Protocol: HSS Hip (Metal, 1.5T Only)
Intravenous Route: None Gain intravenous access using a 20ga, 22ga Angiocath Access central venous devices per protocol
Patient Identification:
Checklist
Safety questionnaire done
Correct patient selected on scanner
Patient ID checked
Sent to PACS
ICD-9 completedProtocol Information
Scanning Technologist:
Contrast : None
Comments
Accession #:
Reviewing Technologist:
RIS completed
Plane Sequence name FOV Siemens 1. 3-Plane Scout 40 gre_r 2. Axial PD Ax fs (TE high 20s) 32-36 tse 3. Coronal STIR Coro 18 tse ir 4. Axial STIR Ax 18 tse ir 4. Coronal PD Coro (TE high 20s) 18 tse 5. Axial PD Ax (TE high 20s) 18 tse 6. Sagittal PD Sag (TE high 20s) 18 tse fs __________________________TIPS_____________________________ - Plug in Cardiac Coil and position it on patient at beginning of study - Series 1 through 3, use Body Coil (not Torso Coil) - Series 4 through 6, use Cardiac Coil - Bandwidth high on sequences due to joint replacement
John Carrino, M.D / T1908
USR > MSK > ORTHO
CORO
SAG
AXIAL (inferior extent to
cover iliopsoas tendon)
Last updated 2/24/11
CD given to patient (if requested)