Hema Star II
Technical information
Hema Star II S -A / rév.1.0 YSMAN067
2
Summary
SUMMARY 2
ELECTRICITY SUPPLY 3
1. GENERAL REMARKS 3 2. FIRST POWER SUPPLY 3 3. SECOND POWER SUPPLY 3 4. SCREEN 3
DESCRIPTION OF ELECTRONIC BOARDS 4
1. SYSTEM BOARD 4 2. PREAMPLIFIER BOARD L10-CI-PRA20 4 3. PROCESSING BOARD L1O-PR-FIL3O 4 4. MECHANICAL / FLUIDIC MANAGEMENT BOARD L10-CI-DIL30 4 5. INTERFACE BOARD L1O-CI-INT-AA 4
SEQUENCING OF MAIN FLUIDIC CYCLES 5
1. DILUTION SEQUENCING 5 2. COUNTING CYCLE 6 3. RINSING AND DRAINAGE AT THE END OF COUNTING CYCLE (DILUTER) 6 4. RINSING AND DRAINAGE AT THE END OF CYCLE (FLUIDIC) 6 5. SEQUENCING FIRST RINSING CYCLE 7 6. SEQUENCING OF RINSING CYCLE EXCEPT THE FIRST 8 7. SEQUENCING OF RINSING CYCLES AFTER 5 MINUTES 9
LIST OF ELECTRONIC SCHEMES 10
LIST OF FLUIDIC SCHEMES 10
Hema Star II S -A / rév.1.0 YSMAN067
3
Electricity supply 1. General remarks The device works on 10w voltage power from two sources in ventilated casing. The main supply element is Iocated at the back of the device preventing any risk of electrocution. Important: These partitioned power supplies do not have any accessible fuses. Contact your technical support. N.B: the presence of the different voltages displayed by leds on the interface board (appendix E3). 2. First power supply PC type power supply giving all boards the following voltages: • +12V • -12V • +5V Power supply linked to the floppy disk driver and the system board, then to all other boards via the expansion slots and cable layers. It works on 50 or 60 Hz and there is a switch between the two EEC jacks for changing to 115 or 230V. 3. Second power supply Works on 30 VI 2 A, feeding the interface board. Accepts from 47 to 440 Hz and 2 voltage ranges, switched over by internal strap or automatic according to the version: • - 95 to 132V • -187 to 265V 4. Screen Works on fixed voltage 220 V, 50 to 60 Hz. On request, it may be supplied with different power supply voltages.
Hema Star II S -A / rév.1.0 YSMAN067
4
Description of electronic boards Refer to block diagram : appendix E1. 1. System board Board of 80386 or 80486 type. The program, written in an advanced language, loaded in the core memory after power on. The dialog with the peripheral boards L10-CI-FIL-30 and L10-CI-DIL30 is in mode 1/0 by reciprocal interruption. 2. Preamplifier board L10-CI-PRA20 It polarizes the two counting probes and amplifies the signals. The hemoglobin determination system is on the preamplifier board. 3. Processing board L1O-PR-FIL3O Made up of three structures with independent Z80 microprocessors. Each structure has a specific task : • counting of red corpuscles • platelets • white corpuscles / hemoglobin. The different signals are sorted out on peak width and height. 4. Mechanical / Fluidic management board L10-CI-DIL30 Two Z80 structures deal with the mechanical and fluidic management of the device. They control all the fluidic functions and all the motors following the program of the different cycles. 5. Interface board L1O-CI-INT-AA Power interface board which is linked to the management board by a 60 lead cable layer. Diluter Inputs / Outputs : • needle mechanism • liquid syringe • from electrovalve EV1 to EV3 • start button • red and green leds to start the system up. Fluidics Inputs / Outputs : • volumetric counting sensor and drainage optic sensor • liquid micro-pump • from electrovalve EV4 to EVI 2 • air syringe
