12B18
7 SeriesMeters
Service Manual
PN 602730August 1996© 1996 Fluke Corporation, All rights reserved. Printed in U.S.A.All product names are trademarks of their respective companies.
riešenia na presné meranie
http://www.elso.sk
LIMITED WARRANTY & LIMITATION OF LIABILITY
Each Fluke product is warranted to be free from defects in material and workmanship undernormal use and service. The warranty period for Models 7-300, 7-600, and 12B is two years andbegins on the date of shipment. The warranty period for Model 18 is three years and begins on thedate of shipment. Parts, product repairs and services are warranted for 90 days. This warrantyextends only to the original buyer or end-user customer of a Fluke authorized reseller, and doesnot apply to fuses, disposable batteries or to any product which, in Fluke’s opinion, has beenmisused, altered, neglected or damaged by accident or abnormal conditions of operation orhandling. Fluke warrants that software will operate substantially in accordance with its functionalspecifications for 90 days and that it has been properly recorded on non-defective media. Flukedoes not warrant that software will be error free or operate without interruption.
Fluke authorized resellers shall extend this warranty on new and unused products to end-usercustomers only but have no authority to extend a greater or different warranty on behalf of Fluke.Warranty support is available if product is purchased through a Fluke authorized sales outlet orBuyer has paid the applicable international price. Fluke reserves the right to invoice Buyer forimportation costs of repair/replacement parts when product purchased in one country is submittedfor repair in another country.
Fluke’s warranty obligation is limited, at Fluke’s option, to refund of the purchase price, free ofcharge repair, or replacement of a defective product which is returned to a Fluke authorizedservice center within the warranty period.
To obtain warranty service, contact your nearest Fluke authorized service center or send theproduct, with a description of the difficulty, postage and insurance prepaid (FOB Destination), tothe nearest Fluke authorized service center. Fluke assumes no risk for damage in transit.Following warranty repair, the product will be returned to Buyer, transportation prepaid (FOBDestination). If Fluke determines that the failure was caused by misuse, alteration, accident orabnormal condition of operation or handling, Fluke will provide an estimate of repair costs andobtain authorization before commencing the work. Following repair, the product will be returned tothe Buyer transportation prepaid and the Buyer will be billed for the repair and returntransportation charges (FOB Shipping Point).
THIS WARRANTY IS BUYER’S SOLE AND EXCLUSIVE REMEDY AND IS IN LIEU OF ALLOTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING BUT NOT LIMITED TO ANYIMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE.FLUKE SHALL NOT BE LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL ORCONSEQUENTIAL DAMAGES OR LOSSES, INCLUDING LOSS OF DATA, WHETHERARISING FROM BREACH OF WARRANTY OR BASED ON CONTRACT, TORT, RELIANCE ORANY OTHER THEORY.
Since some countries or states do not allow limitation of the term of an implied warranty, orexclusion or limitation of incidental or consequential damages, the limitations and exclusions ofthis warranty may not apply to every buyer. If any provision of this Warranty is held invalid orunenforceable by a court of competent jurisdiction, such holding will not affect the validity orenforceability of any other provision.
Fluke Corporation Fluke Europe B.V.P.O. Box 9090 P.O. Box 1186Everett, WA 98206-9090 5602 BD EindhovenU.S.A. The Netherlands
5/94
riešenia na presné meranie
http://www.elso.sk
i
Table of Contents
Chapter Title Page
1 Introduction and Specifications........................................................ 1-1
Introduction ....................................................................................................... 1-1Specifications .................................................................................................... 1-1
2 Theory of Operation........................................................................... 2-1
Introduction ....................................................................................................... 2-1Analog Measurement IC (U1) ........................................................................... 2-1Microcomputer IC (U2)..................................................................................... 2-3Automatic Selection Input Resistance............................................................... 2-3
3 Maintenance ....................................................................................... 3-1
Introduction ....................................................................................................... 3-1Disassembly and Reassembly............................................................................ 3-1Cleaning............................................................................................................. 3-6Performance Tests ............................................................................................. 3-6Calibration Adjustments for Models 12B and 18.............................................. 3-9Calibration Adjustments for Models 7-300 and 7-600...................................... 3-10
4 Parts and Schematics........................................................................ 4-1
Introduction ....................................................................................................... 4-1How to Obtain Parts .......................................................................................... 4-1Manual Status Information................................................................................ 4-2Newer Instruments............................................................................................. 4-2Parts Lists, Drawings, and Schematics.............................................................. 4-2
riešenia na presné meranie
http://www.elso.sk
12/18/7 ServiceService Manual
ii
riešenia na presné meranie
http://www.elso.sk
iii
List of Tables
Table Title Page
1-1. Specifications......................................................................................................... 1-22-1. Electrical Components on Main PCA.................................................................... 2-22-2. U1 Pinout Table..................................................................................................... 2-43-1. Required Equipment .............................................................................................. 3-63-2. Performance Tests for Models 7-300 and 7-600 ................................................... 3-73-3. Performance Tests for Models 12B and 18 ........................................................... 3-84-1. Manual Status Information .................................................................................... 4-24-2. Models 12B, 18, and 18/AL Final Assembly ........................................................ 4-34-3. Models 7-300 and 7-600 Final Assembly.............................................................. 4-54-4. A1 Main PCA ........................................................................................................ 4-7
riešenia na presné meranie
http://www.elso.sk
12/18/7 ServiceService Manual
iv
riešenia na presné meranie
http://www.elso.sk
v
List of Figures
Figure Title Page
2-1. Typical Automatic Selection Input Resistance with DC Volts Function Selected 2-33-1. Models 12B and 18 Disassembled Unit................................................................. 3-33-2. Models 7-300 and 7-600 Disassembled Unit......................................................... 3-43-3. Removing and Reinserting the Printed Circuit Assembly .................................... 3-53-4. Calibration Adjustment.......................................................................................... 3-104-1. Models 12B and 18 Final Assembly...................................................................... 4-44-2. Models 7-300 and 7-600 Final Assembly.............................................................. 4-64-3. Models 12B, 18, 7-300, and 7-600 Main PCA Components ................................. 4-94-4. Model 12B Schematic............................................................................................ 4-104-5. Model 18 Schematic .............................................................................................. 4-124-6. Model 7-300 Schematic ......................................................................................... 4-144-7. Model 7-600 Schematic ......................................................................................... 4-16
riešenia na presné meranie
http://www.elso.sk
12/18/7 ServiceService Manual
vi
riešenia na presné meranie
http://www.elso.sk
MULTIMETER SAFETYThe Fluke 7-300, 7-600, 12B, and 18 Meters have been designed and tested according to IECPublication 1010, Safety Requirements for Electronic Measuring Apparatus. This manual containsinformation and warnings that must be followed to ensure safe operation and keep the meter insafe condition. Use of this equipment in a manner not specified herein may impair the protectionprovided by the equipment.
These multimeters comply with part 15 of the FCC Rules. Operation is subject to the following twoconditions: (1) these meters may not cause harmful interference, and (2) these meters mustaccept any interference received, including interference that may cause undesired operation.
Some common international electrical symbols are shown below:
AC: ALTERNATING CURRENT DANGEROUS VOLTAGE
DC: DIRECT CURRENT EARTH GROUND
EITHER AC OR DC CURRENT SEE EXPLANATION IN MANUAL
X FUSEDOUBLE INSULATION FORPROTECTION AGAINST ELECTRICSHOCK
Before using the meter, read the following safety information carefully. In this manual, "Warning" isreserved for conditions and actions that pose hazard(s) to the user; "Caution" is reserved forconditions and actions that may damage your meter.
• Avoid working alone.
• Follow all safety procedures for equipment being tested.
• Inspect the test leads for damaged insulation or exposed metal. Check test lead continuity.Damaged leads should be replaced.
• Be sure the meter is in good operating condition.
• Select the proper function for your measurement.
• To avoid electrical shock, use caution when working above 60V dc or 30V ac rms.
• Disconnect the live test lead before disconnecting the common test lead.
• Disconnect the power and discharge high-voltage capacitors before testing in resistance anddiodes or continuity.
• When making a current measurement, turn the circuit power off before connecting the meterin the circuit.
• Check meter fuses before measuring transformer secondary or motor winding current. Anopen fuse may allow high voltage build-up, which is potentially hazardous.
• Use clamp-on probes when measuring circuits exceeding 10 amps.
• When servicing the meter, use only the replacement parts specified.
• Do not allow the meter to be used if it is damaged or if its safety features are impaired.
riešenia na presné meranie
http://www.elso.sk
1-1
Chapter 1Introduction and Specifications
Introduction
PWarningService procedures described herein should be performed byqualified personnel only. To avoid electric shock, perform onlythose service procedures described in this manual.
The 12B, 18, 7 Series Service Manual provides the information necessary to service theFluke Model 12B, Model 18, and Models 7-300 and 7-600 meters. This manual providesthe following information:
• Specifications (Chapter 1)
• Basic theory of operation (Chapter 2)
• Disassembly and reassembly (Chapter 3)
• Performance tests (Chapter 3)
• Calibration (Chapter 3)
• Illustrated parts lists and schematic diagrams (Chapter 4)
Refer to the users instruction sheet for operating instructions.
SpecificationsSpecifications are in Table 1-1. Accuracy is specified for a period of one year aftercalibration, at 18ºC to 28ºC (64ºF to 82ºF) with relative humidity to 90%. ACconversions are ac-coupled, average responding, and calibrated to the rms value of a sinewave input.
riešenia na presné meranie
http://www.elso.sk
12B/18/7 SeriesService Manual
1-2
Accuracy specifications are given as follows:
±([% of reading] + [number of least significant digits])
Table 1-1. Specifications
Maximum Voltage Between anyTerminal and Earth Ground(excludes 10% tolerance)
600V ac rms or dc
Display 3-3/4-digits, 4000 counts, updates 4/sec
Operating Temperature -10°C to 50°C
Storage Temperature -30°C to 60°C indefinitely (to -40°C for 100 hrs)
Temperature Coefficient 0.1 x (specified accuracy)/°C (<18°C or >28°C)
Relative Humidity 0% to 90% (-10°C to 35°C)
0% to 70% (35°C to 50°C)
Battery Type 9V, NEDA 1604 or IEC 6F22
Battery Life 650 continuous hours with alkaline; 600 hours for Model 7450 continuous hours with carbon-zinc; 400 hours for Model 7
Shock, Vibration 1 meter shock. Per MIL-T-28800D for a Class 3 Instrument
Size (HxWxL) 1.35 in x 2.75 in x 5.55 in(3.46 cm x 7.05 cm x 14.23 cm)
Weight 10 oz (286g)
EMI Regulations Complies with FCC Part 15, Class B, and VDE 0871B.
Safety Designed to Protection Class II requirement of UL1244,ANSI/ISA-S82.01 - 1988, CSA C22.2 No 231, and VDE 0411, andIEC 1010-1 overvoltage category III (CAT III), 600V.
