WRITTEN RADIATION SAFETY PLAN
Transcript of WRITTEN RADIATION SAFETY PLAN
October 2011
WRITTEN RADIATION SAFETY PLAN
TABLE OF CONTENTS
RADIATION SAFETY OFFICER 1
LOCATION AND RETENTION OF DOCUMENTS 3
PERSONNEL TRAINING POLICY 5
TECHNIQUE CHART 7
WRITTEN RADIATION SAFETY PLAN 8
PROTECTION OF PERSONS IN ROOM 9
MINIMUM EXPOSURE STANDARDS 11
INTRAORAL IMAGE RECEPTOR 12
PATIENT PREGNANCY POLICY 13
MOBILE/PORTABLE EXMAS 13
OPERATOR LOCATION 14
DECLARED EMPLOYEE PREGNANCY 16
MONITORING OF OCCUPATIONAL DOSE 17
REPORTING INCIDENTS/EXPOSURES 21
PRACTICE SPECIFIC PROTOCOLS
Table of Contents
Required Use of PPE 1
Appropriate Use of Dosimeter 2
Material Safety Data Sheets 3
Infection Control Protocols 4
Sample Protocols
Film Storage and Processing
Gendex 765CD Intraoral X-ray
Trophy CCX Digital
Planmeca Intra
Orthomantomography Panorex OP100 Orthoceph
Air Techniques Proverta 70
Note: This section of the Written Radiation Protection Safety Program contains:
• Procedures to minimize exposures to blood, OPIM or chemicals.
• Procedures for operating specific equipment.
• Office specific policies and procedures.
These serve as instructions and guidelines for employees and must be followed as outlined.
Sample protocols are provided as examples. Replace the samples with protocols specific to
instrumentation in use and remove the samples. Use the samples only if they are accurate for your
practice.
January 2012 ©2008 by Total Medical Compliance
1
RADIATION SAFETY OFFICER
The Written Radiation Safety Plan polices, and procedures are reviewed by the Safety
Officer and updated annually and when changes are made:
Name Date Signature
The Radiation Safety Officer (RSO) as required in .0203(b)(2) for our practice is:
Name Date Signature
If the primary Radiation Safety Officer for the practice is located at another office, the
contact for this office/satellite is:
Name Date Signature
The RSO responsibilities include industry monitoring, office performance, daily
duties and equipment safety. The RSO will:
• Monitor radiation safety for our office.
• Report on radiation safety issues and take action to correct the issues.
• Maintain awareness of the radiation safety legislation and keep abreast of the trends in
radiation safety.
The RSO duties on an ongoing basis are to:
• Ensure all personnel are properly trained before operating equipment and on an annual basis
as needed.
• Ensure all personnel utilize appropriate PPE as outlined in the Hazard Assessment
Certification.
• Monitor the safety of the x-ray equipment, premises and protective equipment and devices to
identify that relevant radiation safety standards are in compliance. (for example, compliance
may be affected by damage to, or modification of, the x-ray equipment or premises)
January 2012 ©2008 by Total Medical Compliance
2
• Maintain records of all x-ray equipment and the locations at which it is installed
• Identify ways to minimize radiation doses
• Advise staff on safe working practices
• Ensure regulatory requirements are met for personal monitoring badges for new and existing
employees, devices are properly calibrated, and appropriate reports are maintained and
available for review.
• Investigate radiation incidents.
The RSO will keep management informed:
• Provide reports on radiation incidents and action taken or recommendations for
improvements
• Notify management of any action that needs to be taken to achieve compliance with this plan
or relevant Radiation Safety Standard.
• Provide written assurance that the plan has been reviewed annually as outlined below.
Annually the RSO will ensure completion of the following actions and report the
action to management:
• The equipment inventory list is accurate and all equipment has been accounted for.
• The State registration is current and appropriate. All changes must be reported to the State
immediately.
• All users of the x-ray equipment have been trained and necessary licensure or certifications
are current.
• All maintenance and operational checks have been completed within the stated timeframes
and recorded, and any relevant problems have been appropriately rectified.
• Recommendations for improvement.
The Radiation Safety Officer for the practice has been trained in:
Fundamentals of Radiation Safety
Characteristics of gamma radiation
Hazards of x-ray exposure
Units of radiation dose
Levels of radiation from sources
Methods of controlling radiation dose
Use of personnel monitoring device
January 2012 ©2008 by Total Medical Compliance
3
Location and Retention of Required Documents
Employees can locate the following documents for inspection and review in the
following locations:
The Radiation for Dental Manual is located: __________________________________________
The following documents are located in the Equipment and Facility section:
Plan Reviews [.0603(a)(2)(A)] [.0603(b)]
Letter of Acknowledgement [.0603(a)(2)(A)] [.0603(b)]
Current Notice of Registration [.0603(a)(2)(A)] [.0209]
Post installation Survey as required for dental units utilizing cephalometric equipment, dental
CT, tomography, and i-CAT. Survey is also required when taking radiographs in an open bay
if stated in the Letter of Acknowledgement or during an inspection by the State.
[.0603(a)(2)(B)] [.0603(c)]
FDA 2579 Report of Assembly Form [.0115]
The location of the following documents is listed:
NC Regulation Book [.102(a)(1)] (2002 edition): ____________________________
Notice to Employees [.102(c)]: In the State Regulations tab in this manual and posted in
__________________________________of this facility.
Written Safety Procedures[.0603(a)(1)(D)]: Radiation Safety Program section of this manual
Review of Written Procedures [.1603(c)] [.1636(a)(2)]: Page 1 of Radiation Safety Program
January 2012 ©2008 by Total Medical Compliance
4
Records Retention
Retention schedules are set to consider legal requirements and business requirements.
Record Type Retention Policy
Employee Training
Records
For duration of employment
Patient X-rays Same time as all patient medical records.
In the case of practice ceasing business,
state laws will be followed.
Dosimeter reports Forever
Surveys Latest report will be retained as long as
equipment is in service +5 years
Equipment Maintenance &
Repair Records
Retain for the life of the equipment
Approvals to acquire/
relocate equipment
Permanent
Operational Checklists 6 years
Incident Reports Life of the equipment
Equipment Inventory Permanent
Equipment Registration Most current
Plan Reviews Latest for each piece of equipment used
Letter of
Acknowledgement
Permanent
Post Installation Survey Latest for each piece of equipment
FDA Form 2579 Report of
Assembly
Life of equipment
Notice of Registration Most current for life of equipment
January 2012 ©2008 by Total Medical Compliance
5
PERSONNEL TRAINING POLICY
Certification Requirements-NC
Only individuals certified or otherwise recognized by the State of NC may operate x-ray
equipment.
All RDH, DDS & DMDs are qualified as per current licensure.
A certified dental assistant (one who successfully completes the DANB exam) may take
radiographs without further examination.
Dental assistant who can show evidence of satisfactory performance on an equivalency
examination, recognized by the Board of Dental Examiners, based on seven hours of
instruction in the production and use of dental x rays and an educational program of not
less than seven hours in clinical dental radiology.
Operation of x-ray equipment in this practice must be authorized by the Radiation Safety Officer.
As outlined in [.0603(a)(1)(B))] training must occur PRIOR to operation of equipment.
The following sections of the Radiation Manual must be read by personnel prior to operation of
x-ray equipment:
Introduction
Written Radiation Safety Plan
State Regulations and Guidance
Quality Control Plans
Supplemental Materials
o Infection Control Procedures
o Recycling Plans
Employees Authorized to Operate Equipment
A list of employees authorized to operate the x-ray equipment is located in the training
section of the manual. These employees have met the certification requirements for x-ray
safety training and have been trained on the policies and procedures of this practice.
No one under the age of 18 will be allowed to be involved with radiation services.
Exceptions will be handled by the RSO.
January 2012 ©2008 by Total Medical Compliance
6
Annual Training
Upon hire, all employees will be trained on the policies and equipment of this office. Refresher
training will be conducted at least annually, if the office is remodeled or relocates radiation
equipment or moves to a new facility.
Documentation on training will be maintained for the duration of employment in the training
section of the manual.
A copy of this policy is located in the training section of the manual.
January 2012 ©2008 by Total Medical Compliance
7
TECHNIQUE CHART
The Technique Chart required by [.0603(a)(1)(C)] provides information on all routine
examinations performed by the system and must be used to set up for each routine exam. The
chart gives the mAs or (mAand time) and kVp that must be used for: bitewing, anterior PA,
posterior PA, occlusal, panoramic and cephalometric examinations.
Technique charts are unique to the machine and must be changed when new equipment is
introduced. The technique factors to be used during an exposure should be set before the exposure
begins unless the exposure is automatic by preset buttons.
Types of technique charts utilized – Select ALL that apply
Our wall mounted control has automatic settings that are calculated when the patient size
(large adult, medium adult, small child pictographs) button is activated therefore we do not
have a chart for each patient size.
Instruction to SHIELD EVERY PATIENT will be displayed at the control panel
If settings are being changed, contact your supervisor because a technique
chart will be required.
There is separate chart information for each technique that relates to patient size, kVp, mA,
mAs or exposure time.
The dental unit has a set kV and mA. The technique chart states the time of exposure to be
used in these cases.
Other (Explain) _____________________________________________________
_______________________________________________________________
The information in the chart is provided to help minimize the exposure to the employee and
patient during the examinations to comply with ALARA (the minimum necessary to accomplish
the desired result). The technique is unique to the machine regardless of the operator. The
operator must follow the guidelines on the Technique Chart for each machine at all times.
The chart is posted near the control panel of each x-ray machine, near the control where the
technique is adjusted. Employees must be familiar with the chart and its utilization prior to
operating the equipment. If the Technique Chart is not seen or some recommendation is not
understood, ask the RSO or supervisor before proceeding with the examination.
January 2012 ©2008 by Total Medical Compliance
8
WRITTEN RADIATION PROTECTION SAFETY PROGRAM
The Dental Radiation manual is located _________________________________ and made
available at all times to operators as required by [.0603(a)(1)(D)]. The safe work practices
included are policies of this office and/or are required by regulation.
