A Systematic Approach to Optimize Real-Time Quantitative ... · TapeStation System Application Note Author ... by designing and reporting experimental ... peak with a T m of 80.3

3

Real-time quantitative PCRPrimers were obtained from Sigma-Aldrich as HPLC purified oligonucleotides. Previously published primer sequences were used in this study1. Primers were designed to target both 5’ and 3’ splice variants of two genes: hypoxanthine phosphoribosyl transferase-1 (HPRT1) and tyrosine 3-monooxygenase/tryptophan 5-monooxygenase activation protein, zeta (YWHAZ). A summary of the primer sequence, gene ID, and amplicon sizes are presented in Table 1. The qPCR reactions were carried out using Agilent Brilliant III Ultra-Fast SYBR® Green QPCR Master Mix (600882) in a final reaction volume of 20 µL, and a final primer concentration of 200 nM. ROX was used as a passive reference dye at a final concentration of 30 nM. A fast thermal cycle of 95 °C

Table 1. Details of the primers used in the study.

Gene Ensembl gene ID Primers Primer sequence Amplicon size (bp)HPRT1 ENSG00000165704 HPRT1 5’ forward 5’-CGTCGTGATTAGTGATGATGAACCAG-3’ 130

HPRT1 5’ reverse 5’-AGCAAGACGTTCAGTCCTGTCC- 3’HPRT1 3’ forward 5’- TTCAGGGATTTGAATCATGTTTGTGTC- 3’ 114HPRT1 3’ reverse 5’-GCGATGTCAATAGGACTCCAGATG-3’

YWHAZ ENSG00000164924 YWHAZ 5’ forward 5’-TTGAGACGGAGCTAAGAGATATCTGC- 3’ 142YWHAZ 5’ reverse 5’-GCAACCTCAGCCAAGTAACGG-3’YWHAZ 3’ forward 5’-CTTCCTTTGCTTGCATCCCACAG-3’ 128YWHAZ 3’ reverse 5’-GACAATGACAGACCATTCAGGATAGG-3’

for 10 minutes, 40 cycles of 95 °C for 20 seconds, 60 °C for 20 seconds with measurement at 60 °C was used. A dissociation curve was generated using a thermal profile of 95 °C for 30 seconds, 60 °C for 30 seconds, and 95 °C for 30 seconds with continuous fluorescent measurement along the ramp from 60 °C to 95 °C. Each standard was run in triplicate, with no reverse transcriptase (NRT) and no template control (NTC).

All primers span exon-exon junctions with exception of YWHAZ, where the 5’ assay has the primers in exons 2 and 3 with a 23.6 kb intron in between. All qPCR experiments were performed on an Agilent Mx3005P qPCR system.

DNA amplicon analysisThe amplified products were analyzed using the D1000 ScreenTape and DNA 1000 assays with the 2200 TapeStation and 2100 Bioanalyzer systems for size assessment.

Data analysisThe qPCR data obtained from the Mx3005P system was analyzed using MxPro software and exported to MS Excel for presentation. The quantification cycle (Cq) was calculated based on baseline corrected normalized fluorescence (dRn). If not stated otherwise, all samples and standards were run in triplicate. The data is presented as mean ± S.D.

Page 4: A Systematic Approach to Optimize Real-Time Quantitative ... · TapeStation System Application Note Author ... by designing and reporting experimental ... peak with a T m of 80.3

4

[FU]

25

20

15

10

5

0

B

×103

6

5

4

3,

2

1

0Size [nt]

Lower 18S 28S

25 200

500

1,000

2,000

4,000

6,000

Sam

ple in

tens

ity [F

U]

A

Size [nt]25 200 500 1,000 2,000 4,000

Figure 2. Electropherogram trace of the total RNA from the Agilent 2200 TapeStation system (A) and Agilent 2100 Bioanalyzer system (B).

