Xray imaging F. Grüner

Transcript of Xrayimaging - DESYgarutti/LECTURES/BioMedical/Lecture8_x-ray_imagin… · film-system / detector...

Page 1: Xrayimaging - DESYgarutti/LECTURES/BioMedical/Lecture8_x-ray_imagin… · film-system / detector • geometrical limitation of radiation (aperture) • anti-scatter grid ideal: focused

X-­‐ray  imaging  

F.  Grüner  

Page 2: Xrayimaging - DESYgarutti/LECTURES/BioMedical/Lecture8_x-ray_imagin… · film-system / detector • geometrical limitation of radiation (aperture) • anti-scatter grid ideal: focused

Röntgen’s  first  “medical  imaging”  

Page 3: Xrayimaging - DESYgarutti/LECTURES/BioMedical/Lecture8_x-ray_imagin… · film-system / detector • geometrical limitation of radiation (aperture) • anti-scatter grid ideal: focused

System-specific image processing

Adaptive auto-windowing algorithm

Improvements….  

Page 4: Xrayimaging - DESYgarutti/LECTURES/BioMedical/Lecture8_x-ray_imagin… · film-system / detector • geometrical limitation of radiation (aperture) • anti-scatter grid ideal: focused

Basic  principles  

X-­‐ray  absorp1on/phase  contrast  imaging  

contrast  

signal  („tumor  is  light  source“)  

X-­‐ray  fluorescence  

Page 5: Xrayimaging - DESYgarutti/LECTURES/BioMedical/Lecture8_x-ray_imagin… · film-system / detector • geometrical limitation of radiation (aperture) • anti-scatter grid ideal: focused

absorpAon  vs  phase  contrast  imaging  

refracAve  index  is  wriAen  as  n  =  1  –  d  +  iß  

phase  contrast  absorp1on  

d  and  ß  as  a  func1on  of  energy  in  keV  for  biological  1ssue    

Page 6: Xrayimaging - DESYgarutti/LECTURES/BioMedical/Lecture8_x-ray_imagin… · film-system / detector • geometrical limitation of radiation (aperture) • anti-scatter grid ideal: focused

absorpAon  vs  phase  contrast  imaging  

different  contrast  generaAon  

absorp1on   phase  contrast  

hAp://www.medphys.ucl.ac.uk/research/acadradphys/researchac1vi1es/pci.htm  

Page 7: Xrayimaging - DESYgarutti/LECTURES/BioMedical/Lecture8_x-ray_imagin… · film-system / detector • geometrical limitation of radiation (aperture) • anti-scatter grid ideal: focused

absorpAon  vs  phase  contrast  imaging  

different  contrast  quality  

absorp1on   phase  contrast  

hAp://www.medphys.ucl.ac.uk/research/acadradphys/researchac1vi1es/pci.htm  

Page 8: Xrayimaging - DESYgarutti/LECTURES/BioMedical/Lecture8_x-ray_imagin… · film-system / detector • geometrical limitation of radiation (aperture) • anti-scatter grid ideal: focused

§   high  resolu1on  voxel  models  of  breast  §   created  from  CT-­‐scans          of  anatomical  breast  specimens  §   voxel  size:  60  x  60  x  60  µm³    §   segmenta1on  in  different  1ssues:        -­‐  adipose        -­‐  glandular        -­‐  skin  

§   using  brilliant  undulator  radia1on:        beam  geometry,  spectral  angular  flux,...  §   simula1on  of  absorp1on  and  scaAering  processes          with  Geant4-­‐SoXware-­‐Toolkit  

absorpAon  imaging:  simulaAons  

Page 9: Xrayimaging - DESYgarutti/LECTURES/BioMedical/Lecture8_x-ray_imagin… · film-system / detector • geometrical limitation of radiation (aperture) • anti-scatter grid ideal: focused

Object detail

Surrounding structures

Balance in imaging

processing

absorpAon  imaging:  contrast  reducAon  

Page 10: Xrayimaging - DESYgarutti/LECTURES/BioMedical/Lecture8_x-ray_imagin… · film-system / detector • geometrical limitation of radiation (aperture) • anti-scatter grid ideal: focused

primary   Rayleigh   Compton  

0.2  mGy  average  glandular  dose  at  ~10¹¹  photons    

absorpAon  imaging:  contrast  reducAon  

Page 11: Xrayimaging - DESYgarutti/LECTURES/BioMedical/Lecture8_x-ray_imagin… · film-system / detector • geometrical limitation of radiation (aperture) • anti-scatter grid ideal: focused

Rasterfocus (focal spot)

Object

X ray film layer

Raster Absorber Blade

Primary radiation

Scattered rad.