Trademark of TÜV Product Services.IEC 348 certification; designed tocomply with EN 61010-1: 1993.
riešenia na presné meranie
http://www.elso.sk
Introduction and SpecificationsSpecifications 1
1-3
Table 1-1. Specifications (continued)
Accuracy (50 to 400 Hz)
Function Range Resolution 7-300 7-600 12B/18
f 4000 mV1
4.000V
40.00V
300.0V
400.0V
600.0V
1 mV
0.001V
0.01V
0.1V
0.1V
1V
NA
NS2
±(2.9%+3)
±(2.9%+3)
NA
NA
NA
NS2
±(2.9%+3)
NA
±(2.9%+3)
±(2.9%+3)
±(1.9%+3)
±(1.9%+3)
±(1.9%+3)
NA
±(1.9%+3)
±(1.9%+3)
E 4000 mV1
4.000V
40.00V
300.0V
400.0V
600V
1 mV
0.001V
0.01V
0.1V
0.1V
1V
NA
NS2
±(1.5%+1)
±(1.5%+1)
NA
NA
NA
NS2
±(1.5%+1)
NA
±(1.5%+1)
±(1.5%+1)
±(0.9%+2)
±(0.9%+2)
±(0.9%+1)
NA
±(0.9%+1)
±(0.9%+1)
J 400.0Ω
4.000 kΩ
40.00 kΩ
400.0 kΩ
4.000 MΩ
40.00 MΩ
0.1Ω
0.001 kΩ
0.01 kΩ
0.1 kΩ
0.001 MΩ
0.01 MΩ
±(1.5%+2)
NA
NA
NA
NA
NA
±(1.5%+2)
NA
NA
NA
NA
NA
±(0.9%+2)
±(0.9%+1)
±(0.9%+1)
±(0.9%+1)
±(0.9%+1)
±(1.5%+3)
K 1.000 µF
10.00 µF
100.0 µF
1000 µF
10000 µF
0.001 µF
0.01 µF
0.1 µF
1 µF
10 µF
NA
NA
NA
NA
NA
NA
NA
NA
NA
NA
±(1.9%+2)
±(1.9%+2)
±(1.9%+2)
±(1.9%+2)
±(10%+90)typical
LM3 2.000V 0.001V NA NA ±(0.9%+2)
1. The 4000 mV range can be entered only in the manual range mode. Use the 4000 mV range withaccessories. The 4000 mV range is not available on the Model 7.
2. Not specified for the Model 7.
3. The beeper is guaranteed to come on at <25Ω and turn off at >250Ω. The meter detects opens orshorts of 250 µs or longer. These values are not specified for the Model 7.
riešenia na presné meranie
http://www.elso.sk
12B/18/7 SeriesService Manual
1-4
Table 1-1. Specifications (continued)
FunctionOverload
Protection1
InputImpedance(Nominal)
Common ModeRejection Ratio
(1 kΩUnbalanced)
2Normal Mode
Rejection2
f600V rms or dc >5 MΩ <100pF
2
Automatic Selectionand LoZ = >2 kΩ<200pF(ac coupled)
3
>60 dB at dc 50 or60 Hz
E600V rms or dc >10MΩ <100pF
2
Automatic Selectionand LoZ = >2 kΩ<200pF
3
>100 dB at dc, 50 or60 Hz
>50 dB at 50 Hz or60 Hz
Open CircuitTest Voltage
Full Scale VoltageTo 4.0 MΩ 40 MΩ
Short CircuitCurrent
J600V rms or dc <1.5V dc <450 mV
dc<1.5V dc <500 µA
L2 600V rms or dc 2.4-3.0V dc 2.400V dc 0.95 mA (typical)
1. 3 x 106 V Hz maximum.
2. Does not apply to Model 7.
3. ~2 kΩ input impedance up to 50V. Impedance increases with input voltage to >300 kΩ at 600V.
MIN MAX Recording Accuracy and Response Time (Models 12B and 18)
Specified accuracy of the measurement function ±12 digits for changes >200 ms in duration(±40 digits in ac). Typical 100 ms response to 80%.
MIN MAX Recording with Elapsed Time (Models 12B and 18)
Elapsed Time: 0 to 100 hours (99.59)
Resolution: 1 minute
Accuracy: 0.3% typical
Continuity Capture (Model 12B)(Open/Short Capture Model 18)
Detects opens or shorts of 250 µs or longer.
riešenia na presné meranie
http://www.elso.sk
2-1
Chapter 2Theory of Operation
IntroductionChapter 2 provides a basic theory of operation for the Models 7, 12B, and 18 Meters.Electrical components on the printed circuit assembly (A1 Main PCA) are listed in Table2-1. Refer to Figure 4-3 for the location of these components and Figures 4-4 through4-7 for the schematic diagrams.
The analog/digital IC (U1) performs the electrical measurement functions. See Table 2-2for pin names and descriptions. The microcomputer (U2) controls U1, the LCD (U3),and the user interface. Discrete components support U1 and U2, provide referencestandards for measurements, and provide input overload protection.
Analog Measurement IC (U1)U1 performs the following analog functions: a/d converter, ac to dc converter,Automatic Selection circuitry, active filter, passive filter, power supply, rangeconfiguration circuitry, signal routing circuitry, beeper driver, digital control circuitry,and digital U2 interface circuitry.
The a/d converter is a patented dual-rate, dual-slope converter. The dual-rate conversionallows for MIN MAX (Models 12B and 18 only) and fast autoranging functions. The acto dc converter is full-wave rectified and average-responding. The active and passivefilters are two-pole and one-pole low-pass filters (respectively) that are used for signalfiltering prior to a/d conversion. The internal power supply generates a ground voltagenominally +3V relative to VSS. Range configuration circuitry connects the Z1 resistornetwork as needed for different ranges. Routing circuitry connects the various signalconditioning circuits as needed. The a/d converter and a counter are controlled by a statemachine.
riešenia na presné meranie
http://www.elso.sk
12B/18/7 SeriesService Manual
2-2
Table 2-1. Electrical Components on Main PCA
Circuit and Function Component Designator(s)
Analog Measurement IC
Microcomputer IC
Input Divider and Ohms Reference Resistor Network
Input Divider AC Coupling Capacitor
J2 Input Receptacle Voltage Sense Resistor
J1 Input Receptacle Voltage Sense Resistor
Reference Voltage for Volts Measurements
A/D Converter Integrate Capacitor
A/D Converter Autozero Capacitor
A/D Converter Gain Resistors
Active Filter Components
Passive Filter Components
AC to DC Converter Gain Resistors
AC to DC Converter AC Coupling Capacitor
Bias Current Setting Resistor
DGND-VSS Voltage Setting Resistors
Power Supply Bypass Capacitors
Reverse Battery Protection
System Clock
Automatic Selection Input Current Limiters
Automatic Selection Input Positive Temp. Coef. Thermistor
Automatic Selection Input Voltage Clamp Circuit
Automatic Selection Input Fusible Resistor
Microcomputer Reset Circuit
On/Off and Automatic Selection Select Slide Switch
Beeper
Beeper Current Limiter
Slide Switch Protection
Switch Sense Bypass Capacitor
Input Divider Bypass Capacitor
Zero Ohm Jumpers
U1
U2
Z1
C14
R17
R14
VR1, R4, R5, R6, R26
C2
C1
R1, R3, R15
R7, R8, C5, C6
R9, C7
R11, R12, R13
C8
R2
R10, R24
C3, C15
CR1
Y1
R16, R18, R29
RT1
Q1, Q2, CR2, R28
R19
R23, R25, C10, C13, CR3
S1
LS1
R22
E1
C9
C16
R20, R21, R27
riešenia na presné meranie
http://www.elso.sk
Theory of OperationMicrocomputer IC (U2) 2
2-3
Finally, digital circuitry interfaces with U2 via a parallel address and bidirectional databus.
Voltage is measured using a ratio comparison of the unknown voltage to the referencevoltage (REFI pin). Resistance is calculated using a ratio comparison of the voltageacross the unknown resistor to the voltage across a precision reference resistor, with thesame current in both. Capacitance is measured by determining the amount of chargeadded for a given dc voltage change.
Microcomputer IC (U2)U2 writes range settings and a/d converter information to U1. U2 reads a/d converterresults and status information. This includes the low battery check, slide-switch position,continuity check, and Automatic Selection data. The microcomputer performs mathoperations on the raw data from U1 and configures it for the LCD. U2 also reads push-button inputs. Finally, the 2.1 MHz clock signal at U2 is divided down to 131 kHz andsent to U1 (CLK pin) for the counter.
Automatic Selection Input ResistanceAs shown in Figure 2-1, the input resistance for the Automatic Selection function is non-linear. The data in the graph are for the volts dc function (any range) and are the steadystate values obtained after the PTC thermistor (RT1) has stabilized. The data also applyfor the volts ac function.
500
100
50
10
5
1
0.1 0.5 1.0 5.0 10.0 50.0 500.0
Sm
all S
igna
l Inp
ut T
esis
tanc
e (k
Ω)
|Vin| (Vdc)
Positive DCV Input
Negative DCV Input
as01f.eps
Figure 2-1. Typical Automatic Selection Input Resistance with DC Volts Function Selected
