Technical specification
FLUXUS® Cryo
TSFLUXUS_CYO_V1-0EN_Leu, 2016-10-06 1
Non-invasive flow measurement of cryogenic liquids
FLUXUS Cryo extends the application range of the ultra-sonic clamp-on flow measurement with WaveInjectorCryo to cryogenic temperatures up to -200 °C.
The patented mounting fixture thermally separates the ul-trasonic transducers from the hot or cold pipe and at thesame time ensures good acoustic contact. Therefore,FLEXIM's standard transducers are suitable for long-termoperation even at extremely cold temperatures.
Features• Non-invasive measurement without contact to the fluid
• Fluid temperature range: -200...+80 °C
• Transducers and mounting fixture can be fully integrat-ed into the insulation
• Bidirectional measurement
• Paired and factory-calibrated transducers: no zeroizing necessary, excellent zero-point stability
• High repeatability of measurement results
• Highly reliable: maintenance-free, no moving or vibrat-ing parts, dual-beam redundant measurement
• Operationally safe: no gaskets, no leakage points
• No pipework necessary for installation
ApplicationsFlow measurement of:
• LNG
• Liquid ethane
• Liquid nitrogen
• Liquid oxygen
• Other cryogenic fluids
Transmitters FLUXUS Cryo
Measurement with transducers mounted with WaveInjector Cryo
Measurement with transducers mounted with WaveInjector Cryo and with insulation
F721
F801
F809
FLUXUS® CYO Technical specification
Table of contents
TSFLUXUS_CYO_ V1-0EN_Leu, 2016-10-062
Typical measurement setup ........................................................................................................................... 3
Flow transmitter F721 ..................................................................................................................................... 4Technical data................................................................................................................................................... 4Dimensions ....................................................................................................................................................... 72 " pipe mounting kit (optional).......................................................................................................................... 8Terminal assignment......................................................................................................................................... 9
Flow transmitter ............................................................................................................................................ 10Technical data................................................................................................................................................. 10Dimensions ..................................................................................................................................................... 12Wall and 2 " pipe mounting kit......................................................................................................................... 12Terminal assignment....................................................................................................................................... 13
Flow transmitter F809 ................................................................................................................................... 15Technical data................................................................................................................................................. 15Dimensions ..................................................................................................................................................... 18Wall and 2 " pipe mounting kit......................................................................................................................... 18
Transducers................................................................................................................................................... 20Transducer selection....................................................................................................................................... 20Installation recommendation ........................................................................................................................... 20Technical data................................................................................................................................................. 21
Transducer mounting fixture ....................................................................................................................... 24Coupling materials for transducers ................................................................................................................. 25
Connection systems ..................................................................................................................................... 26Transducer cable ............................................................................................................................................ 26
Junction box .................................................................................................................................................. 27Technical data................................................................................................................................................. 27Dimensions ..................................................................................................................................................... 272 " pipe mounting kit (optional)........................................................................................................................ 28Terminal assignment....................................................................................................................................... 28
Transducer heating system (optional) ........................................................................................................ 29Continous operation ........................................................................................................................................ 29Cycle operation ............................................................................................................................................... 29Technical data................................................................................................................................................. 29
Insulation ....................................................................................................................................................... 30
TSFLUXUS_CYO_ V1-0EN_Leu, 2016-10-06 3
Technical specification FLUXUS® Cryo
Typical measurement setup
Example of a diagonal arrangement with 2 beams and transducer heating system
transmitter
JB0x JB0x
power supply
transducer heating system
FLUXUS® Cryo Technical specification
4 TSFLUXUS_CYO_ V1-0EN_Leu, 2016-10-06
Flow transmitter F721
Technical data
FLUXUS F721**-NNN*A F721**-NNN*SF721**-A20*S
design standard field device field devicewith stainless steel housing
measurementmeasurement principle transit time difference correlation principle,
automatic NoiseTrek selection for measurements with high gaseous or solid contentflow velocity 0.01...25 m/srepeatability 0.15 % of reading ±0.01 m/sfluid all acoustically conductive liquids with < 10 % gaseous or solid content in volume (transit time difference
principle)temperature compensation corresponding to the recommendations in ANSI/ASME MFC-5.1-2011
accuracy1
with standard calibration ±1.6 % of reading ±0.01 m/swith advanced calibration (optional)
±1.2 % of reading ±0.01 m/s
with field calibration2 ±0.5 % of reading ±0.01 m/sflow transmitterpower supply 100...230 V/50...60 Hz or
20...32 V DC or11...16 V DC
power consumption < 15 Wnumber of flow measuring channels
1, optional: 2
damping 0...100 s, adjustablemeasuring cycle (1 channel) 100...1000 Hzresponse time 1 s (1 channel), option: 20 mshousing material aluminum, powder coated stainless steel 316L (1.4404)degree of protection accord-ing to IEC/EN 60529
IP66 IP66
dimensions see dimensional drawingweight 5.4 kg 5.1 kgfixation wall mounting, optional: 2 " pipe mountingambient temperature -40...+60 °C (< -20 °C without operation of the display)display 128 x 64 dots, backlightmenu language English, German, French, Dutch, Spanish, Russian, Polishexplosion protection
ATEX
transmitter - F721**-A20*Szone - 2marking -
Ex nA nC ic IIC T4 GcEx tb IIIC T 120 °C DbTa -40...+60 °C
certification - IBExU11ATEX1015type of protection - gas: non sparking
dust: protection by enclosure1 for transit time difference principle, reference conditions and v > 0.15 m/s2 reference uncertainty < 0.2 %
II3GII2D
0637
TSFLUXUS_CYO_ V1-0EN_Leu, 2016-10-06 5
Technical specification FLUXUS® Cryo
measuring functionsphysical quantities volumetric flow rate, mass flow rate, flow velocity,
heat flow (if temperature inputs are installed)totalizer volume, mass, optional: heat quantitycalculation functions average, difference, sum (2 measuring channels necessary)diagnostic functions sound speed, signal amplitude, SNR, SCNR,
standard deviation of amplitudes and transit timesdata loggerloggable values all physical quantities, totalized values and diagnostic valuescapacity max. 800 000 measured valuescommunication (optional)service/diagnosis measured value transmission, parametrization of the transmitter:
- USB- Ethernet
process integration(max. 1 optional)
- BACnet MS/TP- BACnet IP (nonEx)- M-Bus (nonEx)- RS485 (ASCII sender)
including parametrization of the transmitter:- Modbus RTU- Modbus TCP (nonEx)- HART- Profibus PA- FF H1
serial data kit (optional)software FluxDiag: online diagnostics and report generation (min. Windows 7)сable USB cableoutputs (optional)
The outputs are galvanically isolated from the transmitter.number on request
switchable current outputAll switchable current outputs are switched to active or passive mode at the same time.
