Thermo Fisher Scientific Wax Dispensers Operator Guide 1
Wax Dispensers Operator Guide - English
A81610200 - Issue 1
Wax DispenserHigh Capacity Wax Dispenser
Slimline Wax Dispenser
2 Wax Dispensers Operator Guide Thermo Fisher Scientific
© 2009 Thermo Fisher Scientific. All rights reserved.
Thermo Shandon Limited is an ISO 9001 and TickIT Accredited Company
Thermo Fisher Scientific is the trading name of Thermo Shandon Limited
All other trademarks are the property of Thermo Fisher Scientific and its subsidiaries
Thermo Fisher Scientific makes every endeavour to ensure that the information contained in its support documentation is correct and clearly stated but does not accept responsibility for any errors or omissions. The development of Thermo Fisher products and services is continuous. Make sure that any published information that you use for reference is up to date and relates to the status of the product. If necessary, check with Thermo Fisher or your local Thermo Fisher representative.
This manual may not, in whole or in part, be copied, photocopied, reproduced, translated, or converted to any electronic or machine readable form without prior written consent of Thermo Fisher.
All information contained in this manual is proprietary and confidential, and the exclusive property of Thermo Fisher Scientific. This manual is protected by copyright and any reproduction is prohibited. This manual is for use only by the individuals to whom it has been made available by Thermo Fisher Scientific.
Manufacturer/Contact Address Contact AddressAnatomical PathologyTudor Rd, Manor ParkRuncorn WA7 1TA, UK
Tel: +44 (0)1928 534050Fax: +44 (0)1928 534049www.thermo.com/pathology
Anatomical Pathology4481 Campus DriveKalamazooMI 49008, USA
Tel: 1-800-522-7270Fax: +1 269-372-2674www.thermo.com/pathology
The Thermo Scientific Wax Dispensers meet the following CE Mark requirements:In Vitro Diagnostic Directive 98/79/EC
Low Voltage Directive 2006/95/EC
Thermo Fisher Scientific Wax Dispensers Operator Guide 3
The following symbols and conventions are used throughout this manual and on the instrument.
this syMbol is UsED oN thE EQUiPMENt, oR iN A DoCUMENt, to wARN thAt iNstRUCtioNs MUst bE followED foR sAfE AND CoRRECt oPERAtioN. if this syMbol APPEARs oN thE iNstRUMENt, AlwAys REfER to this oPERAtoR GUiDE.
this syMbol wARNs yoU thAt sURfACEs ARE hot. if this syMbol APPEARs oN thE iNstRUMENt, AlwAys REfER to this oPERAtoR GUiDE.
A wARNiNG is GivEN iN thE DoCUMENt if thERE is A DANGER of PERsoNAl iNjURy oR DAMAGE to sAMPlEs oR EQUiPMENt.
Note Notes give more information about a job or instruction but do not form part of the instructions.
Symbols
4 Wax Dispensers Operator Guide Thermo Fisher Scientific
ContentsSymbols 3
Chapter 1 Introduction 6
1.1 Sample Safety 61.2 Your Safety 71.3 Switching the Instruments On and Off 91.4 Unpacking the Instruments 101.5 Taking out of Operation 111.6 Common Safety Warnings 121.7 If the Thermal Cut-Out Operates 13
Chapter 2 Wax Dispenser 14
2.1 General Description 142.2 Cautionary Notes 162.3 Operating Procedure 172.4 Cleaning Procedure 19
Chapter 3 High Capacity Wax Dispenser 21
3.1 General Description 213.2 Cautionary Notes 233.3 Operating Procedure 253.4 Cleaning Procedure 27
Chapter 4 Slimline Wax Dispenser 28
4.1 General Description 284.2 Cautionary Notes 304.3 Operating Procedure 324.4 Cleaning Procedure 35
Thermo Fisher Scientific Wax Dispensers Operator Guide 5
Appendix A Technical Specification 36
A1. Wax Dispenser 36A2. High Capacity Wax Dispenser 37A3. Slimline Wax Dispenser 38
Index 39
6 Wax Dispensers Operator Guide Thermo Fisher Scientific
Chapter 1 IntroductionBefore operating any of the equipment the user should have read and understood the safety section of the safety and warranty booklet (A79910001).
