3/24/11 1 March 2011 GBM8320 - Dispositifs Médicaux Intelligents 2 Introduction ! Electrogenic cell Electrode/electrolyte interface ! Electrical double layer ! Half-cell potential ! Polarization ! Electrode equivalent circuits Biopotential electrodes ! Body surface electrodes ! Internal electrodes ! Implantable electrodes ! Electrode arrays ! Microfabricated electrodes ! Microelectrodes. GBM8320 - Dispositifs Médicaux Intelligents 3 Many types of cells in the body have the ability to undergo a transient electrical depolarization and repolarization These are either triggered by external depolarization (in the heart) or by intracellular, spontaneous mechanisms Cells that exhibit the ability to generate electrical signals are called electrogenic cells The most prominent electrogenic cells include brain cells or neurons and heart cells or cardiomyocytes. (e.g. cardiac pacemaker cells).

Transcript of • Introduction • Electrode/electrolyte interface • Biopotential electrodes

Page 1: • Introduction • Electrode/electrolyte interface • Biopotential electrodes

3/24/11

1

March 2011

GBM8320 - Dispositifs Médicaux Intelligents 2

•  Introduction !  Electrogenic cell

•  Electrode/electrolyte interface !  Electrical double layer !  Half-cell potential !  Polarization !  Electrode equivalent circuits

•  Biopotential electrodes !  Body surface electrodes !  Internal electrodes !  Implantable electrodes !  Electrode arrays !  Microfabricated electrodes !  Microelectrodes.

GBM8320 - Dispositifs Médicaux Intelligents 3

•  Many types of cells in the body have the ability to undergo a transient electrical depolarization and repolarization

•  These are either triggered by external depolarization (in the heart) or by intracellular, spontaneous mechanisms

•  Cells that exhibit the ability to generate electrical signals are called electrogenic cells

•  The most prominent electrogenic cells include brain cells or neurons and heart cells or cardiomyocytes. (e.g. cardiac pacemaker cells).

Page 2: • Introduction • Electrode/electrolyte interface • Biopotential electrodes

3/24/11

2

GBM8320 - Dispositifs Médicaux Intelligents 4

•  Electrogenic cells such as neurons contain ion channels, selectively enable the permeation of certain ions such as sodium or potassium

Jenkner et al, “Cell-based CMOS sensor …,” IEEE ISSC, V. 39, 2004.

•  In a transient change of conductivity, the overall ion flux generates an action potential, which is the elementary electrical signal in biological systems.

GBM8320 - Dispositifs Médicaux Intelligents 5

•  Electrical activity is explained by differences in ion concentrations within the body (sodium, Na+; cloride, Cl–; potassium, K+)

•  A potential difference occurs between 2 points with different ionic concentrations

•  Cell membranes at rest are more permeable to some ions (e.g. K+, Cl–) than others (e.g. Na+) –  Na+ ions are pumped out of the cells, while K+ ions are pumped in –  Due to a difference in rates of pumping, a difference in positive ion

concentration results –  A negative potential (–70 mV ) is established between the inside and

outside of the cell.

GBM8320 - Dispositifs Médicaux Intelligents 6

•  When a cell is electrically stimulated, the permeability of the cell membrane changes

–  Na+ ions rush into the cell, and K+ ions rush out –  Again, due to a difference in rates of flow, the ion concentration changes (more

positive ions inside cell than outside) –  Cellular potential becomes positive (40 mV) –  Cell is said to be depolarized.

•  After the stimulus, the permeability of the cell membrane returns to its original value, and the rest potential is restored

–  Due to unequal pumping rates of ions –  Time taken for restoration is called the refractory

period –  Cell is said to be repolarized during this time

•  The resulting variation in cellular potential with time is known as the action potential.

Page 3: • Introduction • Electrode/electrolyte interface • Biopotential electrodes

3/24/11

3

GBM8320 - Dispositifs Médicaux Intelligents 7

•  Introduction !  Electrogenic cell

•  Electrode/electrolyte interface !  Electrical double layer !  Half-cell potential !  Polarization !  Electrode equivalent circuits

•  Biopotential electrodes !  Body surface electrodes !  Internal electrodes !  Implantable electrodes !  Electrode arrays !  Microfabricated electrodes !  Microelectrodes.

