With LED Illumination - Medidenta2. Connect contra-angle (no load for the contra angle) 3. Press...

12
Cordless Endodontic Handpiece With LED Illumination Instructions For Use Read Instructions Before Use

Transcript of With LED Illumination - Medidenta2. Connect contra-angle (no load for the contra angle) 3. Press...

Page 1: With LED Illumination - Medidenta2. Connect contra-angle (no load for the contra angle) 3. Press down the Select key until “CAL” arises 4. Press On/Off button once “CAL” appears,

Cordless Endodontic HandpieceWith LED Illumination

Instructions For Use

Read Instructions Before Use

Page 2: With LED Illumination - Medidenta2. Connect contra-angle (no load for the contra angle) 3. Press down the Select key until “CAL” arises 4. Press On/Off button once “CAL” appears,

2

INTENDED USE

PACKING LIST

COMPONENTS

The RootPro CL is a cordless endodontic handpiece that is intended to be used primarily for root canal enlargement.

1. Handpiece2. Contra angle Mini-Head3. Charging Station4. Power Cable for Charging Station

1. 1x Cordless Endo Motor Rootpro CL- Motor Handpiece2. 1x Charging Base3. 1x HP Contra Angle 20:14. 1x Adapter5. 2x LED Lamp Holder6. 1x Motor Cover7. 1x Instructions For Use

1.

3.

2.

4.

Page 3: With LED Illumination - Medidenta2. Connect contra-angle (no load for the contra angle) 3. Press down the Select key until “CAL” arises 4. Press On/Off button once “CAL” appears,

3

OPERATION PANEL

B

F

Program Mode Display

Torque

Speed

Battery

E

D

A

C

B. On/Off Key

C. Select

D. Up/Down Key

E. Program Key

F. LCD Panel

A. Power Key

When the Unit is Powered On, Press On/Off key to start the motor. If you re-press the key, it will stop the motor.

-Used to change Torque (TQ) and Speed (SP)-The Torque and Speed can be changed while the handpiece is running. -Selected items will flash.

Use to Adjust the set value for each parameter. Please use the suitable contra-angle based on gear ratio

Press the Program Key to change the current program selected for operation of the handpiece. The Program numbers are available from 1-6

Switch-onHold down the Power Key for more than one second, and the LCD Panel will light upSwitch-offHold down the Power Key for more than one second to turn power off. Note:If 5 minutes pass without operation the handpiece will power off automatically.

Speed (rpm) Gear Ratio Torque (N.CM)120-500 20:1 0.6-0.4

Page 4: With LED Illumination - Medidenta2. Connect contra-angle (no load for the contra angle) 3. Press down the Select key until “CAL” arises 4. Press On/Off button once “CAL” appears,

4

LCD PANEL

Program Display (P1-P6)

Speed Display

Torque Display

Mode Display

6 Different programs can be preset and saved according to the requirements of the operator.

This Section of the Panel will show your motor speed in RPM

This Section of the Panel will show the amount of torque being produced.

Displays the present mode set.

Forward Rotation - 360° Auto Reverse Forward

Reverse-Reciprocating• Adjustable Speed: 200~400• Reverse Rotation 150°, Forward Rotation 30°

(approx.)

Reverse-Rotation- 360° Auto Reverse Forward

Forward-Reciprocating• Adjustable Speed: 200~400• Reverse Rotation 150°, Forward Rotation 30°

(approx.)

Forward-Rotation 360°• Stopped when blocked (No Reverse)

Reverse-Rotation 360°• Stopped when blocked

Page 5: With LED Illumination - Medidenta2. Connect contra-angle (no load for the contra angle) 3. Press down the Select key until “CAL” arises 4. Press On/Off button once “CAL” appears,

5

CHARGING

Battery RemainingDisplays the amount of power remaining in the battery, the mark is animated when the battery is charging.

Note: It is normal to find the mark animated during the operation, for safety, the motor will shut down when the battery is charged.

Full Charge

Approximately 30%-80% remaining

Less than 30% remaining.• In this case , the auto-reverse function may not

activate

Battery Needs to be Charged

A. Connect the Power Adapter to the charging stand.• Insert the power cable to the backside of the charging unit.

B. Place the handpiece on the charging stand, during this period, please confirm that a beep is heard and the power display flashes.

