WIDE-ANGLE PACKAGED X-RAY TUBE

1
WIDE-ANGLE PACKAGED X-RAY TUBE Ideally suited for end users wanting a plug and play solution. Our packaged tubes are fully radiation shielded, high voltage isolated and include a high voltage cable and a low voltage cable. Our packaging has been designed to provide maximum thermal properties enabling continuous operation, without compromise. The control of the high voltage and the filament current as well as the design of the cooling system and the radiation protection are the responsibility of the original equipment manufacturer (OEM). Careful selection of power supply should assure that the X-Ray tube will be protected against overcurrent, overvoltage and lack of cooling. Otherwise the tube and/or the radiation protection may be damaged and become a hazard. When installing the X-ray tube assembly into X-ray equipment and/or operating X-ray equipment, the responsibility for radiation protection from the X-ray window is with the user. Compliance with local regulatory requirements and limit values must be assured. SPECIFICATIONS GENERAL RADIATION PROTECTION MICRO X-RAY INC, SANTA CRUZ, CA MICROXRAY.COM +1 831 207 4900 Polarity Grounded Cathode Flange Type (6) 8-32 thread High Voltage Range 4-60 kV 1 Anode current 1 mA (3 mA maximum) 1 Continuous rating 50 W (higher on request) 1,2 Focal spot 33-50 μm, 100 μm Filament current, max. 1.7 A Filament voltage, (nominal) 2.5 V Inherent filtration 127 μm Be Target materials Mo, W, Rh, Cr, Ag, Cu, Co Target Angle 20° Radiation coverage 40° Radiation leakage <2 μSv/hr @50mm (60kv, 1mA) 1 Interrelationship between voltage, current and power will affect specs 2 Up to 150 W with external cooling module

Transcript of WIDE-ANGLE PACKAGED X-RAY TUBE

Page 1: WIDE-ANGLE PACKAGED X-RAY TUBE

WIDE-ANGLE PACKAGED X-RAY TUBE

Ideally suited for end users wanting a plug and play

solution. Our packaged tubes are fully radiation shielded,

high voltage isolated and include a high voltage cable

and a low voltage cable. Our packaging has been

designed to provide maximum thermal properties

enabling continuous operation, without compromise.

The control of the high voltage and the filament current as well

as the design of the cooling system and the radiation

protection are the responsibility of the original equipment

manufacturer (OEM). Careful selection of power supply should

assure that the X-Ray tube will be protected against

overcurrent, overvoltage and lack of cooling. Otherwise the

tube and/or the radiation protection may be damaged and

become a hazard.

When installing the X-ray tube assembly into X-ray equipment

and/or operating X-ray equipment, the responsibility for

radiation protection from the X-ray window is with the user.

Compliance with local regulatory requirements and limit values

must be assured.

SPECIFICATIONS GENERAL

RADIATION PROTECTION

MICRO X-RAY INC, SANTA CRUZ, CA MICROXRAY.COM +1 831 207 4900

Polarity Grounded Cathode

Flange Type (6) 8-32 thread

High Voltage Range 4-60 kV 1

Anode current 1 mA (3 mA maximum) 1

Continuous rating 50 W (higher on request) 1,2

Focal spot 33-50 µm, 100 µm

Filament current, max. 1.7 A

Filament voltage, (nominal) 2.5 V

Inherent filtration 127 µm Be

Target materials Mo, W, Rh, Cr, Ag, Cu, Co

Target Angle 20°

Radiation coverage 40°

Radiation leakage <2 µSv/hr @50mm (60kv, 1mA)

1 Interrelationship between voltage, current and power will affect specs2 Up to 150 W with external cooling module

WIDE-ANGLE PACKAGED X-RAY TUBE

Ideally suited for end users wanting a plug and play

solution. These tubes are fully radiation shielded, high

voltage isolated and include a high voltage cable and low

voltage cable. The bronze packaging provides radiation

shielding as well as excellent thermal properties, enabling

continuous operation, without compromise.

The control of the high voltage and the filament current as well

as the design of the cooling system and the radiation

protection are the responsibility of the original equipment

manufacturer (OEM). Careful selection of power supply should

assure that the X-Ray tube will be protected against

overcurrent, overvoltage and lack of cooling. Otherwise the

tube and/or the radiation protection may be damaged and

become a hazard.

When installing the X-ray tube assembly into X-ray equipment

and/or operating X-ray equipment, the responsibility for

radiation protection from the X-ray window is with the user.

Compliance with local regulatory requirements and limit values

must be assured.

SPECIFICATIONS GENERAL

RADIATION PROTECTION

MICRO X-RAY INC, SANTA CRUZ, CA MICROXRAY.COM +1 831 207 4900

Polarity Grounded Cathode

Flange Type (6) 8-32 thread

High Voltage Range 4-60 kV

Anode current, maximum 3.0mA *

Continuous rating 75 W *

Focal spot from 30 µm up to 1000 µm

Filament current, max. 1.7 A

Filament voltage, (nominal) 2.5 V

Inherent filtration 127 µm Be

Target materials Mo, W, Rh, Cr, Ag, Cu, Co

Target Angle 20°

Radiation coverage 40°

Radiation leakage <2 µSv/hr @50mm (60kv, 1mA)

* Up to 150 W with external cooling module

mA

MAX

MAX