3041 Extended Specifications V1.11 Oct 2006 Extended Specifications CE Declartation : V1.11 Page 3...

25
3041 MULTI-PRODUCT CALIBRATOR 25ppm MULTI-PRODUCT CALIBRATOR EXTENDED SPECIFICATIONS TRANSMILLE LTD., UNIT 4, SELECT BUSINESS CENTRE, STAPLEHURST, KENT. TN12 0QW. UK. www.transmille.com : [email protected] : Tel : +44 (0) 1580 890700 : Fax : +44 (0) 1580 890711

Transcript of 3041 Extended Specifications V1.11 Oct 2006 Extended Specifications CE Declartation : V1.11 Page 3...

Page 1: 3041 Extended Specifications V1.11 Oct 2006 Extended Specifications CE Declartation : V1.11 Page 3 3041 MULTI-PRODUCT CALIBRATOR : DCV SPECIFICATIONS 1 Year Total Accuracy Specifications

3041 MULTI-PRODUCT CALIBRATOR25ppm MULTI-PRODUCT CALIBRATOR

EXTENDED SPECIFICATIONSTRANSMILLE LTD., UNIT 4, SELECT BUSINESS CENTRE, STAPLEHURST, KENT. TN12 0QW. UK.

www.transmille.com : [email protected] : Tel : +44 (0) 1580 890700 : Fax : +44 (0) 1580 890711

Page 2: 3041 Extended Specifications V1.11 Oct 2006 Extended Specifications CE Declartation : V1.11 Page 3 3041 MULTI-PRODUCT CALIBRATOR : DCV SPECIFICATIONS 1 Year Total Accuracy Specifications

3041 MULTI-PRODUCT CALIBRATOR :: GENERAL SPECIFICATIONSWarm Up Time Double the time since last used up to 20 minutes maximumStandard Interfaces RS232Optional Interfaces GPIB (IEEE-488) : USB (Universal Serial Bus)Temperature Performance Storage : -5°C to +60°C

Operation : 0°C to +50°CRelative Humidity Operation : <80% to 30°C, <70% to 40°C, <40% to 50°C

Storage : <95%, non-condensingAltitude Operation : 3000m (10,000ft) Maximum

Transit : 12000m (40,000ft) MaximumEMC & Safety The calibrator line input plug must be earthed

See D.O.C for full detailsLine Power Line Voltage Selectable : 110V / 230V

Line Frequency : 50Hz to 60HzLine Voltage Variation : -6% +10%

Power Consumption 28 Watts (Standby) 200 Watts (Maximum)Low Analogue Isolation 100VConnections Voltage / 2 Wire Resistance 1x Black : 1x White 4mm Safety sockets

Low Current (<=2A) 1x Black : 1x Red 4mm Safety socketsHigh current (>2A) 1x Blue : 1x Yellow 4mm Safety socketsEarth Connection 1x Green 4mm Safety SocketOscilloscope Functions 2x BNC terminalFeature (Ext. Pod) 1x Female 'D' type socketRS232 Interface 1x Female 'D' type socket

RS232 Settings Baud Rate 9600Parity NoneData Bits 8Stop Bits 1

Display Information Type Backlit Black on white film STN typeViewing Area 124.3mm * 34mmResolution 256 * 94 dotsBacklight Type Cold fluorescent lampBrightness 70 to 90 cd/m2

Indicators Voltage / Current / High Current Red LED (between terminals)Negative to ground Green LED (left of Earth terminal)Oscilloscope Green LED (right of BNC Connector)Feature Connector (Ext. Pod) Green LED (right of 'D' type connector)

Keyboard Rubber keyFuses Mains Inlet 3.15A A/S (240 Volt)

5A A/S (110 Volt operation)Isolation Outputs are opto-isolated from mains earth and the RS-232 interface

Maximum common mode voltage between earth and thelow terminals 30 Volts ac/dc.

Dimensions & Weights Calibrator Only 14cm x 43cm x 46cm : 12.5kgsCalibrator in Shipping Box 58cm x 56cm x 37cm : 15kgsCalibrator in Soft Carry Case 49cm x 50cm x 19cm : 13.5kgsCalibrator in Hard Transit case 55cm x 56cm x 26cm : 22kgs

Warranty Period 3 Years (Parts & Labour)Recommended Service Interval 1 YearSupplied Connections 1x Serial Interface Connection 1x Mains Lead

1x Adaptor Connection Lead (if at least one adaptor ordered)Optional Lead Set Kit 1x Voltage connection leadset

1x Low Current connection leadset1x High current connection leadset1x AC connection leadset

Mounting Kit (optional) 3U rack mount kitCase Colour Cream (RAL 9002)

Due to continuous development specifications may be subject to change.3041 Extended SpecificationsGeneral Specifications : V1.11 Page 2

Page 3: 3041 Extended Specifications V1.11 Oct 2006 Extended Specifications CE Declartation : V1.11 Page 3 3041 MULTI-PRODUCT CALIBRATOR : DCV SPECIFICATIONS 1 Year Total Accuracy Specifications

DECLARATION OF CONFORMITY

Manufacturer’s Name: Transmille Ltd.Manufacturer’s Address: Unit 4, Select Business Centre

Lodge RoadStaplehurstTN12 0QW

Declares, that the product

Product Name: Multi-product CalibratorModel Number: 3050 / 3041 / 3010Product Options: This declaration covers all options of the above product(s)

Conforms with the following European Directives:

The product herewith complies with the requirements of the Low Voltage Directive 73/73EEC and the EMC Directive 89/336/EEC (including 93/68/EEC) and carries the CE Marking accordingly

Conforms with the following product standards:

EMCIEC616326-1:1997+A1:1998 / EN 61326-1:1997+A1:1998 EN55011:1991

Standard LimitIEC 61000-4-Group 1Class AIEC 61000-4-3:1995 / EN 61000-4-3:1995 4kV CD, 8kV ADIEC 61000-4-4:1995 / EN 61000-4-4:1995 3 V/m, 80-1000 MHzIEC 61000-4-5:1995 / EN 61000-4-5:1995 0.5kV signal lines, 1kV power linesIEC 61000-4-6:1996 / EN 61000-4-6:1996 0.5kV line-line, 1kV line-groundIEC 61000-4-11:1994 / EN 61000-4-11:1994 3V, 0.15-80 MHz I cycle, 100%

Dips: 30% 10ms; 60% 100msInterrupt > 95%@5000ms

SAFETYIEC 61010-1:1990+A1:1992+A2:1995 / EN 61010-1:1993+A2:1995

16/01/2006

Revision N 1 Managing DirectorDate :16/01/2006

Due to continuous development specifications may be subject to change.3041 Extended SpecificationsCE Declartation : V1.11 Page 3

Page 4: 3041 Extended Specifications V1.11 Oct 2006 Extended Specifications CE Declartation : V1.11 Page 3 3041 MULTI-PRODUCT CALIBRATOR : DCV SPECIFICATIONS 1 Year Total Accuracy Specifications

