Minimization of matrix effectswith environmental samples

using ICP-OES

Page 2: Www.spectro.com Minimization of matrix effects with environmental samples using ICP-OES.

www.spectro.com

Agenda

• Introduction

• Matrix effects with environmental applications– Instrumental and analytical approach to avoid/reduce those effects

• Hardware, Optic and introduction systems

• Summary

• Conclusion

2

Page 3: Www.spectro.com Minimization of matrix effects with environmental samples using ICP-OES.

www.spectro.com 3

CONSISTENCY IS KEY!

Know your HARDWARE!

Know your INTRODUCTION SYSTEM SETUP!!

Know the WORST CASE SAMPLE SENARIO!!!

Only then can you OPTIMIZE your methodology….

Page 4: Www.spectro.com Minimization of matrix effects with environmental samples using ICP-OES.

www.spectro.com

Concerns;

• Internal Standard usage• Background Correction Points / Models• Atypical IEC’s (Carbon?)• Sample diversity (drinking waters are not waste waters or brines,

etc.)• SPEED!• Getting the best data WHILE trying to follow a regulated

methodology (EPA)…problems of second and third source standards

4

Page 5: Www.spectro.com Minimization of matrix effects with environmental samples using ICP-OES.

www.spectro.com

Environmental Applications – Challenges

• Emission spectra of environmental samples is typically not extremely line rich and thus environmental work is often described as being “easy”

But

• The matrix composition can vary considerably• The amount of TDS in the final solution can be high vary strongly:

Ca – 3000 mg/l Mg – 500 mg/l

Al – 1000 mg/l Fe – 1000 mg/l

Na – 100 mg/l K – 100 mg/l

P – 1000 mg/l

5

Page 6: Www.spectro.com Minimization of matrix effects with environmental samples using ICP-OES.

www.spectro.com

Axial Plasma Observation – The choice for environmental work

• Axial plasma observation was introduced in the mid 1990s.

• The development driven by requirements for higher sensitivity, particularly for the toxic, heavy metals.

• Today, axial plasma observation still plays an important role as it provides much higher matrix tolerance as compared to ICP-MS.

6

Page 7: Www.spectro.com Minimization of matrix effects with environmental samples using ICP-OES.

www.spectro.com

Axial Plasma Observation – Challenges

• Axial plasma observation improves the sensitivity due to the larger observation volume by sampling the emitted light from the entire excitation channel

• However, it also suffers from stronger influences, since all phenomena present in the excitation zone including the plasma tail above the plasma, are viewed

• The most relevant effects are:– Recombination effect– Influence of the plasma load– Self absorption– Easy ionizable element effect

7

Page 8: Www.spectro.com Minimization of matrix effects with environmental samples using ICP-OES.

www.spectro.com

Recombination effect

• In the cooler temperature regions, within the plasma tail, ions and electrons recombine.

The energy required for their ionization and any excess energy the electrons carry is then again released, which produces additional continuum-like radiation.

8

Page 9: Www.spectro.com Minimization of matrix effects with environmental samples using ICP-OES.

www.spectro.com

Removal/ minimization the recombination effect

• “Removal” of the recombination zone from the optical path

– The OPI to minimizes the effects from the recombination zone by radial

deflection from the light path

Optical light pathto spectrometer

Adjustableargon jet

Recombinationmatrix effects

Plasma

Page 10: Www.spectro.com Minimization of matrix effects with environmental samples using ICP-OES.

www.spectro.com

Removal/ minimization the recombination effect

10

Radial plasma viewing completely

eliminates the effect

Page 11: Www.spectro.com Minimization of matrix effects with environmental samples using ICP-OES.

www.spectro.com

Influence of the plasma load

– With higher plasma loads, the excitation efficiency decreases , which results in lower background to peak rations.