C. If the beep has been heard and the power display shows “FUL”, Then the battery has been fully charged.

Note: Please ensure the handpiece has been fully charged prior to initial use.

Page 6: With LED Illumination - Medidenta2. Connect contra-angle (no load for the contra angle) 3. Press down the Select key until “CAL” arises 4. Press On/Off button once “CAL” appears,

6

CONNECTING AND DISCONNECTING CONTRA-ANGLE

Connecting contra-angle:The Contra-angle can be connected at 6 adjustable head positions. Align the positioning pins of the contra-angle with the positioning slots of the motor handpiece and insert until you hear a click.

Disconnecting Contra-angleWhen removing the contra-angle, pull straight out (as seen above) Note:^Before detaching the contra-angle ensure that the power has been shut off. ^^ Check to ensure that the contra-angle is secured properly to the motor.

SETTING TORQUE LIMIT

ADJUSTING SPEED

1. Press the Select key while power is on until the Torque display begins blinking to select from 17 different torque level settings

2. Press the Up/Down key until the desired torque level is selectedThe handpiece will begin auto-reverse when the preset value is reached.

1. Press the Select key while power is on until the Speed display begins blinking to select from 11 different speed settings

2. Press the Up/Down key until the desired Speed is selectedEnsure that the Speed suggested is in accordance with the file manufacturer recommendations.

.6 .7 .8 .9 1.0 1.2 1.4 1.6 1.8 2.0 2.2 2.4 2.6 2.8 3.0 3.5 4.0

Torque Setting Level (N.CM)

Speed Setting Level (RPM)

120 140 160 180 200 250 300 350 400 450 500

Page 7: With LED Illumination - Medidenta2. Connect contra-angle (no load for the contra angle) 3. Press down the Select key until “CAL” arises 4. Press On/Off button once “CAL” appears,

7

INSERTING AND REMOVING FILE

Note:^Before inserting or removing file, ensure that the power has been shut off to the handpiece. ^ After the file is locked, lightly pull on the file to ensure that it is secure.

File Insertion

File Removal

1. Insert the file into the chuck until it stops.2. Lightly turn the file until it engages with the latch mechanism.3. Push inwards until you hear a click.

Press the push-button and pull out the file.

INSERTING AND REMOVING LED LIGHT

Inserting LED Light:Insert LED light holder to the socket on the main unit, and buckle the upper part to the contra-angle firmly.

Removing LED Light:Pull out the buckle from the contra-angle first then pull the LED light holder upward.

Note: The LED lamp holder can be autoclaved at 135°C.

Page 8: With LED Illumination - Medidenta2. Connect contra-angle (no load for the contra angle) 3. Press down the Select key until “CAL” arises 4. Press On/Off button once “CAL” appears,

8

OPERATION

AFTER OPERATION

INITIALIZATION OF PROGRAM

1. Press the power button to start the unit2. Press the On/Off button to begin rotation and to engage LED light. 3. To stop rotation press the On/Off button again (the LED light will shut off 8

seconds later)

Note:If you press and hold the On/Off button the engine will continue rotation until you release the button. the LED will shut off 8 seconds later.

The program can be returned to the original state configuration. 1. Turn on the power, and choose P1, P3, P5, or P62. Connect contra-angle (no load for the contra angle)3. Press down the Select key until “CAL” arises4. Press On/Off button once “CAL” appears, One “click” will be heard,

cancellation will begin, after finished, two “click” followed, and return to home screen.

5. Initialization is completed, when LCD panel display normally

Note: Initialization is used when the LCD screen does not appear in accordance with the selected mode or settings.

Press the Power button to turn off the unit once operation is completed.

Auto Turn Off Function: The device will automatically shut off after 5 minutes without operation.

SPECIFICATIONS

Battery Voltage 3.7Vdc

Battery Capacity 800mAh

Handpiece Weight 133g

Charging Station Weight 80g

Shipping & Storage Conditions -10°C ~ +50°C | Relative Humidity ≤80%

Operation Environment 5°C ~ 40°C | Relative Humidity ≤80%

Rated Input Voltage AC100-240V | 50HZ/60HZ

Output Voltage DC5V | Output Current: 1A

Page 9: With LED Illumination - Medidenta2. Connect contra-angle (no load for the contra angle) 3. Press down the Select key until “CAL” arises 4. Press On/Off button once “CAL” appears,

9

LUBRICATION

STERILIZATION

• Lubricate the contra angle with Medidenta's I-Spray• Lubricate after each use and prior to each sterilization

• Screw the spray nozzle onto the spray until it is fully seated.