riešenia na presné meranie
http://www.elso.sk
12B/18/7 SeriesService Manual
2-4
Table 2-2. U1 Pinout Table
Pin No. Pin Name Description
1 READ (L) When low (VSS), data from addressed register can be read
2 ADR5 Most significant address bus line
3 ADR4 Bit 4 of parallel address bus
4 ADR3 Bit 3 of parallel address bus
5 ADR2 Bit 2 of parallel address bus
6 ADR1 Bit 1 of parallel address bus
7 ADR0 Least significant address bus line
8 VSS Negative power supply voltage (-3V relative to DGND)
9 VDD Positive power supply voltage for analog only (Vbat-3V=VDD-DGND)
10 K0 1st a/d gain resistor pin, generally for de-integrate
11 K1 2nd a/d gain resistor pin, generally for integrate
12 K2 3rd a/d gain resistor pin, for autozero
13 AZ A/D converter pin for autozero capacitor
14 INT A/D converter pin for integrator capacitor
15 REFI 1.00V reference voltage for de-integrate signal in volts
16 BIAS Pin for analog bias current generator reference resistor
17 REFH Connects 1.235V reference voltage to on-chip circuits
18 BGND Not used
19 AFO Active filter high output pin (2-pole, low-pass filter)
20 FAO Internal active filter node
21 FAI Internal active filter node
22 AFI Active filter high input pin
23 VSS Negative power supply voltage (-3V relative to DGND)
24 PFO Passive filter low output pin (1-pole, low-pass filter)
25 AVAOM Absolute value amp (full wave rectified) negative output pin
26 AVAOP Absolute value amp (full wave rectified) positive output pin
27 AVAM Absolute value amp inverting input (summing node)
28 ACBO AC buffer output
29 VSET Voltage divider sense for power supply
30 PSTEST Pin to disable on-chip power supply for U1 testing
31 DGND Ground power supply pin connected to common (digital ground)
32 AGND Common input sense line. No current flow (analog ground).
33 CLAMP Op amp output voltage for Q2 clamp transistor base drive
34 AMPS Shunt resistor sense pin. Unused in Fluke Model 18.
35 ISRC Current source pin for continuity, ohms, and capacitance
36 SWS Slide-switch position sense pin. Internal pull down.
37 OHMS Ohms sense resistor input pin. Sense voltage at J2.
38 VSS Negative power supply voltage (-3V relative to DGND)
39 ACV AC volts input pin from 10 MΩ resistor
riešenia na presné meranie
http://www.elso.sk
Theory of OperationAutomatic Selection Input Resistance 2
2-5
Table 2-2 U1 Pinout Table (continued)
Pin No. Pin Name Description
40 APV0 DC volts input pin from 10 MΩ resistor
41 DIVLO Input divider (Z1) common (low) pin
42 APV1 10-to-1 voltage divider and 1 MΩ reference resistor input
43 APV2 100-to-1 voltage divider and 100 kΩ reference resistor input
44 APV3 1000-to-1 voltage divider and 10 kΩ reference resistor input
45 APV4 10,000-to-1 voltage divider and 1 kΩ reference resistor input
46 GND Ground power supply pin connected to common (digital ground)
47 DATA3 Most significant bidirectional data bus line
48 DATA2 Bit 2 of parallel data bus
49 DATA1 Bit 1 of parallel data bus
50 DATA0 Least significant bidirectional data bus line
51 N/C No connection
52 BEEPER (L) One of two beeper drive lines. Voltage swings VDD to VSS.
53 VSS Negative power supply line (-3V relative to DGNG)
54 VSS Negative power supply line (-3V relative to DGNG)
55 CLK System clock line from U2. 131,072 Hz.
56 BCLK Beeper frequency (2.3 kHz) clock line from U2
57 TESTCLK Test clock pin for U1 testing
58 BEEPER One of two beeper drive lines. Voltage swings VDD to VSS.
59 N/C No connection
60 WRITE (L) When driven low (VSS), data is written to addressed register
riešenia na presné meranie
http://www.elso.sk
12B/18/7 SeriesService Manual
2-6
riešenia na presné meranie
http://www.elso.sk
Some semiconductors and custom IC's can bedamaged by electrostatic discharge duringhandling. This notice explains how you canminimize the chances of destroying such devicesby:
1. Knowing that there is a problem.2. Leaning the guidelines for handling them.3. Using the procedures, packaging, and bench techniques that are recommended.
The following practices should be followed to minimize damage to S.S. (static sensitive) devices.
1. MINIMIZE HANDLING
2. KEEP PARTS IN ORIGINAL CONTAINERS UNTIL READY FOR USE.
3. DISCHARGE PERSONAL STATIC BEFORE HANDLING DEVICES. USE A HIGH RESIS- TANCE GROUNDING WRIST STRAP.
4. HANDLE S.S. DEVICES BY THE BODY.
static awarenessA Message From
Fluke Corporation
riešenia na presné meranie
http://www.elso.sk
5. USE STATIC SHIELDING CONTAINERS FOR HANDLING AND TRANSPORT.
6. DO NOT SLIDE S.S. DEVICES OVER ANY SURFACE.
7. AVOID PLASTIC,VINYL AND STYROFOAM IN WORK AREA.
8. WHEN REMOVING PLUG-IN ASSEMBLIES HANDLE ONLY BY NON-CONDUCTIVE EDGES AND NEVER TOUCH OPEN EDGE CONNECTOR EXCEPT AT STATIC-FREE WORK STATION. PLACING SHORTING STRIPS ON EDGE CONNECTOR HELPS PROTECT INSTALLED S.S. DEVICES.
9. HANDLE S.S. DEVICES ONLY AT A STATIC-FREE WORK STATION.
10. ONLY ANTI-STATIC TYPE SOLDER- SUCKERS SHOULD BE USED.
11. ONLY GROUNDED-TIP SOLDERING IRONS SHOULD BE USED.
PORTIONS REPRINTEDWITH PERMISSION FROM TEKTRONIX INC.AND GERNER DYNAMICS, POMONA DIV.
Dow Chemical
riešenia na presné meranie
http://www.elso.sk
3-1
Chapter 3Maintenance
IntroductionChapter 3 covers the following:
• Disassembly and reassembly
• Cleaning
• Performance tests
• Calibration
Disassembly and Reassembly
P WarningTo avoid electrical shock, remove test leads and any inputsignals before opening the case.
CautionTo avoid contamination from the fingers, handle the pca by theedges or wear gloves. PCA contamination can cause failures inhumid environments. This meter contains components that canbe damaged by static discharge. To avoid damaging thesecomponents when servicing the meter, take precautionsindicated on the "Static Awareness" at the beginning ofChapter 3.
riešenia na presné meranie
http://www.elso.sk
12B/18/7 SeriesService Manual
3-2
Referring to Figures 3-1 and 3-2 as necessary, disassemble the meter as follows. APhillips-head screwdriver and small flat-blade screwdriver are required.
1. Remove the test leads and set the slide switch or rotary knob to OFF.
2. Remove the Phillips-head screws (H1-4) from the case bottom (MP8).
3. Separate the case top (MP2) from the case bottom.
4. TO REPLACE THE BATTERY: Lift the battery from the case bottom and insert anew 9V battery (NEDA 1604, 6F22, or 006P). Be sure the positive and negativebattery posts are oriented correctly.
5. TO REMOVE THE PCA (A1): Insert a small, flat-blade screwdriver between theedge of the case top and the pca where shown in Figure 3-3. Gently unsnap a side ofthe case top from the pca. Repeat on the other side of the pca. Unsnap the case fromthe top of the pca last.
6. LIFT THE PCA FROM THE CASE TOP BY ITS EDGES. If the elastomeric contactstrips (J3, J4)* for the switch assembly (S2)* and LCD (U3) are stuck to the pca,remove them without touching the conductive edges.
TO REINSERT THE PCA: Important: First make sure that the slide-switchactuator (MP5), the slide switch (S1), and the rotary knob (for the 12B and 18) arein the OFF position. Place the pca over the four screw posts in the case top, thenpress gently on the center of the pca while using the small flat-edge screwdriver toshoehorn the pca under the snap on a side of the case top. Repeat on the other sideand the top.
7. TO REMOVE THE SWITCH SUPPORT (MP6): Use a small, flat-bladescrewdriver to gently unsnap the sides and top of the switch support from the snapsshown in Figure 3-3.
8. The LCD, switch assembly, slide-switch actuator, and elastomeric contact strips(J3, J4) for the LCD and switch assembly are accessible and can be replaced asneeded. Do not allow the LCD to get wet. Before installing a new LCD, make surethat all connector contact points are clean.
CautionDo not touch the conductive edges of the elastomeric strips orthe contacts on the switch assembly. If they are contaminated,clean them with isopropyl alcohol.
9. Reassembling the meter is the reverse of disassembling it. After the meter isreassembled, execute the PERFORMANCE TESTS to confirm that the meter isworking properly.
* J4 and S2 are not present in Models 7-300 and 7-600.
riešenia na presné meranie
http://www.elso.sk
MaintenanceDisassembly and Reassembly 3
3-3
Case, Top
Switch Assembly
Actuator, Switch
Main PCA
Shield, Bottom
Foot
LCD
Conn.Elastomeric
Conn.ElastomericLCD to PCA
Support, Switch Assembly
Battery
Shock Absorber
Case, Bottom
Screw, THD Form, PH.P.STL,5-14 x .750
as07f.eps
Figure 3-1. Models 12B and 18 Disassembled Unit
riešenia na presné meranie
http://www.elso.sk
12B/18/7 SeriesService Manual
3-4
Case, Top
Actuator Switch
Main PCA
Shield, Bottom
Foot
LCD
Conn.ElastomericLCD to PCA
Support, Switch Assembly
Shock Absorber
Battery
Case, Bottom
Screw, THD Form, PH.P.STL,5-14 x .750
as10f.eps
Figure 3-2. Models 7-300 and 7-600 Disassembled Unit
riešenia na presné meranie
http://www.elso.sk
MaintenanceDisassembly and Reassembly 3
3-5
1
2
3
Snap
SnapSnaps
as04f.eps
Figure 3-3. Removing and Reinserting the Printed Circuit Assembly
riešenia na presné meranie
http://www.elso.sk
12B/18/7 SeriesService Manual
3-6
Cleaning
CautionTo avoid damaging the meter, do not use aromatichydrocarbons or chlorinated solvents for cleaning. Thesesolutions will react with the plastics used in the instruments.Do not get the LCD wet.
To clean the case, wipe it with a cloth lightly dampened with water and a mild detergent.
Wash the pca with isopropyl alcohol or hot deionized water and a soft brush. Do not usedetergent of any kind for cleaning the pca. The pca must be completely dry before themeter is reassembled. Dry the pca with clean dry air at low pressure (<20 psi); then bakeit at 50ºC for 2 hours.
Performance Tests
P WarningTo avoid electric shock, do not execute the performance testsprocedures unless the meter is fully assembled.
Use the PERFORMANCE TESTS to confirm that the meter is working properly. If themeter fails any of these tests, it needs calibration (see CALIBRATION) or repair. Theequipment required is specified in Table 3-1.
1. Connect the calibrator to the [+] and COM jacks on the meter.
2. Referring to Table 3-2 for the Fluke 7-300 or 7-600, or Table 3-3 for the Fluke 12Bor 18, put the meter in the function and range shown for Test 1.
3. Apply the input from the appropriate source. The reading on the display should bewithin the MINIMUM and MAXIMUM values shown in Table 3-2 or 3-3.
4. Test the remaining functions and ranges.
Table 3-1. Required Equipment
Equipment Minimum Specifications Recommended Models
DMM Calibrator DC Voltage: 0-600V
Accuracy: ±0.25%
AC Voltage: 0-600V
Accuracy: ±0.5%
Frequency: 50-400 Hz
Fluke Models 5700A, 5500A,5100B, or equivalent
Decade Resistor Resistance: 1.0-40 MΩ
Accuracy: ±0.25%
General Resistance RDS-77Band 41B or Fluke 5500A
Decade Capacitor Capacitance: 0-1.000 µF
Accuracy: ±0.5%
GenRad 1412-BC or Fluke 5500A
riešenia na presné meranie
http://www.elso.sk
MaintenancePerformance Tests 3
3-7
Table 3-2. Performance Tests for Models 7-300 and 7-600
Test No. ApplicableModel(s)
1MeterRange
Input ToMeter
DisplayMinimum
DisplayMaximum
1
2
3
4
5
6
7
8
7-300/7-600
7-300/7-600
7-300/7-600
7-300/7-600
7-300/7-600
7-300
7-600
7-600
40.00V ac
40.00V ac
40.00V ac
400.0V ac(300.0V ac for7-300)
400.0V ac(300.0V ac for7-300)
300V ac
400.0V ac
600V ac
4.40V, 50 Hz
35V, 50 Hz
35V, 400 Hz
100V, 50 Hz
150V, 400 Hz
300V, 50 Hz
350V, 80 Hz
600V, 400 Hz
4.24
33.95
33.95
96.8
145.4
291
339.5
580
4.56
36.05
36.05
103.2
154.6
309
360.5
620
9
10
11
12
13
14
7-300/7-600
7-300/7-600
7-300/7-600
7-300/7-600
7-300/7-600
7-600
40.00V dc
40.00V dc
400.0V dc
400.0V dc(300.0V dc for7-300)
400.0V dc(300.0V dc for7-300)
600V dc
4.40V dc
-35V dc
+100V dc
-150V dc
300V dc
-600V dc
4.32
-35.54
+98.4
-152.3
295.4
-610
4.48
-34.46
+101.6
-147.7
304.6
-590
152
162
173
7-300/7-600
7-300/7-600
7-300/7-600
400.0Ω
400.0Ω
400.0Ω
0.0Ω
1.0Ω
350.0Ω
0.0
0.8
344.5
0.2
1.2
355.5
1. To avoid overload/current limit condition on the calibrator when testing the Model 7, defeat the 50Ω divider on thecalibrator and allow at least 3 seconds at each voltage interval above 40V.
