You Can Measure the Benefits… Training Course SmartScan Setup & Configuration.

46
You Can Measure the Benefits… Training Course SmartScan Setup & Configuration

Transcript of You Can Measure the Benefits… Training Course SmartScan Setup & Configuration.

Page 1: You Can Measure the Benefits… Training Course SmartScan Setup & Configuration.

You Can Measure the Benefits…

Training Course

SmartScan Setup & Configuration

Page 2: You Can Measure the Benefits… Training Course SmartScan Setup & Configuration.

You Can Measure the Benefits…

Main Menu

Password window Enter the password code

(716)

Single sensor

Than

Main Menu

Entering Main Screen

Page 3: You Can Measure the Benefits… Training Course SmartScan Setup & Configuration.

You Can Measure the Benefits…

Main Screen

Main Menu

Page 4: You Can Measure the Benefits… Training Course SmartScan Setup & Configuration.

You Can Measure the Benefits…

Selecting Indication

DIST – To indicate distance to target

LEVEL – To indicate Level of medium

FLOW – To indicate Flow in Open Channel application

Main Menu -SmartScan 50

Page 5: You Can Measure the Benefits… Training Course SmartScan Setup & Configuration.

You Can Measure the Benefits…

Main Menu -SmartScan 50

Setting an indication mode enables you to toggle between other indications types as well, using BACK or NEXT buttons, as described in the table below

Display Indication

Modes

IndicationIndication Toggle Option

Distance/LevelDistance/Level

FlowFlow/Level/Distance/Totalization

(High)/Totalization (Low)

Totalization/Auto toggling

Totalization (High)/Totalization (Low)/Distance/Flow

VolumeVolume/Level/Distance

Page 6: You Can Measure the Benefits… Training Course SmartScan Setup & Configuration.

You Can Measure the Benefits…

For Solid/Liquid application: METER INCH FEETFor Flow application: METER3/HOUR GALLONS Per MINUTE

Main MenuMeasuring Units

Page 7: You Can Measure the Benefits… Training Course SmartScan Setup & Configuration.

You Can Measure the Benefits…

Modes

STORAGE I Slow filling/emptying rate

0.2m/min High accuracy Recommended Where:

Surfaces are wavy

* Not suitable for measuring surfaces with foam

Liquid

Page 8: You Can Measure the Benefits… Training Course SmartScan Setup & Configuration.

You Can Measure the Benefits…

Modes

STORAGE II Twice as fast as STORAGE

I 0.4m/min

Recommended where: Surface conditions are

reasonable

* Not suitable for measuring surfaces with foam

Liquid

Page 9: You Can Measure the Benefits… Training Course SmartScan Setup & Configuration.

You Can Measure the Benefits…

Modes

Process When a reliable reading is

more important than precision 1m/min

Recommended where: Fast reading is required Foamy top surface There is presence of

agitation There is presence of vapor

Liquid

Page 10: You Can Measure the Benefits… Training Course SmartScan Setup & Configuration.

You Can Measure the Benefits…

Main Menu

Includes 10 types of Flume/Weirs, each of them is offered for several sizes.

The measured units can be displayed as either GPM or M3 per Hour.

The methodology used when setting up open channel:X – Particular flume/weir

typeE/U – Flume standardYY- Specific flume/weir

dimensions

Open Channel/Flow

Page 11: You Can Measure the Benefits… Training Course SmartScan Setup & Configuration.

You Can Measure the Benefits…

Flow measurement (SmartScan 50/25 Open Channel)

The default code under PARSH.FLUM mode stands

for custom flume and will lead

you to an additional menu.

Open Channel/Flow

Main Menu

0E.01

Page 12: You Can Measure the Benefits… Training Course SmartScan Setup & Configuration.

You Can Measure the Benefits…

Custom FlumePr. 1 in the additional menu enables you to manually create an accordance table, which can be used for custom flume measurements. The table is used to manually insert up to 24distance/custom flumemeasurement values.

It is recommended to enter a minimum of 8 points.

Open Channel/Flow - Custom Flume

Measuring Flow

Page 13: You Can Measure the Benefits… Training Course SmartScan Setup & Configuration.

You Can Measure the Benefits…

Main Menu

Manual function allows a correction of the measured result, when the sensor is installed either above or below the tank’s cover.

This indication should be added to the measured result.

Manual Function

Page 14: You Can Measure the Benefits… Training Course SmartScan Setup & Configuration.

