A Charge-Based Magneto-Electric Test Procedure
description
Transcript of A Charge-Based Magneto-Electric Test Procedure
![Page 1: A Charge-Based Magneto-Electric Test Procedure](https://reader035.fdocuments.us/reader035/viewer/2022062804/56814932550346895db676bb/html5/thumbnails/1.jpg)
Radiant Technologies, Inc.Magneto-Electric Test Procedure
A Charge-Based Magneto-Electric Test Procedure
Scott P. Chapman & Joseph T. Evans, Jr.
Radiant Technologies, Inc.
Aug 9, 2011
IWPMA 2011
![Page 2: A Charge-Based Magneto-Electric Test Procedure](https://reader035.fdocuments.us/reader035/viewer/2022062804/56814932550346895db676bb/html5/thumbnails/2.jpg)
Radiant Technologies, Inc.Magneto-Electric Test Procedure
Summary
The goal is to describe an experiment to characterize the charge response of a piezoelectric or multiferroic sample in the presence of a magnetic (B) field by:
P = H
B = H
P = /B
For a multiferroic, H induces P directly. For our piezoelectric sample, P results from direct force (dc) or torque (d) applied to the sample ferroelectric with a magnet attached.
![Page 3: A Charge-Based Magneto-Electric Test Procedure](https://reader035.fdocuments.us/reader035/viewer/2022062804/56814932550346895db676bb/html5/thumbnails/3.jpg)
Radiant Technologies, Inc.Magneto-Electric Test Procedure
SummaryI will present:
• Mathematics and theory relating predictive and measured polarization response to the magnetic field and magnetic field geometry.
• Experimental considerations.
• Experimental design and configuration.
• Measured results.
• Measured comparison to predictive.
![Page 4: A Charge-Based Magneto-Electric Test Procedure](https://reader035.fdocuments.us/reader035/viewer/2022062804/56814932550346895db676bb/html5/thumbnails/4.jpg)
Radiant Technologies, Inc.Magneto-Electric Test Procedure
Experiment• Experiment – Stimulate sample with
magnetic field and measure charge. Simple.
• Characterize sample independently, calculate B Field and predict charge. Hard.
• Compare the two.
![Page 5: A Charge-Based Magneto-Electric Test Procedure](https://reader035.fdocuments.us/reader035/viewer/2022062804/56814932550346895db676bb/html5/thumbnails/5.jpg)
Radiant Technologies, Inc.Magneto-Electric Test Procedure
Magnetic Force
• Magnetic force is generated only by moving electric charges.
• For two objects to exert magnetic force both must contain moving charges.
These three statements apply to understanding Magnetic Force:
![Page 6: A Charge-Based Magneto-Electric Test Procedure](https://reader035.fdocuments.us/reader035/viewer/2022062804/56814932550346895db676bb/html5/thumbnails/6.jpg)
Radiant Technologies, Inc.Magneto-Electric Test Procedure
Magnetic Force• Magnetic force calculation proceeds as follows:
Calculate a mathematical field, H, that sums the motion of all charge particles at the point of interest in the field.
Multiply H by the magnetic permeability factor, , to convert it to a force field, B.
Use B to calculate magnetic force on the target. This requires the calculation of both the HH coil force and the target force, and their multiplication.