2. Model 7 beeper should be on.
3. Model 7 beeper may be on or off.
riešenia na presné meranie
http://www.elso.sk
12B/18/7 SeriesService Manual
3-8
Table 3-3. Performance Tests for Models 12B and 18
TestNo.
Meter Range(Automatic Selection 12B Only)
Input ToMeter
SwitchPosition
DisplayMinimum
DisplayMaximum
1
2
3
4
5
6
7
8
92
10
4.000V ac
4.000V ac, MIN MAX
4.000V ac
4.000V ac
4.000V ac
4.000V ac
40.00V ac
40.00V ac
400.0V ac
600V ac
0V
0V
10 mV, 50 Hz
10 mV, 400 Hz
3.5V, 50 Hz
3.5V, 400 Hz
35V, 50 Hz
35V, 400 Hz
350V, 80 Hz
600V, 400 Hz
Center
Center
Center
Right1
Right1
Center
Center
Right1
Right1
Center
0.000
0.000
0.007
0.007
3.430
3.430
34.30
34.30
343.0
586
0.003
0.040
0.013
0.013
3.570
3.570
35.70
35.70
357.0
614
11
12
13
14
15
16
17
18
4.000V dc, MIN MAX
4.000V dc
40.00V dc
400.0V dc
4.000V dc
40.00V dc
400.0V dc
600V dc
0V
0V
+20 mV dc
-200 mV dc
+3.5V
-35V dc
+100V dc
-600V dc
Center
Right1
Right1
Center
Center
Center
Center
Center
-0.012
-0.002
00.01
-000.3
+3.466
-35.33
+99.0
-606
0.012
0.002
00.03
-000.1
+3.534
-34.67
+101.0
-594
19
203
213
22
23
24
25
26
27
28
29
30
315
325
Continuity
Continuity, Automatic Selection
Continuity, Automatic Selection
400.0Ω
400.0Ω
400.0Ω
4.000 kΩ
40.00 kΩ
400.0 kΩ
4.000 MΩ
40.00 MΩ
Ω, Automatic Selection
1.000 µF
1.000 µF
+2V dc3
+3.43V dc3
-0.5V dc3
0.0Ω
1.0Ω
350.0Ω
1.0 kΩ
35 kΩ
100 kΩ
2.9 MΩ
35 MΩ
1.96V dc
0.0 µF
0.95 µF
Right
Right
Right
Right
Right
Right
Right
Right
Right
Right
Right
Right
Right
Right
1.980
3.3974
-0.5074
0.0
0.8
346.6
0.990
34.67
99.0
2.873
34.44
1.9404
-0.001 µF
0.930 µF
2.020
3.4634
-0.4934
0.2
1.2
353.4
1.010
35.33
101.0
2.927
35.56
1.9804
0.001 µF
0.970 µF
1. Center position for Model 18.
2. In the Automatic Selection mode, the UUT uses a low-impedance thermistor (~2.5k for circuit protection and loadtesting (referred to as low-Z input circuitry). When using the 5100B, 5500A, or 5700A to drive the UUT with highvoltages, avoid an overload/current limit condition by gradually stepping the voltage up (waiting two secondsbetween each step) from 90.0V, 120.0V, 180.0V, and 350.0V at 80 Hz each step.
3. Calibrator 50Ω divider override.
4. The dc volts annunciator must be on. Test does not apply to Model 18.
5. Conducting performance tests of the 400Ω, 4 kΩ, 40 kΩ, and 1 µF ranges (tests no. 22, 23, 24, 25, 26, 31, and 32)verifies that the discrete and integrated circuitry needed to support the other capacitance ranges are working withinspecifications. Therefore, the tests indirectly verify that the meter will meet specification in the 10 µF, 100 µF,1000 µF, and 10,000 µF ranges.
riešenia na presné meranie
http://www.elso.sk
MaintenanceCalibration Adjustments for Models 12B and 18 3
3-9
Calibration Adjustments for Models 12B and 18To ensure that the meter performs to specifications, make calibration adjustmentsannually using the following procedure:
1. Set the calibrator for 0V dc. Put the meter in the 4.000V dc range.
2. Connect the calibrator to the [+] and COM jacks on the Meter.
3. Apply an input of +4.000V dc ±0.25%
The meter display should read between 3.997-4.003V. If it does not, adjust R4 (seeFigure 3-4) as described below.
To adjust R4, proceed as follows:
1. Remove any input signals to the meter.
2. Remove the four screws on the back and separate the case bottom and case top.
Notice that when you do so the battery remains in the case bottom and power to themeter is disconnected.
3. Observing correct polarity, connect a 9V battery to the battery contacts using easyhook jumpers or alligator clip leads (see Figure 3-4).
Note To avoid stretching or bending the battery contacts, connect leads to thebase of the contacts as shown in Figure 3-4.
4. Set the calibrator for 0V dc. Put the meter in the 4.000V dc range.
5. Connect the calibrator to the [+] and COM jacks on the meter.
6. Apply an input of +4.000V dc ±0.25%
7. Adjust R4 (see Figure 3-4) so that the meter display reads between 3.997 and4.003V.
riešenia na presné meranie
http://www.elso.sk
12B/18/7 SeriesService Manual
3-10
+ _9V
BATTERY
+
Clip to Base Contacts
Adjust R4 for 3.997-4.003V DC
Input Receptacles +
as05f.eps
Figure 3-4. Calibration Adjustment
Calibration Adjustments for Models 7-300 and 7-600To ensure that the meter performs to specifications, make calibration adjustmentsannually using the following procedure:
1. Set the calibrator for 0V dc. Switch the meter ON.
2. Connect the calibrator to the [+] and COM jacks on the meter. Defeat the 50Ωdivider on the calibrator (if any).
3. Apply an input of 3.500V dc ±0.25%. Wait at least 3 seconds.
4. Apply an input of 3.75V dc ±0.25%. Within 5 seconds the meter should display areading between 3.747 and 3.753V dc. If it does not, adjust R4 as described below.
To adjust R4, proceed as follows:
1. Remove any input signal to the meter.
2. Remove the four screws on the back and separate the case bottom from the case top.
Notice that when you do so, the battery remains in the case bottom and power to themeter is disconnected.
3. Observing the correct polarity, connect a known good 9V battery to the batterycontacts using easy-hook jumpers or alligator clip leads. Refer to Figure 3-4.
Note To avoid stretching or bending the battery contacts, connect the leads tothe base of the contacts, as shown in Figure 3-4.
4. Set the calibrator for 0V dc. Turn the meter ON.
riešenia na presné meranie
http://www.elso.sk
MaintenanceCalibration Adjustments for Models 7-300 and 7-600 3
3-11
5. Connect the calibrator to the [+] and COM input jacks on the meter. Defeat the 50Ωdivider on the calibrator (if any).
6. Apply an input of 3.500V dc ±0.25%. Wait at least 3 seconds.
7. Apply an input of 3.75V dc ±0.25%. Wait 5 seconds.
8. Adjust R4 so that the meter displays a reading between 3.747 and 3.753V dc.
9. Remove the inputs to the meter, disconnect the battery, and reassemble the meter.
riešenia na presné meranie
http://www.elso.sk
12B/18/7 SeriesService Manual
3-12
riešenia na presné meranie
http://www.elso.sk
4-1
Chapter 4Parts and Schematics
IntroductionThis chapter contains an illustrated list of replaceable parts for the 7-300, 7-600, 12B,and 18 Meters. Parts are listed by assembly; alphabetized by reference designator. Eachassembly is accompanied by an illustration showing the location of each part and itsreference designator. The parts lists give the following information:
• Reference designator• An indication if the part is subject to damage by static discharge• Description• Fluke stock number• Total quantity• Any special notes (i.e., factory-selected part)
CautionA * symbol indicates a device that may be damaged by staticdischarge.
How to Obtain PartsElectrical components may be ordered directly from the manufacturer by using themanufacturers part number, or from the Fluke Corporation and its authorizedrepresentatives by using the part number under the heading FLUKE STOCK NO. In theU.S., order directly from the Fluke Parts Dept. by calling 1-800-526-4731. Parts priceinformation is available from the Fluke Corporation or its representatives. Prices are alsoavailable in a Fluke Replacement Parts Catalog which is available on request.
In the event that the part ordered has been replaced by a new or improved part, thereplacement will be accompanied by an explanatory note and installation instructions, ifnecessary.
riešenia na presné meranie
http://www.elso.sk
12B/18/7 SeriesService Manual
4-2
To ensure prompt delivery of the correct part, include the following information whenyou place an order:
• Instrument model and serial number• Part number and revision level of the pca containing the part.• Reference designator• Fluke stock number• Description (as given under the DESCRIPTION heading)• Quantity
Manual Status InformationThe Manual Status Information in Table 4-1 defines the assembly revision levels that aredocumented in the manual. Revision levels are printed on each pca.
Table 4-1. Manual Status Information
Ref. or Option No. Assembly Name Fluke Part No. Revision Level
A1 PCB ASSEMBLY (FLUKE 12) 879002 M
A1 PCB ASSEMBLY (FLUKE 10/18) 878991 M
A1 PCB ASSEMBLY (FLUKE 7-300) 614963 M
A1 PCB ASSEMBLY (FLUKE 7-600) 614966 M
Newer InstrumentsChanges and improvements made to the instrument are identified by incrementing therevision letter marked on the affected pca. These changes are documented on a manualsupplement sheet which, when applicable, is included with the manual.
Parts Lists, Drawings, and SchematicsThis section contains the parts lists, drawings, and schematics for the meters.
Note
This instrument may contain a Nickel-Cadmium battery. Do not mix withthe solid waste stream. Spent batteries should be disposed of by a qualifiedrecycler or hazardous materials handler. Contact your authorized Flukeservice center for recycling information.
