5. TEM - Electron Microscopy and Diffraction
description
Transcript of 5. TEM - Electron Microscopy and Diffraction
Electron Microscopy Electron Microscopy and Diffractionand Diffraction
5. 5. Transmission Electron MicroscopeTransmission Electron Microscope
Do Minh NghiepMaterials Science Center
01.01.2009 Materials Science Center, HUT 2
Electron optics and instrument Image contrast (mass thickness
contrast, phase contrast, diffraction contrast)
Magnification and electron beam adjustment
Sample preparation Application
ContentContent
Introduction Introduction
01.01.2009 3Materials Science Center, HUT
What’s TEM What’s TEM The transmission electron microscope (TEM) is an analytical tool that allows:detailed microstructural examination through high-resolution and high-magnification imaging: magnifications of up to 500,000x and detail resolution below 1 nm are achieved routinely, investigation of crystal structures and orientations through electron diffraction pattern,determination of chemical compositions in phases, precipitates and contaminants through X-ray and electron spectroscopy: qualitative and quantitative elemental analysis can be provided from features smaller than 30 nm.
01.01.2009 Materials Science Center, HUT 4
What’s TEM imageWhat’s TEM image Basic idea of a TEM is to
project a magnified image of the specimen onto a fluorescent screen where it can be viewed by the user.
The image itself is the result in intensity difference of beam electrons that are forward scattered by the specimen vs. those that are not (transmitted unscattered) .
and aperture
Objective
Projective
01.01.2009 5Materials Science Center, HUT
Pros and consPros and cons
AdvantageAdvantage
- TEM is the only technique that will allow the determination of the Burgers vector.
- TEM can also provide atomic scale resolution (2 Å). DisadvantageDisadvantage
- TEM is an expensive and destructive technique.
- Some materials are sensitive to electron beam radiation, resulting in a loss of crystallinity and mass.
- Sample preparation is very time-consuming, sample dimension small (3 mm diameter, less than 100 nm thick)
- Artifacts cause misleading for the interpretation of transmission images.
This usually causes by the fact that the TEM presents us with 2D-images of 3D-specimens viewed in transmission.
01.01.2009 Materials Science Center, HUT 6
3D-object 3D-object 2D-image 2D-image
01/01/2009 Handouts-MSE4346 -K51 KHVL 7
A single projection image is plainly insufficient to infer the structure of
an object.
Each image represents a 2D projection of the 3D object
3D-object 3D-object 2D-image 2D-image
01/01/2009 Handouts-MSE4346 -K51 KHVL 8
Watch out!A cover slide!
InstrumentInstrument
01.01.2009 9Materials Science Center, HUT
TEM components TEM components and electron pathand electron path
A simplified ray diagram of a TEM consists of - an electron source, - condenser lens with aperture,- specimen, - objective lens with aperture, - projector lens and - a fluorescent screen.
01.01.2009 10Materials Science Center, HUT
SuplementarySuplementary
In actuality a modern TEM consists of many more components including- a dual condenser system,- stigmators, - deflector coils, and -a combination of intermediate and dual projector lens.
Sîi ®èt
AnètChao
Khe cè ®ÞnhCuén chØnh nguån
Condenser 1
Condensor 2
Stigmator 1Khe cè ®Þnh
DeflectorKhe tô kÝnh
VËt kÝnh
Khe vËt kÝnh
Gi¸ gi÷ mÉu
Stigmator 2
KÝnh trung gianKhe kÝnh trung gian (nhiÔu x¹)
Dual projector
Mµn huúnh quang
Focus wobble
01.01.2009 11Materials Science Center, HUT
Conventional TEMsConventional TEMs
01.01.2009 Materials Science Center, HUT 12
200 kV (left) and 100 kV (right)
TEM
Modern TEMsModern TEMs
01.01.2009 13Materials Science Center, HUT
High Resolution Transmission High Resolution Transmission Electron Microscope (HRTEM)Electron Microscope (HRTEM)
01.01.2009 14Materials Science Center, HUT
HRTEM - a comparisonHRTEM - a comparison
01.01.2009 15Materials Science Center, HUT
Image contrast Image contrast
01.01.2009 16Materials Science Center, HUT
Imaging mechanismImaging mechanism
01.01.2009 17Materials Science Center, HUT
Phase contrastPhase contrast
18
Bright field (BF) Dark field (DF)
Parallel incident beam
Phase contrast/ image mode
Materials Science Center, HUT01.01.2009
Diffraction contrastDiffraction contrast
01.01.2009 Materials Science Center, HUT 19
Diffraction contrast/ image mode (Electron microdiffraction)
Convergent incident beam
Diffraction contrastDiffraction contrast
01.01.2009 Materials Science Center, HUT 20
If a monochromatic e-beam of known l strikes a crystal at the appropriate Bragg’s angle a number of the diffracted electrons will be forward scattered.
