8500 Training Manual Iss 3
-
Upload
kittiwat-wongsuwan -
Category
Documents
-
view
231 -
download
0
Transcript of 8500 Training Manual Iss 3
-
8/18/2019 8500 Training Manual Iss 3
1/69
8500 Venti lato r Train ing Manual
Blease
8500 Anaesthesia Ventilator
T i i M l
-
8/18/2019 8500 Training Manual Iss 3
2/69
T i i M l
8500 Venti lato r Train ing Manual
Read this manual before servicing the 8500
V til t
-
8/18/2019 8500 Training Manual Iss 3
3/69
8500 Venti lato r Train ing Manual
Contents
1. Mode Descriptions for the 8500
2. Test Procedures
Adjustment and Calibration Document 129AC000
Checklist and Calibration Record Document 129TC000
Final Test Procedure Document 129TP000
3. Circuit Diagrams
Interconnection Diagrams8500 Electrical Interconnection
6500/8500 Pneumatic system DiagramDisplay
Blease Display Interface 10100190
Pressure InterfacePressure Interface 10110088
BAV ControllerBAV Controller Type 2 Alarms 10110077
BAV Controller Type 2 Analogue 10110077
-
8/18/2019 8500 Training Manual Iss 3
4/69
8500 Venti lato r Train ing Manual
The Blease 8500 ventilator must only be serviced by Qualified Service Personnel.
The contents of this manual are not binding. If any significant difference is found betweenthe product and this manual please contact Blease Medical Equipment Limited for further
information. To ensure correct functioning, the equipment must be serviced at regular in-tervals.
Blease Medical Equipment Limited recommends that the machine should be serviced atintervals not exceeding three months. Qualified Service Personnel and genuine spareparts should be used for all servicing and repairs. Blease Medical Equipment Limited willnot otherwise assume responsibility for the materials used, the work performed or any pos-sible consequences of the same.
In communication with Blease Medical Equipment Limited, quote the model and serialnumber of the equipment, with the approximate date of purchase. If the unit is being re-turned for repair, indicate the nature of the fault or the work you require to be carried out.
Contact:
Blease Medical Equipment Limited
Beech House
Chilt C t
-
8/18/2019 8500 Training Manual Iss 3
5/69
8500 Venti lato r Train ing Manual
MODE Descriptions for the 8500s
PEEP
Control Panel
The Control panel has a permanent display of measured PEEP and a button to actuatethe setting of the desired value.
The display box will have the heading “PEEP” and show the “SET” value.
The default is that PEEP will be at the minimum setting ie. the residual that is < 2 cmH2Ocaused by the bellows assembly. In this default condition or if the Set value is subse-quently set below 3 cmH2O the set value display will show as “OFF”.
The method of control is that the user will rotate the Trak Wheel until the desired display is
highlighted, press the Trak Wheel and then increase the set value by rotating the TrakWheel until the desired value is reached, this is then accepted by a further press of theTrak Wheel.
The range of set values will be 3 to 20 cmH2O. An alarm will be implemented to indicateto the user that the set value has not been maintained. The alarm will activate at ± 50 %of the set value.
The acti ation of a set al e ill be made b pressing the Trak Wheel If this is not
-
8/18/2019 8500 Training Manual Iss 3
6/69
8500 Venti lato r Train ing Manual
Future Improvements
In the 8500s ventilator there is a monitor that measures the fresh gas flow of the anaes-thetic machine. The 10 LPM gas flow from the Flow valves in 1 above is an arbitrary fig-ure that has been arrived at empirically. This flow can be reduced substantially at low freshgas flow rates. It only requires high drive gas flows with high fresh gas flows to maintainPEEP. This could obviously save gas which would be helpful when using cylinders for sup-ply.
Volume Control ventilation (CMV)
Volume Control ventilation is a mechanical mode that delivers a tidal volume set by theuser into the patient tubing. This delivered volume is to be independent of the compressionlosses in the absorber, bellows and associated tubing. It will also be independent of anysmall leaks that may be present. Fresh gas flowing into the breathing system will notcause a permanent change in the delivered volume. If a change in fresh gas flow rate oc-
curs during ventilation the ventilator will re-adjust the delivered tidal volume to be correctwithin the next 4 breaths.
8500s Ventilator Fresh Gas and Compliance Compensation
Fresh Gas
Fresh gas flow adds to the delivered Tidal Volume during the inspiratory period. To com-
t f thi d ti i th d li d l d t b d Thi d ti i
-
8/18/2019 8500 Training Manual Iss 3
7/69
8500 Venti lato r Train ing Manual
To calculate the effect of Breathing System compliance on the delivered TV it is necessaryto measure what the capacity or compliance of the system is (Cs). This can only be doneby some form of pre-use check procedure. In essence it is necessary to have the ability tofirst of all select “Compliance Compensation” from a menu. It is then necessary to lead theuser through the process of measuring the dead space within the particular breathing sys-tem for that period of use. It is obvious that should the system be re-configured the testwill need to be repeated.
A possible process is to allow the ventilator to prompt the user to :-
1. Reduce the FG flow to minimum (but Allow for it as above)
2. Occlude the catheter mount. This can be a 15 mm male taper on the gas machine.
3. The ventilator then delivers a breath to pressurize the system to 10 cmH2O.
4. The ventilator records the volume required to achieve this pressure and verifies thata leak is not present.
The dead space is now calculated as follows :-
Volume in ml = Dead space compliance Cs
Pressure
This figure is stored until the ventilator is switched off or a re-test is asked for by the user.
-
8/18/2019 8500 Training Manual Iss 3
8/69
8500 Venti lato r Train ing Manual
So 735 ml is the actual ventilator output into the breathing circuit to give 500 mL at the
catheter mount.
This figure would have to be recalculated in the light of any fresh gas flow change asabove but would just be a variation on the 735 ml figure.
Pressure Control ventilation (PCV)
Pressure control ventilation is a time cycled mode where the ventilator strives to produce
the user set inspiration pressure for the inspiratory period. To accomplish this goal, the in-spiratory flow rate and pressure are set by the user. To do this, the control that is used forthe Set TV function will be reasigned as “I Flow” and calibrated in LPM. The pressureLimit control becomes the “Set Pressure” control. This allows the user to define the deliv-ered wave form. The default pressure will be 30 cmH2O in both Adult and Paediatricmodes.
Inspiratory Pause
A function that is “Inspiratory Pause” is to be implemented. This option will be available involume and PCV mode such that a pause of 25% of the inspiratory time is used as a Pla-teau before starting the expiratory phase. This means the expiratory time is reduced bythe plateau time. If at all possible the pause or plateau time should be made variable from5 to 60 %. The x % should be displayed on the second row up of the display.
Sigh
-
8/18/2019 8500 Training Manual Iss 3
9/69
8500 Venti lato r Train ing Manual
It will also be necessary to define an inspiratory time in which the pressure support is deliv-
ered. In this case the breathing rate and tidal volume are not relevant. The Inspiratoryflow rate will be a fixed value of 40 LPM in Adult and 30 LPM in Paediatric.
The detection of a “Patient Attempt” will be carried out by the control system to aidthe speed of support. A patient attempt will be indicated on screen by the negativeexcursion of the pressure wave form being green rather than blue for the positivephase.
SIMV
This mode is Synchronised Intermittent Mandatory Ventilation. This is an extension of thespontaneous mode in so far as the patient can take breaths on demand with pressure sup-port but some mandatory breaths are included. To achieve this it is necessary to set atidal volume, a breathing rate and all of the features described above for pressure support.To simplify things a little the inspiratory flow rate can be assumed from the TV / Inspiratory
Time set values.
The patient attempt or trigger signal will be used to synchronise the mandatory breaths tothe patients breathing pattern and initiate the pressure supported spontaneous breaths.Thus allowing the patient to establish the breathing rate.
In this way a patient could be breathing at say 12 BPM with the settings such that 4breaths of say 500 ml are mandatory and 8 breaths are taken spontaneously with pressure
t f t 30 H2O b th PEEP l l
-
8/18/2019 8500 Training Manual Iss 3
10/69
8500 Venti lato r Train ing Manual
This can be explained as follows :-
1. The frequency control sets up an internal clock tick in the ventilator. The time be-tween clock ticks is the Set breathing period and is equal to a time in seconds of60/frequency in BPM. ie. 4 BPM = 1 tick every 15 seconds
With no patient triggers detected, the ventilator will deliver a mechanical breath atthe start of each respiratory period on the clock tick.
2. Preceding each clock tick is a time window in which the ventilator will be looking
for a patient trigger. If a trigger occurs within the window the patient will be given asynchronised mechanical breath. If it occurs outside the window it will be a sponta-neous or Pressure Supported spontaneous breath depending on the ventilator set-ting.
In practical use of the SIMV mode, the ventilator should first be set at an adequateRATE, TIDAL VOLUME, I:E RATIO, PEEP and PRESSURE LIMIT for controlledventilation, say 600 ml, 12 BPM and 1.5 secs for an adult patient. When patient
trigger signals are being displayed regularly the mandatory rate can be reducedand the total breathing rate will be controlled by the patient. If this rate is inade-quate the the low MV or Rate alarm will be activated.
-
8/18/2019 8500 Training Manual Iss 3
11/69
8500 Venti lato r Train ing Manual
Vol Rate InspTIME
Press Limit Vol Alm
Lo / Hi
Press Alm
Lo / Hi
Adult 500 ml 12 1.5 50 4 / 10 4 / 55
Paed 150 ml 20 1.5 40 1 / 6 4 / 45
Mode Sigh Pause Press Supp PEEP MV Alm Rate Alm Press Alm
CMV üü üü üü üü üü
Spont üü üü üü üü üü üü
SIMV üü üü üü üü üü üü
PCV üü üü üü üü üü
CMV Defaults at start up
Matrix of Modes, Facilities and Alarms
-
8/18/2019 8500 Training Manual Iss 3
12/69
8500 Venti lato r Train ing Manual
InspFlow
InspTIME
SupportPressure
PressLimit
Rate Alm
Lo / Hi
Vol Alm
Lo / Hi
Press Alm
Lo / Hi
Adult 30L 1.5 10 50 5 / 25 4 / 10 4 / 55
Paed 20L 1.5 10 40 10 / 35 1 / 6 4 / 45
InspFlow
VentilationPressure
Rate Press Alm
Lo / Hi
Adult 30L 30 12 4 / 50
Paed 20L 20 20 4 / 40
Spontaneous with Pressure Support Defaults at start up
Pressure Control Ventilation Defaults at start up
Default for all Modes is “PEEP off” in PCV the maximum PEEP available will be limited to6 cmH2O.
