Post on 23-Feb-2016
description
QA101: 40 CFR 58 App. A -Gaseous Pollutants
Glenn Gehring, Technology Specialist IIIInstitute for Tribal Environmental Professionalsglenn.gehring@nau.eduVoice: 541-612-0899http://www4.nau.edu/itep/http://www4.nau.edu/tams/
40CFR Part 58 Appendix A
The Code of Federal Regulations (CFRs) are dynamic; always check
to be sure you have the latest information.
CFR Title 40 Part 58 Appendix A
3.2.1 One-Point Quality Control Check for SO2, NO2, O3, and CO. A one-point quality control (QC) check must be performed at least once every 2 weeks on each automated analyzer used to measure SO2, NO2, O3and CO. … The QC check is made by challenging the analyzer with a QC check gas of known concentration (effective concentration for open path analyzers) between 0.01 and 0.10 parts per million (ppm) for SO2, NO2, and O3, and between 1 and 10 ppm for CO analyzers.
3.2.1.1 Except for certain CO analyzers described below, point analyzers must operate in their normal sampling mode during the QC check, and the test atmosphere must pass through all filters, scrubbers, conditioners and other components used during normal ambient sampling and as much of the ambient air inlet system as is practicable. …
3.2.1.3 Report the audit concentration … of the QC gas and the corresponding measured concentration … indicated by the analyzer. The percent differences between these concentrations are used to assess the precision and bias of the monitoring data as described in sections 4.1.2 (precision) and 4.1.3 (bias) of this appendix.
Inside the Monitoring Shelter
Gas Analyzer Rack
Data Logger
Calibrator
Continuous Particulate
Monitor Control Units
Zero Air Generator
NOy
NOx
SO2
Ozone
Toxics Flow Controller
Continuous Particulate Sensor Unit
Vertical manifold
Tubing for NOy
EPA Protocol Gas
316 Stainless Steel 1/8 inch tubing connects regulator to calibrator
CGA 660? Fitting must match bottle
Calibration System PneumaticsDon’t pressurize analyzers; you must vent. Total flow from calibrator must exceed combined analyzer demand
Calibrator
Analyzer Analyzer Analyzer
Vent
Zero Air
Cal Bott
le
Exhaust manifold
Cal gasmust pass through filter
Wall
Insect screen
Regulator(2 stage stainless steel match fittings & purge
CGA 660)
EPA Protocol GasMost have 2-year certification
Calibration Manifold – may have one on calibrator
What is Zero Air?From QA-Handbook – Vol – II (Redbook):Zero Air Systems: Zero air systems should be able to deliver 10 liters/min of air that is free of ozone, NO, NO2, and SO2 to 0.001 ppm and CO and non-methane hydrocarbons to 0.1 ppm. There are many commercially available systems; however, simple designs can be obtained by using a series of canisters.
From 40CFR Part 50 Appendix A-1 (this is not Part 58, Appendix A)
SO24.1.6.2 Clean zero air, free of contaminants that could cause a detectable response or a change in sensitivity of the analyzer. Since ultraviolet fluorescence analyzers may be sensitive to aromatic hydrocarbons and O2-to-N2ratios, it is important that the clean zero air contains less than 0.1 ppm aromatic hydrocarbons and O2and N2percentages approximately the same as in ambient air. A procedure for generating zero air is given in reference 1.
From 40CFR Part 50 Appendix C (this is not Part 58, Appendix A)
CO3.2 Dilution gas (zero air). Air, free of contaminants which will cause a detectable response on the CO analyzer. The zero air should contain <0.1 ppm CO. A procedure for generating zero air is given in Reference 1.
From 40CFR Part 50 Appendix F (this is not Part 58, Appendix A)
NOx1.3.2 Zero air. Air, free of contaminants which will cause a detectable response on the NO/NOX/NO2analyzer or which might react with either NO, O3, or NO2in the gas phase titration. A procedure for generating zero air is given in reference 13.