With varying matrices the achievable sensitivity greatly varies. At 1100 W and 10 L/min Ar, the sensitivity difference between a water- and

a 1 % Ca matrix is more than a factor of two

11

1000 1100 1200 1300 1400 1500 1600 1700 1800100

200

300

400

500

600

700

800

1.001.201.401.601.802.002.202.402.602.80

BEC versus Plasma Power - As 189.042 nm

0% Ca

1% Ca

BEC Ratio

Plasma Power [W]

BE

C [

µg/l

]

BE

C-R

atio

Page 12: Www.spectro.com Minimization of matrix effects with environmental samples using ICP-OES.

www.spectro.com

Optimization of the Plasma Power

– Operating of the plasma with robust plasma conditions and appropriate line selection the effect can be minimized

Using robust conditions at 1400 W and 13L/min and As at 193 nm (less influenced by recombination) the sensitivity difference can be reduced to a factor of 1.18

With radial plasma observation the effect is further reduced

12

1000 1100 1200 1300 1400 1500 1600 1700 1800100

200

300

400

500

600

700

800

1.00

1.10

1.20

1.30

1.40

1.50

1.60

1.70

BEC versus Plasma Power - As 193.759 nm

0% Ca

1% Ca

BEC Ratio

Plasma Power [W]

BE

C [

µg/l

]

BE

C-R

atio

Page 13: Www.spectro.com Minimization of matrix effects with environmental samples using ICP-OES.

www.spectro.com

Calibration at 1100 W

13

Page 14: Www.spectro.com Minimization of matrix effects with environmental samples using ICP-OES.

www.spectro.com

Calibration at 1400 W

14

Page 15: Www.spectro.com Minimization of matrix effects with environmental samples using ICP-OES.

www.spectro.com

Self absorption

– Self absorption is the effect where cooler atoms and ions absorb the radiation, they would, under higher temperature conditions, emit

– The effect, reduces the measured light intensity the more of the respective element is present in the sample

Calibration functions become non linear or even reverse, (self reversal of emission lines)

15

Page 16: Www.spectro.com Minimization of matrix effects with environmental samples using ICP-OES.

www.spectro.com

Elimination of self absorption effects

• High dynamic range detector readout 0.1 m sec phase interval to utilize the full potential of the emission line

Does not influence the self absorption behavior, but enables maximum concentration coverage with a minimum number of lines

• Multi line calibration Selection of appropriate lines providing a linear calibration within the

relevant concentration range Automatic, concentration dependent line switching

• Radial plasma observation Less effected by self absorption Further expansion of the linear dynamic range

16

Page 17: Www.spectro.com Minimization of matrix effects with environmental samples using ICP-OES.

www.spectro.com

100 mg/l Na @ 589.592 nm – Axial Plasma Observation

Page 18: Www.spectro.com Minimization of matrix effects with environmental samples using ICP-OES.

www.spectro.com

Multi line calibration - Na @ 589.592 nm and 330.237 using Axial View

18

R2 = 0.9995R2 = 0.9999

Page 19: Www.spectro.com Minimization of matrix effects with environmental samples using ICP-OES.

www.spectro.com 19

Multi line calibration - Na @ 589.592 nm and 330.237 using Radial View

R2 = 0.9997R2 = 0.9999

Page 20: Www.spectro.com Minimization of matrix effects with environmental samples using ICP-OES.

www.spectro.com

Easy ionizable element effect

– Group 1 and 2 elements cause the strongest effect. Those elements are almost exclusively present as ions

– If introduced at “higher” concentrations, this leads to a massive increase of the electron density and thus a shift of the equilibrium

– It often leads to an enhanced emission for atomic lines and respectively higher than normal intensities for the alkali and earth alkali elements

20

Effect of 50 ppm Na on0.2 ppm K using axial plasma observation

Page 21: Www.spectro.com Minimization of matrix effects with environmental samples using ICP-OES.

www.spectro.com

Excellent resolution

21

Higher accuracy in line rich matrices

Al 168 nm

Cd 214 nm

8 pm Resolution 23 pm Resolution

Page 22: Www.spectro.com Minimization of matrix effects with environmental samples using ICP-OES.

www.spectro.com

Ionization buffering and Internal standardization

– The EIE effect can be reduced by providing an excess of easy ionizable

elements in the plasma

– Matrix differences are additionally compensated by the use of an

internal standard• On-line introduction, Sc or Y used as internal standard, Cs used as

ionization bufferPeristalticPump ICP

To nebulizer

From spray chamber drain

Waste

Sample

Buffer/InternalStandard

“orange-orange” peristaltic tubing

“orange-green” peristaltic tubing

Page 23: Www.spectro.com Minimization of matrix effects with environmental samples using ICP-OES.

www.spectro.com

Calibration of Alkali and Earth Alkali Elements

Calibration StandardsElem.