ALARA is a radiation safety principle for minimizing radiation doses by employing all
reasonable methods. ALARA is not only a sound safety principle, but is required by law for
radiation safety programs.
The following practices contained in this manual will ensure all x-ray equipment operators will
keep their exposures as well as, ancillary staff and patients As Low As Reasonably Achievable
(ALARA).
Maximum Annual Occupation Dose Limits:
5 rems (.05 Sv)), total effective dose to whole body.
15 rems (0.15 Sv) eye dose equivalent to the lens of the eye.
50 rems (0.50 Sv) total organ dose equivalent.
50 rems (0.50 Sv) shallow dose equivalent to skin or any extremity
All equipment operators will take reasonable steps to ensure the radiation dose received by any
person, employee, patient or others is as low as reasonably achievable (ALARA).
Key concepts to minimize radiation exposure include:
1. Shorter exposure times to lower the radiation dose
2. Increasing distance from the source of radiation
3. Shielding such as lead aprons and barriers, to reduce actual exposure.
This Plan is reviewed at initial hire, on an annual basis, and if there is a change to activities or
procedures.
January 2012 ©2008 by Total Medical Compliance
9
PROTECTION OF PERSONS IN ROOM DURING EXPOSURE
Presence during exposures:
Based on [.0603(a)(1)(E)] it is the policy of this office that without approval of the RSO, no one
is allowed in the room with the patient during a radiographic exposure. When holding is required
the policy is to use the parent/caregiver whenever possible. If professional staff and ancillary
personnel are required for the procedure then the listed safety measures are used.
Lead Shielding
Lead aprons or non-lead equivalent required by [.0603(a)(1)(F)] are provided and should be
worn by all patients when x-rays are taken.
Thyroid collars will be used, whenever practicable, with all patients but especially
children.
Lead shielding:
Will be inspected before each use for any cracks or damage and if detected the RSO will
be notified.
Will be stored unfolded to help prevent cracking.
Aprons are located:
In the hygiene operatory (ies)
In the dental exam rooms
In the close vicinity of the pan/ceph unit
In every room that houses an x-ray unit
Other _____________________________________________________________
Safety Measures for Professional Staff and Ancillary Support for Holding Patients
and/or Film During Exposure [.0603(a)(1)(E)(i)&(ii)].
No one individual will be used routinely to hold patients or film.
When holding patients, personal monitoring badges are utilized and will be worn at the
collar outside the apron. [.0603(a)(1)(J)][.1614].
Professional staff and ancillary personnel will be protected from the direct scatter
radiation by protective aprons or whole body protective barriers of not less than 0.25 mm
lead equivalent.
Individuals will be positioned so that no part of the body including the extremities which
is not protected by 0.5 mm lead equivalent will be exposed to the useful beam.
Forceps or a film holder will be used for holding film. [.0603(a)(1)(H)(i)].
January 2012 ©2008 by Total Medical Compliance
10
Non-occupationally exposed staff will not be utilized for holding of patients or films. [.0603(a)(1)(E)(i)(ii)&(iv)].
Safety Measures When Others are Permitted in the Room:
Whenever possible, an adult accompanying the patient will be used for holding. This must be
approved by the RSO. Information on the holder must be recorded in the patient file.
The following instructions will be provided to the human holder [.0603(a)(1)(H)(ii)&(iii)]:
If a parent or another person is required to hold the patient, they must be at least 18 years
old and must not be pregnant. [.0603(a)(1)(H)(iv)].
Wear a 0.25 mm lead apron.
Stand as far as possible from the useful beam. Under no circumstances must any person
other than the patient be exposed to the direct beam.
Ordering of Radiographic Exams
X-rays including retakes of patients may be taken only upon the order of the dentist as identified
in [.0603(a)(1)(G)(ii)].
January 2012 ©2008 by Total Medical Compliance
11
MINIMUM EXPOSURE STANDARDS
Minimum Necessary Exposure is the Standard
This practice uses equipment and procedures in all cases to reduce patient and employee exposure
to the minimum necessary.
The radiation exposure time to the patient will be the minimum exposure required to produce
images of valuable diagnostic quality. Exposure techniques listed on Technique Charts or
preprogramed on certain units must not be adjusted to compensate for inadequate film processing.
In order to meet the requirements set forth in [.0603(a)(1)(I)] the following procedures/equipment
is utilized to minimize radiation exposure to patients and personnel:
Film speed is the fastest speed consistent with the diagnostic objective of the exam. Film
must be “E” speed or faster. [.0603(a)(1)(I)(i]
For non-digital panoramic or cephalometric machines, screen and film combinations are
the fastest speed consistent with the diagnostic objective of the exam. Film must be “E”
speed or faster. [.0603(a)(1)(I)(i]
For digital units, the exposure index will be monitored to ensure it is within the range for a
particular body part x-rayed as established by the equipment manufacturer.
[.0603(a)(1)(I)(ii)]
Other: ___________________________________________________________
_________________________________________________________________
_________________________________________________________________
All employees will report conditions to the RSO that compromise the concept of minimum
exposure. Examples would be:
• Slower film than necessary is being used when a better speed is possible.
• Equipment operating conditions that require extra exposure time to accomplish the task due
to a malfunction or less than maximum operating performance.
January 2012 ©2008 by Total Medical Compliance
12
INTRAORAL IMAGE RECEPTOR SAFETY
Intraoral images are not performed in this practice.
The following information on Intraoral Image Receptor Safety applies to this office.
Radiographic systems designed for use with an intraoral image receptor will be provided with
means to limit the x-ray beam such that:
• If the source-skin distance (SSD) is 18 centimeters or more, the x-ray field at the SSD will be
containable in a circle having a diameter of no more than 7 centimeters (2.75 inches); and
• If the SSD is less than 18 centimeters, the x-ray field at the SSD will be containable in a
circle having a diameter of no more than six centimeters (2.4 inches).
Automatic Protection Equipment on Machines
X-Ray machines are equipped with devices to limit the radiation exposure to patients and
employees. These devices include filters that reduce unnecessary low-energy radiation from the
primary beam (PBL- positive beam limiting) and collimators which restrict the size of the x-ray
beam.
No operator is to alter, remove, or tamper with these devices. Any interference in these
protective measures could lead to needless radiation exposure to patients and/or employees.
X-Ray Beam Restriction & Alignment
Use the cone provided on the x-ray machine.
Never take a patient x-ray without a cone on the tube head.
Shielded, lead lined, cylinder, open-ended cones will replace short, black plastic cones.
Measurements have shown a scatter reduction to the patient of up to 75 percent by
changing to the shielded cone.
The beam size may not be larger than the image receptor on stand-alone cephlametric
equipment.
Tube Housing
The tube housing must not drift or move during any exposure as this could result in radiation
exposure to unintended areas. No one is allowed, under any circumstances, to hold the tube
housing or support housing during any exposure.
If a problem exists with the stability of the support arm, notify the RSO and suspend use of the
equipment immediately. There is the possibility of electric shock from improper grounding if the
machine is hand-held.
Place a sign on the equipment that it is OUT OF ORDER to ensure no other person will attempt
to use the equipment until it has been evaluated or repaired.
January 2012 ©2008 by Total Medical Compliance
13
PATIENT PREGNANCY POLICY
A number of studies have suggested that the embryo/fetus may be more sensitive to ionizing
radiation than adults. This is especially true for the first trimester of gestation.
In order to ensure the safety of the pregnant patient this practice will:
Post signs prominently instructing patients to inform the technologist if they are pregnant
or might be pregnant.
Question female patients about the possibility of pregnancy prior to radiation exposure
being sensitive to the privacy rights and feelings of the patient.
If the patient is pregnant or thinks they may be pregnant notify the doctor.
The x-ray procedure will be delayed until the end of the pregnancy if possible.
If necessary to proceed or if the patient chooses to proceed, shield the patient with a 0.5 mm
lead apron or double shield with the regular aprons.
Obtain a signed waiver from the patient if the decision is made to proceed with the x-ray.
MOBILE/PORTABLE EXAMS
Mobile or portable exams are not performed in this practice.
The following information on mobile devices applies to this office.
Use of Mobile or Portable Machines is used only for examinations where it is
impractical for medical reasons to transfer the patient to a stationary radiographic
installation. [.0603(a)(1)(I)(iii)]
During the exposure using a mobile or portable x-ray device the x-ray machine operator:
1. Must be positioned so that his/her exposure to scatter radiation is as low as reasonably
achievable (ALARA) (e.g. 6 feet or more away)
2. Should never be in line with the direct beam.
3. Any person must be able to leave the room unless it is an Operating Room and then all
personnel must be badged and shielded. [.0603(a)(1)(E)(iii)]
January 2012 ©2008 by Total Medical Compliance
14
OPERATOR LOCATION WHILE TAKING THE X-RAY
The employee must not be in the room or subject to scatter radiation due to their position.
Additionally the employee taking the x-ray must be able to control the area so that others will not
be exposed to scatter radiation during the process.
This practice performs the following radiographic studies which requires visual
contact of the patient
Panorex
Cephalometric
CT
Tomography
i-CAT
Visual contact as required in [.0604(b)(1)(C)] is maintained through:
Use of a window of lead equivalent glass.
Use of a mirror system.
Location of operator as required in [.0607(e)(2)(A)].
The operator is required to be in the protected area where the exposure switch is permanently
mounted.
The protected area(s) is located:
________________________________________________________________________
Visual Indicators and Audible Signal [.0607(e)(3)].
The exposure switch of our x-ray system provides a means to terminate an exposure at
any time if the exposure exceeds the devices set time +/- 15%. The switch is located in
the protected area.
Employees must remain in the protected area where access to the switch is possible
during an exposure. When the exposure has been completed the employee will:
o Hear an audible signal
o See the indication that the exposure is complete.
Each operator will see the visual indicator and hear the audible signal before leaving the
protected area following an exposure.
January 2012 ©2008 by Total Medical Compliance
15
In order to keep exposures as low as reasonably achievable (ALARA) as outlined in
[.1603(b)], employees and this practice will protect themselves, and the patient by:
Operator standing behind the protective barrier.
Visually control the area. The operator will verify hallways and other areas are clear in
order to protect others from scatter radiation.