All laboratory staff should also read this Operator Guide thoroughly.
Any inspection or maintenance repairs must be carried out only by a qualified Service Engineer.
thE oPERAtoR MUst ENsURE thAt thE sEt tEMPERAtURE is sUCh thAt No DAMAGE is CAUsED to sAMPlEs withiN ANy of thE DEviCEs. if A DEviCE MAlfUNCtioNs, ChECK All sAMPlEs withiN thE DEviCE foR hARM oR DAMAGE
1.1 Sample Safety
Thermo Fisher Scientific Wax Dispensers Operator Guide 7
i. If the equipment is used in a manner or for a purpose not recommended by Thermo Fisher Scientific, the protection provided by the equipment might be impaired and may introduce a hazard to health or damage the equipment.
ii. The equipment needs no regular servicing, but we recommend an annual inspection to prolong the life of the instrument and ensure continued safety (details are available from your Thermo Supplier).
iii. If the unit has been subjected to humid conditions after storage or transportation, the equipment could fail to meet all the safety requirements of the standards shown in this manual, in which case the equipment should be left to dry out for at least 3 hours.
iv. All the instruments must be sited on surfaces which are non-flammable, level and secure.
v. Make sure that the voltage of the mains supply corresponds with the voltage rating on the rating plate on the back of the instrument.
Note The ~ symbol on the rating plate indicates that the instrument operates on an alternating current supply (ac).
vi. Instruments are supplied with power cords with moulded plugs. If you want a different plug, a technically competent person must remove the moulded plug and fit a suitably rated fused plug, using the wiring convention:
Brown wire – Live (L or L2) terminal
Blue wire – Neutral (N or L1) terminal
Green/yellow wire – Earth terminal (E or )
1.2 Your Safety
8 Wax Dispensers Operator Guide Thermo Fisher Scientific
Insert the mains cable into the mains connector on the rear panel of the instrument and clip the cable restraint over the mains connector. Connect the mains supply cable to the local power supply outlet.
thE iNstRUMENts MUst bE PRotECtivEly EARthED. MAKE sURE thAt thEy ARE PlUGGED iNto A PRoPERly EARthED MAiNs sUPPly
it MUst bE PossiblE to iNtERRUPt thE MAiNs sUPPly At soURCE by REMoviNG thE PlUG fRoM thE MAiNs sUPPly soCKEt
Thermo Fisher Scientific Wax Dispensers Operator Guide 9
To switch the instruments on, press the I (ON) side of the I/O power switch.
To switch the instruments off, press the O (OFF) side of the I/O power switch.
1.3 Switching the Instruments On and
Off
10 Wax Dispensers Operator Guide Thermo Fisher Scientific
If the packaging has been damaged, check the condition of the instrument. Contact your Thermo dealer if there is any damage.
If necessary, get help to lift the instrument from its box and to place it on the bench. To lift or move the instrument, hold it securely under the sides of the unit.
Make sure you have received all the parts listed on the packing list supplied with the instrument. If necessary, contact your Thermo dealer.
Notes Tell your Thermo dealer immediately if there are any breakages or shortages. Quote the instrument Serial Number, and your Order Number, Invoice Number, and Delivery Note (or Packing Slip) Number, and the date.
