Quality Assessment for Routine Clinical MR Spectroscopy

Post on 31-Dec-2015

19 views 1 download

description

Quality Assessment for Routine Clinical MR Spectroscopy. Diana M. Lindquist, PhD Assistant Professor of Radiology. Purpose. Clinical indication: apparently abnormal metabolite ratios in infant with normal imaging findings. - PowerPoint PPT Presentation

Transcript of Quality Assessment for Routine Clinical MR Spectroscopy

Quality Assessment for Quality Assessment for Routine Clinical MR Routine Clinical MR

SpectroscopySpectroscopy

Diana M. Lindquist, PhDAssistant Professor of Radiology

PurposePurpose Clinical indication: apparently abnormal metabolite

ratios in infant with normal imaging findings. Quality control is as important in clinical spectroscopy

as in research Develop a simple short protocol for quality control in

clinical MRS Localization Spectral data

MethodsMethods Standard head coil and

manufacturer’s phantom

Protocol Axial localizer 3 voxels (8 cm3) PRESS

Total time: 30 minutes

Results: Image UniformityResults: Image Uniformity

Results: Voxel LocalizationResults: Voxel Localization Mean voxel size was

20.2 (± 0.8) by 18.9 (± 0.4) mm2 in-plane

One voxel is adequate; scan time reduced from 30 to 15 minutes

Sample spectraSample spectra

TE 35 ms TE 144 ms

Lac

NAA

CrCho

Results: Metabolite RatiosResults: Metabolite RatiosTE

35msTE

144ms

NAA/Cr 1.50 ± 0.03

1.64 ± 0.04

Cho/Cr 0.84 ± 0.01

0.68 ± 0.02

mI/Cr 0.47 ± 0.02

N/A

ConclusionsConclusions Following implementation of this protocol, we

are now able to detect and correct scanner instabilities.

QC data are useful as a monitor of system performance

Scheduling still difficult even with a 15 minute exam.