3041 MULTI-PRODUCT CALIBRATOR : DCV SPECIFICATIONS

1 Year Total Accuracy Specifications at TCal ±5°C & Range ParametersRange 1 Year Total

ppm set uV0-202mV 0.1uV 1mA2 50 Ohms 20 V 30 + 3.60.2-2.02V 1uV 50mA 0.2 Ohms 150V 30 + 52-20.2V 10uV 50mA 0.2 Ohms 150V 25 + 4020-202V 100uV 20mA3 0.5 Ohms 1200V 30 + 400

200-1025V 1mV 20mA30.7 Ohms 1200V 30 + 4000

Stability (Accuracy relative to calibration Standards)Range Noise4 90 day Rel 180 Day Rel 1 year Rel 2 year Rel

ppm Set Rng 1Hz to 10Hz ppm Set uV ppm Set uV ppm Set uV ppm Set uV

0-202mV 2 + 1 60nV 21.6 + 3 24.3 + 3 27 + 3 37.8 + 4.20.2-2.02V 2 + 1 280nV 21.6 + 4 24.3 + 4 27 + 4 37.8 + 5.62-20.2V 2 + 1 2.5uV 17.6 + 39 19.8 + 39 22 + 39 30.8 + 54.620-202V 3.5 + 1 50.7uV 21.6 + 400 24.3 + 400 27 + 400 37.8 + 560

200-1025V 5 + 2 280uV 21.6 + 4000 24.3 + 4000 27 + 4000 37.8 + 5600

NotesNote 1 : Specifications apply up to 10% of maximum load current. Above this level, allowance must be made for output resistance.Note 2 : Limited by 50 Ohm output impedance.Note 3 : Internally adjustable from 2mA to 30mA - Factory set to 20mA as standard. For safety the trip is controlled by a fail-safe circuit independant of the processor which shuts the high voltage output off in the event of an overload.Note 4 : Typical RMS noise figures at 50% of full scale.

High Voltage SafetyHigh voltage output is ramped to allow instrument under test to auto range.Standby is automatically activated when setting voltages greater than 20V or 200V from a lower voltage.Standby is automatically selected for high voltage (>20V) after 20 minutes on the same setting.High voltage (> 20V) output is indicated to user through an audible warning beep.An external high voltage output/standby control switch is available as an option.

2 Wire output / Remote sensing not available.Isolation : Floating or grounded selection available as standard.Maximum floating voltage : 100VSpecifications apply between 17°C and 27°C. Outside this range an allowance of 0.18 x 1 Year Spec. per °C should be added.

Worked Accuracy Calculation of 5V output on the 20V range at 20°C using 1 year specification

25ppm Set (Output setting on calibrator = 5V) 25ppm of 5V = 125uVZero or Floor Allowance 40uVTotal accuracy of calibrator = ± 165uV

OverloadProtection

OutputResistance1

Resolution Max. BurdenCurrent

24 Hour Stability

Due to continuous development specifications may be subject to change.3041 Extended SpecificationsDCV Specifications : V1.11 Page 4

Page 5: 3041 Extended Specifications V1.11 Oct 2006 Extended Specifications CE Declartation : V1.11 Page 3 3041 MULTI-PRODUCT CALIBRATOR : DCV SPECIFICATIONS 1 Year Total Accuracy Specifications

3041 MULTI-PRODUCT CALIBRATOR : DCI SPECIFICATIONS

1 Year Total Accuracy Specifications at TCal ±5°C & Range ParametersRange Resolution Max. Inductive Compliance Overload 1 Year Total

Load Voltage Protection % set uA0-202uA 100pA 10mH 150V 0.01 + 0.03

0.2-2.02mA 1nA 10mH 150V 0.008 + 0.042-20.2mA 10nA 10mH 150V 0.005 + 0.320-202mA 100nA 10mH 150V 0.008 + 30.2-2.02A 1uA 10mH 150V 0.015 + 35

2-30A 10uA 10mH 150V 0.04 + 350

Stability (Accuracy relative to calibration Standards)Range Noise 1 90 day Rel 180 Day Rel 1 year Rel 2 year Rel

0.1-1Hz %Set uA %Set uA %Set uA %Set uA0-202uA 180pA 0.0064 + 0.03 0.0072 + 0.03 0.008 + 0.03 0.0112 + 0.042

0.2-2.02mA 500pA 0.0056 + 0.04 0.0063 + 0.04 0.007 + 0.04 0.0098 + 0.0562-20.2mA 4nA 0.0032 + 0.3 0.0036 + 0.3 0.004 + 0.3 0.0056 + 0.4220-202mA 40nA 0.004 + 3 0.0045 + 3 0.005 + 3 0.007 + 4.20.2-2.02A 1uA 0.0064 + 35 0.0072 + 35 0.008 + 35 0.0112 + 49

2-30A 220uA 0.024 + 350 0.027 + 350 0.03 + 350 0.042 + 490

NotesNote 1 : Typical RMS noise figures at 50% of full scale.Note 2 : Power & temperature sensor on 20A range - microprocessor monitors & protects from overheating.

Higher resistance loads allow a longer ON period. See graphs 1 and 2 for details.Note 3 : Specifications apply to loads of less than 10% of the maximum burden voltage.Note 4: Zero or floor allowance.

Specifications apply between 17°C and 27°C. Outside this range an allowance of 0.18 x 1 Year Spec. per °C should be added.

Worked Accuracy Calculation of 1.2mA output on the 2mA range at 20°C using 1 year spec.

0.008% Set (Output setting on calibrator = 1.2mA) 0.008% of 1.2mA = 96nAZero or Floor Allowance 40nATotal accuracy = ± 136nA

4.2 Volts3.9 Volts

4.2 Volts4.2 Volts4.2 Volts4.2 Volts

Due to continuous development specifications may be subject to change.3041 Extended SpecificationsDCI Specifications : V1.11 Page 5

Page 6: 3041 Extended Specifications V1.11 Oct 2006 Extended Specifications CE Declartation : V1.11 Page 3 3041 MULTI-PRODUCT CALIBRATOR : DCV SPECIFICATIONS 1 Year Total Accuracy Specifications

3041 DC Current Specifications

Graph 1* : Operating time on 20A range with current into a short circuit at 20°CContinuous current in availiable below 7A output.

Graph 2* : Operating time on 20A range with current into a 0.1ohm load at 20°CContinuous current in availiable below 10.5A output.

* Note Timing is started after a minimum period of 7 minutes at zero output.Shorter periods will reduce the output time availiable.

