Discipline: Physics
Subject: Characterization Techniques for Materials II
Unit 3: Surface Morphology
Lesson/ Module: Atomic Force Microscopy-I
Author (CW): Dr. Ajit K. Mahapatro
Department/ University: Department of Physics and Astrophysics,
University of Delhi, New Delhi-110007
Characterization
Techniques for Materials II [e-PG Pathshala]
Department of Physics and Astrophysics, University of Delhi, Delhi 1
Contents
Learning Objectives…………………………………...…………………………………..……….2
1. Introduction ………………………………..……………………………………...........…............3
2. Basic working principle of AFM……..…………….…….………………………………...…………3
2.1. Surface Sensing………….…….……………………………..…………………….………4
2.2. Detection of Cantilever Deflection………….…….……………………………..…………4
2.3. Image Processing………….…….……………………………..…………………………4
2.4 Feedback Mode Operation………………………..……………………………………4
3. Components of AFM……….……………………………………….…………….….……………6
3.1 Piezo ceramic scanner …………………………………………………………..………6
3.1.1 Raster scanning …………………………………………………………..……….…7
3.1.2 Hysteresis …………………………………………………………..……….…7
3.1.3 Creep…………………………………………………………..……….…8
3.2 Probes ……………………………………………………………………………...8
4. Force Measurement ……………………………………………………………………………9
5. Beam Deflection Detection………………………………………………………………………11
6. Scanning of the Sample Surface ……………………………………..………………………….11
7. Comparison between AFM and Electronic Microscopes ……………………………………12
Summary……………………………………………………………………………………..13
Characterization
Techniques for Materials II [e-PG Pathshala]
Department of Physics and Astrophysics, University of Delhi, Delhi 2
Learning Objectives:
From this module students may get to know about the following:
i. Basic concepts and necessary components of AFM
ii. Components of the AFM
iii. Force Measurements and Beam Deflection Detection Leading to Scanning
Sample Surfaces
Characterization
Techniques for Materials II [e-PG Pathshala]
Department of Physics and Astrophysics, University of Delhi, Delhi 3
1. Introduction
Imaging the surface features at nano/micro-meter scale provide realization of systems at the atomic and
molecular level, and helps gaining in-depth knowledge that could lead to new research discoveries and
technological developments in the areas of materials science, polymer science, electrochemistry, life
science, biophysics, biotechnology and nanotechnology. Atomic force microscopy (AFM) was invented
by Binning in 1986. Currently, AFM is used to map and study the three-dimensional surface topography
at micro/nano-meter scale for various types of materials including metals, semiconductors, soft biological
samples, and conductive and non-conductive materials, in different environments of air, liquid, and
vacuum. AFM can generate images at atomic resolution with height resolution at angstrom scale
precession. Recently, AFM is used as a technique for performing manipulations of nano-objects and in
nano-imprint technology for writing features at molecular dimension using molecular ink.
2. Basic Working Principle of AFM
The basic principle is based on capturing the interaction force between an interacting probe (supported on
a flexible cantilever) of tip dimensions at molecular level and surface of the sample. Depending on the
process of interaction, different operation modes could be defined. There are three important stages in
AFM (as schematized in Figure 1) combined to generate the three dimensional topography of the
sample’s surface at atomic resolution, including (1) surface sensing, (2) detection of cantilever deflection,
and (3) image processing.
Figure 1. Schematics of Atomic force microscopy (AFM)
Characterization
Techniques for Materials II [e-PG Pathshala]
Department of Physics and Astrophysics, University of Delhi, Delhi 4
2.1 Surface Sensing
A cantilever with molecular sized tip is used to scan over surface of the sample (as shown in Figure 2),
and measured the forces between the cantilever of dimension < 10 nm and surface at very short distances
(~ 0.2-10 nm). As the tip approaches the surface, attractive force between the surface and tip causes the
cantilever to attract towards the surface. With the cantilever approaching more closer to the surface, the
tip is almost in contact and develops repulsive force, resulting deflection of the cantilever away from the
surface. This process of deflection of cantilever due to surface sensing is performed at every point of the
cantilever-surface positioning during scanning.
Figure 2. Interaction of the cantilever with surface causes bending of cantilever in AFM
2.2 Detection of Cantilever Deflection
A laser beam is used to detect the position of the cantilever throughout the process during approaching
with attractive force of the tip towards the sample surface and deflecting away from the surface due to
repulsive force on reaching more close to the sample.