What is Zero Air? (continued)
Commercial Zero Air Generator
Series of scrubbersDesiccant, Purafil, Molecular sieve, activated carbon, whatever scrubber media is needed
(supply a clean, cooled, dry air flow source that meets pressure needs – always put a 0.1 micron particulate filter downstream of scrubbers)
What are Gas Standards?From QA-Handbook – Vol – II (Redbook):
12.1.2 Gaseous StandardsIn general, ambient monitoring instruments should be calibrated by allowing the instrument to sample andanalyze test atmospheres of known concentrations of the appropriate pollutant in air. The following is an
excerpt from 50 CFR Part 58, Appendix A Section 2.6.1:“Gaseous pollutant concentration standards (permeation devices or cylinders of compressed gas)
used to obtain test concentrations for carbon monoxide (CO), sulfur dioxide (SO2), nitrogenoxide (NO), and nitrogen dioxide (NO2) must be traceable to either a National Institute ofStandards and Technology (NIST) Traceable Reference Material (NTRM) or a NIST-certified
Gas Manufacturer’s Internal Standard (GMIS), certified in accordance with one of theprocedures given in reference 4 of this appendix. Vendors advertising certification with the
procedures provided in reference 4 of this appendix and distributing gasses as ‘‘EPA ProtocolGas’’ must participate in the EPA Protocol Gas Verification Program or not use ‘‘EPA’’ in any
form of advertising.”"Traceable" is defined in 40 CFR Parts 50 and 58 as meaning that a local standard has been compared andcertified, either directly or via not more than one intermediate standard, to a primary standard such as a
National Institute of Standards and Technology Standard Reference Material (NIST SRM) or aUSEPA/NIST-approved Certified Reference Material (CRM)”.
In general, ambient monitoring instruments should be calibrated by allowing the instrument to sample andanalyze test atmospheres of known concentrations of the appropriate pollutant in air. The following is an
excerpt from 50 CFR Part 58, Appendix A Section 2.6.1:“Gaseous pollutant concentration standards (permeation devices or cylinders of compressed gas)
used to obtain test concentrations for carbon monoxide (CO), sulfur dioxide (SO2), nitrogenoxide (NO), and nitrogen dioxide (NO2) must be traceable to either a National Institute ofStandards and Technology (NIST) Traceable Reference Material (NTRM) or a NIST-certified
Gas Manufacturer’s Internal Standard (GMIS), certified in accordance with one of theprocedures given in reference 4 of this appendix. Vendors advertising certification with the
procedures provided in reference 4 of this appendix and distributing gasses as ‘‘EPA ProtocolGas’’ must participate in the EPA Protocol Gas Verification Program or not use ‘‘EPA’’ in any
form of advertising.”"Traceable" is defined in 40 CFR Parts 50 and 58 as meaning that a local standard has been compared andcertified, either directly or via not more than one intermediate standard, to a primary standard such as a
National Institute of Standards and Technology Standard Reference Material (NIST SRM) or aUSEPA/NIST-approved Certified Reference Material (CRM)”.
From 40CFR Part 50 Appendix A-1 (this is not Part 58, Appendix A)
4.1.6.1 SO2 gas concentration transfer standard having a certified SO2 concentration of not less than 10 ppm, in N2, traceable to a NIST Standard Reference Material (SRM).
From 40CFR Part 50 Appendix C (this is not Part 58, Appendix A)
3.1 CO concentration standard(s). Cylinder(s) of CO in air containing appropriate concentrations(s) of CO suitable for the selected operating range of the analyzer under calibration; CO standards for the dilution method may be contained in a nitrogen matrix if the zero air dilution ratio is not less than 100:1. The assay of the cylinder(s) must be traceable either to a National Bureau of Standards (NBS) CO in air Standard Reference Material (SRM) or to an NBS/EPA-approved commercially available Certified Reference Material (CRM). CRM's are described in Reference 2, and a list of CRM sources is available from the address shown for Reference 2. A recommended protocol for certifying CO gas cylinders against either a CO SRM or a CRM is given in Reference 1. CO gas cylinders should be recertified on a regular basis as determined by the local quality control program.