GBM8320 - Dispositifs Médicaux Intelligents 8

•  Biopotential electrodes convert ionic conduction to electronic conduction so that biopotential signals can be viewed and/or stored

•  Different electrodes types include surface macroelectrodes, indwelling macroelectrodes & microelectrodes (cuff or other shapes)

•  Skin and other body tissues act as electrolytic solutions !

GBM8320 - Dispositifs Médicaux Intelligents 9

•  Charge carriers in electrode materials:

–  Metals (e.g. Pt) : electrons –  Semiconductors (e.g. n-Si) :

electrons and holes –  Solid electrolytes (e.g. lanthanum

fluoride - LaF3) : ions –  Insulators (e.g. SiO2): no charge

carriers –  Mixed conductors (e.g. IrOx) : ions

and electrons –  Solution (e.g. 1 M NaCl in H2O):

solvated ions. Inner Helmholtz plane (IHP)

Outer Helmholtz plane (OHP) Gouy-Chapman layer (GCL)

Double layer

Webster, J.G., Medical Instrumentation, Wiley, 4Ed, 2009,

Page 4: • Introduction • Electrode/electrolyte interface • Biopotential electrodes

3/24/11

4

GBM8320 - Dispositifs Médicaux Intelligents 10

•  Electrode discharges some metallic ions into electrolytic solution –  Increase in # free electrons in electrode –  Increase in # positive cations (electric charge) in solution;

OR •  Ions in solution combine with metallic electrodes

–  Decrease in # free electrons in electrode –  Decrease in # positive cations in solution.

•  As a result, a charge gradient builds up between the electrode and electrolyte and this in turn creates a potential difference.

GBM8320 - Dispositifs Médicaux Intelligents 11

!!

!+

+"+"meAAneCC

m

nGeneral Ionic Equations

•  If anion can be oxidized at the electrode to form a neutral atom, one or two electrons are given to the electrode.

•  The dominating reaction can be inferred from the following : - Current flow from electrode to electrolyte : Oxidation (Loss of e-) - Current flow from electrolyte to electrode : Reduction (Gain of e-).

a)

b) where n and m are les valences

•  If the electrode is of the same material as the cations, then this material gets oxidized and enters the electrolyte as a cation and electrons remain at the electrode & flow in the external circuit;

GBM8320 - Dispositifs Médicaux Intelligents 12

•  For both mechanisms, (Oxidation = Loss of e-, and reduction = Gain of e-), two parallel layers of oppositely charged ions are produced; i.e. the electrode double layer :

- e.g. when metal ions recombine with the electrode.

•  The excess of negative anions is replaced with positive cations in the case of metal ions discharging into solution, and Vh is then < 0.

Page 5: • Introduction • Electrode/electrolyte interface • Biopotential electrodes

3/24/11

5

GBM8320 - Dispositifs Médicaux Intelligents 13 Geddes, Principles of Applied Biomedical Instrumentation, Wiley, 1989

GBM8320 - Dispositifs Médicaux Intelligents 14

•  A characteristic potential difference established by the electrode and its surrounding electrolyte which depends on the metal, concentration of ions in solution and temperature.

•  Reason for half-cell potential : Charge separation at interface : Oxidation or reduction reactions at the electrode-electrolyte interface lead to a double-charge layer, similar to that which exists along electrically active biological cell membranes.

•  Half-cell potential cannot be measured without a second electrode. The half-cell potential of the standard hydrogen electrode has been arbitrarily set to zero. Other half cell potentials are expressed as a potential difference with this electrode.

GBM8320 - Dispositifs Médicaux Intelligents 15

* Standard Hydrogen electrode

•  Convention: The hydrogen electrode is defined as having a half-cell potential of zero.

•  The half-cell potentials of all other electrode materials is measured with respect to this hydrogen electrode.

•  Eo : Standard half-cell potential.

*

Page 6: • Introduction • Electrode/electrolyte interface • Biopotential electrodes

3/24/11

6

GBM8320 - Dispositifs Médicaux Intelligents 16

•  Electrode material is metal + salt or polymer selective membrane.

GBM8320 - Dispositifs Médicaux Intelligents 17

•  If there is a current between the electrode and electrolyte, the observed half-cell potential is often altered due to polarization, then an overpotential occurs:

Overpotential Difference between observed and zero-current half-cell potentials

Resistance Current changes resistance

of electrolyte and thus, a voltage drop results.