An incident beam is reflected from the top of the cantilever and monitored the change in the direction of
the reflected beam with respect to the cantilever position. A position sensitive photo diode (PSPD)
capable of detecting changes in Amstrung precision is used to track any slight changes in the cantilever.
During measurement of the surface curvature, the cantilever deflects when the AFM tip passes over the
top of a surface feature, and the subsequent change in position of reflected beam in the scanning direction
is recorded through the PSPD.
2.3 Image Processing
An AFM images the topography of a sample surface by scanning the cantilever over a region of interest.
The raised and lowered features on the sample surface influence the deflection of the cantilever, which is
monitored by the PSPD. By using a feedback loop to control the height of the tip above the surface is
maintained at constant laser position.
2.4 Feedback Mode Operation
Feedback control is used in AFM for maintaining a consistent interaction or force between the
probe and surface. The schematic of a typical feedback system for AFM is given in Figure 3.
Characterization
Techniques for Materials II [e-PG Pathshala]
Department of Physics and Astrophysics, University of Delhi, Delhi 5
Figure 3. Schematic description of AFM feedback loop
The feedback control measures the force between the surface and probe, activates the piezoelectric
ceramic to establish a relative position of the probe and surface to maintain a consistent force between
them at a user-specified set point level. Foe example, in contact mode, when the control signal (the
cantilever deflection) is above the setpoint, the feedback loop tries to reduce the force between the
cantilever and the sample by contracting the piezo in z-axis, and moves the tip a little further away from
the surface. When the deflection is less than the specified setpoint, the feedback loop expands the piezo in
z-direction to move the tip closer to the surface of the sample. Although signal processing varies
according to the image mode used (contact mode, tapping mode, etc), the feedback loop always performs
essentially the same function. The feedback system used to control the tip-sample interactions and the
resulting images are optimized for each new sample. This optimization is accomplished by adjusting
various gains in the feedback circuit of the scanning probe microscopy.
In the AFM, the feedback control electronics take an input from the force sensor and compares the signal
to a setpoint value. An error signal (Zerr) is then sent through a feedback controller. The output of the
feedback controller (Zout) then drives the z-piezo through a high voltage amplifier. The most common
form of feedback control for AFM is the PID controller consisting of proportional (P), integral (I) and
derivative (D) controller, takes the error signal and processes it as:
where P, I, and D are the gain settings appropriately selected for tracking the surface by the probe as it is
scanned. The ‘I’ term facilitates the probe to move over large surface features, and P and D terms allow
the probe to follow the smaller, high-frequency features on a surface. When the PID parameters are
optimized, the error signal image is minimized (Figure 3). The higher the feedback gains, the faster the
feedback loop reacts to changes in topography while scanning. However, since the AFM is not infinitely
fast in responding to the output of the PID controller, the increase of the feedback gains to a certain point
is limited, beyond this the feedback loop becomes unstable. This is the limiting factor for the maximum
achievable scan speed.
Characterization
Techniques for Materials II [e-PG Pathshala]
Department of Physics and Astrophysics, University of Delhi, Delhi 6
3. Components of the AFM
Figure 4. Components of AFM
3.1. Piezo ceramic scanner
Scanning probe microscopy (SPM) scanners are composed of piezoelectric materials. The scanner is
constructed by the independent combination of piezo-electrodes for x, y, and z position coordinates into a
single tube or flexure scanner. The scanner formed can manipulate samples and probes in the three
dimensions with highly accurate precision. These are ceramic materials extend or contract with respect to
the polarity of the applied voltage gradient and conversely develop electrical potential in response to
mechanical pressure. In this way, movements in x, y and z direction are possible.
Figure 5. Schematics showing piezoelectric effect depending on the applied bias.
Characterization
Techniques for Materials II [e-PG Pathshala]
Department of Physics and Astrophysics, University of Delhi, Delhi 7
3.1.1 Raster scanning
A series of voltage ramps are created by the x-y signal generators to drive the x and y piezoelectric
ceramics in the pizotube scanner of AFM (as shown in Figure 6). The scan range is set by adjusting the
minimum and maximum values for the applied voltage. The scanning position is set by offsetting the
voltages to the ceramic. Finally, the scan orientation is rotated by changing the phase between the signals.