P WarningTo avoid fire or explosion, replace the fusible resistor only withFluke PN 867361.
riešenia na presné meranie
http://www.elso.sk
Parts and SchematicsParts Lists, Drawings, and Schematics 4
4-3
Table 4-2. Models 12B, 18, and 18/AL Final Assembly
ReferenceDesignator Description
Fluke StockNumber
TotalQuantity
A1 MAIN PCA (MODEL 12B) 879002 1
A1 MAIN PCA (MODEL 18) 878991 1
BT1 BATTERY, 9V, 0-15MA 696534 1
H1-4 SCREW, PH, P, AM THD FORM, STL, 5-14, .750 832246 4
J3 CONN, ELASTOMERIC, LCD TO PWB, 2.050 L 867247 1
J4 CONN, ELASTOMERIC, SWITCH TO PWB, .500 L 867242 1
MP2 CASE TOP, (PAD XFER & WINDOW, MODEL 12B) 614948 1
MP2 CASE TOP, (PAD XFER & WINDOW, MODEL 18) 614955 1
MP2 CASE TOP, (PAD XFER & WINDOW, MODEL 18/AL) 615192 1
MP5 ACTUATOR, SWITCH 203445 1
MP6 SUPPORT, SWITCH ASSEMBLY 879031 1
MP8 CASE, BOTTOM, YELLOW (MODEL 12B) 614864 1
MP8 CASE, BOTTOM, GRAY (MODEL 18) 614872 1
MP9 SHIELD, BOTTOM 878277 1
MP10 SHOCK ABSORBER 878983 1
MP11 FOOT, NON-SKID 885884 1
MP12 TEST LEADS 855742 1
S2 SWITCH ASSEMBLY (MODEL 12B) 614799 1
S2 SWITCH ASSEMBLY (MODEL 18) 614807 1
U3 LCD, 3.5 DIGIT, MULTIPLEXED 855226 1
TM1 12B/18 INSTRUCTION SHEET 602000 1
riešenia na presné meranie
http://www.elso.sk
12B/18/7 SeriesService Manual
4-4
FLUKE 12B, FLUKE 18, FLUKE 18/ALT&B
MP2
S2
MP5
A1
MP9
MP11
U3
J4
J3
MP6
BT1
MP10
MP8
H1-4
as06f.eps
Figure 4-1. Models 12B and 18 Final Assembly
riešenia na presné meranie
http://www.elso.sk
Parts and SchematicsParts Lists, Drawings, and Schematics 4
4-5
Table 4-3. Models 7-300 and 7-600 Final Assembly
ReferenceDesignator Description
Fluke StockNumber
TotalQuantity
A1 MAIN PCA (MODEL 7-600) 614966 1
A1 MAIN PCA (MODEL 7-300) 614963 1
BT1 BATTERY, 9V, 0-15MA 696534 1
H1-4 SCREW, PH, P, AM THD FORM, STL, 5-14, .750 832246 4
J3 CONN, ELASTOMERIC, LCD TO PWB, 2.050 L 867247 1
MP2 CASE TOP, (PAD XFER & WINDOW, MODEL 7-600) 614930 1
MP2 CASE TOP, (PAD XFER & WINDOW, MODEL 7-300) 614922 1
MP5 ACTUATOR, SWITCH 203452 1
MP6 SUPPORT, SWITCH ASSEMBLY 879031 1
MP8 CASE, BOTTOM, YELLOW 614864 1
MP9 SHIELD, BOTTOM 878277 1
MP10 SHOCK ABSORBER 878983 1
MP11 FOOT, NON-SKID 885884 1
MP12 TEST LEADS 855742 1
U3 LCD, 3.5 DIGIT, MULTIPLEXED 855226 1
TM1 INSTRUCTION SHEET, FLUKE 7 602018 1
riešenia na presné meranie
http://www.elso.sk
12B/18/7 SeriesService Manual
4-6
FLUKE 7-300, FLUKE 7-600T&B
MP2
MP5
A1
MP9
MP11
U3
J3
MP6
MP10
BT1
MP8
H1-4
as09f.eps
Figure 4-2. Models 7-300 and 7-600 Final Assembly
riešenia na presné meranie
http://www.elso.sk
Parts and SchematicsParts Lists, Drawings, and Schematics 4
4-7
Table 4-4. A1 Main PCA
ReferenceFlukeStock Total Quantity
Designator Description Number 7-300 7-600 12B 18
C1 CAP, TA, 0.47 µF, ±20%, 35V 655035 1 1 1 1
C2, C5-7 CAP, POLYES, 0.022 µF, ±10%, 50V 715268 0 0 4 4
C2, C7 CAP, POLYES, 0.022 µF, ±10%, 50V 715268 2 2 0 0
C3 CAP, TA, 22 µF, ±20%, 6V, 6032 876545 1 1 1 1
C4, C13 CAP, CER, 0.1 µF, ±10%, 25V, X7R, 1206 747287 0 2 2 0
C4, C12, C13 CAP, CER, 0.1 µF, ±10%, 25V, X7R, 1206 747287 3 0 0 0
C4, C11, C13 CAP, CER, 0.1 µF, ±10%, 25V, X7R, 1206 747287 0 0 0 3
C5, C6 CAP, CER, 4700 PF, ±20%, 100V 743427 2 2 0 0
C8 CAP, TA, 3.3 µF, ±20%, 16V, 3528 876552 1 1 1 1
C9 CAP, CER, 0.022 µF, ±10%, 50V, X7R, 1206 747279 1 1 1 1
C10 CAP, AL, 22 µF, ±20%, 10V, SOLV PROOF 887245 1 1 1 1
C14 CAP,POLY,.01 µF, ±10%,.1000V 822361 1 1 1 1
C15 CAP, TA, 47 µF, ±20%, 10V, 7343 867580 1 1 1 1
C16 CAP, CER, 10 PF, ±20%, 50V, C0G 721589 1 1 1 1
CR1 DIODE, SI, 100 PIV, 1 AMP, SURFACE MOUNT 912451 1 1 1 1
CR2 DIODE,SI,DUAL, BV = 50V, IO = 100 mA, SOT-23 851659 1 1 1 1
CR3 DIODE,SI, BV = 70V, IO = 50 mA, DUAL, SOT-23 742320 1 1 1 1
E1 SURGE PROTECTOR,3750v,±20% 867361 1 1 1 1
J1, J2 RECEPTACLE, INPUT 878988 2 2 2 2
LS1 AF TRANSD, PIEZO, 20MM 876995 1 1 1 1
Q1 TRANSISTOR, SI, NPN, SELECT IEBO, SOT-23 821637 1 1 1 1
Q2 TRANSISTOR, SI, PNP, SELECT ICER, SOT-23 887179 1 1 1 1
R1 RES, CERM, 162K, ±1%, .125W, 100 PPM, 1206 876198 1 1 1 1
R2 RES, MF, 332K, ±1%, 0.125W, 100 PPM 655217 1 1 1 1
R3 RES, MF, 23.4K, ±0.1%, 0.125W, 100 PPM 876292 1 1 1 1
R4 RES,VAR,CERM,25K, ±25%,.2W 876300 1 1 1 1
R5 RES, CERM, 43.2K, ±1%, .125W, 100 PPM, 1206 887109 1 1 1 1
R6 RES, CERM, 226K, ±1%, .125W, 100 PPM, 1206 876524 1 1 1 1
R7-9, R23, R28 RES, CERM, 1M, ±1%, .125W, 100 PPM, 1206 836387 5 5 5 5
R10 RES, CERM, 309K, ±1%, .125W, 100 PPM, 1206 876201 1 1 1 1
R11, R12 RES, MF, 30.1K, ±0.1%, 0.125W, 100 PPM 887161 2 2 2 2
R13 RES, MF, 27.1K, ±0.1%, 0.125W, 100 PPM 876289 1 1 1 1
R14, R15 RES, CERM, 10K, ±5%, .125W, 200 PPM, 1206 746610 2 2 2 2
R16, R18 RES,CERM, 470, ±5%,3W 887117 2 2 2 2
R17 RES,CERM, 1M, ±5%,1W 912589 1 1 1 1
R19 RES, MF, 450, ±10%, 107398 1 1 1 1
R21 RES JUMPER, 0.02, 0.25W 682575 1 1 1 1
R22, R29, R30 RES, CERM, 300, ±5%, .125W, 200 PPM, 1206 746362 3 3 3 3
R24 RES, CERM, 487K, ±1%, .125W, 100 PPM, 1206 887112 1 1 1 1
R25 RES, CERM, 2.2M, ±5%, .125W, 200 PPM, 1206 811778 1 1 1 1
riešenia na presné meranie
http://www.elso.sk
12B/18/7 SeriesService Manual
4-8
Table 4-4. A1 Main PCA (cont)
ReferenceFlukeStock Total Quantity
Designator Description Number 7-300 7-600 12B 18
R26 RES, CERM, 100K, ±1%, .125W, 100 PPM, 1206 769802 1 1 1 1
R27 RES, CERM, 0, +.05 MAX, .125W, 1206 810747 1 1 1 1
RT1 THERM, POS 1.1K, ±20%, 25C 867192 1 1 1 1
S1 SWITCH, SLIDE, 3 POS, 5 PIN 868117 1 1 1 1
U1 ARTIC, ASSEMBLY TESTED 858472 1 1 1 1
U2 IC, CMOS, 4 BIT MPU, LCD 601575 1 1 0 0
U2 *IC, CMOS, 4 BIT MPU, LCD CNTRLR, 80PNQFP 866913 0 0 1 1
VR1 IC, 1.23V, 150 PPM T.C., BANDGAP V. REF 634451 1 1 1 1
XBT, XBT2 CONTACT, BATTERY 890327 2 2 2 2
Y1 RESONATOR, CERAMIC, PIEZOELEC, 2.10
MHZ
876024 1 1 1 1
Z1 RES NET THK FILLM TESTED 872234 1 1 1 1
riešenia na presné meranie
http://www.elso.sk
Parts and SchematicsParts Lists, Drawings, and Schematics 4
4-9
FLUKE 12-1601
as11f.eps
Figure 4-3. Models 12B, 18, 7-300, and 7-600 Main PCA Components
riešenia na presné meranie
http://www.elso.sk
12B/18/7 SeriesService Manual
4-10
R2
81
M
R2
6
10
0K
R5
43
.2
K
R6
22
6K
R7
1M
R8
1M
R9
1M
R1
3
27
.1
K
R1
7
1M
R1
8
47
0
R1
6
47
0
LS
1
APV4
APV3
APV2
APV1
DIVLO
APV0
ACV
VSS
OHMS
SWS
ISRC
AMPS
CLAMP
AGND
DGND
READ*
ADR5
ADR4
ADR3
ADR2
ADR1
ADR0
VSS
VDD
K0
K1
K2
AZ
INT
REFI
DG
ND
DA
TA
3
DA
TA
2
DA
TA
1
DA
TA
0
N/
C
BE
EP
ER
*
VS
S
VS
S
CL
K
BC
LK
TE
ST
CL
K
BE
EP
ER
N/
C
WR
IT
E*
PS
TE
ST
VS
ET
AC
BO
AV
AM
AV
AO
P
AV
AO
M
PF
O
VS
S
AFI
FAI
FA
O
AF
O
BG
ND
RE
FH
BI
AS
U1
33
2K
R2
VR
1
R1
1
30
.1
KE
1
J1
3.
3
C8
CR
2B
AV
74
10
.0
M
Z1
1.00K
Z1
10.01K
Z1
101K
Z1
1.111M
Z1
10
.0
M
Z1
.4
7C
12
3.
4K
R3
R1
16
2K
.0
22
C2
10
KR
15
22
C3
J2
10
pF
C1
6
CR
11
N4
00
2
R2
7
0
47C
15
BOTTOM SHIELD
TOP SHIELD
S1
BT
1
9V
R3
03
00
TP
9
Q2
.0
22
C5
.0
22
C7
C1
4
.01
.0
22
C6
C9
.0
22
TP
5
TP
4
TP
6
TP
8
TP
7
CW
R4
25
K
Q1
20
%1.
1KRT
1
t+
R2
2
30
0
R1
41
0K
R1
03
09
KR
24
48
7K
R1
23
0.
1K
47
0
R1
9
0
R2
1
R2
93
00
20
55
55
CO
MM
ON
M5
0V
25
%
1%
1%
1%
1%
1%
1%
1%
MF
0.
1%
44
52
1
60
28
39
76
54
32
22
19
32
34
40
42
43
45
27
25
26
13
56
58
18
16
33
55
50
49
48
47
31
46
41
21
20
14
35
10
11
12
51
59
37
24
30
17
15
36
57
9
29
8
23
38
53
54
1%
MF
1.