 Std.1[mg/L]

 

Std.2[mg/L]

 

Std.3[mg/L]

 

Std.4[mg/L]

 

Std.5[mg/L]

 

Std.6[mg/L]

 Na 5000 2500 500 100 10 0K 2500 500 100 10 0 5000Mg 500 100 10 0 5000 2500Ca 100 10 0 5000 2500 500Al 10 0 5000 2500 500 100Fe 0 5000 2500 500 100 10Cs 2000 2000 2000 2000 2000 2000Sc 10 10 10 10 10 10

Elem. 

CS7[mg/L]

 

CS8[mg/L]

 

CS9[mg/L]

 

CS10[mg/L]

 

Na 50 100 3.28 2.34K 1 1 50 100Mg 4000 2000 1000 750Ca 750 4000 2000 1000Al 1000 750 4000 2000Fe 2000 1000 750 4000Cs 2000 2000 2000 2000Sc 10 10 10 10

  Axial RadialPower 1450 W 1450 WCoolant flow 12 L/min 12 L/minAuxiliary flow 0.8 L/min 0.8 L/minNebulizer flow 0.95 L/min 0.95 L/minPlasma Torch Quartz, fixed, 2.5 mm

Injector tubeQuartz, fixed, 1.8 mm Injector tube

Spray Chamber Cyclonic CyclonicNebulizer Modified Lichte Modified LichteSample aspiration rate 2.0 mL/min 2.0 mL/minReplicate read time 50 sec per replicate 50 sec per replicate

Plasma Parameters

Check Standards

Page 24: Www.spectro.com Minimization of matrix effects with environmental samples using ICP-OES.

www.spectro.com

Recoveries of Alkali and Earth Alkali elements usingionization buffering and internal standardization

Sample Al Ca Fe K Mg Na  [mg/l] [mg/l] [mg/l] [mg/l] [mg/l] [mg/l]

Meas.            CS7 928 757 2004 1.15 4020 51CS8 721 3934 976 1.06 1985 98CS9 4157 2157 799 51 984 3.65CS10 1943 982 4036 104 758 2.29             Given            CS7 1000 750 2000 1 4000 50CS8 750 4000 1000 1 2000 100CS9 4000 2000 750 50 1000 3.28CS10 2000 1000 4000 100 750 2.34             Recov. [%] [%] [%] [%] [%] [%]

CS7 93% 101% 100% 115% 101% 103%CS8 96% 98% 98% 106% 99% 98%CS9 104% 108% 106% 102% 98% 111%CS10 97% 98% 101% 104% 101% 98%

24

Sample Al Ca Fe K Mg Na

  [mg/l] [mg/l] [mg/l] [mg/l] [mg/l] [mg/l]Meas            CS7 990 741 1938 1.11 3852 47CS8 758 3891 966 1.09 1986 97CS9 3982 2047 730 49 985 3.25CS10 2007 983 3816 97 733 2.19             Given            CS7 1000 750 2000 1 4000 50CS8 750 4000 1000 1 2000 100CS9 4000 2000 750 50 1000 3.28CS10 2000 1000 4000 100 750 2.34             Recov. [%] [%] [%] [%] [%] [%]

CS7 99% 99% 97% 111% 96% 93%CS8 101% 97% 97% 109% 99% 97%CS9 100% 102% 97% 98% 98% 99%CS10 100% 98% 95% 97% 98% 94%

Axial View Radial View

Slightly better average recoveries are achieved in radial mode (99% versus 102%)

Page 25: Www.spectro.com Minimization of matrix effects with environmental samples using ICP-OES.

www.spectro.com

Influence of an Alkali Matrix (50 ppm Na )on Alkali Elements (K) using Axial and Radial Plasma Observation

25

Axial: Strong Effect

Radial: Effect visible,but greatly reduced

Axial

Radial

Depth of field

Viewing volume

Induction areaInduction coil

Viewing volume

Viewing height

Central channel

Induction coil

Page 26: Www.spectro.com Minimization of matrix effects with environmental samples using ICP-OES.

www.spectro.com

Twin Interface Plasma Observation

• Accurate determination of alkali elements in the presence of a varying alkali/earth alkali matrix (e.g. mineral waters) with axial ICP-OES

The EIE (Easy Ionizable Element) effect is greatly reduced since the alkali/earth alkali elements are measured in radial mode