1.4 Unpacking the Instruments
Thermo Fisher Scientific Wax Dispensers Operator Guide 11
If the instrument is to be taken out of operation for a long period of time, or put into storage:
• Ensure the instrument is thoroughly clean
• Re-pack it carefully into its original packaging
Returning the instrumentIf you need to return the instrument:
• Ensure the instrument is thoroughly clean and where necessary accompanied by a completed Decontamination Certificate
• Re-pack it carefully into its original packaging
• Complete a copy of the PRODUCT RETURN SAFETY DECLARATION found in the Safety and Warranty Booklet (A79910001)
1.5 Taking out of Operation
12 Wax Dispensers Operator Guide Thermo Fisher Scientific
This section refers to all Thermo laboratory support products. Make sure you read and understand before proceeding.
bEfoRE CoNNECtiNG thE EQUiPMENt to thE MAiNs sUPPly, yoU MUst obsERvE stRiCtly thE UsUAl ElECtRiCAl sAfEty ChECKs
Do:• Maintain the instrument in a reasonably clean condition
• Switch the instrument off before removing the plug
• Ensure that replacement fuses are of the correct specification
Do Not:• Do not use the instrument for non-specified purposes,
without prior consultation with the supplier
wE RECoMMEND thAt:• You disconnect the equipment from the mains when it is
not required for use
• In the event of mechanical damage to the power cable or instrument, a service engineer check the instrument before use
• Only a trained Service Engineer carry out repairs
1.6 Common Safety Warnings
Thermo Fisher Scientific Wax Dispensers Operator Guide 13
The red lamp glows when the non-resetting thermal safety cut-out has been triggered. If this happens:
1. Switch the equipment off and disconnect it from the mains.
2. Leave the equipment to cool for one hour and then retry.
3. Consult a Service Engineer if the equipment has not reset.
1.7 If the Thermal Cut-Out Operates
14 Wax Dispensers Operator Guide Thermo Fisher Scientific
Chapter 2 Wax DispenserThe Wax Dispenser holds up to 5 litres of wax at a thermostatically controlled temperature. Wax is delivered via a two-position tap: the first position allows wax to be dispensed only while the lever is held down, and the second position allows the lever to be locked open for continuous flow (this may require you to turn the lever). A 0.5mm mesh at the base of the reservoir prevents coarse debris from blocking the delivery assembly. A separate tap heater ensures an even flow of wax.
Figure 1 shows the main features of the Wax Dispenser.
Figure 1. Main features of the Wax Dispenser
2.1 General Description
Wax reservoir(under lid)
Temperaturecontrol dial
Thermostat indicator lamp (amber)
Wax delivery tap
Mains rocker switch
Thermostat safety cut-outindicator lamp (red) at rear
Thermo Fisher Scientific Wax Dispensers Operator Guide 15
The Wax Dispenser operates over a temperature range of 50°C to 70°C. The outer casing is made from polished stainless steel. The reservoir and lid are made from aluminium, and the tap is chromium-plated brass.
A positive rocker switch controls power to the instrument, and is illuminated when depressed to the I (ON) position. The dial controlling the thermostat setting is graduated to give a guide to the wax temperature, and is fitted with a maximum stop at 70°C. An amber lamp indicates operation of the reservoir heating elements via the thermostat. A red lamp shows you when the thermal safety cut-out has operated.
Note The thermal safety cut-out is a non-resetting secondary thermostat fitted to protect the instrument from overheating.
The power cable has moulded plugs at both terminals. The power inlet socket at the back of the Wax Dispenser has fuses on both the negative line and the positive line for optimum protection. It also incorporates a filter which prevents radio interference.
16 Wax Dispensers Operator Guide Thermo Fisher Scientific
Do Not:• Use the wax dispenser without wax in the reservoir
• Use abrasives to clean the interior
Do:• Maintain the apparatus in a reasonably clean condition
• Switch off before removing the plug
• Ensure that replacement fuses are of the correct specification
• Use a drip tray
• Use on a level surface or the tap may not hold onto drips
2.2 Cautionary Notes
Thermo Fisher Scientific Wax Dispensers Operator Guide 17
Preliminary Check1. Make sure that the mains rocker switch is in the O (OFF)
position.
