What is Residual solvent and its identification

29
RESIDUAL SOLVENTS USP <467> 1 ACI Limited

Transcript of What is Residual solvent and its identification

Page 1: What is Residual solvent and its identification

RESIDUAL SOLVENTS USP <467>

1ACI Limited

Page 2: What is Residual solvent and its identification

What is Residual Solvent? Organic volatile impurities (OVIs), commonly referred to as

residual solvents, are trace level chemical residues in drug substances and drug products that are byproducts of manufacturing or that form during packaging and storage.

Drug manufacturers must ensure that these residues are removed, or are present only in limited concentrations.

The International Conference on Harmonization (ICH) Q3C guideline lists the acceptable amounts of solvent residues that can be present.

2ACI Limited

Page 3: What is Residual solvent and its identification

3

Why Residual Solvent is important?

To ensure safety of the patients it is recommended to use less toxic solvents.

To prevent any side-effects of drugs by any carcinogenic or toxic trace elements.

Testing is to be performed only for solvents used or produced in the final manufacturing step, or used in previous step and not removed by validated procedure. Thus to maintain quality of drugs.

ACI Limited

Page 4: What is Residual solvent and its identification

4

Residual Solvent Classes

ACI Limited

Page 5: What is Residual solvent and its identification

Class 1 (solvents to be avoided) Class 1 residual solvents should not be employed in the

manufacture of drug substances, excipients, and drug

products because of the unacceptable toxicities or deleterious environmental effects of these residual solvents.

5ACI Limited

Page 6: What is Residual solvent and its identification

6

Class 2 Residual SolventsClass 2 residual solvents should be limited in drug substances, excipients, and drug products because of the inherent toxicities of the residual solvents. PDEs are given to the nearest 0.1 mg per day.

ACI Limited

Page 7: What is Residual solvent and its identification

Class 2 Residual Solvents (continued…)

7ACI Limited

Page 8: What is Residual solvent and its identification

8

Class 3 Residual SolventClass 3 residual solvents may be regarded as less toxic and of lower risk to human health than Class 1 and Class 2 residual solvents. Class 3 includes no solvent known as a human health hazard at levels normally accepted in pharmaceuticals.

ACI Limited

Page 9: What is Residual solvent and its identification

RS Limits for DP/DS/Excipients “Option 1”: (Concentration limits, ppm) for Class 1,

2, & 3 Residual Solvents

“Option 2”: (Permitted Daily Exposure/PDE) for Class 2 & 3 Residual Solvents

Option-2 is used when Drug Substance, Drug Product and Excipients exceed Option-1 RS limits, or Drug Product daily dose exceeds 10 g

ACI Limited 9

Page 10: What is Residual solvent and its identification

10

Class1 Solvent (to be avoided): General Requirements All Class – 1 Solvents likely to be present in drug

product components should be identifeid and quantified.

Drug product component manufacturer (API, excipients) should report residual solvent concentration to the Drug product manufacturer.

Option 1 concentration limit (ppm) can be applied to API, excipients, or finished drug product.

Option 2 generally may not be applied to class 1 residual solvents.

ACI Limited

Page 11: What is Residual solvent and its identification

11

Class 2 Solvents (to be limited): General Requirements All Class - 2 Solvents likely to be present in drug

product components should be identified. Option 1 may be applied if all drug product

components have residual solvents concentration below the Table 2 Limits.

Components residual solvents concentrations must be reported if Option 1 limits are exceeded.

Option 2 applies when option 1 fails Daily exposure calculated to determine if drug

product meets the PDE Limits.

ACI Limited

Page 12: What is Residual solvent and its identification

12

Class 3 Solvents (low toxicity): General Requirements All Class 3 solvents likely to be present should be

identified. <731> Loss on drying (LOD) method can be used to

demonstrate the drug product component meets option 1 Limit of NMT 0.5 % when referenced in monograph

FDA recommends the LOD test when only class 3 residual solvents are present.

If Option 1 Limit of NMT 0.5 % is exceeded, residual solvents in Drug Substance, Drug Product and Excipients should be reported

Option 2 (PDE NMT 50 mg/day for drug product) can be applied

ACI Limited

Page 13: What is Residual solvent and its identification

13

Responsibility for Reporting RS

Drug product sponsors are responsible for Quality and safety of their drug products, and Reporting all required RS information.Keeping comprehensive, accurate records and documentation of compliance with all applicable GMP and other regulationsThe Drug Substance, Drug Product and Excipients manufacturer should establish the reliability of such analyses, through verification at appropriate intervals Drug Product manufacturer should verify the Vendor Statements

ACI Limited

Page 14: What is Residual solvent and its identification

Principle for analysis of residual solvents The analysis of residual solvents is commonly performed using

static headspace gas chromatography (HS/GC). The basic premise behind headspace analysis begins with the

addition of an exact, known volume or weight of sample into a closed, sealed vial. This creates two distinct phases in the vial—a sample phase and a gaseous phase, or “headspace”.

Volatile components inside the sample phase, whether a solid or solution, can be extracted, or partitioned, from the sample phase into the headspace.

