Waters LCMSWaters LCMS Troubleshooting
-
Upload
dangerous0 -
Category
Documents
-
view
270 -
download
5
Transcript of Waters LCMSWaters LCMS Troubleshooting
-
8/12/2019 Waters LCMSWaters LCMS Troubleshooting
1/82
Waters Corp. 20042004 Waters Corporation
Troubleshooting CommonMS Problems
by Claude Mallet, Ph.D
presented by
Michael S. Young, Ph.D.
-
8/12/2019 Waters LCMSWaters LCMS Troubleshooting
2/82
2004 Waters Corporation2004 Waters Corporation
Troubleshooting Common MS Problems
Overview of Troubleshooting Strategy
ESI sources parameters
Single and triple Quadrupoles
SIR vs MRM
Ion Suppression
Outline
-
8/12/2019 Waters LCMSWaters LCMS Troubleshooting
3/82
2004 Waters Corporation2004 Waters Corporation
Try to simplify --
assess impact on lab efficiency --
inspect the MS or /MS/MS --
try to categorizetroubleshoot the easiest to fix
items first
CHEMISTRY MECHANICAL IMPROPER SETTINGS
Adducts (Na+, K+)
Multiple chargeIon stability (pH)
Ion suppression
Ion beam instability
Probe clogging
Heater/sensor
N2 gas flow
Loss of vacuum
Power supply
ESI sources parameters
Quadrupoles parameters
Acquisition modes
MS Troubleshooting Strategy
-
8/12/2019 Waters LCMSWaters LCMS Troubleshooting
4/82
2004 Waters Corporation2004 Waters Corporation
Sample
Preparation ChromatographyMass
Spectrometry
Polarity:
Silica- C18, C8, C4, C2Hybrid- C18, C8, C4, C2Polymer- C
18
, C8
, C4
, C2Embedded polar group
Cyano, Phenyl
Particle size:
2.5, 3.5, 5 or 7 m
Internal diameter:4.6, 3.9, 2.1, 1.0,
0.32 mm and 75 m
Length:
150, 100, 50, 30, 20 mm
Source:
ESI
APcI
Nano-ESI
Mass analyzers:
magnetic sectors
electric sectors
time of flight
quadrupoleion trap
FT-ICR
Raw sample:
- CaCO2, microsomes, P450,
hepatocytes etc
- tissue, CSF, plasma, serumurine, tears etc
- water, sediment, food etc
Extracted sampleFor LC/MS/MS
The Total Analysis
-
8/12/2019 Waters LCMSWaters LCMS Troubleshooting
5/82
2004 Waters Corporation2004 Waters Corporation
BA MassSpectrometry
ESI source parameters
Part 1
-
8/12/2019 Waters LCMSWaters LCMS Troubleshooting
6/822004 Waters Corporation2004 Waters Corporation
Quattro UltimaZQ
Quattro Premier
Mass Spectrometers
-
8/12/2019 Waters LCMSWaters LCMS Troubleshooting
7/822004 Waters Corporation2004 Waters Corporation
First, lets take a look at the
first event in an ESI orthogonal
source. The primary function
of the probe is to transform a
liquid (from LC column or other
source) into a gas stream as
shown in the red circle. Three
parameters are used to
optimize the probe, which are
the 1- nebulizer gas, 2- the
desolvation gas flow and 3-the desolvation temperature.
The nebulizer gas is
automatically set at maximum
on the ZQ and manually on
other mass spectrometers (i.e.
Ultima, QToF, LCT etc).
The desolvation gas flow anddesolvation temperature can
be optimized to maximize
signal intensity. Higher
temperatures are required
when using mobile phases
containing high percentage of
water.
2,31
Mass Spectrometers
-
8/12/2019 Waters LCMSWaters LCMS Troubleshooting
8/822004 Waters Corporation2004 Waters Corporation
Nebulizer gas flow off Nebulizer gas flow on
Notice the formation of a liquid drop. It can lead to
source flooding if unattended for a long period of time.
To avoid potential electrical hazard, the source isequipped with a drain valve.
Notice the formation of liquid droplets from
condensation of the sprayer on the probe holder
assembly.
ESI Probe Parameters
-
8/12/2019 Waters LCMSWaters LCMS Troubleshooting
9/822004 Waters Corporation2004 Waters Corporation
Isolation valve
Cone shield and
cone assembly
Baffle
Stainless steel capillary
Desolvation heater
Ion Block
ESI probe
ESI Source
-
8/12/2019 Waters LCMSWaters LCMS Troubleshooting
10/822004 Waters Corporation2004 Waters Corporation
10 mm
5 mm
Cone
Probe
ESI Probe Parameters
-
8/12/2019 Waters LCMSWaters LCMS Troubleshooting
11/822004 Waters Corporation2004 Waters Corporation
Capillary Tip
Make sure capillary
extends approx. 0.5 mm
beyond probe tip.
Any corrosion, deposit
constriction or other flow
restriction will hinder
proper nebulization.
-
8/12/2019 Waters LCMSWaters LCMS Troubleshooting
12/822004 Waters Corporation2004 Waters Corporation
Probe toofar from the
cone?
Probe
extends toofar past
cone?
ESI Probe Parameters
-
8/12/2019 Waters LCMSWaters LCMS Troubleshooting
13/82
2004 Waters Corporation2004 Waters Corporation
Probe tooclose to the
cone?