Due to continuous development specifications may be subject to change.3041 Extended SpecificationsDCI Specifications : V1.11 Page 6

Page 7: 3041 Extended Specifications V1.11 Oct 2006 Extended Specifications CE Declartation : V1.11 Page 3 3041 MULTI-PRODUCT CALIBRATOR : DCV SPECIFICATIONS 1 Year Total Accuracy Specifications

3041 MULTI-PRODUCT CALIBRATOR : ACV SPECIFICATIONS1 Year Total Accuracy Specifications at TCal ±5°C & Range Parameters

Range Frequency Resolution OutputResistance % set uV

0-202mV 10 to 44Hz 1uV 1mA 1 50 Ohms 20 V 0.2 + 5045 to 999Hz 1uV 1mA 1 50 Ohms 20 V 0.04 + 201 to 19.999kHz 1uV 1mA 1 50 Ohms 20 V 0.09 + 3520 to 99.999kHz 1uV 1mA 1 50 Ohms 20 V 0.3 + 70100 to 500kHz 1uV 1mA 1

50 Ohms 20 V 0.8 + 3800.2-2.02V 10 to 44Hz 10uV 50mA 0.2 Ohms 1200V 0.2 + 350

45 to 999Hz 10uV 50mA 0.2 Ohms 1200V 0.04 + 901 to 19.999kHz 10uV 50mA 0.2 Ohms 1200V 0.09 + 13520 to 99.999kHz 10uV 50mA 0.2 Ohms 1200V 0.25 + 2000100 to 500kHz 10uV 50mA 0.2 Ohms 1200V 0.45 + 3800

2-20.2V 10 to 44Hz 100uV 50mA 0.2 Ohms 1200V 0.2 + 300045 to 999Hz 100uV 50mA 0.2 Ohms 1200V 0.035 + 9001 to 19.999kHz 100uV 50mA 0.2 Ohms 1200V 0.07 + 135020 to 100kHz 100uV 50mA 0.2 Ohms 1200V 0.22 + 33000

20-202V 30 to 44Hz 1mV 20mA2 0.5 Ohms 1200V 0.06 + 20mV45 to 999Hz 1mV 15mA2 0.5 Ohms 1200V 0.04 + 7.5mV1Khz to 20kHz 1mV 2mA2

0.5 Ohms 1200V 0.09 + 40mV200-1020V3 30 to 45Hz 10mV 20mA2 0.7 Ohms 1200V 0.06 + 200mV

45 to 999Hz 10mV 15mA2 0.7 Ohms 1200V 0.04 + 75mV1Khz to 10kHz 10mV 2mA2

0.7 Ohms 1200V 0.15 + 400mV

Stability (Accuracy relative to calibration Standards)Range Frequency Frequency 90 day Rel 180 Day Rel 1 year Rel 2 year Rel

Resolution %Set uV %Set uV %Set uV %Set uV0-202mV 10 to 44Hz 1Hz 0.144 + 50 0.162 + 50 0.18 + 50 0.252 + 70

45 to 999Hz 1Hz 0.024 + 20 0.027 + 20 0.03 + 20 0.042 + 281 to 19.999kHz 1Hz 0.056 + 35 0.063 + 35 0.07 + 35 0.098 + 4920 to 99.999kHz 1Hz 0.224 + 70 0.252 + 70 0.28 + 70 0.392 + 98100 to 500kHz 1Hz 0.48 + 380 0.54 + 380 0.6 + 380 0.84 + 532

0.2-2.02V 10 to 44Hz 1Hz 0.144 + 350 0.162 + 350 0.18 + 350 0.252 + 49045 to 999Hz 1Hz 0.024 + 90 0.027 + 90 0.03 + 90 0.042 + 1261 to 19.999kHz 1Hz 0.048 + 135 0.054 + 135 0.06 + 135 0.084 + 18920 to 99.999kHz 1Hz 0.176 + 2000 0.198 + 2000 0.22 + 2000 0.308 + 2800100 to 500kHz 2Hz 0.328 + 3800 0.369 + 3800 0.41 + 3800 0.574 + 5320

2-20.2V 10 to 44Hz 1Hz 0.144 + 3000 0.162 + 3000 0.18 + 3000 0.252 + 420045 to 999Hz 1Hz 0.0224 + 900 0.025 + 900 0.03 + 900 0.0392 + 12601 to 19.999kHz 1Hz 0.04 + 1350 0.045 + 1350 0.05 + 1350 0.07 + 189020 to 100kHz 2Hz 0.144 + 33000 0.162 + 33000 0.18 + 33000 0.252 + 46200

20-202V 30 to 44Hz 1Hz 0.032 + 20mV 0.036 + 20mV 0.04 + 20mV 0.056 + 28mV45 to 999Hz 1Hz 0.024 + 7.5mV 0.027 + 7.5mV 0.03 + 7.5mV 0.042 + 11mV1kHz to 20kHz 1Hz 0.056 + 40mV 0.063 + 40mV 0.07 + 40mV 0.098 + 56mV

200-1020V3 30 to 45Hz 1Hz 0.032 + 200mV 0.036 + 200mV 0.04 + 200mV 0.056 + 280mV45 to 999Hz 1Hz 0.024 + 75mV 0.027 + 75mV 0.03 + 75mV 0.042 + 105mV1kHz to 10kHz 1Hz 0.096 + 400mV 0.108 + 400mV 0.12 + 400mV 0.168 + 560mV

All specifications apply from 10% of full scale. 5

NotesNote 1: Current limited by 50 ohms output resistance.Note 2 : Internally adjustable from 2mA to 30mA - Factory set to 20mA as standard For safety the trip is controlled by a fail-safe circuit independant of the processor which shuts the high voltage output off in the event of an overload.Note 3 : Frequency and voltage combinations are limited. See Volt-Hertz profile in Graph 3Note 4 : Specifications apply up to 10% of maximum load current. Above this level, allowance must be made for output resistance.

Note 5 : Zero or floor allowance.

Overload Protection

1 year AccuracyMax. BurdenCurrent

Due to continuous development specifications may be subject to change.3041 Extended SpecificationsACV Specifications : V1.11 Page 7

Page 8: 3041 Extended Specifications V1.11 Oct 2006 Extended Specifications CE Declartation : V1.11 Page 3 3041 MULTI-PRODUCT CALIBRATOR : DCV SPECIFICATIONS 1 Year Total Accuracy Specifications

3041 MULTI-PRODUCT CALIBRATOR : ACV SPECIFICATIONS2 Wire output / Remote sensing not available. Maximum floating voltage : 100V.Isolation : Floating or grounded selection available as standard.Specifications apply between 17°C and 27°C. Outside this range an allowance of 0.18 x 1 Year Spec. per °C should be added.