Figure 6. Construction of piezotube scanner
3.1.2 Hysteresis
Differences in the properties of material and dimensions of piezoelectric element, develops signals with
scanner responding in a different way to the applied bias. This response is conveniently recorded in terms
of sensitivity, which is the ratio of piezo movement to piezo voltage, i.e. how far the piezo expands or
contracts with the applied voltage. Sensitivity is not a linear relationship with respect to scan size. Piezo
materials have inherent nonlinearities and shows hysteresis. The effect of nonlinearity and hysteresis can
be seen from the curves presented in Figure 7. As the piezo expands and retracts throughout its full range,
its movement with applied bias is reduced in the beginning of the extension but larger at the end. The
same process is repeated when the piezo is retracted from it’s position. This produces forward and reverse
scan directions to show hysteresis behavior between two scan directions. Nonlinearity and hysteresis can
cause feature distortion in SPM images and has to be properly corrected.
Characterization
Techniques for Materials II [e-PG Pathshala]
Department of Physics and Astrophysics, University of Delhi, Delhi 8
Figure 7. Schematics showing piezoelectric effect depending on the applied bias.
3.1.3 Creep
The drift of the piezo displacement after a DC offset voltage is applied to the piezo ceramic is known as
creep (Figure ) and occurs either by changes in x and y offsets, or while the usage of “frame up” and
“frame down” commands such that piezo rapidly travels over a large area to restart the scan. When a large
offset is applied, the scanner stops scanning and a DC voltage is applied to the scanner for the movement
of offset distance. The majority of the offset distance is moved rapidly by the scanner and slows down for
the remaining distance. Resuming the scan, after a large offset distance moves with the scanner moving
slowly in the direction of the offset. Creep is the result of this slow movement of the piezo over the
remaining offset distance once scanning is resumed. Creep appears in the image as an elongation and
stretching of features in the direction of the offset for a short duration of time after the offset.
3.2. Probe
The probe represents a micro machined cantilever with a sharp tip at one end, which is brought into
interaction with the sample surface. Each probe has different specifications and shape. The two most
common geometries for AFM cantilevers are rectangular ("diving-board") and triangular. Generally, V-
shaped cantilevers are used due to low mechanical resistance to vertical deflection and high resistance to
lateral torsion. The dimension of the cantilevers ranges from 100 to 200 µm in length (l), 10 to 40 µm in
width (w), and 0.3 to 2 µm in thickness (t).
Integrated cantilevers are usually made from silicon (Si) or silicon nitride (Si3N4). They are characterized
by their force constant and resonant frequency, which have to be chosen according to the sample to be
studied. Additionally an optical detection system and electronics for the management of scanning
procedures and data acquisition are necessary.
AFM cantilevers are typically made either of silicon or silicon nitride, where silicon nitride is reserved for
softer cantilevers with lower spring constants. The dimensions of the AFM cantilever are very important
as they dictate its spring constant or stiffness; this stiffness is fundamental to governing the
Characterization
Techniques for Materials II [e-PG Pathshala]
Department of Physics and Astrophysics, University of Delhi, Delhi 9
interaction between the tip and the sample and can result in poor image quality if not chosen carefully.
The relationship between the cantilever’s dimensions and spring constant (k) is defined by the equation:
k = Ewt3/4L3
where, w is the cantilever width, t is the cantilever thickness, L is the cantilever length, and E is the
Young’s modulus of the cantilever material.
4. Force Measurement
The cantilever is designed with a very low spring constant (easy to bend), making it highly sensitive to
force. The probe is attached on the end of a cantilever. The amount of force between the probe and
sample depends on the spring constant (stiffness of the cantilever), and distance between the probe and
sample surface. This force is described using Hooke’s Law:
F = - (k. x)
where, k is the spring constant and x is the cantilever deflection
If the spring constant of cantilever (typically ~ 0.1 - 1 N/m) is less than that for surface, the cantilever
bends and the deflection is monitored. The interaction of the probe with the force field associated with the
surface of the sample provides the dependence of the force upon the distance between the tip and the
sample (Fig. 8). The force measured by AFM can be classified into long-range forces and short-range
forces. The long-range forces dominate when scanned at large distances from the surface and can be Van-
der Waals force, capillary forces (due to the water layer often present in an ambient environment). When
the scanning is in contact with the surface the short range forces are very important, in particular the
quantum mechanical forces (Pauli Exclusion Principle forces).
Figure 8. Potential energy curve of a probe and sample.
Characterization
Techniques for Materials II [e-PG Pathshala]
Department of Physics and Astrophysics, University of Delhi, Delhi 10
In the contact mode, the cantilever is less than a few angstroms away from the surface of the sample and
the interatomic force acting between the probe and the sample is repulsive (the purple region shown in
Fig. 8). The magnitude of this repulsive force increases as the probe begins to contact the surface and at a
later time it causes the cantilever to bend up.
