Operating Instructions
Sartorius ME and SE Series
ME and SE Models
Electronic Analytical Balances and Semi-micro-, Micro- and Ultra-Microbalances
98648-008-84
7/25/2019 MAN-ME_SE-e
2/173
ME and SE models are high-resolutionbalances of special accuracy forextremely precise measurement of mass.
These series cover a range from0.001 mg to 610 g.
A broad range of special performancefeatures makes the ME and SE balancesideal for use as measuring and testequipment in ISO or GLP quality man-agement systems.These features include:
The fully automatic self-calibratingand adjustment function, isoCAL(time- and temperature-dependent)
reproTEST for quick determination ofthe standard deviation to check therepeatability of results
ISO/GLP-compliant recording capabilityfor printouts
Password-protected menu lock
Display of maintenance |service intervals when due
ME and SE balances meet the highestrequirements placed on the accuracyand reliability of weighing results
through the following features: Efficient filtering-out of vibration
Fully automatic draft shield with threemotorized, self-teaching draft shieldelements and soft-touch technology
Stable and repeatable results
Excellent readability under anylighting conditions
Rugged design and durable weighingsystem
ME215/235/254/414/614: Integrated static electricity eliminator
feature to neutralize interfering electro-static charges (ionizer)
ME and SE balances save work andspeed up both simple and complexroutine applications through:
Ultrafast response times
Built-in application programs;application level 1: Second weight unit Counting Weighing in percent Animal weighing Recalculation Calculation Density determination Differential weighing Air buoyancy correction Air density determination
for ME5, SE2 Diameter determination
Application level 2: Checkweighing Time-controlled functions
Application level 3: Totalizing Formulation Statistics
with the following additional functions: Second tare memory
Identification codes Product data memory SQmin function Manual data storage in
application level 3 DKD uncertainty of measurement
Automatic initialization when youswitch on the balance
Easy input of IDs forsamples or other weighed objects
If requested: control using anexternal computer
SymbolsThe following symbols are used inthese instructions:
indicates steps you must perform
$ indicates steps you must performonly under certain conditions
> describes what happens after youhave performed a certain step
indicates an item in a list
! indicates a hazard
Conventions Used in These
Operating Instructions: The pictures in these Operating
Instructions are based on theME215S model. On other models,some display readouts and printoutsmay differ slightly from the onesshown. This will be explained incases where this is important foroperation of the balance.
Hotline:For advice on the use of applica-tions, just call or fax your localSartorius office. For the address,please visit our Internet website at:www.sartorius.com
2
Intended Use
7/25/2019 MAN-ME_SE-e
3/173
2 Intended Use
4 Warning and Safety Instructions
5 General Views of the Balances5 ME215/235/254/414/6146 ME36S7 ME5, SE2
8 Operating Design
12 Getting Started12 Storage and Shipping Conditions12 Unpacking the Balance12 Carrying the Balance14 Equipment Supplied15 Installation Instructions15 Remote Operation of the
Display and Control Unit17 Connecting the Balance to AC Power18 Warmup Time19 Leveling the Balance20 Transporting the Balance
23 Configuring the Balance23 Selecting the Language24 Navigating in the Setup Menu25 Entering the Time and Date26 Setting the Balance Functions29 Setting the Device Parameters29 Entering a Password34 Setting the Application Parameters
42 Selecting the Printout Function44 Printout Configuration46 Device Information46 Factory Settings
47 Operating the Balance47 Basic Weighing Function47 General Instructions for
Analytical Weighing48 Below-Balance Weighing51 Device Parameters51 Opening and Closing the Draft Shield53 Static Electricity Eliminator (Ionizer)55 Calibration, Adjustment, Linearization65 Repeatability Test
66 Application Programs67 Toggle between Two Weight Units69 Counting72 Weighing in Percent75 Calculation78 Density Determination83 Differential Weighing95 Air Buoyancy Correction
102 Diameter Determination
3
Contents
105 Time-Controlled Functions108 Statistics113 Extra Functions
113 Second Tare Memory115 Individual Identification Codes119 Saving Values Manually in M+120 Changing the Resolution122 Product Data Memory124 SQmin Function126 DKD Uncertainty of Measurement128 Combining Applications129 Practical Combination of Several
Applications (Example)131 Data Output Functions133 Interfaces136 Printouts140 Serial Communications Port145 Pin Assignment Charts146 Cabling Diagram
147 Error Codes and Messages
150 Care and Maintenance
152 Recycling
153 Overview153 Specifications159 Accessories (Options)160 Dimensions (Scale Drawings)163 Declarations of Conformity166 EC Type-Approval Certificate
168 Plates and Markings171 Index
AppendixEntering the General PasswordBrief Instructions
7/25/2019 MAN-ME_SE-e
4/173
This balance complies with theEuropean Council Directives as well asinternational regulations and
standards for electrical equipment,electromagnetic compatibility, andthe stipulated safety requirements.Improper use or handling, however,can result in damage and/or injury.
Read these operating instructionsthoroughly before using yourbalance to prevent damage to theequipment. Keep these instructions ina safe place.
Follow the instructions belowto ensure safe and trouble-freeoperation of your balance:
! Do not operate in a hazardousarea/location
! Make sure that the voltage ratingprinted on the AC adapter is identicalto your local line voltage
! If you use electrical equipment ininstallations and under ambientconditions requiring higher safetystandards, you must comply with theprovisions as specified in the appli-cable regulations for installation in
your country.
! Exposure to excessive electromagnet-ic interference can cause the readoutvalue to change. Once the distur-bance has ceased, the instrument canbe used again in accordance with itsintended purpose.
The only way to switch the power offcompletely is to disconnectthe AC adapter
The balance housing is protectedagainst the penetration of solid objectswith a diameter of more than 2.5 mm(such as accumulated dust) anddripping water falling vertically (IP32) the housing is not completely dust-and leak-tight, however
Protect the AC adapter from contactwith liquid
Note on Installation:The operator shall be responsible forany modifications to Sartorius equip-ment and for any connections of cablesor equipment not supplied by Sartoriusand must check and, if necessary,correct these modifications and connec-tions. On request, Sartorius will provideinformation on the minimum operatingspecifications (in accordance with theStandards listed above for definedimmunity to interference).
Connect only Sartorius accessories andoptions, as these are optimally designedfor use with your balance
When cleaning your balance, makesure that no liquid enters the balancehousing; use only a slightly moistenedcloth to clean the balance.
Do not open the balance housing.If the seal is broken, this will result inforfeiture of all claims under themanufacturers warranty.
In case you have any trouble with yourbalance:
$ contact your local Sartorius office,dealer or service center
4
Warning and Safety Instructions
7/25/2019 MAN-ME_SE-e
5/173
5
Models ME215/235/254/414/614
General Views of the Balance
Pos. Designation Spare Part Order No.
17 On/off key18 Toggle key for alphabetic input19 Function keys20 Leveling foot 69MA009121 Lug for attaching an antitheft
locking device22 Menu access switch23 Serial printer port (PRINTER)24 Serial communications port
(PERIPHERALS)25 Terminal for connecting an
equipotential bonding conductor
26 DC jackNot shown:Set of dust covers 6960ME01Set of small parts (operating panel) 69ME0007Set of caps 69ME0008
Pos. Designation Spare Part Order No.
1 Draft shield door grips2 Weighing pan 69 ME00013 Shield disk (ME235S/P only)4 Shield plate 69 ME00025 Level indicator6 Operating panel7 Display8 Tare key9 Key for opening/closing 69ME0007
draft shield (set of small parts)10 Decimal point key11 Print key
12 Ionizer on/off key13 CF key (clear function)14 Numeric keys15 Toggle key for changing the application program16 Setup key for configuring the balance
1
2
34
5
6
7
8
910
12
13
1415
1617
18
8
9
19
20
21
20
22
2324
25
26
11
7/25/2019 MAN-ME_SE-e
6/173
6
General Views of the Balance
Pos. Designation
1 Draft shield door grips2 Weighing pan3 Shield disk4 Shield plate5 Level indicator6 Operating unit with control panel7 Serial communications port (PERIPHERALS)8 Serial printer port (PRINTER)9 Menu access switch
10 DC jack11 Female connector for weigh cell
12 Decimal point key13 Print key14 Info key for displaying device information
ME36S
Pos. Designation
15 CF key (clear function)16 Toggle key for changing the application program17 Setup key for configuring the balance18 On/off key19 Toggle key for alphanumeric input20 Numeric keys21 Function keys22 Key for opening/closing draft shield23 Tare key24 Display25 Connector for plugging the weigh cell into the operating panel
26 Leveling foot27 Lug for attaching an antitheft locking device28 Terminal for connecting an equipotential bonding conductor
MES_06.cdr
1817
12
13141516
19
20
21
28
27
26
26
25
24
2223
2322
1
2
3
4
5
6
789
10
11
7/25/2019 MAN-ME_SE-e
7/173
7
Pos. Designation
1 Weighing pan2 Filter pan, 50 mmd3 Interior draft shield (only for SE2)4 Optional filter pan, 75 mmd5 Shield ring6 Optional filter pan, 90 mmd7 Shield disk8 Level indicator9 Leveling foot
10 Draft shield cover11 Female connector for weigh cell
12 DC jack for AC power13 Communications port (PERIPHERALS)14 Printer port (PRINTER)15 Tare key
Pos. Designation
16 Key for opening the draft shield counterclockwise17 Print key18 Info key for displaying device information19 CF key (clear function)20 Numeric keys21 Toggle key for changing to the next application program22 Key for accessing Setup mode (settings)23 On/off key24 Toggle key for alphanumeric input25 Key for opening the draft shield clockwise26 Function keys
27 Display28 Lug for attaching an antitheft locking device29 Female connector for evaluation unit30 Terminal for connecting an equipotential bonding conductor31 Draft shield
General Views of the Balance
Models ME5 and SE2
12
3
4
5
6
7
8
9
10
11
12
13
14
15
17
16
18
19
20
23
24
25
15
28
29
30
7
1
3
31
22
21
27
26
7/25/2019 MAN-ME_SE-e
8/173
8
The balance consists of a weighing cell, adraft shield and a display and control unit.In addition to the choice of power supply,
via AC adapter or external rechargeablebattery pack, your balance also has inter-face ports for connecting additionaldevices, such as a printer, computer, oruniversal remote control switch, etc.
