8/15/2019 Liquid Liquid Extraction Unit
1/49
EXPERIMENTAL MANUAL
SOLTEQ® EQUIPMENT FOR ENGINEERING EDUCATION & RESEARCH
EXCESS FEED
MATERIAL
RECOVERY
EXCESS FEED
MATERIAL
RECOVERY
8/15/2019 Liquid Liquid Extraction Unit
2/49
TABLE OF CONTENTS
LIST OF FIGURES…………………………………………………………………………………………...i
1.0 INTRODUCTION.................................................................................................................. 1
2.0 GENERAL DESCRIPTION .................................................................................................. 3
2.1 Description and Assembly ....................................................................................... 32.2 Valves and Instruments List .................................................................................... 52.3 Experiment Capabilities .......................................................................................... 62.4 General Requirements……………………………………………………………………6
3.0 SUMMARY OF THEORY ..................................................................................................... 7
4.0 OPERATING PROCEDURES ............................................................................................ 10
4.1 General Start-Up Procedure ................................................................................. 104.2 General Shut-Down Procedure..............................................................................11
4.3 Cleaning Procedure …………………………………………………………………….11
5.0 OPERATING PROCEDURES
5 1 Experiment 1: Extraction of Acetone From Water Using Toluene 12
8/15/2019 Liquid Liquid Extraction Unit
3/49
SOLTEQ® CHEMICAL PROCESS PILOT PLANT (EXCESS FEED MATERIAL RECOVERY SYSTEM) (Model: BP533-CPP)
List of FiguresPage
Figure 1 Process Flow Diagram for the Liquid-Liquid Extraction Unit 2
Figure 2 Flow of Streams and Components in a Liquid-Liquid Extraction System 7
Figure 3 A Typical 3-Component System Representation on an EquilateralTriangular Diagram
8
Figure 4 Operating Lines and Tie Lines Constructions for Determination ofExtraction Efficiency
9
8/15/2019 Liquid Liquid Extraction Unit
4/49
SOLTEQ® CHEMICAL PROCESS PILOT PLANT (EXCESS FEED MATERIAL RECOVERY SYSTEM) (Model: BP533-CPP)
1.0 INTRODUCTION
The Excess Feed Material Recovery System of the SOLTEQ® Chemical Process PilotPlant (Model: BP533-CPP) can be used to demonstrate the efficiency of separation andrecovery of recyclable materials. The apparatus requires a cold water, hot liquid from heatexchanger system, and a triple phase electrical outlet to enable a series of simplemeasurements to be made by students needing an introduction to excess feed materialrecovery design and operation. This experiment allows the student to understand someimportant things in this system which are:
Mass and energy balances Concept in separation process mainly in liquid-liquid extraction Liquid sampling on the bottom of extraction column to determine the practicalseparation behavior of the particular mixture.
This experiment consists of two important equipments which are extraction column andsolvent recovery column. Two flowrate transmitter are installed in both inlet of extractioncolumn (top and bottom column) and valves are incorporated in both inlet to regulate theflow. On the other hand, one temperature measuring device are installed in the outlet ofsolvent recovery column.
The process flow diagram for the excess feed material recovery system is given in Figure1
8/15/2019 Liquid Liquid Extraction Unit
5/49
8/15/2019 Liquid Liquid Extraction Unit
6/49
SOLTEQ® CHEMICAL PROCESS PILOT PLANT (EXCESS FEED MATERIAL RECOVERY SYSTEM) (Model: BP533-CPP)
2.0 GENERAL DESCRIPTION
2.1 Description and Assembly
Before operating the excess feed material recovery system, users must familiarizethemselves with all the components of the unit. Please refer to Figure 1 to understand theprocess.