High Voltage SafetyHigh voltage output is ramped to allow instruments under test to auto-rangeStandby is automatically activated when setting voltages greater than 20V or 200V from a lower voltageStandby is automatically selected for high voltage (>20V) after 20 minutes on the same setting for frequenciesup to 5kHz or 3 mins for frequencies above 5kHz.High voltage (> 20V) output is indicated to user through an audible warning beepAn external high voltage output/standby control switch is available as an option

Worked Accuracy Calibration of 1V output at 5kHz on the 2V range at 20°C using 1 Year spec.0.09% Set (Output setting on calibrator = 1V) 0.09% of 1V = 900uVZero or Floor Allowance 135uVTotal accuracy = ± 1.035mV

Graph 3 : Volt-Hertz profile for 1000V AC range

Graph 4 : Time-Hertz profile for voltages above 20VDue to continuous development specifications may be subject to change.3041 Extended SpecificationsACV Specifications : V1.11 Page 8

Page 9: 3041 Extended Specifications V1.11 Oct 2006 Extended Specifications CE Declartation : V1.11 Page 3 3041 MULTI-PRODUCT CALIBRATOR : DCV SPECIFICATIONS 1 Year Total Accuracy Specifications

3041 MULTI-PRODUCT CALIBRATOR : ACI SPECIFICATIONS

1 Year Total Accuracy Specifications at TCal ±5°C & Range ParametersRange Frequency Resolution Maximum Overload Inductive

Burden Voltage Protection Load uA20-202uA 10Hz to 44Hz + 0.25

45Hz to 999Hz 1nA 3 Volts 5mH + 0.251kHz to 10kHz + 0.25

0.2-2.02mA 10Hz to 44Hz + 0.545Hz to 999Hz 10nA 3 Volts 150V 5mH + 0.41kHz to 10kHz + 0.7

2-20.2mA 10Hz to 44Hz + 545Hz to 999Hz 100nA 3 Volts 150V 5mH + 41kHz to 10kHz + 7

20-202mA 10Hz to 44Hz + 5045Hz to 999Hz 1uA 3 Volts 150V 5mH + 401kHz to 10kHz + 70

0.2-2.02A 10Hz to 44Hz + 50045Hz to 999Hz 10uA 3 Volts 150V 5mH + 4001kHz to 5kHz + 700

2-30.0A 30Hz to 44Hz + 500045Hz to 99Hz 100uA 2.8 Volts 150V 0.8mH + 2500100Hz to 1kHz + 4000

All specifications apply from 10% of full scale.

Settling Time: For 50% change in output: Less than 3 second from standby to within specInductive Loads: Up to 1H may be connected without additional protection.

Stability (Accuracy relative to calibration Standards)Range Frequency Frequency 90 day Rel 180 Day Rel 1 year Rel 2 year Rel

Resolution %Set uA %Set uA %Set uA %Set uA20-202uA 10Hz to 44Hz 1Hz 0.136 + 0.25 0.153 + 0.25 0.17 + 0.25 0.238 + 0.35

45Hz to 999Hz 1Hz 0.064 + 0.25 0.072 + 0.25 0.08 + 0.25 0.112 + 0.351kHz to 10kHz 1Hz 0.88 + 0.25 0.99 + 0.25 1.1 + 0.25 1.54 + 0.35

0.2-2.02mA 10Hz to 44Hz 1Hz 0.136 + 0.5 0.153 + 0.5 0.17 + 0.5 0.238 + 0.745Hz to 999Hz 1Hz 0.056 + 0.4 0.063 + 0.4 0.07 + 0.4 0.098 + 0.561kHz to 10kHz 1Hz 0.48 + 0.7 0.54 + 0.7 0.6 + 0.7 0.84 + 0.98

2mA-20.2mA 10Hz to 44Hz 1Hz 0.136 + 5 0.153 + 5 0.17 + 5 0.238 + 745Hz to 999Hz 1Hz 0.056 + 4 0.063 + 4 0.07 + 4 0.098 + 5.61kHz to 10kHz 1Hz 0.32 + 7 0.36 + 7 0.4 + 7 0.56 + 9.8

20-202mA 10Hz to 44Hz 1Hz 0.136 + 50 0.153 + 50 0.17 + 50 0.238 + 7045Hz to 999Hz 1Hz 0.056 + 40 0.063 + 40 0.07 + 40 0.098 + 561kHz to 10kHz 1Hz 0.32 + 70 0.36 + 70 0.4 + 70 0.56 + 98

200-2.02A 10Hz to 44Hz 1Hz 0.136 + 500 0.153 + 500 0.17 + 500 0.238 + 70045Hz to 999Hz 1Hz 0.064 + 400 0.072 + 400 0.08 + 400 0.112 + 5601kHz to 5kHz 1Hz 0.4 + 700 0.45 + 700 0.5 + 700 0.7 + 980

2-30.0A1 30Hz to 44Hz 1Hz 0.12 + 5000 0.135 + 5000 0.15 + 5000 0.21 + 700045Hz to 99Hz 1Hz 0.028 + 2500 0.0315 + 2500 0.04 + 2500 0.049 + 3500100Hz to 1kHz 1Hz 0.2 + 4000 0.225 + 4000 0.25 + 4000 0.35 + 5600

0.20.11.50.2

0.20.090.6

0.090.70.20.09

0.050.3

150V

1 year Accuracy%Set

0.20.10.60.2

0.5

Due to continuous development specifications may be subject to change.3041 Extended SpecificationsACI Specifications : V1.11 Page 9

Page 10: 3041 Extended Specifications V1.11 Oct 2006 Extended Specifications CE Declartation : V1.11 Page 3 3041 MULTI-PRODUCT CALIBRATOR : DCV SPECIFICATIONS 1 Year Total Accuracy Specifications

3041 MULTI-PRODUCT CALIBRATOR : ACI SPECIFICATIONS

NotesNote 1 : Temperature sensor on 30A range - microprocessor monitors & protects from overheating.Higher resistance loads allow a longer ON period. See graphs 5 and 6 for details.Note 2 : Specifications apply to loads of less than 10% of the maximum burden voltage.

Specifications apply between 17°C and 27°C. Outside this range an allowance of 0.18 x 1 Year Spec. per °C should be added.

Graph 5* : Operating time on 20A range with current into a short circuit at 20°CContinuous current in availiable below 7A output.

Graph 6* : Operating time on 20A range with current into a 0.1ohm load at 20°CContinuous current in availiable below 10.5A output.

* Note Timing is started after a minimum period of 7 minutes at zero output.Shorter periods will reduce the output time availiable.

Due to continuous development specifications may be subject to change.3041 Extended SpecificationsACI Specifications : V1.11 Page 10

Page 11: 3041 Extended Specifications V1.11 Oct 2006 Extended Specifications CE Declartation : V1.11 Page 3 3041 MULTI-PRODUCT CALIBRATOR : DCV SPECIFICATIONS 1 Year Total Accuracy Specifications

3041 MULTI-PRODUCT CALIBRATOR : RESISTANCE SPECIFICATIONSFor the highest possible accuracy and dependability of the measured value, regardless ofthe measurement technique used, the 3000 Series calibrators use passive standard resistors,the calibrated value of which is displayed when selected.