Like the transmitted electrons these diffracted electrons will have nearly their same energy but will have been significantly altered from their trajectory.
Diffraction contrastDiffraction contrast
01.01.2009 Materials Science Center, HUT 21
The transmitted electrons will be brought to convergence in the back focal plane of the objective lens (Y).
Likewise the diffracted electrons will also be brought to convergence in the back focal plane of the lens but at a different spot (X).
.
01.01.2009 Materials Science Center, HUT 22
Normally an aperture is placed in the back focal plane of the objective lens to stop widely scattered electrons from reaching the viewing screen, but in the case of diffraction it is these same scattered electrons that contain the information about the diffraction event.
To operate the TEM in diffraction mode
the objective aperture is removed from the beam path and the microscope is adjusted to focus an image of the back focal plane of the objective lens, not the image plane.
Diffraction contrastDiffraction contrast
Diffraction contrastDiffraction contrast
01.01.2009 Materials Science Center, HUT 23
This is most easily accomplished by adjusting the strength of the objective lens so that an image of the back focal plane is projected onto the viewing screen.
Back focal plane
01.01.2009 24
DF imageDF imageBF imageBF image
Mass-thickness contrast (fringes)Mass-thickness contrast (fringes)
Materials Science Center, HUT
Contrast is caused by:- Unscattered electrons coming into screen (bright area)- Scattered electrons not coming into screen (dark area)
Contrast imageContrast image
Scattered eTransmitted e(unscattered)
01.01.2009 25Materials Science Center, HUT
MagnificationMagnification
01.01.2009 26Materials Science Center, HUT
Total magnification in the TEM is a combination of the magnification from the objective lens times the magnification of the intermediate lens times the magnification of the projector lens.
Each of which is capable of approximately 100x.
Mob x Mint x Mproj = Total Mag so Mtotal can be a magnitude of 106
What’s magnificationWhat’s magnification
01.01.2009 27Materials Science Center, HUT
Under different conditions of high and low magnification the same component can have different functions:
MagnificationMagnification
01.01.2009 .28Materials Science Center, HUT
Contrast forming
Image focus
Diffraction focus
Diffraction stigmation
Image stigmation
Area selection
Low magnificationHigh magnification
Objective lens
Diffraction lens
Objective aperture
SA aperture
Object. stigmator
Diffract. stigmator
Image focus
Diffraction focus
Contrast forming
Area selection
Diffraction stigmation
Image stigmation
Components
Depth of Field (Dfi): the range of distance at the specimen parallel to the illuminating beam in which the object appears to be in focus. Depth of Focus (Dfo): the range of distance at the image plane (i.e. the eyepiece, camera, or photographic plate) in which a well focussed object appears to be in focus.
Magnification and focusing Magnification and focusing
01.01.2009 29Materials Science Center, HUT
In TEM thanks to great Dfi and Dfo the image is well focuced even at high magnification
Magnification and focusing Magnification and focusing
The depth of focus is so great for the projector lens system that an image that is in focus on the screen will also be in focus on plate film beneath the screen or even a TV camera below the plate film
Fluorescent screen
Binocular viewing scope
Viewport
Plate camera
01.01.2009 30Materials Science Center, HUT
Adjustment of Adjustment of electron beamelectron beam
01.01.2009 31Materials Science Center, HUT
Adjustment of condenserAdjustment of condenser
The role of the condenser lenses is to make the beam that is striking the specimen as nearly parallel as is possible.
Condenser lens
Specimen
Primary beam
01.01.2009 32Materials Science Center, HUT
Adjustment of condenserAdjustment of condenser
As magnification increases the condenser lens must be adjusted to properly illuminate the specimen. When the lens is brought to its smallest spot the beam is said to be at the crossover point.