ESI Probe Parameters
-
8/12/2019 Waters LCMSWaters LCMS Troubleshooting
14/82
2004 Waters Corporation2004 Waters Corporation
Response ofreserpine shows
a good gaussian
distribution with
baseline
resolution of the
C13 isotopes
Temperature and
gas flow are
parameters thataffect the
desolvation
efficiency of the
probe. Improper
settings can result in
loss of signal. Thesevalues are optimized
according to the
column flow rate.
ESI Probe Parameters (tune page)
-
8/12/2019 Waters LCMSWaters LCMS Troubleshooting
15/82
2004 Waters Corporation2004 Waters Corporation
In this case, a too low
desolvation
temperature resulted
in a 50 % reduction in
signal intensity.
This effect is
compound dependent.
Desolvation Temperature
-
8/12/2019 Waters LCMSWaters LCMS Troubleshooting
16/82
2004 Waters Corporation2004 Waters Corporation
Similar loss of signal
intensity, in this case,
it is due to a too low
setting of thedesolvation gas (113
vs 550 L/hr). The
gas used for the
desolvation is
nitrogen
It must be of high
purity (99.95%) and
oil free. (traps can be
used to increase the
gas purity if needed).
Make sure delivery
pressure is regulated
to 100 psi.
Desolvation Gas Flow
-
8/12/2019 Waters LCMSWaters LCMS Troubleshooting
17/82
2004 Waters Corporation2004 Waters Corporation
Column flow rate Desolvation temp Desolvation gas flow
L/min C liters/hr
< 10 100 to 120 200 to 250
10 to 20 120 to 250 250 to 400
20 to 50 250 to 350 250 to 400
>50 350 to 400 400 to 750
Higher desolvation temperatures give increased sensitivity. However, increasing thetemperature above the range suggested reduces beam stability. Increasing the gas
flow rate higher that the quoted values lead to unnecessary high nitrogen consumption.
Avoid operating the desolvation heater for long periods of time without proper gas flow.
To do so could damage the source.
Suggested Settings
-
8/12/2019 Waters LCMSWaters LCMS Troubleshooting
18/82
2004 Waters Corporation2004 Waters Corporation
Dewar tanks Nitrogen generator
Both setups are widely used and the choice mostly depends on the consumption ofnitrogen per day. Larger laboratories will have a tendency to choose the nitrogen
generator for convenience and cost for long term operation.
Dewar Tank vs Nitrogen Generator
-
8/12/2019 Waters LCMSWaters LCMS Troubleshooting
19/82
2004 Waters Corporation2004 Waters Corporation
At this point lets take a look at the
second event. Once a spray is
stable, ions are produced and
directed toward the mass analyzer.
Five parameters in the orthogonal
source are used for this purpose.These parameters are: 1- capillary
voltage, 2- cone voltage, 3-
extraction voltage and 4- RF lens
(transfer optics) 5- Source
temperature. A high voltage, in
the kV range, is applied to astainless steel capillary tubing in
the probe. This will produce
charged droplets. With the
assistance of the desolvation gas
flow and desolvation temperature,
those droplets will in turn produce
ions in gas phase next to the cone.
The cone voltage attracts positively
charged ions from the spray into a
reduced pressure chamber (ion
block). The extractor and RF lensare used to guide the ion beam into
the mass analyzer
1 2
34
ESI Probe
ESI Source
ESI Source Parameters
-
8/12/2019 Waters LCMSWaters LCMS Troubleshooting
20/82
2004 Waters Corporation2004 Waters Corporation
Clean cone and cone shield
Notice the white residue on the cone
shield, but the aperture of the cone is
still clear. This is an indication that
samples injected on this MS were notclean. In both pictures, the baffle
shows brown spots, which indicates
routine and normal usage. The white
residue can result from long exposure
to poorly prepared samples or from
nonvolatile mobile phase additives.Over time, the aperture of the cone will
become clogged, thus reducing signal
intensity.cone
baffle
Brown spot
ESI Source Parameters
-
8/12/2019 Waters LCMSWaters LCMS Troubleshooting
21/82
2004 Waters Corporation2004 Waters Corporation
These are typical
starting values to
obtain a stable ion
beam with flow
rate ranging from
0.2 to 0.4 ml/min.
The ion block is
heated to avoid
any condensationproblems.
The source has a
maximum setting
of 150 C.
ESI Source Parameters
-
8/12/2019 Waters LCMSWaters LCMS Troubleshooting
22/82
2004 Waters Corporation2004 Waters Corporation
With insufficient
capillary voltage, the
signal shows an 80%
decrease in signalintensity.
Typical optimum
values for most small
molecules are
between 3.0 and 3.5kV.
Higher values usually
have little effect on
signal intensity.
Deviations from
experimentally
optimized value may
indicate problems in
the source.
Capillary Voltage
-
8/12/2019 Waters LCMSWaters LCMS Troubleshooting
23/82
2004 Waters Corporation2004 Waters Corporation
The cone voltage is
applied to a spherical
metal plate, the first gate
between the sprayer (atatmospheric pressure)
and the inside of the
mass analyzer (at 10-6
Torr of pressure). The
cone creates the first
bend of the ion beam in
the orthogonal source.
This slide shows that we
have optimized the cone
voltage at 35 volts andincreased our signal
intensity.
Cone Voltage
-
8/12/2019 Waters LCMSWaters LCMS Troubleshooting
24/82
2004 Waters Corporation2004 Waters Corporation
Poor response can
occur if cone voltage
is set too low. A
sufficient voltage is
required to atract a
high population of
ions into the ion
block.