![Page 7: A Charge-Based Magneto-Electric Test Procedure](https://reader035.fdocuments.us/reader035/viewer/2022062804/56814932550346895db676bb/html5/thumbnails/7.jpg)
Radiant Technologies, Inc.Magneto-Electric Test Procedure
Geometry (Along Field Axis)Bsingle coil = 0NIR2 x 0.5(R2+x2)-3/2 (1)
BHHC = 0.5 0NIR2 x 1/(R2+(x+K/2)2)3/2
+ 0.5 0NIR2 x 1/(R2+(x-K/2)2)3/2 (2)
N = Number of Coils R = Coil Radius (m)I = Current Through Loop (Amps) K = Coil Separation (m)x = Location of interest measured from midway between the coils (m)
B = 0.716 0NI/R (3)
For: K = R and x = 0 (Centered Between Coils)
![Page 8: A Charge-Based Magneto-Electric Test Procedure](https://reader035.fdocuments.us/reader035/viewer/2022062804/56814932550346895db676bb/html5/thumbnails/8.jpg)
Radiant Technologies, Inc.Magneto-Electric Test Procedure
Basic Test ConfigurationOrientation 1 – Centered – m || B
![Page 9: A Charge-Based Magneto-Electric Test Procedure](https://reader035.fdocuments.us/reader035/viewer/2022062804/56814932550346895db676bb/html5/thumbnails/9.jpg)
Radiant Technologies, Inc.Magneto-Electric Test Procedure
Basic Test ConfigurationOrientation 2 – Centered – m B
![Page 10: A Charge-Based Magneto-Electric Test Procedure](https://reader035.fdocuments.us/reader035/viewer/2022062804/56814932550346895db676bb/html5/thumbnails/10.jpg)
Radiant Technologies, Inc.Magneto-Electric Test Procedure
Basic Test ConfigurationOrientation 3 – x = 1.5 R – m B
![Page 11: A Charge-Based Magneto-Electric Test Procedure](https://reader035.fdocuments.us/reader035/viewer/2022062804/56814932550346895db676bb/html5/thumbnails/11.jpg)
Radiant Technologies, Inc.Magneto-Electric Test Procedure
Plot Measured Charge Vs Field
H
P
P is Accurately Measured but H may be inferred
Arbitrary Data
![Page 12: A Charge-Based Magneto-Electric Test Procedure](https://reader035.fdocuments.us/reader035/viewer/2022062804/56814932550346895db676bb/html5/thumbnails/12.jpg)
Radiant Technologies, Inc.Magneto-Electric Test Procedure
Independent Field ValuesThe independent (Field) axis in the data presentation can be
determined by:
• Assumed Field by DRIVE Volts into the Current Amplifier. This experiment presented here uses this approach.
• Assumed Field by Measured Current into the Helmholtz Coil. This reduces the number of error sources in the first option by half.
• Measured Field by magnetic sensor. Most accurate.
![Page 13: A Charge-Based Magneto-Electric Test Procedure](https://reader035.fdocuments.us/reader035/viewer/2022062804/56814932550346895db676bb/html5/thumbnails/13.jpg)
Radiant Technologies, Inc.Magneto-Electric Test Procedure
Some Field Profiles
![Page 14: A Charge-Based Magneto-Electric Test Procedure](https://reader035.fdocuments.us/reader035/viewer/2022062804/56814932550346895db676bb/html5/thumbnails/14.jpg)
Radiant Technologies, Inc.Magneto-Electric Test Procedure
Improved Test Configuration - Measure HH Coil Input Current
![Page 15: A Charge-Based Magneto-Electric Test Procedure](https://reader035.fdocuments.us/reader035/viewer/2022062804/56814932550346895db676bb/html5/thumbnails/15.jpg)
Radiant Technologies, Inc.Magneto-Electric Test Procedure
Improved Test Configuration - Direct Field Measurement at Sample
![Page 16: A Charge-Based Magneto-Electric Test Procedure](https://reader035.fdocuments.us/reader035/viewer/2022062804/56814932550346895db676bb/html5/thumbnails/16.jpg)
Radiant Technologies, Inc.Magneto-Electric Test Procedure
Advanced Test Configuration - Introduce a DC Bias Field
![Page 17: A Charge-Based Magneto-Electric Test Procedure](https://reader035.fdocuments.us/reader035/viewer/2022062804/56814932550346895db676bb/html5/thumbnails/17.jpg)
Radiant Technologies, Inc.Magneto-Electric Test Procedure
Program Entry Parameters