2. Ensure there is at least 5cm of molten or pelletized paraffin wax in the reservoir.
3. Connect the Wax Dispenser to the mains supply and depress the mains rocker switch to the I (ON) position.
4. Set the temperature control dial to 4°C above the melting point of the wax. Allow to equilibrate for one hour.
5. The upper aluminium rim of the reservoir and the tap should feel warm after this period and there should be evidence of the wax melting.
Routine operation1. Make sure that the mains rocker switch is in the O (OFF)
position.
2. Ensure that the plugs on both ends of the power lead are pushed fully into their sockets.
3. Part fill the reservoir bowl with molten or pelletized wax.
4. Depress the mains rocker switch to the I (ON) position.
5. Turn the temperature control dial to a mark of 3-4°C above the melting point of the wax.
6. Leave the instrument for 1-2 hours (depending on the level of wax in the reservoir) to allow the temperature to settle.
7. Deliver wax by pulling the lever. The lever has two positions: while in the upright position, twist the lever one way for continuous operation and the other for demand delivery.
2.3 Operating Procedure
18 Wax Dispensers Operator Guide Thermo Fisher Scientific
Do Not oPERAtE thE lEvER whilE thE wAX DisPENsER is ColD, As this MiGht DisloDGE thE PistoN fRoM thE sEAl RECEss
if the red indicator lamp glowsThe red lamp glows when the non-self resetting thermal safety cut-out has been triggered:
1. Switch the Wax Dispenser off and disconnect from the mains. Leave to cool.
2. Consult a Service Engineer.
Thermo Fisher Scientific Wax Dispensers Operator Guide 19
Do Not UsE sCoURiNG PADs oR DEsCAliNG AGENts
if hAZARDoUs MAtERiAl is sPilt oN oR iNsiDE thE DisPENsER, yoU MUst CARRy oUt APPRoPRiAtE ClEANiNG AND DECoNtAMiNAtioN
bEfoRE yoU UsE ANy ClEANiNG MEthoD othER thAN thosE RECoMMENDED iN this GUiDE, yoU MUst ChECK with yoUR AGENt thAt thE PRoPosED MEthoD will Not DAMAGE thE EQUiPMENt
To ensure that the instrument continues to operate efficiently and safely in normal day-to-day use, it should be cleaned regularly.
1. Disconnect the Dispenser from the mains supply before cleaning.
2. Empty the reservoir by locking the delivery tap into the continuous operation position.
3. Dispose of the wax contents according to local regulations.
4. Remove the residual 3mm depth of wax using absorbent tissues.
5. Wipe the reservoir clean.
6. You can clean the filter while it is in position, or remove it first:
With the filter in position, clean it using a toothbrush or similar item.
To remove the filter, slide it sideways and lift it out, then clean it with solvent. In this case, wear protective gloves.
2.4 Cleaning Procedure
20 Wax Dispensers Operator Guide Thermo Fisher Scientific
7. Clean the outer casing, as required, using wax solvents.
Thermo Fisher Scientific Wax Dispensers Operator Guide 21
Chapter 3 High Capacity Wax DispenserThe High Capacity Wax Dispenser is designed to hold up to 10 litres of wax at a controlled temperature. Wax is delivered via a switch operated solenoid valve. A two-part heating element ensures that the total volume of wax in the reservoir is molten but avoids localised overheating.
Coarse debris is prevented from blocking the delivery assembly by means of a 0.5mm mesh at the base of the reservoir. A separate tap heater ensures an even flow of wax.
Figure 2 shows the main features of the High Capacity Wax Dispenser.
Figure 2. The High Capacity Wax Dispenser
3.1 General Description
Temperature controller
Dispensing valve
Mains rocker switch at rear
Dispensing switch
22 Wax Dispensers Operator Guide Thermo Fisher Scientific
The Wax Dispenser is manufactured to operate over a temperature range of 50°C to 70°C.
The outer casing is made from zinc coated steel with a stoved epoxy finish, the lid is made from polished stainless steel and the reservoir and the tap anodised aluminium.
A positive rocker switch at the rear of the unit controls power to the instrument. The controller has buttons for varying the setting. An LED indicates operation of the heating element via the Controller. The unit is fitted with a thermal safety cut-out which become operative if a serious over heat error occurs.