Once the conevolatage is optimized,
loss of sensitivity may
result from
contamination at the
cone.
Cone Voltage
-
8/12/2019 Waters LCMSWaters LCMS Troubleshooting
25/82
2004 Waters Corporation2004 Waters Corporation
Poor response can also
occur if cone voltage is set
too high. Too much energy
causes a phenomenon
known as In-source
fragmentation.
When ions are accelerated
from the sprayer to the ion
block with very highvelocities, collisions among
ions can create a high
population of daughter ions
at the expense of parent
ions.
In this case, the ion at m/z
609 shows a 90% reduction
in signal intensity.
Cone Voltage
-
8/12/2019 Waters LCMSWaters LCMS Troubleshooting
26/82
2004 Waters Corporation2004 Waters Corporation
The extractor voltageis applied to a second
cone shaped metal
plate that separates
the ion block and the
mass analyzer. This
plate creates a second90 degree angle in the
ion beam, completing
the Z spray shape.
An incorrect voltagesetting of the extractor
resulted in a 70%
reduction in signal
intensity.
Typical extractorvoltage settings range
from 1 to 3 volts;
higher values will not
usually give better
sensitivity. Higher than
expected values mayindicate contamination
in the source block
Extractor Voltage
-
8/12/2019 Waters LCMSWaters LCMS Troubleshooting
27/82
2004 Waters Corporation2004 Waters Corporation
The RF lens focuses
the ion beam as it
passes into the mass
analyzer. In the
tandem mass-spectrometer, it
focuses the beam to
the center of the
transition lens
hexapole assembly.
The RF lens value
should typically be set
to range from 0.1 to
0.5 volts.
RF Lens
-
8/12/2019 Waters LCMSWaters LCMS Troubleshooting
28/82
2004 Waters Corporation2004 Waters Corporation
3.5
3.5
3. 3.
In the example shown,
we needed to increase
the RF lens to achieve
a symmetrical peakshape. This may
indicate that the
source is
contaminated.
RF Lens
-
8/12/2019 Waters LCMSWaters LCMS Troubleshooting
29/82
2004 Waters Corporation2004 Waters Corporation
BA MassSpectrometry
ESI source parameters
Quadrupoles
Part 2
-
8/12/2019 Waters LCMSWaters LCMS Troubleshooting
30/82
2004 Waters Corporation2004 Waters Corporation
The quadrupole mass analyzer, like other type of mass analyzers (I.e. ToF, ion traps, sector etc)
separates ions according to their mass to charge ratio (m/z). The quadrupole is made of 4 highly
polished metal rods positioned at precise angles from one another. These rods are connected to high
voltage power supply (DC, positive/negative) and a radio frequency (RF) generator. The slope ofRF/DC applied to the rods is proportional to a range or a specific mass to charge ratio.
Quadrupole Mass Analyzers
-
8/12/2019 Waters LCMSWaters LCMS Troubleshooting
31/82
2004 Waters Corporation2004 Waters Corporation
Source
DetectorNonresonant
Ion
Resonant Ion
dc and rf voltages
+Udc + V cost
-Udc V cost
Molybdenum Alloy
Quadrupole Schematic
-
8/12/2019 Waters LCMSWaters LCMS Troubleshooting
32/82
2004 Waters Corporation2004 Waters Corporation
Stable ion
Non-resonant
Trajectory
Pre-Filters
Resonant
Trajectory
Quadrupole
Resonant vsNon-Resonant Trajectory
-
8/12/2019 Waters LCMSWaters LCMS Troubleshooting
33/82
2004 Waters Corporation2004 Waters Corporation
Quadrupole Unit Mass Resolution
1
-
8/12/2019 Waters LCMSWaters LCMS Troubleshooting
34/82
2004 Waters Corporation2004 Waters Corporation
572.8339
570 571 572 573 574 575 576 577m/z0
100
%
0
100
%
573.9185
574.8116
573.2997
574.3072
575.3155
QuadrupoleResolution: 1000
Q-ToF
Resolution: 10 000
[M+H]+
Isotopes
Bradykinin Frag 1-5:Arg-Pro-Pro-Gly-Phe
-
8/12/2019 Waters LCMSWaters LCMS Troubleshooting
35/82
L d
-
8/12/2019 Waters LCMSWaters LCMS Troubleshooting
36/82
2004 Waters Corporation2004 Waters Corporation
The low and high
mass resolution are
arbitrary values that
are calculated from
the RF/DC ratio.
The LM setting
affects the resolution
of ions at the low
mass range of the
quadrupole; the HM
setting at the high
mass range of the
quadrupole.
The quadrupole canonly achieve mass
unit resolution,
which means that
multiple charged
peaks are not fully
resolved.
Low andHigh Mass Resolution
-
8/12/2019 Waters LCMSWaters LCMS Troubleshooting
37/82
2004 Waters Corporation2004 Waters Corporation
.
Udc
V
(DC voltage)
(Rfvoltage)
Correct V/U ratio
mass 1 & 2 are resolved
R: 1000
V/U slope
Stable trajectory
Unstable trajectory
Quadrupole Stability Diagram
-
8/12/2019 Waters LCMSWaters LCMS Troubleshooting
38/82
-
8/12/2019 Waters LCMSWaters LCMS Troubleshooting
39/82
2004 Waters Corporation2004 Waters Corporation
.
