Analyzer SpecificationsDimensions (W x D x H)
Weight
Operating temperature
Setup time
Required sample volume
Data memory
Cartridge information
238 x 326 x 310 [mm]
Approximately 7 kg
15℃ - 30℃ (59°F - 86°F)3 min
basic test : 100 µLdilution test : 20 µL
100 tests
Information about test parameters is embedded in the QR code attached to the cartridge.
Reagent Cartridges SpecificationsTypes of sample
Dynamic Range
Package Unit
Measurement time
Reagent Storage
v-T4vc-TSHv-CORv-BA
v-T4vc-TSHv-COR
v-BA
10 cartridges/box
Approximately 10 min (without dilution)
: Canine and Feline Plasma/Serum: Canine Plasma/Serum: Canine Plasma(dilution)/Serum: Canine and Feline Plasma/Serum
: 0.50 - 8.00μg/dL(6.4 - 103.0 nmol/L): 0.25 - 5.00 ng/mL: 1.0 - 30.0 μg/dL(27.6 - 828.0 nmol/L) 1.0 - 50.0 μg/dL(27.6 - 1380.0 nmol/L)(with dilution): 2.0 - 150.0 μmol/L
2℃ - 8℃ (35.6°F - 46.4°F)
26-30, NISHIAZABU 2-CHOME, MINATO-KU, TOKYO 106-8620, JAPANhttp://www.fujifilm.com/products/medical/ Ref. No. FDCAU10V R(17·11·F1079·F8598) Printed in Japan ©2011 FUJIFILM Corporation
This equipment complies with FDA performance standards for laser products except for deviations pursuant to Laser Notice No. 50, dated (June 24, 2007).This equipment is a Class 1 laser product (designed according to IEC60825-1:2014 / EN60825-1:2014)
Options
Sample barcode reader Printer
Automated Fluorescence Immunoassay Analyzer
Innovation in Immunodiagnostic Testing
Bile Acid Testv-BA
Top
Front
Side
IMMUNO AU10V DRI-CHEM analyzers (NX500, FDC4000, FDC7000)
FUJI PLAIN/HEPARIN TUBE and FUJI DRI-CHEM AUTO TIPScan be used in both systems.
System Configuration Diagram
W 238 mm
D 326 mm
100 mm
100 mm
W 238 mm
H 310 mm
D 326 mm
H 310 mm100 mm
DIMENSIONS
NEW
Perform automatic measurements simply by inserting a tube (for samples), a tip (for sampling and injection), and a cartridge (reagents), and press the START key!
Simple 2 - step procedure
AU10V has an automatic dilution function for v-COR measurements to accommodate serum samples with high cortisol levels. (Plasma samples are to be measured in dilution mode regardless of cortisol level.)
Automatic Dilution Function
All reagents and materials are already prepared inside the cartridge.
No need to add reagents.
All calibrations and measurements are conducted automatically based on the QR code* attached to the bottom of the cartridge, which eliminates the need for lot calibration.
Automated calibration for reagents using attached QR code
1.5 mL or 0.5 mL sample tube can be used, allowing flexibility when collecting bloodsamples in accordance with the measurement situation.
Flexibility in selecting the most appropriate sample tube (1.5 mL or 0.5 mL)
Top
Bottom
Workflow Time
Measurement completedand results displayed.
Fully automated measurement in approximately 10 min!
Insert materials.1
Press the START key.2
Easy To UseEasy To Use
“QR code” is a registered trademark of DENSO WAVE INC.
FUJI DRI-CHEM IMMUNO AU10VA new in-clinical Immunodiagnostic Analyzer realized by FUJIFILM’s cutting edge technologiesFUJIFILM's new Surface Plasmon enhanced Fluorescence (SPF) technology has brought about a compact sized Fluorometry Immunodiagnostic Analyzer of high reliability and accuracy. Ideal for in-clinic immunodiagnostic testing.
Real time diagnosis and monitoring for endocrineand liver diseases made possible.
