QA101: 40 CFR 58 App. A - Gaseous Pollutants

34
QA101: 40 CFR 58 App. A - Gaseous Pollutants Glenn Gehring, Technology Specialist III Institute for Tribal Environmental Professionals [email protected] Voice: 541-612-0899 http://www4.nau.edu/itep / http://www4.nau.edu/tams/

description

QA101: 40 CFR 58 App. A - Gaseous Pollutants . Glenn Gehring, Technology Specialist III Institute for Tribal Environmental Professionals g [email protected] Voice: 541-612-0899 http://www4.nau.edu/itep / http://www4.nau.edu/tams/. - PowerPoint PPT Presentation

Transcript of QA101: 40 CFR 58 App. A - Gaseous Pollutants

Page 1: QA101: 40  CFR 58 App.  A - Gaseous  Pollutants

QA101: 40 CFR 58 App. A -Gaseous Pollutants

Glenn Gehring, Technology Specialist IIIInstitute for Tribal Environmental [email protected]: 541-612-0899http://www4.nau.edu/itep/http://www4.nau.edu/tams/

Page 2: QA101: 40  CFR 58 App.  A - Gaseous  Pollutants

40CFR Part 58 Appendix A

The Code of Federal Regulations (CFRs) are dynamic; always check

to be sure you have the latest information.

Page 3: QA101: 40  CFR 58 App.  A - Gaseous  Pollutants
Page 4: QA101: 40  CFR 58 App.  A - Gaseous  Pollutants

CFR Title 40 Part 58 Appendix A

3.2.1 One-Point Quality Control Check for SO2, NO2, O3, and CO. A one-point quality control (QC) check must be performed at least once every 2 weeks on each automated analyzer used to measure SO2, NO2, O3and CO. … The QC check is made by challenging the analyzer with a QC check gas of known concentration (effective concentration for open path analyzers) between 0.01 and 0.10 parts per million (ppm) for SO2, NO2, and O3, and between 1 and 10 ppm for CO analyzers.

3.2.1.1 Except for certain CO analyzers described below, point analyzers must operate in their normal sampling mode during the QC check, and the test atmosphere must pass through all filters, scrubbers, conditioners and other components used during normal ambient sampling and as much of the ambient air inlet system as is practicable. …

3.2.1.3 Report the audit concentration … of the QC gas and the corresponding measured concentration … indicated by the analyzer. The percent differences between these concentrations are used to assess the precision and bias of the monitoring data as described in sections 4.1.2 (precision) and 4.1.3 (bias) of this appendix.

Page 5: QA101: 40  CFR 58 App.  A - Gaseous  Pollutants

Inside the Monitoring Shelter

Gas Analyzer Rack

Data Logger

Calibrator

Continuous Particulate

Monitor Control Units

Zero Air Generator

NOy

NOx

SO2

Ozone

Toxics Flow Controller

Continuous Particulate Sensor Unit

Vertical manifold

Tubing for NOy

EPA Protocol Gas

316 Stainless Steel 1/8 inch tubing connects regulator to calibrator

CGA 660? Fitting must match bottle

Page 6: QA101: 40  CFR 58 App.  A - Gaseous  Pollutants

Calibration System PneumaticsDon’t pressurize analyzers; you must vent. Total flow from calibrator must exceed combined analyzer demand

Calibrator

Analyzer Analyzer Analyzer

Vent

Zero Air

Cal Bott

le

Exhaust manifold

Cal gasmust pass through filter

Wall

Insect screen

Regulator(2 stage stainless steel match fittings & purge

CGA 660)

EPA Protocol GasMost have 2-year certification

Calibration Manifold – may have one on calibrator

Page 7: QA101: 40  CFR 58 App.  A - Gaseous  Pollutants

What is Zero Air?From QA-Handbook – Vol – II (Redbook):Zero Air Systems: Zero air systems should be able to deliver 10 liters/min of air that is free of ozone, NO, NO2, and SO2 to 0.001 ppm and CO and non-methane hydrocarbons to 0.1 ppm. There are many commercially available systems; however, simple designs can be obtained by using a series of canisters.