1 Year Total Accuracy Specifications at TCal ±5°C & Range ParametersRange Maximum

Current % set Ohms0Ω 0.5A - 0.005

0.1Ω 0.5A 0.01 + 0.0051Ω 0.4A 0.01 + 0.005

10Ω 0.3A 0.01 + 0.005100Ω 0.1A 0.005 + 0.0051kΩ - 0.004 + 0.04

10kΩ - 0.004 + 0.4100kΩ - 0.004 + 41MΩ∗ - 0.01 + 40

10MΩ∗ - 0.035 + 400100MΩ∗ - 0.5 + 4000

1GΩ∗- 1 + 40000

* 2-Wire onlyStability (Accuracy relative to calibration Standards)

Range 90 day Rel 180 Day Rel 1 year Rel 2 year Rel% Ohms % Ohms % Ohms % Ohms

0Ω - 0.005 - 0.005 - 0.005 -0.1Ω 0.0112 + 0.005 0.0126 + 0.005 0.014 + 0.005 0.0196 + 0.0051Ω 0.0072 + 0.005 0.0081 + 0.005 0.009 + 0.005 0.0126 + 0.005

10Ω 0.0072 + 0.005 0.0081 + 0.005 0.009 + 0.005 0.0126 + 0.005100Ω 0.0036 + 0.005 0.00405 + 0.005 0.0045 + 0.005 0.0063 + 0.0051κΩ 0.0028 + 0.04 0.00315 + 0.04 0.0035 + 0.04 0.0049 + 0.04

10κΩ 0.0028 + 0.4 0.00315 + 0.4 0.0035 + 0.4 0.0049 + 0.4100κΩ 0.0024 + 4 0.0027 + 4 0.003 + 4 0.0042 + 41ΜΩ 0.0072 + 40 0.0081 + 40 0.009 + 40 0.0126 + 40

10ΜΩ 0.024 + 400 0.027 + 400 0.03 + 400 0.042 + 400100ΜΩ 0.32 + 4000 0.36 + 4000 0.4 + 4000 0.56 + 4000

1000ΜΩ 0.72 + 40000 0.81 + 40000 0.9 + 40000 1.26 + 40000For 2-Wire connection allow 35mΩ on all resistance specifications.The 2 and 4 Wire value for each resistor is calibrated. The 2-Wire value is measured at the terminalsThe 4-Wire values are taken using the zero position to NULL the measuring system.Specifications apply between 17°C and 27°C. Outside this range an allowance of 0.18 x 1 Year Spec. per °C should be added.

Worked Examples100 ohms 4-wire output at 31°C using 2 year specification.0.0063% of setting (100Ω) 6.3mΩ5mOhm Floor 5mΩTemp coefficient (31-27) * 0.18 * 0.0045% * 100Ω 3.24mΩ

Total accuracy of calibrator only = 14.54mΩ

MaximumVoltage

1 Year Total Accuracy

-----

100V100V100V

10V50V100V100V

Due to continuous development specifications may be subject to change.3041 Extended SpecificationsDC Resistance : V1.11 Page 11

Page 12: 3041 Extended Specifications V1.11 Oct 2006 Extended Specifications CE Declartation : V1.11 Page 3 3041 MULTI-PRODUCT CALIBRATOR : DCV SPECIFICATIONS 1 Year Total Accuracy Specifications

3041 MULTI-PRODUCT CALIBRATOR : CAPACITANCE SPECIFICATIONS

For the highest possible accuracy and dependability of the measured value, regardless ofthe measurement technique used, the 3000 Series calibrators use passive standard capacitors,the calibrated value of which is displayed when selected.

General SpecificationsRange Maximum D Rs

Voltage1nF 50V 0.006 N/A10nF 50V 0.006 N/A20nF 50V 0.006 N/A50nF 50V 0.006 N/A100nF 50V 0.006 N/A1uF 30V 0.002 N/A10uF 20V 0.014 0.2Ω

Specifications apply at 1kHz. Allow 20pF for lead effects.No appreciable variation is noticable at frequencies below 1kHz.

Total AccuracyRange 90 day 180 Day 1 year 2 year

% % % %1nF 0.2 0.225 0.25 0.3510nF 0.2 0.225 0.25 0.3520nF 0.2 0.225 0.25 0.3550nF 0.2 0.225 0.25 0.35100nF 0.2 0.225 0.25 0.351uF 0.2 0.225 0.25 0.3510uF 0.48 0.54 0.6 0.84

Measurement methodsCp up to 1uFCs above 1uF

Capacitance is calibrated as value at the terminalsie. displayed value incorporates capacitance of circuit up to and including the terminals

Specifications apply between 17°C and 27°C. Outside this range an allowance of 0.18 x 1 Year Spec. per °C should be added.

Worked Accuracy Calculation of 50nF output at 20°C using 90 day specification.

0.2% of setting ( 50nF ) = 100pFZero or floor allowance ( Lead effects ) = 20pFTotal accuracy = 120pF

Due to continuous development specifications may be subject to change.3041 Extended SpecificationsCapacitance Specifications : V1.11 Page 12

Page 13: 3041 Extended Specifications V1.11 Oct 2006 Extended Specifications CE Declartation : V1.11 Page 3 3041 MULTI-PRODUCT CALIBRATOR : DCV SPECIFICATIONS 1 Year Total Accuracy Specifications

3041 MULTI-PRODUCT CALIBRATOR : SIMULATED CAPACITANCE SPECIFICATIONS

General SpecificationsRange Maximum

Voltage100uF 5V1mF 5V10mF 5V

Total AccuracyRange 90 day 180 Day 1 year 2 year

% % % %100uF 0.48 0.54 0.6 0.841mF 0.8 0.9 1 1.410mF 0.8 0.9 1 1.4

Measurement methodsCs 100uF to 1mF

Capacitance is calibrated as value at the terminalsie. displayed value incorporates capacitance of circuit up to and including the terminals

Specifications apply between 17°C and 27°C. Outside this range an allowance of 0.18 x 1 Year Spec. per °C should be added.

Worked Accuracy Calculation of 50nF output at 20°C using 90 day specification.

0.6% of setting ( 100uF ) = 600nFZero or floor allowance ( Lead effects ) = 0nFTotal accuracy = 600nF

Due to continuous development specifications may be subject to change.3041 Extended SpecificationsSimulated Capacitance : V1.11 Page 13

Page 14: 3041 Extended Specifications V1.11 Oct 2006 Extended Specifications CE Declartation : V1.11 Page 3 3041 MULTI-PRODUCT CALIBRATOR : DCV SPECIFICATIONS 1 Year Total Accuracy Specifications

3041 MULTI-PRODUCT CALIBRATOR : SIMULATED RESISTANCE SPECIFICATIONS

Total AccuracyRange 1 year 1 year

% (Rng) Zero 30R to 100R 0.03 50mΩ 101R to 1kR 0.03 50mΩ

1.01kR to 10kR 0.03 50mΩ 10.1kR to 100kR 0.03 50mΩ 101kR to 1MR 0.03 50mΩ

1.01MR to 10MR 0.03 50mΩ

Specifications apply between 17°C and 27°C.

Outside this range an allowance of 0.18 x 1 Year Spec. per °C should be added.

Worked Accuracy Calculation of 10kOhm output at 20°C using 90 day specification.

0.03% of setting ( 10kR ) = 3Ω Zero or floor allowance ( Lead effects ) = 50mΩ

Total accuracy = 3.05Ω

Due to continuous development specifications may be subject to change.3041 Extended SpecificationsSimulated Resistance : V1.11 Page 14

Page 15: 3041 Extended Specifications V1.11 Oct 2006 Extended Specifications CE Declartation : V1.11 Page 3 3041 MULTI-PRODUCT CALIBRATOR : DCV SPECIFICATIONS 1 Year Total Accuracy Specifications

3041 Series Inductance Option Specifications

General SpecificationsRange Maximum DC Q

Current Resistance1mH 30mA 7.8Ω 110mH 25mA 24Ω 2.819mH 20mA 33Ω 3.829mH 20mA 41Ω 4.750mH 20mA 54Ω 6.1100mH 20mA 78Ω 8.6

1H 10mA 260Ω 2910H 1mA 950Ω 110

All Inductance specifications ± 50uH.

