©Ian Sommerville 2004 Software Engineering, 7th edition. Insulin Pump Slide 1
The portable insulin pump
Developing a dependabilityspecification for the insulin
pump
8/3/2019 Insulin Pump Depend Spec
http://slidepdf.com/reader/full/insulin-pump-depend-spec 2/13
©Ian Sommerville 2004 Software Engineering, 7th edition. Insulin Pump Slide 2
Dependability attributes
Availability The pump should have a high level of availability but the nature
of diabetes is such that continuous availability is unnecessary
Reliability Intermittent demands for service are made on the system
Safety The key safety requirements are that the operation of the
system should never result in a very low level of blood sugar. Afail-safe position is for no insulin to be delivered
Security Not really applicable in this case
8/3/2019 Insulin Pump Depend Spec
http://slidepdf.com/reader/full/insulin-pump-depend-spec 3/13
©Ian Sommerville 2004 Software Engineering, 7th edition. Insulin Pump Slide 3
System availability
In specifying the availability, issues that must beconsidered are: The machine does not have to be continuously available as
failure to deliver insulin on a single occasion (say) is not a
problem However, no insulin delivery over a few hours would have an
effect on the patient¶s health
The machine software can be reset by switching it on and off hence recovery from software errors is possible withoutcompromising the usefulness of the system
Hardware failures can only be repaired by return to themanufacturer. This means, in practice, a loss of availability of atleast 3 days
8/3/2019 Insulin Pump Depend Spec
http://slidepdf.com/reader/full/insulin-pump-depend-spec 4/13
©Ian Sommerville 2004 Software Engineering, 7th edition. Insulin Pump Slide 4
Availability
A general specification of availability suggeststhat the machine should not have to be returnedto the manufacturer more than once every year
years (this repair time dominates everythingelse) so System availability = 727/730 *100 = 0.99
It is much harder to specify the software
availability as the demands are intermittent. Inthis case, you would subsume availability under reliability
8/3/2019 Insulin Pump Depend Spec
http://slidepdf.com/reader/full/insulin-pump-depend-spec 5/13
©Ian Sommerville 2004 Software Engineering, 7th edition. Insulin Pump Slide 5
Reliability metric
Demands on the system are intermittent (several
times per hour) and the system must be able to
respond to these demands
In this case, the most appropriate metric is
therefore Probability of Failure on Demand
Other metrics
Short transactions so MTTF not appropriate
Insufficient number of demands for ROCOF to be
appropriate
8/3/2019 Insulin Pump Depend Spec
http://slidepdf.com/reader/full/insulin-pump-depend-spec 6/13
©Ian Sommerville 2004 Software Engineering, 7th edition. Insulin Pump Slide 6
System failures
Transient failures
can be repaired by user actions such as resetting or
recalibrating the machine. For these types of failure, a relatively
low value of POFOD (say 0.002) may be acceptable. Thismeans that one failure may occur in every 500 demands made
on the machine. This is approximately once every 3.5 days.
Permanent failures
require the machine to be repaired by the manufacturer. The
probability of this type of failure should be much lower. Roughlyonce a year is the minimum figure so POFOD should be no
more than 0.00002.
8/3/2019 Insulin Pump Depend Spec
http://slidepdf.com/reader/full/insulin-pump-depend-spec 7/13
©Ian Sommerville 2004 Software Engineering, 7th edition. Insulin Pump Slide 7
System hazard analysis
Physical hazards
Hazards that result from some physical failure of the
system
Electrical hazards
Hazards that result from some electrical failure of the
system
Biological hazards
Hazards that result from some system failure that
interferes with biological processes
8/3/2019 Insulin Pump Depend Spec
http://slidepdf.com/reader/full/insulin-pump-depend-spec 8/13
©Ian Sommerville 2004 Software Engineering, 7th edition. Insulin Pump Slide 8
insulin overdose or underdose (biological)
power failure (electrical)
machine interferes electrically with other medical
equipment such as a heart pacemaker (electrical)
parts of machine break off in patient¶s body(physical)
infection caused by introduction of machine (biol.)
allergic reaction to the materials or insulin used in the
machine (biol).
Insulin system hazards
8/3/2019 Insulin Pump Depend Spec
http://slidepdf.com/reader/full/insulin-pump-depend-spec 9/13
©Ian Sommerville 2004 Software Engineering, 7th edition. Insulin Pump Slide 9
Risk analysis example
Identifiedhazard
Hazardprobability
Hazardseverity
Estimatedrisk
Acceptability
Insulin overdose Medium High High Intolerable
Insulin
underdose
Medium Low Low Acceptable
Power failure High Low Low Acceptable
Machineincorrectly fitted
High High High Intolerable
Machine breaksin patient
Low High Medium ALARP
Machine causes
infection
Medium Medium Medium ALARP
Electricalinterference
Low High Medium ALARP
Allergic reaction Low Low Low Acceptable
8/3/2019 Insulin Pump Depend Spec
http://slidepdf.com/reader/full/insulin-pump-depend-spec 10/13
©Ian Sommerville 2004 Software Engineering, 7th edition. Insulin Pump Slide 10
Software-related hazards
Only insulin overdose and insulin underdose are
software related hazards
The other hazards are related to the hardwareand physical design of the machine
Insulin underdose and insulin overdose can be
the result of errors made by the software in
computing the dose required
8/3/2019 Insulin Pump Depend Spec
http://slidepdf.com/reader/full/insulin-pump-depend-spec 11/13
©Ian Sommerville 2004 Software Engineering, 7th edition. Insulin Pump Slide 11
Software problems
Arithmetic error
Some arithmetic computation causes a representation failure
(overflow or underflow)
Specification may state that arithmetic error must be detectedand an exception handler included for each arithmetic error.
The action to be taken for these errors should be defined
Algorithmic error
Difficult to detect anomalous situation
May use µrealism¶ checks on the computed dose of insulin
8/3/2019 Insulin Pump Depend Spec
http://slidepdf.com/reader/full/insulin-pump-depend-spec 12/13
©Ian Sommerville 2004 Software Engineering, 7th edition. Insulin Pump Slide 12
Insulin pump fault tree
8/3/2019 Insulin Pump Depend Spec
http://slidepdf.com/reader/full/insulin-pump-depend-spec 13/13
©Ian Sommerville 2004 Software Engineering, 7th edition. Insulin Pump Slide 13
General dependability requirements
SR1: The system shall not deliver a single dose of insulin that is
greater than a specified maximum dose for a system user.
SR2: The system shall not deliver a daily cumulative dose of insulin
that is greater than a specified maximum for a system user.
SR3: The system shall include a hardware diagnostic facility that
should be executed at least 4 times per hour.
SR4: The system shall include an exception handler for all of the
exceptions that are identified in Table 3.
SR5: The audible alarm shall be sounded when any hardware
anomaly is discovered and a diagnostic message as defined inTable 4 should be displayed.