The display and control unit is fastenedto the weighing cell. Operation of thebalance follows a uniform philosophy,which is described in this manual.
Where not expressly indicated otherwise,the uses described in this manual apply toverified balance versions (indicated by thesuffix -.0CE in the model number), aswell as the standard version.
Combination of Several ApplicationsYou can combine the use of variousapplication programs to meet your morecomplicated requirements.
To select application programs one afterthe other, press d (toggle function).
KeysYou can operate the balance either by
using the keys on the display andcontrol unit or from an on-line PC.This manual describes operation usingthe balance keys.
Labeled KeysThese keys always have the functionindicated by their label, but are notavailable at all times. Availability of theirfunctions depends on the currentoperating status of the balance and themenu settings.
Meaning
a Alphabetic keysPlease see section on Text Input
e On | off keyTurns the balance on and offor switches it to the standby mode
oMenu settingsAccesses and exits the Setup menu
d Toggles to the nextapplication program
c Clear functionDeletes keypad input
Interrupts a calibration andadjustment routine in progressQuits application programs
Operating Design
ME215/235/254/414/614:J Turns the ionizer on and off
ME36S, ME5, SE2:JDisplays device information
rPrint keyOutputs displayed valuesor data logs to the serialcommunications and | orprinter port
.Enters a decimal point
1 ... 9 0 keysSee the section on Numeric Input
w Tares the balance
K, O, P Opens | closes the draft shield
Numeric InputTo enter numbers: press1 9 0 .
To store numbers entered: press the
corresponding function key directlybelow the soft key label
To delete an entire numeric input digitby digit: press the c key
Text Input To enter numbers:
see the section on Numeric Input
To enter letters or characters:press the a key
> Letters are displayed in the bottom linefor selection
To select a different letter:
press the corresponding soft keyto change the letter shown
To select the letter | character shown:press the corresponding function keybelow the soft key label
> The selected letter is shown onthe display
$ Enter the next letter | character,if desired, as described above
$ To exit the letter input mode (e.g., if thelast character entered is a letter): press
the a key
To store a word: press thecorresponding function key (soft key),such as ID
To delete an input character by character:press the c key
To delete user data: enter . ora space and save
7/25/2019 MAN-ME_SE-e
9/173
9
Function Keys (Soft Keys)The current function of soft keys isindicated in the bottom line of the display
(footer).
Texts (abbreviations) or symbols can bedisplayed.
Texts (Examples)Cal: Start calibration | adjustmentS ID: Save ID
The function keys are numbered from right(F1) to left (F6).
SymbolsThe bottom line shows the followingsymbols:ooBack to the initial state
(in the Setup menu: exit Setup)
o Go to the higher selection level
O Show sub-items under the active item
Q Move upward in the input |output window
q Move downward in the input |output window
l Set the selected menu parameter
F6 F5 F4 F3 F2 F1
Operation
Display for Weights andCalculated ValuesThis display is subdivided into 9 areas.
Line for Metrological Data:When the balance is used in legal metrol-ogy, the following metrological specifi-cations of the balance are shown here:
Max Maximum capacity (upper rangelimit) of the balance
Min Minimum capacity (lowerrange limit) of the balance
e Verification scale interval
d Readability | scale interval
On standard balances, onlyMax
andd
are displayed.
Operating the Balance as a LegalMeasuring Instrument (Legal forTrade)
Remove the cover plate from the backof the balance housing
Move switch1 in the directionof the arrow
> Switch left:Standard/not legal for tradeSwitch right:
Legal for trade
> Note:Do not move switch 2
2
1
Line for metrological dataBar graph
Measured value line
Text lineSoft key labels
Plus/minus sign UnitStability indicator Tare memory
Calculated valueApplication pictograms
Bar Graph:The bar graph indicates how much of thebalances capacity is used up by the
current load; during checkweighing, itindicates the control limits.
The following symbols may be displayed:
0% Lower load limit
100% Upper load limit
Bar graph showing 10% intervals
- Minimum for checkweighing
= Target for checkweighing
+ Maximum for checkweighing
Plus/Minus Sign, Stability Symbol:A plus or minus sign (F or H) is shownhere for a weight (or a calculated value,such as that for counting), or the Ssymbol indicating that a verified balance*has been zeroed or tared.
Line for Measured Values:This area shows the weighed or calculatedvalue and the alphanumeric input.
Unit and Stability:
When the balance reaches stability, theweight unit or calculated unit is displayedhere.
When the a symbol is displayed here,the value indicated in the readout cannotbe used in legal metrology.
* = Verification scale interval e =scale interval d
Operating Design
7/25/2019 MAN-ME_SE-e
10/173
10
Display for Menu ParameterSettings (Setup)This display is divided into three sections.
Status Line:The status line of shows the functionof the display screen page. In the Setupmenu, the current menu path isshown here.
Setup Menu Example: Balance/scale
functions:
Input and Output WindowThis window contains either detailedinformation (e.g., on the activeapplication) or a pick list. A selecteditem is displayed inversely (whitecharacters on a black background).You can also enter information in anactive field in this window using thealphabetic and numeric keys.
Setup Menu Example,Device parameters, Adapt filter:
The following symbol may be displayedin the input and output window:
d this symbol marks the savedmenu setting
Soft Key LabelsSee the description Function Keys(Soft Keys) on the previous page
Minimum vibration
o Normal vibration
Strong vibration
Extreme vibration
SETUP BAL.FUNC.
Line for Operating State
Input and Output Window
Soft Key Labels
To set a parameter: Press the Q or q soft key repeatedly
until the desired setting is selected
(displayed inversely)
Confirm your selection:press the l soft key
To change the numeric value ofa parameter:
Press the Q or q soft key repeatedly,if necessary, until the desired setting isselected (displayed inversely)
Enter a new value or character:use the 0 1 9 . keysor the a key and enterthe desired letters
Confirm your selection:press the l soft key
To exit Setup: press the oo soft key
Tare Memory, Calculated Values:The symbols displayed here indicate whenthere is a value in one of the tare
memories or when the value shown is aresult of calculation rather than directmeasurement.
These symbols are as follows:
a Calculated value
s1 Net value | tare memory
s2 used by an application program(e.g., formulation, second tarememory)
Application Pictograms:The pictograms displayed here indicatethe application(s) selected. The pictogramis displayed inversely (white on a blackbackground) when the correspondingapplication is active.
For example, the following symbols may bedisplayed simultaneously:
A The counting application is active
H Checkweighing is also active
S Print
T Data record
Text Line:Additional information is displayed here(e.g., operator guidance prompts, name ofthe active program, etc.)
Soft Key Labels:The current functions of the soft keysabove the function keys (arrow keys) areindicated here; during calibration | adjust-ment, this line shows up- and down-arrows(Q and q) for selecting calibration andadjustment functions.
7/25/2019 MAN-ME_SE-e
11/173
11
Input
Bar Code Scanner or Keyboard Input
You can use a bar code scanner or anexternal keyboard to input alphanumericvalues. These inputs are processed inthe same manner as keypad inputs onthe display and control unit of the bal-ance. Bar code and keyboard inputs areonly displayed; they cannot activate anyfunction.
To assign a bar code scanner orkeyboard input to a function, press oneof the following soft keys:
Lot
Samples Measured values Sample number Tare value Initial weight Backweighed value Sample ID
Foot or Hand Switch InputYou can connect a foot switch ora hand switch to the balance to havethis device perform a keypadfunction (such as c or w).
PC InputYou can use a computer to control thefunctions of the balance and displayand control unit via the communica-tions port (see the Data OutputFunction section in the chapterentitled Operating the Balance).
Data Output
The balance provides two interfaceports for outputting weights, calculatedvalues and parameter settings:
Serial communications port(PERIPHERALS Serial I/O)
Serial printer port(PRINTER Serial Out)
Serial Printer PortIn addition to Sartorius printers (such asthe YDO03-0CE), you also have thechoice of connecting a remote displayor an external checkweighing display to
the printer port.You can configure the data outputfunctions in the Setup menu to meetyour various requirements, includingISO | GLP requirements.
