Alarm Systems Management Webinar Series
How to Identify the Most Important Alarm Signals
to ManageOctober 30, 2013
10/30/2013 1
Co-Conveners
10/30/2013 2
Alarm Systems Management Webinar Series
Thank You to Our SponsorsThis webinar series is offered at no charge thanks to commercial grants from the companies shown here. HTSI and its co-convening organizations appreciate their generosity. AAMI and HTSI are managing all costs for the series. None of the co-convening organizations participated in seeking, discussing, or otherwise facilitating the commercial grants. The companies had no role in content development, and the webinars do not contain commercial content. Webinar presenters were selected based on topic expertise without regard to industry affiliation by a multi-disciplinary HTSI volunteer planning committee.
10/30/2013 3
Speaker Introductions• Shawn Forrest, MS, Biomedical Engineering Reviewer, Cardiac
Diagnostic Devices Branch, Division of Cardiovascular Devices, Office of Device Evaluation, CDRH, U.S. Food and Drug Administration
• Shashi Avadhani, CBET, CCE, MBA, Resident Regional Manager, Crothall Clinical Equipment Solutions
• Melanie E. Quinton, MS, Solution Consultant/Consultant Specialist, CD BIO MDI, Kaiser Permanente
• Lisa A. Pahl, RN, BSN, MSN, Sr. Manager, Enterprise Consulting, Philips Healthcare Americas
• Marjorie Funk, PhD, RN, FAHA, FAAN, Professor, Yale University School of Nursing (Moderator)
10/30/2013 4
Alarm Systems Management Webinar Series
Alarm-Related Adverse Events
Shawn Forrest, MSBiomedical Engineering Reviewer
Cardiac Diagnostic Devices BranchDivision of Cardiovascular DevicesOffice of Device Evaluation, CDRH U.S. Food and Drug Administration
10/30/2013 5
Adverse Event Reporting
• Manufacturer and User Facility Device Experience (MAUDE) Database• Reports of suspected device-associated:
• Deaths• Serious injuries• Malfunctions
• Mandatory reporting:• Manufacturers, importers, user facilities
• Voluntary reporting:• Healthcare professionals, patients, consumers
10/30/2013 6
Alarm Systems Management Webinar Series
MAUDE Limitations
• Passive Surveillance• Submission of incomplete, inaccurate, or biased
reports• Under-reporting of events
• Cannot establish rates of events• Cannot compare event rates over time or between devices
• Often difficult to establish causality from event descriptions
• Variations in trade, product, and company names complicate searches
• Some information protected from public disclosure
10/30/2013 7
Alarm-Related Reports
• Each adverse event report is assigned one or more:• Device Product Code(s)• Device Problem Code(s)• Patient Problem Code(s)
• Codes are assigned by the reporter• Alarm-related reports were identified by
searching by Device Problem Codes containing the term “alarm” for 2010-2012
10/30/2013 8
Alarm Systems Management Webinar Series
Types of Alarm Problems
10/30/2013 9
DeviceAlarm
Systemissue
NotAudibleAlarm
Alarm,Audible
DelayedAlarm
ImproperAlarm
FalseAlarm
LowAudibleAlarm
DefectiveAlarm
AlarmNot
Visible
Alarm,Failure ofWarning
306
63
5 5 4 3 3 2 1 1
Medical Device Reports by Device Problem Codewith Alarm-Related Adverse Events
January 1, 2010 - December 31, 2012n=393 Occurrences
(of the 383 Death Reports)
Devices With Reported Alarm Problems
10/30/2013 10
215
3820 12 11 9 8
Medical Device Reports by Product Typewith Alarm-Related Adverse Events
January 1, 2010 - December 31, 2012n=313 Occurrences
(of the total 383 Death Reports)
Alarm Systems Management Webinar Series
Initial Steps to Meet the Goal of Reducing Alarm Fatigue
Shashi Avadhani, CBET, CCE, MBA Resident Regional Manager
Crothall Clinical Equipment Solutions
10/30/2013 11
National Patient Safety Goal 06.01.01 EP1
Elements of Performance EP1• Leaders establish alarm system safety as a
hospital priority• Formation of a multidisciplinary team• Representation from administration, nursing, medical
staff, clinical engineering• Identification of alarms in the clinical patient care
setting
10/30/2013 12
Alarm Systems Management Webinar Series
National Patient Safety Goal 06.01.01 EP2
Elements of Performance EP2• Identification of alarms• Risk assessment to patient care if alarms
unattended• Input from medical staff and clinical departments• Assessment of potential for patient harm based