If PEEP is set, the default “PEEP Alarm” will be ± 50 % in the range 0 to 30 cmH2O
Th R t BPM l i t b th t l 50 % i SIMV d L 10 hi 20 BPM
-
8/18/2019 8500 Training Manual Iss 3
13/69
Blease Test Procedure
Title: 8500 Adjustment & Calibration Document No:129AC000 Sheet 1 of 12
1.0 OBJECTIVES
To set up and calibrate the Blease 8500 Anaesthesia Ventilator. A final test (129TP000) must
also be followed to obtain a set of calibration results.
2.0 SCOPE
This test document is applied to new manufactured units before burn-in and after warm up,
and/or when any repairs have been completed on the ventilator.
This procedure is only valid for software versions 8.20 and 8.60 onwards.
3.0 OVERVIEW
The following procedure describes the adjustments and checks necessary to calibrate the
ventilator and set all the pre-set controls to achieve the specified performance. Inability to
make these adjustments, or failure to meet the specification, after these adjustments have been
-
8/18/2019 8500 Training Manual Iss 3
14/69
Blease Test Procedure
Title: 8500 Adjustment & Calibration Document No:129AC000 Sheet 2 of 12
4.1 Test Apparatus
4.2 Initial Checks
4.3 Reference Voltage and Electrical Supplies
4.4 BAV Controller PCB
4.5 Display Interface PCB
4.6 Main Pneumatic Regulator Calibration
4.7 Calibration
4.8 Flow Control Set-up
4.9 Zero Offset Connection4.10 Pressure Sensor Gain
4.11 Delivered TV Calibration
4.12 Hardware Pressure Overload Backup
4.13 Fresh Gas Flow Sensor Calibration
4.14 Saving Data and Printer Check
4.15 O2 Monitor Calibration
4.16 PEEP Calibration
-
8/18/2019 8500 Training Manual Iss 3
15/69
Title: 8500 Adjustment & Calibration Document No: 129AC000 Sheet 3 of 12
4.2 Initial Checks
BAV Controller Board 10110077
Check all links are fitted correctly and securely.
• JP2 link pins1&2
• JP5 link pins 2&3
• GND & AGND
• No links to be fitted to JP1 & JP3
PSU Board 10100076
• LK2 and LK3 on PSU board both have links fitted.
Display Interface Board 10100190
• Ensure J15 has pins 3 & 4 linked
BAV Interface Board 10110075
4.3 Reference Voltage and Electrical Supplies
With nominal mains input applied, switch ventilator ON in standby mode.
Ch k th f ll i lt t b f di f th
-
8/18/2019 8500 Training Manual Iss 3
16/69
Title: 8500 Adjustment & Calibration Document No: 129AC000 Sheet 4 of 12
4.5 Display Interface PCB
First time program
Ensure that JP2 has a link fitted & JP1 has a link fitted between pins 1 & 2, prior to switch on.
Connect a serial cable (12600016) between J10 & a PC & run the Hitachi flash programming
utility to load the H8/3048 microcontroller.
Select connect from the command menu, select BOOT mode and then OK to load the boot
kernel.
Once the kernel has been downloaded select program for the command menu, the
LOAD386.A20 absolute opcode file will be loaded.
Switch off 8500 remove the link on JP2 & move the link on JP1 to between pins 2 & 3.
Switch on the 8500 and run the flash for front and back program, select the A20 absolute opcode
file, followed by open to download the file.
After the first time programming of the display interface board, ensure JP1 has a link between
pins 2 & 3 and JP2 has no link fitted.
Re-program
Connect the ventilator serial cable (12600015) between the calibration port connector on the
back panel of the 8500 & a PC.
R h fl h f f d b k ili d i d h PC l h A20 b l
-
8/18/2019 8500 Training Manual Iss 3
17/69
Title: 8500 Adjustment & Calibration Document No: 129AC000 Sheet 5 of 12
Press CTRL+E on the PC followed by the up cursor key. Page through the diagnostics using the
TAB key & scroll through using the UP/DOWN cursor keys.
Select Blow off valve ref. Pressure. Set a flow of 1 LPM, the flow rate can be adjusted using
the Tidal Volume control.
Set main regulator output pressure = 34.0 ± 0.5 PSI while ventilator is in standby mode.
Turn ventilator off and remove gas supply.
Remove pressure gauge and replace plug.
-
8/18/2019 8500 Training Manual Iss 3
18/69
Title: 8500 Adjustment & Calibration Document No: 129AC000 Sheet 6 of 12
Adjust RV1 on the Interface board for 0V +/-10mV at PFZ test pin.
Press Space Bar to save.
Press Ctrl+U to save zeros.
Leak Test and Absorber Switching
Absorber switching
The function of this will have been checked during compliance compensation and the
presence of an absorber will have been displayed.Fresh gas leak test
Connect spirometry tubing 10110089 from TJ92 to the fresh gas connections on the rear of the
ventilator. Each side of the circuit is checked by pressurising it to 80 mmHg and checking that
h i i i d
-
8/18/2019 8500 Training Manual Iss 3
19/69
Title: 8500 Adjustment & Calibration Document No: 129AC000 Sheet 7 of 12
Fine calibration
Ensure that unit is in cal standby.
Note: to run the Blease Medical Test I/O Diagnostics connect the ventilator serial cable between
the calibration port connector on the back panel of the 8500 & a PC. Run HyperTerminal on the
PC using the stored 8500 settings file.
Press CTRL+E on the PC followed by the up cursor key. Page through the diagnostics using
the TAB key & scroll through using the UP/DOWN cursor keys.
On page 1 set gas to that specified by the model number, use right cursor, set gas convert to
Norm.
4.8 Flow Control Set-up
Connect Gas supply, Flow sensor, Fresh gas sensor and Mains power supply to the rear of the
ventilator.
With no fresh gas supply and the flow sensor disconnected from the test lung, select the
relevant fields; on the flowmeter, watch peak flow on constant not peak.
Select Flow valve just off ref. value.
Press space bar to start auto-find. A series of dots will appear in the value field. Once the lowest
achievable flow has been discerned, the relevant value will replace the dots.
Select Min flow available (use right cursor) enter peak value from the monitor.
-
8/18/2019 8500 Training Manual Iss 3
20/69
Title: 8500 Adjustment & Calibration Document No: 129AC000 Sheet 8 of 12
4.10 Pressure Sensor Gain
Wind both pots fully anti-clockwise on PEEP control board.
Select Blow off valve ref. Pressure. Set a flow of 1 LPM, the flow rate can be adjusted using
the Tidal Volume control.
Tee in to the drive gas outlet the pressure monitor and then block the flow sensor outlet. With
the flow sensor outlet blocked the internal pressure will build up until the Blow off valve
operates. The pressure will then stabilise at this point, shown on the pressure monitor, this
should be 75 ± 3 cmH2O.
If the reading is other then 75 +/- 3 cmH2O then adjust the valve.
Select Patient airway pressure gain.
When measured patient airway pressure reaches set value press the Space bar to set gain value.
Select Drive gas pressure gain.When measured patient airway pressure reaches set value press the Space bar to set gain value.
Press Ctrl+U to save values.
4.11 Delivered TV Calibration
With the flow sensor connected to the test lung (C20 compliance & R20 resistance) via the high
fl h i h k h d f l l TV 500 l 12 BPM d
-
8/18/2019 8500 Training Manual Iss 3
21/69
Title: 8500 Adjustment & Calibration Document No: 129AC000 Sheet 9 of 12
adjust flow for a peak pressure of 80 cmH2O. Wind R99 on the BAV board fully clockwise.
Adjust R98 on the BAV board until pressure limiting is observed. This can be checked byobserving that the can of Q1 goes high to low each time the limit is reached.
Once triggering has been achieved, decrease delivered volume to reduce the peak pressure to
find the exact trigger point. The trigger point should be such that it will not have any effect on
the ventilator functions. It should be set to greater than 80cmH2O.
4.13 Fresh Gas Flow Sensor CalibrationEnsure that the unit is standby.
Run the Blease Medical Test I/O Diagnostics, connect the ventilator serial cable between the
calibration port connector on the back panel of the 8500 & a PC. Run HyperTerminal on the PC
using the stored 8500 settings file. Press Ctrl+E to connection to 8500 then press up arrow to
enter calibration page.
Ensuring that there is no fresh gas flowing, highlight the Fresh gas flow zero field and press
spacebar to set value.
Set 10 LPM fresh gas flow, highlight Fresh gas flow gain (10 L/M) field and press spacebar to
set value. Press Ctrl+U to save values.
Press Ctrl+E to exit calibration page.
Ensure that the unit is standby.
Press the memory button, then select configuration.
-
8/18/2019 8500 Training Manual Iss 3
22/69
Title: 8500 Adjustment & Calibration Document No: 129AC000 Sheet 10 of 12
4.15 02 Monitor Calibration
Hardware Calibration
Short TP2 And TP3
Rotate VR2 fully anti clockwise when viewed from the back looking at the adjusting screw.
Adjust VR1 until just greater than 2.52V on TP1 and display turns to OFF.
Remove link and connect O2 sensor simulator to rear panel socket.
Set O2 sensor simulator to 100 then set VR2 fully clockwise and note that display reads over100%. Also TP1 must be lower then 1.3V
Adjust VR2 until display reads approximately 1.28V on TP1.
Set O2 sensor simulator to 0. Check that no more than 2.49V on TP1. Check that display reads
off when the sensor has been unplugged.
Software Calibration
Put 8500 into standby.
