A Charge-Based Magneto-Electric Test Procedure

Post on 15-Jan-2016

21 views 3 download

description

A Charge-Based Magneto-Electric Test Procedure. Scott P. Chapman & Joseph T. Evans, Jr. Radiant Technologies, Inc. Aug 9, 2011 IWPMA 2011. Summary. - PowerPoint PPT Presentation

Transcript of A Charge-Based Magneto-Electric Test Procedure

Radiant Technologies, Inc.Magneto-Electric Test Procedure

A Charge-Based Magneto-Electric Test Procedure

Scott P. Chapman & Joseph T. Evans, Jr.

Radiant Technologies, Inc.

Aug 9, 2011

IWPMA 2011

Radiant Technologies, Inc.Magneto-Electric Test Procedure

Summary

The goal is to describe an experiment to characterize the charge response of a piezoelectric or multiferroic sample in the presence of a magnetic (B) field by:

P = H

B = H

P = /B

For a multiferroic, H induces P directly. For our piezoelectric sample, P results from direct force (dc) or torque (d) applied to the sample ferroelectric with a magnet attached.

Radiant Technologies, Inc.Magneto-Electric Test Procedure

SummaryI will present:

• Mathematics and theory relating predictive and measured polarization response to the magnetic field and magnetic field geometry.

• Experimental considerations.

• Experimental design and configuration.

• Measured results.

• Measured comparison to predictive.

Radiant Technologies, Inc.Magneto-Electric Test Procedure

Experiment• Experiment – Stimulate sample with

magnetic field and measure charge. Simple.

• Characterize sample independently, calculate B Field and predict charge. Hard.

• Compare the two.

Radiant Technologies, Inc.Magneto-Electric Test Procedure

Magnetic Force

• Magnetic force is generated only by moving electric charges.

• For two objects to exert magnetic force both must contain moving charges.

These three statements apply to understanding Magnetic Force:

Radiant Technologies, Inc.Magneto-Electric Test Procedure

Magnetic Force• Magnetic force calculation proceeds as follows:

Calculate a mathematical field, H, that sums the motion of all charge particles at the point of interest in the field.

Multiply H by the magnetic permeability factor, , to convert it to a force field, B.

Use B to calculate magnetic force on the target. This requires the calculation of both the HH coil force and the target force, and their multiplication.

Radiant Technologies, Inc.Magneto-Electric Test Procedure

Geometry (Along Field Axis)Bsingle coil = 0NIR2 x 0.5(R2+x2)-3/2 (1)

BHHC = 0.5 0NIR2 x 1/(R2+(x+K/2)2)3/2

+ 0.5 0NIR2 x 1/(R2+(x-K/2)2)3/2 (2)

N = Number of Coils R = Coil Radius (m)I = Current Through Loop (Amps) K = Coil Separation (m)x = Location of interest measured from midway between the coils (m)

B = 0.716 0NI/R (3)

For: K = R and x = 0 (Centered Between Coils)

Radiant Technologies, Inc.Magneto-Electric Test Procedure

Basic Test ConfigurationOrientation 1 – Centered – m || B

Radiant Technologies, Inc.Magneto-Electric Test Procedure

Basic Test ConfigurationOrientation 2 – Centered – m B

Radiant Technologies, Inc.Magneto-Electric Test Procedure

Basic Test ConfigurationOrientation 3 – x = 1.5 R – m B

Radiant Technologies, Inc.Magneto-Electric Test Procedure

Plot Measured Charge Vs Field

H

P

P is Accurately Measured but H may be inferred

Arbitrary Data

Radiant Technologies, Inc.Magneto-Electric Test Procedure

Independent Field ValuesThe independent (Field) axis in the data presentation can be

determined by:

• Assumed Field by DRIVE Volts into the Current Amplifier. This experiment presented here uses this approach.

• Assumed Field by Measured Current into the Helmholtz Coil. This reduces the number of error sources in the first option by half.

• Measured Field by magnetic sensor. Most accurate.