Accuracy Relative to Calibration Standards Specifications

Range 90 day Rel 180 Day Rel 1 year Rel 2 year Rel% % % %

1mH 0.4 0.45 0.5 0.710mH 0.4 0.45 0.5 0.719mH 0.4 0.45 0.5 0.729mH 0.4 0.45 0.5 0.750mH 0.4 0.45 0.5 0.7100mH 0.4 0.45 0.5 0.7

1H 0.4 0.45 0.5 0.710H 0.4 0.45 0.5 0.7

Measurement methodsLs up to 1HLp from 1H to 10H

Specifications apply between 17°C and 27°C. Outside this range an allowance of 0.18 x 1 Year Spec. per °C should be added.

Due to continuous development specifications may be subject to change.3041 Extended SpecificationsInductance : V1.11 Page15

Page 16: 3041 Extended Specifications V1.11 Oct 2006 Extended Specifications CE Declartation : V1.11 Page 3 3041 MULTI-PRODUCT CALIBRATOR : DCV SPECIFICATIONS 1 Year Total Accuracy Specifications

3041 DC Power Option Specifications

General SpecificationsVoltage Range 1V to 1000V DCCurrent Range 0.3A to 30A DCOutput Terminals Voltage output from top (Black & White) terminals

300mA to 2A current output from middle 2A (Black & Red) terminals2.01A to 30A current output from bottom 30A (Blue & Yellow) terminalsNote : Indicator LEDs for both sets of terminals will illuminate to indicate DC Power mode

1 Year Accuracy Relative to Calibration standards Current Range Resolution Setting Zero

0.3A to 2A 200uA 0.015% 35uA2.01A to 30A 2mA 0.04% 350uA

1 Year Accuracy Relative to Calibration standards Voltage Range Resolution Setting Zero

20V 1uV 0.0025% 40uV200V 10uV 0.0030% 400uV1000V 100uV 0.0030% 4000uV

High Voltage SafetyHigh voltage output is ramped to allow instruments to auto rangeStandby is automatically activated when setting voltages greater than 20V or 200V from a lower voltageStandby is automatically selected for high voltage (>20V) after 20 minutes on the same settingHigh voltage (> 20V) output is indicated to user through an audible warning beepAn external high voltage output/standby control switch is available as an option

30A available as standard - external amplifier not requiredSpecifications apply between 17°C and 27°C. Outside this range an allowance of 0.18 x 1 Year Spec. per °C should be added.

Due to continuous development specifications may be subject to change.3041 Extended SpecificationsDC Power : V1.11 Page16

Page 17: 3041 Extended Specifications V1.11 Oct 2006 Extended Specifications CE Declartation : V1.11 Page 3 3041 MULTI-PRODUCT CALIBRATOR : DCV SPECIFICATIONS 1 Year Total Accuracy Specifications

3041 AC Power Option Specifications

General SpecificationsVoltage Range 1V to 1000V ACCurrent Range 0.3A to 30A ACFrequency Range 40 to 400HzOutput Terminals Voltage output from top (Black & White) terminals

300mA to 2A current output from middle 2A (Black & Red) terminals2.01A to 30A current output from bottom 30A (Blue & Yellow) terminalsNote : Indicator LEDs for both sets of terminals will illuminate to indicate AC Power mode

1 Year Accuracy Relative to Calibration standards Current Range Resolution Setting Zero

0.2A to 2A 200uA 0.1% 400uA2.01A to 30A 2mA 0.05% 2.5mA

1 Year Accuracy Relative to Calibration standards Voltage Range Resolution Setting Zero

20V 1uV 0.035% 900uV200V 10uV 0.04% 7.5mV1000V 100uV 0.04% 75mV

Frequency range 45Hz to 400HzPower Factor = 1

Phase SpecificationsPhase Angle Resolution

0° to 359.9° 0.1°

3041 calibrators automatically correct for any errors in the phase caused by inductive loading, for example when using the clamp coil adaptor.

High Voltage SafetyHigh voltage output is ramped to allow instruments to auto rangeStandby is automatically activated when setting voltages greater than 20V or 200V from a lower voltageStandby is automatically selected for high voltage (>20V) after 20 minutes on the same settingHigh voltage (> 20V) output is indicated to user through an audible warning beepAn external high voltage output/standby control switch is available as an option

30A available as standard - external amplifier not requiredSpecifications apply between 17°C and 27°C. Outside this range an allowance of 0.18 x 1 Year Spec. per °C should be added.

Due to continuous development specifications may be subject to change.3041 Extended SpecificationsAC Power : V1.11 Page17

Page 18: 3041 Extended Specifications V1.11 Oct 2006 Extended Specifications CE Declartation : V1.11 Page 3 3041 MULTI-PRODUCT CALIBRATOR : DCV SPECIFICATIONS 1 Year Total Accuracy Specifications

3041 Series Oscilloscope Calibration Option Specifications

AmplitudeRanges 2mV/Div : 5mV/Div : 10mV/Div : 20mV/Div : 50mV/Div : 100mV/Div

200mV/Div : 500mV/Div : 1V/Div : 2V/Div : 5V/Div : 10V/Div : 20V/Div : 50V/DivSequence 1, 2, 5Waveshapes Square Wave (positive going from ground) : DCFrequency 1kHzFrequency Accuracy 30ppmGraticule Height 6 Graticules Rise Time 2usFall Time 2usOutput Terminal Front BNC (Green LED indicates terminal active)

Range 90 Day Rel. 180 Day Rel. 1 Year Rel. 2 Year Rel. @ 1Mega Ohm load % ± uV % ± uV % ± uV % uV

2mV to 50V/Div 0.009 ± 20 0.01 ± 20 0.01 ± 5 0.014 20

High Voltage Safety

High voltage output is ramped to allow instruments to auto range

Auto standby is activated when passing through 20V or 200V output values

High voltage (> 20V) output is indicated to user through an audible warning beep

An external high voltage output/standby control switch is available as an option

Amplitude DeviationDeviation Range ±10%Deviation Resolution 3041 : Better than 1ppm • 2041A : Better than 10ppm

Range 90 Day Rel. 180 Day Rel. 1 Year Rel. 2 Year Rel.% ± uV % ± uV % ± uV % uV

-10% to +10% 0.008 20 0.01 ± 20 0.01 ± 20 0.014 20

TimebaseRanges 2ns/Div : 5ns/Div : 10ns/Div : 20ns/Div : 50ns/Div : 100ns/Div : 200ns/Div

500ns/Div : 1ms/Div : 2ms/Div : 5ms/Div : 10ms/Div : 20ms/Div : 50ms/Div100ms/Div : 200ms/Div : 500ms/Div : 1s/Div : 2s/Div : 5s/Div

Sequence 1, 2, 5Waveshape Comb below 100ns

Sine Wave above 100nsOscillator Internal Crystal TCXOOutput Terminal Front BNC (Green LED indicates terminal active)

Range 90 Day Rel. 180 Day Rel. 1 Year Rel. 2 Year Rel.ppm ppm ppm ppm

2ns/Div to 5s/Div 4.5 4.75 5 6

Timebase DeviationDeviation Range ±10% in 0.05% StepsDeviation Resolution Better than 0.05%

Range 90 Day Rel. 180 Day Rel. 1 Year Rel. 2 Year Rel.