Press the menu button, and then select oxygen calibration.
Set the O2 sensor simulator to 100%.
Select the oxygen figure; use the trakwheel to change the number to 100. Push the Trak wheel to
accept. Return to main screen.
-
8/18/2019 8500 Training Manual Iss 3
23/69
Title: 8500 Adjustment & Calibration Document No: 129AC000 Sheet 11 of 12
4. 17 Front Panel Calibration Adjustment
Set-up the vetilator in a full circuit including an absorber (See below). Run the ventilaor and
compare the values displayed on the front panel of the ventilator with the values displayed by the
calibration analyser.
20RResistor
20 LCompliance12600010
Calibration Analyser
Fresh gas is set for minimum requiredto maintain full bellows
-
8/18/2019 8500 Training Manual Iss 3
24/69
Title: 8500 Adjustment & Calibration Document No: 129AC000 Sheet 12 of 12
4.18 Parameter checking
Check that the ventilator monitored Vte &Vti are within tolerance at the following settings
Set Tv 0.2, BPM 25, I:E 1:2 Measured Tv = Set Tv +/- 0.03
Set Tv 0.5, BPM 12, I:E 1:2 Measured Tv = Set Tv +/- 0.05
Set Tv 0.8, BPM 10, I:E 1:2 Measured Tv = Set Tv +/- 0.05
Set Tv 1.0, BPM 6, I:E 1:1.0 Measured Tv = Set Tv +/- 0.05
Also confirm that the ventilator monitored peak pressure = the externally monitored peak
pressure +/- 2cmH2O.
4.19 Assembly Completion and burn in
At this point the case wrap should be put on the unit and the unit should be put on burn
in for a further 24 hours before following final test procedure 129TP000.
-
8/18/2019 8500 Training Manual Iss 3
25/69
Checklist and Calibration Record for 8500 Anaesthesia Ventilator
Front PanelPart Numbers: Issue: 4 Software
version:BAV
Serial No: Document No: 129TC000 Supply Gas: Fresh Gas:
Works Order No: Date: Signature:
Ref Description Value Units Notes
5.2 Initial Inspect, mains in locked-plug type, finger proof
exhaust, gas inlet, orientation
Gas Inlet Fixing =
Inlet (USA O2 green, USA Air yellow,
UK 02 White, UK Air White/Black) Inlet ID =
Bellows =
P/F
USA/UK
USA/UK
Ad/Paed
P/F
⇒
ColourAd/Paed
USA/UK
5.3 Leakage/Continuity mains supply volts
Mains insulation (500v dc)Earth Continuity
Earth equipot Continuity (≤ 0.2R)
Earth Leakage Normal (≤ 500 µA)
Reverse (≤ 500 µA)
Supply (sfc) Normal (≤ 1000 µA)
Reverse (≤ 1000 µA)
-------------
-------------
-------------
-------------
-------------
-------------
-------------
-------------
Vac
Mohm
Ohms
Ohms
µA
µA
µA
µA
-
8/18/2019 8500 Training Manual Iss 3
26/69
Vent Measured
Ref Description BPM TV(L) RatioDel Freq
(BPM)Del Ratio Del TV/L VTi VTe
5.10 Vent Control
Functions
20C+20R
Adult Mode
Min Fresh Gas
6
6
10
20
20
20
40
60
90
1.0
1.0
1.0
0.6
0.4
0.2
0.1
0.2
0.1
1:1.0
1:Max
1:2.0
1:3.0
1:Min
1:2.0
1:2.0
1:1.0
1:2.0
-------------
-------------
-------------
-------------
-------------
1:
1:
1:
-------------
1:
1:
-------------
1:
1:
-------------
-------------
-------------
-----------
-----------
-----------
-----------
-----------
-----------
Ref Operation\Description Value 2 Value 1 Units
5.11 Expired Volume + Alarms, 1000 ml / 10 BPM, 1:2.0 Ventilator
Monitored MV (Value 2) & TV (Value 1)
MV/TV = 4L/200ml, Frq = 20 BPM
Decrease gas, MV low alarm operation 6.0L for 30 seconds
-------------
-------------
P/F
P/F
VTi
VTe
-------------
-------------
L
-------------
P/F
P/F
5.12 Pat pressure + Alarms, 1000 ml / 12 BPM, 1:1.0
Delivered peak pressure (2)/vent monitored peak (1)
vent monitored mean
Pressure high (2) and low (1) limit alarm settings
-------------
-------------
------------- -------------
cm H2O
cm H2O
cm H2O
5.13 Adult 50 cm H2O pressure limit, 1200 ml, 10 BPM, 1:2.0 ------------- cm H2O
-
8/18/2019 8500 Training Manual Iss 3
27/69
Ref Operation\Description Value 2 Value 1 Units
5.19 PCV Test – Default Settings
Monitored Peak (Value 2) delivered Peak (Value 1)
Monitored Peak (Value 2) delivered Peak (Value 1)
Return to default
PCV Test – Inspired Flow
Monitored Peak (Value 2) delivered Peak (Value 1)
Monitored Peak (Value 2) delivered Peak (Value 1)
Return to default
-------------
-------------
-------------
-------------
-------------
-------------
-------------
cm H20
cm H20P/F
-------------
cm H2O
cm H2O
P/F
5.20 PEEP 500ml, 12 BPM, 1:2.0
Delivered minimum pressure (value 2), monitored PEEP (value 1)
Set 3 cmH20
Set 20 cmH20
Set 10 cmH20
PEEP set to OFF
cm H2O
cm H2O
cm H2O
P/F
5.21 Hardware Pressure Overload Cm H2O
5.22 Fresh Gas
0
1
3
5
LPM
LPM
LPM
LPM
-
8/18/2019 8500 Training Manual Iss 3
28/69
Blease Test Procedure
Title: Final Test 8500 Document No: 129TP000 Sheet 1 of 17
1.0 OBJECTIVES
To final test and provide a set of results for the 8500 Anaesthesia Ventilator.
2.0 SCOPE
The final test in conjunction with the checklist and calibration record (129TC000) should be
completed whenever an 8500 Anaesthesia Ventilator is being readied for shipment from the
factory.
This may be for Demonstration or Production purposes. It should also be used on ventilators
which are going into stock, and are to be fully packed and for shipping within a few days.
WARNING - Under no circumstances should ANY part of the ventilator be opened orremoved after this test procedure has been completed and signed off. This procedure is
only valid for software versions 8.20 & 8.60 onwards.
3.0 OVERVIEW
The remaining sections of this document are:
-
8/18/2019 8500 Training Manual Iss 3
29/69
Title: Final Test 8500 Document No: 129TP000 Sheet 2 of 17
DURING TESTING, ALWAYS BE AWARE OF CORRECT VENTILATOR OPERATION AND
LISTEN FOR ANY UNEXPECTED OR UNUSUAL NOISES FROM THE MACHINE AND
ASSOCIATED TUBING, AS THIS MAY INDICATE A POTENTIAL PROBLEM.
ALWAYS COMPLETE COMPLIANCE COMPENSATION AFTER SWITCHING THE
VENTILATOR OFF AND BACK ON AGAIN. ALWAYS WAIT FOR COMPLIANCE
COMPENSATION TO COME BACK ON BEFORE TAKING ANY READINGS.
4.0 TEST LIMITS SPECIFICATION
Sched. Function set Parameter Tolerance Units
Ref. or tested Measured
5.3 Mains connected Mains Supply 90-265 Vac
(500 Vdc) Mains insul Res 50 Mohms
(25 A ac) Equip cont Res ≤ 0.2 Ohms
Norm leakage ≤ 500 µA
Revs leakage ≤ 500 µA
O/C norm leakage ≤ 1000 µA
O/C revs leakage ≤ 1000 µA
-
8/18/2019 8500 Training Manual Iss 3
30/69
Title: Final Test 8500 Document No: 129TP000 Sheet 3 of 17
≈ 3 ml/cm H 2O, and vent paediatric compliance ≈ 2 ml/cm H 2O.
Total TV = Set TV + Fresh gas in L/m
BPM Rate x (I + E)
At higher frequencies time delays due to compliances greatly affect delivered volumes and
ratios.
5.10 Frequency tolerance at all BPM ± 1%
5.10 I:E Ratio tolerance at all BPM ± 10% or ± 0.1 whichever is larger.
Sched. Function set Parameter Tolerance
Ref. or tested Measured
5.11 Exp Vol 1 L Exp TV Delivered TV ± 10%
10 L Exp MV 10 x Exp TV ± 0.2
5.12 Patient Pressure vent monitor peak Delivered peak ± 1.5
Calibration vent monitor mean 20 ± 3 cmH2O
Pressure high alarm High alarm setting Delivered peak ± 2
Pressure low alarm Low alarm setting Delivered peak ± 2
5.13 Adult Pressure Limit 50 cm H 2O set 50 ± 2 cm H2O
-
8/18/2019 8500 Training Manual Iss 3
31/69
Title: Final Test 8500 Document No: 129TP000 Sheet 4 of 17
5.0 TEST PROCEDURE
5.1 Test Equipment and Setup
ENSURE ALL TEST EQUIPMENT HAS A CURRENT CALIBRATION CERTIFICATE.
The following items will be needed to perform all the checks in this procedure:
• Timeter Calibration Analyser RT-200 with following options fitted: RT-201 Control
Module, RT-202 Press Module, RT-203 Flow Module, or another Calibration Analyser.
The Timeter is supplied with its own pneumatic adapters for connection to hoses. See
section 6.0 for details of Calibration Analyser usage.
• Hyperterminal should be set up as follows: 9600 baud, 1 start and stop bit, 8 data bits. Nohandshaking is used. File should be saved as 8500.HT for further ref.
• Electrical Safety Tester for earth leakage and continuity checks.
• Stop Watch or other timepiece to measure up to 180 seconds.
• 5 ± 0.25 L compliance, 20 ± 1 L compliance, 50 ± 2.5 L compliance, 20 ± 2 cm H2O
resistance.