ISO: International Organizationfor Standardization
GLP:Good Laboratory Practice
You can have printouts generatedautomatically, or by pressing r;generation can be dependent on or
independent of the stability or timeparameters.
See the section on Data OutputFunctions in the chapter entitledOperating the Balance for a detaileddescription.
Serial Communications PortYou can connect a PC, a remote display,an external checkweighing displayor a standard (non-verifiable) printerto this port.
Request messages are sent via theinterface to initiate functions in theweighing cell and in the display andcontrol unit. Some of the functionsgenerate response messages.
See the section on Data OutputFunctions in the chapter entitledOperating the Balance for a detaileddescription.
Error CodesIf you press a key that has no function,or which is blocked at a certain point in
an application program, this error isindicated as follows:
a double beep is sounded as an acousticsignal if the key has no function
a double beep is sounded and themessage No function is displayed inthe text line if the key function is notavailable at that time
The response to an operator error isidentical in all operating modes. See thechapter entitled Error Codes for adetailed description.
Storing SettingsSaving Parameter SettingsThe settings configured remain storedin the balances non-volatile memory. Inaddition, you can reload the factory set-tings.
Saving SettingsUnder Setup > Device parameters >password you can assign passwords inorder to block access to:
Balance | scale functions
Device parameters
Application parameters
Printout
Factory settings
7/25/2019 MAN-ME_SE-e
12/173
Storage and Shipping ConditionsAllowable storage temperature: 5C to 40C (41F to 104F)The packaging has been designed to ensure that the A/D converter will not be
damaged even if it is dropped from a height of 80 cm (approx. 31 inches). Do notexpose the balance to extreme temperatures, moisture, shocks, blows or vibration.
Unpacking the Balance
Lift the inner package containing the balance out of the outer packaging by the strap.
Loosen and remove the strap.
Remove the cardboard sleeve.
12
Getting Started
7/25/2019 MAN-ME_SE-e
13/173
13
Remove the following parts from the recessed spaces at the top of theinner packaging:
AC adapter (in cardboard packaging)
Weighing pan Shield plate
Remove the two padding blocks that make up the inner packaging by pulling outward.
Remove the shield disk (included with ME235S/P models only) and dust cover.
Remove the retainer securing the front draft shield panel.
Open the plastic wrapping.
Place one hand under the display and control unit and the other under the back of thebalance and lift the balance out of the lower packaging.
7/25/2019 MAN-ME_SE-e
14/173
14
! Do not lift the balance by the draft shield or the front panel, as this can resultin damage.
Set up the balance at the place of installation. Open the draft shield doors.
Remove the retainers (1)securing the display and control unit and remove the frontdraft shield panel.
Remove the foam padding (2)from the draft shield.
! Save the box and all parts of the packaging in case it should become necessary totransport the balance over a long distance. Only the original packaging provides thebest protection for shipment (see also Transporting the Balance on page 20).Before packing the balance, unplug all connected cables to prevent damage.
Equipment SuppliedThe following individual componentsare supplied:
ME215/235/254/414/614 Balance AC adapter with power cord Weighing pan with hanger for below-
balance weighing
Shield disk Dust cover for the balance housing Dust cover for the display and control
unit Instruction manual
ME5, SE2 Weigh cell Draft shield Electronic evaluation unit Connecting cable AC adapter with power cord Accessories kit
The accessories kit includes:
Weighing pan Shield disk Interior draft shield (only for SE2) Brush Forceps Cloth
ME36S Weigh cell Electronic evaluation unit AC adapter with power cord Weighing pan
Shield disk
ME5-F, SE2-F Weigh cell Draft shield cover Shield ring Electronic evaluation unit Connecting cable AC adapter with power cord Accessories kit
The accessories kit includes: Filter pan, 50 mm d Weighing pan Shield disk
Interior draft shield (only for SE2) Brush Forceps Cloth
1
1
2
7/25/2019 MAN-ME_SE-e
15/173
15
Installation InstructionsThe ME/SE balances are designed to provide reliable weighing results under normal ambientconditions in the laboratory and in industry. Choose the right location to set up your balance
by observing the following so that you will be able to work with added speed and accuracy: Set up the balance on a completely even surface on a low-vibration balance table or
wall console Avoid placing the balance in close proximity to a heater or otherwise exposing the
balance to heat or direct sunlight, as this can considerably increase the temperatureinside the draft shield (greenhouse effect), resulting in incorrect readouts dueto convection currents, turbulence and buoyancy effects.
Protect the balance from drafts that come from open windows or doors Avoid brief fluctuations in room temperature Protect the balance from aggressive chemical vapors Do not expose the balance to extreme moisture
Linearization after TransportAfter transporting the balance, its linearity may be outside the allowable tolerances(please refer to the Specifications in the Overview chapter). After transporting thebalance, be sure to perform internal linearization. Repeat this process to obtain optimalaccuracy. For directions on this procedure, please refer to the section on Linearization.
Conditioning the Balance
Moisture in the air can condense on the surfaces of a cold balance whenever it is broughtinto a substantially warmer place. If you transfer the balance to a warmer area, makesure to condition it for about 2 hours at room temperature, leaving it unplugged fromAC power. Afterwards, if you keep the balance connected to AC power, the continuouspositive difference in temperature between the inside of the balance and the outsidewill practically rule out the effects of moisture condensation.
Setting Up the Balance ME215/235/254/414/614 Place the components listed below inside the weighing chamber in the order given: 1) Shield plate 2) Position the weighing pan and turn to the left or right until it snaps into place 3) Shield disk (ME235S/P only)
$ Note: the shield disk minimizes the effects of drafts within the weighing chamber
Remote Operation of the Display and Control Unit Unplug the cables, turn the balance on its side and lay it on a padded surface
to avoid damaging the weighing system and draft shield
Use an Allen wrench to remove the three fastening screws
Remove the display unit and attach the connecting cable
> Length of the connecting cable: 44 cm (17 inches)
$ For information on longer cables, please see the Accessories section
$ If you wish to use a longer cable, it must be installed by authorized
Sartorius service technicians
3
2
1
ME215/235/254/414/614:
7/25/2019 MAN-ME_SE-e
16/173
16
Setting Up the ME36S Balance Place the components below in the weighing chamber in the order given:
Shield plate
Weighing pan
Setting Up the ME5 or SE2 Balance Place the components below on the weigh cell base in the order given:
Shield disk
Weighing panPlease note: after placing the weighing pan on the base, turn the pan slightly towardsthe left then the right while pressing down on it to lock.
Interior draft shield (only for SE2)
Draft shield: center the hole over the pan (see arrows)
Setting Up the ME5-F or SE2-F Balance Place the components below on the weigh cell base in the order given:
Shield disk
Interior draft shield ring
50 mm dia. filter pan or weighing pan (or, optionally, the 75 or 90 mm dia. filter pan)Please note: after placing the pan on the base, turn the pan slightly towards the leftthen the right while pressing down on it to lock
Interior draft shield (only for SE2)
! If the weighing pan is removed during operation, turn the balance off and then on againafter you return it to the balance.
Draft shield cover
Setting Up the Filter Balance for Left-handed Persons:
Remove the draft shield cover
Detach the pin on the right and re-attach on the left
ME36S:
ME5, SE2:
ME5-F, SE2-F:
7/25/2019 MAN-ME_SE-e
17/173
17
Turn draft shield assembly by approx. 90 degrees toward the left(loosen knurled knob to turn)
Connect the weigh cell to the evaluation unit Use a screwdriver to tighten the screws to the female connector on the weigh cell
Connecting the Balance to AC PowerThe wide-range AC adapter is designed for 100 V to 240 V.
Check the plug design of the power cord If it does not fit your wall outlet (mains supply), please contact your
Sartorius office or dealer
Use only Original Sartorius AC adapters and power cords
AC adapters with a registered approval rating from a national testing laboratory
$ To use a main feeder cable from the ceiling or to mount a CEE plug,have a certified electrician install it
$ To use an external rechargeable battery pack, refer to the Accessories in theOverview chapter
Insert the AC adapter plug with the angle facing downward into the jack on the balance
7/25/2019 MAN-ME_SE-e
18/173
18
Plug power cord into the AC adapter
To power the balance with AC current, plug the power cord into
a wall outlet (mains supply)
Charging the Rechargeable Battery for Saving Data:All data is saved in the battery-backed memory. When initially operating the balance,leave it connected to AC power for one day to charge the battery. When the balanceis disconnected from AC power, the balance-generated data will remain stored forapproximately three months. In the standby mode, data is retained in the memory viathe power supply. Be sure to print out data before storing your balance for a relativelylong period.
Safety PrecautionsThe AC adapter rated to Class 2 can be plugged into any wall outletwithout requiring any additional safety precautions. The ground or earth terminal isconnected to the scale housing, which can be additionally grounded, if required.The data interface is also electrically connected to the balance housing (ground).
Note:This equipment has been tested and found to comply with the limits for a Class A digitaldevice, pursuant to Part 15 of the FCC rules. These limits are designed to providereasonable protection against harmful interference when the equipment is operated ina commercial environment. This equipment generates, uses and can radiate radiofrequency energy and, if not installed and used in accordance with the instructionmanual, may cause harmful interference to radio communications. Operation of thisequipment in a residential area is likely to cause harmful interference in which case theuser will be required to correct the interference at his own expense. Changes ormodifications not expressly approved by Sartorius AG could void the users authorityto operate the equipment.