No need for the use of an ionization buffer Cost reduction, reduced risk of contamination

Toxic elements can still be determined with great sensitivity since they are measured in axial mode

Dynamic range and linearity can be further expanded

Page 27: Www.spectro.com Minimization of matrix effects with environmental samples using ICP-OES.

www.spectro.com

Twin Interface (TI) – (Dual View)

27

Page 28: Www.spectro.com Minimization of matrix effects with environmental samples using ICP-OES.

www.spectro.com

SPECTROBLUE Twin Interface - Principle

28

Coupling mirror Optic window

Control mirror

Mirror, plain Mirror, concave

Torch

Entrance slit optic

Radial view

Axial view

Page 29: Www.spectro.com Minimization of matrix effects with environmental samples using ICP-OES.

www.spectro.com

SPECTROBLUE Twin Interface - Design

29

Page 30: Www.spectro.com Minimization of matrix effects with environmental samples using ICP-OES.

www.spectro.com

Twin Interface - Design

30

• Small, light throughput optimized light path

• Periscope optic rigidly connected to the optic and the torch.

• Plasma sided, easy to clean/change, window to avoid the contamination of optical components

• Self adjusting, pneumatic control mirror

• All components easily exchangeable

Page 31: Www.spectro.com Minimization of matrix effects with environmental samples using ICP-OES.

www.spectro.com

New OPI Flange and Locking Mechanism

The OPI can be disassembled without removal of the torch

Novel, easy to use locking mechanism

Page 32: Www.spectro.com Minimization of matrix effects with environmental samples using ICP-OES.

www.spectro.com

Twin Interface - Properties

• Switchable 3-Mirror Periscope• Optimized for smallest size and maximum light throughput using

“Reverse Ray Tracing”• Self adjusting, pneumatic control mirror• Robust construction. Rigidly coupled to the optical system and the

plasma torch• 45°- viewing angle to reduce contamination• Plasma sided protection windows to avoid contamination of optical

components• All components easy to maintain• Separate, computer controlled light path purge flows

32

Page 33: Www.spectro.com Minimization of matrix effects with environmental samples using ICP-OES.

www.spectro.com

Axial Plasma Observation – The choice for trace analytical work

• Axial plasma observation was introduced in the mid 1990s

• The development driven by requirements for higher sensitivity, particularly for the toxic, heavy metals

• Axial plasma observation still plays an important role as it provides much higher matrix tolerance as compared to ICP-MS

33

Page 34: Www.spectro.com Minimization of matrix effects with environmental samples using ICP-OES.

www.spectro.com

Axial Plasma Observation – Challenges

• Axial plasma observation improves the sensitivity by sampling the emitted light from the entire excitation channel

• However, it also suffers from stronger influences, since all phenomena present in the excitation are viewed

• Pros and Cons:– High sensitivity– Stronger matrix effects– Less suitable for high TDS and

applications with organic solutions

34

Viewing volume

Page 35: Www.spectro.com Minimization of matrix effects with environmental samples using ICP-OES.

www.spectro.com

Radial Plasma Observation – The choice for high stability, high matrix loads and organic solutions

• Radial plasma observation provides lower sensitivity since the central channel is only partially view

• However, it provides high stability and freedom from matrix effects since the affected zones in the plasma are blanked out

• Pros and Cons:– High stability– High matrix tolerance– High linear dynamic range – Freedom from matrix effects– Lower sensitivity

Viewing volume

Page 36: Www.spectro.com Minimization of matrix effects with environmental samples using ICP-OES.

www.spectro.com

“Dual View” – really the best of both worlds?

• “Dual View” appears to solve the dilemma having to make a choice between axial and radial plasma observation

But!• Only one view uses the direct light path, the other is compromised

Sensitivity loss due to additional opticalcomponents

Sensitivity loss in the UV since thelight path cannot be purged effectively

36

Coupling mirror Optic window

Mirror, plain Mirror, concave

Torch

Entrance slit optic

Page 37: Www.spectro.com Minimization of matrix effects with environmental samples using ICP-OES.

www.spectro.com

“Dual View” – really the best of both worlds?