- range 4...20 mA (3.2...22 mA)- accuracy 0.04 % of reading ±3 μA- active output Rext < 350 Ω- passive output Uext = 8...30 V, depending on Rext, Rext < 1 kΩ
current outputcurrent output- range 0/4...20 mA- accuracy 0.1 % of reading ±15 μA- active output Rext < 500 Ω- passive output Uext = 4...24 V, depending on Rext, Rext < 1 kΩcurrent output I1 in HART mode- range 4...20 mA- active output Uint = 24 V- passive output Uext = 10...24 V
voltage outputrange 0...1 V or 0...10 Vaccuracy 0...1 V: 0.1 % of reading ±1 mV
0...10 V: 0.1 % of reading ±10 mVinternal resistance Rint = 500 Ω
frequency outputrange 0...5 kHzopen collector 24 V/4 mA, Rint = 66.5 Ω
binary outputReed relay 48 V/100 mA, Rint = 22 Ωoptorelay 26 V/100 mAbinary output as alarm output- functions limit, change of flow direction or errorbinary output as pulse output mainly for totalizing- pulse value 0.01...1000 units- pulse width optorelay: 1...1000 ms
Reed relay: 80...1000 ms3 outside of explosive atmosphere (housing cover open)
FLUXUS F721**-NNN*A F721**-NNN*SF721**-A20*S
3
3
FLUXUS® Cryo Technical specification
6 TSFLUXUS_CYO_ V1-0EN_Leu, 2016-10-06
inputs (optional)The inputs are galvanically isolated from the transmitter.
number max. 4, on requesttemperature input
type Pt100/Pt1000connection 4-wirerange -150...+560 °Cresolution 0.01 Kaccuracy ±0.01 % of reading ±0.03 K
current inputaccuracy 0.1 % of reading ±10 μAactive input Uint = 24 V, Rint = 50 Ω, Pint < 0.5 W, not short-circuit proof- range 0...20 mApassive input Rint = 50 Ω, Pint < 0.3 W- range -20...+20 mA
voltage inputrange 0...1 Vaccuracy 0.1 % of reading ±1 mVinternal resistance Rint = 1 MΩ
binary inputswitching signal 5...30 V, 1 mAfunctions - resetting the measured values
- resetting the totalizers- stopping the totalizers- activation of the measuring mode for highly dynamic flows
FLUXUS F721**-NNN*A F721**-NNN*SF721**-A20*S
TSFLUXUS_CYO_ V1-0EN_Leu, 2016-10-06 7
Technical specification FLUXUS® Cryo
Dimensions
FLUXUS F721**-****A
FLUXUS F721**-****S
in mm
94
320
thread: 6x M20 x 1.5 cable gland: max. 6x M20
255
165
.6
190.2
wall mount
Ø10
.4
Ø8.5
255
240
cable gland: max. 6x M20 with flat gasket and counter nut
280
87
fixing holes for wall mounting
320
FLUXUS® Cryo Technical specification
8 TSFLUXUS_CYO_ V1-0EN_Leu, 2016-10-06
2 " pipe mounting kit (optional)