-
8/18/2019 8500 Training Manual Iss 3
32/69
Title: Final Test 8500 Document No: 129TP000 Sheet 5 of 17
20R
Resistor
20 LCompliance12600010
Calibration Analyser
Fresh gas is set for minimum requiredto maintain full bellows
-
8/18/2019 8500 Training Manual Iss 3
33/69
Title: Final Test 8500 Document No: 129TP000 Sheet 6 of 17
Test
1. Electrical safety tester test: Check the electrical safety tester meter rises to a value within the
indicated 'test' region.
2. Supply: Note the mains supply voltage reading.
3. Insulation: Press the test switch on the electrical safety tester and note the insulation resistance
reading.
4. Earth Continuity: put the test probe on one of the two screws adjacent to the mains inlet at the
rear of the ventilator and repeat the test. Note the meter reading.
5. Earth/equipotential continuity: Place the electrical safety tester test probe on the equipotential
connector at the rear of the ventilator. Press the test switch and again note the meter reading.
6. Normal Earth Leakage: Set the electrical safety tester Normal/Operate/Reverse switch to
Normal and note the earth leakage current. Then set to Reverse and again note the earthleakage current.
7. Supply o/c Earth Leakage: Obtain the two earth leakage readings as in operation 6. (Normal
and Reverse.
5.4 Fresh gas leak test
-
8/18/2019 8500 Training Manual Iss 3
34/69
Title: Final Test 8500 Document No: 129TP000 Sheet 7 of 17
5.6 Battery Back Up
Turn the ventilator OFF for at least 5 seconds and isolate from the mains supply.
Turn the ventilator ON to adult mode and listen for the vent mains fail alarm to activate.
Switch off the ventilator.
Reconnect the mains supply to the ventilator and observe that it starts up as normal.
Isolate from the mains supply again and verify the battery on charge indicator goes off
after one minute and the mains fail alarm activates for a few seconds.
Verify that after one minute of mains failure the mains fail alarm activates again for a few
seconds and the screen backlight goes off. Check the SET UP screen comes on and remains
on until the next alarm sounds.
Reconnect the mains supply and check that the battery on charge indicator turns on and the
mains fail indicator turns off.
5.7 Compliance test
Conect the ventilator using the relevant tubes, occlude the patient airway sensor. Switch the
ventilator on and say yes compliance compensation. Follow the on screen instructions. If the
compliance and leak tests are passed it is unlikely that there are any internal leaks.
-
8/18/2019 8500 Training Manual Iss 3
35/69
Title: Final Test 8500 Document No: 129TP000 Sheet 8 of 17
5.10 Ventilatory Control Functions
With the correct supply gas (oxygen/air), note the following delivered parameters at the stated
control settings whilst ensuring no pressure limiting occurs: (Patient = 20C+20R, mode = adult
run & min fresh gas.) After changing any settings wait for compliance compensation to come
back on before taking any readings.
BPM TV(L) Ratio Delivered and Monitored Parameters to Note6 1.0 1:1.0 Freq Ratio TV
6 1.0 1:Max Ratio
10 1.0 1:2.0 Ratio TV
20 0.6 1:3.0 Freq TV
20 0.4 1:Min Ratio
20 0.2 1:2.0 Ratio TV
40 0.2 1:2.0 Freq60 0.2 1:1.0 Ratio TV
90 0.1 1:2.0 Freq Ratio TV
Max ratio for the 8500 is 1:5. Min ratio for the 8500 is inverse 2:1
-
8/18/2019 8500 Training Manual Iss 3
36/69
Title: Final Test 8500 Document No: 129TP000 Sheet 9 of 17
5.12 Patient Pressure and Alarms
Set ventilatory controls to: MV/TV = 10L/1000ml
Freq = 10 BPM
Ratio = 1:1.0.
After several breaths, note the delivered peak pressure, the ventilator monitored peak and mean
pressures.
Reduce the pressure high alarm limit until a pressure high alarm just occurs and note the
control setting. Restore control to default.
Increase the pressure low alarm limit until a pressure low alarm just occurs and note the
control setting. Restore control to default.
5.13 Adult Pressure Limit
Set ventilatory controls to: MV/TV = 12L/1200ml
Freq = 12 BPM
Ratio = 1:2.0
Set pressure limit to 50 cm H20 and note the delivered peak pressure after several breaths.
Verify the pressure high alarm activates for a short period as the ventilator pressure cycles
into expiration.
-
8/18/2019 8500 Training Manual Iss 3
37/69
Title: Final Test 8500 Document No: 129TP000 Sheet 10 of 17
Switch the ventilator to paediatric mode and carry out compliance compensation as directed
on screen:
Set ventilatory controls to: TV = 0.5L/50ml
Freq = 10 BPM
Ratio = 1:2.5
Note bellows movement.
Set ventilatory controls to: TV = 2L/200ml
Freq = 10 BPM
Ratio = 1:2.0
Note bellows movement.
5.17 Paediatric Pressure Limit
Set ventilatory controls to: TV = 6L/200ml
Freq = 30 BPM
Ratio = 1:2.0
Set pressure limit to 20 cm H2O and note the delivered peak pressure after several breaths.
-
8/18/2019 8500 Training Manual Iss 3
38/69
-
8/18/2019 8500 Training Manual Iss 3
39/69
Title: Final Test 8500 Document No: 129TP000 Sheet 12 of 17
5.20 PEEP Test
Return circuit to as shown in 5.1. Set ventilatory controls to TV = 500ml , Freq = 12 BPM
Ratio = 1:2.0.
Set peep to 3cm H2O, check on the Calibration Analyser this is achieved within 5 breaths
Set peep to 20cm H2O, check on the Calibration Analyser this is achieved within 5 breathsSet peep to 10cm H2O, check on the Calibration Analyser this is achieved within 5 breaths
Set peep to OFF, check on the Calibration Analyser this is achieved within 2 breaths
Allowable tolerances Set = measured ± 0.5cm H 20
5.21 Hardware Pressure Overload
With the ventilator in Adult CMV mode with settings at default, connect to a C5 lung.
Run the Test I/O diagnostics as described in Section 5.4. Position the cursor on line 1 of
page 1, to disable software limiting.
Increase delivered volume until the hardware pressure limit is triggered. Note the peak
delivered pressure.
-
8/18/2019 8500 Training Manual Iss 3
40/69
Title: Final Test 8500 Document No: 129TP000 Sheet 13 of 17
5.24 Dealer specific information
The following options can be set:
Absorber Present (default) / Not present
Language English (default), German, French, Spanish, Turkish, Italian
O.E.M Blease (default), A.M.S, Acoma, Finesa, User select
The Absorber switch setting can be found under setup enter the password “TECHY”
The Language and O.E.M settings can also be found under setup enter the password “LINGO”
If anything other than the default options are required then these should be set
All options should be recorded on the calibration record even if they are default.
5.25 Test Completion
Restore ventilator controls to default. Ensure mode on/off switch is in the OFF.
Check that all the noted readings are within the tolerances as specified in section 4.
Sign off the 'Checklist and Calibration Record' and the customer printout. Ensure both have the
ventilator serial number, date and signature of final tester, etc.
If the original sticker over the hole in the calibration port on the rear of the ventilator was
removed, replace with a new ‘calibration void’ sticker.
-
8/18/2019 8500 Training Manual Iss 3
41/69
Title: Final Test 8500 Document No: 129TP000 Sheet 14 of 17
5.27 Ventilator Catalogue Number/Model Summary
Cat No Description Notes
12900001 8500 90 - 260 Vac Adult USA O2 Driven
12900002 8500 90 - 260 Vac Paed USA O2 Driven
12900003 8500 90 - 260 Vac Adult UK O2 Driven12900004 8500 90 - 260 Vac Paed UK O2 Driven
12900005 8500 90 - 260 Vac Adult USA Air Driven
12900006 8500 90 - 260 Vac Paed USA Air Driven
12900007 8500 90 - 260 Vac Adult UK Air Driven
12900008 8500 90 - 260 Vac Paed UK Air Driven
-
8/18/2019 8500 Training Manual Iss 3
42/69
Title: Final Test 8500 Document No: 129TP000 Sheet 15 of 17
6.0 USING THE CALIBRATION ANAYSER
This section is intended to be guide in using the Calibration Analyser with this test procedure.
It should be read in conjunction with the operating manual supplied with the Calibration
Analyser.
6.1 Volume Measurement Correction For volume measurements, the patient circuit is connected as shown in section 6.4, that is,
with compliance connected to the High Flow Outlet of the Calibration Analyser. In this
configuration, all Calibration Analyser volume measurements MUST HAVE A
CORRECTION FIGURE ADDED TO THE INDICATED VALUE.
A simple correction formulae is:
At 500 ml ≈ 25cm H 2O peak pressure ∴ mean pressure ≈ 12.5
Calibration Analyser reads 0.48 Lpm
Actual = (0.48 x (1+0.0125)) = 0.486 Lpm
Where MEAN PRES is the breath mean pressure in cm H2O while there is measurable flow.
This formulae must be applied to ALL the Calibration Analyser volume readings BEFORE
-
8/18/2019 8500 Training Manual Iss 3
43/69
Title: Final Test 8500 Document No: 129TP000 Sheet 16 of 17
It is not unusual when measuring volume that the Timeter throws up an error message like
'HELP R' or 'HELP F'. All this means is that the Timeter had difficulty in identifying the start
or end of a breath, possibly due to some pneumatic valve noise or reverse flow through the
Timeter.
If it occurs a lot, it might be because there is a leak in the patient circuit, or the Timeter
needs zeroing, or the patient circuit is moving/vibrating with the breath. These are a fewthings to check before assuming there is anything wrong with the ventilator.
6.2 Timeter Peak Pressure Measurements
If the ventilator has anything more than a few cmH2O of CPAP pressure, the Peak Pressure
function on the Timeter will NOT operate. It will sit there forever showing the last peak
pressure measured.
6.3 Timeter Zeroing and Power Up
When the Timeter is first switched ON, its sensors and electronics will need to be zeroed
before the Function Codes will operate. This is simply done by pressing the Zero button
provided after a Function has been selected and there are NO PNEUMATIC CONNECTIONS
TO THE TIMETER.