Udc
V
(DC voltage)
(Rfvoltage)
Low V/U ratio
mass 1 & 2 merge together
R: 10V/U slope
Stable trajectory
Unstable trajectory
Quadrupole Stability Diagram
-
8/12/2019 Waters LCMSWaters LCMS Troubleshooting
40/82
2004 Waters Corporation2004 Waters Corporation
In this case, the LH
and HM resolution
were set too high.The ion beam falls in
the nonresonant
portion of the stability
diagram shown
earlier.
Under these
conditions, the ion
beam will not reach
the multiplier at the
back of the massspectrometer and
produce a signal.
Low and High Mass Resolution
-
8/12/2019 Waters LCMSWaters LCMS Troubleshooting
41/82
2004 Waters Corporation2004 Waters Corporation
.
Udc
V
(DC voltage)
(Rfvoltage)Incorrect V/U ratio
mass 1 & 2 are over resolvedV/U slope
Unstable trajectory
Stable trajectory
Quadrupole Stability Diagram
-
8/12/2019 Waters LCMSWaters LCMS Troubleshooting
42/82
2004 Waters Corporation2004 Waters Corporation
The ion energy is
applied to a small
lens positioned
between the
quadrupole and
the multiplier. This
lens is used to
refocus the beamtoward the
multiplier. Typical
values range from
0.3 to 0.6.
As shown here,higher values will
produce distortion
and loss of
resolution between
the peak and
isotopes.
Ion Energy
-
8/12/2019 Waters LCMSWaters LCMS Troubleshooting
43/82
2004 Waters Corporation2004 Waters Corporation
The multiplier is the
last step in the signal
production. The ions
produced by the ESI
source and filtered bythe quadrupole are
converted by the
multiplier into a
measurable current.
If the multiplier is settoo low, as shown
here, the signal
intensity will be
considerably reduced.
Too high a multipliersetting produces
saturation (flat-top
peaks) and poor
quantitation.
Multiplier
-
8/12/2019 Waters LCMSWaters LCMS Troubleshooting
44/82
2004 Waters Corporation2004 Waters Corporation
BA MassSpectrometry
MS/MS
Part 3
-
8/12/2019 Waters LCMSWaters LCMS Troubleshooting
45/82
2004 Waters Corporation2004 Waters Corporation
Single Ion Recording
(SIR Mode)
Static
Full Scan
(MS mode)
Scanning
LOQ = 500 pg (quantity injected) LOQ = 5 pg (quantity injected)
Note: A quadrupole mass spectrometer is typically available with amass range of 2000 Daltons or 4000 Daltons
Single Q Mode of Acquisition
-
8/12/2019 Waters LCMSWaters LCMS Troubleshooting
46/82
2004 Waters Corporation2004 Waters Corporation
Full Scan Acquisition
-
8/12/2019 Waters LCMSWaters LCMS Troubleshooting
47/82
2004 Waters Corporation2004 Waters Corporation
0.00 0.25 0.50 0.75 1.00 1.25 1.50 1.75 2.00 2.25 2.50 2.75 3.00 3.25 3.50 3.75 4.00Time0
100
%
Scan ES+TIC
2.06e9
2.76
0.00 0.25 0.50 0.75 1.00 1.25 1.50 1.75 2.00 2.25 2.50 2.75 3.00 3.25 3.50 3.75 4.00Time0
100
%
Scan ES+TIC
3.12e9
2.76
2.56
3.182.98
3.29
0.00 0.25 0.50 0.75 1.00 1.25 1.50 1.75 2.00 2.25 2.50 2.75 3.00 3.25 3.50 3.75 4.00Time0
100
%
Scan ES+TIC
5.60e93.18
2.962.54 2.743.27
[ ] = 50 ng/mL
[ ] = 5 ng/mL mixture of 5 basic compounds
[ ] = 500 ng/mL
Full Scan Acquisition
-
8/12/2019 Waters LCMSWaters LCMS Troubleshooting
48/82
2004 Waters Corporation2004 Waters Corporation
Single Ion Recording (SIR)
-
8/12/2019 Waters LCMSWaters LCMS Troubleshooting
49/82
2004 Waters Corporation2004 Waters Corporation
0.00 0.25 0.50 0.75 1.00 1.25 1.50 1.75 2.00 2.25 2.50 2.75 3.00 3.25 3.50 3.75 4.00Time0
100
%
SIR of 5 Channels ES+TIC
3.77e7
2.76
2.55
0.85 2.26
3.182.97
3.28
0.00 0.25 0.50 0.75 1.00 1.25 1.50 1.75 2.00 2.25 2.50 2.75 3.00 3.25 3.50 3.75 4.00Time0
100
%
Scan ES+TIC
2.06e9
[ ] = 5 ng/mL
[ ] = 5 ng/mL
Scan mode
SIR mode
Single Ion Recording (SIR)
Tandem Mass Spectrometry
-
8/12/2019 Waters LCMSWaters LCMS Troubleshooting
50/82
2004 Waters Corporation2004 Waters Corporation
A single quadrupole mass analyzer can be operated in two distinct modes, SCAN and SIR. A triple
quadrupole mass spectrometer can offer 4 types of acquisition; 1- Daughter scan, 2- Multiple Reaction
Monitoring (MRM), 3- Parent scan and 4- Constant neutral loss or gain scan. These types of scans
rely on the middle quadrupole called the collision cell. The collision cell is in fact a hexapole (6 rods) that
operates in RF mode only (no resolution capacity). The cell can be pressurize with argon gas. This
provides a physical surface onto which ions filtered by MS1 can be fragmented by collision, hence the termcollision induced dissociation. Depending if MS1 and MS2 are set in scan or park mode will determine the
desired type of acquisition mentioned earlier.