Hema Star II S -A / rév.1.0 YSMAN067
5
Sequencing of main fluidic cycles 1. Dilution sequencing
PHASE SYRINGE NEEDLE HOLDER EV1 EV2 EV3 BLOOD SAMPLE
Sample – 32 µl Amount – 11 ml
Total ascent
EXT. RINSING OF NEEDLE
3 ml ext. needle Descent 40 mm Init → rinsing û FIRST DILUTION
2 ml ext. needle 2 ml int. needle 1.5 ml ext. needle 1.5 ml int. needle to init
Ascent 15 mm Descent 15 mm Ascent 5 mm Ascent 5 mm
Rinsing → WBC
û
û û
û
û
û û
û DROP EXTRACTOR
Flush elimination
Descent 15 mm Ascent 15 mm
SAMPLE IN DILUTION 1
Sample –43.7 µl Amount –10 ml
Descent 23 mm Ascent 29 mm
EXT. RINSING OF NEEDLE
1.5 ml ext. needle
WBC → rinsing
û
SECOND DILUTION
Descent 9 mm
Rinsing → RBC
REQUEST TRANSFER CHAMBER WBC + LYSIS 6.5 ml ext. needle û û
1.5 ml int. needle û û û RINSING LIQUID IN CHAMBER WC
Amount – 7 ml 7 ml ext. needle
Ascent 5 mm RBC → WBC
û
REQUEST TRANSFER CHAMBER RBC RINSING LIQUID IN CHAMBER RC
Amount – 7 ml 6 ml ext. needle 1 ml int. needle
WBC → RBC
û û
û û
REQUEST COUNTING CYCLE POSITION AWAIT OF COUNTING
To switch Descent 20 mm
RBC → rinsing
Hema Star II S -A / rév.1.0 YSMAN067
6
2. Counting cycle
PHASE ACTION FIRST COUNTING CHART Pressure lowering with air syringe
Injection of a bubble into the capillary Bubble passes in front of the start sensor Bubble passes in front of the end sensor Control of counting time Transfer of results WC, RC, PLT to PC
SECOND COUNTING CHART Counting of duration determined by first chart Transfer of results WBC, RBC, PLT to PC
THIRD COUNTING CHART Pressure lowering with air syringe Injection of a bubble into the capillary Bubble passes in front of the start sensor Bubble passes in front of the end sensor Control of counting time Transfer of results WC, RC, PLT to PC
FOURTH COUNTING CHART Counting of duration determined by third chart Hemoglobin measure Transfer of results WBC, RBC, PLT to PC
3. Rinsing and drainage at the end of counting cycle (diluter)
PHASE SYRINGE NEEDLE HOLDER EV1 EV2 EV3 AFTER THE END OF COUNTING
Backlash elimination
Total ascent Descent 75 mm
To slide Slide return Complete return
READY FOR ANOTHER CYCLE
4. Rinsing and drainage at the end of cycle (fluidic)
CHAMBER DIL. WBC CHAMBER DIL. RBC CHAMBER WBC CHAMBER RBC Drainage Transfer
CHAMBER READY Drainage Drainage Transfer
CHAMBER READY Drainage Drainage control Drainage control
Hema Star II S -A / rév.1.0 YSMAN067
7
5. Sequencing first rinsing cycle PHASE SYRINGE NEEDLE HOLDER EV1 EV2 EV3
REQUEST CAPILLARY RINSING REQUEST DRAINAGE OF CHAMBERS WBC AND RBC
1st DISTRIBUTION TO RINSING CHAMBER