• For highest sensitivity, instrument using a “Dual View” technique typically have a horizontal torch orientation Disadvantages with higher matrix loads and organic solutions

• Compared to a dedicated radial interface the sampling volumeusing a periscope is smaller Ultra high stability and precision as compared to a dedicated radial

systems cannot be achieved

The “dual view” technique serves a purpose, but does not provide the highest possible performance

37

Page 38: Www.spectro.com Minimization of matrix effects with environmental samples using ICP-OES.

www.spectro.com

Other hardware options?

Turn the plasma!

• Using MultiView, the plasma orientation can be changed

The instrument can be used as a dedicated radial as well as a dedicated axial instrument

38

Page 39: Www.spectro.com Minimization of matrix effects with environmental samples using ICP-OES.

www.spectro.com

MultiView - Radial/Axial Plasma in one Instrument

39

Analytical performance without compromise

Highest sensitivity Highest stability and precision Highest dynamic range Highest matrix compatibility Freedom from matrix interferences

Other techniques or instruments are not anymore required

Page 40: Www.spectro.com Minimization of matrix effects with environmental samples using ICP-OES.

www.spectro.com

TI Analytical Performance: LODs TI axial mode

40

Line LOD Appl. Report

EOP

TVO-LOD Limit

nm ppb ppb ppbAs 189.042 0.79 0.81 1Sn 189.991 0.26 0.28Tl 190.864 0.58 0.55Se 196.09 0.96 0.98 1Sb 206.833 0.88 0.87 1Zn 213.856 0.04 0.05Cd 214.438 0.04 0.04 0.5Pb 220.351 0.63 0.68 1Ni 231.604 0.17 0.19 2Cr 267.716 0.14 0.15 5Cu 324.778 0.28 0.31Na 589.592 0.29 0.4Li 670.784 0.04 0.04K 766.491 1.50 1.7K 769.896 2.10

• Integration time 60s per replicate• TI-Torch with slit

– No visible wear after 200h of operation

• TI/DV in axial mode achieves the exact same sensitivity as a dedicated axial (EOP) system BUT the radial mode is less sensitive and throughput reduced!

Page 41: Www.spectro.com Minimization of matrix effects with environmental samples using ICP-OES.

www.spectro.com

TI Analytical Performance: Linearity, K 766.491 nm

41

• The linearity is drastically reduced for elements effected by the EIE

• Linearity: R2=0.99• Rel. deviation of the standards up to 25%

• TI Radial: Linearity, R2=0.99998• Rel. deviation of the standards < 5%

Axial mode Radial mode

Page 42: Www.spectro.com Minimization of matrix effects with environmental samples using ICP-OES.

www.spectro.com

Analytical Performance: Linearity radial mode

42

• Linearity: R2=0.99994• Rel. deviation of the standards < 7%

Page 43: Www.spectro.com Minimization of matrix effects with environmental samples using ICP-OES.

www.spectro.com

Summary

• Matrix effects with environmental applications can effectively addressed– Recombination effects can be minimized by a deflection of the

recombination zone (OPI) or eliminated by radial plasma observation– Robust plasma conditions minimize the differences in excitation

properties in the plasma caused by a varying and higher TDS matrix– Self absorption effects can be avoided by multi line calibration and an

appropriate line selection.• A high dynamic range read out helps to reduce the required number of lines• Self absorption effects are drastically reduces by radial plasma observation

– The EIE effect can be well reduced by the use of an ionization buffer– Twin Interface Plasma Observation eliminates the EIE and further

expands the dynamic range

43

Page 44: Www.spectro.com Minimization of matrix effects with environmental samples using ICP-OES.

www.spectro.com

Conclusion

• EOP, the axial interface version, provides high sensitivity and detection limits for superior analysis of industrial and environmental trace elements.

• SOP radial interface version offers precise performance at higher sample concentrations, exhibiting excellent tolerance for high saline and organic fractions plus superb analysis of suspensions and slurries.

• TI twin-interface version automatically performs both axial and radial viewing of the plasma. This “eliminates” the EIE effect, optimizes linearity and dynamic range, while enabling high-sensitivity measurement of toxic elements.

• MV, Multi View interface allows the end user to SWITCH between a DEDICATED Radial AND Axial system (without the drawback of a periscope or complicated entrance optics). This offers the best of both worlds in a systems that is optimized for Axial and Radial viewing, something NO DUAL VIEW system can achieve!

44

Page 45: Www.spectro.com Minimization of matrix effects with environmental samples using ICP-OES.

www.spectro.com 45

Thank You for Your Attention!