Tandem Mass Spectrometry
Collision Induced Dissociation (CID)
-
8/12/2019 Waters LCMSWaters LCMS Troubleshooting
51/82
2004 Waters Corporation2004 Waters Corporation
Parent Ion Scanning
MS1 MS2
Collision
Cell
StaticScanning
Triple Q Modes of Acquisition
-
8/12/2019 Waters LCMSWaters LCMS Troubleshooting
52/82
2004 Waters Corporation2004 Waters Corporation
Daughter Ion Scanning
MS1 MS2
Collision
Cell
Static Scanning
Triple Q Modes of Acquisition
-
8/12/2019 Waters LCMSWaters LCMS Troubleshooting
53/82
2004 Waters Corporation2004 Waters Corporation
Constant Neutral Loss or Gain
MS1 MS2
Collision
Cell
ScanningScanning
Triple Q Modes of Acquisition
-
8/12/2019 Waters LCMSWaters LCMS Troubleshooting
54/82
2004 Waters Corporation2004 Waters Corporation
Multiple Reaction Monitoring
MS1 MS2
Collision
Cell
Static Static
Triple Q Modes of Acquisition
-
8/12/2019 Waters LCMSWaters LCMS Troubleshooting
55/82
2004 Waters Corporation2004 Waters Corporation
MS1 MS2Daughter
MS1
CID
MS2
A triple quadrupole mass
spectrometer offers lower
sensitivity and
reproducable
fragmentation. With
Multiple ReactionMonitoring (MRM), up to
1000x in sensitivity can be
achieved in comparison to
scan mode. The next
slides will describe some
of the common problems
associated with MRM and
a guide on how to optimize
MRM transitions.
We infused a basic drug(clemastine) and opened
windows for MS1, daughter
and MS2. Notice the mass
unit resolution of the
parent mass and isotopes
on both MS1 and MS2.
Erratum: the optimized values of IE1 and IE2 are 0.4 and 0.8 respectively
Optimizing an MRM
-
8/12/2019 Waters LCMSWaters LCMS Troubleshooting
56/82
2004 Waters Corporation2004 Waters Corporation
Next, the LM/HM (1)
values are lowered to the
point that the first isotope
and the parent ion are
both passed into the
collision cell. The peaks inthe MS1 windows (2) will
broaden and show loss of
resolution. Conequently,
the ion beam passing from
MS1
to the collision cell is
also increased (3). In the
daughter scan window
(middle window in the tune
page), the parent peak is
offscale and one isotope
of the molecule is evident.Since MS2 is set with unit
mass resolution setting
(LM/HM = 15 ), good
resolution is seen in the
third window among the
parent peak and theisotopes.
1
2
3
Erratum: the optimized values of IE1 and IE2 are 0.4 and 0.8 respectively
Optimizing an MRM
-
8/12/2019 Waters LCMSWaters LCMS Troubleshooting
57/82
2004 Waters Corporation2004 Waters Corporation
1
2
In this slide, LM/HM on MS1is slightly increased ( small
gain in resolution) just to the
point that the isotope is notseen. This step is crucial, if
LM/HM on MS1 are too low,
additional ions will enter the
collision cell and will create
additional daughter ions foreach isotope of the parent
molecule.
Erratum: the optimized values of IE1 and IE2 are 0.4 and 0.8 respectively
Optimizing an MRM
-
8/12/2019 Waters LCMSWaters LCMS Troubleshooting
58/82
2004 Waters Corporation2004 Waters Corporation
12
3
Then, by decreasing the
LM/HM on MS2 (1), the signal in
the daughter scan window has
increased (3). The resolution
on MS2 also decreases as aconsequence of lowering the
LM/HM values (2).
Erratum: the optimized values of IE1 and IE2 are 0.4 and 0.8 respectively
Optimizing an MRM
-
8/12/2019 Waters LCMSWaters LCMS Troubleshooting
59/82
2004 Waters Corporation2004 Waters Corporation
Lets take a look at a common
problem when optimizing an MRM
transition. If we look at the LM/HM
values (1,2) on both MS1 and MS2,
the quadrupoles are set at unit
mass resolution. This can be
verified in the MS1 and MS2window in the tune page. The
peaks shows a gaussian
distribution and resolution with the
isotopes. However, the daughter
window in the tune page shows no
signal (4). The answer is quitesimple; choosing a correct MRM
transition also requires us to park
MS1 on the top of the parent peak.
The parent peak has a molecular
weight of 344.2 Da (see previous
slide). In this example, the setting
was incorrect, 343.7 Da. The
difference of 0.5 Da (3) was
enough to miss the parent peak
completely in MS1, thus leading to
a total loss of signal in MS2.
1
2
3
4
Erratum: the optimized values of IE1 and IE2 are 0.4 and 0.8 respectively
Optimizing an MRM
-
8/12/2019 Waters LCMSWaters LCMS Troubleshooting
60/82
2004 Waters Corporation2004 Waters Corporation
1
2
3
4
5
At this point, the quadrupoles
are optimized to give maximum
signal intensity (1,2) and MS1 is
correctly set at 344,2 Daltons
(4). In this tune page both MS1and MS2 windows wereremoved so we can
concentrate on the daughter
ion scan (5). As we can see,
the tune page only shows the
parent ion without any daughterions. This is because the
collision gas was not activated
(6) and the collision cell was
not optimized to produce
daughters ions from collision
with argon gas (3).