FLUXUS F721**-****A
FLUXUS F721**-****S
TSFLUXUS_CYO_ V1-0EN_Leu, 2016-10-06 9
Technical specification FLUXUS® Cryo
Terminal assignment
1 The number, type and terminal assignment of the outputs and inputs will be customized.
FLUXUS F721
power supplyterminal strip KL4
terminal connection (AC) connection (DC)PE earth earthN(-) neutral -L(+) phase +
transducersterminal strip KL11, KL12
extension cable (transducers *****8*, ****LI*, *****52)transducer cable (transducers *****8*, ****LI*)
transducer cable (transducers *****52)
measuring channel A measuring channel B measuring channel A measuring channel Bterminal connection terminal connection terminal connection
AV signal BV signal X_AV X_BV SMB connectorAVS shield BVS shield X_AR X_BR SMB connectorARS shield BRS shieldAR signal BR signal
outputs1 RS485, Modbus RTU, BACnet MS/TP, Profibus, FF H1, M-Bus (optional)
terminal strip KL9, KL10 terminal strip KL9, KL10terminal connection terminal connection
P1+...P4+, P1-...P4- current output, voltage output, frequency output, binary output (Reed relay)
A+ signal +
P5a...P7a, P5b...P7b binary output (optorelay) B- signal -S shield
analog inputs1
terminal strip KL7, KL8temperature probe passive current source active current source
terminal direct connection connection with extension cable
connection of anactive input
connection of a passive input
T1a...T4a red red not connected not connectedT1A...T4A red/blue grey - +T1b...T4b white/blue blue + not connectedT1B...T4B white white not connected -S1, S3 shield shield not connected not connected
binary inputs1
terminal strip KL9, KL10terminal
P1+...P2+, P1-...P2-
A+B-P1+P2+P3+P4+P5aP6aP7a
S S P1-P2-P3-P4-P5bP6bP7bX2 X3
X_AV
KL11
KL12
KL7
KL8
KL9
KL10 KL4
X_AR X_BV X_BR
T1A
T1B
S1T2A
T2B
T3A
T3B
S3
T4B
T4A
T1a
T1b
S1T2a
T2b
T3a
T3b
S3
T4b
T4a
N(-)
PE L(+)
AV AVS
AGN
ARS
AR
BVBVS
BGN
BRS
BR
F721**-****S F721**-****Aequipotential bonding terminal
FLUXUS® Cryo Technical specification
10 TSFLUXUS_CYO_ V1-0EN_Leu, 2016-10-06
Flow transmitter
Technical data
FLUXUS F801**-A10 F801C24model code F801**-A10****-*A
F801**-A10****-*PF801**-A10****-FF F801**-A1B
design explosion proof offshore device
measurementmeasurement principle transit time difference correlation principle,
automatic NoiseTrek selection for measurements with high gaseous or solid contentflow velocity 0.01...25 m/srepeatability 0.15 % of reading ±0.01 m/sfluid all acoustically conductive liquids with < 10 % gaseous or solid content in volume (transit time difference
principle)temperature compensation corresponding to the recommendations in ANSI/ASME MFC-5.1-2011
accuracy1
with standard calibration ±1.6 % of reading ±0.01 m/swith advanced calibration (optional)
±1.2 % of reading ±0.01 m/s
with field calibration2 ±0.5 % of reading ±0.01 m/sflow transmitterpower supply 100...230 V/50...60 Hz
or20...32 V DCor on request:11...16 V DC
24 V DC ±10 %
power consumption < 8 W < 4 Wnumber of flow measuring channels
1, optional: 2
damping 0...100 s, adjustablemeasuring cycle (1 channel) 100...1000 Hzresponse time 1 s (1 channel), option: 70 mshousing material stainless steel 316/316L (1.4401, 1.4404, 1.4432)degree of protection accord-ing to IEC/EN 60529
IP66
dimensions see dimensional drawingweight 6.6 kgfixation wall mounting, 2 " pipe mountingambient temperature -20...+60 °C -20...+50 °Cdisplay 2 x 16 characters, dot matrix, backlightmenu language English, German, French, Dutch, Spanishexplosion protection
ATEX/IECEx
zone 1 1marking
Ex d e IIC T6 GbEx tb IIIC T 100 °C DbTa -20...+60 °C
0637 II2G Ex d e [ib] IIC T4 GbTa -20...+50 °C
certification ATEX IBExU05ATEX1078 IBExU05ATEX1078certification IECEx IECEx IBE 12.0020 -type of protection electronics compartment: flameproof enclosure
connection compartment: increased safety
electronics compartment: flame-proof enclosure
connection compartment: increa-sed safety
output circuits: intrinsic safetyintrinsic safety parameters - Um = 250 V AC
intrinsically safe outputs: Ui = 28.2 VPi = 0.76 WLi, Ci negligible
1 for transit time difference principle, reference conditions and v > 0.15 m/s2 reference uncertainty < 0.2 %
II2GII2D
0637
TSFLUXUS_CYO_ V1-0EN_Leu, 2016-10-06 11
Technical specification FLUXUS® Cryo
measuring functionsphysical quantities volumetric flow rate, mass flow rate, flow velocitytotalizer volume, masscalculation functions average, difference, sum
(2 measuring channels necessary)diagnostic functions sound speed, signal amplitude, SNR, SCNR,
standard deviation of amplitudes and transit timesdata loggerloggable values all physical quantities, totalized values and diagnostic valuescapacity > 100 000 measured valuescommunication
interface - process integration (optional):RS485 (sender) or Modbus RTU or HART
- diagnosis: RS2323
- diagnosis: RS2323 - diagnosis: RS2323
serial data kit (optional)
software (all Windows™ ver-sions)
- FluxData: download of measurement data, graphical presentation,conversion to other formats (e.g. for Excel™)
- FluxDiag (optional): online diagnostics and report generation
- FluxSubstanceLoader: upload of fluid data sets
сable RS2323
adapter RS232 - USB3
outputs (optional)The outputs are galvanically isolated from the transmitter.