23
5V
0.
1%
MF
20
%3
75
0V
20
%1
6V
T
23
48
476 4
45
21
20
%3
5V
T
0.
1%
MF
1%
50
VM
T20
%6
V
VOLT/OHMS
50
V
32
54
1
50
V M
M1K
V
10
%5
0V
1%
1%
MF
0.
1%
VD
D
SW
S
VS
S1
BC
LK
VB
T+
DA
TA
0
DA
TA
1
DA
TA
2
DA
TA
3
VS
S2
VS
S2
DG
ND
DG
ND
VS
S2
VD
D
VD
D
RE
AD
*
AD
R5
AD
R4
AD
R3
AD
R2
AD
R1
AD
R0
CL
K
DG
ND
DG
ND
OH
MS
IS
RC
VS
S1
RE
FH
AF
O FA
O
RE
FI
AV
AO
M
AG
ND
BG
ND
AP
V0
FAI
VB
T+
AC
V
AG
ND
VD
D
WR
IT
E*
DG
ND
DG
ND
10
V2
0%
T
1W
50
VM
3W
3W
5%
MG
MG5
%3
W MG5
%
FLU
KE
12-
1201
(1 o
f 2)
as20f.eps
Figure 4-4. Model 12B Schematic
riešenia na presné meranie
http://www.elso.sk
Parts and SchematicsParts Lists, Drawings, and Schematics 4
4-11
TP
2T
P1
TP
32
.1
MH
Z
Y1
U2
P70P63P62P61P60X2X1VPPXT2XT1VDDP33P32P31P30P23P22P21P20P13P12P11P10P03
SO
SC
KP
00
P5
3P
52
P5
1P
50
VS
SP
43
P4
2P
41
P4
0V
LC
2V
LC
1V
LC
0B
IA
S
S12S13S14S15S16S17S18S19S20S21S22S23S24S25S26S27S28S29S30S31
COM0COM1COM2COM3
P7
1P
72
P7
3R
ES
ET
*S
0S
1S
2S
3S
4S
5S
6S
7S
8S
9S
10
S1
1
12
34
56
78
91
01
11
21
31
41
51
61
71
81
92
02
12
22
32
42
52
62
72
8
U3
CR
3B
AV
99
R2
3
1.
0M
2.
2M
R2
5
C1
3
0.1
22
C1
0
C4
0.1
S2
S2
S2
S2
MIN
/MA
X
CA
PA
CIT
AN
CE
RA
NG
E
SE
LE
CT
AZ
EN
D
SC
LK
SO
UT
1%
DG
ND
AD
R0
AD
R1
AD
R2
AD
R3
AD
R4
AD
R5
RE
AD
*W
RI
TE
*
VS
S1
BC
LK
CL
K
DA
TA
3
DA
TA
2
DA
TA
1
DA
TA
0
VD
D
VS
S2
25
V
10
%
25
V1
0%
.5
%
67
68
25
38
29
30
31
32
34
35
36
37
65
66
69
70
79
80
71
72
73
74
75
76
77
78
39
40
26
27
28
33
47
41
43
2
5655
595857
54
20191817161514131211109876543
1 6463626160
535251504948
464544
4224232221
12
72
28
34
56
78
91
01
11
21
31
41
51
61
71
81
92
02
12
22
32
42
52
6
AL
10
V2
0%
51
52
53
54
FLU
KE
12-
1201
(2 o
f 2)
AL
UM
IN
UM
AL
CA
RB
ON
F
IL
MC
FC
AR
BO
N
CO
MP
CC
CO
MP
ON
EN
T
TY
PE
A
BB
RE
VIA
TIO
N
CO
DE
S
SY
MC
AP
AC
ITO
RS
YM
RESI
STOR
PO
LY
PR
OP
YL
EN
E
FI
LM
M C T PC
PP
MY
LA
R/
PO
LY
ES
TE
RC
ER
AM
IC
TA
NT
AL
UM
PO
LY
C
AR
BO
NA
TE
MG
WW
MF
ME
TA
L
GL
AZ
EW
IRE
W
OU
ND
META
L FI
LM
5
NOTE
S:UN
LESS
OTH
ERW
ISE
SPEC
IFIE
D:
1. 2.A
LL
C
AP
AC
ITA
NC
E
VA
LU
ES
A
RE
IN
M
ICR
OF
AR
AD
S,
+/
-
10
%,
C
ER
AM
IC.
3.AL
L RE
SIST
ANCE
VAL
UES
ARE
IN O
HMS,
1/8
WAT
T,+
/-
5%
,
CE
RM
ET
.
+/
-1
%
RE
SIS
TO
RS
A
RE
1
/8
WA
TT
,C
ER
ME
T.
4.
R19
IS A
FUS
IBLE
RES
ISTO
R. E
XACT
RE
PL
AC
EM
EN
T
ON
LY
.
6.R
11
A
ND
R
12
M
AY
H
AV
E
+/
-0
.2
5%
TO
LE
RA
NC
E.
7.R
23
M
AY
H
AV
E
+/
-5
%
TO
LE
RA
NC
E.
8.C
4,
9
,
11
&
1
3
MA
Y
HA
VE
+
/-
20
%
TOL
ER
AN
CE
.
SE
LE
CT
PIN
A
SS
IGN
ME
NT
SM
EM
BR
AN
E
TA
IL
S2
1 2 6543
VS
S
VS
S
RA
NG
E
as21f.eps
Figure 4-4. Model 12B Schematic (cont)
riešenia na presné meranie
http://www.elso.sk
12B/18/7 SeriesService Manual
4-12
R2
81
M
R2
6
10
0K
R5
43
.2
K
R6
22
6K
R7
1M
R8
1M
R9
1M
R1
3
27
.1
K
R1
7
1M
R1
8
47
0
R1
6
47
0
LS
1
APV4
APV3
APV2
APV1
DIVLO
APV0
ACV
VSS
OHMS
SWS
ISRC
AMPS
CLAMP
AGND
DGND
READ*
ADR5
ADR4
ADR3
ADR2
ADR1
ADR0
VSS
VDD
K0
K1
K2
AZ
INT
REFI
DG
ND
DA
TA
3
DA
TA
2
DA
TA
1
DA
TA
0
N/
C
BE
EP
ER
*
VS
S
VS
S
CL
K
BC
LK
TE
ST
CL
K
BE
EP
ER
N/
C
WR
IT
E*
PS
TE
ST
VS
ET
AC
BO
AV
AM
AV
AO
P
AV
AO
M
PF
O
VS
S
AFI
FAI
FA
O
AF
O
BG
ND
RE
FH
BI
AS
U
1
33
2K
R2
VR
1
R1
1
30
.1
KE
1
J1
3.
3
C8
CR
2B
AV
74
10
.0
M
Z1
1.00K
Z1
10.01K
Z1
101K
Z1
1.111M
Z1
10
.0
M
Z1
.4
7C
12
3.
4K
R3
R1
16
2K
.0
22
C2
10
KR
15
22
C3
J2
10
pF
C1
6
CR
11
N4
00
2
R2
7
0
47C
15
BOTTOM SHIELD
TOP SHIELD
S1
BT
1
9V
R3
03
00
TP
9
Q2
.0
22
C5
.0
22
C7
C1
4
.01
.0
22
C6
C9
.0
22
TP
5
TP
4
TP
6
TP
8
TP
7
CW
R4
25
K
Q1
20
%1
.1
K
RT
1
t +
R2
2
30
0
R1
41
0K
R1
03
09
KR
24
48
7K
R1
23
0.
1K
50
0
R1
9
R2
1
R2
93
00
20
5
CO
MM
ON
25
%
M5
0V
T
1W
M
T2
0%
10
V
DG
ND
DG
ND
WR
IT
E*
VD
D
AG
ND
AC
V
VB
T+
FAI
AP
V0
BG
ND
AG
ND
AV
AO
M
RE
FI
FA
O
AF
O
RE
FH
VS
S1
IS
RC
OH
MS
DG
ND
DG
ND
CL
K
AD
R0
AD
R1
AD
R2
AD
R3
AD
R4
AD
R5
RE
AD
*
VD
D
VD
D
VS
S2
DG
ND
DG
ND
VS
S2
VS
S2
DA
TA
3
DA
TA
2
DA
TA
1
DA
TA
0
VB
T+
BC
LK
VS
S1
SW
S
VD
D
0.
1%
MF
1%
1%
50
V1
0%
1K
V
M5
0V
1
45
23
50
V
VOLT/OHMS
6V
20
%T
M50
V
1%
MF
0.
1%
T
35
V2
0%
12
5 446
7 4
8 4
32
16
V2
0%
37
50
V2
0%
MF
0.
1%
1.
23
5V
MF
1%
54
53
38
23
8
29
9
57
36
15
17
30
24
37
59
51
12
11
10
35
14
20
21
41
46
31
47
48
49
50
55
33
16
18
58
56
13
26
25
27
45
43
42
40
34
32
19
22
23
45
67 3
9
28
60
1
52
44
0.
1%
MF
1%
1%
1%
1%
1%
1%
1%
50
V M
0
MG
5%
3W
1W
MF
+0
%-
20
%M
G5
%3
W
FLU
KE
10-
1201
(1 o
f 2)
as16f.eps
Figure 4-5. Model 18 Schematic
riešenia na presné meranie
http://www.elso.sk
Parts and SchematicsParts Lists, Drawings, and Schematics 4
4-13
TP
2T
P1
TP
32
.1
MH
Z
Y1
U2
P70P63P62P61P60X2X1VPPXT2XT1VDDP33P32P31P30P23P22P21P20P13P12P11P10P03
SO
SC
KP
00
P5
3P
52
P5
1P
50
VS
SP
43
P4
2P
41
P4
0V
LC
2V
LC
1V
LC
0B
IA
S
S12S13S14S15S16S17S18S19S20S21S22S23S24S25S26S27S28S29S30S31
COM0COM1COM2COM3
P7
1P
72
P7
3R
ES
ET
*S
0S
1S
2S
3S
4S
5S
6S
7S
8S
9S
10
S1
1
12
34
56
78
91
01
11
21
31
41
51
61
71
81
92
02
12
22
32
42
52
62
72
8
U3
CR
3B
AV
99
R2
3
1.
0M
0.1
C1
1
2.