![Page 18: A Charge-Based Magneto-Electric Test Procedure](https://reader035.fdocuments.us/reader035/viewer/2022062804/56814932550346895db676bb/html5/thumbnails/18.jpg)
Radiant Technologies, Inc.Magneto-Electric Test Procedure
Measurement Configuration
![Page 19: A Charge-Based Magneto-Electric Test Procedure](https://reader035.fdocuments.us/reader035/viewer/2022062804/56814932550346895db676bb/html5/thumbnails/19.jpg)
Radiant Technologies, Inc.Magneto-Electric Test Procedure
Data Presentation Configuration
![Page 20: A Charge-Based Magneto-Electric Test Procedure](https://reader035.fdocuments.us/reader035/viewer/2022062804/56814932550346895db676bb/html5/thumbnails/20.jpg)
Radiant Technologies, Inc.Magneto-Electric Test Procedure
Direct Force (dc) Measurement
![Page 21: A Charge-Based Magneto-Electric Test Procedure](https://reader035.fdocuments.us/reader035/viewer/2022062804/56814932550346895db676bb/html5/thumbnails/21.jpg)
Radiant Technologies, Inc.Magneto-Electric Test Procedure
Direct Force (dc) Response
![Page 22: A Charge-Based Magneto-Electric Test Procedure](https://reader035.fdocuments.us/reader035/viewer/2022062804/56814932550346895db676bb/html5/thumbnails/22.jpg)
Radiant Technologies, Inc.Magneto-Electric Test Procedure
Torque (d) Measurement
![Page 23: A Charge-Based Magneto-Electric Test Procedure](https://reader035.fdocuments.us/reader035/viewer/2022062804/56814932550346895db676bb/html5/thumbnails/23.jpg)
Radiant Technologies, Inc.Magneto-Electric Test Procedure
Torque (d) Response
![Page 24: A Charge-Based Magneto-Electric Test Procedure](https://reader035.fdocuments.us/reader035/viewer/2022062804/56814932550346895db676bb/html5/thumbnails/24.jpg)
Radiant Technologies, Inc.Magneto-Electric Test Procedure
Measured Piezo Constants
The constants we measured: Numbers are high (~10 x) due to substrate bending.
• Parallel to the magnetic axis (dc):
* 612 pC/N (10 g = 0.098 N)
* 688 pC/N (20 g = 0.196 N)
* 653 pC/N (50 g = 0.49 N)
* 650 ±40 pC/N (Average)
![Page 25: A Charge-Based Magneto-Electric Test Procedure](https://reader035.fdocuments.us/reader035/viewer/2022062804/56814932550346895db676bb/html5/thumbnails/25.jpg)
Radiant Technologies, Inc.Magneto-Electric Test Procedure
Measured Piezo Constants
• Torque (d): 765 pC/N
When applying magnetic torque, the force must be calculated from the lever arm length and then multiplied by the equivalent torque piezoelectric constant
![Page 26: A Charge-Based Magneto-Electric Test Procedure](https://reader035.fdocuments.us/reader035/viewer/2022062804/56814932550346895db676bb/html5/thumbnails/26.jpg)
Radiant Technologies, Inc.Magneto-Electric Test Procedure
Primary Measurement Error SourcesThere are three primary sources of error:
• Frequency response of the current amplifier with the attached HH coil. Slow the measurement to ensure the amplifier can provide the requested HH coil input power.
• Parasitic charge resulting from magnetic induction in the RETURN cable. This effect is reduced by slowing the measurement. Measure the effect and subtract from the final measurement.
• Charge measurement accuracy reduced by charge deterioration over long tests. This effect is reduced by limiting the length of the test.
1000.0 ms is optimal and well within equipment capabilities. Significantly limits inductive charge.
![Page 27: A Charge-Based Magneto-Electric Test Procedure](https://reader035.fdocuments.us/reader035/viewer/2022062804/56814932550346895db676bb/html5/thumbnails/27.jpg)
Radiant Technologies, Inc.Magneto-Electric Test Procedure
Pre-Measurement Steps
• Calculate the magnetic field at the point where the sample is located.
• Measure the induced current in the cable, under measurement test conditions, and retain to subtract from the measured data.