Results and DiscussionTotal RNA integrity analysisThe extraction procedure, sample processing, and sample integrity are essential information to be documented in qPCR. This ensures that only the highest quality samples are used in the study to obtain meaningful results. Sample integrity was determined using the 2200 TapeStation and 2100 Bioanalyzer systems before the qPCR validation. Total RNA was analyzed in replicates of six. The gel images and electropherogram traces showed two intact peaks (28S and 18S) with no degradation products, confirming intact RNA (Figures 1 and 2). The integrity numbers for the samples were expressed as average RINe and RIN of 10 and 9.7 by the 2200 TapeStation and 2100 Bioanalyzer systems respectively. Additionally, purity analysis on the extracted RNA by the NanoDrop spectrophotometer gave an absorbance ratio (A260/A280) of 2.0 for all samples, confirming the absence of contaminants and average yields of 25 µg per pellet. The samples were quantified as 111.8 ng/µL (26 µg per pellet) and 99.5 ng/µL (23 µg per pellet) by the 2200 TapeStation and 2100 Bioanalyzer systems respectively.

6,000

[nt]A

A1 (L) B1 C1 D1 E1 F1 G1 [nt]

BLadder

L 1 2 3 4 5 6

RNA 1 RNA 1 RNA 2 RNA 2 RNA 3 RNA 3

4,000

2,000

1,000

500

200

25

RINe

10.0RINe

10.0RINe

10.0RINe

10.0RINe

10.0RINe

10.0RIN9.8

RIN9.6

RIN9.9

RIN9.7

RIN9.6

RIN9.7

4,000

2,000

1,000

500

200

25

Figure 1. Gel image of total RNA analysis carried out by the Agilent 2200 TapeStation system (A) and Agilent 2100 Bioanalyzer system (B) with an average RINe and RIN of 10 and 9.7 respectively.

Page 5: A Systematic Approach to Optimize Real-Time Quantitative ... · TapeStation System Application Note Author ... by designing and reporting experimental ... peak with a T m of 80.3

5

QPCR validation and amplicon analysisThe primer specificity was validated following the MIQE guidelines by generating a five-point standard curve with 10 fold dilutions with each standard run in triplicate. The slope of the standard curve (R2), PCR efficiency, Cq, and dissociation curve of the controls were evaluated. Primer validation for specificity was carried out using intact total RNA with RINe above 9.0 as template.

Figure 3. Amplification plots (A and C) and dissociation curve (B and D) data from YWHAZ-5’ and HPRT1-3’ primer pairs, respectively.

0

1,000

2,000

3,000

4,000

5,000

6,000

60 65 70 75 80 85

Fluo

resc

ence

(–R'

(T))

Fluo

resc

ence

(–R'

(T))

Temperature (°C)

1 pg load10 pg load100 pg load1 ng load10 ng loadNRTNTC

B

0

1,000

2,000

3,000

4,000

5,000

6,000 1 pg load10 pg load100 pg load1 ng load10 ng loadNRTNTC

D

-0.3

0.7

1.7

2.7

3.7

4.7

5.7

Fluo

resc

ence

(dRn

)

1 pg load10 pg load100 pg load1 ng load10 ng loadNRTNTC

C

-0.3

0.7

1.7

2.7

3.7

4.7

5.7

1 4 7 10 13 16 19 22 25 28 31 34 37 40

Fluo

resc

ence

(dRn

)

Cycles

60 65 70 75 80 85Temperature (°C)

1 4 7 10 13 16 19 22 25 28 31 34 37 40Cycles

YWHAZ-5' Assay-amplification plot YWHAZ-5' Assay-dissociation curve

HPRT1-3' Assay-amplification plot HPRT1-3' Assay-dissociation curve

1 pg load10 pg load100 pg load1 ng load10 ng loadNRTNTC

A

Efficiency 90.9 %, Slope –3.561, R2 = 0.987

Efficiency 93 %, Slope –3.501, R2 = 0.978

A five-point standard curve with total RNA input load ranging from 0.001 ng to 10 ng was generated for each primer pair. The RNA was reverse transcribed into cDNA using oligo (dT) primer. The synthesized cDNA was used as template for qPCR amplifications. NRT was used to indicate genomic DNA (gDNA) contamination and NTC was used as negative control.