From 40CFR Part 50 Appendix F (this is not Part 58, Appendix A)1.3.1 NO concentration standard. Gas cylinder standard containing 50 to 100 ppm NO in N2with less than 1 ppm NO2. This standard must be traceable to a National Bureau of Standards (NBS) NO in N2Standard Reference Material (SRM 1683 or SRM 1684), an NBS NO2Standard Reference Material (SRM 1629), or an NBS/EPA-approved commercially available Certified Reference Material (CRM).
Basically, if you purchase EPA Protocol Gas at the right concentrations things should be good. BUT, there are factors and calculations that impact what concentration you can get out of your gas calibrator. Let’s talk more about generating a test gas atmosphere to better understand this.
Creating a Test Atmosphere
Cal Gas
Zero AirTest Atmosphere
Basically, it’s mostly about mixing ratios.
Gas/Titration/ Ozone
Concentration & Flow
2 parts this
15 parts this
FlowConcentration & FlowFlow must be correct!
Environics 6100 gas calibrator with 100 ml/min Mass flow Controller (MFC) for the calibration gas and a 10 liter/min (10,000 ml/min) zero air flow.Flow controller limits – not trusted below 10% of range (10 ml/minute is the minimum allowable flow on the cal gas MFC – 8ml/min won’t work)
These flow controllers are calibrated at EPA Standard flow (25˚C at 760 mmHg)
Note: most of the gas in a calibration bottle is actually N2. Gasses flow differently and calibrators sometimes ask what the source gas is in order to make small adjustments to flow. If the cal gas is 50 ppm NO and 50 ppm SO2, almost all the gas in the bottle is actually the mostly inert carrier gas, N2. In this case flow would be set for N2 and not SO2 or NO. The concentrations entered for concentration calculations would be NO and SO2.
NO2 is typically generated by reacting ozone with the nitrogen oxide gas (NO) from a cal bottle in the calibrator. This is called Gas Phase Titration (GPT)
Step 1. How much total flow do I want?
I need enough flow for each analyzer and some excess to ensure outside air won’t mix with my test atmosphere
Analyzer 1 draws 0.5 slpm – Analyzer 2 draws 0.6 slpm - analyzer 3 draws 1.0 slpm
0.5 + 0.6 + 1.0 = a total analyzer draw of 2.1 slpm - add excess
Let’s make total flow 3.1 slpm. 3.1 standard liters per minute is the same as 3100 standard cubic centimeters per minute
(A rotameter in the vent line can assure excess flow – it will measure flow and you just look – tap the floating ball to be sure it isn’t stuck)
(Flow of cal gas) X [cal gas bottle concentration] [concentration of gas out ] = Flow of zero air + Flow of cal gas + other flows (note: the denominator adds up to total flow)
What do we know? Say we want a cal gas out of .090 ppm; we can pick that and “concentration of gas out” is then known. We decided we wanted a total flow of 3.1 slpm (3100 sccm) so we know the total of the denominator, total flow. We also know the calibration gas concentration (you can read it from the cal bottle – let’s say it reads 50.1 ppm). What don’t we know? We need to know the flow of the cal gas. Think algebra and solve for flow of cal gas in the numerator in sccm units. WATCH your units!
Step 2 – Calculate from this formula
(Flow of cal gas) X (50.1 ppm) 0.090 ppm = 3100 sccm
(0.090 ppm) X (3100 sccm) = (Flow of cal gas) = 5.569 sccm PROBLEM!!!!!!!! 50.1 ppm
We did the calculation right so what’s the problem?
Our 100 ml/min calibration gas mass flow controller in the 6100 isn’t accurate below 10% of their flow maximum - 10% of 100 is 10 and we ended up with 5.569. We need to have a higher total flow, a smaller mass flow controller for the cal gas (a 50 sccm flow controller would work in this instance) or a lower concentration cal gas. The easiest might be to just increase total flow, but there are limits. It’s a balancing act between total flows, cal gas concentrations and the flow limits of your calibrator mass flow controllers. Let’s explore this further.