To prepare for the Magneto-Electric Response Task measurement, perform these steps:
![Page 28: A Charge-Based Magneto-Electric Test Procedure](https://reader035.fdocuments.us/reader035/viewer/2022062804/56814932550346895db676bb/html5/thumbnails/28.jpg)
Radiant Technologies, Inc.Magneto-Electric Test Procedure
Pre-Measurement Steps
• Reduce the test speed to reduce inductive current, but no slower than 1000.0 ms.
• Determine through experimentation the maximum frequency and ensure 1000/Test Period (ms) does not exceed this value.
![Page 29: A Charge-Based Magneto-Electric Test Procedure](https://reader035.fdocuments.us/reader035/viewer/2022062804/56814932550346895db676bb/html5/thumbnails/29.jpg)
Radiant Technologies, Inc.Magneto-Electric Test Procedure
Predictive Modelm || B - Centered in HH Coil
Define, for our force-inducing magnet:
m = MV
M = Magnetization of Magnet V = Volume
For B || m
F = [m ·B] (1)
For constant m, as with our reference sample magnets:
F = m B/ x (2)
For constant B, as in the center of the Helmholtz coil:
F = 0 => Q = 0 pC
![Page 30: A Charge-Based Magneto-Electric Test Procedure](https://reader035.fdocuments.us/reader035/viewer/2022062804/56814932550346895db676bb/html5/thumbnails/30.jpg)
Radiant Technologies, Inc.Magneto-Electric Test Procedure
Predictive Modelm B - Centered in HH Coil
Piezo Constant:
d = 0.75 V x 100 pC/10g (Sense Capacitor) =
75 pC/0.098 N = 765 pC/N
Ftorque ():
m = 4 x 1.08 T/4x10-7 x (0.00252 x 0.006) = 0.4 A/m
Estimated Charge (Q) at 45.0 Gauss:
Q = d x 0.4 A/m x B / Height
= 765 pC/N x 0.4 A/m x 45 e-4 T / 0.006 m
= 229.5 pC
![Page 31: A Charge-Based Magneto-Electric Test Procedure](https://reader035.fdocuments.us/reader035/viewer/2022062804/56814932550346895db676bb/html5/thumbnails/31.jpg)
Radiant Technologies, Inc.Magneto-Electric Test Procedure
Predictive Model m || B - At 1 K From Closest Coil
x = 1.5 K = 1.5 R
B = -0.319 0NI/R2
Q = dc x 0.4 A/m x B
= 650 pC/N x 0.4 A/m x -0.093I T/m
At 45 Gauss, I = 1.68 Amps
Q = 650 pC/N x 0.4 A/m x -0.093 x 1.68 A T/m
= -4.06 pC
![Page 32: A Charge-Based Magneto-Electric Test Procedure](https://reader035.fdocuments.us/reader035/viewer/2022062804/56814932550346895db676bb/html5/thumbnails/32.jpg)
Radiant Technologies, Inc.Magneto-Electric Test Procedure
Predictive Modelm B - At 1 K From Closest Coil
At x = 0:
B = 0.716 0NI/R
=> 0NI/R = B/0.716 = 45.0/0.716 = 62.85 G
At x = 1.5 K = 1.5 R:
BHHC = 0.5 0NIR2/(R2+(x+K/2)2)3/2+ 0.5 0NIR2/(R2+(x-K/2)2) 3/2 G
= 0.5 0NIR2/(R2+(1.5R + R/2)2)3/2+ 0.5 0NIR2/(R2+(1.5R-K/2)2)3/2
= 0.1727 0NI/R G = 10.855 G
Q = d x 0.4 A/m x B / Height
= 765 pC/N x 0.4 A/m x 10.855 e-4 T / 0.006 m
= 55.36 pC
![Page 33: A Charge-Based Magneto-Electric Test Procedure](https://reader035.fdocuments.us/reader035/viewer/2022062804/56814932550346895db676bb/html5/thumbnails/33.jpg)
Radiant Technologies, Inc.Magneto-Electric Test Procedure
Experiment
![Page 34: A Charge-Based Magneto-Electric Test Procedure](https://reader035.fdocuments.us/reader035/viewer/2022062804/56814932550346895db676bb/html5/thumbnails/34.jpg)
Radiant Technologies, Inc.Magneto-Electric Test Procedure
Measured Data - Centered || B