All appropriate doors are closed and appropriate shields are used during x-ray operation
as documented in the operating procedures and Plan Review.
Cone positioning is adjusted to ensure that no beam is aimed at a non-shielded area.
Use of lead aprons and thyroid collars when possible on patients for all exposures.
Appropriate construction of the walls, floor and ceiling areas exposed to the useful beam
in accordance with regulatory requirements to provide appropriate protection to all
employees taking x-rays.
All equipment is maintained in good operating order. Maintenance and calibration
records are stored behind the Equipment and Facility tab for the lifetime of the
equipment.
Other ____________________________________________________________
_________________________________________________________________
_________________________________________________________________
Additional measures taken in order to ensure that members of the general public do
not receive a total effective dose that exceeds 0.1 rem (1mSv) in a year as outlined in
[.1611(a)(1)] and that the dose in any unrestricted area of this practice will not
exceed 0.002 rem (0.02 mSv) in any hour [.1611(a)(2)] :
No member of the general public is allowed in controlled areas.
As outlined in [.1604(a)(1)] the Maximum Annual Occupation Dose Limits:
5 rems (.05 Sv)), total effective dose to whole body.
15 rems (0.15 Sv) eye dose equivalent to the lens of the eye.
50 rems (0.50 Sv) total organ does equivalent.
50 rems (0.50 Sv) shallow dose equivalent to skin or any extremity
January 2012 ©2008 by Total Medical Compliance
16
VOLUNTARY DECLARED EMPLOYEE PREGNANCY POLICY
For the declared pregnant worker, this practice has the responsibility to take appropriate action to
ensure the dose to a fetus due to occupational exposure [.1610]:
Does not exceed 0.5 rem (5 mSv or 500 millirems) during the entire pregnancy
Does not exceed more than 0.5 mSv in any month
The dose to the fetus is measured as the sum of:
The deep-dose equivalent to the declared pregnant woman and
The dose to the embryo from radionuclides in the embryo and radionuclides in the
pregnant woman.
Verbal notification is not an acceptable form of notification. Declaration of pregnancy must be
submitted in writing and must include the following information:
Pregnant employee’s name.
Estimated due date or date of conception.
The date the declaration was signed.
Upon receipt of written notification of pregnancy the following apply:
Fetal Badging:
The use of a fetal badge is required at any time the potential for radiation exposure
exists. The badge will be worn at waist level of the pregnant worker.
Declared pregnant worker:
This practice has performed baseline testing which has documented workers are not
likely to receive, in one year, in excess of 10 percent of the annual limits which is
0.05 rem (.5mSv) therefore personal badging is not required.
The use of a personal badge at any time there is the likelihood the pregnant women
would receive a deep dose equivalent in excess of 0.1 rem (1mSv) as required in
[.1614(1)(c)] is required.
Information will be provided to the employee on the dangers of fetal exposure to
radiation, as well as techniques and procedures to minimize exposure to the fetus.
Job reassignment or other actions depending on the potential for exposure may be
considered. The employee may also request reassignment of duties if desired.
Exposure history will be reviewed.
Record Keeping [.1640(f)]
The practice will maintain the employee declaration of pregnancy records of dose to an
embryo/fetus with the records if indicated of dose to the declared pregnant woman for 30 years
past the last date of employment or until the practice license or registration is terminated by The
North Carolina Department of Health and Human Services, Radiation Protection Section,
whichever is longer.
January 2012 ©2008 by Total Medical Compliance
17
MONITORING OF OCCUPATIONAL DOSE
Radiation Monitoring for Employees
Personnel monitoring devices to monitor radiation dose are required for workers using dental x-
ray machines if an employee is likely to receive 10 percent of the annual limit allowed by
regulation. If previous baseline radiation monitoring records show that it is unlikely that a person
will receive a dose of this amount, then ongoing personal monitoring is not necessary except in
outlined situations.
For the safety of employees and to meet regulations outlined in [.1614], this office
will provide personal monitoring badges for workers as outlined below:
Personal monitoring badging is not required. This practice has performed baseline testing
which has documented workers are not likely to receive, in one year, in excess of 10
percent of the annual limits which is 0.05 rem (.5mSv).
This practice will perform baseline testing to document workers are not likely to receive, in
one year, in excess of 10 percent of the annual limits which is 0.05 rem (.5mSv).
• All employees will wear the badges for ________________________ (time) even
when operating only digital equipment.
Badges will be required for an additional ______________________(time) period if new
equipment is installed or if the office layout is changed.
Any employee required to hold patients must wear a badge.
Any employee must wear a badge if instructed to do so by the RSO.
Other _________________________________________________________
Personal Monitoring Devices Placement and Recommendations:
Follow manufacturer’s guidelines for use.
Whole body monitoring devices are placed at the neck level or upper torso. If a protective
apron is worn the monitoring device is placed at collar level outside of the protective
apron.
Fetal monitoring devices are placed at the waist and always on the outside of any
protective apron.
Badges are issued to a specific employee and record the personal exposure of that
employee. Do not share your badge with another employee it will distort both records.
Remove badge when receiving personal radiographic studies.
Changing of Personal Monitoring Badges
The schedule for badge changing is set by the RSO.
Badges will be changed: _____________ (must be at least every 3 months).
This practice will change badges on the first working day of the month.
January 2012 ©2008 by Total Medical Compliance
18
If any employee in the practice reaches the maximum annual allowed dose of 5 mSv, then
all workers will change badges monthly until the cause for the high reading has been
resolved.
Fetal monitoring badges will be changed on a monthly basis.
Notify the RSO if badge is lost or misplaced.
Storage of Personal Monitoring Badges
Store in a safe place, at work away from sources of radiation and heat.
Do not store badge clipped to lab coat or jacket as someone could accidentally pick up
your badge or it could fall off of the jacket and become lost, or laundered accidentally.
Storage location for this office: __________________________________________
Control badge monitoring – A control personal monitoring device must be stored in an area
where only background radiation will be measured. The control badge will not be used for area
monitoring or assigned to a person.
Storage location for this office: ____________________________________________.
The monitoring badge service provider is ___________________________________.
The provider is accredited by the National Voluntary Laboratory Accreditation Program
(NVLAP) of the National Institute of Standards and Technology (NIST).
When the defined issuing period has passed, all badges, including the control, must be returned to
the service provider for measurement. Any badges that have not been issued, or if issued have not
been worn, must also be returned to the service provider.
Report of Individual Monitoring Results
Individual reports of dose levels reported by the dosimeter will be reviewed by the RSO and with
employees on a quarterly basis when badging is in process. These reports will be retained in the
employees file and are available for review by the employee upon request. The employee will
initial each report to acknowledge the review process.
Reports of Excessive Radiation Exposure
In this practice the RSO will investigate if a report is received that indicates any employee badge
has a reading that exceeds the following ALARA investigational limits.
Quarterly Investigational Limits for Monthly Wear Dates
125 mRem for whole body badges
30 mRem for fetal badges
January 2012 ©2008 by Total Medical Compliance
19
Each employee must be aware of their own accumulated dose as registered on their badge.
Accumulated exposure, even from a prior employer, must be considered in the overall exposure
limit.
If at any time the employee for any reason (accident, etc.) believes they have received an
exposure that exceeds the limit the employee must notify the RSO immediately.
Reassignment of duties may be necessary if dose limits are met or exceeded for any reason.
Prior Records for Occupational Dose for New Worker [.1638(a)(1)&(2)],
This practice has performed baseline testing which has documented workers are not
likely to receive, in one year, in excess of 10 percent of the annual limits which is 0.05
rem (.5mSv).
This practice will attempt to obtain previous occupational dose history for the current year
from the previous employer(s) on all new employees and will attempt to obtain the
records of lifetime cumulative occupational radiation dose.
This practice will accept any of the following:
• Signed statement from the individual, or from the individual's most recent employer
for work involving radiation exposure, that discloses the nature and the amount of
any occupational dose that the individual may have received during the current
year;
• Accept, as the record of lifetime cumulative radiation dose, an up-to-date agency
form for recording occupational radiation dose history, or equivalent, signed by the
individual and counter-signed by an appropriate official of the most recent
employer for work involving radiation exposure, or the individual's current
employer if the individual is not employed by this practice (example – staffing
service).
• Reports of the individual's dose equivalent(s) obtained by telephone, telegram,
electronic media, or letter from the most recent employer for work involving
radiation exposure, or the individual's current employer if the individual is not
employed by this practice (example – staffing service). This practice will request a
written verification of the dose data if the authenticity of the transmitted report
cannot be established.
Working for Multiple Employers
It is the employee’s responsibility to provide dose information to each employer from the other
place of employment. Employees working for multiple employers with the possibility of radiation
exposure should talk with the RSO about the best method of ensuring the total radiation dose is
recorded.
January 2012 ©2008 by Total Medical Compliance
20
Information in Personnel and Exposure Records: Records will be maintained for 30 years past the last date of employment or until the
practice license or registration is terminated by The North Carolina Department of Health
and Human Services, Radiation Protection Section, whichever is longer
[.1640(a)(1)&(g)] and will include:
Full name, sex and date of birth.
Home address.
The date of commencement of employment and the last date of employment.
Description of work performed by the employee.
Details of the types of ionizing radiation to which the employee may have been exposed
in the course of their employment.
Details must be provided of any radiation accidents in which the employee has been
involved or by which they may have been affected.
Details of the personal monitoring badge worn by the employee. This may include the
type of monitor, where on the body the monitor was worn and the name of the monitoring
service provider.
Radiation exposure dose results for the employee.
Radiation exposure dose for the fetus.
Radiation Exposure Records for Departing Employees
This practice will provide the following:
A copy of their radiation exposure records. This will include all annual dose records
and any subsequent periodic reports received after the employee has ceased
employment if available.
An additional copy of the radiation exposure records will be provided to the new
employer if requested by the employee. These records are required to be given to a
new employer so that an assessment can be made of possible future doses that can be received to keep an employee under their annual limit.