Solvent Tank (V-113)
Material : SS 304Capacity: approx. 25-LDimensions: 0.290m (D) x 0.400m (H)
Feed Tank (V-114)
Material: SS 304Capacity: approx. 25-LDimensions: 0.290m (D) x 0.400m (H)
Settling Tank (V-115)
Material: SS 304Capacity: approx. 70-LDimensions: 0.400m (D) x 0.560m (H)
R ffi t T k (V 116)
8/15/2019 Liquid Liquid Extraction Unit
7/49
SOLTEQ® CHEMICAL PROCESS PILOT PLANT (EXCESS FEED MATERIAL RECOVERY SYSTEM) (Model: BP533-CPP)
Product Pump (P-115)
Material: SS 304Type: Vertical Centrifugal pumpCapacity: 10 LPH @ 35m headRating: 0.37 kW
Transfer Pump (P-117)
Material: SS 304Type: Vertical Centrifugal pumpCapacity: 10 LPM @ 35m headRating: 0.37 kW
Reboiler 2 (E-407)
Material: SS 304Capacity: 20 LElectrical Heater: 2 x 2 kW
Condenser 3 (E-408)
Type: Coil Heat ExchangerMaterial: SS 304
A 0 52
8/15/2019 Liquid Liquid Extraction Unit
8/49
SOLTEQ® CHEMICAL PROCESS PILOT PLANT (EXCESS FEED MATERIAL RECOVERY SYSTEM) (Model: BP533-CPP)
2.2 Valves and Instruments List
Valves list:
Tag Location
HV 601 V-113 drain valve
HV 602 V-113 outlet valve to pump, P-113
HV 603 T-501 (bottom) inlet valve
HV 604 V-114 drain valve
HV 605 V-114 outlet valve to pump, P-114HV 606 T-501 (top) inlet valve
HV 607 T-501 (bottom) drain valve
HV 608 T-501 (top) drain valve
HV 609 V-115 drain valve
HV 610 E-407 inlet valve
HV 611 V-115 inlet valve from HEXHV 612 V-115 cooling water drain valve
HV 613 E-407 drain valve
HV 614 E-408 sampling/drain valve
HV 615 V-116 drain valve
HV 616 E-408 cooling water outlet valve
HV 617 V 117 tl t l
8/15/2019 Liquid Liquid Extraction Unit
9/49
SOLTEQ® CHEMICAL PROCESS PILOT PLANT (EXCESS FEED MATERIAL RECOVERY SYSTEM) (Model: BP533-CPP)
Additional Note:
All piping are 3/4" stainless steel 304 tubing unless stated otherwise.
2.3 Experimental Capabilities
a)
Determination of the mass and energy balance of excess feed material recoverysystem.
b) Demonstration of the effect of feed flow rates on the flowrate efficiency.c) Study concept of thermodynamics of liquid-liquid system.d) Study the operation of liquid-liquid extraction and solvent recovery system.e) Demonstration of the effect of solvent to feed ratio on the extraction efficiency.f) Demonstration of the effect on of organic and aqueous phase as continuous phase.g) Determination the number of theoretical plates for the extraction column.
2.4 General Requirements
Electrical : 415 VAC/ 50Hz (3-phase)Cooling water : Laboratory tap water
8/15/2019 Liquid Liquid Extraction Unit
10/49
SOLTEQ® CHEMICAL PROCESS PILOT PLANT (EXCESS FEED MATERIAL RECOVERY SYSTEM) (Model: BP533-CPP)
3.0 SUMMARY OF THEORY
When separation by distillation is ineffective or very difficult, liquid-liquid extraction is one ofthe main alternatives to consider. Close boiling mixtures or substances that cannotwithstand the temperature of distillation, even under a vacuum, may often be separatedfrom impurities by extraction, which utilizes chemical differences instead of vapor pressuredifferences.
Separation by liquid-liquid extraction can be defined as the selective removal of one ormore components either from a homogeneous liquid mixture or from a solution, using asecond liquid or solvent, which is partially or wholly immiscible with the first.
The following terms are widely used to describe the different streams in a liquid-liquidextraction system:
F : FeedS : Solvent
R : RaffinateE : Extract
solvent,S
extract, E
l t
EXTRACTION PROCESS
8/15/2019 Liquid Liquid Extraction Unit
11/49
SOLTEQ® CHEMICAL PROCESS PILOT PLANT (EXCESS FEED MATERIAL RECOVERY SYSTEM) (Model: BP533-CPP)
The liquid used to strip the solute from the feed is referred to as the solvent. However,solvent is a general term and should not be confused with the feed solvent or diluent. The
solvent after becoming enriched in solute leaves the system as the extract.It is also important to define the composition of the four streams and this may be done inseveral ways:
a) Molar concentration = mol/dm3 (note numerically this is the same as kmol/m3)b) Molal concentration = mol/kg
c) Mole fraction, x A =
molesof no.Total
Amolesof No.
d) Mole ratio, x A/B =Bmolesof No.
Amolesof No.
e) Mass fraction, x A =massTotal
Aof Mass
f) Mass ratio, x A/B =Bof Mass
Aof Mass
Note: Fractions and ratios are dimensionless and may also be expressed in percent (%).The common basis used in most literatures are either mole or mass fraction.
