A pixel chamber as beam monitor for IMRT S. Belletti 4, A.Boriano 6, F.Bourhaleb 5,7, R. Cirio 6, M....
-
Upload
tamsyn-chase -
Category
Documents
-
view
215 -
download
0
Transcript of A pixel chamber as beam monitor for IMRT S. Belletti 4, A.Boriano 6, F.Bourhaleb 5,7, R. Cirio 6, M....
A pixel chamber as A pixel chamber as beam monitor for beam monitor for
IMRT IMRT S. Belletti S. Belletti 44, A.Boriano , A.Boriano 66, F.Bourhaleb , F.Bourhaleb 5,75,7, R. Cirio, R. Cirio 6 6, M. Donetti , M. Donetti
5,65,6, B. Ghedi , B. Ghedi 44, E. Madon , E. Madon 22, F. Marchetto , F. Marchetto 66, U. Nastasi , U. Nastasi 33, C. , C. Peroni Peroni 66, C.J. Sanz Freire , C.J. Sanz Freire 1,61,6 , E. Trevisiol , E. Trevisiol 22
11 IBA, Ion Beam Applications, Louvain la-Neuve, Belgium IBA, Ion Beam Applications, Louvain la-Neuve, Belgium22 OIRM S. Anna, Torino, Italy OIRM S. Anna, Torino, Italy
33 Servizio di Fisica Sanitaria, S. Giovanni A.S. Hospital, Torino, Italy Servizio di Fisica Sanitaria, S. Giovanni A.S. Hospital, Torino, Italy44 Servizio di Fisica Sanitaria, Spedali Civili, Brescia, Italy Servizio di Fisica Sanitaria, Spedali Civili, Brescia, Italy
55 TERA Foundation, Novara, Italy TERA Foundation, Novara, Italy66 University and INFN, Torino, Italy University and INFN, Torino, Italy
77 University Mohamed First, Oujda, Morocco University Mohamed First, Oujda, Morocco
1st Euro-mediterranean meeting on radiotherapy –Oujda (Morocco) [email protected]
OutlineOutline
IMRT measurementsIMRT measurements
Mechanical structure of the pixel ionisation chamberMechanical structure of the pixel ionisation chamber
Read-out: electronics and performancesRead-out: electronics and performances
Characterization of the detector with hadron beamsCharacterization of the detector with hadron beams
Characterization of the detector with photon beamsCharacterization of the detector with photon beams
1st Euro-mediterranean meeting on radiotherapy –Oujda (Morocco) [email protected]
The pixel The pixel chamberchamber
parallel plate ionization chamber
anode segmented in 1024 square pixels
pixel dimension = 7.5 x 7.5 mm2
sensitive area = 24 x 24 cm2
1 mm water equivalent thickness
front-end electronics, located around the chamber, to perform analog to digital conversion
1st Euro-mediterranean meeting on radiotherapy –Oujda (Morocco) [email protected]
Detector designDetector design
When used in plastic phantom the detector can be thick.
Anode and cathode are glued to a plastic slab, which is a grid of 1024 holes.
Each hole can be considered as an independent sensitive volume and the material around provides lateral electronic equilibrium.