Radiant Technologies, Inc.Magneto-Electric Test Procedure

Some Field Profiles

Radiant Technologies, Inc.Magneto-Electric Test Procedure

Improved Test Configuration - Measure HH Coil Input Current

Radiant Technologies, Inc.Magneto-Electric Test Procedure

Improved Test Configuration - Direct Field Measurement at Sample

Radiant Technologies, Inc.Magneto-Electric Test Procedure

Advanced Test Configuration - Introduce a DC Bias Field

Radiant Technologies, Inc.Magneto-Electric Test Procedure

Program Entry Parameters

Radiant Technologies, Inc.Magneto-Electric Test Procedure

Measurement Configuration

Radiant Technologies, Inc.Magneto-Electric Test Procedure

Data Presentation Configuration

Radiant Technologies, Inc.Magneto-Electric Test Procedure

Direct Force (dc) Measurement

Radiant Technologies, Inc.Magneto-Electric Test Procedure

Direct Force (dc) Response

Radiant Technologies, Inc.Magneto-Electric Test Procedure

Torque (d) Measurement

Radiant Technologies, Inc.Magneto-Electric Test Procedure

Torque (d) Response

Radiant Technologies, Inc.Magneto-Electric Test Procedure

Measured Piezo Constants

The constants we measured: Numbers are high (~10 x) due to substrate bending.

• Parallel to the magnetic axis (dc):

* 612 pC/N (10 g = 0.098 N)

* 688 pC/N (20 g = 0.196 N)

* 653 pC/N (50 g = 0.49 N)

* 650 ±40 pC/N (Average)

Radiant Technologies, Inc.Magneto-Electric Test Procedure

Measured Piezo Constants

• Torque (d): 765 pC/N

When applying magnetic torque, the force must be calculated from the lever arm length and then multiplied by the equivalent torque piezoelectric constant

Radiant Technologies, Inc.Magneto-Electric Test Procedure

Primary Measurement Error SourcesThere are three primary sources of error:

• Frequency response of the current amplifier with the attached HH coil. Slow the measurement to ensure the amplifier can provide the requested HH coil input power.

• Parasitic charge resulting from magnetic induction in the RETURN cable. This effect is reduced by slowing the measurement. Measure the effect and subtract from the final measurement.

• Charge measurement accuracy reduced by charge deterioration over long tests. This effect is reduced by limiting the length of the test.

1000.0 ms is optimal and well within equipment capabilities. Significantly limits inductive charge.

Radiant Technologies, Inc.Magneto-Electric Test Procedure

Pre-Measurement Steps

• Calculate the magnetic field at the point where the sample is located.

• Measure the induced current in the cable, under measurement test conditions, and retain to subtract from the measured data.

To prepare for the Magneto-Electric Response Task measurement, perform these steps:

Radiant Technologies, Inc.Magneto-Electric Test Procedure

Pre-Measurement Steps

• Reduce the test speed to reduce inductive current, but no slower than 1000.0 ms.

• Determine through experimentation the maximum frequency and ensure 1000/Test Period (ms) does not exceed this value.

Radiant Technologies, Inc.Magneto-Electric Test Procedure

Predictive Modelm || B - Centered in HH Coil

Define, for our force-inducing magnet:

m = MV

M = Magnetization of Magnet V = Volume

For B || m

F = [m ·B] (1)

For constant m, as with our reference sample magnets:

F = m B/ x (2)

For constant B, as in the center of the Helmholtz coil:

F = 0 => Q = 0 pC

Radiant Technologies, Inc.Magneto-Electric Test Procedure

Predictive Modelm B - Centered in HH Coil

Piezo Constant:

d = 0.75 V x 100 pC/10g (Sense Capacitor) =

75 pC/0.098 N = 765 pC/N

Ftorque ():

m = 4 x 1.08 T/4x10-7 x (0.00252 x 0.006) = 0.4 A/m

Estimated Charge (Q) at 45.0 Gauss:

Q = d x 0.4 A/m x B / Height

= 765 pC/N x 0.4 A/m x 45 e-4 T / 0.006 m

= 229.5 pC

Radiant Technologies, Inc.Magneto-Electric Test Procedure

Predictive Model m || B - At 1 K From Closest Coil

x = 1.5 K = 1.5 R

B = -0.319 0NI/R2

Q = dc x 0.4 A/m x B

= 650 pC/N x 0.4 A/m x -0.093I T/m

At 45 Gauss, I = 1.68 Amps

Q = 650 pC/N x 0.4 A/m x -0.093 x 1.68 A T/m

= -4.06 pC

Radiant Technologies, Inc.Magneto-Electric Test Procedure

Predictive Modelm B - At 1 K From Closest Coil

At x = 0:

B = 0.716 0NI/R

=> 0NI/R = B/0.716 = 45.0/0.716 = 62.85 G

At x = 1.5 K = 1.5 R:

BHHC = 0.5 0NIR2/(R2+(x+K/2)2)3/2+ 0.5 0NIR2/(R2+(x-K/2)2) 3/2 G

= 0.5 0NIR2/(R2+(1.5R + R/2)2)3/2+ 0.5 0NIR2/(R2+(1.5R-K/2)2)3/2

= 0.1727 0NI/R G = 10.855 G

Q = d x 0.4 A/m x B / Height

= 765 pC/N x 0.4 A/m x 10.855 e-4 T / 0.006 m

= 55.36 pC

Radiant Technologies, Inc.Magneto-Electric Test Procedure

Experiment

Radiant Technologies, Inc.Magneto-Electric Test Procedure

Measured Data - Centered || B

Measured: 1 pCInductive: 1 pCCorrected: 2 pC

Signal Smaller than Noise

Radiant Technologies, Inc.Magneto-Electric Test Procedure

Measured Data - Centered B

Measured: 100 pCInductive: -1 pCCorrected: 101 pC

Radiant Technologies, Inc.Magneto-Electric Test Procedure

Measured Data - x = 1.5 R || B

Measured: -0.75 pCInductive: -1.5 pCCorrected: 0.5 pC

Signal Smaller than Noise

Radiant Technologies, Inc.Magneto-Electric Test Procedure

Measured Data - x = 1.5 R B

Measured: 30 pCInductive: -2 pCCorrected: 32 pC

Radiant Technologies, Inc.Magneto-Electric Test Procedure

Summarize Results @ 45 GExperiment Predicted

Results (pC)Measured

Results(pC)Centered || 0.0 0.0

Centered 229 101

At x = 1.5 R || -4 Indeterminate

At x = 1.5 R 55 32

Radiant Technologies, Inc.Magneto-Electric Test Procedure

Possible Error Sources• Amps/DRIVE Volts conversion for the KEPCO 36-6M

current amplifier. -1.75 Volts/Amp used. Expected current = 45.0 G X 0.0373 Amps/Gauss = 1.68 Amps. Post-data measurement showed 1.799 Amps. Generated 48.15 G.

• Current/Gauss conversion for the Lakeshore MH-6 Helmholtz coil. Used the Lakeshore published conversion of 26.76 G/A => 0.0373 A/G. Did not measure the actual ratio.

Radiant Technologies, Inc.Magneto-Electric Test Procedure

Possible Error Sources

• Manual dc and d measurements. Primary

• Unstable measurement surface.

• Unfixed sample subject to bending an shear. Primary

Radiant Technologies, Inc.Magneto-Electric Test Procedure

Conclusion• Radiant successfully tested the magneto-electric response

of a piezoelectric force sensor coupled to a magnet using Radiant’s Magnetoelectric Response Task.

• The system was able to cleanly capture the measurements that generated 30 pC of response.

• The sample response differed from our predictions most likely due to several possible error sources in the test fixture and piezo constants.

• To properly utilize the MT Task and achieve accurate results, these error sources must be mitigated.