number current output: 1...2andbinary output (open collector): 1...2
or
current output: 1...2andbinary output (open collector): 1andbinary output (Reed relays): 1
frequency output: 1andbinary output (open collector): 1
current output: 1andbinary output (open collector): 1
current outputcurrent output I1, I2- range 0/4...20 mA - 4...20 mA- accuracy 0.1 % of reading ±15 μA - 0.1 % of reading ±15 μA- active output - -
- passive output - Uext = 4...28.2 V,depending on RextRext < 1 kΩ
intrinsic safetycurrent output I1 in HART mode- range 4...20 mA - -- passive output Uext = 10...24 V - -
frequency outputrange - 0...5 kHz -open collector - 30 V/100 mA
Ioff = 0.8 mA
optional: 8.2 VDIN EN 60947-5-6 (NAMUR)
-
binary outputReed relay 48 V/100 mA - -open collector 24 V/4 mA 30 V/100 mA
Ioff = 0.8 mA24 V/4 mAintrinsic safety
binary output as alarm output- functions limit, change of flow direction or erroropen collector as pulseoutput
mainly for totalizing
- pulse value 0.01...1000 units- pulse width 1...1000 ms
3 connection of the interface RS232 outside of explosive atmosphere (housing cover open)
FLUXUS F801**-A10 F801C24
Rext < 500 Ω
Uext = 4...26.4 V,depending on RextRext < 1 kΩ
FLUXUS® Cryo Technical specification
12 TSFLUXUS_CYO_ V1-0EN_Leu, 2016-10-06
Dimensions
Wall and 2 " pipe mounting kit
FLUXUS F801
in mm
FLUXUS F801
277
Ø 1
32
189
178
thread: 4x M25cable gland: max. 4x M20
TSFLUXUS_CYO_ V1-0EN_Leu, 2016-10-06 13
Technical specification FLUXUS® Cryo
Terminal assignment
FLUXUS F801**-A10****-*AFLUXUS F801**-A10****-*P
power supply
AC DCterminal connection terminal connectionPE earth PE earthN neutral L- -L1 phase L+ +
transducers
measuring channel A measuring channel Bterminal connection terminal connectionAV transducer , signal BV transducer , signalAVS transducer , internal shield BVS transducer , internal shieldARS transducer , internal shield BRS transducer , internal shieldAR transducer , signal BR transducer , signalcable gland external shield cable gland external shield
outputs
F801**-A10****-*A F801**-A10****-*Pterminal connection terminal connection1(-), 2(+) active current output I1 1(+), 2(-) passive current output I13(-), 4(+) active current output I2 (optional) 3(+), 4(-) passive current output I2 (optional)5(-), 6(+) binary output B1 (open collector)7(-), 8(+) binary output B2 (open collector, optional)9(a), 10(b) binary output B1 (open collector, Reed relay, optional)11(a), 12(b) binary output B2 (open collector, Reed relay, optional)13(B-), 14(A+),15 (shield)
communication interface
power supply
outputs
transducers
equipotential bonding terminal
FLUXUS® Cryo Technical specification
14 TSFLUXUS_CYO_ V1-0EN_Leu, 2016-10-06
FLUXUS F801C24FLUXUS F801**-A10****-FF
power supply
AC DC(F801**-A10****-FF)terminal connection terminal connectionPE earth PE earthN neutral L- -L1 phase L+ +
transducers
measuring channel A measuring channel Bterminal connection terminal connectionAV transducer , signal BV transducer , signalAVS transducer , internal shield BVS transducer , internal shieldARS transducer , internal shield BRS transducer , internal shieldAR transducer , signal BR transducer , signalS not connected S not connectedcable gland external shield cable gland external shield
outputs
F801C24 F801**-A10****-FFcolour of terminals blue (intrinsic safety) green
terminal connection1(-), 2(+) current output I1 frequency output F15(-), 6(+) binary output B1 (open collector) binary output B1 (open collector)
power supply
outputs
transducers
equipotential bonding terminal
partition(F801C24)
TSFLUXUS_CYO_ V1-0EN_Leu, 2016-10-06 15
Technical specification FLUXUS® Cryo
Flow transmitter F809
Technical data
FLUXUS F809**-A10 F809**-A1Adesign explosion proof field device
1 or 2 measuring channelszone 1 zone 1 (intrinsically safe current output)
measurementmeasurement principle transit time difference correlation principle,
automatic NoiseTrek selection for measurements with high gaseous or solid contentflow velocity 0.01...25 m/srepeatability 0.15 % of reading ±0.01 m/sfluid all acoustically conductive liquids with < 10 % gaseous or solid content in volume (transit time difference
principle)temperature compensation corresponding to the recommendations in ANSI/ASME MFC-5.1-2011accuracy1
with standard calibration ±1.6 % of reading ±0.01 m/swith advanced calibration (optional)
±1.2 % of reading ±0.01 m/s
with field calibration2 ±0.5 % of reading ±0.01 m/sflow transmitterpower supply 100...230 V/50...60 Hz or
20...32 V DC20...32 V DC
power consumption < 8 Wnumber of flow measuring channels
1, optional: 2
damping 0...100 s, adjustablemeasuring cycle (1 channel) 100...1000 Hzresponse time 1 s, option: 70 mshousing material cast aluminum, special offshore coatingdegree of protection accord-ing to IEC/EN 60529
IP66
dimensions see dimensional drawingweight 7.1 kgfixation wall mounting, 2 " pipe mountingoperating temperature -30...+60 °C (< -20 °C without operation of the display)display 2 x 16 characters, dot matrix, backlightmenu language English, German, French, Dutch, Spanish
1 for transit time difference principle, reference conditions and v > 0.15 m/s2 reference uncertainty < 0.2 %
FLUXUS® Cryo Technical specification
16 TSFLUXUS_CYO_ V1-0EN_Leu, 2016-10-06
explosion protection
ATEX/IECEx
zone 1marking
Ex db e IIC T6 GbEx tb IIIC T 100 °C DbTa -40...+60 °C
Ex db e ia IIC T6 GbEx tb ia IIIC T 100 °C DbTa -40...+60 °C
certification ATEX IBExU11ATEX1022 X IBExU11ATEX1022 X
certification IECEx IECEx IBE 11.0006X IECEx IBE 11.0006Xtype of protection gas: electronics compartment: flameproof enclo-
sure, connection compartment: increased safetydust: protection by enclosure
gas: electronics compartment: flameproof enclo-sure, connection compartment: increased safetydust: protection by enclosure
intrinsic safety parameters - Um = 250 VUi = 30 V DCIi = 100 mAPi = 0.75 WCi = 3 nF
measuring functionsphysical quantities volumetric flow rate, mass flow rate, flow velocitytotalizer volume, masscalculation functions average, difference, sum (2 measuring channels necessary)diagnostic functions sound speed, signal amplitude, SNR, SCNR, standard deviation of amplitudes and transit timesdata loggerloggable values all physical quantities, totalized values and diagnostic valuescapacity > 100 000 measured valuescommunication
interface - process integration (optional): RS485 (sender) or Modbus RTU or HART
- diagnosis: RS2323
- diagnosis: RS2323
serial data kit (optional)software (all Windows™ versions)
- FluxData: download of measurement data, graphical presentation,conversion to other formats (e.g. for Excel™)
- FluxDiag (optional): online diagnostics and report generation
- FluxSubstanceLoader: upload of fluid data setsсable RS2323
adapter RS232 - USB3
3 connection of the interface RS232 outside of explosive atmosphere (housing cover open)
FLUXUS F809**-A10 F809**-A1A
II2GII2D
0637 II2GII2D
0637
TSFLUXUS_CYO_ V1-0EN_Leu, 2016-10-06 17
Technical specification FLUXUS® Cryo
outputsThe outputs are galvanically isolated from the transmitter.