Connecting Electronic Peripheral Devices Make absolutely sure to unplug the balance from AC power before you connect or
disconnect a peripheral device (printer or PC) to or from an interface port
Warmup TimeEach time you move your balance to another location, you must condition it for at least12 hours to the new location. To deliver exact results, the balance must warm up for atleast 12 hours after initial connection to AC power. Only after this time will the balance
have reached the required operating temperature.
Using Balances Verified as Legal Measuring Instruments in the EU*:$ The balance must warm up for at least 24 hours after initial connection
to AC power
$ Warmup time each time power is turned on the ME614S-0CE, ME414S-0CE:at least 30 minutes
$ Always wait for the power-on adjustment routine to be completed:for requirements see page 63.
* including the Signatories of the Agreement on the European Economic Area
7/25/2019 MAN-ME_SE-e
19/173
19
Antitheft Locking DeviceTo fasten an antitheft locking device, use the lug located on the rear panelof the balance.
Secure the balance at the place of installation, e.g., with a chainor a lock
Leveling the BalancePurpose:
To compensate for unevenness at the place of installation
To achieve perfectly horizontal positioning of the balance for consistent repeatabilityof the weighing results
Always level the balance again any time it has been moved.
Only the 2 front feet are used for leveling.
Turn the leveling feet as shown in the diagram until the air bubble is centered exactlywithin the circle of the level indicator
> Several leveling steps are usually required
Setting the Language> See the section on Setting the Language in the chapter Configuring the Balance
Setting the Date and Time> See the example on page 25, in the chapter Configuring the Balance
7/25/2019 MAN-ME_SE-e
20/173
$ The balance must be disconnected from power before preparing it for shipment.Disconnect the AC adapter and all interface cables from the balance.
Transport Over Short Distances Place one hand under the display and control unit and the other under the back of the
balance. Lift the balance carefully and carry it to the new location.
$ Avoid subjecting the balance to vibration or shocks (impact).
! Do not lift the balance by the draft shield or the front panel, as this can result
in damage.
Transport or Shipping Over Long DistancesMake sure you use all components of the original packaging in the following cases:
for transport or shipping over long distances. if it is not certain that the balance will remain upright during transport or shipping.
Remove the following parts: Shield disk (included with ME235S/P models only) Weighing pan and shield plate:
Reach beneath the shield plate and lift it carefully, together with the weighing pan,to avoid damaging the weighing system.
Open the draft shield doors and carefully position the balance in the lower packaging. Press the foam padding against the housing.
20
Shipping the Balance
7/25/2019 MAN-ME_SE-e
21/173
21
Attach the retainers to secure the front draft shield panel and the display andcontrol unit.
Place the dust cover on the display and control unit. Models ME235S/P only: Place the shield disk in a bag and lay it on the dust cover.
Position the two padding blocks that make up the inner packaging and press inward.
Place the following parts in the recessed spaces at the top of the inner packaging: Shield plate Weighing pan AC adapter (in cardboard packaging)
Wrap the cardboard sleeve around the inner packaging.
7/25/2019 MAN-ME_SE-e
22/173
Place the strap around the package and tighten it.
Lift the packaged balance by the strap and place it in the shipping box, with the
bottommost padding already in the box.
Place the uppermost padding on top of the inner package.
Close the shipping box and seal it appropriately for the intended transportor shipment.
22
7/25/2019 MAN-ME_SE-e
23/173
23
Configuring the Balance
PurposeYou can configure your ME/SE balanceto meet individual requirements by
entering user data and settingparameters in the Setup menu.
The Setup menu is divided into thefollowing items:
Balance/scale functions Device parameters Application parameters Printout Device information Language Factory settings
Setting the LanguageYou can choose from 5 languages forthe information display:
German English (factory setting) English with U.S. date | time format French Italian Spanish
Example: Selecting the Language U.S. Mode
Step Press key(s) (or follow instructions) Display | Printout
1. Select Setup menu o
2. Select Language Repeatedlyand confirm press q soft key,
thenO soft key
3. Select U.S. mode q soft key
4. Save language l soft key
5. Exit the Setup menu oo soft key
SETUP LANGUAGE
Deutsch
oEnglish
U.S.-Mode
Franais
Italiano
oo o q l
SETUP LANGUAGE
Deutsch
English
oU.S.-Mode
Franais
Italiano
oo o q l
7/25/2019 MAN-ME_SE-e
24/173
24
Navigating in the Setup Menu (Examples):
Example: Adapt the balance to Extreme vibration by selecting this setting
Step Press key(s) (or follow instructions) Display | Printout
1. Select Setup menu o
2. Confirm Balance/scale functions O soft key
3. Select menu item Adapt filter q soft key, thenand confirm O soft key
4. Select menu item Extreme vibration q soft key
5. Confirm menu item l soft key
Extreme vibration
6. If required, select further menu items q Q soft keys
7. Save setting and exit Setup menu oo soft key
Exiting the Setup MenuIf you use the oo soft key:
The software will be restarted if you have changed a setting. The software will not be restarted if you have kept the same settings.
In this case, the program will return to its initial state before you pressthe o key.
If you press the o key: When you exit o, the software is generally restarted.
7/25/2019 MAN-ME_SE-e
25/173
25
Example: Entering the time and date
Step Press key(s) (or follow instructions) Display | Printout
1. Select Setup menu; o, thenselect Device parameters q and O soft keys
2. Set clock press q repeatedly,then press O
3. Enter the time 1 1 . 1 2
. 3 0
4. Set the time according toyour local clock l soft key
5. Enter the date 1 3 . 0 3. 00
6. Store the date l soft key
7. Enter other data, if desired q Q soft keys
8. Exit Setup menu oo soft key
* = not on ME36S, ME5, SE2
7/25/2019 MAN-ME_SE-e
26/173
26
PurposeThis menu item enables you toconfigure the balance functions, i.e.,
to meet individual requirements byselecting predefined parameters in theSetup menu. You can block access tothe menu by assigning a password.
FeaturesThe balance functions are combined inthe following groups(1st menu level):
Calibration | adjustment Adapt filter Application filter Stability range Taring Auto zero Weight unit 1 Display accuracy 1 Tare/zero with power on Factory settings: only wgh. param. (only
the balance functions)For legal metrology, the selectionof individual parameters is limited.
Factory SettingsParameters: The factory settings areidentified by the symbol o in the liststarting on page 27.
PreparationShow available balance functions:
Select Setup menu: press the o key
> SETUP is displayed
Select Balance functions:press the O soft key
If you already assigned a password:
> The password prompt is displayed
$ If access is blocked by a password:enter the password using the numeric |alphabetic keys
$ If the last character of the password isa letter: conclude input by pressing a
Confirm your password and havethe balance functions displayed:Press the l soft key
> Balance functions are displayed:
$ To select the next group:press the q soft key (down arrow)
$ To select the previous item ofa group: press the Q soft key(up arrow)
$ To select the next sub-item withina group:press the O soft key (right arrow)
$ To select the previous group:press the o soft key (left arrow)
$ To confirm: press the l soft key
Extra Functions Exit the Setup menu:
press the oo soft key
> Restart your application
Print parameter settings: When the balance functions are
displayed, press r> Printout (example)
Texts with more than20 characters are cut off
SETUPBAL. FUNC.
--------------------Calibration/adjustmCAL/isoTST key fun
Selection modeCal/adjustment seq
Calibrate, then autoadjust
isoCAL functionOn without resetting
app.Start automatic ad
isoCALPrint GLP/GMP adju
Automatic if GLP isselected
Parameter for exteWt. ID (W ID):
Cal./adj. wt:200.00000 g
Adapt filterNormal vibration
Application filterFinal readout
Stability range2 digits
Auto zeroOn
Weight unit 1Grams /g
Display accuracy 1All digits
--------------------
Setting the Balance Functions (BAL.FUNC.)
7/25/2019 MAN-ME_SE-e
27/173
27
Balance Functions (Overview) factory setting user-defined setting(s)
Setup Calibration | Cal | isoTST External calibration | adjustment** withBalance Adjustment key function factory-defined calibration weightfunctions External calibration | adjustment**;
user-defined weightInternal calibration | adjustmentInternal linearizationCal key blockedreproTEST
Selection mode for cal. | adjustment
Calibration | Calibrate, then auto adjustadjustment sequence Calibrate, then manual adjust
isoCAL function Off Only adjustment promptisoCAL and linearization onOn and reset application
On without resetting application
Start automatic User-defined adjustment Enter adjustment time 1: (hh.mm)adjustment times Enter adjustment time 2: (hh.mm)
Enter adjustment time 3: (hh.mm) isoCAL
Print GLP Automatic if GLP is selectedadjustment record On request,
from record memory
Parameter for Weight set ID (W ID): Enter 14 characters max.external weight Calibration |
adjustment weight: Enter exact weight
Adapt filter Minimum vibration(ambient Normal vibrationconditions) Strong vibration
Extreme vibration
Application filter Final readoutFilling mode
Stability range 14 digit12 digit1 digit
2 digits4 digits8 digits*
* = not applicable to verified balances** = only external calibration is possible for verified balances
Factor
ysetting
Factor
ysetting
7/25/2019 MAN-ME_SE-e
28/173
7/25/2019 MAN-ME_SE-e
29/173
29
Setting the Device Parameters (Device)
PurposeThis menu item enables you toconfigure the balance, i.e., to meet
individual requirements by selectingpredefined menu parameters in theSetup menu. You can block access tothe menu by assigning a password.