Test Parameter
v-T4(Thyroxine (T4) )
vc-TSH(Thyroid-Stimulating Hormone)
v-BA(Bile Acids)
Purpose
Screening and diagnosis for Hypo/Hyperthyroidism
Diagnosis for Hypothyroidism
v-COR(Cortisol)
Diagnosis for Addison’s Disease and Cushing’s Syndrome
Screening for liver diseases and diagnosis for portosystemic shunt
Reagent Cartridges Specifications
v-T4
vc-TSH
v-COR
v-BA
NEW
Principle of the system
TechnologyTechnology
High ReliabilityHigh Reliability
Fluorescent beads
AntigenAntibodyThin gold film
Sample Injection Fluorescence MeasurementLaser Beam IrradiationBinding ReactionThe sample, which is mixed with the reagent, will be dispensed into the sample flow channel of the cartridge.
The analyte binds with the antibody, which will be attached to the thin gold film.
A SPR is generated upon irradiation by the laser beam.
The fluorescent intensity, which is commensurate with the level of analyte concentration, is detected.
Reagent Cup
• The equipment performs fluorometry using Surface Plasmon enhanced Fluorescence(SPF), a technology based on Surface Plasmon Resonance (SPR), within a small cartridge. • SPF has two major advantages which made the realization of a compact sized immunodiagnostic analyzer.• First advantage is it does not require a washing process for surplus fluorescent beads, thus the mechanical process of the equipment is simple. • Second, since the fluorescence is enhanced by SPR, a low power beam is sufficient enough to generate substantial fluorescence intensity for detection.
AU10V Key FeaturesAU10V Key Features
New test parameter: v-BA (canine and feline)
PLASMA samples measurable in all tests
FELINE samples measureable in v-T4
・The only bench top analyzer with TSH test
・Easy to use
・Approximately 10 min measurement
Excellent correlation with standard laboratory analyzerAU10V provides reliable and accurate results. It has excellent correlations with its comparison methods in serum samples.
AU10V can also accurately measure PLASMA samples as shown in the good correlations of serum and plasma.
NEW
NEW
NEW
v-T4 Caniney = 1.034x - 0.002r = 0.967n = 92
CLEIA (µg/dL)
AU10
v-T
4 (µ
g/dL
)
v-T4 Feliney = 0.965x + 0.024r = 0.962n = 90
CLEIA (µg/dL)
AU10
v-T
4 (µ
g/dL
)
v-T4 Caniney = 0.984x - 0.129r = 0.996n = 60
AU10V v-T4 Serum (µg/dL)
AU10
V v-
T4 P
lasm
a (µ
g/dL
)
vc-TSH Caniney = 1.041 x - 0.019 R = 0.991n = 50
AU10V vc-TSH Serum (ng/mL)
AU10
V vc
-TSH
Pla
sma
(ng/
mL)
v-T4 Feliney = 1.005x - 0.003r = 0.997n = 57
AU10V v-T4 Serum (µg/dL)
AU10
V v-
T4 P
lasm
a (µ
g/dL
)
v-COR Caniney = 0.996 x + 0.824 r = 0.994n = 24
AU10V v-COR Serum (µg/dL)
AU10
V v-
CO
R Pl
asm
a (µ
g/dL
)
vc-TSH Caniney = 0.941x + 0.077r = 0.983n = 72
CLEIA (ng/mL)
AU10
V vc
-TSH
(ng/
mL)
v-COR Caniney = 0.999 x + 0.285r = 0.963n = 125
CLEIA (µg/dL)
AU10
V v-
CO
R (µ
g/dL
)
v-BA Caniney = 0.987x - 0.537r = 0.998n = 53
AU10V v-BA Serum (µmol/L)
AU10
V v-
BA P
lasm
a (µ
mol
/L)
v-BA Feliney = 1.039x - 0.714r = 0.999n = 50
AU10V v-BA Serum (µmol/L)
AU10
V v-
BA P
lasm
a (µ
mol
/L)
v-BA Caniney = 0.970x - 0.600r = 0.960n = 53
Enzymatic Method (µmol/L)
AU10
V v-
BA (µ
mol
/L)
v-BA Feliney = 0.933x + 0.643r = 0.967n= 62
Enzymatic Method (µmol/L)
AU10
V v-
BA (µ
mol
/L)