Amount 9 ml 4 ml ext. needle
Ascent 50 mm Descent 50 mm
Init → rinsing
û
û
REQUEST DRAINAGE OF NEEDLE RINSING CHAMBER 4 ml int. needle û û
REQUEST DRAINAGE OF NEEDLE RINSING CHAMBER 2nd DISTRIBUTION TO RINSING CHAMBER
Amount 9 ml 4 ml ext. needle
û
û
REQUEST DRAINAGE OF NEEDLE RINSING CHAMBER 4 ml int. needle û û û
REQUEST DRAINAGE OF RINSING CHAMBER REQUEST DRAINAGE OF COUNTING CHAMBER WBC
1st DISTRIBUTION TO CHAMBER WBC
Amount 8 ml 5 ml ext. needle 2 ml int. needle
Ascent 30 mm Descent 30 mm
Rinsing → WBC
û û
û REQUEST TRANSFER CHAMBER WBC
REQUEST DRAINAGE OF COUNTING CHAMBER WBC 2nd DISTRIBUTION TO CHAMBER WBC
Amount 8 ml 5 ml ext. needle 2 ml int. needle
Ascent 30 mm Descent 30 mm
WBC → RBC
û û
û û
1st DISTRIBUTION TO CHAMBER RBC
Amount 9 ml 6 ml ext. needle 2 ml int. needle
û û
û û
REQUEST TRANSFER CHAMBER RBC REQUEST DRAINAGE OF COUNTING CHAMBER RBC
2nd DISTRIBUTION TO CHAMBER RBC
Amount 9 ml 6 ml ext. needle 2 ml int. needle
û û
û û
REQUEST TRANSFER CHAMBER RBC REQUEST DRAINAGE OF COUNTING CHAMBER RBC
3rd DISTRIBUTION TO CHAMBER WBC
Amount 9 ml
REQUEST TRANSFER CHAMBER WBC + LYSIS 6 ml ext. needle
2 ml int. needle û
û
û û
û û
REQUEST TRANSFER CHAMBER RBC REQUEST COUNTING CYCLE
Hema Star II S -A / rév.1.0 YSMAN067
8
6. Sequencing of rinsing cycle except the first PHASE SYRINGE NEEDLE HOLDER EV1 EV2 EV3
REQUEST CAPILLARY RINSING REQUEST DRAINAGE OF CHAMBERS WBC AND RBC
1st DISTRIBUTION TO RINSING CHAMBER
Amount 5 ml 2 ml ext. needle 2 ml ext. needle
Total ascent Descent 50 mm
Init → rinsing
û û
û û
REQUEST DRAINAGE OF NEEDLE RINSING CHAMBER DISTRIBUTION TO CHAMBER WBC
Amount 9 ml 6 ml ext. needle 2 ml int. needle
Ascent 30 mm Descent 30 mm
Rinsing → WBC
û û
û û
DISTRIBUTION TO CHAMBER RBC
Amount 9 ml
Ascent 30 mm Descent 30 mm
WBC → RBC
REQUEST TRANSFER CHAMBER WBC 6 ml ext. needle
2 ml int. needle û
û
û û
û û
REQUEST TRANSFER CHAMBER RBC REQUEST COUNTING CYCLE
Hema Star II S -A / rév.1.0 YSMAN067
9
7. Sequencing of rinsing cycles after 5 minutes PHASE SYRINGE NEEDLE HOLDER EV1 EV2 EV3
REQUEST CAPILLARY RINSING REQUEST DRAINAGE OF CHAMBERS WBC AND RBC
1st DISTRIBUTION TO CHAMBER WBC
Amount 8 ml 5 ml ext. needle 2 ml ext. needle
Total ascent Descent 50 mm
Init → WBC
û û
û û
REQUEST TRANSFER CHAMBER WBC 2nd DISTRIBUTION TO CHAMBER WBC
Amount 8 ml
û
REQUEST DRAINAGE OF CHAMBER WBC 5 ml ext. needle
2 ml int. needle û
û
û û
DISTRIBUTION TO CHAMBER RBC
Amount 8 ml Ascent 30 mm Descent 30 mm
WBC → RBC
REQUEST TRANSFER CHAMBER WBC 5 ml ext. needle
2 ml int. needle û
û
û û
û û
REQUEST TRANSFER CHAMBER RBC REQUEST CAPILLARY RINSING REQUEST CAPILLARY RINSING MECHANICAL INITIALIZATIONS
Hema Star II S -A / rév.1.0 YSMAN067
10
List of electronic schemes Appendix D1 : Preamplifier board L10-CI-PRA20 Appendix D2 : A/D conversion board L10-ANA20 Appendix D21 : Component location on conversion board Appendix D3 : Processing board L10-CI-FIL30 Appendix D31 : Component location on processing board Appendix D4 : Mechanical / Fluidic management board L10-CI-DIL30 Appendix D41 : Component location on mechanical / fluidic management