on history• Identification of possible steps based on
published best practices
10/30/2013 13
Monitor
Ventilator
Infusion Pump
Bed Exit
Feeding Pump
Sequential Compression Device IABP
Devices With Alarms
Pulse Oximeter
1410/30/2013
Alarm Systems Management Webinar Series
Risk Assessment
Clinical Equipment Alarms Central Monitor
PRIORITY
A=Highest; could result in death if
unattendedB=High priority; may lead to unintended
consequence if unattended
C- Low priority; little risk if unattended
RISK
Level of Oversight
needed
A-High
B-Moderate
C-Low
Level of oversight typically available
Telemetry monitors ICU, PICU, NICU, ED (both), PEDS, 3G A A
A (enhance-ment on-
going)
10/30/2013 15
Priority of Alarm
• A = Highest; could result in death if unattended• B = High priority; could lead to unintended
consequence if unattended• C = Low priority; little risk if unattended
10/30/2013 16
Alarm Systems Management Webinar Series
Level of Oversight Needed Based on Priority
• A = High • To prevent harm to the patient
• B = Moderate• C = Low
10/30/2013 17
Current Status
• Level of oversight available, based on risk• e.g. monitor tech in case of central monitoring may
reduce risk, level of oversight available would be A
• Input from all areas of patient care• Based on multidisciplinary team’s input• Identify alarms that are high priority, that
currently do not possess the level of oversight required
10/30/2013 18
Alarm Systems Management Webinar Series
Clinical Equipment Alarms Central Mon.
PRIORITY
A=Highest; could result in death if unattendedB=High priority; may lead to unintended consequence if unattendedC- Low priority; little risk if unattended
RISK
Level of Oversight needed
A-High
B-Moderate
C-Low
Level of oversight typically available
Telemetry monitors ICU, PICU, NICU, ED (both), PEDS, 3G A A
A (enhancement
on-going)
Telemetry Monitors
10/30/2013 19
Clinical Equipment Alarms Central Monitor
PRIORITY
A=Highest; could result in death if unattended
B=High priority; may lead to unintended
consequence if unattended
C- Low priority; little risk if unattended
RISK
Level of Oversight
needed
A-High
B-Moderate
C-Low
Level of oversight typically available
IV Infusion pump/syringe pump B *B patient
specific
B *clinical judgment
to determine oversight
IV Infusion Pump/Syringe Pump
10/30/2013 20
Alarm Systems Management Webinar Series
Clinical Equipment Alarms Central Monitor
PRIORITY
A=Highest; could result in death if unattended
B=High priority; may lead to unintended
consequence if unattended
C- Low priority; little risk if unattended
RISK
Level of Oversight
needed
A-High
B-Moderate
C-Low
Level of oversight typically available
PCA Pump C C C
PCA Pump
10/30/2013 21
Other Steps and Best Practices
• Rounding to identify areas that have high noise levels due to alarms
• Follow published best practices, such as • Skin prep for ECG monitoring• Periodic electrode replacements• Regular battery changes on telemetry transmitters
10/30/2013 22
Alarm Systems Management Webinar Series
Methods of Collection of Alarm Data
• Manual• Continuous observation• Periodic rounding to assess most common alarms not
silenced and adding to background noise
• Automated• Manufacturer-provided tools• Middleware/secondary notification systems
10/30/2013 23
How Do We Get the Data and What Do We Do With It?
It’s All About the DataMelanie Quinton, MS
Solution Consultant/Consultant SpecialistCD BIO MDI
Kaiser Permanente
10/30/2013 24
Alarm Systems Management Webinar Series
How Do We Get the Data and What Do We Do With It?
• To ensure quality clinical decision support making we need to look at the data
• You do not need to be a clinician or understand medical terminology to retrieve data. You simply need to take ownership
• Where is the alarm data stored? – in primary system or in secondary system
• Who has access to the data? – work with your vendor and business partner to access
• How can the data be retrieved?
10/30/2013 25
How Do We Get the Data and What Do We Do With It?
• How will the data be presented?• Reports• Dashboards
• Begin with a baseline that is simple, and build from there• How many alerts are the nurses receiving per day?
• Compare day-to-day, month-to-month, and even season-to-season• What are the top 3 alerts the nurses are receiving?• Are the top 3 alerts actionable?