1st Euro-mediterranean meeting on radiotherapy –Oujda (Morocco) [email protected]
Detector designDetector design
1st Euro-mediterranean meeting on radiotherapy –Oujda (Morocco) [email protected]
Front-end electronicsFront-end electronics
located around the chamber
we have developed a full custom chip (TERA05)
every chip has 64 channels. Each one converts the charge collected by a pixel into counts
1st Euro-mediterranean meeting on radiotherapy –Oujda (Morocco) [email protected]
Linearity at 100 fC charge Linearity at 100 fC charge quantumquantum
20 pA < I < 0.6 A
Linearity better than 0.7 %
1st Euro-mediterranean meeting on radiotherapy –Oujda (Morocco) [email protected]
Linearity at 600 fC charge Linearity at 600 fC charge quantumquantum
10 pA < I < 2 A
Linearity better than 0.3 %
1st Euro-mediterranean meeting on radiotherapy –Oujda (Morocco) [email protected]
Reproducibility of charge quantumReproducibility of charge quantum
I = 49.96 nA Qc = 600 fC 24 ºC < T < 27 ºC 1 month 6 measures
1st Euro-mediterranean meeting on radiotherapy –Oujda (Morocco) [email protected]
Beam test at PSI (CH)Beam test at PSI (CH)
PSI test with a 138 MeV proton PSI test with a 138 MeV proton beambeam
7 mm FWHM beam7 mm FWHM beam Spot-scanningSpot-scanning
Uniformity of the Uniformity of the response response ~0.9%~0.9%
Profiles measurement okProfiles measurement ok
1st Euro-mediterranean meeting on radiotherapy –Oujda (Morocco) [email protected]
Beam test at GSI (D)Beam test at GSI (D)
Beam characteristics
raster-scan delivery system
C+6, beam dimension = 8.8 mm (FWHM)
data acquisition synchronized with raster-scan
Aims
spatial resolution
homogeneity of response
1st Euro-mediterranean meeting on radiotherapy –Oujda (Morocco) [email protected]
Spatial resolution: 16 spotSpatial resolution: 16 spot
1st Euro-mediterranean meeting on radiotherapy –Oujda (Morocco) [email protected]
Spatial Spatial resolutionresolution
Spatial resolution < 0.2 mm
1st Euro-mediterranean meeting on radiotherapy –Oujda (Morocco) [email protected]
Homogeneity of the Homogeneity of the responseresponse
18 × 18 cm2 uniform field18 × 18 cm2 uniform field
= 1.1 % = 1.1 %
= 2.0 % = 2.0 %
1st Euro-mediterranean meeting on radiotherapy –Oujda (Morocco) [email protected]
Q.A. with photon beamsQ.A. with photon beams
A Quality Assurance protocol A Quality Assurance protocol establish:establish:
Physic and dosimetric parameters Physic and dosimetric parameters to be measuredto be measured
Procedures followed for the Procedures followed for the measurementsmeasurements
Frequency of the controlsFrequency of the controls
Reference valuesReference values
1. Homogeneity, symmetry, penumbra, energy stability
2. Calibration of the monitor system, stability.
Conventional instrumentation:Conventional instrumentation:
Thimble ion chamber (Farmer)Thimble ion chamber (Farmer) One point One point measurementmeasurement
Diode array 1D measurementDiode array 1D measurement
1st Euro-mediterranean meeting on radiotherapy –Oujda (Morocco) [email protected]
Q.A. with photon beamsQ.A. with photon beamsS. Giovanni A.S. and S. Anna hospitals – Torino
Aims
Depth dose profile measurements and comparison to a reference ion chamber (Farmer)
Profile measurements and comparison Measurements to a reference ion chamber
Stability of the response
To perform these kind of measurements, there are no time constrains. The only information needed is the final measurement
1st Euro-mediterranean meeting on radiotherapy –Oujda (Morocco) [email protected]
Depth dose curveDepth dose curve
Depth dose profiles have been measured for different photon energy.We have used a Farmer ion chamber in a plastic phantom as a referenceAn excellent agreement has been found for each photon energy.
1st Euro-mediterranean meeting on radiotherapy –Oujda (Morocco) [email protected]
ProfilesProfilesProfile comparison with a Farmer in a water phantom
Field 10 X 10 cm2
Depth 10 cm
Field 20 X 20 cm2
Depth 10 cm
1st Euro-mediterranean meeting on radiotherapy –Oujda (Morocco) [email protected]
ProfilesProfiles
Field 5 X 5 cm2
Depth 10 cm
1st Euro-mediterranean meeting on radiotherapy –Oujda (Morocco) [email protected]
Long-term stabilityLong-term stability
Stability measured during a one month periodStability measured during a one month period Stability withinStability within 1.5%1.5% Mechanical problem caused variations in the sensitive Mechanical problem caused variations in the sensitive
volumevolume1.050
0.950
1.000
1.050
0.950
1.000
±1% ±5%
1st Euro-mediterranean meeting on radiotherapy –Oujda (Morocco) [email protected]
IMRT measurementsIMRT measurements
Dynamic wedge case• By moving each side of the
collimator with different speed one obtains a wedge shape of the dose
• At the end of the treatment one has to check the delivered dose in different points
The pixel ion chamber can measure precisely the 2D dose in 1024 points.