6
Erratum: the optimized values of IE1 and IE2 are 0.4 and 0.8 respectively
Optimizing an MRM
-
8/12/2019 Waters LCMSWaters LCMS Troubleshooting
61/82
2004 Waters Corporation2004 Waters Corporation
Prior to introduction of
collision cell gas (1). the
pressure on the collision
cell pirani guage
indicates 1.0 e-4 mbar(2). Also, since there is
no argon gas in the
collision cell, the analyzer
penning guage should
show a pressure in the
vicinity of 1-2 e-5 mbar
(3). This pressure
indicates that the entire
mass analyzer is under
optimum vacuum.
12
3
Optimizing an MRM
Optimizing an MRM
-
8/12/2019 Waters LCMSWaters LCMS Troubleshooting
62/82
2004 Waters Corporation2004 Waters Corporation
1
2
When the collision gas
button is activated (1),
argon gas will flow
freely into the collision
cell located inside the
mass analyzer(between MS1 and
MS2). Notice that the
pressure on the
collision cell gage will
increase (2), typicalvalues are between 2
to 3e-4 mbar.
p gTune Page with Gas Cell Pressure
Optimizing an MRM
-
8/12/2019 Waters LCMSWaters LCMS Troubleshooting
63/82
2004 Waters Corporation2004 Waters Corporation
1 2
3
Once the argon gas pressure is
optimized in the collision cell, it
requires some energy to
produce fragments. In this case,
the collision energy is set at 15
(1) (arbitrary units). The result isthe production of two major
fragments at 215 Da and 128 Da
(2) of the parent ion of mass
344.20 Da. Notice that the
energy level is still low enoughto see a small fraction of the
parent ion (3).
Erratum: the optimized values of IE1 and IE2 are 0.4 and 0.8 respectively
p gSetting Collison Cell Energy
Optimizing an MRM
-
8/12/2019 Waters LCMSWaters LCMS Troubleshooting
64/82
2004 Waters Corporation2004 Waters Corporation
1
2
3
In this scenario, the collision
energy was purposely
increased to higher values (1)
that gives a 100 % conversion
of the parent ion (3) into
fragments ions. However, thelevel of energy is also high
enough to produce further
fragmentation of smaller
daughter ions (2) and to reduce
the intensity to the larger
fragments. This type of settingis not favored for trace
analysis. The optimum for
sensitivity is to use conditions
that will produce a 100 %
conversion of the parent ion
into one or two majors
fragments. Hhowever, the
production of more than two
fragments may be desirable for
verification of unknowns.
Erratum: the optimized values of IE1 and IE2 are 0.4 and 0.8 respectively
p gSetting Collison Cell Energy
-
8/12/2019 Waters LCMSWaters LCMS Troubleshooting
65/82
2004 Waters Corporation2004 Waters Corporation
100 120 140 160 180 200 220 240 260 280 300 320 340 360 380 400m/z0
100
%
0
100
%
0
100
%
344.2
215128
215
128
344.2
215
128
CID 0 volts
CID 10 volts
CID 20 volts
[M+H]+
NCH3
O
CH3
Cl215
128
Clemastine
(Different scale)
Daughter Ion Spectrum
-
8/12/2019 Waters LCMSWaters LCMS Troubleshooting
66/82
2004 Waters Corporation2004 Waters Corporation
Note: typical value of dwell times are between 0.2 and 0.05 seconds
Multiple MRM
M l i l MRM
-
8/12/2019 Waters LCMSWaters LCMS Troubleshooting
67/82
2004 Waters Corporation2004 Waters Corporation
0.00 0.25 0.50 0.75 1.00 1.25 1.50 1.75 2.00 2.25 2.50 2.75 3.00 3.25 3.50 3.75 4.00Time0
100
%
MRM of 5 Channels ES+TIC
2.91e5
2.95
2.54 2.76
3.18
3.27
0.00 0.25 0.50 0.75 1.00 1.25 1.50 1.75 2.00 2.25 2.50 2.75 3.00 3.25 3.50 3.75 4.00Time0
100
%
SIR of 5 Channels ES+TIC
2.29e6
2.272.75
2.56
3.18
2.96 3.27
0.00 0.25 0.50 0.75 1.00 1.25 1.50 1.75 2.00 2.25 2.50 2.75 3.00 3.25 3.50 3.75 4.00Time0
100
%
Scan ES+TIC
2.04e9
[ ] = 0.1 ng/mL
Scan mode
[ ] = 0.1 ng/mLSIR mode
[ ] = 0.1 ng/mLMRM mode
Multiple MRM
P t 4
-
8/12/2019 Waters LCMSWaters LCMS Troubleshooting
68/82
2004 Waters Corporation2004 Waters Corporation
BA MassSpectrometry
Ion suppression
Causes of Ion SuppressionTroubleshooting Ion Supression
Part 4
-
8/12/2019 Waters LCMSWaters LCMS Troubleshooting
69/82
-
8/12/2019 Waters LCMSWaters LCMS Troubleshooting
70/82
2004 Waters Corporation2004 Waters Corporation
Various type of additives can increase or decrease the signal of a target analyte. Furthermore, since
ESI is compound dependent, it is expected to see variation in signal intensity as well as suppression
or enhancement effect. At this point, lets take a look at common additives used in LC and the
response profile of various SPE extraction protocols.