You Can Measure the Benefits…

The correction value should be entered only in mm. The value may be any figure between: -200mm to

+200mm. Manual function will correct the display result and

current output. It has no influence on the value of data transferred via digital connection.

Note: Manual correction should be entered after the set up of the device.

Main MenuManual Function

Page 15: You Can Measure the Benefits… Training Course SmartScan Setup & Configuration.

You Can Measure the Benefits…

Defining interfering

signals (Scan Distance process)

“Static interference” is a well

known phenomenon in ultrasound

technology.

Use Scan Distance function to

“map” your tank from these

obstacles and to memorize the

interfering signals.

Note: Measurement results appears as Level

Main MenuScan Distance

Page 16: You Can Measure the Benefits… Training Course SmartScan Setup & Configuration.

You Can Measure the Benefits…

SmartScan can locate and store up to 8 obstructions.

After storing 8 interference signals the memory acts as FIFO (first in first out), the eight signal will erase the first signal and so on.

Note:Each obstruction will extend calculation time. Defining interfering signals is

done while the tank is empty.

Main MenuScan Distance

Page 17: You Can Measure the Benefits… Training Course SmartScan Setup & Configuration.

You Can Measure the Benefits…

Main Menu – SmartScan 25

You can set up to five relay switches for single‑sensor versions of SmartScan.

Each relay enables you to define open and close values for the switch, enabling its use for functions such as triggering an alarm.

Note: The values for the relays should be set up as Level

mode.

Relays

Page 18: You Can Measure the Benefits… Training Course SmartScan Setup & Configuration.

You Can Measure the Benefits…

Main Menu – SmartScan 25

The relay values function as follows:

Open value: (Default = 0) The relay opens if the level measured in the tank is higher than the entered open value.

Close value: (Default = 0) The relay closes if the level measured in the tank is lower than the entered close value.

Note:The close value must be lower

than the open value for each relay.

Relays

Page 19: You Can Measure the Benefits… Training Course SmartScan Setup & Configuration.

You Can Measure the Benefits…

Main Menu – SmartScan 50

Relay setup by distance, volume, flow and totalization value (and not with level values only).

Assign an alarm relay for reporting on errors. Assign a relay as totalization pulse output to send a pulse to a

sample activator. The duration of this momentary closure is adjustable to meet different batch sample activators. Each pulse represent a totalization of any value between 1 to

100,000M^3 or GPM The width of the pulse can be any value between 20ms to

2000ms

Relay features

Page 20: You Can Measure the Benefits… Training Course SmartScan Setup & Configuration.

You Can Measure the Benefits…

Main Menu – SmartScan 50

You can set the relays to five different configuration modes: Level, Distance and Flow (where applicable), Volume and Totalization.

The last two modes should be configured only after Volume/Totalization options were enabled in the Additional Menu).

Relays Configuration

Page 21: You Can Measure the Benefits… Training Course SmartScan Setup & Configuration.

You Can Measure the Benefits…

Main Menu – SmartScan 50Relays Setup Options

Following are the relays setup options. Relay values can be viewed in various indication modes according to the selected relay mode. For example, if the selected relay mode is FLOW, then relay values can be viewed in either, Flow, Distance or Level.

Non Flow Application

Flow Application

Indication Mode

Dist.LevelVol.FlowTotal.

Relay Mode

Dist., LevelVol.Flow, Dist., Level

Flow, Total.

Page 22: You Can Measure the Benefits… Training Course SmartScan Setup & Configuration.

You Can Measure the Benefits…

Main Menu – SmartScan 50

You can configure relay 4 to report errors or to remain in normal set-up mode.

Relay 4 – Error Report Configuration

Page 23: You Can Measure the Benefits… Training Course SmartScan Setup & Configuration.

You Can Measure the Benefits…

Main Menu – SmartScan 50

Once relay 4 is configured to report

error, the following messages will be shown in these cases: F.F.F.F” - Whenever the level

reaches up to 10cm inside dead zone or overflow occurred, “F.F.F.F” will appear on the MonoScan’s display.

“E.E.E.E” - Whenever the echo is not received by the transducer because of empty tank or when the measured distance is greater than permitted, “E.E.E.E” will appear on the MonoScan’s display.

Relay 4 – Error Report Configuration

Page 24: You Can Measure the Benefits… Training Course SmartScan Setup & Configuration.

You Can Measure the Benefits…

Main Menu – SmartScan 50

You can choose to set relay number five (5) for flow totalization pulse indication or to remain in normal set-up mode.