-
8/18/2019 8500 Training Manual Iss 3
44/69
-
8/18/2019 8500 Training Manual Iss 3
45/69
Blease Test Procedure
Title: B.A.V. POWER SUPPLY PCB DocumentNo:129TP076 Sheet 1 of 5
1.0 OBJECTIVE
To set up and test power supply board 10100076 used with the Blease 8500 Anaesthetic
Ventilator and record a set of test results using 129TC000 Checklist and Calibration Record
2.0 SCOPE
The test procedure should be used with newly manufactured boards. When successfully passedthey can be placed into Blease stock after inspection.
3.0 OVERVIEW
All testing is done with the aid of a test jig (TJ 70), this provides power resistors and switching
that allows the correct test loading to be applied. Five 3½ digit LCD panel meters are also
incorporated into the jig which are connected to the input and output power rails of the PSU
board via its connectors.
The jig also contains a mains transformer and rechargable battery with circuits identical to that
used in an 8500 Anaesthetic Ventilator. Hence any results noted will accurately reflect the
board operation when it is eventually fitted into a ventilator.
Most of the testing uses an external DC variable supply, this avoids the need for a 'Variac' and
also allows current limiting to be used.
WARNING - WHEN POWERING UP A BOARD FOR THE FIRST TIME ALWAYS USE
-
8/18/2019 8500 Training Manual Iss 3
46/69
Blease Test Procedure
Title: B.A.V. POWER SUPPLY PCB DocumentNo:129TP076 Sheet 2 of 5
4.0 TEST DETAILS
4.1 Equipment
1. TJ 70 - Blease test jig (The jigs are serialised TJ 70A, TJ 70B, etc)
2. Variable DC power supply, 17-30 Vdc @ 2.5 Amp. The supply must have an adjustablecurrent limit of 100 mA to 2.5 Amps.
3. Pair of short 4mm plug to plug leads for external DC supply to jig connection.
4. Miscellaneous short connection leads for attachment of test BAV supply board to jig (6 off).
These will vary according to the issue of board tested.
4.2 Connections and Set up
Connections for the test set up are shown below. (Not to scale)
1. Test jig TJ 70 (viewed from Top)
Jig Interface Board (Fixed to jig) - 10100177
PL0
PL7 PL4PL5PL8PL1PL2PL3PL6
-
8/18/2019 8500 Training Manual Iss 3
47/69
Blease Test Procedure
Title: B.A.V. POWER SUPPLY PCB DocumentNo:129TP076 Sheet 3 of 5
4.4 Test Limits
These are provided on the Checklist and Calibration Record.
The battery volts for ops. 2, 3 & 4 may fall below the stated limits if the temperature of the
board under test (IC3) rises above 21°C.
For op. 12 the test limits (z+?) are referenced to the noted reading (z) for the unregulatedsupply voltage.
5.0 TEST SCHEDULE
See 4.0 TEST DETAILS for information on the test equipment and its use. Also covered are
connection details, set ups, methods and test limits.
All operations in this section must be done in the order as written.
5.1 Preliminary
The board must NOT be attached to the test jig until operations 1 to 3 below are successfully
completed.
1. Set all Test Jig switches to uppermost toggled position.
2. Set external DC supply to 0 volts with a 100 mA current limit.
3. Perform a visual quality check of the new board assembly. In particular:
-
8/18/2019 8500 Training Manual Iss 3
48/69
Blease Test Procedure
Title: B.A.V. POWER SUPPLY PCB DocumentNo:129TP076 Sheet 4 of 5
5.3 On Load Volts
Set the external supply current limit to maximum (≥ 2.5 Amp) and then toggle the following
five jig switches to ON:
TOP CHARGE ⋅ +5V LOAD ⋅ +6V LOAD ⋅ FLOW VALVE ⋅ DUMPVALVE
Note the four supply voltages stated. (UNREG +25V = 17.0 Vdc)
5.4 Regulation Volts
Increase the external DC supply voltage until the jig UNREG +25V meter reads 29.0 ± 1 Vdc
and note the four supply voltages again.
5.5 Off/On Check
Toggle the jig PCB MODE switch to OFF for a few seconds and then return to the ON
position.
Apart from the BATT VOLTS supply, verify that the other three supplies turn OFF and are
then restored to there previous value when the PCB MODE switch is turned back ON.
5.6 Battery Operational Check
Toggle the following three jig switches:
BATT to CONNECTED ⋅ BATT RELAY to ON ⋅ +6V SUPPLY to HOLD
-
8/18/2019 8500 Training Manual Iss 3
49/69
Blease Test Procedure
Title: B.A.V. POWER SUPPLY PCB DocumentNo:129TP076 Sheet 5 of 5
5.11 Mains Operation
Toggle the following three jig switches:
ALARM SOUND to OFF ⋅ +24V RELAY to NORM ⋅ PCB SUPPLY to AC
After toggling the jig PCB MODE switch to ON, verify all supplies are restored to their
nominal ON LOAD values. The BATT VOLTS may be a little low as charging is initiated dueto previous testing.
Verify on the jig that the PCB SUPPLY lamp is similarly illuminated as the adjacent JIG
SUPPLY lamp.
Note - the supply voltage to the board under test as shown on the UNREG +25V meter is
protected by thermistors. These may trip if the board is allowed to run at full load for
any length of time, particularly when running from AC.
To regain normal operation, turn OFF the jig external DC and mains supplies for a
couple of minutes.
5.12 Unregulated Supply
Results for the following are affected by the mains supply voltage which should be at its
nominal value of 240 Vac.
-
8/18/2019 8500 Training Manual Iss 3
50/69
Blease Test Procedure
Title 8500 B.A.V. Front Panel Document No: 129TP078 Sheet 1 of 1
The front panel for the 8500 Anaesthesia Ventilator
is functionally checked as part of the test procedure for the controller.
REMOVE ALLBURRS ANDSHARP CORNERS DO NOT SCALE IF YOU HAVE ANY DOUBTS PLEASE ASK
-
8/18/2019 8500 Training Manual Iss 3
51/69
1011SM02
REMOVE ALLBURRS ANDSHARP CORNERS
Blease DRG.No.
DO NOT SCALE IF YOU HAVE ANY DOUBTS PLEASE ASK
-
8/18/2019 8500 Training Manual Iss 3
52/69
REMOVE ALLBURRS AND SHARP CORNERS
1
2
3
4
A B
D R G . N o .IFYOU HAVEANY DOUBTSPLEASEASK
C D FE G
DONOT SCALE
AUTOZEROVALVE
AUTOZEROVALVE
36psiDETECTOR
PRESSURE
SUPPLY
LOW
POWERSUPPLY BOARDCONTROLLERBOARD
SWITCH
PRESSURE
10110077
PL5VALVES
FLOW
10100076
2.5-7BAR
36-101psi
INLET
SUPPLY
GAS
REGULATOR
MAIN
34.5psi
FLOWCONTROL VALVES
PT4
10110075PRESSUREINTERFACE BOARD
PRESSUREFLOW
INSPIRATORY DRIVE
PT2
PNEUMOTACHOGRAPH
VALVE
EXPIRATORY
SOLENOID
VALVE
DUMP
PRESSURE
PATIENT
PT5
75cmH2O
VALVE
RELIEF
EXHAUST
PORT
PORT
OUTLET
GAS
DRIVE
-
-
-
-
? ? ? ?
1:1
1-00
JLB
1011SM03
1
PL4 PL8
VALVE
PEEP
AIRWAYFLOW
PATIENT
GASFLOW
FRESH
PT1PT3
SENSORINLET
AIRWAYFLOW
SENSORINLET
FRESHGAS
VALVEPEEP
FILTER
DUMPVALVE
PL5
-
8/18/2019 8500 Training Manual Iss 3
53/69
1 2 3 4 5 6 7 8
A
B
C
D
87654321
D
C
B
A
Title
Number RevisionSize
A3
D at e: 4 -F eb -2 00 2 S he et o fFile: \\Cad 3\c\1 JONS ARCHIVE\8500archive\circuit diagrams\10100076-9\10100076.ddbDrawn By:
TR1, IC1, IC2, IC4, D1 D2 and D10 are all mounted on the heatsink block.