From 40CFR Part 50 Appendix A-1 (this is not Part 58, Appendix A)

SO24.1.6.2 Clean zero air, free of contaminants that could cause a detectable response or a change in sensitivity of the analyzer. Since ultraviolet fluorescence analyzers may be sensitive to aromatic hydrocarbons and O2-to-N2ratios, it is important that the clean zero air contains less than 0.1 ppm aromatic hydrocarbons and O2and N2percentages approximately the same as in ambient air. A procedure for generating zero air is given in reference 1.

Page 8: QA101: 40  CFR 58 App.  A - Gaseous  Pollutants

From 40CFR Part 50 Appendix C (this is not Part 58, Appendix A)

CO3.2 Dilution gas (zero air). Air, free of contaminants which will cause a detectable response on the CO analyzer. The zero air should contain <0.1 ppm CO. A procedure for generating zero air is given in Reference 1.

From 40CFR Part 50 Appendix F (this is not Part 58, Appendix A)

NOx1.3.2 Zero air. Air, free of contaminants which will cause a detectable response on the NO/NOX/NO2analyzer or which might react with either NO, O3, or NO2in the gas phase titration. A procedure for generating zero air is given in reference 13.

What is Zero Air? (continued)

Page 9: QA101: 40  CFR 58 App.  A - Gaseous  Pollutants

Commercial Zero Air Generator

Series of scrubbersDesiccant, Purafil, Molecular sieve, activated carbon, whatever scrubber media is needed

(supply a clean, cooled, dry air flow source that meets pressure needs – always put a 0.1 micron particulate filter downstream of scrubbers)

Page 10: QA101: 40  CFR 58 App.  A - Gaseous  Pollutants

What are Gas Standards?From QA-Handbook – Vol – II (Redbook):

12.1.2 Gaseous StandardsIn general, ambient monitoring instruments should be calibrated by allowing the instrument to sample andanalyze test atmospheres of known concentrations of the appropriate pollutant in air. The following is an

excerpt from 50 CFR Part 58, Appendix A Section 2.6.1:“Gaseous pollutant concentration standards (permeation devices or cylinders of compressed gas)

used to obtain test concentrations for carbon monoxide (CO), sulfur dioxide (SO2), nitrogenoxide (NO), and nitrogen dioxide (NO2) must be traceable to either a National Institute ofStandards and Technology (NIST) Traceable Reference Material (NTRM) or a NIST-certified

Gas Manufacturer’s Internal Standard (GMIS), certified in accordance with one of theprocedures given in reference 4 of this appendix. Vendors advertising certification with the

procedures provided in reference 4 of this appendix and distributing gasses as ‘‘EPA ProtocolGas’’ must participate in the EPA Protocol Gas Verification Program or not use ‘‘EPA’’ in any

form of advertising.”"Traceable" is defined in 40 CFR Parts 50 and 58 as meaning that a local standard has been compared andcertified, either directly or via not more than one intermediate standard, to a primary standard such as a

National Institute of Standards and Technology Standard Reference Material (NIST SRM) or aUSEPA/NIST-approved Certified Reference Material (CRM)”.