Since liquid-liquid extraction involves a 3-component system, a graphical representation ofthe liquid mixtures is usually employed. There are two graphical systems, namely:
) R t l di t
8/15/2019 Liquid Liquid Extraction Unit
12/49
SOLTEQ® CHEMICAL PROCESS PILOT PLANT (EXCESS FEED MATERIAL RECOVERY SYSTEM) (Model: BP533-CPP)
Figure 3: A Typical 3-Component System Representation on an Equilateral Triangular Diagram The individual components of the chosen ternary system are usually represented bysymbols as follows:
A = SoluteB = DiluentC = Solvent
The assignment of symbols are arbitrary. Confusion shall not arise as long as the apex forthe solute is at the top while the bottom apexes are for the solvent and diluent.
Referring to Figure 3, the curve on the triangular diagram is called the binodal or solubilitycurve of the ternary system. The curve, which is essentially a combination of diluent-richand solvent-rich curves, separates the upper region of stable single phase mixture from thelower region of unstable mixture, which tends to separate into two phases.
The compositions of extract, E and raffinate, R are connected by tie-lines having differentslopes (i.e. not parallel). The tie lines get shorter as they get away from the x-axis, andeventually they converge at a point P (plait point). The point P is not normally the highestpart of the curve.
The number of theoretical stages can be determined graphically by constructing operatinglines and tie-lines on the binodal/solubility curves, as outlined in Figure 4. Consequently,
h ff f diff i d d f d fl h i ffi i
8/15/2019 Liquid Liquid Extraction Unit
13/49
SOLTEQ® CHEMICAL PROCESS PILOT PLANT (EXCESS FEED MATERIAL RECOVERY SYSTEM) (Model: BP533-CPP)
4.0 OPERATING PROCEDURES
The following chemicals will be used in the experiment:
a) Acetone (solute)b) Water (diluent)c) Toluene (solvent)
4.1 General Start-Up Procedure
1. Prepare the following chemicals:
a) Feed (heavy phase): mixture of acetone-water solution at the desiredcomposition
b) Solvent (light phase): pure toluene
2. Ensure that all valves are closed.
3.
Turn on the power of the control panel and SOLDAS® Software.
4. Open valves HV620 and HV621.
5. Fill solvent V-113 with pure toluene solvent and feed tank V-114 with acetone-water solution.
6. Close all valves except outlet vessel valves (HV 602 and HV 605).
7. Open valve HV606. Switch on pump P-114. Allow the acetone-water solution
8/15/2019 Liquid Liquid Extraction Unit
14/49
SOLTEQ® CHEMICAL PROCESS PILOT PLANT (EXCESS FEED MATERIAL RECOVERY SYSTEM) (Model: BP533-CPP)
4.2 General Shut-Down Procedure
1.
Make sure turn off all pumps (P-113, P-114, P-115 and P-117) and the heaterof reboiler 2 E-407.
2. Open valves HV 615, HV 617, HV 613 and HV 609 to drain all chemicals fromthe tanks.
Note: Make sure temperature TIC-113 in the reboiler, E-407 is drop to roomtemperature before drain the chemicals.
3.
If the equipment will not use for a long period, follow procedure 4.3 to cleanthe equipment.
4. Turn off the power for the control panel.
4.3 Cleaning Procedure
1. Drain all chemicals in tanks, columns and reboiler 2.
2.
Fill feed tanks (V-113 and V-114) with filtered water.
3. Open HV 602, HV 603, HV605 and HV 606.
4. Switch on pump P-113 and P114 to pump all water to the extraction column,T-501.
5. Switch off P-113 and P-114.
6. Repeat steps 1 to 5 using alcohol (methanol) if necessary.
8/15/2019 Liquid Liquid Extraction Unit
15/49
SOLTEQ® CHEMICAL PROCESS PILOT PLANT (EXCESS FEED MATERIAL RECOVERY SYSTEM) (Model: BP533-CPP)
5.0 OPERATING PROCEDURES
8/15/2019 Liquid Liquid Extraction Unit
16/49
SOLTEQ® CHEMICAL PROCESS PILOT PLANT (EXCESS FEED MATERIAL RECOVERY SYSTEM) (Model: BP533-CPP)
5.1 EXPERIMENT 1: Extraction of Acetone from Water Using Toluene
In this experiment, there are two feeds and two outlet flows from the column.Feed flows:Solvent: Toluene (light phase)Feed: Acetone-water mixture (heavy phase)
Outlet flows:Extract : Toluene rich phaseRaffinate: Water rich phase
OBJECTIVE:
1. To study mass balance of the system2. To study the effect of feed flow rates on the extraction efficiency
PROCEDURES:
1.