21
January 2012 ©2008 by Total Medical Compliance
REPORTING RADIATION INCIDENTS AND EXPOSURES IN NC
A radiation incident is an incident adversely affecting, or likely to adversely affect, the health or
safety of any person because of the emission of radiation and must be reported to the RSO. The
most likely radiation incident at this practice is an actual or suspected malfunction of the x-ray
equipment.
In this case of equipment malfunction, the following procedure must be observed:
The user must switch off the x-ray unit as quickly as possible at the main supply.
The user must take precautions to prevent the use of the x-ray unit by:
1. Posting a sign which states that the x-ray unit must not be used.
2. Removing the key to the x-ray unit or taking other appropriate action.
The x-ray unit must not be used until it has been repaired (and if necessary compliance
tested) and the possession licensee authorizes its use.
The RSO must determine the radiation dose to each person involved in the incident.
Overexposure Incidents
If a patient may have received radiation in excess of the normal x-ray or a member of the general
public may have received radiation in excess of the dose limits, notify the RSO immediately. This
office is required to report to the State Radiation Protection area any individual exposure in
excess of the limit set by regulation. The timeframe for reporting of overexposures by this
practice depends on the exposure the patient received.
Records Retention
Unless otherwise indicated, all reports will be retained for 3 years.
The following incidents will be reported by the RSO immediately as outlined in
[.1646]:
Any event involving a source of radiation possessed by this practice that may have caused or
threatens to cause any of the following conditions:
An individual to receive:
o a total effective dose equivalent of 25 rems (0.25 Sv) or more
o an eye dose equivalent of 75 rems (0.75 Sv) or more
o a shallow-dose equivalent to the skin or extremities of 250 rads (2.5 Gy) or more
Loss of one working week or more of the operation of any facilities affected
Damage to property in excess of $200,000.
Reports required immediately should be called to:
The Division of Radiation Protection
(919) 571-4141
8:00 a.m. to 5:30 p.m. on workdays
22
January 2012 ©2008 by Total Medical Compliance
The following incidents must be reported by the RSO within 24 hours as required in
[.1646]:
Any event involving a source of radiation possessed by this practice that may have caused or
threatens to cause any of the following conditions:
An individual to receive, in a period of 24 hours
o a total effective dose equivalent exceeding five rems (0.05 Sv)
o an eye dose equivalent exceeding 15 rems (0.15 Sv)
o a shallow-dose equivalent to the skin or extremities exceeding 50 rems (0.5 Sv);
Loss of one day or more of the operation of any facilities affected
Damage to property in excess of $2,000
Reports required within 24 hours should be called to:
The Division of Radiation Protection
(919) 571-4141
8:00 a.m. to 5:30 p.m. on workdays
A written report will be submitted by the RSO within 30 days as required in [.1647]
for the following incidents:
Any incidents which were reported by phone immediately or within a 24 hour timeframe
as required by [.1646]
For the dose limits that exceed:
o Occupational dose limit for adults [.1604], and minors[.1609]
o Limits for an embryo/fetus[.1610)]
o Limits for a patient or general public[.1611]
o The ALARA constraints for air emissions[.1603]
The report using the Radiation and Exposure Report form will outline the extent of exposure of
the individual to radiation and will include the following [.1647(b)]:
Estimates of each individual's dose.
Levels of radiation involved.
Cause of the elevated exposures, dose rates, or concentrations.
Corrective steps taken or planned to ensure against a recurrence, including the schedule
for achieving conformance with applicable limits as outlined in ALARA.
For each occupationally overexposed individual, the report as required by [.1647(c)] will be
prepared so that the following information is stated in a separate and detachable part of the
report and will include
Name
Social security account number
Date of birth
With respect to the limit for the embryo/fetus, the identifying information shall be that of
the declared pregnant woman.
23
January 2012 ©2008 by Total Medical Compliance
A copy of the completed report will be provided as directed in [.1647(d)] and [.0111] to:
The North Carolina Department of Health and Human Services
Radiation Protection Section
1645 Mail Service Center
Raleigh, North Carolina 27699-1645
Or delivered to the agency at its office located at:
3825 Barrett Drive
Raleigh, North Carolina 27609-7221
The practice will provide a report to the exposed individual no later than when the report is
provided to The North Carolina Department of Health and Human Services, Radiation Protection
Section as required in [.1647(e)]
Radiation exposure reports will be maintained for 30 years past event date or until the practice
license or registration is terminated by The North Carolina Department of Health and Human
Services, Radiation Protection Section, whichever is longer.
PRACTICE SPECIFIC PROTOCOLS
Table of Contents
Required Use of PPE 1
Appropriate Use of Dosimeter 2
Material Safety Data Sheets 3
Infection Control Protocols 4
Sample Protocols
Film Storage and Processing
Gendex 765CD Intraoral X-ray
Trophy CCX Digital
Planmeca Intra
Orthomantomography Panorex OP100 Orthoceph
Air Techniques Proverta 70
Note: This section of the Written Radiation Protection Safety Program contains:
• Procedures to minimize exposures to blood, OPIM or chemicals.
• Procedures for operating specific equipment.
• Office specific policies and procedures.
These serve as instructions and guidelines for employees and must be followed as outlined.
Sample protocols are provided as examples. Replace the samples with protocols specific to
instrumentation in use and remove the samples. Use the samples only if they are accurate for your
practice.
1
REQUIRED USE OF PERSONAL PROTECTIVE EQUIPMENT
Personal protective equipment (PPE) such as gloves, eye protection, masks and gowns are used
by this office for infection control purposes.
As part of new employee and annual training, the appropriate use of PPE, is reviewed with all
employees involved in radiology services.
Training includes:
Indication for use of PPE
Appropriate donning and doffing of PPE
Appropriate disposal of PPE
The appropriate PPE for use with each procedure is indicated on:
The Hazard Assessment Certification which is located at the end of this section.
Other (Indicate where the information is recorded and located)
_______________________________________________________________
2
APPROPRIATE USE OF DOSIMETER
The following are reminders of the appropriate use of the dosimeter. The RSO will provide clear
instruction when dosimeters are to be utilized outside of the following situations:
Baseline monitoring of radiation exposure in the practice
Declared pregnancy
When holding patient/film during exposure
• DO WEAR IT when working. It has no value in your locker or purse.
• DO TURN IT IN on time. Time gaps make analysis more difficult, less accurate and reduces
legal and historical value of the reports.
• DO PLACE the control dosimeter in a radiation-safe area; the dose to the control is subtracted
from each dosimeter and needs to be accurate.
• DO REPORT LOST OR DAMAGED dosimeters immediately. Prevent damage by not
leaving your dosimeter in areas of high temperature such as your dashboard or in the clothes
dryer.
• DON’T WEAR IT when you are receiving x-rays for your own health care.
• DON’T WEAR IT away from the workplace. Leave your dosimeter in the same place every
day when you leave the office so you know where it is.
• DON’T WEAR IT under your apron unless you are wearing two dosimeters, one at the neck
level outside the apron and one under the apron. This applies to pregnant workers.
• DON’T PLACE a dosimeter in an area for testing of stray radiation. Additional dosimeters
can be assigned for testing.
• DON’T SHARE dosimeters. An average total for a shared dosimeter is meaningless to each
individual.
• DON’T TAMPER with your dosimeter or anyone else’s. The reports are legal documents and
are regarded as real exposures received.
3
MATERIAL SAFETY DATA SHEETS This practice recognized the employees “right to know” about chemicals in which they may come
in contact with during the performance of their work responsibilities.
As part of new employee orientation, on an annual basis, and when new chemicals are added
employees are trained on appropriate use, spill response, and appropriate disposal of chemical
utilized in the X-ray process.
Employees will follow manufacturers recommendation for use of all chemicals associated with
the X-ray process.
Material Safety Data Sheets (MSDS) are available at all times to employees.
MSDS are accessed via the internet at _________________________________________
In the event internet access is unavailable employees may find the needed MSDS by
________________________________________________________________________
MSDS are available in hard copy form in ______________________________________
4
INFECTION CONTROL PROTOCOLS
Preparation before taking x-rays
Utilize appropriate PPE as outlined in the Hazard Assessment
Protect radiography equipment (radiograph tube head and control panel) with surface
barriers that are changed after each patient.
Aseptically dispense film to be used into a clean disposable container and transport to the
treatment area.
Protect the patient by using a lead apron with a thyroid collar even when performing
digital studies.
Panoramic/Extraoral x-rays
Wash hands.
Handle extraoral cassettes with ungloved hands.
The bite holder required on some panoramic x-ray units (or any intraoral film holder)
should be autoclaved if possible.
Disposable or barrier protected bite holders may be used instead. If a barrier is used have
the patient remove the barrier and place it in the trash.
Decontaminate bite holder based on manufacturer’s recommendation.
Digital Studies
Utilize barrier device as listed by equipment manufacturer to reduce gross contamination.
Clean and disinfect device with EPA approved hospital disinfectant as outlined by
manufacturer.
Taking x-rays:
Wear gloves when taking intraoral radiographs. Use eyewear, mask and gowns if spatter
is likely.
Stay behind the lead shield.
Dry the film with disposable gauze or paper towel to remove blood or excess saliva.
Place film in a disposable container, such as a plastic or paper cup.
Using Film-Holders
Use film-holders that have been autoclaved or sterilized with a cold sterilant if they
cannot take the heat of an autoclave.
Assemble the film-holder for specific region to be viewed/exposed.
Insert the correct film size and insert into block with ‘dot in the slot.’
Position the film-holder in the mouth.
Use cotton roll as needed for comfort of patient.
Slide aiming ring down the indicator rod to near the skin surface.
Align BID/PID flat against the aiming ring.
Select correct exposure settings on machine and make the exposure.
Gently remove the film-holder.
Place exposed film in a place where it will not be accidently mistaken for unexposed
films.
5
After completing the x-rays:
If film barrier pouches are used, remove the film packets from the pouch and allow them
to drop into a clean container for transport to the developing area. Take care to avoid
contamination of the outside packet and container. If barrier pouches are not used,
transport the film to the developing without further handling in the patient area.
Remove and discard contaminated barriers. Discard other contaminated disposable items.
Remove and discard gloves and wash hands.
After each patient, clean and disinfect any equipment, not protected by barriers, that was
touched by gloved hands or contaminated film packets. Use an EPA-registered
disinfectant.