number max. 4 1current output
number max. 2 -current output I1, I2 -- range 0/4...20 mA -- accuracy 0.1 % of reading ±15 μA -- active output Rext < 500 Ω -- passive output Uext = 4...26.4 V, depending on Rext, Rext < 1 kΩ -current output I1 in HART mode- range 4...20 mA -- passive output Uext = 7...30 V DC -- active output Uint = 24 V -
current output (intrinsic safety)number - 1current output I1- range - 4...20 mA- accuracy - 0.04 % of reading ±3 μA- passive output - Uext = 7...30 V, depending on Rext, Rext < 1 kΩcurrent output I1 in HART mode- range - 4...20 mA- passive output - Uext = 7...30 V DC
frequency outputnumber max. 1 -range 0...5 kHz -
open collector 30 V/100 mA
or
8.2 V DIN EN 60947-5-6 (NAMUR)
or
24 V/4 mA (on request)
-
binary outputnumber max. 2 -Reed relay 48 V/100 mA -
open collector 24 V/4 mA
optional: 30 V/100 mA or8.2 V DIN EN 60947-5-6 (NAMUR)
-
binary output as alarm output- functions limit, change of flow direction or error -binary output as pulse output mainly for totalizing- pulse value 0.01...1000 units -- pulse width 80...1000 ms -
FLUXUS F809**-A10 F809**-A1A
FLUXUS® Cryo Technical specification
18 TSFLUXUS_CYO_ V1-0EN_Leu, 2016-10-06
Dimensions
Wall and 2 " pipe mounting kit
FLUXUS F809
in mm
243
174
136
286
material: stainless steel 304 (1.4301)
F809**-A1: thread, cable gland: 4x M20 x 1.5
F809**-A1: thread, cable gland: 2x M20 x 1.5
TSFLUXUS_CYO_ V1-0EN_Leu, 2016-10-06 19
Technical specification FLUXUS® Cryo
Terminal assignment
power supply
transducers
outputs
FLUXUS F809
AC DCterminal connection terminal connectionL phase L+ +N neutral L- -PE earth PE earth
measuring channel A measuring channel Bterminal connection terminal connectionAV transducer , signal BV transducer , signalAVS transducer , internal shield BVS transducer , internal shieldARS transducer , internal shield BRS transducer , internal shieldAR transducer , signal BR transducer , signalcable gland or equipo-tential bonding termi-nal (transducers)
external shield cable gland or equipo-tential bonding termi-nal (transducers)
external shield
terminal connection1(-), 2(+) current output I1 frequency output F13(-), 4(+) current output I25(-), 6(+) binary output B1 (open collector)7(-), 8(+) binary output B2 (open collector)9(-), 10(+) binary output B1 (Reed relay) binary output B1 (open collector)A+, B-, S communication interface
L N PE(+) (-)
AR ARS AVAVS BR BRS BVBVS
equipotential bonding terminal
lower housing,front view
upper housing,back view
equipotential bonding terminal
power supply
transducers
outputs
equipotential bonding terminal (transducers)
FLUXUS® Cryo Technical specification
20 TSFLUXUS_CYO_ V1-0EN_Leu, 2016-10-06
Transducers
Transducer selection
• for LNG, others on request
• outer pipe diameter 40...70 mm, > 1000 mm on request
Installation recommendation
transducer order code
FSG 1000 4000 6500
FSK 180 200 1500
FSM 70 100 200
FSP 65 100
5 10 50 100 500 1000 5000inner pipe diameter [mm]
transducer order code
FSG 11
FSK 5
FSM 2
FSP 1
5 10 15 20 25 30pipe wall thickness [mm]
recommended possible
inner pipe diameter measurement arrangement number of sound paths min. number of measuring channels
mm65...100 diagonal arrangement 3 1>100...180 reflection arrangement 2 1>180 diagonal arrangement 1 2
TSFLUXUS_CYO_ V1-0EN_Leu, 2016-10-06 21
Technical specification FLUXUS® Cryo
Technical data
Shear wave transducers (zone 1)
technical type CDG1N81 CDK1N81 CDM2N81 CDP2N81order code FSG-NA1TS
FSG-NA1TS/OSFSK-NA1TSFSK-NA1TS/OS
FSM-NA1TSFSM-NA1TS/OS
FSP-NA1TSFSP-NA1TS/OS
transducer frequency MHz 0.2 0.5 1 2inner pipe diameter see Transducer selectionpipe wall thicknessmin. mm 11 5 2 1materialhousing PEEK with stainless steel
cap 304 (1.4301), option OS: 316L (1.4404)
PEEK with stainless steel cap 304 (1.4301), option OS: 316L (1.4404)
PEEK with stainless steel cap 304 (1.4301), option OS: 316L (1.4404)
PEEK with stainless steel cap 304 (1.4301), option OS: 316L (1.4404)
contact surface PEEK PEEK PEEK PEEKdegree of protection according to IEC/EN 60529