Acidic additive Buffers SPE extracts
Trifluoroacetic acid Ammonium formate protein precipitation
Acetic acid Ammonium bicarbonate Oasis HLB 1-D
Formic Acid Ammonium biphosphate Oasis HLB 2-D
Oasis MCX
Basic additive Ion pairing additive
Ammonium hydroxide Tetraethylammonium hydroxide
Pyrrolidine Dimethylhexylamine
Detergents
Triton X100
SDS
What is ion suppression or enhancement ?
Experimental design aimed to look for a better solution
-
8/12/2019 Waters LCMSWaters LCMS Troubleshooting
71/82
2004 Waters Corporation2004 Waters Corporation
2795
ESI-MS
Infusion pump
Used to add range of modifiers,
salts, ion pairs, pH additives,Matrix extracts
50/50 ACN/ H2O
8 compounds
0.2mL/min
0.2mL/min
ES+
260.2 Propranolol
291.3 Trimethoprim
354.4 Pipenzolate*
411.4 Resperidone
472.6 Terfenadine
485.6 Methoxy-Verapamil
591.6 Benextramine609.6 Reserpine
*quaternary amine drug
Compare 50/50 ACN/ H2O
to additive stream signal
(triplicates) blank, matrix, blank
p gremoval of suppression
S f t t
-
8/12/2019 Waters LCMSWaters LCMS Troubleshooting
72/82
2004 Waters Corporation2004 Waters Corporation
250 300 350 400 450 500 550 600 650 700 750 800 850 900 950m/z0
100
%
0
100
%
Scan ES+
6.41e9
354.4
291.3
669.78609.6
581.71
472.6
411.4
537.60
485.6
713.74
757.83
801.86
845.88
889.91933.93
977.95
Scan ES+7.38e9354.4
260.2
291.3
609.6
485.6472.6
411.4
591.6
260.2
260.2 - 80 %
291.3 - 38 %
354.4 - 13 %
411.4 - 78 %472.6 - 59 %
485.6 - 80 %
591.6 - 71 %
609.6 - 63 %
50/50 water/ACN Blank
0.5 % Triton X 100
260.2 Propranolol
291.3 Trimethoprim
354.4 Pipenzolate *
411.4 Resperidone472.6 Terfenadine
485.6 Methoxy-Verapamil
591.6 Benextramine
609.6 Reserpine
* Quaternary amine molecule
250 300 350 400 450 500 550 600 650 700 750 800 850 900 950m/z0
100
%
0
100
%
Scan ES+
6.41e9
354.4
291.3
669.78609.6
581.71
472.6
411.4
537.60
485.6
713.74
757.83
801.86
845.88
889.91933.93
977.95
Scan ES+7.38e9354.4
260.2
291.3
609.6
485.6472.6
411.4
591.6
260.2
260.2 - 80 %
291.3 - 38 %
354.4 - 13 %
411.4 - 78 %472.6 - 59 %
485.6 - 80 %
591.6 - 71 %
609.6 - 63 %
50/50 water/ACN Blank
0.5 % Triton X 100
260.2 Propranolol
291.3 Trimethoprim
354.4 Pipenzolate *
411.4 Resperidone472.6 Terfenadine
485.6 Methoxy-Verapamil
591.6 Benextramine
609.6 Reserpine
* Quaternary amine molecule
250 300 350 400 450 500 550 600 650 700 750 800 850 900 950m/z0
100
%
250 300 350 400 450 500 550 600 650 700 750 800 850 900 950m/z0
100
%
0
100
%
Scan ES+
6.41e9
354.4
291.3
669.78609.6
581.71
472.6
411.4
537.60
485.6
713.74
757.83
801.86
845.88
889.91933.93
977.95
0
100
%
Scan ES+
6.41e9
354.4
291.3
669.78609.6
581.71
472.6
411.4
537.60
485.6
713.74
757.83
801.86
845.88
889.91933.93
977.95
Scan ES+7.38e9354.4
260.2
291.3
609.6
485.6472.6
411.4
591.6
354.4
260.2
291.3
609.6
485.6472.6
411.4
591.6
260.2
260.2 - 80 %
291.3 - 38 %
354.4 - 13 %
411.4 - 78 %472.6 - 59 %
485.6 - 80 %
591.6 - 71 %
609.6 - 63 %
50/50 water/ACN Blank
0.5 % Triton X 100
260.2 Propranolol
291.3 Trimethoprim
354.4 Pipenzolate *
411.4 Resperidone472.6 Terfenadine
485.6 Methoxy-Verapamil
591.6 Benextramine
609.6 Reserpine
* Quaternary amine molecule
Surfactant
A idi Additi
-
8/12/2019 Waters LCMSWaters LCMS Troubleshooting
73/82
2004 Waters Corporation2004 Waters Corporation
260 280 300 320 340 360 380 400 420 440 460 480 500 520 540 560 580 600 620m/z0
100
%
0
100
%
Scan ES+
609.6354.4
260.3
291.3
485.6472.6
411.5 591.7
Scan ES+354.3
260.3
291.2
609.6
485.6472.6
411.4
591.7
50/50 Water/ACN + 0.5 % FA
50/50 Water/ACN
260.3 + 5 %