An aliquot of the headspace can then be taken and delivered into a GC system for separation and detection.

14ACI Limited

Page 15: What is Residual solvent and its identification

If we look at the anatomy of a headspace vial we can begin to see the relationship of the vial components and how we can control these parameters to create analytical methods.

Volatile components partition from the sample phase and equilibrate in the vial headspace.

15ACI Limited

Page 16: What is Residual solvent and its identification

Headspace Vial

ACI Limited 16

Page 17: What is Residual solvent and its identification

Principle for analysis of residual solvents (continued…)

Residual solvent analysis by static HS/GC can be enhanced by careful consideration of two basic concepts—partition coefficient (K) and phase ratio (β).

Partition coefficients and phase ratios work together to determine the final concentration of volatile compounds in the headspace of sample vials.

The partition coefficient (K) is defined as the equilibrium distribution of an analyte between the sample and gas phases. Compounds that have low K values will tend to partition more readily into the gas phase, and have relatively high responses and low limits of detection.

The phase ratio (β) is defined as the volume of the headspace over the volume of the sample in the vial. Lower values for β (i.e., larger sample sizes) will yield higher responses for compounds with inherently low K values.

17ACI Limited

Page 18: What is Residual solvent and its identification

Principle for analysis of residual solvents (continued…)

Striving for the lowest values for both K and β when preparing samples will result in higher concentrations of volatile analytes in the gas phase and, therefore, better sensitivity

18ACI Limited

Page 19: What is Residual solvent and its identification

Overview of Method for Residual Solvent Testing USP <467> is divided into two separate sections based upon

sample solubility: water-soluble and water-insoluble articles. The methodology for both types of articles is similar, but the diluent used in both standard and sample preparations differs based upon the solubility of the test article.

The test method consists of three procedures (A, B, and C), that are designed to identify, confirm, and then quantify residual solvents in drug substances and products .

The revised USP <467> method consists of a static headspace extraction coupled with a gas chromatographic separation and flame ionization detection.

19ACI Limited

Page 20: What is Residual solvent and its identification

20ACI Limited

Page 21: What is Residual solvent and its identification

Procedure A - Identification

ACI Limited 21

Procedure A is the first step in the identification process and is performed on a G43 column to determine if any residual solvents are present in the sample at detectable levels.

First, Class 1 standard and system suitability solutions and Class 2 Mix A standard solutions are assayed under the method-specified operating conditions to establish system suitability.

All peaks in the Class 1 system suitability solution must have a signal-to-noise ratio not less than 3, the Class 1 standard solution must have a 1,1,1-trichloroethane response greater than 5, and the resolution of acetonitrile and dichloromethane must be not less than 1 in the Class 2 Mixture A solution.

When system suitability has been achieved, the test solutions are assayed along with the Class 1 and Class 2 Mixtures A and B standard solutions. If a peak is determined in the sample that matches a retention time and has a greater response than that of a corresponding reference material, then Procedure B is performed for verification of the analyte.

Page 22: What is Residual solvent and its identification

a G43 Column

ACI Limited 22

Page 23: What is Residual solvent and its identification

Identification of residual solvent in sample by comparing with reference standard

ACI Limited 23

Page 24: What is Residual solvent and its identification

Procedure B - Confirmation Once a residual solvent is identified and found to be above the percent

daily exposure limit, Procedure B is performed to confirm analyte identity.

A G16 capillary column is used here as a confirmation column, because it yields an alternate selectivity compared to a G43 column. The same standard and system suitability preparations are used in Procedures A and B.

The system suitability requirements differ here in that the Class 1 standard solution must have a benzene response greater than 5 and the resolution of acetonitrile and cisdichloroethene must not be less than 1 in the Class 2 Mixture A solution, a change from the original version.

If the analyte identified in Procedure A again matches the retention time and exceeds the peak response of the reference materials (with the same exception to 1,1,1-trichloroethane), the analyst must quantify the analyte using Procedure C.

24ACI Limited

Page 25: What is Residual solvent and its identification

Confirmation of residual solvent in sample by comparing retention time and peak response

ACI Limited 25

Page 26: What is Residual solvent and its identification

Procedure C – Quantification Once a residual solvent has been identified and verified,

Procedure C is used to quantify the analyte by analyzing the sample against compound-specific reference materials.

Individual standards are prepared by diluting the analyte in solution to a concentration of 1/20 of the concentration limit given under concentration limit Table 1 or 2 of the method.

Following the procedure and instrument conditions in either Procedure A or B (whichever provides the most definitive results), a quantifiable result is produced.

For water-insoluble articles, the same procedure is followed, except dimethylformamide or dimethylsulfoxide is used as the diluent.

26ACI Limited

Page 27: What is Residual solvent and its identification

Confirmation of residual solvent in sample by analyzing the sample against compound-specific reference materials.

ACI Limited 27

Page 28: What is Residual solvent and its identification

ACI Limited 28

QUESTIONS ?

Page 29: What is Residual solvent and its identification

THANK YOU

ACI Limited 29

By

Hasan Al BannaQuality Control Officer

Quality Assurance Department

ACI Limited (Pharma Plant)