• Who will analyze the data?• What will be accomplished with the analyzed data?• Always continue to improve your alarm management program by
analyzing the numbers and through Key Performance Indicators
10/30/2013 26
Alarm Systems Management Webinar Series
Examples of How You Can Present the Data
• Dashboards• Reports/Charts
The following data is for illustration purposes only and is not real data. Some items are blurred
intentionally.
10/30/2013 27
Dashboards – Enterprise View
10/30/2013 28
Alarm Systems Management Webinar Series
Dashboards – Enterprise View
10/30/2013 29
Dashboards – Enterprise View
10/30/2013 30
Alarm Systems Management Webinar Series
Dashboards – Individual Medical Center View
10/30/2013 31
Dashboards – Set a Baseline, Look for Outliers
10/30/2013 32
Alarm Systems Management Webinar Series
Dashboards – Set a Baseline, Look for Outliers
10/30/2013 33
Dashboards – Make It Simple to Understand
10/30/2013 34
Alarm Systems Management Webinar Series
Create Reports List
10/30/2013 35
Filtering ReportsOur reports can be run by date range, criticality, and/or unit type, and we can drill
all the way down to the actual patient bed. There is no PHI.
10/30/2013 36
Alarm Systems Management Webinar Series
Charts – Number of Alerts in 24hrs Based on Criticality
10/30/2013 37
Charts – Type of Alerts Received by the Nurse, by Unit Type
10/30/2013 38
Alarm Systems Management Webinar Series
10/30/2013 39
Charts – Number of Alerts and Dispatches per Occupied Bed
Charts – Number of Heart Rate Alerts and Dispatches
10/30/2013 40
Alarm Systems Management Webinar Series
Reports – % Heart Rate Alerts and Dispatches
10/30/2013 41
What the Data Can Tell You• We have confirmed (by analyzing the number of alerts at Kaiser
Permanente Medical Centers) that the industry-leading physiological alerts are consistent with our results1. High Heart Rate = 24%2. Low Heart Rate = 12%3. Sp02 = 12%
• A majority of our heart rates1. 44% >120, but <1252. 33% >45, but <50
• We have also discovered that electrode replacement/hygiene has decreased our alerts by 22%* at our first pilot medical center
• 60 alerts per patient, per day (2.5 alerts per hour)* There may be other factors that account for the decrease, i.e. increased awareness and attention to the issue
10/30/2013 42
Alarm Systems Management Webinar Series
What You Can Do With the Data
10/30/2013 43
Alarm Management Assessment
Lisa A. Pahl, RN, BSN, MSNSr. Manager, Enterprise Consulting
Philips Healthcare Americas
10/30/2013 44
Alarm Systems Management Webinar Series
TechnologyProcessesPeople
Alarm Management AssessmentInterviews and Observation, Review, and Data
DATA
CULTURE
10/30/2013 45
Frequent Alarm SignalsOne patient’s alarms in 24 hours
46
633 Alarms*
A disruption every 2 minutes for the nurse and for the patient!
Impacts work flow, patient care, and healing*Does not include technical alarms
10/30/2013
Alarm Systems Management Webinar Series
One Patient’s Pages in 2.5 HoursJust for irregular heart rate
8:088:10
8:188:20
8:288:30
8:408:50
9:009:10
9:139:15
9:179:19
9:219:23
9:259:27
9:299:31
9:339:35
9:419:43
9:459:47
9:499:51
9:539:55
9:579:59
10:0210:04
10:0610:08
10:1010:12
10:2210:24
10:2610:28
10 irregular heart rate alarm signals translated into 43 alerts
because of delays in acknowledging the alarms signals
10/30/2013 47
What Alarm Signal Data Can You Capture?
Potential Categories of Monitoring Alarm Signals – Priority level may vary
48
You can select which alarm signals to route to a secondary notification system
10/30/2013
Life-Threatening Arrhythmia Alarm
Signals
• Asystole• Vfib/Tach• Extreme Tachy• Extreme Brady• Vtach
Other Arrhythmia Alarm Signals
• Ventricular• Beat Detection • Rate/ Rhythm
Detection• HR Limit
Violations
Bed/Non-Arrhythmia Alarm
Signals
• Apnea• SpO2 Desat• Invasive
Pressure Line Disconnect
• High or Low Limit Violations:
• SpO2• Resp• NBP• Invasive Pressure• Temp• CO2• Other
Technical Alarm Signals
• Multiple Types• Sensor
disconnect• Cannot analyze
ECG• Leads Off• Etc.—refer to IFU
Alarm Systems Management Webinar Series
Data: What Stands Out? • Understand the difference between alarm signal data
and secondary notification alert data—what stands out for each may vary
• Look at the big picture first and then drill down—total numbers of alarm signals compared to individual alarm signal occurrences
• Which unit(s) have the most alarm signals per patient bed?