In the non-contact mode, the cantilever is held on the order of 10-100 Å away from the sample’s surface.
Thus as result of long range Van der Waals interaction, the interatomic force between the probe and the
sample is attractive (the blue area in the Fig. 8). Attractive force near the surface is caused by a nanometer
sized layer of contamination present over the surfaces at ambient conditions. The amount of
contamination depends on the environment in which the microscope is being operated.
One complete cycle of the probe approaching towards the surface of the material, getting into contact, and
bending away from the sample is demonstrated in Fig. 9. At the right side of the curve (position A) the
scanner is fully retracted and the cantilever remains undeflected, since the tip is not in contact with the
sample. As the scanner experiences the attractive Van der Waals force, the cantilever comes close enough
to the sample surface. In the position B, the cantilever suddenly bends slightly towards the surface. As
the scanner continues to extend, the probe experiences a repulsive force (position D) and the cantilever
deflects away from the surface, approximately linearly (blue color). After full extension, at the extreme
left of the plot (position C, black color), the scanner begins to retract. The cantilever deflection retraces
the same curve. In the position E, the scanner retracts enough to make the tip spring-free.
Figure 9. Force dependence on the tip-sample distance
Characterization
Techniques for Materials II [e-PG Pathshala]
Department of Physics and Astrophysics, University of Delhi, Delhi 11
5. Beam Deflection Detection
The laser is focused to reflect from the cantilever and onto the sensor. The position of the beam in the
sensor measures the deflection of the cantilever, and hence the force between the tip and sample. The
probe of the AFM scans the sample and the readings for the displacement of the cantilever is collected by
the photodiode through a laser beam reflected from the back of the cantilever. The photo diode is divided
into four parts as shown in the Fig 10. When the laser is displaced vertically along the positions top (1, 2)
and bottom (3, 4), a bending due to topography is developed, while if the displacement is horizontally
along the left (1, 4) and right (2, 3), a lateral force is produced i.e. torsion due to friction.
Finally, the topographic information = Photodiode signals [(1)+(2)]-[(3)+(4)]
Figure 10. Beam deflection detection
6. Scanning of the Sample Surface
The tip passes back and forth in a straight line across the sample. In a typical imaging mode, the tip
sample force is held constant by adjusting the vertical position of the tip (feedback). Tip is brought within
nanometer range of the sample.
A topographic image is built up recording the vertical position when the tip is rastered across the sample,
using a computer (Figure 11). During contact with the sample, the probe predominately experiences
repulsive Van der Waals forces (contact mode). This leads to the tip deflection as described below. As the
tip moves further away from the surface, attractive Van der Waals forces are dominant (non-contact
mode).
Characterization
Techniques for Materials II [e-PG Pathshala]
Department of Physics and Astrophysics, University of Delhi, Delhi 12
Figure 11. Raster motion during Scanning in AFM
Stiffer cantilevers protect against sample damage because they deflect less in response to a small force.
This means a more sensitive detection scheme is needed. Radius of the tip of cantilever limits accuracy
for the image analysis and resolution, as shown in Fig. 12 below.
Figure 12. Tip radius dependence on image resolution.
7. Comparison between AFM and Electronic Microscopes
Optical and electron microscopes can easily generate two dimensional images of a sample surface. The
magnification is as large as 1000 for an optical microscope, and a few hundreds thousands ~100,000
for an electron microscope. However, these microscopes can’t measure the vertical dimension (z-
direction) of the sample, the height (e.g. particles) or depth (e.g. holes, pits) of the surface features. AFM,
which uses a sharp tip to probe the surface features by raster scanning, can image the surface topography
with extremely high magnifications, up to 1,000,000 , comparable or even better than electronic
microscopes. In AFM, the measurement is performed in three dimensions, the horizontal X-Y plane and
the vertical Z dimension. Resolution (magnification) at Z-direction is normally higher than X-Y.
Characterization
Techniques for Materials II [e-PG Pathshala]
Department of Physics and Astrophysics, University of Delhi, Delhi 13
Summary
The AFM can generate an accurate topographic map of the surface features.
The interaction of the probe with the force field associated with the surface of the
sample provides the dependence of the force upon the distance between the tip
and the sample.
A topographic image is built up on the computer by recording the vertical
position as the tip is rastered across the sample.
Force Measurements and Beam Deflection Detection Leading to Scanning
Sample Surfaces.