IP65 IP66 IP66 IP66
transducer cabletype 1699 1699 1699 1699length m 5 5 4 4dimensionslength l mm 129.5 126.5 64 64width b mm 51 51 32 32height h mm 67 67.5 40.5 40.5dimensional drawing
weight (without cable) kg 0.47 kg 0.36 kg 0.066 kg 0.066 kgambient temperaturemin. °C -40 -40 -40 -40max. °C +130 +130 +130 +130temperature compen-sation
x x x x
explosion protection
ATEX/IECEx
categoryEPLzone
gas: 2G dust: 2DGb Db1 21
gas: 2G dust: 2DGb Db1 21
gas: 2G dust: 2DGb Db1 21
gas: 2G dust: 2DGb Db1 21
explosion protection temperature (pipe surface)min. °C -55 -55 -55 -55max. °C +180 +180 +180 +180marking
Ex e q IIC T6...T3 GbEx tb IIIC TX Db
Ex e q IIC T6...T3 GbEx tb IIIC TX Db
Ex e q IIC T6...T3 GbEx tb IIIC TX Db
Ex e q IIC T6...T3 GbEx tb IIIC TX Db
certification ATEX IBExU07ATEX1168 X IBExU07ATEX1168 X IBExU07ATEX1168 X IBExU07ATEX1168 Xcertification IECEx IECEx IBE 08.0007X IECEx IBE 08.0007X IECEx IBE 08.0007X IECEx IBE 08.0007Xtype of protection gas: increased safety,
powder fillingdust: protection by enclosure
gas: increased safety, powder fillingdust: protection by enclosure
gas: increased safety, powder fillingdust: protection by enclosure
gas: increased safety, powder fillingdust: protection by enclosure
transducer mounting fixture necessary
x x x x
remark on request
l
hb
l
hb
l
hb
l
hb
II2GII2D
0637 II2GII2D
0637 II2GII2D
0637 II2GII2D
0637
FLUXUS® Cryo Technical specification
22 TSFLUXUS_CYO_ V1-0EN_Leu, 2016-10-06
Shear wave transducers (zone 2 or not explosion proof)
technical type CDG1N52 CLG1N52 CDK1N52 CLK1N52order code FSG-NA2TS
FSG-NA2TS/OSFSG-NNNTSFSG-NNNTS/OS
FSG-NA2TS/LCFSG-NA2TS/LC/OSFSG-NNNTS/LCFSG-NNNTS/LC/OS
FSK-NA2TSFSK-NA2TS/OSFSK-NNNTSFSK-NNNTS/OS
FSK-NA2TS/LCFSK-NA2TS/LC/OSFSK-NNNTS/LCFSK-NNNTS/LC/OS
transducer frequency MHz 0.2 0.2 0.5 0.5inner pipe diameter see Transducer selectionpipe wall thicknessmin. mm 11 11 5 5materialhousing PEEK with stainless steel
cap 304 (1.4301), option OS: 316L (1.4404)
PEEK with stainless steel cap 304 (1.4301), option OS: 316L (1.4404)
PEEK with stainless steel cap 304 (1.4301), option OS: 316L (1.4404)
PEEK with stainless steel cap 304 (1.4301), option OS: 316L (1.4404)
contact surface PEEK PEEK PEEK PEEKdegree of protection according to IEC/EN 60529
IP67 IP67 IP67 IP67
transducer cabletype 1699 1699 1699 1699length m 5 9 5 9dimensionslength l mm 129.5 129.5 126.5 126.5width b mm 51 51 51 51height h mm 67 67 67.5 67.5dimensional drawing
weight (without cable) kg 0.47 kg 0.47 kg 0.36 kg 0.36 kgambient temperaturemin. °C -40 -40 -40 -40max. °C +130 +130 +130 +130temperature compen-sation
x x x x
explosion protection
ATEX/IECEx
order code FSG-NA2TSFSG-NA2TS/OS
FSG-NA2TS/LCFSG-NA2TS/LC/OS
FSK-NA2TSFSK-NA2TS/OS
FSK-NA2TS/LCFSK-NA2TS/LC/OS
categoryEPLzone
gas: 3G dust: 2DGc Db2 21
gas: 3G dust: 2DGc Db2 21
gas: 3G dust: 2DGc Db2 21
gas: 3G dust: 2DGc Db2 21
explosion protection temperature (pipe surface)min. °C -55 -55 -55 -55max. °C gas: +190, dust: +180 gas: +190, dust: +180 gas: +190, dust: +180 gas: +190, dust: +180marking
Ex nA IIC T6...T2 GcEx tb IIIC TX Db
Ex nA IIC T6...T2 GcEx tb IIIC TX Db
Ex nA IIC T6...T2 GcEx tb IIIC TX Db
Ex nA IIC T6...T2 GcEx tb IIIC TX Db
certification ATEX IBExU10ATEX1163 X IBExU10ATEX1163 X IBExU10ATEX1163 X IBExU10ATEX1163 Xcertification IECEx IECEx IBE 12.0005X IECEx IBE 12.0005X IECEx IBE 12.0005X IECEx IBE 12.0005Xtype of protection gas: non sparking
dust: protection by enclosure
gas: non sparkingdust: protection by enclosure
gas: non sparkingdust: protection by enclosure
gas: non sparkingdust: protection by enclosure
transducer mounting fixture necessary
x x x x
l
hb
l
hb
l
hb
l
hb
II3GII2D
0637 II3GII2D
0637 II3GII2D
0637 II3GII2D
0637
TSFLUXUS_CYO_ V1-0EN_Leu, 2016-10-06 23
Technical specification FLUXUS® Cryo
Shear wave transducers (zone 2 or not explosion proof)
technical type CDM2N52 CDP2N52order code FSM-NA2TS
FSM-NA2TS/OSFSM-NNNTSFSM-NNNTS/OS
FSP-NA2TSFSP-NA2TS/OSFSP-NNNTSFSP-NNNTS/OS