1st Euro-mediterranean meeting on radiotherapy –Oujda (Morocco) [email protected]
IMRT measurementsIMRT measurements
The pixel ion chamber has a fast read out.Every 1ms a complete read out can be performed.It is possible to monitor the 2D dose in real time.
Dynamic multi leaves collimator (DMLC) case:• independent leaves can be moved to shape the beam field• one needs to check as a function of time the 2D delivered dose
1st Euro-mediterranean meeting on radiotherapy –Oujda (Morocco) [email protected]
IMRT measurementsIMRT measurements
Total dose delivered
Example: small square dose distribution going to large square dose distribution
1st Euro-mediterranean meeting on radiotherapy –Oujda (Morocco) [email protected]
IMRT measurementsIMRT measurements
Delivered dose in function of time
Time: 0 secTime: 1 secTime: 2 secTime: 3 secTime: 4 secTime: 5 secTime: 6 secTime: 7 secTime: 8 secTime: 9 secTime: 10 secTime: 11 secTime: 12 secTime: 13 secTime: 14 sec
1st Euro-mediterranean meeting on radiotherapy –Oujda (Morocco) [email protected]
1st Euro-mediterranean meeting on radiotherapy –Oujda (Morocco) [email protected]
Detector Detector designdesign
Thin version
When used as monitor the detector has to be thin to minimize the changes on the beam energy and shape.For this configuration, anode and cathode are glued to frames in order to assure the mechanical rigidity and the gap thickness .Water equivalent thickness < 1 mm.
1st Euro-mediterranean meeting on radiotherapy –Oujda (Morocco) [email protected]
Read out electronicsRead out electronics
Recycling Integrator architecture.
1st Euro-mediterranean meeting on radiotherapy –Oujda (Morocco) [email protected]
TERA05TERA05Very Large Scale Integration (VLSI)if conversionoutput Qint
100 fC<charge quantum<800 fCImax= 4 A64 channels16 bit wide countersmultiplexed digital outputdead-time free readoutmax read out rate = 10 MHz
channel # 1
channel # 64
1st Euro-mediterranean meeting on radiotherapy –Oujda (Morocco) [email protected]
Fast Data Fast Data acquisitionacquisition
Signals (RS422 standard) are delivered via twisted pair flat cables (100 m maximum length)
max transfer rate = 10 MHz = 40 Mbyte/s
read out transfer time = 50 s
read out cycle total time = 100 s
real time operating system
<100 m
1st Euro-mediterranean meeting on radiotherapy –Oujda (Morocco) [email protected]
Slow Data acquisitionSlow Data acquisition
• connection by twisted pair flat cables (100 m max)
• max rate transfer = 1 MHz =
2 Mbyte/s
• read out transfer time = 1 ms
• PCI DAQ card
• LabVIEW software
• cheap solution
• easy to handle
Slow data acquisition
1st Euro-mediterranean meeting on radiotherapy –Oujda (Morocco) [email protected]
ReproducibilityReproducibility
Several measurements have been performed to evaluate the reproducibility.After a pre-irradiation reproducibility is better than 1%
100 MU per irradiationfield bigger than sensitive volume
1st Euro-mediterranean meeting on radiotherapy –Oujda (Morocco) [email protected]
OUTPUT FACTOROUTPUT FACTOR
10
..D
DFO r
3.4% No solution so far:No solution so far:
Higher sensitivity of Higher sensitivity of the pixel chamber to the pixel chamber to low energy photons low energy photons due to a different due to a different structurestructure
Disagreement Disagreement between the pixel between the pixel chamber and the chamber and the FarmerFarmer