Put away reusable items such as film-holding devices and lead apron.
Transport film to the developing area with ungloved hands.
Developing film:
Use barriers to protect equipment or clean and disinfect surfaces that become
contaminated.
To process film without barrier pouches:
o Don gloves.
o Open each film and drop it out of its wrapping onto a flat surface covered by a
paper towel. Discard lead foil into container for recycling.
o After opening all of the exposed film, dispose of container and packets.
o Remove and disposed of contaminated gloves. Wash hands.
o Place the film into the processor slot or film holder for developing.
When an automatic film processor with daylight loader is used, contamination of the
fabric shield is likely a problem. There is no practical way to disinfect this material. The
following procedures will prevent contamination.
o With clean, ungloved hands, open the lid. Place a paper towel, paper cup and
powder-free gloves inside the loader’s compartment.
o Place the container with the film next to the cup.
o Place ungloved hands through the light shield and don gloves. Unwrap the film
packet and drop the film onto paper towel or processor film feed slot.
o Place the film wrapping into the second cup. Remove gloves, turn them inside
out and place them into the cup.
o Place the film in the chute for processing handling only by the edges.
o Remove hands from the loader and wash hands.
o Lift the lid and dispose of the paper cups and waste.
FILM STORAGE AND PROCESSING This section does not apply to facilities with digital imaging and no “wet chemistry” processing
capability or Endodontic facilities that use “hot” chemistry at the chairside.
Manufacturer’s instructions for processing and storing film are located in the Equipment and
Facility section of this manual for reference.
Film Storage Unexposed x-ray film must be stored in accordance with the manufacturer’s instructions. The
manufacturer of our film is __________________________________.
Store daily use supply of film in a container in each operatory or
_____________________________________________________________.
Larger quantities, more than a month’s supply, are stored
______________________________________________________________.
Film is stored in light-tight containers with shielding to prevent fogging of
undeveloped film from stray radiation. If boxes are lead-lined be sure the lead has
been treated.
Do not store in untreated lead-lined boxes. Some older wooden boxes, usually the
size of a shoebox, used untreated lead. The FDA recommends discontinuing their use
immediately.
The area of storage is not near excessive heat of any type or excessive humidity.
Film may be frozen for up to two years usually (check the manufacturer’s
instructions.) Thaw at room temperature at least 24 hours before using.
The film container is not stored where chemical contamination might occur.
Film processing chemicals are stored in a manner to protect the film.
The film storage area is far enough away from the X-Ray unit to protect it from
ionizing radiation. (Do not store in the direct beam of the x-ray tube) and is protected
by ______________________________________________.
Do not use expired film. Remove expired film and check with the Safety Officer for recycling or disposal instructions.
Film Processing Tips
Chemicals: • The manufacturer of the processing fluid is
____________________________________________________.
• Test processing fluid before use by: ____________________________________ as specified
in the manufacturer’s instructions.
• Replace processing fluid every __________________________________________ as
specified in the manufacturer’s instructions.
• Processing fluid should not be used past the expiration date on the container.
• Do not use premix chemistry that is more than 30 days old.
• Do not allow processing fluid to freeze.
Lighting: • Lighting in the film processing area should be retained at:
• Filter type on the safe light ___________________
• Bulb wattage in the safe light _________________
• Distance from work surface for mounting of safe light _____________ feet.
• If you see leaks around the door or other openings in the darkroom notify the RSO.
The person responsible for replacing the fluid (chemistry) and cleaning the processor:
Name Date Name Date
Film Processing – Manual
Processing tanks are mechanically rigid and corrosion resistant.
The temperature of the solution is adequate will be verified using a non Mercury
thermometer.
Temperature guides will be followed as indicated by the film manufacturer’s chart
located in the Equipment section of the manual and ______________________.
Temperatures must be between 60F and 80F (16C to 27C)
The recommended developing time for that temperature as shown on the time-
temperature developing charts will be utilized.
The temperature will be checked at the beginning of the day and throughout the day each
time film is developed.
Sight developing of radiographs is not acceptable for processing films.
Caution:
o When film holders and pan cassettes are used it is important that each be marked
so it is identifiable by the exposure operator if it is possible for a 2nd film
cassette to be placed in developer or fixer at the same time. Failure to mark them
could result in removing the wrong one and possibly ruining both.
Film Processing – Automatic
The temperature of the developer will be verified to ensure it is at the level recommended
by the manufacturer.
The temperature will be checked at the beginning of the day and throughout the day. The
temperature should be at ___________.
The accuracy of the automatic temperature control will be verified periodically.
Automatic developers monitor and adjust the temperature only and do not monitor the
chemistry condition.
Clean-up or blank films will be run daily or more often as specified by the manufacturer.
Processor will be maintained according to the instructions in the manufacturer’s
operation manual. The manufacturer’s manual is located in the Manufacturer’s section of
the Radiation Manual or ____________________________________.
Film Processing Records
Records of the weekly processor testing will not be retained, but the weekly films will be
retained until the chemistry is changed and the process begins again.
The processor chemistry change and darkroom maintenance log sheet will be posted in
the film processing area.
GENDEX 765DC INTRAORAL X-RAY
1. Switch the power switch to the “On” position. The Ready indicator lamp on the front of the
control panel will light.
2. Select or verify that the unit is set for either film or digital imaging. The indicator light should
be lit next to the appropriate graphic icon. Press the button between the two icons to toggle to
the correct selection.
3. Select or verify that the system is set for the appropriate patient selection (i.e. an adult or
child). Press the button between the two icons to toggle to the correct selection. The
technique factors to be used during an exposure should be set before the exposure begins
except when the exposure is automatic. In this case note the settings.
4. Press the selector button in the center of the control to toggle the indicator lamp until it lights
next to the tooth (or bitewing) symbol for the area that you wish to x-ray.
Note: The exposure time defaults (the times displayed when the system is initially turned on)
can be modified per the instructions in the Configuring Control Settings.
Setting exposure times and types before placement of film holders in the patient’s mouth
increases the chances of a better exposure. The least amount of time the film is in the mouth,
the better the odds the patient will not move to blur your image.
5. Position the tubehead to the patient using standard accepted positioning procedures.
Note: Care should be taken to not place fingers in areas where they may be potentially
pinched during the movement of the arm. Also, be careful to not have the tubehead hit the
wall after returning the arm to the storage position. Restrict the beam to the area of clinical
interest. The beam should not be larger than the image receptor. The method used for
restricting the beam is collimation or _________________________________.
6. Prior to making the exposure, verify or modify the time settings as necessary using the
secondary tubehead controls. Note: The unit is supplied with an 8” (20cm) cone. An optional
longer 12” (30cm) cone can be used to sharpen the image as when the paralleling technique is
used. A rectangular cone is available as an option to reduce the x-ray field size to that of the
film and reduce the radiation to the patient. Each cone must have a separate technique chart.
Notify your supervisor or the RSO if it is not available.
7. Optional coil-cord exposure switch is not allowed in North Carolina or South Carolina. If the
equipment comes with this option it must be disconnected even though you do not use it.
TROPHY CCX DIGITAL
1. Switch the power switch to the “On” position. The switch is on the control panel. The
power indicator light will come on.
2. Verify that the unit is set for film. The film speed indicator should be 1. Adjust as
necessary.
3. Select or verify that the system is set for the appropriate patient selection (i.e. an adult or
child) and tooth or bite wings. Press the button to make this selection.
4. The exposure time defaults can be modified to accommodate a large adult if necessary.
Always consult the dentist before changing the exposures. You will adjust this exposure
time by increasing the film speed selector number.
5. Drape the patient with a lead apron.
6. Place the phosphorus plate in the XCP device and position it in the patient’s mouth.
7. Position the tubehead to the patient using standard accepted positioning procedures. Note:
Care should be taken to not place fingers in areas where they may be potentially pinched
during the movement of the arm. Also, be careful to not have the tubehead hit the wall
after returning the arm to the storage position. Restrict the beam to the area of clinical
interest.
8. Step outside room to the control panel. Press the exposure buttons on the left and right
simultaneously. Watch the exposure move to o. You will hear a beep and see the X-ray
indicator light showing the x-ray was completed. Release the buttons. The machine will
automatically stop when the proper exposure time has passed. However if you release the
buttons early you will see the operator error light and hear a beeping noise.
9. Retrieve the plate form the patient’ mouth.
10. Turn the machine off.
PLANMECA INTRA
1. Turn the equipment on. The switch is in the patient room. Switch is on the base where the
arm is attached to the wall.
2. Go to the control panel and verify that the ready indicator light is lit.
3. Verify the unit is set on Plate mode. Plate mode will show P in the Exposure time
display.
4. Select or verify that the system is set for the appropriate patient selection (i.e. an adult or
child).
5. Select the appropriate X-ray area. You may have to press the button multiple times to get
to the proper area (upper, lower, etc.).
6. Drape the patient with a lead apron.
7. Place the phosphorus plate in the XCP device and position it in the patient’s mouth.
8. Position the tubehead to the patient using standard accepted positioning procedures. Note:
Care should be taken to not place fingers in areas where they may be potentially pinched
during the movement of the arm. Also, be careful to not have the tubehead hit the wall
after returning the arm to the storage position. Restrict the beam to the area of clinical
interest.
9. Return to the control panel. Press the buttons in the middle and the one on the left side to
activate the X-Ray. The Exposure warning indicator light will flash and you will hear a
beep during the exposure. Both will stop on completion. If you press only one button the
equipment will not activate.
10. Retrieve the plate form the patient’ mouth.
11. Turn the machine off.
ORTHOMANTOMOGRAPH PANOREX OP100 ORTHOCEPH
1. Turn the equipment on. The switch is under the base.
2. To set the unit to manual mode first press the down key once to move the flashing light
from standard panoramic to the AEC (A) position. Then press the right over key once to
move the flashing light to Manual (M) position.
3. At this time the light over the juvenile of the programmed exposure should be lit. To
change the program exposure first press the down key two times until the flashing light is
over the patient size symbol.