In general, ambient monitoring instruments should be calibrated by allowing the instrument to sample andanalyze test atmospheres of known concentrations of the appropriate pollutant in air. The following is an

excerpt from 50 CFR Part 58, Appendix A Section 2.6.1:“Gaseous pollutant concentration standards (permeation devices or cylinders of compressed gas)

used to obtain test concentrations for carbon monoxide (CO), sulfur dioxide (SO2), nitrogenoxide (NO), and nitrogen dioxide (NO2) must be traceable to either a National Institute ofStandards and Technology (NIST) Traceable Reference Material (NTRM) or a NIST-certified

Gas Manufacturer’s Internal Standard (GMIS), certified in accordance with one of theprocedures given in reference 4 of this appendix. Vendors advertising certification with the

procedures provided in reference 4 of this appendix and distributing gasses as ‘‘EPA ProtocolGas’’ must participate in the EPA Protocol Gas Verification Program or not use ‘‘EPA’’ in any

form of advertising.”"Traceable" is defined in 40 CFR Parts 50 and 58 as meaning that a local standard has been compared andcertified, either directly or via not more than one intermediate standard, to a primary standard such as a

National Institute of Standards and Technology Standard Reference Material (NIST SRM) or aUSEPA/NIST-approved Certified Reference Material (CRM)”.

Page 11: QA101: 40  CFR 58 App.  A - Gaseous  Pollutants

From 40CFR Part 50 Appendix A-1 (this is not Part 58, Appendix A)

4.1.6.1 SO2 gas concentration transfer standard having a certified SO2 concentration of not less than 10 ppm, in N2, traceable to a NIST Standard Reference Material (SRM).

From 40CFR Part 50 Appendix C (this is not Part 58, Appendix A)

3.1 CO concentration standard(s). Cylinder(s) of CO in air containing appropriate concentrations(s) of CO suitable for the selected operating range of the analyzer under calibration; CO standards for the dilution method may be contained in a nitrogen matrix if the zero air dilution ratio is not less than 100:1. The assay of the cylinder(s) must be traceable either to a National Bureau of Standards (NBS) CO in air Standard Reference Material (SRM) or to an NBS/EPA-approved commercially available Certified Reference Material (CRM). CRM's are described in Reference 2, and a list of CRM sources is available from the address shown for Reference 2. A recommended protocol for certifying CO gas cylinders against either a CO SRM or a CRM is given in Reference 1. CO gas cylinders should be recertified on a regular basis as determined by the local quality control program.

Page 12: QA101: 40  CFR 58 App.  A - Gaseous  Pollutants

From 40CFR Part 50 Appendix F (this is not Part 58, Appendix A)1.3.1 NO concentration standard. Gas cylinder standard containing 50 to 100 ppm NO in N2with less than 1 ppm NO2. This standard must be traceable to a National Bureau of Standards (NBS) NO in N2Standard Reference Material (SRM 1683 or SRM 1684), an NBS NO2Standard Reference Material (SRM 1629), or an NBS/EPA-approved commercially available Certified Reference Material (CRM).

Basically, if you purchase EPA Protocol Gas at the right concentrations things should be good. BUT, there are factors and calculations that impact what concentration you can get out of your gas calibrator. Let’s talk more about generating a test gas atmosphere to better understand this.

Page 13: QA101: 40  CFR 58 App.  A - Gaseous  Pollutants

Creating a Test Atmosphere

Page 14: QA101: 40  CFR 58 App.  A - Gaseous  Pollutants

Cal Gas

Zero AirTest Atmosphere

Basically, it’s mostly about mixing ratios.

Gas/Titration/ Ozone

Concentration & Flow

2 parts this

15 parts this

FlowConcentration & FlowFlow must be correct!

Page 15: QA101: 40  CFR 58 App.  A - Gaseous  Pollutants

Environics 6100 gas calibrator with 100 ml/min Mass flow Controller (MFC) for the calibration gas and a 10 liter/min (10,000 ml/min) zero air flow.Flow controller limits – not trusted below 10% of range (10 ml/minute is the minimum allowable flow on the cal gas MFC – 8ml/min won’t work)

These flow controllers are calibrated at EPA Standard flow (25˚C at 760 mmHg)

Page 16: QA101: 40  CFR 58 App.  A - Gaseous  Pollutants

Note: most of the gas in a calibration bottle is actually N2. Gasses flow differently and calibrators sometimes ask what the source gas is in order to make small adjustments to flow. If the cal gas is 50 ppm NO and 50 ppm SO2, almost all the gas in the bottle is actually the mostly inert carrier gas, N2. In this case flow would be set for N2 and not SO2 or NO. The concentrations entered for concentration calculations would be NO and SO2.