The power cable has moulded plugs at both terminals. The power inlet socket at the rear of the instrument has fuses on both negative and positive lines for optimum protection, it also incorporates a filter preventing radio interference.
Note The thermal safety cut-out is a non-self-resetting secondary thermostat fitted to protect the instrument from overheating.
Thermo Fisher Scientific Wax Dispensers Operator Guide 23
hANDlE with CAUtioN At All tiMEs, As this APPliANCE is liKEly to CoNtAiN A hot liQUiD
bEfoRE MoviNG this APPliANCE REMovE thE PlUG fRoM thE MAiNs sUPPly soCKEt AND Allow wAX to Cool
Do Not: • Use without wax in the reservoir
• Use abrasives to clean the interior
• Use for non-specified purposes without prior consultation with the supplier
Do:• Maintain the apparatus in a reasonably clean condition
• Switch off before removing the plug
• Ensure that replacement fuses are of the correct specification
• Use a drip tray
• Use on a level surface
wE RECoMMEND thAt:• Due to the large volume of wax involved the unit is not
turned off if required during the following 24 hour time period
• The unit is used on a level surface or the tap may not hold onto drips
• The equipment is disconnected from mains supply when not required for use for extended periods
3.2 Cautionary Notes
24 Wax Dispensers Operator Guide Thermo Fisher Scientific
• In the event of mechanical damage to the instrument or power cable the apparatus should be checked by a Service Engineer before use
• No repair should be undertaken except by a trained Service Engineer
Thermo Fisher Scientific Wax Dispensers Operator Guide 25
Preliminary Check1. Make sure that the mains rocker switch is in the O (OFF)
position.
2. Part fill the reservoir bowl with molten or pelletized paraffin wax.
3. Connect the Wax Dispenser to mains supply and depress the rocker switch to the I (ON) position.
4. Set the temperature controller to the required setting (press and hold down the left button and press the Up or Down buttons to change the temperature).
5. Allow to equilibrate for (n) hours (dependent on the level of the wax in the reservoir).
6. When all the wax is molten hold a pellet of wax against the outlet spout. If this melts freely the tap heater is working.
Routine operation1. Ensure that both plugs on the power lead are pushed fully
into their sockets.
2. Part fill the reservoir bowl with molten or pelletized paraffin wax.
3. Depress the mains rocker switch to the I (ON) position.
4. Set the temperature controller to the required setting (press and hold down the left (*) button and press the Up or Down buttons to change the temperature). Allow to equilibrate for (n) hours (dependent on the level of the wax in the reservoir).
5. Wax is delivered by depressing the switch on the front panel.
3.3 Operating Procedure
26 Wax Dispensers Operator Guide Thermo Fisher Scientific
Note Ensure that the wax has completely melted before attempting to dispense wax.
6. The tap may not hold onto drips if the unit is not on a level surface.
thE DElivERy switCh shoUlD Not bE oPERAtED UNtil thE wAX DisPENsER hAs wARMED UP As thE DisPENsiNG MEChANisM MAy bECoME stUCK UNtil wARM ENoUGh
Regular cleaning of the instrument in accordance with the instructions contained within this guide, will ensure that the instrument continues to operate efficiently and safely in normal day to day use. Contact your agent if any part of the instrument appears to be damaged in any way.
in the event a fault occurs1. Switch the Wax Dispenser off and disconnect from the
mains.
2. Consult a Service Engineer.
Thermo Fisher Scientific Wax Dispensers Operator Guide 27
if hAZARDoUs MAtERiAl is sPilt oN oR iNsiDE thE iNstRUMENt, yoU MUst CARRy oUt APPRoPRiAtE ClEANiNG AND DECoNtAMiNAtioN
bEfoRE yoU UsE ANy ClEANiNG MEthoD othER thAN thosE RECoMMENDED iN this GUiDE, yoU MUst ChECK with yoUR AGENt thAt thE PRoPosED MEthoD will Not DAMAGE thE EQUiPMENt
The instrument must be disconnected from the mains for cleaning.