transducer frequency MHz 1 2inner pipe diameter see Transducer selectionpipe wall thicknessmin. mm 2 1materialhousing PEEK with stainless steel
cap 304 (1.4301), option OS: 316L (1.4404)
PEEK with stainless steel cap 304 (1.4301), option OS: 316L (1.4404)
contact surface PEEK PEEKdegree of protection according to IEC/EN 60529
IP67 IP67
transducer cabletype 1699 1699length m 4 4dimensionslength l mm 64 64width b mm 32 32height h mm 40.5 40.5dimensional drawing
weight (without cable) kg 0.066 kg 0.066 kgambient temperaturemin. °C -40 -40max. °C +130 +130temperature compen-sation
x x
explosion protection
ATEX/IECEx
order code FSM-NA2TSFSM-NA2TS/OS
FSP-NA2TSFSP-NA2TS/OS
categoryEPLzone
gas: 3G dust: 2DGc Db2 21
gas: 3G dust: 2DGc Db2 21
explosion protection temperature (pipe surface)min. °C -55 -55max. °C gas: +190, dust: +180 gas: +190, dust: +180marking
Ex nA IIC T6...T2 GcEx tb IIIC TX Db
Ex nA IIC T6...T2 GcEx tb IIIC TX Db
certification ATEX IBExU10ATEX1163 X IBExU10ATEX1163 Xcertification IECEx IECEx IBE 12.0005X IECEx IBE 12.0005Xtype of protection gas: non sparking
dust: protection by enclosure
gas: non sparkingdust: protection by enclosure
transducer mounting fixture necessary
x x
l
hb
l
hb
II3GII2D
0637 II3GII2D
0637
FLUXUS® Cryo Technical specification
24 TSFLUXUS_CYO_ V1-0EN_Leu, 2016-10-06
Transducer mounting fixtureOrder code
1...6 7 8 9 10 11...14 15 16 17 no. of character
Wa
veIn
ject
or
tran
sduc
er
- mea
sure
men
t ar
rang
emen
t
size - fixat
ion
oute
r pi
pe
diam
ete
r1
- coup
ling
foil
tool / optio
n
description
WI-CYO type
K shear wave transducers with transducer frequency G, K
M shear wave transducers with transducer frequency M, P
D reflection arrangement or diagonal arran-gement
L large
C chains
017 70...170 mm
038 80...380 mm
056 350...560 mm
085 560...850 mm
100 600...1000 mm
D coupling foil -200...+80 °C
A automatic WI tool
M mechanical WI tool (pipe planer)
O WIT-R tool 110 V
R WIT-R tool 230 V
N without tool
B cryo insulation boxes for measuring chan-nel (without transducer heating system)
example
WI-CYO M - D L - C 017 - D M / L WaveInjector Cryo for shear wave trans-ducers with transducer frequency M, with chains, coupling foil and mechanical WI tool, transducer heating system (240 V)
- - -
1 outer pipe diameter > 1000 mm on request
WaveInjector WI-CYO
dimensions:
- length: 2 . l + lcpl = 273 mmlcp = depending on application
- width:outer pipe diameter + 32 mm(min. 203 mm)
- height:outer pipe diameter + 2 . hh = 285 mm
material: stainless steel 304 (1.4301)
lcp
h
l
TSFLUXUS_CYO_ V1-0EN_Leu, 2016-10-06 25
Technical specification FLUXUS® Cryo
Coupling materials for transducers
Technical data
type ambient temperature material position°C
coupling foil type SI -40...+80 silicone coupling plate - transducer
coupling foil type D -200...+80 indium pipe - coupling plate
FLUXUS® Cryo Technical specification
26 TSFLUXUS_CYO_ V1-0EN_Leu, 2016-10-06
Connection systems
Transducer cable
Technical data
connection system TSconnection with extension cable direct connection transducers
technical type*****8*
*****52
transducer frequency(3d character of transducer
order code)
F, G, H, K M, P Q S
TS
x l x l x l x lcable length m 5 ≤ 300 4 ≤ 300 3 ≤ 90 2 ≤ 40
cable length (option LC) m 9 ≤ 300 - - - - - -
x - transducer cable lengthl - max. length of extension cable
transducer cable extension cabletype 1699 2615 5245ambient temperature °C -55...+200 -30...+70 -30...+70properties halogen free
fire propagation test according to IEC 60332-1
combustion test according to IEC 60754-2
halogen free
fire propagation test according to IEC 60332-1
combustion test according to IEC 60754-2
cable jacketmaterial PTFE PUR PURouter diameter mm 2.9 12 12thickness mm 0.3 2 2colour brown black blackshield x x xsheathmaterial stainless steel 304 (1.4301) - steel wire braid with copolymer
sheathoption OS: 316Ti (1.4571)outer diameter mm 8 - 15.6
tran
smitt
er
x
l
JB01
tran
smitt
er
x
tran
smitt
er
x
l
JB02, JB03
tra
nsm
itter
x