NO2 is typically generated by reacting ozone with the nitrogen oxide gas (NO) from a cal bottle in the calibrator. This is called Gas Phase Titration (GPT)

Page 17: QA101: 40  CFR 58 App.  A - Gaseous  Pollutants

Step 1. How much total flow do I want?

I need enough flow for each analyzer and some excess to ensure outside air won’t mix with my test atmosphere

Analyzer 1 draws 0.5 slpm – Analyzer 2 draws 0.6 slpm - analyzer 3 draws 1.0 slpm

0.5 + 0.6 + 1.0 = a total analyzer draw of 2.1 slpm - add excess

Let’s make total flow 3.1 slpm. 3.1 standard liters per minute is the same as 3100 standard cubic centimeters per minute

(A rotameter in the vent line can assure excess flow – it will measure flow and you just look – tap the floating ball to be sure it isn’t stuck)

Page 18: QA101: 40  CFR 58 App.  A - Gaseous  Pollutants

(Flow of cal gas) X [cal gas bottle concentration] [concentration of gas out ] = Flow of zero air + Flow of cal gas + other flows (note: the denominator adds up to total flow)

What do we know? Say we want a cal gas out of .090 ppm; we can pick that and “concentration of gas out” is then known. We decided we wanted a total flow of 3.1 slpm (3100 sccm) so we know the total of the denominator, total flow. We also know the calibration gas concentration (you can read it from the cal bottle – let’s say it reads 50.1 ppm). What don’t we know? We need to know the flow of the cal gas. Think algebra and solve for flow of cal gas in the numerator in sccm units. WATCH your units!

Step 2 – Calculate from this formula

(Flow of cal gas) X (50.1 ppm) 0.090 ppm = 3100 sccm

(0.090 ppm) X (3100 sccm) = (Flow of cal gas) = 5.569 sccm PROBLEM!!!!!!!! 50.1 ppm

Page 19: QA101: 40  CFR 58 App.  A - Gaseous  Pollutants

We did the calculation right so what’s the problem?

Our 100 ml/min calibration gas mass flow controller in the 6100 isn’t accurate below 10% of their flow maximum - 10% of 100 is 10 and we ended up with 5.569. We need to have a higher total flow, a smaller mass flow controller for the cal gas (a 50 sccm flow controller would work in this instance) or a lower concentration cal gas. The easiest might be to just increase total flow, but there are limits. It’s a balancing act between total flows, cal gas concentrations and the flow limits of your calibrator mass flow controllers. Let’s explore this further.

Let’s find out what the minimum total flow would need to be in order to keep the cal gas flow at 10% of the 100 sccm mass flow controller. This time we’ll solve for total flow

(Flow of cal gas is 10 sccm) X (50.1 ppm) 0.090 ppm = Total Flow

(10 sccm) X (50.1 ppm) Total Flow = = 5567 sccm 0.090 ppm

I can adjust my gas calibrator for a total flow of 5570 sccm instead of 3100 sccm and get .090 out; we will have additional flow to vent

Page 20: QA101: 40  CFR 58 App.  A - Gaseous  Pollutants

What is the lowest concentration we can get from a 50.1 ppm cal gas bottle concentration with 10 liter zero air mass flow controller and a 100 sccm cal gas mass flow controllers in a calibrator?

Lets only use 100% at the top end of the zero air mass flow controller, or 10000 sccm (10 slpm) and the minimum flow of 10 sccm for the cal gas mass flow controller (10% of 100 sccm).