-9.5% to +9.5%% % % %

0.01 0.01 0.01 0.01

Due to continuous development specifications may be subject to change.3041 Extended SpecificationsOscilloscope : V1.11 Page18

Page 19: 3041 Extended Specifications V1.11 Oct 2006 Extended Specifications CE Declartation : V1.11 Page 3 3041 MULTI-PRODUCT CALIBRATOR : DCV SPECIFICATIONS 1 Year Total Accuracy Specifications

3041 Series Oscilloscope Calibration Option Specifications

Levelled SweepSweep Range 5MHz to 600MHz Waveform Sine WaveLevelled Sweep 600mV pk-pk into 50 OhmsReference Level 50kHzOutput Terminal Front BNC (Green LED indicates terminal active)

Range 90 Day Rel. 180 Day Rel. 1 Year Rel. 2 Year Rel.db db db db

5MHz to 350MHz 0.8 0.90 1 1.45MHz to 600MHz 0.8 0.90 1 1.4

Levelled SweepFrequency Accuracy See Time markers

50kHz ReferenceAccuracy 90 Day Rel. 180 Day Rel. 1 Year Rel. 2 Year Rel.

Frequency Accuracy 27 ppm 29 ppm 30 ppm 36 ppmLevel Accuracy 0.4 % 0.45 % 0.5 % 0.7 %

Specifications apply between 17°C and 27°C. Outside this range an allowance of 0.18 x 1 Year Spec. per °C should be added.

Due to continuous development specifications may be subject to change.3041 Extended SpecificationsOscilloscope : V1.11 Page19

Page 20: 3041 Extended Specifications V1.11 Oct 2006 Extended Specifications CE Declartation : V1.11 Page 3 3041 MULTI-PRODUCT CALIBRATOR : DCV SPECIFICATIONS 1 Year Total Accuracy Specifications

3041Resistance Thermometer (PRT) Option Specifications

General SpecificationsRange Actual Max. Power Maximum Maximum

Value (Ohms) Rating (Watts) Voltage (V) Current (mA)-100°C 60.25 0.2 3.47 57.62

0°C 100.00 0.2 4.47 44.72+30°C 111.67 0.2 4.73 42.32+60°C 123.24 0.2 4.96 40.28+100°C 138.50 0.2 5.26 38.00+200°C 175.84 0.2 5.93 33.73+300°C 247.04 0.2 7.03 28.45+800°C 375.51 0.2 8.67 23.08

4-Wire connection. Allow 1mΩ on all resistance specifications.

Accuracy Relative to Calibration Standards SpecificationsRange Actual 90 day Rel 180 Day Rel 1 year Rel 2 year Rel

Value (Ohms) % % % %-100°C 60.25 0.009 0.008 0.01 0.014

0°C 100.00 0.009 0.008 0.01 0.014+30°C 111.67 0.009 0.008 0.01 0.014+60°C 123.24 0.009 0.008 0.01 0.014+100°C 138.50 0.009 0.008 0.01 0.014+200°C 175.84 0.009 0.008 0.01 0.014+300°C 247.04 0.009 0.008 0.01 0.014+800°C 375.51 0.009 0.008 0.01 0.014

Specifications apply between 17°C and 27°C.

Outside this range an allowance of 0.18 x 1 Year Spec. per °C should be added.

Due to continuous development specifications may be subject to change.3041 Extended SpecificationsPRT : V1.11 Page20

Page 21: 3041 Extended Specifications V1.11 Oct 2006 Extended Specifications CE Declartation : V1.11 Page 3 3041 MULTI-PRODUCT CALIBRATOR : DCV SPECIFICATIONS 1 Year Total Accuracy Specifications

3041 Thermocouple Simulation Adaptor Specifications

General SpecificationsAdaptor connection Connects to male 'D' type feature connector on 3041 Series front panelIndicators Incorporates an 'Active' LED mounted in the adaptor caseAdaptor Dimensions 12.5cm x 6.5cm x 2.5cmConnections 1 x 9 way male 'D' type connector

1 x Neutral (copper) industry standard thermocouple plugwith integral temperature sensor

Connection to Calibrator Via supplied 9 Way male to female serial lead (straight through connection)Sensor Type Thermistor

Sensor incorporated into thermocouple plug to eliminate any lead effectsOutput impedance 10 Ohms

Thermocouple Range 90 Day1 180 Year1 1 Year1 2 Year1

Type Rel. (°C) Rel. (°C) Rel. (°C) Rel. (°C)J -180°C to 150°C 0.04 0.05 0.05 0.07

150°C to 750°C 0.24 0.27 0.30 0.42K -140°C to 200°C 0.08 0.09 0.10 0.14

200°C to 1340°C 0.28 0.32 0.35 0.49T -250°C to 400°C 0.16 0.18 0.20 0.28R -50°C to 500°C 0.16 0.18 0.20 0.28

500°C to 1700°C 0.80 0.90 1.00 1.40S -50°C to 1200°C 0.48 0.54 0.60 0.84

1200°C to 1700°C 1.28 1.44 1.60 2.24B 0°C to 1200°C 0.08 0.09 0.10 0.14

1200°C to 1820°C 1.04 1.17 1.30 1.82N -270°C to 260°C 0.08 0.09 0.10 0.14

260°C to 1300°C 0.32 0.36 0.40 0.56E 0°C to 800°C 0.64 0.72 0.80 1.12

Note 1 : Does not include cold junction compensation errorsSpecifications apply between 17°C and 27°C. Outside this range an allowance of 0.18 x 1 Year Spec. per °C should be added.

Due to continuous development specifications may be subject to change.3041 Extended SpecificationsThermocouple : V1.11 Page21

Page 22: 3041 Extended Specifications V1.11 Oct 2006 Extended Specifications CE Declartation : V1.11 Page 3 3041 MULTI-PRODUCT CALIBRATOR : DCV SPECIFICATIONS 1 Year Total Accuracy Specifications

3041 Optical Tachometer Adaptor Specifications

General SpecificationsAdaptor connection Connects to male 'D' type feature connector on 3041 front panelIndicators Incorporates red 8mm 20° Spread Ultra bright LED mounted in the adaptor Brightness 1000mcd (wavelength 660nm)Duty cycle 20% (5:1)Adaptor Dimensions 12.5cm x 6.5cm x 2.5cmConnections 1 x 9 way male 'D' type connectorConnection to 3041 Via supplied 9 Way male to female serial lead (straight through connection)

Optical Tachometer Adaptor AccuracyRange Resolution 180 Day Rel. 1 Year Rel. 2 Year Rel.