FeaturesThe device parameters are combined inthe following groups(1st menu level):
Draft shield
Ionizer*
Password
User ID
Clock
Interfaces
Display
Keys
Extra functions
Factory settings: only device parameters
Factory SettingsParameters: The factory settings areidentified by the symbol in the liststarting on page 31.
* = not on ME36S, ME5, SE2
PreparationDisplay available device parameters
Select the Setup menu:press o
> SETUP is displayed:
Select Device parameters:use the q and O soft keys
If no password has been assigned,anyone can access the Setup menudevice parameters
If a password has already been assigned:
> The password prompt is displayed
$ If access is blocked by a password: enterthe password using the numeric and |or alphabetic keys
$ If the last character of the passwordis a letter:conclude input by pressingthe a key
Press l to confirm the password
> Device parameters are now displayed:
$ To select the next group:press the q soft key (down arrow)
$ To select the previous menu itemof a group: press the Q soft key(up arrow)
$ To select the next sub-item within
a group: press the O soft key(right arrow)
$ To select the previous group:press the o soft key (left arrow)
$ Press l to confirm the selectedmenu item
Entering or Changing a Password
Lets assume that a password with 8
characters max. has already beenassigned to access the Setup deviceparameters
Select the Setup menu:press o
> SETUP is displayed
Select device parameters:Use the q and O soft keys
If you have already assigneda password:
> The password prompt is displayed:
$ Enter the password
$ Press the l soft key to confirmyour password and view the deviceparameters
Write down your password here for easyreference:
Password = .............................
If you assign a password and then for-get what the word is:
$ Enter the General Password(see Appendix)
$ Press the l soft key to confirm and dis-play the password
> The device parameters are displayed
Select the device parameter Password:If necessary, repeatedly press q or Qand O, until you see
> Password: and any existing password
7/25/2019 MAN-ME_SE-e
30/173
30
New password: Enter the numbersand/or letters for the new password(8 characters max.)
If none is displayed, this meansno password has been assignedTo delete the user password:Press . or c and confirm
To confirm:press the l soft key
Exit the Setup menu:press the oo soft key
> Restart the application
Extra Functions
Exit the Setup menu:
press the oo soft key
> Restart the application
Print the parameter settings: If the device parameters
are displayed: press r
> Printout (example)
--------------------SETUP
DEVICE--------------------Draft shieldLeft/right keySame function
Automatic modeOff
Weight resolutionShow all decimal pla
cesIonizerOnAuto-off time:
10 secUser IDUser ID:
InterfacesSerial communicatiSBIBaudrate
1200 baudNumber of data b
7 data bitsParity
OddNumber of stop b
1 stop bitHandshake mode
Hardware handshakeafter 1 char
Serial printer (PR
YDP03Baudrate1200 baud
ParityOdd
Handshake modeHardware handshake
after 1 charFunction: external
Print keyFunction: control
OutputDisplayContrast
2
etc.
7/25/2019 MAN-ME_SE-e
31/173
31
Device Parameters (Overview) factory setting user-defined setting(s)
Setup Deviceparameters
Draft shield Keys Same functionLeft/right key Separate function
Off
Automatic mode OffClose >function > openClose >
perform function
Weight Show all decimal placesresolution Do not show last decimal placewith door open
Ionizer* Off On Auto-off time: 10 s 1 .. 99 seconds
Password: None Enter 8 charactersmax.
User ID: None Enter 20 charactersmax.
Clock Time: Enter
hh.mm.ss
Date: Enter dd.mm.yyor mm.dd.yy (U.S. mode only)(01.01.97)
Interfaces Serial SBI Baud rate 150 baudcommunication 300 baud(PERIPHERALS) 600 baud
1200 baud2400 baud4800 baud9600 baud19,200 baud
Number of data bits 7 bit 1)
8 bit
Parity Space 2)
o OddEvenNone3)
Number of stop bits 1 stop bit2 stop bits
Handshake mode Software handshake Hardware handshake,
1 character after CTSxBPI Network address: Enter any number from 0 to 31
For the display, keys and extra functions, see next pages
* = not on ME36S, ME5, SE21) not if None parity is selected2) only if 7 data bits selected3) only if 8 data bits selected
Factor
ysetting
Factor
ysetting
Factor
ysetting
Factor
ysetting
7/25/2019 MAN-ME_SE-e
32/173
32
Setup Deviceparameters
Interfaces Serial YDP01ISprinter(PRINTER) YDP02 see YDP03 (not with 19,200 baud)
YDP03 Baud rate 1200 baud2400 baud4800 baud9600 baud19,200 baud
Parity Space Odd
Even
Handshake Software handshakemode Hardware handshake,
1 character after CTS
YDP01IS Label(label printer)
YDP02IS
YDP02IS Label(label printer)
Universal Baud rate 150 baud300 baud
600 baud1200 baud2400 baud4800 baud
9600 baud19,200 baud
Number of data bits 7 bits 1)
8 bits
Parity OddEven
None 2)
Number of stop bits 1 stop bit
2 stop bits
Handshake Software handshake modeHardware handshake,1 character after CTS
YDP04IS
YDP04IS Label
For the display, keys and extra functions, see next page
1) not if None parity is selected2) only if 8 data bits selected
Factor
ysetting
Factor
ysetting
7/25/2019 MAN-ME_SE-e
33/173
33
Setup Deviceparameters
Interfaces Function: Print key rexternal switch Tare key
Calibration keyF1 function key(Right soft key)Clear function key cF2 function key(2nd soft key from the right)Bar code scanner/extra keyboardIonizer keyRight draft shield keyLeft draft shield key
Function: control ports Input Output
Display Contrast Select contrast (0 to 4): 01
o 234
Background WhiteBlack
Digit size 10mm + bar graph + text display13mm + bar graph
13mm + text display13mm
Application Offsymbols On
Keys CF function Clear all applicationsin application Clear only selected applications
CF function Delete entire inputfor input Delete last character
Block key All keys unblockedfunctions All blocked except
for o, e
Alphanumerickeys blocked
Extra functions Acoustic signal OnOff
Power-on Off/on/standbymode Auto on
Factory settings: Noonly Yesdevice parameters
Factor
ysetting
Factor
ysetting
7/25/2019 MAN-ME_SE-e
34/173
7/25/2019 MAN-ME_SE-e
35/173
35
Application Parameters (Overview) factory settingsuser-defined setting(s)
Setup Application 1 Off (weighing only)Application (Basic settings)parameters Toggle wt. units Weight unit 1 Grams /g
Kilograms /kg*Carats /ctPounds /lb*Ounces /oz*Troy ounces /ozt*Hong Kong taels /tlh*Singapore taels /tls*Taiwanese taels /tlt*Grains /GN*Pennyweights /dwt*
# Milligrams /mgParts per pound //lb*Chinese taels /tlc*Momme /mom*Austrian carats /K*Tola /tol*Baht /bat*Mesghal /MS*
Display accuracy 1 All digitsFewer for weightchangesLast digit off*
Weight unit 2 # Grams /gKilograms /kg*Carat s/ctPounds /lb*Ounces /oz*Troy ounces /ozt*Hong Kong taels /tlh*Singapore taels /tls*Taiwanese taels /tlt*Grains /GN*
Pennyweights /dwt* Milligrams /mgParts per pound //lb*Chinese taels /tlc*Momme /mom*Austrian carats /K*Tola /tol*Baht /bat*Mesghal /MS*
Display accuracy 2 All digitsFewer digits forweight changeLast digit off*
For counting through air buoyancy correction, see the following pages
Applications 2 and 3 see following pages
* not applicable to verified balances# = factory setting on ME36S, ME5, SE2
Factor
ysetting
Factor
ysetting
7/25/2019 MAN-ME_SE-e
36/173
36
Setup Application 1 Counting Accuracy Average Display accuracyApplication (Basic settings) piece weight calculation + 1 decimal placeparameters (resolution) + 2 decimal places
Average piece Off weight updating Manual
Automatic
Percent weighing Weight storage accuracy Display accuracy(resolution) + 1 decimal places
+ 2 decimal places
Decimal places for Nonereadout in percent 1 decimal place
2 decimal places3 decimal places4 decimal places5 decimal places6 decimal places7 decimal places
Display calculated value ResidueLossRatio 1 (DR)Ratio 2 (OR)
Animal weighing* Animal activity Calm(averaging) Normal
Active0.1% of the animal/object0.2% of the animal/object0.5% of the animal/object1% of the animal/object2% of the animal/object
5% of the animal/object10% of the animal/object20% of the animal/object50% of the animal/object100% of the animal/object
Start Manual mode Automatic mode
Minimum load for automatic Nonestorage 10 digits
20 digits50 digits
100 digits200 digits500 digits
1000 digits
Decimal places in Noneresult display 1 decimal place
2 decimal place3 decimal places4 decimal places5 decimal places6 decimal places7 decimal places
Printout None Average weight only
Average and calculated value
Recalculation*
For calculation through air buoyancy correction, see following pages
Applications 2 and 3 see following pages
* = How to run this application is described in detail in our Masterpro LA. Installation and Operating Instructions.Request your copy directly from Sartorius or download it from the Internet (www.sartorius.com; see download)
Factor
ysetting
7/25/2019 MAN-ME_SE-e
37/173
37
Setup Application 1 Calculation Decimal places in NoneApplication (Basic settings) calculated result 1 decimal placeparameters 2 decimal places
3 decimal places4 decimal places5 decimal places6 decimal places7 decimal places
Density Method Density of liquid Buoyancy
DisplacementPycnometer
Liquid causing Waterbuoyancy Ethanol
User-definable
Decimal places Nonefor disp. of vol./ 1 decimal placedensity 2 decimal places
3 decimal places4 decimal places
Printout NoneAll data
Differential weighing Weighing sequence* Individual weighingConsecutive individual weighing
Combined weighingSerial weighing
Tare weighing No Yes
Result with Nonedecimal point 1 decimal place
2 decimal places3 decimal places4 decimal places5 decimal places6 decimal places7 decimal places
Autosave Offvalues On, first value at stability
On, last value at stability
On, value bet.70 130% at stabil.