(Flow of cal gas is 10 sccm) X (50.1 ppm) Output ppm = = .050 ppm 10010 for total flow

Page 21: QA101: 40  CFR 58 App.  A - Gaseous  Pollutants
Page 22: QA101: 40  CFR 58 App.  A - Gaseous  Pollutants

NOx Analyzer Theory of Operation: Chemiluminescence

Page 23: QA101: 40  CFR 58 App.  A - Gaseous  Pollutants

Illustration is from Teledyne API 200E Operator’s manual

Sample gas w/NOO3

Light leaks can impact measurement

PMTs are sensitive and can be damaged by light

Page 24: QA101: 40  CFR 58 App.  A - Gaseous  Pollutants

Illustration is from Teledyne API 200E Operator’s manual

Page 25: QA101: 40  CFR 58 App.  A - Gaseous  Pollutants

MolyConMolybdenum Converter

NO2 Converter

NO

NO2 NO2 converted to NO NO

NO

Rx Cell

NOx Cycle goes through converterOnly NO reacts with ozone

NO Cycle passes sample air to reaction chamberOzone reacts with NOOnly NO reacts – NO2 isn’t measuredNO2 is calculated: NOx – NO = NO2

Sample AirNOx = NO + NO2

NOx – NO = NO2

NOx

NO

NO Cycle

Valves

Dark Cycle bypasses RX Cell

Ozone generator

Excess Ozone

Dry filtered airhv in reaction cell is measured by PMT

PMT

Page 26: QA101: 40  CFR 58 App.  A - Gaseous  Pollutants

CFR Title 40 Part 58 Appendix A3.2.2.2 (a) NO2audit gas for chemiluminescence-type NO2analyzers must also contain at least 0.08 ppm NO. NO concentrations substantially higher than 0.08 ppm, as may occur when using some gas phase titration (GPT) techniques, may lead to evaluation errors in chemiluminescence analyzers due to inevitable minor NO–NOXchannel imbalance. Such errors may be atypical of routine monitoring errors to the extent that such NO concentrations exceed typical ambient NO concentrations at the site. These errors may be minimized by modifying the GPT technique to lower the NO concentrations remaining in the NO2audit gas to levels closer to typical ambient NO concentrations at the site.

[This is in the Annual Performance Audit Section, but is relevant to GPT]

Gas Phase Titration (GPT) – Creating a NO2 test atmosphere

In the next slide, [Norem] should be at least .080 ppm, but not be excessively higher than that

Page 27: QA101: 40  CFR 58 App.  A - Gaseous  Pollutants

NO2 actual = % NO2 ConvertedNO2 expected

Gas Phase Titration (GPT)

Page 28: QA101: 40  CFR 58 App.  A - Gaseous  Pollutants

28

Some Other Important CFR Requirements

• QMPs/QAPPs/SOPs a must• Independent quality management• Establishment of a Primary Quality Assurance Organization• Use of NIST Traceable Standards• Use of FRM/FEMs at the appropriate operating ranges and with appropriate

settings– PART 53—Ambient Air Monitoring Reference and Equivalent Methods

• Siting, Probe placement & acceptable probe materials– Appendix E to Part 58—Probe and Monitoring Path Siting Criteria for Ambient Air

Quality Monitoring– Appendix D to Part 58—Network Design Criteria for Ambient Air Quality Monitoring

• AQS submittal and annual data certification– 58.15 Annual air monitoring data certification– 58.16 Data submittal and archiving requirements

Page 29: QA101: 40  CFR 58 App.  A - Gaseous  Pollutants
Page 30: QA101: 40  CFR 58 App.  A - Gaseous  Pollutants

SO2 Validation Template continued

Page 31: QA101: 40  CFR 58 App.  A - Gaseous  Pollutants
Page 32: QA101: 40  CFR 58 App.  A - Gaseous  Pollutants

NO2 Validation Template continued

Page 33: QA101: 40  CFR 58 App.  A - Gaseous  Pollutants
Page 34: QA101: 40  CFR 58 App.  A - Gaseous  Pollutants

CO Validation Template continued