Once relay 5 is set for this option, the relay will generate a pulse per Xm^3 or Gallons of flow depend on the value that you have selected from the following list of optional values (you can define the X value):1, 10, 100, 1000, 10000, 100000.

Relay 5 – Totalization Pulse Indication

Page 25: You Can Measure the Benefits… Training Course SmartScan Setup & Configuration.

You Can Measure the Benefits…

Main Menu – SmartScan 50

The electrical pulse will be generated whenever the relay total flow value will be larger than the value selected from the list.

The pulse width varies between 20 to 2000 ms with a resolution of 10 ms.

Relay 5 – Totalization Pulse Indication

Page 26: You Can Measure the Benefits… Training Course SmartScan Setup & Configuration.

You Can Measure the Benefits…

Main Menu

SmartScan enables you to set height, volume or flow values to be used as 20 mA and 4 mA marks.

These values can be used for remote monitoring of tank level, volume or flow using an analog meter

4-20mA

Page 27: You Can Measure the Benefits… Training Course SmartScan Setup & Configuration.

You Can Measure the Benefits…

You can configure 4mA and 20mA parameters to volume values.

The value of volume should be entered in two separated screens for high and low numbers.

Five (5) - enables digits for ‘low’ numbers. Four (4) - enables digits for ‘high’ numbers. for a volume of up to 99,999 configure ‘low’ numbers only and

leave ‘high’ numbers set to zero. For volume above 99,999 (six digits required), the low five

digits should be configured in the ‘low’ screen, and high numbers (starting the sixth digit) should be entered in the ‘high’ screen.

Main MenuConfiguring 4/20mA for Volume

Page 28: You Can Measure the Benefits… Training Course SmartScan Setup & Configuration.

You Can Measure the Benefits…

Programs

Additional Menu

Page 29: You Can Measure the Benefits… Training Course SmartScan Setup & Configuration.

You Can Measure the Benefits…

Entering additional menu

Password window Enter the password code

(716)

Single sensorThan

Additional menu

Additional Menu

Page 30: You Can Measure the Benefits… Training Course SmartScan Setup & Configuration.

You Can Measure the Benefits…

Selecting an Indication

Additional Menu

SmartScan’s display can be modified to show the

following indications: Ind 1: Volume mode. Ind 2: Totalization mode.

(for SmartScan 25O and 50O).

Ind 4: Auto toggle (level, distance, flow, totalization)

Page 31: You Can Measure the Benefits… Training Course SmartScan Setup & Configuration.

You Can Measure the Benefits…

Manually Inserting Strapping Table Values

Additional Menu

The Pr 1 function enables you

to manually create a strapping

table of distance points for either volume calculations

or custom flume

measurements.

Page 32: You Can Measure the Benefits… Training Course SmartScan Setup & Configuration.

You Can Measure the Benefits…

Custom Tank

Measuring Volume

Measuring volume in tanks can bedone in two ways: manual (Pr. 1) semi-automatic. (Pr. 2)In either way two parameters should be entered/programmed accordingly: oneparameter to represent level (distance) and the other to represent volume.For accurate results you may enter upto 24 linear points. If you do not need all 24 points, you may exit at any stage. However, we recommend entering no less than 8 linear points.

Page 33: You Can Measure the Benefits… Training Course SmartScan Setup & Configuration.

You Can Measure the Benefits…

Manual Mode

Measuring Volume

By pressing <NEXT> or <BACK> you can move from one point to another.

Press <ENT> to insert distance and volume value corresponding to the entered distance.

Note:Volume should be entered in two steps: first the number before the decimal point, and then the number after the decimal point.

Page 34: You Can Measure the Benefits… Training Course SmartScan Setup & Configuration.

You Can Measure the Benefits…

Semi-automatic mode

Measuring Volume

This function is based on a “theoretical” calculation, and assumes that you know the corresponding volume for each (critical) distance point.

Page 35: You Can Measure the Benefits… Training Course SmartScan Setup & Configuration.

You Can Measure the Benefits…

Inserting a Coefficient for Readings

The Pr 3 function enables you to enter a coefficient value, K, which can be used for three different options:

Measurement in kilograms Alternative measurement

unit Default volume calculation

For each option, the entered K value is used by SmartScan in the following formula:

X = K (Y) Y- calculated measurement

Additional Menu

Page 36: You Can Measure the Benefits… Training Course SmartScan Setup & Configuration.

You Can Measure the Benefits…

Erasing Strapping Table Values

The Pr 4 function erases all values entered in the strapping table using functions Pr 1 or Pr 2.