Ensure that all equilibrium and calibration data are obtained (Appendix H).
2. Perform the general start-up procedures (Section 4.1).
3. Start both pumps P-113 and P-114 and set the desired feed and solvent flowrates by adjusting the pump speed controller or/and HV 603 and HV 606respectively.
4. Allow both liquids to flow into the collection tanks V-115 and V-116
8/15/2019 Liquid Liquid Extraction Unit
17/49
SOLTEQ® CHEMICAL PROCESS PILOT PLANT (EXCESS FEED MATERIAL RECOVERY SYSTEM) (Model: BP533-CPP)
ANALYSIS AND DISCUSSION:
1. Determine the required theoretical extraction stage from the liquid-liquidextraction experiment results. Refer to Appendix B for a step-by-stepcalculation guide.
2. Calculate the height equivalent theoretical plates (HETP) of the extractioncolumn for different rotating disc speed.
HETP = platesltheoreticaof No.
heightcolumnEffective
3. Calculate the separation efficiency for different solvent to feed ratios androtating speed.
Efficiency =feedinsoluteof Amount
soluteextractedof Amount
=ncompositioFeedflowrateFeedncompositioExtractflowrateExtract
4. Compare the HETP values or separation efficiency for different solvent to feedratio.
8/15/2019 Liquid Liquid Extraction Unit
18/49
SOLTEQ® CHEMICAL PROCESS PILOT PLANT (EXCESS FEED MATERIAL RECOVERY SYSTEM) (Model: BP533-CPP)
5.2 EXPERIMENT 2: RECOVERY OF SOLVENT
Objective:
1. To study the thermodynamics of liquid-liquid systems
2. To study the operation of liquid-liquid extraction and solvent recoverysystem
Procedure
1. Open HV 610 and HV 622.
2. Switch on pump, P-115 to transfer extracted solution to reboiler 2 E-407.
3. Observe the level of extracted solution. Turn off pump, P-115 immediatelywhen the level of extracted solution is ¾ volume of total volume of thereboiler, E-407.
4. Close HV 610 after transfer extracted solution to the reboiler 2 E-407.
5. Open HV 623 and HV 624 to supply cooling water to condenser, E-408 bycontrolling the flowrate of water using HV 616 to 15 L/min.
6. Switch on the heater in the reboiler, E-407 and adjust the power of heater
8/15/2019 Liquid Liquid Extraction Unit
19/49
SOLTEQ® CHEMICAL PROCESS PILOT PLANT (EXCESS FEED MATERIAL RECOVERY SYSTEM) (Model: BP533-CPP)
6.0 REFERENCES
1.
McCabe, Smith & Harriott, “Unit Operations of Chemical Engineering”, McGraw Hill,(Singapore), 2001.
2. Robert E. Treybal, “Mass Transfer Operations”, Third Edition, McGraw Hill Book Co.(Singapore), 1980.
3. Robert E. Treybal, “Liquid Extraction”, Second Edition, McGraw Hill Series, 1963.
4.
Institution of Chemical Engineers, “Standard Test Systems for Liquid Extraction”,Second Edition, 1985.
8/15/2019 Liquid Liquid Extraction Unit
20/49
SOLTEQ® CHEMICAL PROCESS PILOT PLANT (EXCESS FEED MATERIAL RECOVERY SYSTEM) (Model: BP533-CPP)
7.0 MAINTENANCE AND SAFETY PRECAUTIONS
1.
The unit must be operated under the supervision of trained personnel.2. All operating instructions supplied with the unit must be carefully read and understood
before attempting to operate the unit.
3. Always check and rectify any leak.
4. Always make sure that the heater is fully submerged in the water.
5. Do not touch the hot components of the unit.
6.
Be extremely careful when handling liquid at high temperature.
7. Always switch off the heater and allow the liquid to cool down before draining.
8. Restore the system to operating conditions after any repair job.
9. Only properly trained personnel shall be allowed to carry out any servicing.Manufacturer's manual must always be observed.
10.
Before servicing, shut down the whole operation and let the system to cool down.