Minimum load Nonefor autosave 10 digits
20 digits50 digits100 digits200 digits500 digits1000 digits
Save statistics NoYes
Air buoyancy correction: See next page
Applications 2 and 3: see following pages
* = Setting can only be changed when the program is initially run and when the Wg.seq. key option is set to No
Factor
ysetting
7/25/2019 MAN-ME_SE-e
38/173
38
Setup Application 1 Differential weighing Generate printout NoneApplication (Basic settings) Automatic after backweighingparameters Auto. after initial weigh.
and backweighingAutomatic after tare-, initial-and backweigh.
Include sample ID in text line NoYes
Wg.seq key No Yes
Clear sample after individual Noweight, result + and unload Yes
Last residual weight is stored No
as initial weight (ashing) Yes
Air buoyancy correction Air density determination NoYes
Change steel/aluminum Noreferences Yes
Diameter determination
Application 2 Off(Control functions) Checkweighing* Activation of port lines Within checkweighing range
Always onStability andcheckweighing rangeAt stabilityStability + checkweigh.range on
Type of checkweighing input Target, min., max. weightMinimum, maximum weightTarget, min. in %, max. in %
Weight display mode Absolute valueDifference from the target
Automatic printout Yesof OK values No
Time-controlled Function after time interval Beep (acoustic signal)functions Lock in readout Automatic printout of values
Store value in applicat. 3 memory(totalizing, formulation, statistics)
Automatic function restart OnOff
Storage mode Without stabilityAfter stabilityAfter higher stability
Print then tare Off On
* = How to run this application is described in detail in our Masterpro LA. Installation and Operating Instructions.Request your copy directly from Sartorius or download it from the Internet (www.sartorius.com; see download)
Factor
ysetting
Factor
ysetting
7/25/2019 MAN-ME_SE-e
39/173
39
Setup Application 3 OffApplication (Data records)parameters Totalizing* Automatic storage Off
On, first value at stabilityOn, last value at stabilityOn, value bet. 70% 130%at stability
Minimum load for Noneautomatic storage 10 digits
20 digits50 digits100 digits200 digits500 digits1000 digits
Source of data for auto storage Application 1
Application 2
Evaluated values NetCalculatedNet + calculated
Evaluation mode, MR function Intermediate evaluation, printFinal evaluation, printIntermediate evaluation,display+printFinal evaluation,display+print
M+/M function, then tare OffOn
Printout of individual Nocomponents Yes
Formulation* Automatic storage OffOn, first value at stability
Minimum load for Noneautomatic storage 10 digits
20 digits50 digits100 digits200 digits500 digits
1000 digits
Source of data for Application 1automatic storage Application 2
Evaluated values NetCalculatedNet + calculated
Evaluation mode,MR function Intermediate evaluation, printFinal evaluation, print
Printout of individual Nocomponents Yes
* = How to run this application is described in detail in our Masterpro LA. Installation and Operating Instructions.Request your copy directly from Sartorius or download it from the Internet (www.sartorius.com; see download)
Factor
ysetting
Factor
ysetting
7/25/2019 MAN-ME_SE-e
40/173
40
Setup Application 3 Statistics Automatic storage o Off Application (Data records) On, first value at stabilityparameters On, last value at stability
On, value 70% 130%at stability
Minimum load for Noneautomatic storage 10 digits
o 20 digits50 digits100 digits200 digits500 digits1000 digits
Source of data for o Application 1automatic storage Application 2
Evaluated values o NetCalculatedNet + calculated
Evaluation mode, o Intermediate evaluation, printMR function Final evaluation, print
Intermediate evaluation,display + printFinal evaluation,display + print
M+/M function, then tare o OffOn
Printout of individual Nocomponents o Yes
Extra Functions for other extra functions, see next page
Factor
ysetting
Factor
ysetting
7/25/2019 MAN-ME_SE-e
41/173
41
Setup Extra Functions OffApplication F4 keyparameters Extra Functions 2nd tare memory Container tare weight No
F5 key Yes
Automatic printout Net valueTare/preset tare
Off
Identification Printout Automatic, if configuredcodes Once after pressing r(IDs) if configured
Each time the print key(r) is pressedOnce for M+ function(application 3 memory)
Manual storagein app. 3memory M+ (totalizing,formulation, statistics)
Changing the resolution
Product data memory
SQmin Function Display o Text displayBargraph
Print in o OffGLP Header On
DKD uncertainty Display
1)
U (absolute)of measurement U* (relative)PG (process accuracy)
Print1) U (absolute)U* (relative)PA (process accuracy)
Process accuracy Input: 1100,5 decimal places
Auto-start Onapplication when Offpower goes on
Factory settings Noonly for Yesapplication-parameters
1) an asterisk (*) indicates an activated menu item. You can select up to 3 items.
Factor
ysetting
7/25/2019 MAN-ME_SE-e
42/173
Selecting the Printout Function (Printout)
42
PurposeThis menu item enables you toconfigure the printout to meet your
individual requirements by selectingpredefined menu parameters in theSetup menu. Printouts of weights andother measured or calculated values andIDs enable you to document your data.You can select the particular data youwish to print. To prevent changes toyour settings, you can block access tothe menu by assigning a password.
FeaturesThe device parameters are combined inthe following groups(1st menu level):
Application-defined output Automatic output of displayed values Output to interface ports Line format ISO/GLP printout Identification # (identifier) Factory settings only printout
Factory SettingsParameters: The factory settingsare identified by the symbol inthe list on the next page.
PreparationDisplay available printout parameters
Select the Setup menu:press o
> SETUP is displayed:
Select Printout:use the q and O soft keysIf no password has been assigned,anyone can access the printoutparameters in the Setup menu
If a password has already been assigned:> The password prompt is displayed
$ If access is blocked by a password: enterthe password using the numeric and/oralphabetic keys
$ If the last character of the passwordis a letter: conclude input by pressingthe a key
Press l to confirm the password
> Printout parameters are now displayed:
$ To select the next group:press the q soft key (down arrow)
$ To select the previous item ofa group: press the Q soft key(up arrow)
$ To select the next sub-item withina group: press the O soft key (rightarrow)
$ To select the previous group:press the o soft key (left arrow)
$ To confirm: press the l soft key
Extra Functions
Exit the Setup menu:
press the oo soft key> Restart your application
Print parameter settings: When the printout parameters
are displayed, press r
> Printout (Example)
--------------------SETUP
PRINTOUT--------------------Application-definedStability paramete
With stabilityPrint on request t
OffAuto print upon in
All valuesConfigured printouIndiv.: Printout
Automatic output ofStability parameteWitout stability
Stop auto printNot possible
Time-dependent aut1 display update
Output to interfaceSerial communicati
Application-definedoutput
Serial printer (PRApplication-defined
outputLine format
For other apps/GLP (22 characters)
ISO/GLP/GMP printouOff
Identification #Lot (L ID):
ID1:ID1
etc.
7/25/2019 MAN-ME_SE-e
43/173
43
Printout Parameters (Overview) factory settinguser-defined setting(s)
SETUP Application- Stability parameter Without stability 1)
Printout defined output o With stability
Print on request, then tare o OffOn
Auto print upon Off initialization o All values
(of the application) Only main values
Configured printout See own chapter startingon the next page
Automatic Stability parameter o Without stabilityoutput At stabilityof displayed Automatic aftervalue weight change 2)
Stop auto print Use print key ro Not possible
Time-dependent o 1 display updateauto print 2 display updates
10 display updates *100 display updates *
Output to Serial communication Offinterface ports (PERIPHERALS) o Application-defined output
Automatic output of displayed value
Serial printer) Off (PRINTER) o Application-defined output
Line format For raw data (16 characters)o For other apps/
GLP (22 characters)
ISO/GLP o Offprintout Only for calibration/adjustmentAlways on
Identification # Lot ID (L ID): Enter 20 characters max.ID1: ID1ID2: ID2ID3: ID3ID4: ID4
Factory settings: Noprintout only Yes
* = changing settings not applicable to verified balances1)
= Information on use in legal metrology: Only permitted for control purposes; printouts are not allowed2) = auto print when load change is > 10 d and stability is reached: no printout until residual difference in load value is < 5 d
Factor
ysetting
Factor
ysetting
7/25/2019 MAN-ME_SE-e
44/173
Printout Configuration
44
PurposeThis menu item enables you toconfigure individual printout formats.
