Field Portable Methods for the Determination of Arsenic
in Environmental Samples
2University of Massachusetts at Amherst Department of Chemistry
Presentation Outline 1. Research Goals, the Arsenic Problem, and Field Kits. 2. The Chemical Methods
• The Gutzeit Method: Hydride Generation of Arsenic.• The Molybdenum Blue Method.
3. Experimental• Project 1: 24 Hour Field Kit Sensitivity• Project 2: Measuring Arsenic in Soils with the Gutzeit Method• Project 3: Silver Nitrate as a Detection Reagent for the Gutzeit Method• Project 4: Molybdenum Blue and the Detection of Arsenic with
Cameras• Project 5: Flow Injection and the Determination of Arsenic• Project 6: The Stoichiometry of Heteropolyacids
4. SNPs Research 5. Future Work 6. Questions
3University of Massachusetts at Amherst Department of Chemistry
What is the Size of the Arsenic Problem?
Millions of people worldwide are chronically exposed to arsenic through drinking water, including 35—77 million people in Bangladesh.
Argos, M. et al. The Lancet, Early Online Publication, 2010
4University of Massachusetts at Amherst Department of Chemistry
The Goals of this PhD. Research Project
Goal: To develop a more reliable field portable chemical method to measure arsenic in environmental samples at, or below, 10 µg L-1 (ppb).
Areas of investigation: (1) Improvement in Gutzeit methodology for water and soil testing with digital image analysis and use of silver nitrate as a reagent (2) Optimization of the molybdenum blue chemistry (3) The single nucleotide polymorphism study to understand the health consequences of arsenic exposure.
5University of Massachusetts at Amherst Department of Chemistry
The Challenges of Laboratory Instruments:
(1) High Cost (2) Materials and Maintenance (3) Trained Technician
Is There a Need for Field Portable Instruments?
The Current Reliability of Field Portable Methods:
“Accurate, fast measurement of arsenic in the field remains a technical challenge. Technological advances in a variety of instruments have met with varying success. However, the central goal of developing field assays that reliably and reproducibly quantify arsenic has not been achieved”
Melamed, D. Anal. Chim. Acta, 2005, 532, 1-13.
Kinniburgh, D.G.; Kosmus,W. Talanta, 2002, 58, 165-180.
The Need for Field Kits: “The only feasible approach (for the measurement of the tube wells, which are estimated to be more than 10 million) is through the use of field kits.”
6University of Massachusetts at Amherst Department of Chemistry
The Gutzeit Test
Reaction 1 (aq): arsenite + zinc + acid produces AsH3 ,which rises into head space of reaction container. Reaction 2 (g): AsH3 reacts with mercuric bromide impregnated test strip. Measurement: Yellow-brown color produced after set time is compared with preprinted chart.
7University of Massachusetts at Amherst Department of Chemistry
The Gutzeit Method Chemistry
Zn(0) Zn2+ + 2e-
The Formation of Arsine (AsH3)
2H+ + 2e- H2
As(III) + 3e- As(0)
As(0) + 3e- +3H+ AsH3
AsH3 (g) + 3HgBr2 (aq) As(HgBr)3 (aq) + 3HBr
AsH3 (g) + 3AgNO3 (s) AsAg3 (s) + 3HNO3
The reaction of Arsine
Brindle, I. D. “Vapour-generation analytical chemistry: from Marsh to multimode sample-introduction system” Analytical Bioanalytical Chemistry 388, 2007, 735-741.
8University of Massachusetts at Amherst Department of Chemistry
Molybdenum Blue Method
Ammonium molybdate, sulfuric acid, a reducing agent and a catalyst are combined; the molybdate forms an inorganic polymer, which is then reduced and turns from yellow to blue.
9University of Massachusetts at Amherst Department of Chemistry
Molybdenum Blue Chemistry
Molybdate reacts with the +5 species of P, As, Sb, and Bi.
12 MoO42- + AsO4
3- + 24H+ → AsMo12O403-+ 12H2O
Chemical Reaction: for formation of molybdenum blue
Matsunaga, H.; Kanno, C.; Toshishige, M. Suzuki, T.M. Talanta, 2005, 66, 1287-1293.
Analytes which react with the molybdenum blue chemistry
The Stages: of molybdenum blue formation
1. Complex only reacts in a solution containing arsenic (V).2. After reduction, the complex’s Max is near 850nm.
10University of Massachusetts at Amherst Department of Chemistry
The “Molybdenum” Blue Complex
1. Arsenate + molybdate + acid + reducing agent gives blue color due to formation of heteropoly species containing both Mo (IV) and Mo (VI).