4. Now press side arrow keys to select the appropriate size ( child, juvenile, adult, large
adult)
5. Bring the patient to the machine.
6. Drape the patient.
7. Position patient appropriately.
8. On the side panel, lower right, press the down button. This will position the plate.
9. Step into the hall and push exposure button and hold. Watch your patient in the mirror.
10. Have the patient release the bite block.
AIR TECHNIQUES PROVERTA 70
1. Switch the power to the “On” position. The time in seconds will appear on the front of
the control panel.
2. Select or verify that the unit is set for either film or digital imaging. The indicator light
should be lit next to the appropriate graphic icon. Press the button between the two icons
to toggle to the correct selection.
3. Select or verify that the system is set for the appropriate patient selection (i.e. adult or
child). Press the button between the two icons to toggle to the correct selection. The
technique factors to be used during an exposure should be set before the exposure except
when the exposure is automatic. In this case note the settings.
4. Press the selector button in the center of the control to toggle the indicator lamp until it
lights next to the tooth (or bitewing) symbol for the area you wish to X-ray.
Note: The exposure time defaults (the times displayed when the system is initially turned
on) can be modified per the instructions in the Configuring Control Settings.
Setting exposure times and types before placement of film holders in the patient’s mouth
increases the chances of a better exposure. The least amount of time the film is in the
mouth, the better the odds the patient will not move to blur the image.
5. Position the tubehead to the patient using standard accepted positioning procedures.
Take care to not place fingers in areas where they may be pinched during the movement
of the arm. Also be careful to not have the tubehead hit the wall after returning the arm
to the storage position. Restrict the beam to the area of clinical interest.
6. Prior to making the exposure, verify or modify the time settings as necessary using the
secondary tubehead controls.
7. Optional coil-cord exposure switch is not allowed in North Carolina. If the equipment
comes with this option it must be disconnected even though you do not use it.
1
DOCUMENTATION REQUIREMENTS
Shielding Design/Plan Review - Prior to construction of structural modifications, the
floor plans and equipment arrangement of all installations utilizing x-rays for diagnostic
or therapeutic purposes must be reviewed by a qualified expert. The practice will submit
the recommendations to The North Carolina Department of Environment and Natural
Resources, Division of Radiation Protection. [.0603(a)(2)(A)] [.0306(b)]
Letter of Acknowledgement - Written acknowledgement of receipt by North Carolina
Department of Environment and Natural Resources, Division of Radiation Protection
received PRIOR to installation, of any facility and shielding design.
State Registration Forms – Required by the "North Carolina Regulations for Protection
Against Radiation" rules .0203, each registrant will complete proper application to
register radiation machines and facilities within 30 days following initial operation of that
facility and each radiation machine. The registration of the first radiation machine at a
facility constitutes registration of the facility itself.
Application: Registration: Radiation Machines: Facilities
Application: Registration of Service
Post Installation Survey – Required for dental units utilizing cephalometric equipment
or taking radiographs in an open bay which must be performed within 30 days following
the initial operation of each radiation machine. This survey must be performed by a
person approved through The North Carolina Department of Environment and Natural
Resources, Division of Radiation Protection [.0603(a)(2)(B)] [.0603(c)]
Current Notice of Registration – Most current registration as outlined in
[.0603(a)(2)(A)] The registrant shall notify the North Carolina Department of
Environment and Natural Resources, Division of Radiation Protection in writing when
any change will render the information contained in the application for registration or
notice of registration no longer accurate. [.0209].
Maintenance and Calibration Reports - Each registrant will keep records of
maintenance or modifications which affect the useful beam, along with the names of
persons performing the service. [.0603(a)(2)(C)]
Inventory - Listing of all X-ray, development and support equipment essential to the
operation of X-ray units.
FDA Form 2579 - Report of Assembly - Manufacturers of diagnostic x-ray systems
intended for human use are required to file reports of assembly upon installation of a
certifiable system or component(s). The report of assembly represents the assemblers
certification that the system or component(s) are of the type called for by the Standard
(i.e., certified), have been assembled according to the instructions provided by the
manufacturer, and meets the requirements of the applicable Federal standards contained
in 21 CFR 1020.30 through 1020.33. Reports must be filed with FDA's Center for
Devices and Radiological Health (CDRH) within 15 days of completion of the assembly.
Copies must also be provided to State agencies and the purchaser.
QUALITY CONTROL
Table of Contents
QUALITY ASSURANCE PLAN FOR FILM RADIOLOGY
OPERATIONAL CHECKLISTS 1
VENDOR QUALITY ASSURANCE TESTING 3
QUARTERLY CHECKLIST – FILM 4
DAILY CONSISTENCY TESTING 6
PROCESSOR CHEMISTRY TEST 7
DARKROOM QC – SCREEN-FILM UNITS 10
SCREEN FILM UNITS
INTRAORAL
QUALITY ASSURANCE PLAN FOR DIGITAL RADIOLOGY
OPERATIONAL CHECKLISTS 13
VENDOR QUALITY ASSURANCE TESTING 15
QUARTERLY CHECKLIST – DIGITAL 16
LEAD APRONS 18
RETAKES AND REPEATS 19
Revised October 2011 1
QUALITY ASSURANCE PLAN FOR FILM RADIOGRAPHY
Operational Checklists
Forms are available in the Forms section of the manual to use for recording routine
operational checks. Indicate the QC functions performed by this office
Daily
Confirm the processing and intraoral x-ray unit is operating in a consistent
manner every day. Check the quality of a test film (e.g. first clinical image of the
day) by comparing it with a reference radiograph. (Consistency Test).
Run blank or clean-up films.
Verify temperature of processing chemicals each morning.
Check developer, rinse and fixer levels.
Clean processor feed tray and counter tops.
Clean darkroom.
Ensure visual and audible indication of exposure is functional.
Monthly
Perform a retake/reject analysis to determine the reason for the retake or reject
(which may include problems with the film, developing process, the X-ray unit
or the user).
Check processing box for light leaks.
Clean the viewboxes with glass cleaner. Replace bulbs and tubes.
Check the condition of personal protective equipment.
Technique Charts - exposure factors for specific examinations are readily
available.
Processor maintenance procedures are displayed in a prominent location
adjacent to the film processor.
Instructions for mixing chemicals and processing films are available.
Time/temperature chart, a timer and a thermometer are available for manual
processing.
Radiation Caution signs are properly displayed.
Visibly inspect lead aprons and shields for rips or tears.
Revised October 2011 2
Quarterly or Every Six month
X-ray equipment warning signs are displayed on each control panel and are in
good condition.
Film is within the expiration date specified by the manufacturer.
X-ray film is stored in accordance with this plan.
Oldest x-ray film is used first.
Verify accuracy of the information on the technique charts.
Cephalometric BLD insert in place.
Panoramic tube and film holder working properly.
Cables and wires meet OSHA requirements and are not in direct beam.
Cassettes stored properly.
Equipment allows employees to be shielded properly – Exposure switch
placement and patient can be viewed (except intra-oral).
Cone secured.
Film artifacts not present on processed film.
Control panel without operational issues.
Annual
Notice to Employees Poster is current and displayed.
Radiation Plan has been reviewed, updated and signed.
Employees have been trained and documentation of training recorded.
All operators have the appropriate and current certification/training.
Review incident reports for the year to determine if a trend exists and to define
any additional corrective action is needed to avoid future incidents.
Check lead aprons and shields to ensure adequate protection is provided. Aprons
must be x-rayed to detect problems. Suppliers can check aprons when inspecting
equipment.
Revised October 2011 3
Equipment Quality Assurance Testing to be conducted by vendors:
Adherence to the approved shielding plan, if applicable.
Half-value layer, exposure reproducibility.
mA/mAs linearity.
Integrity of pass through interlocks.
kVp accuracy.
Timer reproducibility and accuracy.
X-ray beam size.
Proper indication of multiple tubes on units so equipped.
Minimum source to skin distance.
Patient exposure at skin entrance (bitewing and/or periapicals).
Mechanical support of tubehead.
Visual and audible indication of exposure.
Exposure reproducibility.
Revised October 2011 4
QUARTERLY CHECKLIST ITEMS – Film Radiography
Each of the items below should pass the visual check or corrective action should be
taken. Use the Quarterly Inspection Record form to note any action taken.
Some of the items on the visual checklist are operator convenience features. However,
many of the items are essential for patient safety and high quality diagnostic images. It
may be necessary to add additional items to the list that are specific to particular
equipment or procedures. These should be included on the checklist and in each
evaluation.
1. Cephalometric beam limiting device (BLD) insert in place and used. Determine if the
unit has built-in collimation or if manual beam limiting devices are being used. Assure
that the collimators are functioning correctly and smoothly. Determine that manual
beam limiting devices are sufficient for blocking the beam. Assure that collimators or
beam limiting devices are being utilized correctly for the area being radiographed.
2. Panoramic tube and film holder motion smoothness. Determine that during motion the
tube and film holder moves smoothly without catches or interruptions.
3. Cone or cylinder secured to tube. Visually inspect the connection between the tube
and the cone or cylinder. Determine the connection is secure and stable. Could cause
retakes. Operators may not support tube in any way.
4. Tube or generator oil leakage. Visually inspect areas around tubes and generators for
oil or abnormal collections of dust attaching to oil leaks. Equipment problems and a
risk of slips.
5. Condition of cables and wires. All cables and wires should be inspected for frayed
coverings, kinks, and should be free from catching on other objects. Must meet OSHA
standards for safety. Cords should be outside of area. X-ray can damage cords.
6. Film artifacts. Examine exposed processed films closely to determine if artifacts are
visible. Density variations, small specks from dust particles, fingerprints, water marks,
and other artifacts should not be visible.
7. Cassettes and screens condition. Cassettes and screens should be cleaned regularly.
Check screen condition for dust particles, scratches, and areas of discoloration. Assure
screens are properly fitted and attached to cassettes (if applicable). Check cassette
latches to make sure they are functioning properly and are not broken. Cassettes and
screens should be replaced if necessary.
8. Film storage. Determine that film is stored in a cool, dry place, and is protected from
exposure to stray radiation. Assure that film boxes that have been opened are stored in
a light-tight container. Determine that film being stored has not expired; check the
expiration dates.