2PL6-2PLM
1PL6-1PLM
TR2MPSA13
L41R0
EARTH-TRM3
TR10ZTX650
D16IN4448
2PL8-2
1PL8-1
TP4
TP3TP2
TP1
D13BAT42
TP7
TP6
TP10
TP9
TP5+29V Min
R38
10K0
TR9ZTX650
R373K32
TP8
RL1
RY6-10-006
24 PL1-24
12345678 9
10111213141516
IC3
12 PL1-12
R41R00
10 PL1-10
TO202-3P TO220-2P TO220-5P TO220-5P TO220-3P
20 PL1-20
18 PL1-18
7 PL1-7
11 PL1-11
16 PL1-16
13 PL1-13
14 PL1-14
5 PL1-5
R32100R
C1910u
D141N4002
R314K75
R3010K0LK1
LK2LK3
R2910K0
R282K21
R27475R
ALM1
TR8ZTX650
E-LINE
22 PL1-22
23 PL1-23
21 PL1-21
R262K21
RN2D2K2
RN2C2K2
RN2B2K2
RN2A2K2
RN1D2K2
RN1C2K2
RN1B2K2
RN1A2K2
D51N4448
C17100n
C1610u
R24100R
D121N4002
D111N4002
R16100R
TR7ZTX650
TR5ZTX650
R2347K5
D10PBYR745
TR6ZTX650
C1510u
R221K00
D6BAT42
R211R00
R2047R5
C14100n
R1910R0
9 PL2-9
3 PL2-3
2 PL2-2
2 PL3-2
1 PL3-1
D41N4448
1 2 3
45IC2
LM2577TADJ
9 PL1-9
19 PL1-19
C7100n
C6100n
R10365K
R930K1
R839K2
R7110K
D3BAT49
R61K00
R51K00
R31R00
TR1TIP30A
C22200u
1 2
3 4 5
IC1LM2576TADJ
D1 C3470u
L1220uH
26 PL1-26
15 PL1-15
17 PL1-17
3 PL1-3
1 PL1-1
25 PL1-25
B R1 G BP C8 02
1
2REG-SINK
2
1
BAT1
C8100n
RT1MFR300
2 PL7-2
1 PL7-1
C5470u
C9470U
R332K74
C10220n
R112K00
R12
47K5
D2PBYR745
L5220uH
C1100n
R132K00
R14
7K87
R15
10K0
L2100uH
C412u
L3100uH
C1110u
3 2
1
IC4LM317AT
R174K75
R18240R
C1310u
TR3ZTX650
TR4ZTX650
1 PL4-1
2 PL4-2
1 PL5-1
2 PL5-2
D7IN4002
D8IN4002
D91N4448
13
2
IC5LM341P5
C1210u
1 PL2-1
3 PL3-3
4 PL3-4
R11K50
R21K50
D151N4002
C1810u
4 PL2-4
6 PL2-6
8 PL2-8
10 PL2-10
2 PL1-2
4 PL1-4
6 PL1-6
8 PL1-8
R25475R
TO220-3P
RL2RY6-10-006
R34100R
R351K00
C201n
C211n
R361K00
1
CHASSIS
RELAY-8ATO92
4
T1A
T1A
Battery
ICCFLW
ICCSOL
ALMSIG
ALM+5V
+24V
+25V
VBAT
ON-LEK
ON-FLW
ON-SOL
ON-24V
SBAT
ON-BAT
ON-PSU
/OVRCH
GND
GND
GND
+6V
+6V
SWBAT
OFFLOW
/AD-ON
/AD-ST
/PD-ST
/PD-ON
/PD-ON
/PD-ST
/AD-ST
/AD-ON
OFFLOW
GND
SW-POL
SW-WAF
GND
+6V
LED-
LED+
DRIVE
DRIVE
+24V
+24V
to
BATT-
BATT+
OUT
+24V
CHASSIS
Sec
Sec
3W
101CD076
+24V
controller
Transition
Vol
Alarm
/OVRCH
/OVRCH
2W5
C H A S S I S
SCHRACK RY6-10-006
RL1&2 Top View
IN
GND
OUT
GND TOP VIEW
LM341P5
ICCFLW
ICCSOL
ALMSIG
ALM+5V
+24V
+25V
VBAT
SBAT
SBAT
GND
+
ALMSIG
SBAT
BASE
C
B
E
ZTX650
CMP
TOP VIEW
LM2577T-ADJ
GND
GND
FB
OUT
IN
GND
ON-FLW
ON-SOL
ON-24V
+25V
VBAT
VBAT +6V
ICCFLWICCSOL
+
+24V
ON-24V
ON-SOL
ON-FLW
MODE SW
+6V
+25V
1 6 0 R
NC
TIP30A
TOP VIEW
A
K
PBYR745
TOP VIEWK
IN
OUT
OFF
FB
GND
GND
LM2576T-ADJ
TOP VIEW
LM317AT
TOP VIEWOUT
OUT
ADJ
IN
745
PBYR
9
B.A.V. POWER SUPPLY
1 1
A.J.W./A.A
o14 C/W
MAINS
MAINS
MAINS
LIVE
NEUTRAL
EARTH G R N / Y E L
+25V
GND
12V
I= 5.1A
3.0A
E
C
C
B
BATT CHARGER
DEVICE PINOUTS
+ +
+6V
+25V
GND
+
Flow Valve
Dump Valve
2 A/Hr
NC
1 6 0 R
ALM+5V
GND
+
FRONT PNL
17-35V Mains
9-14V Batt
MAINS INLET BLOCK
BASE
MPSA13
E
B
C
EQUI-
POTENTIALGRN/YEL
L
N
E
1
E
(+26V)
80200000
10100080
IC6
CA3130E
C2210µF
VR15K
R46
10K
R42
12K
PEEP Valve
R43
1K
R44
220R
R45
220R
VR21K
+6V
GND
C23
33P
1
2
3
4
6
78
R4147R5
R40
12K
R39
12K
SWITCH MODE POWER SUPPLY
LPS45-M
-4V
25V
0V
-
8/18/2019 8500 Training Manual Iss 3
54/69
1 2 3 4 5 6
A
B
C
D
654321
D
C
B
A
Title
Number RevisionSize
B
D at e: 4 - Fe b -2002 S hee t o f
File: \\Cad 3\c\1 JONS ARCHIVE\8500archive\circuit diagrams\10110077-4\10110077-4.DDBDrawn By:
ALARM[1..7]
RC2
C1
A4
B5
CLR 3
Q6
Q7
U34A
MC74HC4538AD
1
2
3
U51A
MC74HC132AD
4
5
6
U51B
MC74HC132AD
8
9
10
U51C
MC74HC132AD
11
12
13
U51D
MC74HC132AD
1
2
3
U54A
MC74HC132AD
4
5
6
U52B
MC74HC132AD
8
9
10
U52C
MC74HC132AD
11
12
13
U52D
MC74HC132AD
1
2
3
U53A
MC74HC132AD
4
5
6
U53B
MC74HC132AD
8
9
10
U53C
MC74HC132AD
11
12
13
U53D
MC74HC132AD
1
2
3
U52A
MC74HC132AD
4
5
6
U54B
MC74HC132AD
1011
U42E
MC74HC14AD
RESET
INODSP
D15
BAR42
D16
BAR42
ALARM1 Pulse Audio
ALARM2 Continuous Audio
D17
BAR42
1213
U42F
MC74HC14AD
PRS_OVRLD
Pressure High DisplayALARM7
ALARM6
D18
BAR42
1 2
U42A
MC74HC14AD
APNOEA
3 4
U42B
MC74HC14AD
+ C19
2u2 63V
GND
1TP9
TPOINT
ALARM5
/PWR_FAIL
5 6
U42C
MC74HC14AD
ALARM4
Pressure Low Display
Power Fail LED
Mute LED
1
2
3
U40A
MC74HC02D
ALARM3 Mute Audio
+ C20100u 25V
GND
D19
BAR42
VCC
VCC
3
2
1
8
4
U32A
LMC662CM
+5VALM
GND
GND
GND
+ C21
2u2 63V
1
TP10
TPOINT
89U42D
MC74HC14AD
8
9
10
U54C
MC74HC132AD
11
12
13
U54D
MC74HC132AD
4
5
6
U40B
MC74HC02D
8
9
10
U40C
MC74HC02D
11
12
13
U40D
MC74HC02D
D21BAR42
+5VALM
GND
D22
BAR42
5
6
7
U32B
LMC662CM
+5VALM
GND
15Hz
1TP11
TPOINT
RC14
C15
A12
B11
CLR 13
Q10
Q9
U34B
MC74HC4538AD
CLRBAK
VCC
ALMCLR
VCC
GND
1TP12
TPOINT
OVERLOAD_INT
PHIDSP
PL2[1..50]
COMALM
PLODSP
PL1[1..26]
PL244
PL243
PL2[1..50]
PL17
PL1[1..26]
ALARM[1..7]
R69
62K
R70
10K
R71
68K
R72
47K
R73
160K
R74
100R
R75
2M2
R76
100K C36
10n
R77
47K
R78
47K
R79560K
R80
470K
R81
100K
R82
100K
R83
10M0
R84
100K
R85
100K C37
6n8
R86
1K0
R87
560K
R88
470K
R89
0R0D25
BAS16
REF1
LT1004CZ12
D20
BAR42+5VALM
1 TP18
TPOINT
R15
390R
PL113 +25V
BAV CONTROLLER TYPE 2 ALARMS
10110077 4
NOTFITTED
-
8/18/2019 8500 Training Manual Iss 3
55/69
1 2 3 4 5 6
A
B
C
D
654321
D
C
B
A
Title
Number RevisionSize
B
D at e: 4 - Fe b -2002 S hee t o f
File: \\Cad 3\c\1 JONS ARCHIVE\8500archive\circuit diagrams\10110077-4\10110077-4.DDBDrawn By:
ANALOGSIG
A1
B2
C3
E14
E25
E36
Y015
Y114
Y213
Y312
Y411
Y510
Y69
Y77
U17
MC74HC138AD
X013
X114
X215
X312
X41
X55
X62
X74
INH6
A11
B10
C9
VEE7
X3
U20
MC74HC4051DW
X013
X114
X215
X312
X41
X55
X62
X7
4
INH
6
A11
B10
C9
VEE7
X3
U21
MC74HC4051DW
X013
X114
X215
X312
X41
X55
X62
X74
INH6
A11
B10
C9
VEE7
X3
U22
MC74HC4051DW
X013
X114
X215
X312
X41
X55
X62
X74
INH6
A11
B10
C9
VEE7
X3
U23
MC74HC4051DW
OC1
CLK 11
1D 2
2D3
3D4
4D5
5D6