With the formulation, totalizing andstatistics application, you can alsodefine the values to be included onthe total printout when the MR keyis pressed.
Under Setup > Printout > Application-defined output > Configured printout,you can configure individual, compo-nent or total data records that containthe items in each application that areavailable for printouts. Configure theseprintouts after you have configured theapplications, because some entries inthe data record depend on the particu-lar application.
Features Maximum items in a data record: 60
Separate configuration of printoutformats for individual weights,components, total, backweighing andstatistics
Individual printout generation:press the r key
Automatic printout of application data:e.g., results from animal weighing ordensity application (Setup menu: Appli-cation 1: Density: Printout: All data) OKvalues from checkweighing application,time-controlled printouts, 2nd tarememory
Component printout:For results from totalizing, formulationor statistics applications, pressM+ or M- (Setup: Application 3: ...,Printout of individual components: On)
Total printout: For totalizing,formulation or statistics applications,press MR
Backweighing printouts or records:automatically generated after back-weighing or manually by pressing ther
key when the result is displayedat the end of backweighing
Statistics printout or output:To generate, press the r key whenthe statistics are displayed
Printouts for Differential Weighing:These printouts can be generated asstandard or configured (user-defined)reports.
You can configure the followingprintouts:
Individual printout
Backweighing printout
Statistics printouts
Printouts are generated in one oftwo ways:
at the request of the user by pressingthe r key (print on request)
automatically, if configuredin the Setup menu [Applicationparameters: Application 1: Differentialweighing: Generate printout: Auto]
You can turn off automatic printoutgeneration in the Setup menu[Application parameters:
Application 1: Differential weighing:Generate printout: None]
Data records are deleted after you havedswitched to a different application oractivated or de-activated an extrafunction in the application parametersof the Setup menu
A new pick list for a data record iscreated based on the currently activeapplication programs and extrafunctions
Printout items can be deletedindividually
No printout is generated when thefollowing setting is configured:Setup: Printout: Line format:For raw data (16 characters)
Print item Form feed for footer:Advance to beginning of next label inthe YDP01IS-Label and YDP02IS-Label [printer] interface mode
Extra Functions
Exit printout configuration:press oo soft key
> Restart application
Printing Select and List Settings
LIST: print the currently selected listSELECT: printout items that can stillbe selected
When the select bar is onLIST or SELECT:press the r key
> Printout (Example)
BACKW. PRINT.LIST====================Sample dateNet initial wt.Backweighed resLoss in %====================etc.
7/25/2019 MAN-ME_SE-e
45/173
45
Example:Configure an Individual Printout for Counting Application to Include Dotted Line, Date/Time, Piece Count and Net Weight
Settings (changes in the factory settings required for this example):Setup: Application parameters: Application 1: CountingExit the Setup menu: press the oo soft keyThen call Setup again: Printout: Application-defined output: Configured printout
Step Press key(s) (or follow instructions) Display/Output
1. Select Setup menu, o, thenthen Printout q repeatedly and
O soft key
2. Confirm Application-defined O soft keyoutput
3. Select and confirm q soft key 3x andConfigured printout O soft key
4. Confirm Indiv. printout O soft key
5. Select Blank line O, q, l soft keys
6. Select Date/time q soft key twice, thenl soft key
7. Select Piece count q soft key repeatedly,then l soft key
8. Select net weight Q soft key repeatedly,then l soft key
9. Exit Printout configuration oo soft key
10. Perform weighing operations, rthen print
--------------------14.01.2000 09:19Qnt + 598 pcs
N + 2003.13 g
7/25/2019 MAN-ME_SE-e
46/173
Device Information
46
PurposeThis menu item enables you to haveinformation displayed about the specific
balance (device).
Display Device Information onME215/235/254/414/614
Select the Setup menu:press the o key
> SETUP is displayed:
Select Info:Repeatedly press the q soft key, thenpress the O soft key
Select Device information:press the O soft key
> Device information is displayed:
Display Device Informationon ME36S, ME5 or SE2
Select Device Information:Press the J key
> Device information is displayed
oo o q
SETUP INFO
Version no: 01-41-05
Wgh.sys. ver. #: 00-21-09
Draft sh. ver.#: 05-01-03
Model: ME215S
Serial no: 91205355
Print device information:Press the r key
> Printout (Example)
--------------------23.12.2001 13:02Model ME215SSer. no. 91205355Ver. no. 01-41-05(Version of the operating program)ID BECKER123(User ID)--------------------L ID LOT 23(Lot ID)SETUP
INFO
--------------------Version no.:
01-41-05(Version of the operating program)Wgh. sys. ver. #:
00-21-09(Version no. of the weighing cell)Dft. shield v. no.:
05-01-03(Program version no. of thedraft shield)Model:
ME215SSerial no.:
91205355
Next mainten.:01.01.2003
Service phone:00495513080
SQmin:0.0300 g
--------------------
Return to SETUP overview:press the o soft key
Exit Setup menu:Press the oo soft key
> Original settings are restored
Factory SettingsEach parameter category has a factory
setting. In the Setup menu, you canrestore all factory settings by confirmingthe selection YES.
The following settings are not restored: Language Password Display contrast Time (clock)
7/25/2019 MAN-ME_SE-e
47/173
47
Basic Weighing Function
PurposeThe basic weighing function is alwaysaccessible and can be used aloneor in combination with an applicationprogram (Toggle between Weight Units,Counting, Weighing in Percent, etc.).
Features Taring the balance Assigning IDs to weights Printing weights Printing ID codes for weights
Soft Key FunctionsCal Start calibration/adjustmentisoCAL
Press when necessary tostart calibration andadjustment
S ID Stored ID entered
General Instructionsfor Analytical Weighing
Handling Samples and ContainersThe sample should be conditionedto the temperature inside the balance.This is the only way to avoid errorscaused by air buoyancy and deviations
caused by convection currents at thesurface of the sample.
Since these effects increase proportion-ally to the volume and surface of thesample, make sure that the size of thetare vessel selected is appropriate forthe initial sample.
Never use your bare hands to touchsamples to be weighed. In addition tothe effect of the temperature, theextremely hygroscopic behaviorof fingerprints left on the sample willotherwise cause considerable inter-
ference during weight measurement.
Use forceps or a similar utensil that isappropriate to place your sample care-fully on the pan. Working with yourbalance requires a steady hand and asmooth, uninterrupted technique.
If the weighing chamber has not beenopened for a relatively long period,it may have a temperature differentfrom that of the balances surroundingenvironment. When you open theweighing chamber, a change in temp-erature will inevitably occur, due to the
laws of physics, and may show up asa change in the weight readout.
Operating the Balance
Therefore, we recommend that beforeyou begin the actual weighing series youopen and close the weighing chamber at
the same rate as you will be doing duringweighing. After the weighing chamberhas been closed, the weight readout willusually stabilize after about 8 seconds.The accuracy of the weight readouts willincrease as you continue weighing withgreater consistency.
Weighing Electrostatically ChargedSamples and ContainersMajor measuring errors can occur whenelectrostatically charged samples andcontainers are weighed. This problemparticularly involves samples that haveextremely poor conductivity (glass, plastic,filters) since they can discharge electro-static i.e., friction-induced charges onlyover a relatively long period of time. Theresult is an interaction of forces among thecharges adhering to the sample and thestationary components of the balance(weighing chamber base plate, draft shieldconstruction, balance housing). This isnoticeable when the weight readout drifts.At high humidity, this effect is notvery pronounced or may not occur at all,due to the thin layer of water thatcondenses on the sample and, through
conductive discharge, counteractsinterfering static electricity.
In addition to taking purely mechanicalcounteractive measures (protecting thesample using a special antistatic weighingpan see the Accessories), you canneutralize the surface charges bybombarding them with ions of theopposite polarity (see page 53 forinstructions on activating the ionizer).This is a highly effective procedurefor eliminating static electricity.
The balances environment, including the
operator, can considerably interfere withweighing results, due to static electricity.The balances of the ME/SE series havebeen designed to counteract this phenom-enon: the glass surfaces of the draft shieldhave a special metallic coating.
The rear panel of the balance has aterminal for connecting an equipotentialgrounding conductor. It is used foradditional grounding of a peripheraldevice (for example, a vibrating spatula).This terminal is designed for singlegrounding wires up to .25 standardgauge or 6 mm2 stranded wires and for .18
standard gauge or 4mm2
stranded wires.
Weighing Magnetic or MagnetizableSamples
It is technically impossible to avoid the useof magnetizable materials in the manufac-ture of balances. Ultimately, the operatingprinciple of high-resolution balances isbased on electromagnetic force compensa-tion of the load placed on the weighingpan.
When magnetic or magnetizable samplesor containers (e.g., a beaker with a stirrer)are weighed, interactions among theabove-mentioned components of thebalance may occur, distorting weightreadouts.
Unlike deviations caused by electrostaticcharges, magnetic interference is usuallyconstant over time. However, it is sensitiveto and dependent on the position of thesample container on the weighing pan andis also characterized by poor repeatability.