9. Loaded cassette storage. Determine that loaded cassettes are stored in an area that is
properly shielded from radiation to prevent exposure. They should be stored off the
ground and kept free from dust.
Revised October 2011 5
10. Control panel indicators. Assure all control panel switches, lights, and meters are
functioning correctly.
11. Technique chart. Make sure technique charts are available, current, and appropriate for
all procedures normally performed.
12. Equipment allows operators to remain behind barriers during X-rays as required.
Patient view ability. Check for the availability of a method to permit continuous
observation of the patients during the x-ray exposures, except with intraoral films.
Exposure switch placement. Assure the exposure switches are mounted or
exposure switch cables are long enough that operators are not exposed to
unnecessary levels of scatter radiation during x-rays.
Revised October 2011 6
QC - DAILY CONSISTENCY TESTING
A reference radiograph is an image that exhibits the desired qualities of contrast,
density and sharpness. Radiographs should be compared to the reference radiograph.
Changes in appearance may indicate problems with equipment, exposure or processing.
The reference radiograph should last for the life of the processor.
Establish a reference radiograph for daily comparison use. Be sure to use new solutions
of developing films to establish the reference radiograph.
1. Place the Dental Radiographic Normalizing and Monitoring Device on a flat
surface and insert unexposed intraoral film under the copper square.
2. Center the cone on the copper square and touching device.
3. Use posterior bitewing technique factors and expose the film.
4. Process the film as normal.
5. Insert the dried test film into the film slot on the test tool. Slide the test film until
one of the density steps match the test film density.
6. If the test film is above or below the mid-density steps (Steps 3-5), adjust the x-
ray timer settings and repeat the steps to establish the reference radiograph.
7. Record the technique factors used and the density setting matched by the film.
Conduct a daily test prior to the first patient x-ray and record the results on the daily QC
form. The daily test should be within one step of the first test film.
Resolving Problems
If problems are encountered, refer to the troubleshooting guide for the device. If
problems persist call the service engineer for the equipment.
If you are using extraoral screen-film and are doing daily sensitometry to monitor the
film processor, it may not be necessary to complete the daily consistency test. This
process is recommended by the Conference of Radiation Control Program Directors,
Inc. (CRCPD) www.crcpd.org.
Revised October 2011 7
QC – PROCESSOR CHEMISTRY TEST METHOD
Processor Chemistry Test Method
Test processor chemistry activity at least once a week using a density comparison
technique, such as the Dental Radiographic Quality Control Device (Crabtree)®* or a
density step wedge such as the 76-025-4000 Dental Aluminum Step Wedge from
Cardinal Health/Nuclear Associates**.
The DRQCD ® is a density comparison tool that is constructed with a copper strip to
create a density on a dental intra oral film. The film is inserted into the pocket on the
device behind the copper plate and exposed to a typical bitewing x-ray technique.
Create a "master" film when the chemistry is fresh.
A new "master" is created each time the chemistry is changed. This master is used
to compare with the weekly test film.
Insert the processed "master" into the slot above the density comparison scale and
identify the "step" closest to the film density by pulling or pushing the slide back
and forth.
Once you find the matching density, label the "master" with that number (usually
3, 4,or 5).
The subsequent processed image is compared to the density scale to determine
whether there are changes in the density or contrast due to changes in the
chemistry when compared to the "master". Lighter films could indicate developer
that is getting exhausted and needs to be changed or replenished or could indicate
that the developer temperature is low.
By testing at least once per week, you may be able to extend the chemistry change
time and save resources by not changing as frequently.
You may find that the automatic replenishment system (if the unit has one) is not
functioning properly and you need to hand replenish the developer or adjust the
replenishment rate.
Films that are cloudy may indicate that the fixer solution has become
contaminated and needs to be changed or replenished or that the wash water is not
turned on.
If the difference between the "master" film created when the chemistry was fresh
and the weekly comparison film is one step or more when comparing the same
step, the chemistry needs to be changed or replenished before patient films are
processed.
DO NOT CHANGE THE TECHNIQUE TO MAKE THE TEST FILM
MATCH THE MASTER! If you use the new technique on patients, all you
are doing is increasing the radiation exposure. Find out what the problem
with the processor is before processing patient films.
Revised October 2011 8
Large film processors (ceph and pan)
If a large film processor is used for ceph or pan films, the 11 step aluminum
wedge, such as the Cardinal Health/Nuclear Associates Model 76-025-4000, may
be used with a flat cassette such as a cephalometric cassette. If the cassette has a
screen block for the nasal bones, be sure to place the wedge on the portion of the
cassette where the screen is not blocked.
If only a curved cassette or flexible film holder for a pan unit is available, the test
cannot be performed and you will have to rely on the processor manufacturer
instruction for frequency of chemistry changes.
If a flat cassette is available, expose the wedge on the cassette to a technique of
about 70 kV at about one second or 60 pulses. This can be done with the cassette
in the ceph frame by taping the wedge to the cassette. This should produce a film
that shows all 11 steps on the wedge. The darker steps (thinner portion of the
wedge) may be difficult to discern.
Label the first film on fresh chemistry as the "master" and retain this film.
Compare all other films to the "master".
If the density difference between the "master" and the test film is more than one
step, change the chemistry or replenish the chemistry if the temperature is correct.
The test film should be exposed using the same x-ray machine and same
technique each time. This will provide more consistent results and prevent
unnecessary chemistry changes.
Hand processing
• Proper processing time is critical with hand tanks and a timer is needed in the
darkroom.
• A stem-type thermometer should be in the developer tank to monitor the temperature.
Mercury thermometers must never be used in film processors.
• Process the film for the length of time required by the developer temperature.
• A time/temperature chart should be posted in the darkroom.
Alternative Methods
Alternative test methods, such as anatomical phantoms, may be used for consistency
testing so long as they are capable of producing images of sufficient contrast variation to
tell when the chemistry needs attention.
Step wedges may also be hand-made by using dental lead foil from film packets. Stagger
a group of four lead foils to create steps of different thicknesses to create a wedge shape.
*A Dental Radiographic Quality Control Device© (Crabtree) device is a tool used to
compare the density of dental films exposed using the device to a standard film strip
which comes with the device. A copper plate in the device creates a density when the
film is exposed to x-ray. The films taken at various times of the month can be compared
to see if the chemistry is becoming exhausted and needs to be replaced.
Other density comparison devices, such as an aluminum step wedge, may be used as
well. The wedge is placed on the film and exposed to the same technique from the same
tube head each time. The films are processed and compared with the master film created
Revised October 2011 9
when the processor chemistry is first replaced. The master film is stored and used for the
weekly comparison. A master film needs to be made each time the chemistry is changed.
Contact your supplier for more information.
Mention of a device, product or service does not constitute an endorsement of a product
or manufacturer. It serves only as a representation of the types of devices, products or
services which are available.
Revised October 2011 10
QC - DARKROOM
General Housekeeping
Clean processor feed tray and counter top.
Mop the floor of the darkroom, clean counters and cabinets to remove any dust.
Clean film hangers.
Developer Temperature Control
Manual Processing:
Measure the developer temperature daily with a non-mercury
thermometer.
Compare the temperature with the time-temperature for the equipment.
Replenish the fluid following the manufacturer’s guidelines for manual
processing and replace rinse water.
Automatic Processing
Measure the temperature to confirm it is within the range established by
the manufacturer and that the display is accurate.
If completing the daily consistency of film test, daily temperature
measurement may not be necessary.
Minimize Darkroom Fog – Screen-Film Units
A suggested standard by CRCPD is an optical density increase of 0.05 or less which
should be conducted with each type of film used on a semiannual or annual basis.
Supplies Needed:
Manila file folder
Watch
Densitometer
Attenuation block (aluminum step wedge, phantom, acrylic block) to create a
medium optical density of about 1.0 on the film.
Procedure:
1. Load a cassette with film and place on a flat surface.
2. Center the attenuation block and expose the film using an x-ray technique that
will result in an optical density of about 1.0 after the film is processed.
3. With the safelights on, place the exposed film on the work area in the darkroom.
Cover half the film with opaque material, bisecting the latent image parallel to the
long axis of the film.
4. Leave exposed film on the counter for 2 minutes, then process as usual.
Revised October 2011 11
5. While waiting 2 minutes for darkroom fog test, look for any sources of extraneous
light. Any light leaks identified should be repaired as soon as possible.
6. Inspect the processed film. If there is no discernible delineation between the
shielded and unshielded sides of the film, there is no fog problem.
7. If a line is evident, measure the optical densities of both sides of the line with the
densitometer. If the density difference is greater than 0.05, corrective action
should be taken.
Corrective Action if Necessary
Repeat the test with the safelight off.
If the results remain the same, the problem may be caused by a light leak or
extraneous light.
If the fog level disappears, the fog was due to the safelight and remedial action
must be taken to correct the problem.
Possible sources of fog
Safelight filters (old or compromised)
Safelight housing
Safelight too close to work area
Light bulb of incorrect wattage or type
Ancillary indicator lights on processor
Timers
Radios
Fluorescent light afterglow
Light leaks
Suspended ceilings
Any place there is a hole cut in the wall
Excessive ambient light through the tinted viewing windows of daylight loading
systems
Minimize Darkroom Fog – Intraoral (without densitometer)
The goal is to maintain no visible difference in density between the shielded and
unshielded sides of the fog test film. With no defined density differences present the use
of a densitometer is unnecessary.
Conduct this test semiannually, for each type of film used, after bulb or filter replacement
and after changing or adding types of film.
Supplies Needed:
Manila file folder
Watch
Attenuation block (aluminum step wedge, phantom, acrylic block) to create a
medium optical density of about 1.0 on the film.
Revised October 2011 12
Procedure:
1. Expose the film so that the background density will produce a medium optical
density.
2. With the safelights on, remove exposed film from its light tight covering and
place the exposed film on the work area in the darkroom. Cover half the film with
opaque material, bisecting the latent image parallel to the long axis of the film.
For a daylight processor, place film under the tinted viewing window in the glove
box.