1. As required, the reservoir can be emptied by operation of the delivery switch. Do not leave the unit unattended whilst the tap is on as when empty the cut out may operate.
2. Dispose of the wax contents according to local regulations.
3. The residual 5mm depth of wax can be removed using absorbent tissues and the reservoir wiped clean.
4. The filter can be cleaned in situ using a toothbrush or similar item, or slid out for solvent cleaning.
5. The outer casing can be cleaned as required using a mild detergent if necessary.
sCoURiNG PADs oR DEsCAliNG AGENts MUst Not bE UsED
3.4 Cleaning Procedure
28 Wax Dispensers Operator Guide Thermo Fisher Scientific
Chapter 4 Slimline Wax Dispenser The Slimline Wax Dispenser is designed to hold up to 5 litres of wax at a controlled temperature. The prime role of this instrument is in the preparation of larger wax blocks especially where a high flow rate is essential.
Wax is delivered via a two-position tap: the first position allows wax to be dispensed only whilst the lever is held down and the second position allows the lever to be locked open for continuous flow (this may require turning the lever).
Coarse debris is prevented from blocking the delivery assembly by means of a 0.5mm mesh at the base of the reservoir. A separate tap heater ensures an even flow of wax.
Figure 3 shows the main features of the Slimline Wax Dispenser.
Figure 3. The Slimline Wax Dispenser
4.1 General Description
Stainless steel lid
Temperature controller
Heated dispensing tap
Mains power switch and fused inlet at rear
Electrical inspectionat rear
Thermo Fisher Scientific Wax Dispensers Operator Guide 29
The Slimline Wax Dispenser is manufactured to operate within a temperature range of 50°C to 70°C. The outer casing is made from epoxy-coated steel. The reservoir and lid are made from stainless steel, whilst the tap is chromium-plated brass.
A positive rocker switch at the rear controls power to the instrument, up is the I (ON) position. The Controller has a dual LED display showing the Set temperature in green and the sensed temperature in red and has a maximum setting of 70°C. Red indicators show the operation of the reservoir heating elements.
The instrument is fitted with a thermal safety cut-out which will cut all power if an overheat fault occurs (see the note below).
The power cable has moulded plugs at both terminals. The power inlet socket at the rear of the instrument has fuses on both negative and positive lines for optimum protection; it also incorporates a filter preventing radio interference.
Note The thermal safety cut-out is a non-self resetting secondary thermostat fitted to protect the instrument from overheating.
30 Wax Dispensers Operator Guide Thermo Fisher Scientific
hANDlE thE sliMliNE wAX DisPENsER with CAUtioN At All tiMEs, As this APPliANCE is liKEly to CoNtAiN A hot liQUiD
bEfoRE MoviNG this APPliANCE REMovE thE PlUG fRoM thE MAiNs sUPPly soCKEt AND Allow wAX to Cool foR 1 hoUR
Do Not:• Use without wax in the reservoir
• Use abrasives to clean the interior
• Use for non-specified purposes without prior consultation with the supplier
• Operate the tap while the wax is cold as this may dislodge the piston from the seal recess
• Touch the inside of the bowl, it can be hot
Do:• Maintain the apparatus in a reasonably clean condition
• Switch off before removing the plug
• Ensure that replacement fuses are of the correct specification
• Use in a safe and stable location where the tap cannot be knocked by accident
4.2 Cautionary Notes
Thermo Fisher Scientific Wax Dispensers Operator Guide 31
wE RECoMMEND thAt:• The instrument should be disconnected from mains supply
when not required for use
• In the event of mechanical damage to the instrument or power cable the apparatus should be checked by a service engineer before use
• No repair should be undertaken except by a trained Service Engineer
32 Wax Dispensers Operator Guide Thermo Fisher Scientific
setting the temperatureTo set the temperature controller to the required temperature:
1. Switch the Wax dispenser on using the mains rocker switch at the rear of the instrument.
2. Press the P key and release it to display a flashing ‘sP1’.
3. To set the temperature press the Up arrow key to increase the value or the Down arrow key to decrease it. These keys increase or decrease the value one digit at a time, but if the button is pressed for more than one second the value increases or decreases rapidly.