6D7
7D8
8D9
1Q19
2Q18
3Q17
4Q16
5Q15
6Q14
7Q13
8Q12
U24
MC74HC574ADW
GND
GND
-5VA
-5VA
-5VA
-5VA
D0D1
D2
D3
D4
D5
D6
D7
D[0..15]D[0..15]
GND
ANALOGSEL
D1
BZX79C4V7
Solenoid Current
Self-Test 0
D2
BZX79C4V7
Flow Valve Current
CPP Set
Pressure Limit Set
Ratio Set
Frequency Set
Volume Set
AGND
PL2[1..50]
PL1[1..26]
PL118
PL120
PL27
PL26
PL25
PL24
PL23
PL1[1..26]
+
C5
10u 50V
+
C6
10u 50V
+
C7
10u 50V
+
C8
10u 50V
+
C9
10u 50V
D3 BAR42
AGND
+
C10
10u 50V
+
C11
10u 50V
AGND
Pressure High Set
Pressure Low Set
Patient FlowPL326 1
TP2
TPOINT
Temperature
Self-Test 1
Freshgas Flow
PL322
PL321EXTPRS
EXTFLW
AGND
+5VA
-5VA
+5VMAG
+5VALM
VCC
+6V
PL3[1..26]
PL15
PL124
SBAT
SWBAT
+25V
+24VPL113
PL116
PL3[1..26]
PL2[1..50]
PL320
PL319
PL318 Patient Pressure
Drive Pressure
Flow Pressure
D4
BAR42
D5
BAR42
D6
BAR42
AGND
PL22
PL21
3
2
1
8
4
U27A
LT1078CN8
5
6
7
U27B
LT1078CN8
5
6
7
U28B
TLC27L2CD
3
2
1
8
4
U28A
TLC27L2CD
D7
BAR42
4VREF
AGND
+5VA
-5VA
AGND
4VREF
GND
RC2
C1
A4
B5
CLR 3
Q6
Q7
U33A
MC74HC4538AD
RC14
C15
A12
B11
CLR 13
Q10
Q9
U33B
MC74HC4538AD
CLK 3
D2
S D
4
C D
1
Q5
Q6
U9A
MC74HC74AD
CLK 11
D12
S D
1 0
C D
1 3
Q9
Q8
U9B
MC74HC74AD
+ C13
47u 16V
GND
D8
BAR42
APNOEA
VCC
VCC
VCC
R6
33K
R7160K
+ C14
2u2 63V
AGND
AGND
4VREF
+5VA
-5VA
1TP3
TPOINT
+ C15
1u0 63V
D9BAR42
VCC
GNDD10
BAR42
+
C16
1u0 63V
AGND
GND
VCC
PL26
Pressure Limit Set
D11
BAR42
D12
BAR42
D13
BAR42
ALMCLR
PL14
PL16
GND
PRS_OVRLD
1
TP4
TPOINT
PL317
TEST3
TEST2
TEST1
TEST0
TEST[0..4]
RESET
1
3
5
7
2
4
6
8
RP10
10K
AGND
1
3
5
7
2
4
6
8
RP11 100K
1
3
5
7
2
4
6
8
RP1210K
1
3
5
7
2
4
6
8
RP13
10K
1
3
5
7
2
4
6
8
RP14
10K
1
3
5
7
2
4
6
8
RP15
10K
AGND
R17
1K0
R181K0
R19
1M0
R20
1M0
R21
1M0
R22
1M0
R23
1M0
R24
1M0
R25
1M0
R26
1M0
R27
47K
R28
470K
R29
10K
R30
10K
R31
10K
R32
10K
Q1
2N2222
R33
10K
R34
1M0
R35
10K
R36
33K
R37
470K
R38
33K
R39
16K0
R40
1M0
R4147K
R42
120K
R43
100K
C28
470n16V
C29
10n
R44
10K
R45
1M0
R46
4K7
1
TP1
TPOINT
R98
100K
R99
10K
TEST[0..4] 1
TP19
TPOINT
+25V
+24V
R60
10K
R96
10K
BAV CONTROLLER TYPE 2 ANALOGUE
10110077 4
-
8/18/2019 8500 Training Manual Iss 3
56/69
1 2 3 4 5 6
A
B
C
D
654321
D
C
B
A
Title
Number RevisionSize
B
D at e: 4 - Fe b -2002 S hee t o f
File: \\Cad 3\c\1 JONS ARCHIVE\8500archive\circuit diagrams\10110077-4\10110077-4.DDBDrawn By:
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
3031
32
33
34
35
36
37
38
39
40
41
42
43
44
45
46
47
48
49
50
PL2
CON50
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
PL1
CON26
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
PL3
CON26
GND
PL12
GND
PL14
PL15PL16
PL17
PL18
PL19
PL110
PL112
PL116
PL118
PL119
PL120
PL121
PL122
PL123
PL124
PL125
PL3[1..26]
PL1[1..26]
PL2[1..50]
PL1[1..26]
PL3[1..26]
PL31GND
PL33
PL34
PL37
AGND
PL39
GND
PL311
PL312
PL313
PL314
PL315
AGND
PL317
PL318
PL319
PL320
PL321
PL322
4VREF
-5VA
+5VA
PL326
+5VALM
PL244
PL243
PL242
PL241
PL240
VCC
VCC
GND
PL236
PL235
GND
PL233
PL232PL231
PL230
PL229
PL228
PL227
AGND
PL224
4VREF
+6V
PL222
PL221
PL220
PL219
PL217
PL218
PL216
PL215
PL214
PL213
PL212
PL211
PL210
PL29
PL27
PL26
PL25
PL24
PL23
PL22
PL21
PL2[1..50]
VIN8
FDBK 7
SHTDN3
TAP6
SENS2
VOUT1
ERR 5
G N D
4
U55
LP2951ACM
IN1
G N D
2
OUT3
U56
LM2937ET-5
IN1
G N D
2
OUT3
U57
LP2950CZ-5
VCC8
/ST7
/RST6
RST5
GND4
TOL3
TD2
/PBRST1
U58
LT1054CN
+ C22
100u 25V
+ C23
10u 50V
+6V
GND
VCC
+5VA
4VREF
-5VA
C38
100n
C39100n
C40100n
+6VGND
UPDOWN
PHIDSP
PLODSP
INODSP
Lamp Backlight
R90
47K
VCC
C41100nD23
BAR42
4VREF
+5VA
VCC
GND
R91
10K 0.1%
R92
300R
R93
22K1 0.1%C42
100n
C43
100n
+ C24
10u 50V
+ C25
10u 50V
+ C26
10u 50V R94
100K
R95
22K
C44
2n2
+
C27
100u 25V
-5VA
C45
100n
GND
+6V
+6V
I11
I22
I33
I44
I55
I66
I77
GND8
VCC9
O710
O611
O512
O413
O314
O215
O1 16
U59
ULN2003
1
TP13
TPOINT
1
TP14
TPOINT
1
TP15
TPOINT
1
TP16
TPOINT
1
TP17
TPOINT
UC1
1n
UC21n
UC3A
1n
UC4
1n
UC5
1n
UC6
1n
UC7
1n
UC8
1n
UC9
1n
UC10
1n
UC11
1n
UC12
1n
UC13
1n
UC14
1n
UC15
1n
UC16
1n
UC17
1n
UC18
1n
UC20A
1n
UC21A
1n
UC22A
1n
UC23A
1n
VCC
GND
Power supply decoupling - adjacent to I.C of same number
VCC
GND
UC26
1n
UC33
1n
UC34
1n
UC36
1n
UC37
1n
UC38
1n
UC39
1n
UC40
1n
UC41
1n
UC42
1n
UC43
1n
UC44
1n
UC45
1n
UC46
1n
UC51
1n
UC52
1n
UC53
1n
UC54
1n
UC55
1n
UC59
1n
UC24
1n
UC25
1n
UC50
1n
UC47
1n
UC48
1n
UC49
1n
VCC
GND
+5VA
AGND
-5VA
FP[0..2]FP[0..2]
FP1
FP2
FP0
R101
10K
R102
10K
R103
10K
R104
10K
GND VCC
GNDVCC
GND
UC6010n
UC3B
1n
UC3C
100n
UC27A
100n
UC27B
100n
UC28A
100n
UC28B
100n
UC29B
100n
UC29A
100n
UC30A
100n
UC30B
100n
UC31B
100n
UC31A
100n
UC32A
100n
+5VALM
PL113
1,2EN4
1A2
1B1
1Y3
2A6
2B7
2Y5
U60A
MC3486
1,2EN4
1A1
1Y2
1Z
3
2A7
2Y6
2Z5
U61A
MC3487
UC61
10n
PL224PL240
PL241
PL242
GND
+6V
UC20B
1n
UC21B
1n
UC22B
1n
UC23B
1n
VCC
+5VMAG
+25V
COMALM
VCC
+6V
+5VA
AGND
+5VALM
+5VALM
GND
1
TP20
TPOINT
1
TP21
TPOINT
1
TP22
TPOINT
1
TP23
TPOINT
1
TP24
TPOINT
GND
AGND
Q2ZTX650
R1054K7
GND 1,2EN12
1A10
1B9
1Y11
2A14
2B15
2Y13
U60B
MC3486
VCC
VCC
GND
GND
1,2EN12
1A9
1Y10
1Z11
2A15
2Y14
2Z13
U61B
MC3487
GND
GND
GND
1
23
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
32
33
34
PL4
CON34
422RxD
422RxD
422RxD#
422DTR
422DTR#
PL26
422RxD#
422TxD
422TxD#
422DTR 422DTR#
422CTS
422CTS#
APNOEA
PRS_OVRLD
GND
PSU+
PSU+GND
GND
422CTS
422CTS#
422TxD
422TxD#
PL317
PL318
PL319
PL320
PL321
BATONCHG
AGND
PL326GND
PL110
PL15
PL18
PL16
PL14
PL54PL12
1
2
3
4
PL5
CON4
PL54
BATONCHG
PSU+GND
BAV CONTROLLER TYPE 2 CONNECT
10110077 4
PL247
PL248PL249
PL250
PL246
-
8/18/2019 8500 Training Manual Iss 3
57/69
1 2 3 4 5 6
A
B
C
D
654321
D
C
B
A
Title