2. Octahedral subunits form the structure. 3. Arsenic substitutes for a molybdenum or trapped in the interior of the
larger polymer.
Gouzerh, P.; Proust, A. Main-group element, organic, and organometallic derivatives of polyoxometalates. Chem. Rev. 1998, 98, 77.
11University of Massachusetts at Amherst Department of Chemistry
Color Measurement and Tristimulus Colorimetry
Photons come in different wavelengths
According to tristimulus colorimetry theory, the human eye interacts with three regions of the electromagnetic spectrum
Detection methods measure light using tristimulus theories
Konica Minolta, the essentials of imaging web site http://www.konicaminolta.com/instruments/knowledge/light/concepts/08.html, (accessed August, 2010)
12University of Massachusetts at Amherst Department of Chemistry
Reflectance Spectroscopy
I = I0*e-kx
Reflectance spectroscopy operates according to Beer’s Law
Where I is observed light I0 is the original light intensity The value k is the absorption coefficient specific
for that substance at a specific wavelength. The value x is the distance the photons travel
through the substance
USGS, about reflectance spectroscopy website, http://speclab.cr.usgs.gov/aboutrefl.html, (accessed August, 2010)
13University of Massachusetts at Amherst Department of Chemistry
Quantification of Molecules Using Reflectance Spectroscopy
I = I0*e-kx
These methods used a tristimulus scanner I and Io are known, the precision of the emission of incident
wavelengths and their detection have to be further established The value k is the absorption coefficient specific for that
substance at a specific wavelength. The k value is not known because the reaction products are not homogeneous or characterized
The value x is the distance the photons travel through the substance because the thickness of the mercuric bromide is not known and are not uniform
The USGS reflectance spectroscopy places samples on glass, this experiment uses white plastic
14University of Massachusetts at Amherst Department of Chemistry
Project 1: Improving Field Kit Sensitivity using Digital Image Analysis
Time: five replicate measurements at 20, 30, 40 minutes and 24 hours at the concentrations of 10, 25, 50, 100, 250, 500 µg L-1 (ppb).
Temperature: five replicate measurements at 35° C and the concentrations of 10, 25, 50, 100, 250, 500 µg L-1 (ppb).
Determinations of (1) s (standard deviation of field kit measurements), (2) S0 (Standard deviation at zero concentration), (3) k (constant relative error) with time and temperature variations plus scanner use.
The Red, Green and Blue values were measured using computer software.
15University of Massachusetts at Amherst Department of Chemistry
The Research or Kinniburgh and Kosmus
Kinniburgh, D.G., Kosmus, W. Talanta, 2002, 58, 165-180
Thompson, M. Howarth, R.J. Analyst, 1976, 690
16University of Massachusetts at Amherst Department of Chemistry
Current Analytical Precision with the Gutzeit Method
Kinniburgh, D.G., Kosmus, W. Talanta, 2002, 58, 165-180
17University of Massachusetts at Amherst Department of Chemistry
Results: Tables of Standard Deviation Values and Mean Blue Values at Different Times
18University of Massachusetts at Amherst Department of Chemistry
The Standard Plot of Color Versus Concentration
Concentration of As(III) g L-1
19University of Massachusetts at Amherst Department of Chemistry
The Determination of Standard Deviation in Concentration
Concentration of As (III)
20University of Massachusetts at Amherst Department of Chemistry
The Values so and k at 24 Hours of Reaction Time
Concentration As (III) g L-1 Standard Deviation in Concentration
10 2.382725 4.147850 6.5894
As(III) g L-1
21University of Massachusetts at Amherst Department of Chemistry
Results: Table of So and K Values at Different Times
22University of Massachusetts at Amherst Department of Chemistry
Results: Graphs Comparing Time and Temperature
Concentration of As (III) g L-1
23University of Massachusetts at Amherst Department of Chemistry
Conclusions of DIA Experiment
1. Scanning improves precision compared to naked eye determination at 20 minutes.
• Naked eye (Kinniburg) k = 0.3 and So = 7.
• This method produced k = 0.2 and So = 2.6.
2. Increasing the reaction time from 20 minutes to 40 minutes also decreases k (for 20, 30, and 40 minutes) and increases So(for 20,30, and 40 minutes).
3. Running the reaction at 35°C produces results similar to 40 minutes and 24 Hours.
24University of Massachusetts at Amherst Department of Chemistry
Conclusions of DIA Experiment Table of s Values at 10 and 50 g L-1