• Which alarm signals are the most frequently occurring—across the institution and within a unit?
• Are there ‘patient outliers’?
10/30/2013 49
Unit Per Patient ComparisonPatient outliers
0 50 100 150 200 250 300 350 400 450
1
2
3
4
5
6
7
8
9
10
11
Number of Alarm Signals
Patie
nt Id
entif
ier
24 Hour Alarm Signal Totals Per Patient
10/30/2013 50
Alarm Systems Management Webinar Series
Numbers Alone Don’t Tell the Whole Story
3842
2617
1681
0
500
1000
1500
2000
2500
3000
3500
4000
4500
Alarm Signal TotalsICU PCU STEP DOWN
119112
94
0
20
40
60
80
100
120
140
Alarm Signals Per Patient BedPCU STEP DOWN ICU
10/30/2013 51
After the Data: Actionable Versus Non-Actionable?
Actionable Alarm Signal – Requires clinical intervention or some type of action• Life threatening, immediate response and action required• Change in patient status, requires action to reverse or prevent further deterioration• Requires action to correct a technical problem to assure proper patient monitoring,
e.g., leads off, SpO2 sensor disconnected
Non-Actionable Alarm Signal – Does not require a clinical intervention or action• Short duration, self correcting, e.g., SpO2 alarm signal• Intentional, e.g., suctioning or positioning/moving a patient• Triggered due to tight limits rather than actionable ones• False alarm:
• System itself incorrectly identifies an alarm condition• Something interferes with system causing it to detect an alarm, e.g., artifact or
low voltage-triggered asystole
5210/30/2013
Alarm Systems Management Webinar Series
Start With the BasicsElectrode Selection• Select electrodes and lead sets that provide the best continuing signal and adherence
with the least irritation• Electrodes for specific patient populations, e.g., diaphoretic patients
ECG Signal Quality• Select the best lead for monitoring purposes:
• Signal quality (gain, waveform criteria, learned and labeled appropriately)• Patient diagnosis, history, and reason for monitoring (e.g., ischemia or
arrhythmia detection)• Pacemaker identification• Single versus multi-lead monitoring• Assign “Leads Off” as a high-priority alarm
5310/30/2013
Start With the BasicsArtifact Prevention• Skin prep prior to electrode attachment• Electrode storage• Proactive electrode replacement and battery replacement (for telemetry
devices)
Patient and Family Education• Basic but often overlooked or limited• Explain why the patient is being monitored and what the patient can expect• Need to notify staff if electrode becomes loose or electrode or lead set
becomes detached• Importance of not ‘self’ disconnecting or silencing alarms• Talk with nurse if perception that alarms are occurring frequently and/or
impacting sleep or stress level
5410/30/2013
Alarm Systems Management Webinar Series
High-Priority Arrhythmia Alarms ICU Impact of artifact
0%10%20%30%40%50%60%70%80%90%
100%
DAY 1 DAY 2
TRUEFALSE
10/30/2013 55
TechnologyReview and Revise Default Settings/Configurations• Use data to prioritize which settings to address first• Evaluate limits for all parameters—balance limits, not too tight, not too wide—Melanie’s
example of majority of alerts between 120-125, high of 125• Assess if some medium-priority arrhythmia alarms can be defaulted off• Identify redundant settings and determine when they should be utilized or eliminated• Additional settings to review: delay times before an alarm is triggered, e.g., SpO2 low
limit, Desat, Apnea
Optimization• Proactive use of alarm pause/suspend when you know you will trigger an alarm• Monitor standby use• Utilization of NBP alarm suppression for SpO2
• Understand functionality and when to use and when not to, e.g., ST-segment monitoring
• Customization capabilities
5610/30/2013
Alarm Systems Management Webinar Series
Protocols to Guide Customization Practice and Decrease Non-Actionable Alarms
• These are designed to reduce non-actionable alarms by focusing on when and how to customize alarm settings
• Others to consider:• Monitoring admission criteria• 24-hour monitoring renewal• Transports• Change-of-shift report/handoffs• Daily huddles/rounding with
physician, nurse, and others to do a quick review of current settings and alarms in the past 24 hours
• New hire education and on-going training and competency assessments
Elements of Performance for NPSG.06.01.01
10/30/2013 57
Potential Impact of Customizing Alarms
0%10%20%30%40%50%60%70%80%90%
100%
Arrhythmia Alarms
One ICU Patient: 242 Alarm Signals in 24 Hours
AfibMultiform PVCsPVC Rate>10/minR-On-T PVCPair PVCs
112
92
2
333
10/30/2013 58
Alarm Systems Management Webinar Series
Eliminate Non-Actionable Alarm Signals
0%10%20%30%40%50%60%70%80%90%
100%
Arrhythmia Alarms
One ICU patient: 130 Alarms in 24 Hours?