Let’s find out what the minimum total flow would need to be in order to keep the cal gas flow at 10% of the 100 sccm mass flow controller. This time we’ll solve for total flow
(Flow of cal gas is 10 sccm) X (50.1 ppm) 0.090 ppm = Total Flow
(10 sccm) X (50.1 ppm) Total Flow = = 5567 sccm 0.090 ppm
I can adjust my gas calibrator for a total flow of 5570 sccm instead of 3100 sccm and get .090 out; we will have additional flow to vent
What is the lowest concentration we can get from a 50.1 ppm cal gas bottle concentration with 10 liter zero air mass flow controller and a 100 sccm cal gas mass flow controllers in a calibrator?
Lets only use 100% at the top end of the zero air mass flow controller, or 10000 sccm (10 slpm) and the minimum flow of 10 sccm for the cal gas mass flow controller (10% of 100 sccm).
(Flow of cal gas is 10 sccm) X (50.1 ppm) Output ppm = = .050 ppm 10010 for total flow
NOx Analyzer Theory of Operation: Chemiluminescence
Illustration is from Teledyne API 200E Operator’s manual
Sample gas w/NOO3
Light leaks can impact measurement
PMTs are sensitive and can be damaged by light
Illustration is from Teledyne API 200E Operator’s manual
MolyConMolybdenum Converter
NO2 Converter
NO
NO2 NO2 converted to NO NO
NO
Rx Cell
NOx Cycle goes through converterOnly NO reacts with ozone
NO Cycle passes sample air to reaction chamberOzone reacts with NOOnly NO reacts – NO2 isn’t measuredNO2 is calculated: NOx – NO = NO2
Sample AirNOx = NO + NO2
NOx – NO = NO2
NOx
NO
NO Cycle
Valves
Dark Cycle bypasses RX Cell
Ozone generator
Excess Ozone
Dry filtered airhv in reaction cell is measured by PMT
PMT
CFR Title 40 Part 58 Appendix A3.2.2.2 (a) NO2audit gas for chemiluminescence-type NO2analyzers must also contain at least 0.08 ppm NO. NO concentrations substantially higher than 0.08 ppm, as may occur when using some gas phase titration (GPT) techniques, may lead to evaluation errors in chemiluminescence analyzers due to inevitable minor NO–NOXchannel imbalance. Such errors may be atypical of routine monitoring errors to the extent that such NO concentrations exceed typical ambient NO concentrations at the site. These errors may be minimized by modifying the GPT technique to lower the NO concentrations remaining in the NO2audit gas to levels closer to typical ambient NO concentrations at the site.
[This is in the Annual Performance Audit Section, but is relevant to GPT]
Gas Phase Titration (GPT) – Creating a NO2 test atmosphere
In the next slide, [Norem] should be at least .080 ppm, but not be excessively higher than that
NO2 actual = % NO2 ConvertedNO2 expected
Gas Phase Titration (GPT)
28
Some Other Important CFR Requirements
• QMPs/QAPPs/SOPs a must• Independent quality management• Establishment of a Primary Quality Assurance Organization• Use of NIST Traceable Standards• Use of FRM/FEMs at the appropriate operating ranges and with appropriate
settings– PART 53—Ambient Air Monitoring Reference and Equivalent Methods
• Siting, Probe placement & acceptable probe materials– Appendix E to Part 58—Probe and Monitoring Path Siting Criteria for Ambient Air
Quality Monitoring– Appendix D to Part 58—Network Design Criteria for Ambient Air Quality Monitoring
• AQS submittal and annual data certification– 58.15 Annual air monitoring data certification– 58.16 Data submittal and archiving requirements
SO2 Validation Template continued
NO2 Validation Template continued
CO Validation Template continued