Screen before focal point: large beam (under focused)
Screen at focal point: minimum beam
size/crossover, maximum brightness (focused)
Screen behind focal point: large beam
(over focused)
(-) Change in condenser lens current (+)
01.01.2009 33Materials Science Center, HUT
Holey (not Holy) Formvar is used to critically adjust the stigmation of a TEM. When the beam is under or over focused on the specimen a Fresnel fringe becomes visible due to the effects of diffraction around the edges of the whole. When this Fresnel fringe is evenly distributed then the beam is said to be stigmated.
StigmatorStigmator
01.01.2009 34Materials Science Center, HUT
Deflection coils
In older TEMs functions such as gun and beam alignment were accomplished by physically moving components in the column.
Today they are achieved by use of electromagnetic deflection coils that are positioned throughout the column.
01.01.2009 35Materials Science Center, HUT
Shift / tilt of beam by deflecton coilsShift / tilt of beam by deflecton coils
Using the deflection coils the beam can be: shifted so that the focused beam is centered in the back focal plane of the lens and tilted so that the beam is centered on the specimen. - deflection angle
01.01.2009 36Materials Science Center, HUT
Pivot pointPivot point
We call these centering spots “pivot points” and as the beam is shifted or tilted back and forth there should be no apparent movement if the microscope is properly aligned
Pivot point
01.01.2009 37Materials Science Center, HUT
Specimen stage positionSpecimen stage position
Likewise the position of the specimen relative to the beam is of critical importance. To adjust the height or “Z” position of the specimen it is physically rocked back and forth until an object in the center no longer moves.
Specimen stage
Specimen
01.01.2009 38Materials Science Center, HUT
SpecimenSpecimen stagestage
In-column motion of specimen: Z-Y-Z shift, rotation/ tilt around a common axis by coresponding mechanism.
01.01.2009 39Materials Science Center, HUT
Sample tilt in single obj. lensSample tilt in single obj. lens A major problem with TEM
lens design is the fact that there is very little space between the pole pieces of the objective lens to accommodate the specimen and objective aperture.
This also puts severe constraints on how far the specimen can be tilted within the lens.
01.01.2009 40Materials Science Center, HUT
Twin/double objective lensTwin/double objective lens
A clever solution to this problem came in the invention of the double objective or “twin” lens.
In this lens design the specimen actually lies between two separate lens fields allowing for tilt angles as large a 60o from perpendicular to the optical axis.
01.01.2009 41Materials Science Center, HUT
Focus Wobbler shifts the image back and forth, and when the movement is stable the image is focused.
01.01.2009 42Materials Science Center, HUT
Specimen Specimen preparationpreparation
01.01.2009 43Materials Science Center, HUT
Wedge sample Wedge sample -”windows” technique-”windows” technique
01.01.2009 44Materials Science Center, HUT
ElectropolishingElectropolishing
01.01.2009 45Materials Science Center, HUT
Sample sections after Sample sections after electropolishing and sample gridselectropolishing and sample grids
01.01.2009 46Materials Science Center, HUT
Cutting thin sample Cutting thin sample by microtomy by microtomy
01.01.2009 47Materials Science Center, HUT
Replica foilsReplica foils
01.01.2009 48Materials Science Center, HUT
01.01.2009 49
TEM grids
3 mm
Instrumentation Instrumentation
Materials Science Center, HUT
01.01.2009 50
Sample positions
Cooling
Standard
Heating
Sample holdersSample holders
Materials Science Center, HUT
ApplicationsApplications
01.01.2009 51Materials Science Center, HUT
Dislocation structureDislocation structure(diffraction contrast) (diffraction contrast)
01.01.2009 52Materials Science Center, HUT
Contrast image Contrast image and and
microdiffraction microdiffraction pattern of two pattern of two
screw dislocation screw dislocation groups with groups with
different Burgers different Burgers vectors vectors
01.01.2009 53Materials Science Center, HUT
BF and DF BF and DF images of images of
dispersive dispersive precipitates precipitates
in Al-Cr-Zr in Al-Cr-Zr alloyalloy
01.01.2009 54Materials Science Center, HUT
Lattice Lattice image of image of
(111) (111) planes in planes in
Ga-AsGa-As
01.01.2009 55Materials Science Center, HUT
HR image HR image of of
phase in phase in Al alloysAl alloys
01.01.2009 56Materials Science Center, HUT
HR contrast HR contrast image of image of
metal atom metal atom arrangement arrangement
in 1,97 nm in 1,97 nm thick thick
specimen specimen of Nb-W-O of Nb-W-O
01.01.2009 57Materials Science Center, HUT