4. Press P to confirm the temperature setting.
Preliminary Check1. Connect the power cable to the instrument, but do not
connect to the power supply.
2. Part fill the reservoir bowl with molten or pelletized paraffin wax.
3. Connect the power cable to a properly earthed power supply.
4. Depress the mains rocker switch at the rear of the instrument to the I (ON) position.
5. Set the temperature control to 50°C (see Setting the Temperature). Allow the instrument time to heat the wax, depending on the amount of wax in the reservoir bowl.
6. The wax should start to melt.
4.3 Operating Procedure
Thermo Fisher Scientific Wax Dispensers Operator Guide 33
Routine operation1. Ensure that both plugs on the power lead are pushed fully
into their sockets.
2. Part fill the reservoir bowl with molten or pelletized paraffin wax.
3. Connect the power cable to a properly earthed power supply.
4. Depress the mains rocker switch at the rear of the instrument to the I (ON) position.
5. Set the temperature control to a mark of 3-4°C above the melting point of the wax. Allow to equilibrate for 1-2 hours (dependent on the level of the wax in the reservoir).
6. Wax is delivered by pulling the lever. This has two positions: for continuous operation or demand delivery.
Note: The lever should not be operated while the Wax Dispenser is cold as this may dislodge the piston from the seal recess.
It should be noted that incorrect use by the operator may introduce a hazard to health or may damage the equipment.
Regular cleaning of the instrument in accordance with the instructions contained within this guide will ensure that the instrument continues to operate efficiently and safely in normal day-to-day use. Contact your supplier if any part of the instrument appears to be damaged in any way.
34 Wax Dispensers Operator Guide Thermo Fisher Scientific
if the wax dispenser power shuts downThe Unit will power down when the thermal safety cut-out has been triggered:
1. Switch the Wax Dispenser off and disconnect from the mains. Leave to cool.
2. Consult a Service Engineer.
Thermo Fisher Scientific Wax Dispensers Operator Guide 35
if hAZARDoUs MAtERiAl is sPilt oN oR iNsiDE thE DRyER, yoU MUst CARRy oUt APPRoPRiAtE ClEANiNG AND DECoNtAMiNAtioN
bEfoRE yoU UsE ANy ClEANiNG MEthoD othER thAN thosE RECoMMENDED iN this GUiDE, yoU MUst ChECK with yoUR AGENt thAt thE PRoPosED MEthoD will Not DAMAGE thE EQUiPMENt
The instrument must be disconnected from the mains supply for cleaning.
1. As required, the reservoir can be emptied by locking the delivery tap into the continuous operation position.
2. Dispose of the wax contents according to local regulations.
3. The residual 3mm depth of wax can be removed using absorbent tissues and the reservoir wiped clean.
4. The filter can be cleaned in situ using a toothbrush or similar item, or slid sideways and lifted out for solvent cleaning (if removing the filter we recommend protective gloves are worn).