8/15/2019 Liquid Liquid Extraction Unit
21/49
SOLTEQ® CHEMICAL PROCESS PILOT PLANT (EXCESS FEED MATERIAL RECOVERY SYSTEM) (Model: BP533-CPP)
APPENDICES
8/15/2019 Liquid Liquid Extraction Unit
22/49
SOLTEQ® CHEMICAL PROCESS PILOT PLANT (EXCESS FEED MATERIAL RECOVERY SYSTEM) (Model: BP533-CPP)
APPENDIX A
SAMPLE TABLE FOR EXPERIMENT EXPERIMENT 1
Run No.Feed
flow rate(L/min)
Solventflow rate(L/min)
Solventto feedratio
RotatingDisc
Speed(RPM)
Time(min)
Raffinate Extract
HETP(mm)
SeparationefficiencyRefractive
index
Solutecomposition
(wt%)
Refractiveindex
Solutecomposition
(wt%)
8/15/2019 Liquid Liquid Extraction Unit
23/49
SOLTEQ® CHEMICAL PROCESS PILOT PLANT (EXCESS FEED MATERIAL RECOVERY SYSTEM) (Model: BP533-CPP)
EXPERIMENT 2: RECOVERY OF SOLVENT
Time (min) Refractive Index (RI) Purity of Solvent (%)
8/15/2019 Liquid Liquid Extraction Unit
24/49
SOLTEQ® CHEMICAL PROCESS PILOT PLANT (EXCESS FEED MATERIAL RECOVERY SYSTEM) (Model: BP533-CPP)
APPENDIX B
STEP BY STEP CALCULATION GUIDE
1. Obtain the equilibrium data from Appendix D or from literature. Plot the binodal/solubility curveand the corresponding tie lines on a triangular plot (Appendix G).
2. Locate the points for:Feed, FSolvent, SRaffinate, RExtract, EFeed and solvent mixture, M
acetone
watertoluene
F
S
R
E
M
8/15/2019 Liquid Liquid Extraction Unit
25/49
SOLTEQ® CHEMICAL PROCESS PILOT PLANT (EXCESS FEED MATERIAL RECOVERY SYSTEM) (Model: BP533-CPP)
4. Draw a tie line from point E to its equilibrium point on the binodal plot. Connect an operatingline from that equilibrium end of the tie line to point ΔR. Extend that operating line so that it
intersects with the binodal plot at the other end. Draw another tie line from that new intersectionand continue until a tie line has passed the operating line connecting R and S.
acetone
watertoluene
F
S
R
E
ΔR
8/15/2019 Liquid Liquid Extraction Unit
26/49
SOLTEQ® CHEMICAL PROCESS PILOT PLANT (EXCESS FEED MATERIAL RECOVERY SYSTEM) (Model: BP533-CPP)
APPENDIX C
CALIBRATION PROCEDURES FOR LIQUID-LIQUID EXTRACTION EXPERIMENT
C.1 Determination Of The Binodal/Solubility Curve
OBJECTIVE:
To obtain the binodal/solubility curve for both toluene and water rich phases.
APPARATUS:
1. Conical flasks, measuring cylinders, titration apparatus.2. Refractometer for calibration.
PROCEDURES:
1. Record the average ambient temperature throughout the experiment and obtain the
pure component densities for acetone, water and toluene at the experimenttemperature.
Toluene rich phase
2. Prepare 6 homogenous mixtures of acetone and toluene in 6 different conical flasksaccording to the recommended composition listed in the first table in Appendix D.
3. Prepare the apparatus for titration and fill the burette with water.
8/15/2019 Liquid Liquid Extraction Unit
27/49
SOLTEQ® CHEMICAL PROCESS PILOT PLANT (EXCESS FEED MATERIAL RECOVERY SYSTEM) (Model: BP533-CPP)
amount of toluene that has been added. This is known as the cloud point where allthree components are in equilibrium and this represents a point on the ternary
binodal/solubility curve.5. Take about 10 mL sample of the ternary mixture and measure the refractive index
using a refractometer. Take 3 series of readings for the same sample and record theaverage.
6. Repeat the experiment (steps 9 to 12) for all 6 mixtures. Calculate the mass fraction ofeach component during equilibrium and record the results in the second table in
Appendix D.
ANALYSIS AND DISCUSSION:
1. Use the sample tables in Appendix D for data collection.
a) For the toluene rich phase, after titration with water, enter the volume of waterrequired for ternary equilibrium. From the knowledge of pure densities for eachcomponent, calculate their mass fraction in the mixture. Enter the refractive index
for each mixture in the appropriate column.b) For the water rich phase, after titration with toluene, enter the volume of toluene
required for ternary equilibrium. Again, from the knowledge of pure densities foreach component, calculate their mass fraction in the mixture. Enter the refractiveindex for each mixture in the appropriate column.