2M
R2
5
C1
3
0.1
22
C1
0
C4
0.1
S2
S2
SE
LE
CT
RA
NG
E
AZ
EN
D
SC
LK
SO
UT
20
%1
0V
AL
26
25
24
23
22
21
20
19
18
17
16
15
14
13
12
11
10
98
76
54
32
82
27
1
21222324
42
444546
484950515253
60616263641
34567891011121314151617181920
54
575859
5556
2
43
41
47
33
28
27
26
40
39
78
77
76
75
74
73
72
71
80
79
70
69
66
65
37
36
35
34
32
31
30
29
38
25
68
67
.5
%
10
%
25
V
VD
D
DA
TA
0
DA
TA
1
DA
TA
2
DA
TA
3
CL
K
BC
LK
VS
S1
WR
IT
E*
RE
AD
*A
DR
5A
DR
4
AD
R3
AD
R2
AD
R1
AD
R0
DG
ND
25
V2
0%
25
V1
0%
1%
25
VS
S21
5
FLU
KE
10-
1201
(2 o
f 2)
AL
UM
IN
UM
AL
CA
RB
ON
F
IL
MC
FC
AR
BO
N
CO
MP
CC
CO
MP
ON
EN
T
TY
PE
A
BB
RE
VIA
TIO
N
CO
DE
S
SY
MC
AP
AC
ITO
RS
YM
RESI
STOR
PO
LY
PR
OP
YL
EN
E
FI
LM
M C T PC
PP
MY
LA
R/
PO
LY
ES
TE
RC
ER
AM
IC
TA
NT
AL
UM
PO
LY
C
AR
BO
NA
TE
MG
WW
MF
ME
TA
L
GL
AZ
EW
IRE
W
OU
ND
META
L FI
LM
SE
LE
CT
PIN
ASSI
GNME
NTS
ME
MB
RA
NE
TA
ILS
2
CL
OS
ED
OP
EN
OP
EN
CL
OS
ED
OP
EN
OP
EN
CL
OS
ED
CL
OS
ED
OP
EN
LO
W
Z
HIG
H
Z
OF
F
4-
52
-3
1-
2
CO
NT
AC
TS
POSI
TIO
NS1 C
ONN
ECTI
VITY
TAB
LE
5
RA
NG
E
VS
SV
SS
62 3 4 51
NOTE
S:UN
LESS
OTH
ERW
ISE
SPEC
IFIE
D:
1. 2.A
LL
C
AP
AC
ITA
NC
E
VA
LU
ES
A
RE
IN
M
ICR
OF
AR
AD
S,
+/
-
10
%,
C
ER
AM
IC.
3.AL
L RE
SIST
ANCE
VAL
UES
ARE
IN O
HMS,
1/8
WAT
T,+
/-
5%
,
CE
RM
ET
.
+/
-1
%
RE
SIS
TO
RS
A
RE
1
/8
WA
TT
,C
ER
ME
T.
4.
R19
IS A
FUS
IBLE
RES
ISTO
R. E
XACT
RE
PL
AC
EM
EN
T
ON
LY
.
6.R
11
A
ND
R
12
M
AY
H
AV
E
+/
-0
.2
5%
TO
LE
RA
NC
E.
7.R
23
M
AY
H
AV
E
+/
-5
%
TO
LE
RA
NC
E.
8.C
4,
9
,
11
&
1
3
MA
Y
HA
VE
+
/-
20
%
TOL
ER
AN
CE
.
as17f.eps
Figure 4-5. Model 18 Schematic (cont)
riešenia na presné meranie
http://www.elso.sk
12B/18/7 SeriesService Manual
4-14
R2
81
M
R2
6
10
0K
R5
43
.2
K
R6
22
6K
R7
1M
R8
1M
R9
1M
R1
3
27
.1
K
R1
7
1M
R1
8
47
0
R1
6
47
0
LS
1
APV4
APV3
APV2
APV1
DIVLO
APV0
ACV
VSS
OHMS
SWS
ISRC
AMPS
CLAMP
AGND
DGND
READ*
ADR5
ADR4
ADR3
ADR2
ADR1
ADR0
VSS
VDD
K0
K1
K2
AZ
INT
REFI
DG
ND
DA
TA
3
DA
TA
2
DA
TA
1
DA
TA
0
N/
C
BE
EP
ER
*
VS
S
VS
S
CL
K
BC
LK
TE
ST
CL
K
BE
EP
ER
N/
C
WR
IT
E*
PS
TE
ST
VS
ET
AC
BO
AV
AM
AV
AO
P
AV
AO
M
PF
O
VS
S
AFI
FAI
FA
O
AF
O
BG
ND
RE
FH
BI
AS
U
1
33
2K
R2
VR
1
R1
1
30
.1
KE
1
J1
3.
3
C8
CR
2B
AV
74
10
.0
M
Z1
1.00K
Z1
10.01K
Z1
101K
Z1
1.111M
Z1
10
.0
M
Z1
.4
7C
12
3.
4K
R3
R1
16
2K
.0
22
C2
10
KR
15
22
C3
J2
10
pF
C1
6
CR
11
N4
00
2
R2
7
0
47C
15
BOTTOM SHIELD
TOP SHIELD
S1
BT
1
9V
R3
03
00
TP
9
Q2
.0
22
C7
C1
4
.01
C9
.0
22
TP
5
TP
4
TP
6
TP
8
TP
7
CW
R4
25
K Q1
20
%1.
1KRT
1
t+
R2
2
30
0
R1
41
0K
R1
03
09
KR
24
48
7K
R1
23
0.
1K
50
0
R1
9
0
R2
1
R2
93
00
5
20
CO
MM
ON
MG
5%
3W
T2
0%
10
V
1%
1%
1%
1%
1%
1%
1%
MF
0.
1%
44
52
1
60
28
39
76
54
32
22
19
32
34
40
42
43
45
27
25
26
13
56
58
18
16
33
55
50
49
48
47
31
46
41
21
20
14
35
10
11
12
51
59
37
24
30
17
15
36
57
9
29
8
23
38
53
54
1%
MF
1.
23
5V
0.
1%
MF
20
%3
75
0V
16
V
23
48
476 4
45
21
20
%3
5V
T
0.
1%
MF
1%
50
VM
T20
%6
V
VOLT/OHMS
50
V
32
54
1
M1K
V
10
%5
0V
1%
1%
MF
0.
1%
VD
D
SW
S
VS
S1
BC
LK
VB
T+
DA
TA
0
DA
TA
1
DA
TA
2
DA
TA
3
VS
S2
VS
S2
DG
ND
DG
ND
VS
S2
VD
D
VD
D
RE
AD
*
AD
R5
AD
R4
AD
R3
AD
R2
AD
R1
AD
R0
CL
K
DG
ND
DG
ND
OH
MS
IS
RC
VS
S1
RE
FH
AF
O FA
O
RE
FI
AV
AO
M
AG
ND
BG
ND
AP
V0
FAI
VB
T+
AC
V
AG
ND
VD
D
WR
IT
E*
DG
ND
DG
ND
25
%
MG
5%
3W
1W
MF
+0
%-
20
% 1W
20
%T
50
VM
FLU
KE
7-3
00-1
201
(1 o
f 2)
47
00
PF
C6 0
2%
01
0V
47
00
PF
C5
02
%0
10
V
as12f.eps
Figure 4-6. Model 7-300 Schematic
riešenia na presné meranie
http://www.elso.sk
Parts and SchematicsParts Lists, Drawings, and Schematics 4
4-15
TP
2T
P1
TP
32
.1
MH
Z
Y1
U2
P70P63P62P61P60X2X1VPPXT2XT1VDDP33P32P31P30P23P22P21P20P13P12P11P10P03
SO
SC
KP
00
P5
3P
52
P5
1P
50
VS
SP
43
P4
2P
41
P4
0V
LC
2V
LC
1V
LC
0B
IA
S
S12S13S14S15S16S17S18S19S20S21S22S23S24S25S26S27S28S29S30S31
COM0COM1COM2COM3
P7
1P
72
P7
3R
ES
ET
*S
0S
1S
2S
3S
4S
5S
6S
7S
8S
9S
10
S1
1
12
34
56
78
91
01
11
21
31
41
51
61
71
81
92
02
12
22
32
42
52
62
72
8
U3
CR
3B
AV
99
R2
3
1.
0M
2.
2M
R2
5
C1
3
0.1
0.1
C1
2
22
C1
0
C4
0.1
S2
S2
S2
SE
LE
CT
AZ
EN
D
SC
LK
SO
UT
CA
PA
CI
TA
NC
E
RA
NG
E
DG
ND
AD
R0
AD
R1
AD
R2
AD
R3
AD
R4
AD
R5
RE
AD
*W
RI
TE
*
VS
S1
BC
LK
CL
K
DA
TA
3
DA
TA
2
DA
TA
1
DA
TA
0
VD
D
VS
S2
25
V
10
%
25
V1
0%
.5
%
67
68
25
38
29
30
31
32
34
35
36
37
65
66
69
70
79
80
71
72
73
74
75
76
77
78
39
40
26
27
28
33
47
41
43
2
5655
595857
54
20191817161514131211109876543
1 6463626160
535251504948
464544
4224232221
12
72
28
34
56
78
91
01
11
21
31
41
51
61
71
81
92
02
12
22
32
42
52
6
AL
10
V2
0%
51
25
V1
0%
1%
53
52
FLU
KE
7-3
00-1
201
(2 o
f 2)
AL
UM
IN
UM
AL
CA
RB
ON
F
IL
MC
FC
AR
BO
N
CO
MP
CC
CO
MP
ON
EN
T
TY
PE
A
BB
RE
VI
AT
IO
N
CO
DE
S
SY
MC
AP
AC
IT
OR
SY
MR
ES
IS
TO
R
PO
LY
PR
OP
YL
EN
E
FI
LM
M C T PC
PP
MY
LA
R/
PO
LY
ES
TE
RC
ER
AM
IC
TA
NT
AL
UM
PO
LY
C
AR
BO
NA
TE
MG
WW
MF
ME
TA
L
GL
AZ
EW
IR
E
WO
UN
DM
ET
AL
F
IL
M
J CR
ELS
C
RE
FE
RE
NC
E
DE
SIG
NA
TIO
N
LA
ST
US
ED
NO
T
US
ED
QCR UBT
S TP
Y ZVR
RT
SE
LE
CT
PI
N
AS
SIG
NM
EN
TS
ME
MB
RA
NE
T
AI
L
S2
CL
OS
ED
OP
EN
OP
EN
CL
OS
ED
OP
EN
OP
EN
CL
OS
ED
CL
OS
ED
OP
EN
LO
W
Z
HI
GH
Z
OF
F
4-
52
-3
1-
2
CO
NT
AC
TS
PO
SI
TI
ONS
1
CO
NN
EC
TIV
ITY
TA
BL
E
5
20
11
11 11 3 22 2 3 1116 1
NO
TE
S:
UN
LE
SS
O
TH
ER
WI
SE
S
PE
CI
FI
ED
:
1. 2.A
LL
C
AP
AC
IT
AN
CE
V
AL
UE
S
AR
E
IN
M
IC
RO
FA
RA
DS
,+
/-
1
0%
,
CE
RA
MI
C.
3.A
LL
R
ES
IS
TA
NC
E
VA
LU
ES
A
RE
I
N
OH
MS
,
1/
8
WA
TT
,+
/-
5%
,
CE
RM
ET
.
+/
-1
%
RE
SI
ST
OR
S
AR
E
1/
8
WA
TT
,C
ER
ME
T.
4.
R1
9
IS
A
F
US
IB
LE
R
ES
IS
TO
R.
E
XA
CT
RE
PL
AC
EM
EN
T
ON
LY
.
6.R
11
A
ND
R
12
M
AY
H
AV
E
+/
-0
.2
5%
T
OL
ER
AN
CE
.
7.R
23
M
AY
H
AV
E
+/
-5
%
TO
LE
RA
NC
E.
8.