5. The outer casing can be cleaned as required using a mild detergent if necessary.
sCoURiNG PADs oR DE-sCAliNG AGENts ARE Not RECoMMENDED
4.4 Cleaning Procedure
36 Wax Dispensers Operator Guide Thermo Fisher Scientific
Appendix A Technical Specification
A1. Wax Dispenser
Dimensions (O.D. x H) Weight
21.5 x 41cm 3.7kg
Electrical Voltages Frequency Power
110–120V (~) 220–240V (~) 50/60 Hz 250 VA
Maximum supply voltage fluctuation must not exceed 10% of nominal
Fuses Mains plug Mains input
5A 240V (where applicable) 230V (~) units F1.6A H250V 115V (~) units F3.15A H250V
Note: Fuses must only be replaced by a technically competent person
Switch conventions I O
Power On Power Off
Environmental Requirements
General Indoor use only
Temperature (operating limits)
+5°C to +40°C (+41°F to +104°F)
Temperature (Transport & Storage)
-25°C to +55°C (-13°F to 131°F)+70°C (158°F) for short exposure
Relative Humidity Max. 80% RH up to 31°CDecreasing linearly to 50% RH at 40°C
Altitude Up to 2000 m (6,500 ft)
Pollution Degree 2
Over Voltage II
Thermo Fisher Scientific Wax Dispensers Operator Guide 37
A2. High Capacity Wax Dispenser
Physical (W x D x H)
CapacityWeight
275mm x 385mm x 525mmTap from bench: Height 150mm10 litres8.5kg
Electrical Voltages Frequency Power
110-120 V (~)220-240 V (~)50/60 Hz460 VA
Maximum supply voltage fluctuation must not exceed 10% of nominal
Fuses Mains plug Mains input
5A 240v (where applicable)230V (~) F3.15A H250V 115V (~) F6.3A H250V
Note: Fuses must only be replaced by a technically competent person
Switch conventions I O
Power On Power Off
Environmental Requirements General Indoor use only
Temperature (operating limits)
+5°C to +40°C (+41°F to +104°F)
Temperature (Transport & Storage)
-25°C to +55°C (-13°F to 131°F)+70°C (158°F) for short exposure
Relative Humidity Max. 80% RH up to 31°CDecreasing linearly to 50% RH at 40°C
Altitude Up to 2000 m (6,500 ft)
Pollution Degree 2
Over Voltage II
38 Wax Dispensers Operator Guide Thermo Fisher Scientific
A3. Slimline Wax Dispenser
Physical (D x W x H)
CapacityWeight
425mm (inc. tap) x 168mm x 405mmTap from bench: Height 135mm5 litres (approx)6.8kg
Electrical Voltages Frequency Power
110-120 V (~)220-240 V (~)50/60 Hz200 VA
Maximum supply voltage fluctuation must not exceed 10% of nominal
Fuses Mains plug Mains input
5A 240V (where applicable)230V (~) Units F1.6A H250V 120V (~) Units F2A H250V
Note: Fuses must only be replaced by a technically competent person
Switch conventions I O
Power On Power Off
Environmental Requirements General Indoor use only
Temperature (operating limits)
+5°C to +40°C (+41°F to +104°F)
Temperature (Transport & Storage)
-25°C to +55°C (-13°F to 131°F)+70°C (158°F) for short exposure
Relative Humidity Max. 80% RH up to 31°CDecreasing linearly to 50% RH at 40°C
Altitude Up to 2000 m (6,500 ft)
Pollution Degree 2
Over Voltage II
Thermo Fisher Scientific Wax Dispensers Operator Guide 39
Index
C
Cautionary notes 16, 23, 30Cleaning procedures 19, 27, 35Common safety warnings 12
D
Drying out recommendation 7
G
General safety 7
H
High capacity wax dispenserDescription 21Main features 21Operating 25Preliminary check 25Set the temperature controller 25Temperature range 22
I
InstrumentsReturning 11
Siting 7Switching On/Off 9Unpacking 10
Introduction 6
M
Mains power supply 8
P
Plug wiring convention 7
S
Safety and Warranty booklet 6Sample safety 6Setting the temperature controller
High capacity wax dispenser 25Slimline wax dispenser 32
Slimline wax dispenserDescription 28Main features 28Preliminary check 32Setting the temperature 32Temperature range 29
Symbols 3
T
Taking out of operation 11Technical specifications 36Thermal cut-out 13
40 Wax Dispensers Operator Guide Thermo Fisher Scientific
U
Unpacking the instruments 10
V
Voltage rating 7
W
Wax dispenserCleaning 19Description 14Main features 14Operating 17Preliminary check 17Temperature range 15