3. Leave exposed film on the counter for 2 minutes, then process as usual.
4. While waiting 2 minutes for darkroom fog test, look for any sources of extraneous
light; any light leaks identified should be repaired as soon as possible.
5. Inspect the processed film. If there is no discernible delineation between the
shielded and unshielded sides of the film, there is no fog problem.
6. If a line is evident, repeat the test to determine the source of darkroom fog. A test
film, which shows no delineation between the shielded and unshielded sides, must
be obtained to assure the source of the darkroom fog has been resolved.
Corrective action if Necessary
Repeat the test with the safelight off. If the results remain the same, a light leak or
extraneous light may be the cause of the problem. If the fog level disappears, the fog was
due to the safelight and remedial action must be taken to correct the problem.
Possible sources of fog
Safelight filters (old or compromised)
Safelight housing
Safelight too close to work area
Light bulb of incorrect wattage or type
Ancillary indicator lights on processor
Timers
Radios
Fluorescent light afterglow
Light leaks
Suspended ceilings
Any place there is a hole cut in the wall.
Excessive ambient light through the tinted viewing windows of daylight loading systems
Revised October 2011 13
QUALITY ASSURANCE PLAN FOR DIGITAL RADIOGRAPHY
Operational Checklists
Forms are available in the Forms section of the manual to use for recording routine
operational checks. Indicate the QC functions performed by this office.
Daily
Confirm the intraoral x-ray unit is operating in a consistent manner every day.
Check the quality of a test digital image (e.g. first clinical image of the day) by
comparing it with a reference radiograph. (Consistency Test)
Visual and audible indication of exposure is functional
Monthly
Perform a retake/reject analysis to determine the reason for the retake or reject
(which may include problems with the X-ray unit, software or the user)
Inspect the condition of personal protective equipment
Technique Charts – Ensure exposure factors for specific examinations are
readily available
Radiation Caution signs are properly displayed.
Visibly inspect lead aprons and shields for rips or tears.
Quarterly or Every Six month
X-ray equipment warning signs are displayed on each control panel and are in
good condition.
Technique Charts - Verify accuracy of the information on the technique charts.
Cephalometric BLD insert in place
Panoramic tube and sensor holder are working properly.
Cables and wires meet OSHA requirements and are not in direct beam.
Cassettes stored properly.
Equipment allows employees to be shielded properly – Exposure switch
placement and patient can be viewed except intra-oral.
Cone secured.
Control panel without operational issues.
Annual
Notice to Employees Poster is current and displayed
Radiation Plan has been reviewed, updated and signed
Employees have been trained and documentation of training recorded.
Revised October 2011 14
Verify all operators have the appropriate and current certification/training.
Review all incident reports for the year to determine if a trend exists and to
define any additional corrective action is needed to avoid future incidents.
Check lead aprons and shields to ensure adequate protection is provided. Aprons
must be x-rayed to detect problems. Suppliers can check aprons when inspecting
equipment.
Document verification of PPE.
Revised October 2011 15
Equipment Quality Assurance Testing to be conducted by vendors:
Adherence to the approved shielding plan, if applicable.
Half-value layer, exposure reproducibility.
mA/mAs linearity.
Integrity of pass through interlocks.
kVp accuracy.
Timer reproducibility and accuracy.
Minimum source to skin distance.
X-ray beam size
Proper indication of multiple tubes on units so equipped.
Patient exposure at skin entrance (bitewing and/or periapicals).
Mechanical support of tubehead.
Visual and audible indication of exposure.
Exposure reproducibility.
Revised October 2011 16
QUARTERLY CHECKLIST ITEMS – Digital Radiography
Each of the items below should pass the visual check or corrective action should be
taken. Use the Quarterly Inspection Record form to note any action taken.
Some of the items on the visual checklist are operator convenience features. However,
many of the items are essential for patient safety and high quality diagnostic images. It
may be necessary to add additional items to the list that are specific to particular
equipment or procedures. These should be included on the checklist and in each
evaluation.
1. Cephalometric beam limiting device (BLD) insert in place and used. Determine if
the unit has built-in collimation or if manual beam limiting devices are being
used. Assure that the collimators are functioning correctly and smoothly.
Determine that manual beam limiting devices are sufficient for blocking the
beam. Assure that collimators or beam limiting devices are being utilized
correctly for the area being radiographed.
2. Panoramic tube and film holder motion smoothness. Determine that during
motion the tube and film holder moves smoothly without catches or interruptions.
3. Cone or cylinder secured to tube. Visually inspect the connection between the
tube and the cone or cylinder. Determine the connection is secure and stable.
Could cause retakes. Operators may not support tube in any way.
4. Tube or generator oil leakage. Visually inspect areas around tubes and generators
for oil or abnormal collections of dust attaching to oil leaks. Equipment problems
and a risk of slips.
5. Condition of cables and wires. All cables and wires should be inspected for frayed
coverings, kinks, and should be free from catching on other objects. Must meet
OSHA standards for safety. Cords should be outside of area. X-ray can damage
cords.
6. Film artifacts. Examine exposed processed films closely to determine if artifacts
are visible. Density variations and other artifacts should not be visible.
7. Cassettes if being utilized, and screens condition. Cassettes and screens should be
cleaned regularly. Check screen condition for dust particles, scratches, and areas
of discoloration. Assure screens are properly fitted and attached to cassettes (if
applicable). Cassettes and screens should be replaced if necessary.
8. Loaded cassette storage. Determine that loaded cassettes are stored in an area that
is properly shielded from radiation to prevent exposure. They should be stored off
the ground and kept free from dust.
9. Control panel indicators. Assure all control panel switches, lights, and meters are
functioning correctly.
10. Technique chart. Make sure technique charts are available, current, and
appropriate for all procedures normally performed.
11. Equipment allows operators to remain behind barriers during X-rays as required.
o Patient view ability. Check for the availability of a method to permit
continuous observation of the patients during the x-ray exposures, except
with intraoral films.
Revised October 2011 17
o Exposure switch placement. Assure the exposure switches are mounted or
exposure switch cables are long enough that operators are not exposed to
unnecessary levels of scatter radiation during x-rays.
Revised October 2011 18
LEAD APRONS
Lead aprons must be utilized for all X-ray studies, regardless of whether the study is
digital or film. Thyroid protection should be provided with each exposure, especially
when films are taken on children.
Lead aprons should never be folded as folds can cause cracks in the lead lining. In order
to protect the useful life of the lead apron always hang or lay flat the apron with no folds.
Always inspect the lead apron visually with each use. If any damage is seen discontinue
use of that apron.
Conduct a more thorough examination of the apron annually:
Preferred Method of Inspection – Image Intensified Fluoroscopy Unit
• Lay the apron or other item on the table
• Examine the entire item with the fluoroscope
• Record results on the Annual QC checklist
Alternate Inspection Method – Fluoroscopy not Available
• Inspect closely for irregularities and kinks
• Take a radiograph of suspect areas
• Review film for breaks in the lead lining
• Record results on the Annual QC Checklist
See the Recycling information in the Supplemental Section of the manual if aprons
need to be discarded. Lead is a hazardous waste.
Revised October 2011 19
RETAKES and REPEATS
Retakes of x-rays regardless of the reasons are expensive for the practice and result in
unnecessary patient exposure. There is no standard or regulation on the number of repeat
rates that are acceptable but the CRCPD in their Quality Control Recommendations
suggests a best practice of a retake rate no greater than 5 to 7 %.
Maintaining records on retakes will allow a practice to calculate the rates, determine the
reasons and take action to reduce the numbers.
1. Record the number of repeats by category on the Repeat X-ray Log.
2. Determine the total number of repeated films and the total number of films
exposed. The overall repeat rate is the total of repeated films divided by the total
number of films exposed during the test period.
3. By dividing the number of repeats per category by the total number of repeated
films, a facility can calculate the repeat rate per category.
Focus corrective action to those areas causing the most retakes. Films that are too light or
too dark may be due to processing problems, equipment problems that require repair or
re-calibration, issues with software or technique charts may need updating.
HAZARD ASSESSMENT CERTIFICATION - DENTAL
PRACTICE NAME: _______________________________________________ DATE: ___________________________
EVALUATED BY:_________________________________________________POSITION:_______________________
POTENTIAL EXPOSSURE
Biological Mechanical Chemical X-Ray
LASER
PPE
Personal Protective
Equipment
Check
All
Tasks
That
Apply
TASKS
BLO
OD
OPI
M
NO
N IN
TAC
T SK
IN
MU
CO
US
MEM
BR
AN
ES
FLY
ING
OB
JEC
TS
SHA
RPS
AB
RA
SIV
ES
BU
RN
S
POIS
ON
S
AB
SOR
PTIO
NS
INH
ALA
TIO
NS
SKIN
/EY
ES
INH
ALA
TIO
NS
GLO
VES
GO
WN
S/A
PRO
NS
MA
SKS
EYEW
EAR
AM
BU
BA
GS
MIC
RO
SH
IELD
S
EXAM (no A/W syringe) √ √ √ √ √ √
EXAM (with A/W syringe) √ √ √ √ √ √ √ √ √ √
PROPHY √ √ √ √ √ √ √ √ √ √ √ √
X-RAY √ √ √
RESTORATIONS √ √ √ √ √ √ √ √ √ √ √ √
CROWN AND BRIDGE √ √ √ √ √ √ √ √ √ √ √ √
ENDODONTICS √ √ √ √ √ √ √ √ √ √ √ √ √ √
PERIODONTICS √ √ √ √ √ √ √ √ √ √ √
ORAL SURGERY √ √ √ √ √ √ √ √ √ √ √
LAB PROCEDURES √ √ √ √ √ √ √ √ √ √ √ √ √
INJECTIONS √ √ √ √ √ √
INSTRUMENT DECONTAMINATION √ √ √ √ √ √ √ √ √ √ √ √ √
OPERATORY DECONTAMINATION √ √ √ √ √ √ √
WASTE HANDLING √ √ √ √ √ √ √ √ √
CPR √ √ √ √
SCALING √ √ √ √ √ √ √ √ √ √
ORTHODONTICS
NOTE: Please review all categories that pertain to your practice to insure the proper fields are checked.
ECP 551 D (Revised 2/2004)