2. Plot the refractive index of each mixture as a function of acetone (solute) composition.Do a separate plot for the toluene rich phase and water rich phase. This will serve as
8/15/2019 Liquid Liquid Extraction Unit
28/49
SOLTEQ® CHEMICAL PROCESS PILOT PLANT (EXCESS FEED MATERIAL RECOVERY SYSTEM) (Model: BP533-CPP)
C.2 Determination of Equilibrium Solute Distribution (Tie Lines)
OBJECTIVE:To determine the tie lines connecting the binodal curves in the toluene-water-acetonesystem.
APPARATUS:
1. Separating funnels for settling and separating the mixtures.2. Conical flasks, measuring cylinders, and standard laboratory apparatus.
3. Refractometer.
PROCEDURES:
1. Obtain the calibration curve from Appendix C.1.
2. Record the average ambient temperature throughout the experiment and obtain thepure component densities for acetone, water and toluene at the experimenttemperature.
3. Prepare 6 heterogeneous mixtures of acetone, toluene and water in 6 different conicalflasks according to the compositions recommended in the table in Appendix E.
4. Agitate the mixtures for about 60 minutes.
Note: To prevent emulsification, do not agitate the mixtures too vigorously.
5. Pour each of the mixture into a separating funnel. Let the mixture settle into two distinct
8/15/2019 Liquid Liquid Extraction Unit
29/49
SOLTEQ® CHEMICAL PROCESS PILOT PLANT (EXCESS FEED MATERIAL RECOVERY SYSTEM) (Model: BP533-CPP)
APPENDIX D
SAMPLE TABLES FOR APPENDIX C.1
Toluene Rich Phase
No
Amount in mixture (mL) Wateradded(mL)
Composition (wt%)Refractive
indexAcetone Toluene Acetone Water Toluene
1 160 mL 40 mL
2 140 mL 60 mL
3 120 mL 80 mL
4 100 mL 100 mL
5 80 mL 120 mL
6 60 mL 140 mL
8/15/2019 Liquid Liquid Extraction Unit
30/49
SOLTEQ® CHEMICAL PROCESS PILOT PLANT (EXCESS FEED MATERIAL RECOVERY SYSTEM) (Model: BP533-CPP)
APPENDIX E
SAMPLE TABLE FOR APPENDIX C.2
No.
Amount in mixture (mL) Water rich phase Toluene rich phase
Acetone Water TolueneRefractive
index
Acetonecomposition
(wt%)
Refractiveindex
Acetonecomposition
(wt%)
1 20 mL 90 mL 90 mL
2 40 mL 80 mL 80 mL
3 60 mL 70 mL 70 mL
4 80 mL 60 mL 60 mL
5 100 mL 50 mL 50 mL
6 120 mL 40 mL 40 mL
8/15/2019 Liquid Liquid Extraction Unit
31/49
SOLTEQ® CHEMICAL PROCESS PILOT PLANT (EXCESS FEED MATERIAL RECOVERY SYSTEM) (Model: BP533-CPP)
APPENDIX F
TRIANGULAR PLOT FOR TERNARY DIAGRAM
acetone
8/15/2019 Liquid Liquid Extraction Unit
32/49
SOLTEQ® CHEMICAL PROCESS PILOT PLANT (EXCESS FEED MATERIAL RECOVERY SYSTEM) (Model: BP533-CPP)
APPENDIX G
PHYSICAL PROPERTIES OF COMPONENTS
G.1 Physical Properties of Pure Components in the Ternary System
ComponentTemp(oC)
Density(kg/m3)
Viscosity(10-3 Pa.s)
SurfaceTension
(10-3 Nm-1)Refractive Index
Water 20 998.2 1.003 72.75 1.333
25 997.02 0.8903 72.58 1.332
Toluene 20 866.7 0.586 28.4 1.497
25 862.3 0.552 27.6 1.494
Acetone 20 790.5 0.322 23.4 1.359
25 784.4 0.304 23.2 1.356
Acetone (CH3COCH3) : A colourless, volatile, extremely flammable liquid, miscible with water,used as a solvent and reagent. Also known as 2-propanone.