30 8
1 2 6543
VS
S
VS
S
CA
PA
CI
TA
NC
E
RA
NG
E
C4
,
9,
1
2
&
13
M
AY
H
AV
E
+/
-2
0%
T
OL
ER
AN
CE
.
as13f.eps
Figure 4-6. Model 7-300 Schematic (cont)
riešenia na presné meranie
http://www.elso.sk
12B/18/7 SeriesService Manual
4-16
R2
81
M
R2
6
10
0K
R5
43
.2
K
R6
22
6K
R7
1M
R8
1M
R9
1M
R1
3
27
.1
K
R1
7
1M
R1
8
47
0
R1
6
47
0
LS
1
APV4
APV3
APV2
APV1
DIVLO
APV0
ACV
VSS
OHMS
SWS
ISRC
AMPS
CLAMP
AGND
DGND
READ*
ADR5
ADR4
ADR3
ADR2
ADR1
ADR0
VSS
VDD
K0
K1
K2
AZ
INT
REFI
DG
ND
DA
TA
3
DA
TA
2
DA
TA
1
DA
TA
0
N/
C
BE
EP
ER
*
VS
S
VS
S
CL
K
BC
LK
TE
ST
CL
K
BE
EP
ER
N/
C
WR
IT
E*
PS
TE
ST
VS
ET
AC
BO
AV
AM
AV
AO
P
AV
AO
M
PF
O
VS
S
AFI
FAI
FA
O
AF
O
BG
ND
RE
FH
BI
AS
U
1
33
2K
R2
VR
1
R1
1
30
.1
KE
1
J1
3.
3
C8
CR
2B
AV
74
10
.0
M
Z1
1.00K
Z1
10.01K
Z1
101K
Z1
1.111M
Z1
10
.0
M
Z1
.4
7C
12
3.
4K
R3
R1
16
2K
.0
22
C2
10
KR
15
22
C3
J2
10
pF
C1
6
CR
11
N4
00
2
R2
7
0
47C
15
BOTTOM SHIELD
TOP SHIELD
S1
BT
1
9V
R3
03
00
TP
9
Q2
47
00
PF
C6
.0
22
C7
C1
4
.01
C9
.0
22
TP
5
TP
4
TP
6
TP
8
TP
7
CW
R4
25
K
Q1
20
%1
.1
K
RT
1
t +
R2
2
30
0
R1
41
0K
R1
03
09
KR
24
48
7K
R1
23
0.
1K
50
0
R1
9
0
R2
1
R2
93
00
5
CO
MM
ON
02
%0
10
V4
70
0P
FC
5
02
%0
10
V
M5
0V
MG
5%
3W
1W
T2
0%
10
V
DG
ND
DG
ND
WR
IT
E*
VD
D
AG
ND
AC
V
VB
T+
FAI
AP
V0
BG
ND
AG
ND
AV
AO
M
RE
FI
FA
O
AF
O
RE
FH
VS
S1
IS
RC
OH
MS
DG
ND
DG
ND
CL
K
AD
R0
AD
R1
AD
R2
AD
R3
AD
R4
AD
R5
RE
AD
*
VD
D
VD
D
VS
S2
DG
ND
DG
ND
VS
S2
VS
S2
DA
TA
3
DA
TA
2
DA
TA
1
DA
TA
0
VB
T+
BC
LK
VS
S1
SW
S
VD
D
0.
1%
MF
1%
1%
50
V1
0%
1K
VM
1
45
23
50
V
VOLT/OHMS
6V
20
%T
M50
V
1%
MF
0.
1%
T
35
V2
0%
12
5 446
7 4
8 4
32
T1
6V
20
%
37
50
V2
0%
MF
0.
1%
1.
23
5V
MF
1%
54
53
38
23
8
29
9
57
36
15
17
30
24
37
59
51
12
11
10
35
14
20
21
41
46
31
47
48
49
50
55
33
16
18
58
56
13
26
25
27
45
43
42
40
34
32
19
22
23
45
67 3
9
28
60
1
52
44
0.
1%
MF
1%
1%
1%
1%
1%
1%
1%
25
%
1W
MF
+0
%-
20
%M
G5
%3
W
FLU
KE
7-6
00-1
201
(1 o
f 2)
as18f.eps
Figure 4-7. Model 7-600 Schematic
riešenia na presné meranie
http://www.elso.sk
Parts and SchematicsParts Lists, Drawings, and Schematics 4
4-17
TP
2T
P1
TP
32
.1
MH
Z
Y1
U2
P70P63P62P61P60X2X1VPPXT2XT1VDDP33P32P31P30P23P22P21P20P13P12P11P10P03
SO
SC
KP
00
P5
3P
52
P5
1P
50
VS
SP
43
P4
2P
41
P4
0V
LC
2V
LC
1V
LC
0B
IA
S
S12S13S14S15S16S17S18S19S20S21S22S23S24S25S26S27S28S29S30S31
COM0COM1COM2COM3
P7
1P
72
P7
3R
ES
ET
*S
0S
1S
2S
3S
4S
5S
6S
7S
8S
9S
10
S1
1
12
34
56
78
91
01
11
21
31
41
51
61
71
81
92
02
12
22
32
42
52
62
72
8
U3
CR
3B
AV
99
R2
3
1.
0M
2.
2M
R2
5
C1
3
0.1
22
C1
0
C4
0.1
S2
S2
S2
S2
SO
UT
SC
LK
AZ
EN
D
SE
LE
CT
RA
NG
E
CA
PA
CI
TA
NC
EMI
N/
MA
X4
5
35
25
15
20
%1
0V
AL
26
25
24
23
22
21
20
19
18
17
16
15
14
13
12
11
10
98
76
54
32
82
27
1
AL
UM
INU
MA
LC
AR
BO
N
FIL
MC
FC
AR
BO
N
CO
MP
CC
CO
MPO
NE
NT
T
YP
E
AB
BR
EV
IAT
ION
C
OD
ES
SY
MC
AP
AC
ITO
RS
YM
RE
SI
ST
OR
PO
LY
PR
OP
YL
EN
E
FIL
M
M C T PC
PP
MY
LA
R/
PO
LY
ES
TE
RC
ER
AM
ICT
AN
TA
LU
MP
OL
Y
CA
RB
ON
AT
E
MG
WW
MF
ME
TA
L
GL
AZ
EW
IR
E
WO
UN
DM
ET
AL
F
ILM
J CR
ELS
C
RE
FE
RE
NC
E
DE
SIG
NA
TIO
N
LA
ST
US
ED
NO
T
US
ED
QCR UBT
S TP
Y ZVR
RT
SE
LE
CT
PI
N
AS
SIG
NM
EN
TS
ME
MB
RA
NE
T
AIL
S2
CLO
SE
D
OP
EN
OP
EN
CLO
SE
D
OP
EN
OP
EN
CLO
SE
D
CLO
SE
D
OP
EN
LOW
Z
HIG
H
Z
OF
F
4-
52
-3
1-
2
CO
NTA
CT
SP
OS
ITIO
N
S1
CO
NN
EC
TIV
ITY
TA
BL
E
5
20
11,
12
11 11 3 22 2 3 1116 1
NO
TE
S:
UN
LE
SS
O
TH
ER
WI
SE
S
PE
CI
FI
ED
:
1. 2.A
LL
C
AP
AC
IT
AN
CE
V
AL
UE
S
AR
E
IN
M
IC
RO
FA
RA
DS
,+
/-
1
0%
,
CE
RA
MI
C.
3.A
LL
R
ES
IS
TA
NC
E
VA
LU
ES
A
RE
I
N
OH
MS
,
1/
8
WA
TT
,+
/-
5%
,
CE
RM
ET
.
+/
-1
%
RE
SI
ST
OR
S
AR
E
1/
8
WA
TT
,C
ER
ME
T.
4.
R1
9
IS
A
F
US
IB
LE
R
ES
IS
TO
R.
E
XA
CT
RE
PL
AC
EM
EN
T
ON
LY
.
6.R
11
A
ND
R
12
M
AY
H
AV
E
+/
-0
.2
5%
T
OL
ER
AN
CE
.
7.R
23
M
AY
H
AV
E
+/
-5
%
TO
LE
RA
NC
E.
8.
30 8
1 2 6543
VS
S
VS
S
C MIN
/MA
X
AP
AC
IT
AN
CE
RA
NG
E
C4
,
9,
1
2
&
13
M
AY
H
AV
E
+/
-2
0%
T
OL
ER
AN
CE
.
21222324
42
444546
484950515253
60616263641
34567891011121314151617181920
54
575859
5556
2
43
41
47
33
28
27
26
40
39
78
77
76
75
74
73
72
71
80
79
70
69
66
65
37
36
35
34
32
31
30
29
38
25
68
67
.5
%
10
%2
5V
10
%
25
V
VS
S2
VD
D
DA
TA
0
DA
TA
1
DA
TA
2
DA
TA
3
CL
K
BC
LK
VS
S1
WR
IT
E*
RE
AD
*A
DR
5A
DR
4
AD
R3
AD
R2
AD
R1
AD
R0
DG
ND
1%
FLU
KE
7-6
00-1
201
(2 o
f 2)
as19f.eps
Figure 4-7. Model 7-600 Schematic (cont)
riešenia na presné meranie
http://www.elso.sk
12B/18/7 SeriesService Manual
4-18
riešenia na presné meranie
http://www.elso.sk
Service Centers
PN 850941 2/96
riešenia na presné meranie
http://www.elso.sk
Service Centers (cont)
2/96
riešenia na presné meranie
http://www.elso.sk
Manual Supplement
© 1999 Fluke Corporation. All rights reserved. Printed in the U.S.A.
Manual Title: 12B,18,7 Series Service Supplement Issue: 3Part Number: 602730 Issue Date: 12/99Print Date: August 1996 Page Count: 1Revision/Date:
This supplement contains information necessary to ensure the accuracy ofthe above manual. Enter the corrections in the manual if either one of thefollowing conditions exist:
1. The revision letter stamped on the indicated PCA is equal to or higherthan that given with each change.
2. No revision letter is indicated at the beginning of the change.
riešenia na presné meranie
http://www.elso.sk
12B,18,7 Series Service Manual Supplement
12/99 1
Change #1On page 3-7, Table 3-2, Test No. 11,
Change: 400.0V dcTo: 400.0V dc4
Add footnote 4 to the bottom of the table:
4. Model 7-300 meter range is 300 Vdc.
On page 3-8, Table 3-3, Test No. 4, 12, and 13, change Switch Position,
From: RightTo: Center
On pages 4-3 and 4-5, Tables 4-2 and 4-3:
Delete both A1 entries in their entirety.On all MP2 entries, remove the word WINDOW from the Description.
On page 4-8, Table 4-4, change the part number for the first U2,
From: 601575To: 601757
Change #2On page 3-6, replace Table 3-1 with the following:
Equipment Minimum Specifications Recommended Model
DMM Calibrator DC Voltage: 0-600VAccuracy: ± 0.25%
AC Voltage: 0-600VAccuracy: ± 0.5%
Frequency: 50-400 Hz
Resistance: 1.0Ω– 40 MΩAccuracy: ± 0.25%
Capacitance: 0-1.000 µF
Accuracy: ± 0.5%
Fluke 5520A
ALTERNATE
Equipment Minimum Specifications Recommended Models
DMM Calibrator DC Voltage: 0-600VAccuracy: ± 0.25%
AC Voltage: 0-600VAccuracy: ± 0.5%
Frequency: 50-400 Hz
Fluke 5700A
Decade Resistor Resistance: 1.0 – 40 MΩAccuracy: ± 0.25%
General Resistance RDS-77Band 41B
Decade Capacitor Capacitance: 0-1.000 µFAccuracy: ± 0.5%
GenRad 1412-BC
riešenia na presné meranie