% % %240 to 60,000 RPM 6 RPM 0.0029 0.0030 0.0036

Frequency RPMInput Simulation40 240100 600200 1200400 2400600 3600800 48001000 60002000 120004000 240008000 3200010000 60000

Specifications apply between 17°C and 27°C.

Outside this range an allowance of 0.18 x 1 Year Spec. per °C should be added.

Due to continuous development specifications may be subject to change.3041 Extended SpecificationsOptical Tachometer : V1.11 Page22

Page 23: 3041 Extended Specifications V1.11 Oct 2006 Extended Specifications CE Declartation : V1.11 Page 3 3041 MULTI-PRODUCT CALIBRATOR : DCV SPECIFICATIONS 1 Year Total Accuracy Specifications

3041 Power Supply Calibration Adapter Specifications

General SpecificationsAdaptor connection Connects to male 'D' type feature connector on 3041front panelIndicators Incorporates 1x Power indicator LED : 1 x Overload indicator LEDAdaptor Dimensions 10.5cm x 13.2cm x 5.5cmConnections 1 x 9 way male 'D' type connector

2x Voltage input 4mm terminals (Red & Black)2x Current load 4mm terminals (Red & Black)

Connection to 3041 Via supplied 9 Way male to female serial lead (straight through connection)

DC Voltage Measurement AccuracyRange Resolution 1 Year

% ± mV0 to 63 Volts 0.01V 0.020 ± 10

DC Current Load AccuracyRange Resolution 1 Year

% ± mA0 to 3 Amps 0.01A 0.050 ± 1.0

Specifications apply between 17°C and 27°C.

Outside this range an allowance of 0.18 x 1 Year Spec. per °C should be added.

Due to continuous development specifications may be subject to change.3041 Extended SpecificationsPower Supply Adaptor : V1.11 page23

Page 24: 3041 Extended Specifications V1.11 Oct 2006 Extended Specifications CE Declartation : V1.11 Page 3 3041 MULTI-PRODUCT CALIBRATOR : DCV SPECIFICATIONS 1 Year Total Accuracy Specifications

EA002 Clamp Coil Adapter Specifications

General SpecificationsAdapter Connection 4x 4mm safety sockets mounted on the rear of the unitAdapter Dimensions 28cm x 12cm x 6cmColour BlackConnection to Calibrator Via supplied 4mm to 4mm connection leads (x2) to 3041 series current terminalsCoil Configuration 2 Turn (LHS) : 10 Turn (RHS) : 50 Turn (CENTRE)Coil Type High accuracy balanced configurationMin. internal jaw dimensions 10mm (2 Turn & 10 Turn) : 25mm (50 Turn)Maximum Current 40AMaximum RMS voltage 4VFrequency Range DC to 500HzConstruction Loose wound coil (for heat dissipation) in moulded ABS enclosureDurability Fully enclosed coil for maximum protection from mechanical damageCompatibility Designed for use with Transmille 3041 calibrators and ProCal Software

2 Turn Coil Accuracy (Input 0 to 40A : Freq. DC - 30Hz to 60Hz : Effective Output 0 to 80A)90 Day Rel. 180 Day Rel. 1 Year Rel. 2 Year Rel.% A % A % A % A

Effective accuracy - Coil only (wound clamps) 0.35 + 0.008 0.35 + 0.008 0.35 + 0.008 0.35 + 0.008Effective accuracy - Coil only (hall effect clamps) 0.48 + 0.07 0.48 + 0.07 0.48 + 0.07 0.48 + 0.07Total uncertainty with 3041 (wound clamps) 0.36 + 0.028 0.36 + 0.028 0.36 + 0.028 0.38 + 0.036Total uncertainty with 3041 (hall effect clamps) 0.49 + 0.090 0.49 + 0.090 0.49 + 0.090 0.50 + 0.098

10 Turn Coil Accuracy (Input 0 to 40A : Freq. DC - 30Hz to 60Hz : Effective Output 0 to 400A)90 Day Rel. 180 Day Rel. 1 Year Rel. 2 Year Rel.% A % A % A % A

Effective accuracy - Coil only (wound clamps) 0.41 + 0.01 0.41 + 0.01 0.41 + 0.01 0.41 + 0.01Effective accuracy - Coil only (hall effect clamps) 0.59 + 0.11 0.59 + 0.11 0.59 + 0.11 0.59 + 0.11Total uncertainty with 3041 (wound clamps) 0.42 + 0.03 0.42 + 0.03 0.42 + 0.03 0.43 + 0.04Total uncertainty with 3041 (hall effect clamps) 0.60 + 0.13 0.60 + 0.13 0.60 + 0.13 0.61 + 0.14

50 Turn Coil Accuracy (Input 0 to 40A : Freq. DC - 30Hz to 60Hz : Effective Output 0 to 2000A)90 Day Rel. 180 Day Rel. 1 Year Rel. 2 Year Rel.% A % A % A % A

Effective accuracy - Coil only (wound clamps) 0.24 + 0.04 0.24 + 0.04 0.24 + 0.04 0.24 + 0.04Effective accuracy - Coil only (hall effect clamps) 0.45 + 0.42 0.45 + 0.42 0.45 + 0.42 0.45 + 0.42Total uncertainty with 3041 (wound clamps) 0.25 + 0.06 0.26 + 0.06 0.26 + 0.06 0.28 + 0.07Total uncertainty with 3041 (hall effect clamps) 0.46 + 0.44 0.46 + 0.44 0.46 + 0.44 0.47 + 0.45

Accuracy is dependant on proper alignment of the clamp meter within the coilCertain clamp meters have alignment marks which should be aligned with the centre of the coil.Certain types of clamp meter may have additional errors, or be outside the range which can be driven by the 3041 directlyUse of the Transmille EA3012 Transconductance amplifier may be used to provide additional drive.

Uncertainty calculated as the square root of the square of coil accuracy + square of calibrator accuracy using empirical data obtained for both wound & hall effect instruments from a wide range of manufacturersClamp coil adaptor is supplied complete with workstation incorporating alignment marks (size 275x295x45mm)

Due to continuous development specifications may be subject to change.3041 Extended SpecificationsEA002 : V1.11 Page24

Page 25: 3041 Extended Specifications V1.11 Oct 2006 Extended Specifications CE Declartation : V1.11 Page 3 3041 MULTI-PRODUCT CALIBRATOR : DCV SPECIFICATIONS 1 Year Total Accuracy Specifications

EA002 Clamp Coil Adapter Specifications (Cont'd)

DC ResistanceAt Coil 0.14ΩWith Connection Leads 0.18Ω

Duty Cycle10A Continuous20A 2mins on ~ 5mins off40A 30secs on ~ 5mins off

InductanceCoil Only 120uHCoil with typical clamp meter on 50 Turn coil 200uHCoil with typical clamp meter on 10 Turn coil 50uHCoil with typical clamp meter on 2 Turn coil 5uH

Specifications apply between 17°C and 27°C.

Outside this range an allowance of 0.18 x 1 Year Spec. per °C should be added.

Due to continuous development specifications may be subject to change.3041 Extended SpecificationsEA002 : V1.11 Page25