To reduce the effect described above,we recommend increasing the distancebetween the sample and the weighing panby inserting a non-magnetizable materialbetween them (the reduction in force isproportional to the quadrate of the distance).In special cases, soft-magnetic plates
should be used to shield against interferingmagnetic effects.
In the presence of extremely strongmagnetic fields for instance, whenmeasuring the susceptibility of a samplein an electromagnet you should usethe below-balance weighing port, whichcomes standard on your balance.
7/25/2019 MAN-ME_SE-e
48/173
7/25/2019 MAN-ME_SE-e
49/173
7/25/2019 MAN-ME_SE-e
50/173
50
Example W2
Enter ABC123 as the sample ID
Note: The sample ID generally applies to one weighing operation only The ID is deleted after data output
Step Press key(s) (or follow instructions) Display/Output
Initial state (balance unloaded)(ID can also be entered whilebalance is loaded)
1. Select alphabetic input a
2. Select the required letter group A ABCDEF soft key
3. Enter the letter A A soft key
(To delete a letter:c
,c
)
4. Select the letter group and ABCDEF soft keyenter B B soft key
5. Select the letter group and ABCDEF soft keyenter C C soft key
(If you enter only letters,conclude input a)
6. Enter the numbers 1, 2 and 3 1 2 3
7. Store the ID S ID soft key(20 characters max.)
The next printout will include
the sample ID
g
Max 210 g
0%
d=0.01mg
100%
000000SCal
Max 210 g
0% AH
L
S
T
d=0.01mg
100%
FS
as
ABCDEFGHIJKLMNOPQRSTUVWXYZ/=-?:#*"&U
OOOg000000
Max 210 g
0% A
H
L
S
T
d=0.01mg
100%
FS
as
A B C D E F
OOOg000000
Max 210 g
0% A
HL
S
T
d=0.01mg
100%
FS
as
ABCDEFGHIJKLMNOPQRSTUVWXYZ/=-?:#*"&U
A
Max 210 g
0% A
H
L
S
T
d=0.01mg
100%
F
S
a
s
ABCDEFGHIJKLMNOPQRSTUVWXYZ/=-?:#*"&U
AB
Max 210 g
0% A
H
LS
T
d=0.01mg
100%
F
S
a
s
ABCDEFGHIJKLMNOPQRSTUVWXYZ/=-?:#*"&UABC
Max 210 g
0% A
H
L
S
T
d=0.01mg
100%
FS
as
S ID
ABC123
g
Max 210 g
0%
d=0.01mg
100%
000000
S
Cal
7/25/2019 MAN-ME_SE-e
51/173
51
Device Parameters
This function can be deactivated.If not, the draft shield doors will closeautomatically if the balance has not
been used for 2 minutes (protectschamber from dust).
The function Close draft shield auto-matically when function is activatedcan be combined with functions andapplications that require the withstability parameter in order for weightsto be accepted: Turn on the balance (tare when the
power goes on) Tare after stability Print on request after stability Start all adjustment functions 2nd tare memory Manual weight storage mode for the
following functions: Counting, weighing in percent,
recalculation, density determination,differential weighing
Checkweighing, time-controlledfunctions with storage mode
Totalizing, formulation, statistics
A lower weight resolution is possiblewhen the draft shield doors are open.
The left and right K keys for operatingthe draft shield doors can have the same function have separate functions be switched off.
Draft Shield onME215|235|414|614|36S:
You can define which draft shielddoor(s) will open and close when youpress the left or right K key (the Geniusbalance has a self-teaching capability)
If a door encounters an obstacle whilemoving, the following will happen: While opening: the door will stop While closing: the door will re-open
Factory Settings of the ParametersLeft/right draft shield keys:Same function
Automatic mode: Off
Weight resolution when door is open:Show all decimal
places
Opening and Closingthe Draft Shield
PurposeThe ME/SE balance is a high-resolutionweighing instrument. It has a draftshield so that convection currents can-not affect the weighing result. To load asample on and remove it from theweighing pan, a draft shield door mustbe opened and closed. You can do thisin various ways, depending on themenu setting you select.
Features The draft shield doors can be opened
and closed at any time, regardless of the
application used
The draft shield doors can be openedand closed by pressing the respectivekeys, by activating an external switch orby sending a command to the balancesinterface port
The draft shield doors can be set so thatthey operate automatically in one of thefollowing ways when a specific balancefunction is performed, such as taring(zeroing): Close automatically Close automatically, then open
Draft Shield on ME5 | SE2
You can define the function of the draft shield
Setup, draft shield keys:Keys Same function Different (separate) function
O , P key Opens according to Opens 100 clockwise predefined Closes opening position Numeric input of the aperture Closes
Numeric keys + O , P key Enter and save value for No function aperture 44 181: Opens counter- clockwise 182 316: Opens clockwise 0 43: Deletes stored value
Numeric keys + Counterclockwise opening No functionP key aperture between 46 and 180
Learning mode Yes; define desired opening Noposition manually
7/25/2019 MAN-ME_SE-e
52/173
7/25/2019 MAN-ME_SE-e
53/173
53
Interfaces
Purpose
This item enables you to set theparameters for the following interfaces:
Serial communications port Serial printer port External switch function Control port function
Serial Communications PortYou can set the serial communicationsport to use for the following modes:
SBI XBPI Sartonet
Serial Printer PortYou can set the serial printer port to usefor the following printers:
YDPO1IS YDP02 YDP03 YDPO1IS Label YDPO2IS YDP02IS Label Universal YDP04IS YDP04IS Label
External Universal Remote SwitchYou can connect an external universal
remote switch (foot switch or bar codescanner or an external keyboard) to oneof the two serial ports. Then you canassign one of the following functions tobe performed whenthe switch is activated:
Print key
Tare key
Cal key
F1 function key
CF key
F2 function key
Bar code scanner/extrakeyboard (requires a specialconnecting cable)
Ionizer key
Right draft shield key
Left draft shield key
ME215/235/414/614 Models:Static Electricity Eliminator(Ionizer) J
PurposeIonization of air (i.e., charging the airwith ions of both positive and negativecharges) causes the air surrounding asample to become electrically conductive.Charges are neutralized in the air streamor grounded.
Be especially careful when weighingelectrostatically charged samples.
Features The ionizer can be turned on and off
independently of the position of thedraft shield doors
When the draft shield doors are open,the static electricity eliminator (ionizer)will shut off after the preset time haselapsed. However, the ionizing time willstart over if the static electricityeliminator is still running when a doorcloses.
The static electricity eliminator (ionizer)can be configured in the Setup menu sothat it will stay on only for a certaintime.
When the ionizer is active, this isindicated by the flashing symbol J(flashes from the outside toward theinside and vice versa)
Factory Setting for the Parameters
Static electricity eliminator (ionizer):On; Auto off after 10 sec.
Configuring the Static Electricity Elimi-nator (Ionizer)
Ionizer Off
On Auto-off time:1 to 99 seconds
Function keyJ Ionizer key
PasswordYou can enter a password to block accessto the menu parameter settings and to ID
code and exact calibration weight inputs.See the detailed description in the chapteron Configuring the Balance: Setting theDevice Parameters.
User IDYou can enter your own personal password(20 characters max.)
ClockISO/GLP printouts in particular must begenerated with the date and time stampof the specific measurement. This date andtime stamp is optional on other printouts.
See the chapter on Configuring theBalance: Setting the Device Parameters.
7/25/2019 MAN-ME_SE-e
54/173
54
KeysYou can assign different functions tothe c key for deleting input and
applications.
When you delete applications, you candelete either the data stored for allapplications or just selected data.CF function
in application
When you delete input, you can eitherdelete all the data input in a field, oronly the last character entered.CF function for inputs
You can block key functions; you canchoose whether to block all keys (excepte, o, draft shield left/right andionizer) or just the alphanumeric keys.Block key functions
Extra Functions
Acoustic SignalAn acoustic signal is emitted whenyou press a key. When the key pressed isallowed, the signal is a single beep-tone; when it is not allowed, this is sig-naled by a double-beep (key does notinitiate a function). In the Setup menu,you can configure whether
the acoustic signal should sound (On
) the acoustic signal should not sound(Off)
Power-On ModeYou can configure the balance so thatwhen a power supply is connected,
the balance is off(Off/on/standby orOff/on)
the balance switches on automatically(Auto on)
You can also set the configurations so
that when the balance is turned offafter use, it is off (Off/on)
in the standby mode(Off/on/standby)
After you turn on the balance,a self-test of the functions is run(TEST is displayed in the text line;the bar graph is shown)
Control Port FunctionYou can connect either a checkweighingdisplay or an external universal switch to
the serial communications port on thebalance (factory setting).
To do so, you need to configure theinterface for input or output.
Pin Assignment Chart of the FemaleInterface Connector
Pin Function: Input15 r key; see
Universal switch16 Left K key17 Soft key 6 (Cal)fgh Q18 Soft key 1 (F)19 w key
Pin Function: Output15 External switch
(see above)16 Control port 1: lighter17 Control port 2: equal18 Control port 3: heavier19 Control port 4: set
For further information on thepin assignment chart, see the section onPin Assignment Charts in the chapter