board Appendix D5 : Interface board L10-CI-INT-AA Appendix D51 : Component location on interface board Appendix E1 : Machine block diagram Appendix E22 : Trimmer location on preamplifier board Appendix E3 : Trimmer location on interface board Appendix E44 : Trimmer location on A/D conversion board
List of fluidic schemes Appendix F1 : Fluidic cabling Appendix F2 : Fluidic skeleton diagram Appendix F21 : WBC dilution transfer Appendix F22 : RBC dilution transfer Appendix F23 : Volumetric sensor rinsing (detergent circuit) Appendix F24 : Counting Appendix F25 : Drainage of counting chambers Appendix F26 : Diluter sequences Appendix F3 : Dismounting of ruby apertures Appendix F4 : Air and liquid syringe greasing
Hema Star II S -A / rév.1.0 YSMAN067
11
Appendix D1
PREAMPLIFIER BOARD L10-CI-PRA20
Hema Star II S -A / rév.1.0 YSMAN067
12
Appendix D2
A/D CONVERSION BOARD L10-ANA20
Hema Star II S -A / rév.1.0 YSMAN067
13
Appendix D21
COMPONENT LOCATION ON CONVERSION BOARD
Hema Star II S -A / rév.1.0 YSMAN067
14
Appendix D3
PROCESSING BOARD L10-CI-FIL30
Hema Star II S -A / rév.1.0 YSMAN067
15
Appendix D31
COMPONENT LOCATION ON PROCESSING BOARD
Hema Star II S -A / rév.1.0 YSMAN067
16
Appendix D4
MECHANICAL / FLUIDIC MANAGEMENT BOARD L10-CI-DIL30
Hema Star II S -A / rév.1.0 YSMAN067
17
Appendix D41
COMPONENT LOCATION ON MECHANICAL / FLUIDIC MANAGEMENT BOARD
Hema Star II S -A / rév.1.0 YSMAN067
18
Appendix D5
INTERFACE BOARD L10-CI-INT-AA
Hema Star II S -A / rév.1.0 YSMAN067
19
Appendix D51
COMPONENT LOCATION ON INTERFACE BOARD
Hema Star II S -A / rév.1.0 YSMAN067
20
Appendix E1
MACHINE BLOCK DIAGRAM
Hema Star II S -A / rév.1.0 YSMAN067
21
Appendix E22
TRIMMER LOCATION ON PREAMPLIFIER BOARD
Hema Star II S -A / rév.1.0 YSMAN067
22
Appendix E3
TRIMMER LOCATION ON INTERFACE BOARD
Hema Star II S -A / rév.1.0 YSMAN067
23
Appendix E44
TRIMMER LOCATION ON A/D CONVERSION BOARD
Hema Star II S -A / rév.1.0 YSMAN067
24
Appendix F1
FLUIDIC CABLING
Hema Star II S -A / rév.1.0 YSMAN067
25
Appendix F2
FLUIDIC SKELETON DIAGRAM
Hema Star II S -A / rév.1.0 YSMAN067
26
Appendix F21
WBC DILUTION TRANSFER
Tubes concernés EV contrôlées x
Hema Star II S -A / rév.1.0 YSMAN067
27
Appendix F22
RBC DILUTION TRANSFER
Tubes concernés EV contrôlées x
Hema Star II S -A / rév.1.0 YSMAN067
28
Appendix F23
VOLUMETRIC SENSOR RINSING (DETERGENT CIRCUIT)
Tubes concernés EV contrôlées x
Hema Star II S -A / rév.1.0 YSMAN067
29
Appendix F24
COUNTING
Tubes concernés EV contrôlées x
Hema Star II S -A / rév.1.0 YSMAN067
30
Appendix F25
DRAINAGE OF COUNTING CHAMBERS
Tubes concernés EV contrôlées x
Hema Star II S -A / rév.1.0 YSMAN067
31
Appendix F26
DILUTER SEQUENCES
Needle external distribution Needle internal distribution
Blood or 2nd
dilution extraction Diluent loading
Appendix F3
DISMOUNTING OF RUBY APERTURES
Rubis S -A / rév.2
33
Appendix F4
AIR AND LIQUID SYRINGE GREASING