Figure 3 shows an amplification plot and dissociation curve data from YWHAZ-5’ and HPRT-3’ to illustrate the results. The flat line for NRT shows that there is no gDNA carryover from the Agilent total RNA isolation mini kit, and the NTC shows that the reaction is free from cross contaminations. The dissociation curve of YWHAZ-5’ and HPRT1-3’ shows a single peak with a Tm of 80.3 °C and 79 °C for all samples across the concentration range. The dissociation curve of the NRT and NTC also shows no presence of amplicons or primer dimers.

Page 6: A Systematic Approach to Optimize Real-Time Quantitative ... · TapeStation System Application Note Author ... by designing and reporting experimental ... peak with a T m of 80.3

6

The amplicon sizes were further analyzed using the 2200 TapeStation and 2100 Bioanalyzer systems with the D1000 ScreenTape and DNA 1000 assay. The sizing data from both systems were compared to the theoretical sizes, and are presented in Figure 4 with a positive correlation between both systems with a sizing accuracy of ± 5 %. Typical electropherogram traces presented in Figure 5 show an overlay of YWHAZ-5’ products from 0.001 ng load with average size of 140 bp, NRT, and NTC controls from the 2200 TapeStation and 2100 Bioanalyzer systems respectively.

5

4

3

2

1

[FU]

30

40

50

60

70

80

20

10

0

-10

Size (bp)

Size (bp)

25 50 100

200

300

400

500

700

1,000

1,500

Sam

ple in

tens

ity [F

U]

A

B

15 100

200

300

400

500

700

1,500

1 pg YWHAZ-5’NRTNTCF2

E2B1

1: 1 pg YWHAZ-5’11: NRT12: NTC

×103

Figure 5. Electropherogram trace overlay from the YWHAZ-5’ assay, showing amplicon data from a 1 pg load (red), NRT (blue), and NTC (green) using the Agilent 2200 TapeStation system (A) and Agilent 2100 Bioanalyzer system (B).

0

50

100

150

200

HPRT1-5' HPRT1-3' YWHAZ-5' YWHAZ-3'Si

ze (b

p)

Amplicon size determination using theAgilent 2200 TapeStation and Agilent 2100 Bioanalyzer systems

TheoreticalAgilent 2200 TapeStation systemAgilent 2100 Bioanalyzer system

Figure 4. Amplicon sizing data from the Agilent 2200 TapeStation (dark blue) and Agilent 2100 Bioanalyzer (light blue) systems. The theoretical sizes are shown in green. Error bars indicate standard deviation.

Page 7: A Systematic Approach to Optimize Real-Time Quantitative ... · TapeStation System Application Note Author ... by designing and reporting experimental ... peak with a T m of 80.3

7

Figure 7B. The Cq was determined for all primer pairs. The Cq value of the untreated and highly intact RNA sample was taken as a reference. The relative Cq difference of the treated RNA templates was calculated, and is presented in Figure 8. The figure suggests that the 5’ end of the HPRT1 gene is more susceptible to degradation compared to the respective 3’ end. YWHAZ shows a similar level of impact for both 5’ and 3’ end mRNA transcripts. These results clearly show that differential degradation is specific to the respective gene products.

systems and designated as high, medium, and low integrity based on RINe.

Figure 6 shows gel images from the 2200 TapeStation and 2100 Bioanalyzer systems showing RNA with three different levels of degradation and their respective integrity number. An electropherogram trace overlay of the three RNA samples from the 2200 TapeStation system is presented in Figure 7A.

qPCR amplifications were carried out using different integrities of total RNA. The amplification plots are shown in

Monitoring the impact of RNA degradation on qPCR experiments

The importance of using only high integrity RNA samples for RT-qPCR experiments as recommended by MIQE guidelines was shown by carrying out RT-qPCR amplification using heat-degraded RNA as a template. The differentially heat-degraded RNA and untreated RNA were analyzed using the 2200 TapeStation and 2100 Bioanalyzer systems and designated as high, medium, and low integrity based on RINe. The samples were analyzed using the 2200 TapeStation and 2100 Bioanalyzer

Figure 6. Gel image showing the respective RNA integrity numbers of untreated and heat degraded RNA samples. RINe for the Agilent 2200 TapeStation (A), and RIN for the Agilent 2100 Bioanalyzer (B) systems.

