Microwave near-field scanning microscope
Abstract We present the development of a novel near-field scanning
microwave microscope based on a dielectric resonator. This probe allows local characterization of conducting and insulating films. Our probe has proved the capability of contactless mapping of local thickness, conductivity, and spatially-resolved Electron Spin Resonance (ESR). As the probe is also polarization sensitive, it allows locally mapping the Hall Effect of semiconductors and the magnetization of ferromagnetic thin films. Recently, we designed various probes operating in the frequency range 4-26 GHz with high quality factor, very good sensitivity and high spatial resolution of micron and sub-micron. Our probe also allows the integration of an optical path to the sample deeming it suitable for optically detected magnetic resonance and other optical-microwave measurements.
F. Sakran, M. Abu-Teir, A. Copty, M.Golosovsky and D. DavidovRacah Institute of Physics, The Hebrew University of Jerusalem, Jerusalem, Israel
References :[1] C. Gao and X. –D. Xiang, Rev. Sci. Instrum. 69, 3846 (1998)
[2] M. Golosovsky and D. Davidov, Appl. Phys. Lett. 68, 1579 (1996)
[3] J. Gallop, L. Hao, and F. Abbas, Physica C 282-287, 1579 (1997)
[4] Abu-Teir M, Golosovsky M, Davidov D, Near-field scanning microwave probe based on a dielectric resonator,REV. SCI. INSTRUM. 72 (4): 2073-2079 APR 2001
[5] Abu-Teir M, Sakran F, Golosovsky M, Local contactless measurement of the ordinary and extraordinary Hall effect using near-field microwave microscopy APPL PHYS LETT 80 (10): 1776-1778 MAR 11 2002
[6] Sakran F, Copty A, Golosovsky M, Electron spin resonance microscopic surface imaging using a microwave scanning probe APPL PHYS LETT 82 (9): 1479-1481 MAR 3 2003
a) Conductivity/Thickness of thin conducting films:
Measurement Setup
00
0s 21
ZtyReflectivi
Z
Z
ZZ
Z s
s
yresistivit the is -
typermeabili spacefree the is -
frequency tmeasuremen the is - f
fZ
0
s
0
thickness. film the is - dd
Z s
Z0 - is the impedance of free space; Zs - is the effective surface impedance;
Thick films:
Thin films:
Near-field reflectivity (S11) of thin silver films of different thickness d.Note the increase of S11 upon increasing film thickness.
The reflectivity Vs. film sheet resistance.The inset shows the a Q factor Vs. sheet resistance.
Probe spatial resolution: X-scan over 0.1 mm chromium strips. The resonator includes slot width of 60 µm. Higher resolution can be obtained by narrowing the slit width.
b) Hall effect:
MRHRS exy 012
-1
0
1
2
3
0 0.1 0.2 0.3
n-Si
p-Si
H(T)
S 12 (
x 1
04 )
)
0
1
2
3
4
5
6
7
0 0.1 0.2 0.3 0.4
80 nm
Fig.5 M. Abu-Teir et al.
S12 (
x10
4)
30 nm
10 nm
Ni
H(T)
-0.5
0
0.5
1
1.5
0 0.1 0.2 0.3
increasing fielddecreasing field
H(T)
S12 (
x1
04)
Ni d=30nm
-1
-0.5
0
0.5
1
-0.15
-0.1
-0.05
0
0.05
0.1
0.15
-15 -10 -5 0 5 10 15
X (mm)
NdFeB magnet
South pole up
North pole up
S12
(x
104 )
μ 0
H (
T)
Contactless measurement of the Hall effect in Si wafers on a metal substrate.
c) Electron Spin Resonance (ESR) :
Measurement Setup
Hgh
10 mm
Spatially-resolved ESR signal from four DPPH grains. An XY-scan, at fixed magnetic field and at fixed frequency, over four small grains of DPPH. We observe four clearly defined ‘‘hills,’’ corresponding to the four DPPH grains. the inset shows the optical image of the sample.
-200
-100
0
100
200
300
2990 3000 3010 3020
Magnetic Field (G)
-100
-50
0
50
100
150
8.43 8.44 8.45 8.46 8.47 8.48
Frequency (GHz)
-200
-100
0
100
200
300
2990 3000 3010 3020
Magnetic Field (G)
-100
-50
0
50
100
150
8.43 8.44 8.45 8.46 8.47 8.48
Frequency (GHz)
Local ESR signal from a 120- μm-thick DPPH layer measured by a 9 GHz probe. The inset shows the ESR signal (using a different DPPH sample) obtained via a frequency sweep and a field modulation.
(a) Probe design and (b) measurement setup.
Microwave Hall effect in ferromagnetic Ni films. (Extraordinary Hall effect)
Mapping of the perpendicular magnetic field of a NdFeB permanent magnet. The solid curve yields the calculated field of the magnet.
XYZ stage
Probe
2” Si wafer
Permanent Magnet
NS
Introduction
Results
Our Microwave Near-Field Scanning Microscope based on dielectric resonator and narrow slit. (a) The 9 GHz probe design, (b) The actual probe, (c) the probe’s resonance as measured by a Network Vector Analyzer. The probe transmit more than 90% of the incident microwave power.
8.865 8.870 8.875 8.880 8.885 8.890-60
-50
-40
-30
-20
-10
0
Ref
lect
ion
(dB
)
(Frequency (GHz))
slit
coaxial connector
sapphire resonator
tuning screwair-gap
sapphiretransducer
coax-to waveguide adaptor
SMA connector
(a)
(b)
(c)
Probe design
g – is the g-factor
μB - is the Bohr magneton
H – is the magnetic field
ν – is the resonant frequency of the probe
h – is plank constant
S12 is the microwave reflectivity measured at port 2.
ρxy is the Hall resistivity H is the magnetic field
M is the magnetization.
The first term here, RoH, represents the ordinary Hall effect.
The second term ReH represaents the Extra ordinary Hall effect.
Microwave scanning probes for local characterization of conducting and insulating films attract considerable interest since they are contactless, versatile, and provide high spatial resolution. Recently several microwave scanning probes have been developed, namely coaxial tip [1], slot aperture [2], and dielectric resonator [3]. In our work, there are two important requirements: spatial resolution and sensitivity. In order to achieve high spatial resolution, one needs to use small aperture radiators for near-field scanning. The combination of the dielectric resonator and small aperture, which we report here, provide a highly efficient and sensitive microwave scanning probe.We have developed a variety of Near-Field Scanning Microwave Microscopes working in the 4-26 GHz frequency range [4]. Our probes already proved their efficiency of measuring thickness of conducting layers in the range of 0.1-1 μm mostly applicable to the semiconductor industry.Our near-field microscope allows measurement of the (a) Hall effect [5] (b) Ferromagnetic Resonance (FMR) and (c) Magnetoresistance of magnetic thin films with micron spatial resolution. Moreover, our probe can perform localized electron spin resonance (ESR) measurements [6].
Conclusions :
1. Contactless measurements of conductivity/thickness of thin conducting films in the range 0.1- 1μm.
2. Local, sensitive and contactless Hall effect in semiconductors and magnetization of thin ferromagnetic films. Possibility for low temperature measurements.
3. Our ESR spectrometer is local, sensitive, contactless and non limited to sample size. The sensitivity of our present probe is already better than the sensitivity of a conventional ESR spectrometer.