Quality Assessment for Routine Clinical MR Spectroscopy
-
Upload
cheyenne-turner -
Category
Documents
-
view
19 -
download
1
description
Transcript of Quality Assessment for Routine Clinical MR Spectroscopy
Quality Assessment for Quality Assessment for Routine Clinical MR Routine Clinical MR
SpectroscopySpectroscopy
Diana M. Lindquist, PhDAssistant Professor of Radiology
PurposePurpose Clinical indication: apparently abnormal metabolite
ratios in infant with normal imaging findings. Quality control is as important in clinical spectroscopy
as in research Develop a simple short protocol for quality control in
clinical MRS Localization Spectral data
MethodsMethods Standard head coil and
manufacturer’s phantom
Protocol Axial localizer 3 voxels (8 cm3) PRESS
Total time: 30 minutes
Results: Image UniformityResults: Image Uniformity
Results: Voxel LocalizationResults: Voxel Localization Mean voxel size was
20.2 (± 0.8) by 18.9 (± 0.4) mm2 in-plane
One voxel is adequate; scan time reduced from 30 to 15 minutes
Sample spectraSample spectra
TE 35 ms TE 144 ms
Lac
NAA
CrCho
Results: Metabolite RatiosResults: Metabolite RatiosTE
35msTE
144ms
NAA/Cr 1.50 ± 0.03
1.64 ± 0.04
Cho/Cr 0.84 ± 0.01
0.68 ± 0.02
mI/Cr 0.47 ± 0.02
N/A
ConclusionsConclusions Following implementation of this protocol, we
are now able to detect and correct scanner instabilities.
QC data are useful as a monitor of system performance
Scheduling still difficult even with a 15 minute exam.