Additional Menu

Page 37: You Can Measure the Benefits… Training Course SmartScan Setup & Configuration.

You Can Measure the Benefits…

Defining 22mA/3.7mA Signal Error

Pr. 6 enables you to define whether signal error indications would be active when the current output reaches 22mA or 3.7mA.

F.F.F.F” - Whenever the level reaches up to 10cm inside dead zone or overflow occurred, “F.F.F.F” will appear on the MonoScan’s display.

“E.E.E.E” - Whenever the echo is not received by the transducer because of empty tank or when the measured distance is greater than permitted, “E.E.E.E” will appear on the MonoScan’s display.

Additional Menu

Page 38: You Can Measure the Benefits… Training Course SmartScan Setup & Configuration.

You Can Measure the Benefits…

Defining 22mA/3.7mA Signal Error

To enable 22mA indications, select 0 and <Ent>

To enable 3.7mA indications, select 1 and <Ent>

To disable these indications, select 2 and <Ent>

PR1 - SmartLite

Additional Menu

Page 39: You Can Measure the Benefits… Training Course SmartScan Setup & Configuration.

You Can Measure the Benefits…

Entering Factor for Gas Compensation

Function Pr.07 enables you to compensate for sound velocity changes in different types of gas.

You can enter the appropriate factor for each type of gas listed on the ‘Gas Factor Table (Appen.B in the User’s manual).

PR2 - SmartLite

Additional Menu in SmartScan 25

Page 40: You Can Measure the Benefits… Training Course SmartScan Setup & Configuration.

You Can Measure the Benefits…

Pr.8 Blocking distance shift – smart 25

“Shift on blocking distance” function is recommended in the following circumstances: When the sensor is installed via a pipe, above the tank’s cover.

(In such cases the device defines the pipe’s end as an echo) Once the sensor’s auto-oscillation time is increasing.

(In such cases it causes an inaccurate display of “FFFF”) When there are several interferences signals in the first

measured meters. (In such cases it is recommended to perform initially “shift on” followed by “scan distance”)

PR7 – SmartScan 50 PR3 - SmartLite

Additional menu

Page 41: You Can Measure the Benefits… Training Course SmartScan Setup & Configuration.

You Can Measure the Benefits…

“Shift blocking distance” permitted values:

When the sensor installed via pipe: SBD = h+10cm where: h – Distance from the sensor surface to pipe edge.

Additional menuPr.8 Shift Blocking Distance (SBD)

Extension pipeSBD= h+10cm

Page 42: You Can Measure the Benefits… Training Course SmartScan Setup & Configuration.

You Can Measure the Benefits…

Troubleshooting

Whenever an illegal value is entered, an error code will be displayed.

Wait till the marker on the bar graph changes from 0 to 100%.

Re-enter the desired values. The default is “000000”.

SmartScan 25, 50 & SmartLite

Page 43: You Can Measure the Benefits… Training Course SmartScan Setup & Configuration.

You Can Measure the Benefits…

Error list

User’s Mode: Err. 1 - the entered

value is greater than the Max permitted value.

Err. 2 - the entered value is lower than the Min permitted value.

Err. 3 – the CLOSED value for relay is greater than its OPEN value

SmartScan 25 and 50

Page 44: You Can Measure the Benefits… Training Course SmartScan Setup & Configuration.

You Can Measure the Benefits…

Err4 – 4 mA value is equal to 20mA value

Err5 -The selected function/option is not applicable for the SmartScan model in use.

Err7 – value entered in 4mA , 20mA or relay is greater than the tank height.

SmartScan 25 and 50Error list

Page 45: You Can Measure the Benefits… Training Course SmartScan Setup & Configuration.

You Can Measure the Benefits…

Err8 – The value entered in “ tank height” is greater than the height value of the specific unit.

Err9 - The measurement unit selected is not applicable for the SmartScan model in use.

Err10 - The Open value entered for the relay is greater than the Close value entered for the relay (in Distance mode).

SmartScan 25 and 50Error list

Page 46: You Can Measure the Benefits… Training Course SmartScan Setup & Configuration.

You Can Measure the Benefits…

Err11 - A strapping table was not entered under volume mode or a flume/weir type was not entered under flow mode.

EEEE - The value entered for the tank height is smaller than the actual height, as measured by SmartScan.

AUTO - If displayed at the base of the display screen, this indicates a problem with the current procedure, for example, acoustic interference.

SmartScan 25 and 50Error list