TSFLUXUS_CYO_ V1-0EN_Leu, 2016-10-06 27
Technical specification FLUXUS® Cryo
Junction box
Technical data
Dimensions
technical type JB01S4E3M JB02 JB03dimensions see dimensional drawing see dimensional drawing see dimensional drawingweight kg 1.2 kg 1.2 kg 1.2 kgfixation wall mounting, optional: 2 " pipe
mountingwall mounting, optional: 2 " pipe mounting
wall mounting, optional: 2 " pipe mounting
materialhousing stainless steel 316L (1.4404) stainless steel 316L (1.4404) stainless steel 316L (1.4404)gasket silicone silicone siliconedegree of protection according to IEC/EN 60529
IP67 IP67 IP67
ambient temperaturemin. °C -40 -40 -40max. °C +80 +80 +80explosion protection
ATEX/IECEx
zone 1 2 -marking
Ex e mb IIC (T6)...T4 GbEx tb IIIC T 100 °C DbTa -40...+(70)80 °C
II3G Ex nA IIC (T6)...T4 GcII3D Ex tc IIIC T 100 °C DcTa -40...+(70)80 °C
-
certification ATEX IBExU06ATEX1161 - -certification IECEx IECEx IBE 08.0006 - -type of protection gas:
• increased safety• decoupled network:
encapsulation dust: protection by enclosure
gas: non sparkingdust: protection by enclosure
-
in mm
II2GII2D
0637
174
1192
wall mounting holder
70163.5
156
Ø 9
thread: 3x M20 x 1.5cable gland: max. 2x M20
FLUXUS® Cryo Technical specification
28 TSFLUXUS_CYO_ V1-0EN_Leu, 2016-10-06
2 " pipe mounting kit (optional)
Terminal assignment
JB01
JB02, JB03
transducers
terminal strip KL1terminal connection
V transducer , signalVS transducer , internal shieldRS transducer , internal shieldR transducer , signalcable gland external shield
extension cable
terminal strip KL2terminal connection
TV signalTVS internal shieldTRS internal shieldTR signalshield terminal external shield
shield terminal
equipotential bonding terminal(at wall mounting holder)
transducers
terminal connectionXV transducer , SMB connectorXR transducer , SMB connectorcable gland external shield
extension cable
terminal strip KL2terminal connection
TV signalTVS internal shieldTRS internal shieldTR signalshield terminal external shield
shield terminal
equipotential bonding terminal(at wall mounting holder)
TSFLUXUS_CYO_ V1-0EN_Leu, 2016-10-06 29
Technical specification FLUXUS® Cryo
Transducer heating system (optional)
Continous operation
fluid temperature < -40 °C: transducer heating system required
Cycle operation
operating time in cycle operation > 12 h: transducer heating system requiredoperating time in cycle operation < 12 h and rest period ≥ 2x operating time: see table below
Technical data
Cable junction box PSB - heating element
fluid temperature
ambient temperature
°C °C-30 -20 -10 0 +10 +20
-200 x x x x x x-190 x x x x x x-180 x x x x x-170 x x x x x-160 x x x x-150 x x x x-140 x x x x-130 x x x-120 x x x-110 x x-100 x x-90 x x-80 x-70 x
x - transducer heating system required
type BARTEC PSBorder code ACC-PE-Fxxx-CY1 (1 measuring channel)
ACC-PE-Fxxx-CY2 (2 measuring channels)ACC-PE-Fxxx-CY3 (1 measuring channel)ACC-PE-Fxxx-CY4 (2 measuring channels)
power supply 240 V AC 120 V ACconsisting of- heating element PSB 33, 07-5801-2335 PSB 33, 07-5801-1335- connection system PLEXO TCS 27-159P1-1010 PLEXO TCS 27-159P1-1010- junction box PSB 07-5103-2201/2090 07-5103-2201/2090explosion protection ATEX, IECEx, TR TS
type H05SS-Flength m 10max. length m on requestweight kg/m 0.11ambient temperature °C -60...+180installation temperature
°C -20...+50
bend radius 7.5 Dpropertiescable jacketmaterial rubberouter diameter D mm 8...10.4colour blackshield -
FLUXUS® Cryo Technical specification
30 TSFLUXUS_CYO_ V1-0EN_Leu, 2016-10-06
Insulation
insulation height:hi = max. 185 mm
height h:
- with transducer heatingsystem: h < 30 mm
- without transducer heating system: h ≥ 100 mm
hi
h
FLUXUS® Cryo Technical specification
117 TSFLUXUS_CYO_ V1-0EN_Leu, 2016-10-06
FLEXIM GmbH
Wolfener Str. 36
12681 Berlin
Germany
Tel.: +49 (30) 93 66 76 60
Fax: +49 (30) 93 66 76 80
internet: www.flexim.com
e-mail: info@flexim.com
Subject to change without notification. Errors excepted.
FLUXUS® is a registered trademark of FLEXIM GmbH.