Anti-Scatter grid is placed directly on the film /screen-film-system / detector

•  geometrical limitation of radiation (aperture)

•  anti-scatter grid ideal: focused anti-scatter grid

absorpAon  imaging:  contrast  enhancement  

Page 12: Xrayimaging - DESYgarutti/LECTURES/BioMedical/Lecture8_x-ray_imagin… · film-system / detector • geometrical limitation of radiation (aperture) • anti-scatter grid ideal: focused

a) Frequency of examinations b) Collective effective dose from medicine

absorpAon  imaging:  applied  dose  

Page 13: Xrayimaging - DESYgarutti/LECTURES/BioMedical/Lecture8_x-ray_imagin… · film-system / detector • geometrical limitation of radiation (aperture) • anti-scatter grid ideal: focused

analy1cal  model   simulaAon  for  3  different  breast  models  

→  requires  tunable  source  

absorpAon  imaging:  applied  dose  

criterion  =  signal-­‐difference-­‐to-­‐noise  ra1o  versus  averaged-­‐glandular  dose  

Page 14: Xrayimaging - DESYgarutti/LECTURES/BioMedical/Lecture8_x-ray_imagin… · film-system / detector • geometrical limitation of radiation (aperture) • anti-scatter grid ideal: focused

k=693

18

22

26

30

34

38

42

46

50

0% 10% 20% 30% 40% 50% 60% 70%Transmission of Main Absorber p

num

ber

of p

hoto

ns (n

) / 1

06

minimum is always at ca. 20%

transmission through main

absorber

Typically used number of photons in clinical

systems

Together with MAP-AG Florian Gruener

Together with Felix Schöfer, Schöfer et al. in prep.

absorpAon  imaging:  applied  dose  

Page 15: Xrayimaging - DESYgarutti/LECTURES/BioMedical/Lecture8_x-ray_imagin… · film-system / detector • geometrical limitation of radiation (aperture) • anti-scatter grid ideal: focused

Lung perfusion dual energy CT, courtesy of LMU

With especially designed „monoenergetic“ radiation low dose individualised medicine will be developed together with MAP II

Actual: dual energy CT with contrast media for e.g. lung perfusion

è New diagnostic tool

è  at least two scans

è  high dose

è  registration problems

è  non-optimal image information due to overlapping spectra

è Limitations for suitable markers

dual  energy  

Page 16: Xrayimaging - DESYgarutti/LECTURES/BioMedical/Lecture8_x-ray_imagin… · film-system / detector • geometrical limitation of radiation (aperture) • anti-scatter grid ideal: focused

dual  energy  

Page 17: Xrayimaging - DESYgarutti/LECTURES/BioMedical/Lecture8_x-ray_imagin… · film-system / detector • geometrical limitation of radiation (aperture) • anti-scatter grid ideal: focused

•  -­‐  Differen1a1on  between  1ssue  structures  

-   most  notably  between  malign  and          benign  carcinomas  by    

•  different  scaAer  intensity  caused  by  change  in  the  collagen  structure  of  malign  1ssue  

•  different  scaAer  distribu1on      

è DifferenAaAon  between  malign  and  benign  Assue  without  the  need  for  biopsy  would  be  a  great  benefit  

 Image of scatter distribution in tumour

coherent  scaIer  imaging  

Page 18: Xrayimaging - DESYgarutti/LECTURES/BioMedical/Lecture8_x-ray_imagin… · film-system / detector • geometrical limitation of radiation (aperture) • anti-scatter grid ideal: focused

X-­‐ray  fluorescence  imaging  

e.g.  100  keV  incident,    69  keV  Au  line  

Page 19: Xrayimaging - DESYgarutti/LECTURES/BioMedical/Lecture8_x-ray_imagin… · film-system / detector • geometrical limitation of radiation (aperture) • anti-scatter grid ideal: focused

fluorescence  imaging:  experiments  

Page 20: Xrayimaging - DESYgarutti/LECTURES/BioMedical/Lecture8_x-ray_imagin… · film-system / detector • geometrical limitation of radiation (aperture) • anti-scatter grid ideal: focused

fluorescence  imaging:  experiments  

sinogram  

Page 21: Xrayimaging - DESYgarutti/LECTURES/BioMedical/Lecture8_x-ray_imagin… · film-system / detector • geometrical limitation of radiation (aperture) • anti-scatter grid ideal: focused

Compton  

fluorescence  

input  energy  

(simulation by B. Müller, HZM/MAP)

max. sensitivity at 1mGy / 5mm pixel / CNR=5 ~ 10µg/ml ~ 100 times more sensitive than transmission-CT

fluorescence  imaging:  background  

Page 22: Xrayimaging - DESYgarutti/LECTURES/BioMedical/Lecture8_x-ray_imagin… · film-system / detector • geometrical limitation of radiation (aperture) • anti-scatter grid ideal: focused

Physicists’  impacts  

•  new  methods:    •  phase-­‐contrast  with  low-­‐brillance  sources  •  absorp1on/x-­‐ray  fluorescence  with  high-­‐brillance  sources  

 •  pushing  the  limits:    

•  dose  reduc1on  •  lowering  the  concentra1on  of  contrast  media  •  enhancing  the  sensi1vity  •  enforcing  persoanlized  medicine  

•  cooperaAons  between  medicine  and  physics:  •  possibly  compara1ve  studies  between  UKE  and  my  group  

(magne1c  par1cle  imaging  vs  X-­‐ray  fluorescence  imaging  •  new  detectors  for  medicine  (Erika!!)