Measured: 1 pCInductive: 1 pCCorrected: 2 pC
Signal Smaller than Noise
![Page 35: A Charge-Based Magneto-Electric Test Procedure](https://reader035.fdocuments.us/reader035/viewer/2022062804/56814932550346895db676bb/html5/thumbnails/35.jpg)
Radiant Technologies, Inc.Magneto-Electric Test Procedure
Measured Data - Centered B
Measured: 100 pCInductive: -1 pCCorrected: 101 pC
![Page 36: A Charge-Based Magneto-Electric Test Procedure](https://reader035.fdocuments.us/reader035/viewer/2022062804/56814932550346895db676bb/html5/thumbnails/36.jpg)
Radiant Technologies, Inc.Magneto-Electric Test Procedure
Measured Data - x = 1.5 R || B
Measured: -0.75 pCInductive: -1.5 pCCorrected: 0.5 pC
Signal Smaller than Noise
![Page 37: A Charge-Based Magneto-Electric Test Procedure](https://reader035.fdocuments.us/reader035/viewer/2022062804/56814932550346895db676bb/html5/thumbnails/37.jpg)
Radiant Technologies, Inc.Magneto-Electric Test Procedure
Measured Data - x = 1.5 R B
Measured: 30 pCInductive: -2 pCCorrected: 32 pC
![Page 38: A Charge-Based Magneto-Electric Test Procedure](https://reader035.fdocuments.us/reader035/viewer/2022062804/56814932550346895db676bb/html5/thumbnails/38.jpg)
Radiant Technologies, Inc.Magneto-Electric Test Procedure
Summarize Results @ 45 GExperiment Predicted
Results (pC)Measured
Results(pC)Centered || 0.0 0.0
Centered 229 101
At x = 1.5 R || -4 Indeterminate
At x = 1.5 R 55 32
![Page 39: A Charge-Based Magneto-Electric Test Procedure](https://reader035.fdocuments.us/reader035/viewer/2022062804/56814932550346895db676bb/html5/thumbnails/39.jpg)
Radiant Technologies, Inc.Magneto-Electric Test Procedure
Possible Error Sources• Amps/DRIVE Volts conversion for the KEPCO 36-6M
current amplifier. -1.75 Volts/Amp used. Expected current = 45.0 G X 0.0373 Amps/Gauss = 1.68 Amps. Post-data measurement showed 1.799 Amps. Generated 48.15 G.
• Current/Gauss conversion for the Lakeshore MH-6 Helmholtz coil. Used the Lakeshore published conversion of 26.76 G/A => 0.0373 A/G. Did not measure the actual ratio.
![Page 40: A Charge-Based Magneto-Electric Test Procedure](https://reader035.fdocuments.us/reader035/viewer/2022062804/56814932550346895db676bb/html5/thumbnails/40.jpg)
Radiant Technologies, Inc.Magneto-Electric Test Procedure
Possible Error Sources
• Manual dc and d measurements. Primary
• Unstable measurement surface.
• Unfixed sample subject to bending an shear. Primary
![Page 41: A Charge-Based Magneto-Electric Test Procedure](https://reader035.fdocuments.us/reader035/viewer/2022062804/56814932550346895db676bb/html5/thumbnails/41.jpg)
Radiant Technologies, Inc.Magneto-Electric Test Procedure
Conclusion• Radiant successfully tested the magneto-electric response
of a piezoelectric force sensor coupled to a magnet using Radiant’s Magnetoelectric Response Task.
• The system was able to cleanly capture the measurements that generated 30 pC of response.
• The sample response differed from our predictions most likely due to several possible error sources in the test fixture and piezo constants.
• To properly utilize the MT Task and achieve accurate results, these error sources must be mitigated.