Number RevisionSize
Orcad B
D a te : 4 - Fe b -2002 S hee t o f
File: \\Cad 3\c\1 JONS ARCHIVE\8500archive\circuit diagrams\10110077-4\10110077-4.DDBDrawn By:
MASTER
A[1..19]
D[0..15]
CS[0..3]
C[0..4]
CLK
WATCHDOG
D[0..15]
D15
D14
D13
D12
D11
D10
D9
D8
D7D6
D5
D4
D3
D2
D1
D0
CS0CS1
CS2
CS3
RESET
HALT
RESET
HALT
ICTL0
ICTL1
ICTL2
ICTL0
ICTL1
ICTL2
WATCHDOG
CLK
C[0..4]
C0
C1
C2
C3
C4
C0
C1
C2
C3
DELAY
CS[0..3]
C5
C7
C7
FC0
FC1FC2
AVEC
AVEC
BERR
BERR
BCLR
BCLR
NVHOLD NVWP
NVHOLD
NVWP
FC1
FRZ
FRZ
FC2
A[1..19]
FC0
SPEN
DELAY
C5
VCC
VCC
VCC
VCC
GND
/ST7
/RST6
RST5
TOL3
TD2
/PBRST1
U1
MAX1232CPA
1
2
JP1
HEADER2
1
2
3
JP2
34
U4B
SN74HC05D
A 2 3
2 7
A 2 2
2 6
A 2 1
2 5
A 2 0
2 4
A 1 9
2 2
A 1 8
2 1
A 1 7
2 0
A 1 6
1 9
A 1 5
1 7
A 1 4
1 6
A 1 3
1 5
A 1 2
1 4
A 1 1
1 2
A 1 0
1 1
A 9
1 0
A 8
9
A 7
8
A 6
7
A 5
6
A 4
5
A 3
3
A 2
2
A 1
1
D1530
D1431
D1332
D1233
D1135
D1036
D937
D838
D740
D641
D542
D443
D345
D246
D147
D048
EXTAL100
XTAL101
CLKO98
/RESET92
/HALT91
/BERR 94
BUSW74
DISCPU73
/FRZ72
/AS104
R/~W103
/UDS-A0106
/LDS-DS105
/DTACK 85
/RMC-IOUT1123
IAC122
/BCLR 86
/BR 90
/BG87
/BGACK 88
/IPL097
/IPL196
/IPL295
FC0132
FC1130
FC2129
/AVEC-IOUT093
R X D 1
5 2
T X D 1
8 0
R C L K 1
8 2
T C L K 1
8 1
/ C T S 1
5 1
/ R T S 1
7 9
/ D C D 1
5 0
B R G 1
7 6
R X D 2 - P
A 0
5 3
T X D 2 - P
A 1
5 4
R C L K 2 - P
A 2
5 5
T C L K 2 - P
A 3
5 6
/ C T S 2 - P
A 4
5 8
/ R T S 2 - P
A 5
5 9
/ D C D 2 - P
A 6
6 0
B R G 2 - S
D S 2 - P
A 7
6 1
R X D 3 - P
A 8
6 3
T X D 3 - P
A 9
6 4
R C L K 3 - P
A 1 0
6 5
T C L K 3 - P
A 1 1
6 6
/ C T S 3 - S
P R X D
4 9
/ R T S 3 - S
P T X D
7 8
/ D C D 3 - S
P C L K
7 7
B R G 3 - P
A 1 2
6 8
/DREQ-PA1369
/DACK-PA1470
/DONE-PA1571
/ I A C K 7 - P
B 0
1 0 8
/ I A C K 6 - P
B 1
1 0 9
/ I A C K 1 - P
B 2
1 1 0
T I N 1 - P
B 3
1 1 1
/ T O U T 1 - P
B 4
1 1 3
T I N 2 - P
B 5
1 1 4
/ T O U T 2 - P
B 6
1 1 5
/ W D O G - P
B 7
1 1 7
/CS0-IOUT2128
/CS1127
/CS2125
/CS3124
P B 8
1 1 8
P B 9
1 1 9
P B 1 0
1 2 0
P B 1 1
1 2 1
U3
MC68302FC16C
R1
1K2
R2
1K2
12
U4A
SN74HC05D
GND
R97
10K
VCC
8
U5
12.288MHz
S C C 1 0
S C C 1 1
S C C 1 4
S C C 1 6
S C C 2 0
S C C 2 1
S C C 2 5
S C C 2 6
S C C 3 4
S C C 3 6
SCC20
SCC21
SCC25
SCC24
+ C1
4u7 63V
+
C210u 50V
+C3
10u 50V
+ C4
4u7 63V
GND
VCC
PL3[1..26]
PL312
PL314
PL311
PL313
TXD
DTR
RXD
CTS
CLK 3
D2
S D
4
C D
1
Q5
Q6
U7A
MC74HC74AD
VCC
VCC
1
3
5
7
2
4
6
8
RP1
10K
1
3
5
7
2
4
6
8
RP2
10K
1
3
5
7
2
4
6
8
RP3
10K
1
3
5
7
2
4
6
8
RP4
10K
VCC
1
3
5
7
2
4
6
8
RP5
10K GND
13
5
7
24
6
8
RP6
4K7
VCC
R4O17
R3O22
R2O4
R1O6
T4I21
T3I19
T2I18
T1I5
T1O2
T2O1
T3O24
T4O20
R1I7
R2I3
R3I23
R4I16
C2-14
C2+13
V-15
C1-12
C1+10
V+11
U10
MAX238CNG
1 2
3 4
5 6
7 8
9 10
PL6
HEADER5X2
GND
DB9 - Recorder
DB25 - RS232
Auxiliary Serial Connector
VCC
VCC
SCC11
SCC10
PULFLW
FLWSIG
+5VA
AGND C L K
PWMDIVCLK
OVERLOAD_INT
S C C 2 4
SCC24
PWMCLK
4VREF
ANALOGSIG
UPDOWN
PL37
PL39
PL33
PL235
PL121
PL1[1..26]PL1[1..26]
PL2[1..50]PL2[1..50]
VCC
GND
Calibrate Protection
S C C 1 2
S C C 1 3
13
5
7
24
6
8
RP7
10K
SCC12SCC13
SCC14
SCC16
GND
S C C 2 2
S C C 2 3
SCC22
SCC23
SCC26
1
35
7
2
46
8
RP8
10K GND
SCC34
SCC36
PL214
RESET
TEST[0..4]TEST[0..4]
TEST0
A 1
A 2
A 3
A 4
A 5
A 6
A 7
A 8
A 9
A 1 0
A 1 1
A 1 2
A 1 3
A 1 4
A 1 5
A 1 6
A 1 7
A 1 8
A 1 9
R3
10K
R4
10K GND
R5
10K
Output data enable
Output data
Input data
FP[0..2]FP[0..2]
FP0
FP1
FP2
R8
10K VCC
GND
Pressure Switch
56
U4C
SN74HC05D
89
U4D
SN74HC05D
10 11
U4E
SN74HC05D
12 13
U4F
SN74HC05D
CLK 11
D12
S D
1 0
C D
1 3
Q9
Q8
U7B
MC74HC74AD
VCC
VCC
GND
R9
4K7VCC
C4
PL315
VREF1
+IN2
-IN3
GND4
CS5
VCC8
DOUT6
CLK 7
U11
LTC1286CS8
R11
1K2
R12
4K7
R13
4K7
R14
4K7
VCC VCC
VCC
VCC
1
2
JP4
HEADER2
PL3[1..26]
/WP3
/HOLD7
SCK 6
SI 5
SO2
/CS1
U6
25C020N
SPEN
VCC
ON_24V
ON_24V
VDD
R16
10K
VCC
12 13
U41F
MC74HC14AD
R106
20K
C4710n
GND
VDD
R107
100R
BAV CONTROLLER TYPE 2 CPU
10110077 4
R108
10K
PL249
SWITCH TO VCC FORCE FLASH MODE
DTR used to drive
Autozero Valve
ALARMS
ALARMS-4.SCH
CONNECT
CONNECT-4.SCH
ANALOGUE
ANALOGUE-4.SCH
IO
IO-4.SCH
MEMORY
MEMORY-4.SCH
PWM
PWM-4.SCH
S2
-
8/18/2019 8500 Training Manual Iss 3
58/69
1 2 3 4 5 6
A
B
C
D
654321
D
C
B
A
Title
Number RevisionSize
B
D at e: 4 - Fe b -2002 S hee t o f
File: \\Cad 3\c\1 JONS ARCHIVE\8500archive\circuit diagrams\10110077-4\10110077-4.DDBDrawn By:
1A12
1A24
1A36
1A48
2A111
2A213
2A315
2A417
1Y118
1Y216
1Y314
1Y412
2Y19
2Y27
2Y35
2Y43
1G1
2G19
U44
MC74HC244ADW
1A12
1A24
1A36
1A48
2A111
2A213
2A315
2A417
1Y118
1Y216
1Y314
1Y412
2Y19
2Y27
2Y35
2Y43
1G1
2G19
U45
MC74HC244ADW
D0
D1
D2
D3
D4
D5
D6
D7
D8
D9
D10
D11
D12
D13
D14
D15
PL18
PL16
PL110
PL14
PL12
Adult On
Adult Standby
Off
Paed Standby
Paed On
PL1[1..26]
VCC
PL232 /PRSSEL
/PWR_FAIL
VCC
PL112 /OVRCH
PL231 /MUTE
/TSTRSTPL230
/VOLSELPL233
/INTLOCPL229
DISCON
+ C18
10u 50V
GND
PL236
PL2[1..50]
PL1[1..26]
PL2[1..50]
A1
B2
C3
E1 4E2
5E3
6
Y015
Y114
Y213
Y312
Y411
Y510Y6
9Y7
7
U18
MC74HC138AD
ANALOGSEL
DACSTRB
DACCS
A3
A4
A5
CS2
D15D14
D13
D12
D11
D10
D9
D8
D7
D6
D5
D4
D3
D2
D1
D0
ALARM7
ALARM6
ALARM5
ALARM4
ALARM3
ALARM2
ALARM1
ALARM[1..7]ALARM[1..7]
TEST1
PL125
PL119
ON_PSU
ON_BAT
ON_LEK PL19
CLRBAK
EN_SOL
EN_FLW
TEST2
CS3
A1
A3
A2
A1
B2
C3
E14
E25
E36
Y015
Y114
Y213
Y312
Y411
Y510
Y69
Y77
U46
MC74HC138AD
D12
D23
D34
D4
5
D56
D67
D78
D89
Q118
Q217
Q316
Q4
15
Q5 14
Q613
Q712
Q811
E11
E219
U47
MC74HC540DW
GND
D7
D6
D5
D4
D3
D2
D1
D0
PL2[1..50]PL2[1..50]
PL29
PL210
PL211
PL212
PL215
PL216
PL217
PL218
PL219
PL220
PL221
PL222
GND
PL213
STRB1
STRB2
STRB3
STRB4
STRB5
GND
CS[0..3]
A[1..19]
CS[0..3]
A[1..19