AfibMultiform PVCsPVC Rate>10/minR-On-T PVCPair PVCs
92
2
333
10/30/2013 59
Eliminate Non-Actionable Alarm Signals
0%10%20%30%40%50%60%70%80%90%
100%
Arrhythmia Alarms
One ICU patient: 5 Alarms in 24 Hours?
AfibMultiform PVCsPVC Rate>10/minR-On-T PVCPair PVCs
2
3
A decrease of 98%
10/30/2013 60
Alarm Systems Management Webinar Series
Partnering Within Your Facility: Default Settings
• Compare units to assess what/if any variations currently • Use your Alarm Management Committee to evaluate and revise unit
defaults based on patient populations, care models, and data, and to determine which units should have the same defaults
• Test changes incrementally on a pilot unit to assess impact and outcomes
• Establish unit-based Alarm/Technology Champions to assist in implementing final changes, to function as a resource for the staff, to provide a feedback loop for the committee, and to share best practices across units
• Include the nurse managers, and physician and staff representatives from each unit in the decision-making process and to ensure definitions of actionable/clinically relevant are aligned
10/30/2013 61
Partnering Within Your Facility: Who Needs to Hear the Alarm Signals?
• TJC expectation that alarm signals are audible and recognizable above other unit noise
• All staff (clinical, ancillary, physicians, etc.) accountable to ensure alarm signals are acknowledged and responded to – role responsibilities and escalation processes clearly outlined
• With primary alarm systems and notification:• Do walk-throughs to validate that alarm signals can be heard and recognized
anywhere within the unit• If not, explore options (additional audible displays/clients, bedside monitor
capabilities that may allow other patient’s alarm signals to be observed in other rooms, monitor watchers, secondary notification systems)
• With monitor watchers: • What do the units still need to hear?• What alarm signals should the monitor watchers notify staff directly of and how
do they notify them (unit phone, ‘red phone’, overhead page, etc.)?• Escalation pathway for techs if no staff response
6210/30/2013
Alarm Systems Management Webinar Series
Partnering Within Your Facility: Who Needs to Hear the Alarm Signals?
• With secondary notification systems:• Well-thought-out process as to which alerts staff receive (e.g., high-
priority only)? Charge nurse to receive alerts for patients who are inadvertently not assigned in the secondary notification system?
• Decide who receives the alerts (nursing assistant receives leads off and replace battery alerts, respiratory therapy receives ventilator alerts, etc.)
• Does your system allow programmable delays before alerts are received or escalations to others if an alert is not addressed within a certain time frame?
• Use your Alarm Management Committee and other internal experts and resources to assist in evaluating and determining who needs to hear which alarm signals
6310/30/2013
In Conclusion
Any Alarm Management Strategy must be carefully thought out, tested, and continuously evaluated to ensure achieving the right balance of patient safety and quality of care with the reduction/elimination of nuisance alarms and alarm fatigue.
10/30/2013 64
Alarm Systems Management Webinar Series
Closing Reminders
Thank you for your time and attention!
Mark Your Calendars!How to Manage Alarms at the Bedside
Tuesday, December 3, 20131:00-2:00 pm Eastern
www.aami.org/htsi/events.html
6510/30/2013
Continuing Nursing Education1.0 contact hour
For those desiring CNE, please visit the link below for the test, evaluation form, and certificate:
http://www.aacn.org/DM/CETests/Overview.aspx?TestID=1025&mid=2864&ItemID=1017&menu=CETests
The American Association of Critical-Care Nurses (AACN) is accredited as a provider of continuing nursing education by the American Nurses Credentialing
Center’s Commission on Accreditation.
10/30/2013 66