6,000

[nt]A

A1 (L) B1 C1

T = 0 min T = 5 min T = 10 min

D1 E1 F1 G1 H1 A2 B2 [nt]

BLadder T=0 RNA T=0 RNA T=0 RNA T=5 RNA T=5 RNA T=5 RNA T=10 RNA T=10 RNA

4,0002,000

1,000500200

25RINe

9.8RINe

9.8RINe

9.8RINe

4.5RINe

4.5RINe

4.6RINe

2.3RINe

2.2RINe

2.2RIN9.6

RIN9.7

RIN9.6

RIN4.5

RIN4.7

RIN4.5

RIN3.3

RIN2.3

RIN2.9

4,000

2,000

1,000

500200

25

T=10 RNA

Figure 7. A) An Agilent 2200 TapeStation system electropherogram trace overlay for RNA samples that are highly intact (green), medium level of degradation (blue), and highly degraded (red). B) Amplification plot obtained from the RNA of different integrity using HPRT1-3’ primers.

4

5

3

2

1

0Size [nt]

25 200

500

1,00

0

2,00

0

4,00

06,

000

Sam

ple

inte

nsity

[FU]

A

-1

0

1

2

3

4

5

6

0 10 20 30 40

Fluo

resc

ence

(dRn

)

Cycles

Amplification plots HPRT1-5’RNA degradation

High intactMedium intactDegradedNRTNTC

BT = 0 minsT = 5 minsT = 10 minsH1

E1B1

×103

Page 8: A Systematic Approach to Optimize Real-Time Quantitative ... · TapeStation System Application Note Author ... by designing and reporting experimental ... peak with a T m of 80.3

www.agilent.com/genomics/tapestation

For Research Use Only. Not for use in diagnostic procedures.

This information is subject to change without notice.

© Agilent Technologies, Inc., 2015, 2017 Published in the USA, August 1, 2017 5991-4971EN

ConclusionsThe performance of the RNA ScreenTape assay using the Agilent 2200 TapeStation system in assessing the integrity of total RNA was compared with the Agilent 2100 Bioanalyzer system and shown to be equal. Following the experimental outline of a previous study, the impact of the RNA integrity on the qPCR application was assessed by artificial degradation of total RNA samples1. The results from this study recapitulate and confirm the previous findings that RNA integrity impacts the amplification of the target gene, and that directionality is gene dependent. Thus, this study demonstrates that the RNA ScreenTape assay can be reliably used in determining the integrity of total RNA samples. In addition to RNA quality control, the 2200 TapeStation system can be used for subsequent sizing analysis of PCR products and provide an ideal system to fit into any qPCR workflow in accordance with the MIQE guidelines.

-6.0

-5.0

-4.0

-3.0

-2.0

-1.0

0.05' Assay 3' Assay 5' Assay 3' Assay

HPRT1 YWHAZ

Rela

tive

C q diff

eren

ce

Relative Cq difference between high integrity RNA versus medium integrity and degraded RNA

Medium integrityDegraded

Figure 8. Relative Cq for different degradation stages of template RNA.

References1. Mueller, S. Optimizing real-time

quantitative PCR experiments with the Agilent 2100 Bioanalyzer, Agilent Technologies Application Note, publication number 5989-7730EN, 2008. http://www.chem.agilent.com/library/applications/5989-7730EN.pdf.

2. Bustin, S.; et al. The MIQE guidelines: minimum information for publication of quantitative real-time PCR experiments. Clinical chemistry 2009, 55, pp 611-22.

3. Imbeaud, S.; et al. Towards standardization of RNA quality assessment using user-independent classifiers of microcapillary electrophoresis traces. Nucleic Acids Research 2005, 33, e56-e56.

4. Connelly, M., et al. Performance of the Agilent RNA ScreenTape and the High Sensitivity RNA ScreenTape Assay for the Agilent 2200 TapeStation System, Agilent Technologies Application Note, publication number 5991-3426EN, 2013. http://www.chem.agilent.com/library/technicaloverviews/Public/5991-3426EN.pdf