Toluene (C6H5CH3) : A colourless, aromatic liquid derived from coal tar or from the catalyticreforming of petroleum naphthas; insoluble in water, soluble in alcohol
8/15/2019 Liquid Liquid Extraction Unit
33/49
SOLTEQ® CHEMICAL PROCESS PILOT PLANT (EXCESS FEED MATERIAL RECOVERY SYSTEM) (Model: BP533-CPP)
G.2 Fire and Safety Properties
Property Acetone Toluene
Boiling Point (oC)
Melting Point (oC)
Flash Point (oC) – closed cup
Explosive Limits in Air
lower (% vol.)
upper (% vol.)
Ignition Temperature (oC)
Vapour Pressure @ 20 oC (mmHg)
56.5
-94.6
-19
2.6
12.8
538
185
110.4
-95
4
1.2
7.0
508
22
8/15/2019 Liquid Liquid Extraction Unit
34/49
SOLTEQ® CHEMICAL PROCESS PILOT PLANT (EXCESS FEED MATERIAL RECOVERY SYSTEM) (Model: BP533-CPP)
G.3 Binodal Compositions for Acetone-Toluene-Water System
Acetone(wt%)
Toluene(wt%)
Water(wt%)
Acetone(wt%)
Toluene(wt%)
Water(wt%)
0.5 0.026 99.474 66 19 15
0 0.029 99.971 66 19.8 14.2
1 0.03 98.97 65.9 20.63 13.47
1.5 0.043 98.457 65.01 24.9 10.092.499 0.05 97.451 65 25 10
4.997 0.054 94.949 64.07 26.94 8.99
10 0.06 89.94 62.4 30.6 7
19.97 0.16 79.87 62.18 31.53 6.29
29.91 0.3 69.79 59.49 34.43 6.08
18.9 0.4 80.7 58.4 36.6 5
34.5 0.5 65 55.4 40.6 4
35.71 0.62 63.67 54.88 41.47 3.65
39.56 0.92 59.52 54.6 41.58 3.82
42 1 57 51.1 45.9 3
42.68 1.03 56.29 48.76 48.76 2.48
42.7 1.12 56.18 48.1 49.4 2.5
8/15/2019 Liquid Liquid Extraction Unit
35/49
SOLTEQ® CHEMICAL PROCESS PILOT PLANT (EXCESS FEED MATERIAL RECOVERY SYSTEM) (Model: BP533-CPP)
G.4 Tie Line Compositions for Acetone-Toluene-Water System
Acetone Composition (wt%)
Water RichPhase
Toluene RichPhase
Water RichPhase
Toluene RichPhase
0.27 0.21 12.42 11.3
0.59 0.36 13.05 11.91
0.95 0.59 14.1 13.11.08 0.78 14.41 13.52
1.54 1.13 17.32 16.98
1.72 1.11 19.3 18.8
2.32 1.52 20 20.2
2.41 1.69 20.5 21.35
2.43 1.74 21.44 22.62
2.7 1.85 25.7 27.4
3.13 2.38 25.8 29.05
3.42 2.45 31.51 38.12
4.2 3.15 32 37.5
4.5 2.8 32.3 39.6
5.1 3.84 32.82 41.06
8/15/2019 Liquid Liquid Extraction Unit
36/49
SOLTEQ® CHEMICAL PROCESS PILOT PLANT (EXCESS FEED MATERIAL RECOVERY SYSTEM) (Model: BP533-CPP)
APPENDIX H
TYPICAL CALIBRATION DATA
Determination Of The Binodal/Solubility Curve
Refractive index for pure acetone = 1.3581
Refractive index for pure water = 1.3332
Refractive index for pure toluene = 1.4954
Toluene Rich Phase
No
Amount in mixture (mL) Wateradded(mL)
Composition (wt%)Refractive
indexAcetone Toluene Acetone Water Toluene
1 12.7 104.1 2.5 9.79 2.42 87.78 1.4753
2 25.2 91.8 3.0 19.48 2.92 77.60 1.4649
3 54.0 64.6 4.1 41.61 3.97 54.42 1.4304
4 67.4 50.0 5.4 52.30 5.27 42.43 1.4147
8/15/2019 Liquid Liquid Extraction Unit
37/49
SOLTEQ® CHEMICAL PROCESS PILOT PLANT (EXCESS FEED MATERIAL RECOVERY SYSTEM) (Model: BP533-CPP)
8/15/2019 Liquid Liquid Extraction Unit
38/49