Concentration Changes in distribution of ions at the electrode-

electrolyte interface

Activation The activation energy barrier depends on the direction of current and

determines kinetics

Note: Polarization and impedance of the electrode are two of the most important electrode properties to consider.

V = V + V + V + E P R C A 0

Eo : Standard half-cell potential

GBM8320 - Dispositifs Médicaux Intelligents 18

!"

#$%

&+=

'(

)*

BA

DC

aaaa

nFRTEE ln0

•  When two aqueous ionic solutions of different concentration are separated by an ion-selective semi-permeable membrane, an electric potential exists across this membrane.

•  For the general oxidation-reduction reaction !++"+ neDCBA dgba

•  The Nernst equation for half-cell potential is

where Eo and E are Standard & half-cell potentials, a : Ionic activity (generally same as concentration)", and n : Number of valence electrons involved.

Note: for a metal electrode, 2 processes can occur at the electrolyte interfaces: –  A capacitive process resulting from the redistribution of charged and polar particles with no

charge-transfer between the solution and the electrode

–  A component resulting from the electron exchange between the electrode and a redox species in the solution termed faradaic process.

Page 7: • Introduction • Electrode/electrolyte interface • Biopotential electrodes

3/24/11

7

GBM8320 - Dispositifs Médicaux Intelligents 19

• Perfectly Polarizable Electrodes These are electrodes in which no actual charge crosses the electrode-electrolyte interface when a current is applied. The current across the interface is a displacement current and the electrode behaves like a capacitor.

Example : Platinum Electrode (Noble metal)

Used for recording

Used for stimulation

• Perfectly Non-Polarizable Electrode These are electrodes where current passes freely across the electrode-electrolyte interface, requiring no energy to make the transition. These electrodes see no overpotentials.

Example : Ag/AgCl electrode

GBM8320 - Dispositifs Médicaux Intelligents 20

!+ +" eAgAg!"+ #+ AgClClAg

AgCl- Cl2

Relevant ionic equations

Governing Nernst Equation

!!"

#

$$%

&+=

'Cl

sAg a

KnFRTEE ln0

Solubility product of AgCl

Fabrication of Ag/AgCl electrodes

1.  Electrolytic deposition of AgCl

2.  Sintering process forming pellet electrodes

GBM8320 - Dispositifs Médicaux Intelligents 21

What •  If a pair of electrodes is in an electrolyte and one moves with

respect to the other, a potential difference appears across the electrodes known as the motion artifact. This is a source of noise and interference in bio-potential measurements.

Why •  When the electrode moves with respect to the electrolyte, the

distribution of the double layer of charge on polarizable electrode interface changes. This changes the half-cell potential temporarily.

Note •  Motion artifact is minimal for non-polarizable electrodes.

Page 8: • Introduction • Electrode/electrolyte interface • Biopotential electrodes

3/24/11

8

GBM8320 - Dispositifs Médicaux Intelligents 22

•  Cd : Capacitance of electrode-electrolyte interface •  Rd : Resistance of electrode-electrolyte interface •  Rs : Resistance of electrode lead wire •  Ecell : Half-cell potential for electrode.

Frequency Response

Corner frequency Rd+Rs

Rs

GBM8320 - Dispositifs Médicaux Intelligents 23

Biopotential

Recording Interface Shunt

Capacitances

Interconnect Resistance

Recording Amplifier

•  Recording/Stimulating Sites: Thin-film materials such as gold, platinum, and iridium.

GBM8320 - Dispositifs Médicaux Intelligents 24

•  Extracellular action potentials have amplitude in the range of 50-500#V ! Very low-level input signals

•  Total system input-referred noise should be < 20#Vrms.

•  Biological frequency band: 100Hz-10kHz

•  System noise= Electrode noise + Preamplifier noise

•  Main source of electrode noise is thermal noise:

- RN is noise resistance (real part of probe impedance magnitude).

- !f is recording bandwidth.

fkTRV Nne != 42

Page 9: • Introduction • Electrode/electrolyte interface • Biopotential electrodes

3/24/11

9

GBM8320 - Dispositifs Médicaux Intelligents 25

•  A body-surface electrode is placed against skin, showing the total electrical equivalent circuit obtained in this situation.

•  Each circuit element on the right is at approximately the same level at which the physical process that it represents would be in the left-hand diagram.

Webster, Medical instrumentation: application and design. 3Ed, Wiley 1998.