• Hold the contra-angle securely to prevent it from detaching with the pres-sure of the spray.

• Insert the spray nozzle into the rear part of the contra-angle and lubricate for 2-3 seconds until oil comes out of the contra angle head.

• Before attaching the lubricated contra angle to the motor handpiece. Wipe off excess oil with a dry cloth. Place it on its end or lean it at an angle for gravity draining. re-attach after excess oil has been drained.

• Do Not: Lubricate the motor handpiece

Insert the contra-angle into a sterilization pouch. Seal the pouch, bag individually.

Do Not exceed 135°C while autoclaving accessories (121°C for 20 Minutes or 132°C for 15 minutes. Please seal the unused device into sterilization bags and store in a clean area.

Please Note:Please Autoclave contra-angle , files, and LED holder only. Place vessel in the center or upper part of the disinfection cabinet due to higher tempatures near the bottom of the disinfection stand.

I-Spray

Page 10: With LED Illumination - Medidenta2. Connect contra-angle (no load for the contra angle) 3. Press down the Select key until “CAL” arises 4. Press On/Off button once “CAL” appears,

10

TROUBLESHOOTINGWhen Issues arise, check the following prior to contacting your dealer. If none of the following are applicable or if the error is not remedied even after action has been taken, the product may have failed. Contact your dealer.

Error Cause Solution

The power is not turning on. The battery is low Charge the battery.

The power button was not held long enough.

Press and hold the power button.

The power button is damaged. Contact your dealer.

The battery wont charge.

The adapter isnt connected well. Re-check your connection.

The unit was not placed on the charger correctly. Re-check your connection.

The battery is damaged. Contact your dealer.

Fault Code

Function Description

E-1 Overcurrent protection

The motor will shut off when overcurrent happens during operation.

E-2 Overheating protection

The motor will shut off when overheating happens during long periods of heavy operation.

E-3 Energy consumption

protection

The motor will shut off in 15 minutes if the situation doesnt change

E-4 Correcting protection

Fault calibration. Exit calibration directly when current over theoretical value.

Note: Press any button to eliminate fault codes except for the Power button.

Page 11: With LED Illumination - Medidenta2. Connect contra-angle (no load for the contra angle) 3. Press down the Select key until “CAL” arises 4. Press On/Off button once “CAL” appears,

11

WARNING• The charging station is not water proof .• The handpiece is only water resistant. Do not immerse the handpiece in

water or any liquid. • Do not lubricate or autoclave the motor. • Do not apply disinfecting agents containing chemicals such as

glutaraldehyde on the plastic parts of this unit as this will cause them to discolor and break down.

• Damages caused by use of such products will not be covered under warranty. • Only use the power adapter provided to charge the unit. • Follow the required guidelines to dispose of the unit, accessories, waste

products, residues, etc.• No modification of this unit is permitted.

WARRANTY

Do not modify or disassemble the device.

This unit has been fully inspected and conforms to the current products specification.

This instrument is guaranteed for its designated use, against original defects in materials and workmanship for a period of 12 months from date of purchase.

Products warranty or service will not be extended if (1) the product is repaired, modified, misused, disassembled, or using the parts are not provided by the manufacturer, (2) The serial number of the product is defaced or missing.

The guarantee for accessories is 6 months. If accessories of the device are damaged or needed to be replaced the user can purchase from the manufacturer.

Please contact sales representative from whom you have bought this device for user’s record and further after-sale service.

Damage to the product resulting from acts of god, faulty installation, misuse, tampering, accident, abuse, negligence, alterations, unauthorized or problems unrelated to materials and workmanship are not covered by this warranty.

Page 12: With LED Illumination - Medidenta2. Connect contra-angle (no load for the contra angle) 3. Press down the Select key until “CAL” arises 4. Press On/Off button once “CAL” appears,

This device is Manufactured by:Medidenta3325 W. Ali Baba Ln STE 603Las Vegas NV, 89118

Website: www.medidenta.comEmail: [email protected]: 800.221.0750Fax: 702.749.8610

Please Contact sales representative from whom you have bought this device for user’s record and further after-sale service.