1.3300
1.3350
1.3400
1.3450
1.3500
1.3550
1.3600
1.3650
1.3700
0 5 10 15 20 25 30 35 40 45 50 55 60 65 70
R I
Acetone composition (wt%)
Calibration Curve for Water Rich Phase
®
8/15/2019 Liquid Liquid Extraction Unit
39/49
SOLTEQ® CHEMICAL PROCESS PILOT PLANT (EXCESS FEED MATERIAL RECOVERY SYSTEM) (Model: BP533-CPP)
Binodal/Solubility Curve on the Triangular Diagram
acetone
8/15/2019 Liquid Liquid Extraction Unit
40/49
SOLTEQ® CHEMICAL PROCESS PILOT PLANT (EXCESS FEED MATERIAL RECOVERY SYSTEM) (Model: BP533-CPP)
8/15/2019 Liquid Liquid Extraction Unit
41/49
0
10
20
30
40
50
60
70
80
0 5 10 15 20 25 30 35 40 45 50 55 60
A c e t o n e c o m p o
s i t i o n i n t o l u e n e r i c h p h a s e ( w t % )
Acetone composition in water rich phase (wt%)
Tie Line Compositions
SOLTEQ® CHEMICAL PROCESS PILOT PLANT (EXCESS FEED MATERIAL RECOVERY SYSTEM) (Model: BP533-CPP)
8/15/2019 Liquid Liquid Extraction Unit
42/49
APPENDIX I: RESULTS
EXPERIMENT 1
RunNo.
Feed flowrate
(L/min)
Solventflowrate
(L/min)
Solvent tofeed ratio
Time(min)
Raffinate Extract
HETP(mm)
Separationefficiency
(%)Refractive
index
Solutecomposition
(wt%)
Refractiveindex
Solutecomposition
(wt%)
1 0.12 0.12 1:1 10 1.3341 1.4511
20 1.3342 1.4508
30 1.3354 1.4411
40 1.3358 1.4435
50 1.3363 1.4461
60 1.3371 1.4461
70 1.3363 5.0 1.4464 32.1 760 64.0
2 0.12 0.12 1:1 0 1.3369 1.4431
10 1.3373 1.4445
20 1.3374 1.4444
30 1.3371 7.0 1.4463 32.6 797 65.2
3 0.12 0.24 1:2 10 1.4098 1.339520 1.4121 1.3425
30 1.3422 1.4162
40 1.3413 1.4211
SOLTEQ® CHEMICAL PROCESS PILOT PLANT (EXCESS FEED MATERIAL RECOVERY SYSTEM) (Model: BP533-CPP)
8/15/2019 Liquid Liquid Extraction Unit
43/49
50 1.3422 1.4631
60 1.3402 1.4655
70 1.3390 9.5 1.4651 19.5 889 39.0
SOLTEQ® CHEMICAL PROCESS PILOT PLANT (EXCESS FEED MATERIAL RECOVERY SYSTEM) (Model: BP533-CPP)
8/15/2019 Liquid Liquid Extraction Unit
44/49
Composition Diagram for Run 1 Data
SOLTEQ® CHEMICAL PROCESS PILOT PLANT (EXCESS FEED MATERIAL RECOVERY SYSTEM) (Model: BP533-CPP)
8/15/2019 Liquid Liquid Extraction Unit
45/49
Composition Diagram for Run 2 Data
SOLTEQ® CHEMICAL PROCESS PILOT PLANT (EXCESS FEED MATERIAL RECOVERY SYSTEM) (Model: BP533-CPP)
8/15/2019 Liquid Liquid Extraction Unit
46/49
Composition Diagram for Run 3 Data
8/15/2019 Liquid Liquid Extraction Unit
47/49
SOLTEQ® CHEMICAL PROCESS PILOT PLANT (EXCESS FEED MATERIAL RECOVERY SYSTEM) (Model: BP533-CPP)
8/15/2019 Liquid Liquid Extraction Unit
48/49
=
=32.6%
50%
= 0.652
For Run 3 data:
= ℎ
.
=2000
2.25
= 889
=
=19.5%
50%
= 0.390
SOLTEQ® CHEMICAL PROCESS PILOT PLANT (EXCESS FEED MATERIAL RECOVERY SYSTEM) (Model: BP533-CPP)
8/15/2019 Liquid Liquid Extraction Unit
49/49
EXPERIMENT 2: RECOVERY OF SOLVENT
POWER OF HEATER REFRACTIVE INDEX PURITY OF TOLUENE (%)
1000 W 1.4850 90
2000 W 1.4901 93