291.3 - 5 %
354.4 - 5 %
411.5 - 54 %
472.6 - 7 %
485.6 - 2 %
591.7 - 52 %
609.6 + 17 %
260 280 300 320 340 360 380 400 420 440 460 480 500 520 540 560 580 600 620m/z0
100
%
0
100
%
Scan ES+
609.6354.4
260.3
291.3
485.6472.6
411.5 591.7
Scan ES+354.3
260.3
291.2
609.6
485.6472.6
411.4
591.7
260 280 300 320 340 360 380 400 420 440 460 480260 280 300 320 340 360 380 400 420 440 460 480 500 520 540 560 580 600 620m/z0
100
%
500 520 540 560 580 600 620m/z0
100
%
0
100
%
Scan ES+
609.6354.4
260.3
291.3
485.6472.6
411.5 591.7
Scan ES+354.3
260.3
291.2
609.6
485.6472.6
411.4
591.7
50/50 Water/ACN + 0.5 % FA
50/50 Water/ACN
260.3 + 5 %
291.3 - 5 %
354.4 - 5 %
411.5 - 54 %
472.6 - 7 %
485.6 - 2 %
591.7 - 52 %
609.6 + 17 %
Acidic Additive
B i Additi
-
8/12/2019 Waters LCMSWaters LCMS Troubleshooting
74/82
2004 Waters Corporation2004 Waters Corporation
260 280 300 320 340 360 380 400 420 440 460 480 500 520 540 560 580 600 620
m/z0
100
%
0
100
%
Scan ES+354.5
260.3291.4
471.6
411.5
609.6
485.6 591.7
Scan ES+354.4
260.3291.3
609.6
485.6472.6
411.6
591.7
260.3 + 10 %
294.4 + 4 %
354.4 0 %
411.5 + 16 %
471.6 + 57 %485.6 + 46 %
594.7 + 37 %
609.6 - 6 %
50/50 Water/ACN + 0.5 % NH4OH
50/50 Water/ACN
260 280 300 320 340 360 380 400 420 440 460 480 500 520 540 560 580 600 620
m/z0
100
%
0
100
%
Scan ES+354.5
260.3291.4
471.6
411.5
609.6
485.6 591.7
Scan ES+354.4
260.3291.3
609.6
485.6472.6
411.6
591.7
260 280 300 320 340 360 380 400 420 440 460 480 500 520 540 560 580 600 620
m/z0
100
%
500 520 540 560 580 600 620
m/z0
100
%
0
100
%
Scan ES+354.5
260.3291.4
471.6
411.5
609.6
485.6 591.7
Scan ES+354.4
260.3291.3
609.6
485.6472.6
411.6
591.7
260.3 + 10 %
294.4 + 4 %
354.4 0 %
411.5 + 16 %
471.6 + 57 %485.6 + 46 %
594.7 + 37 %
609.6 - 6 %
50/50 Water/ACN + 0.5 % NH4OH
50/50 Water/ACN
Basic Additive
Ion Pairing Reagent
-
8/12/2019 Waters LCMSWaters LCMS Troubleshooting
75/82
2004 Waters Corporation2004 Waters Corporation
260 280 300 320 340 360 380 400 420 440 460 480 500 520 540 560 580 600 620 640 660
m/z0
100
%
Scan ES+1.95e8
354.44
260 280 300 320 340 360 380 400 420 440 460 480 500 520 540 560 580 600 620 640 660m/z0
100
%
Scan ES+7.38e9354.44
260.28
291.31
609.55
485.62472.57
411.49591.66
591.66
50/50 Water/ACN Blank
50 mM Tetraethylammonium hydroxide
260.2 - 100 %
291.3 - 100 %354.4 - 88 %
411.4 - 100 %
472.5 - 100 %
485.5 - 100 %
591.6 - 94 %
609.5 - 100 %
260 280 300 320 340 360 380 400 420 440 460 480 500 520 540 560 580 600 620 640 660
m/z0
100
%
Scan ES+1.95e8
354.44
260 280 300 320 340 360 380 400 420 440 460 480 500 520 540 560 580 600 620 640 660m/z0
100
%
Scan ES+7.38e9354.44
260.28
291.31
609.55
485.62472.57
411.49591.66
591.66
260 280 300 320 340 360 380 400 420 440 460 480 500 520 540 560 580 600 620 640 660
m/z0
100
%
Scan ES+1.95e8
Scan ES+1.95e8
354.44
260 280 300 320 340 360 380 400 420 440 460 480 500 520 540 560 580 600 620 640 660m/z0
100
%
Scan ES+7.38e9354.44
260.28
291.31
609.55
485.62472.57
411.49591.66
591.66
50/50 Water/ACN Blank
50 mM Tetraethylammonium hydroxide
260.2 - 100 %
291.3 - 100 %354.4 - 88 %
411.4 - 100 %
472.5 - 100 %
485.5 - 100 %
591.6 - 94 %
609.5 - 100 %
Ion-Pairing Reagent
Salt Adducts
-
8/12/2019 Waters LCMSWaters LCMS Troubleshooting
76/82
2004 Waters Corporation2004 Waters Corporation
260 280 300 320 340 360 380 400 420 440 460 480 500 520 540 560 580 600 620m/z0
100
%
0
100
%
Scan ES+
354.4
291.3260.3
591.7
472.6411.5485.6 609.6
Scan ES+354.4
260.3
291.3
609.6
485.6472.6
411.5591.7
50/50 Water/ACN + 0.1M NaCl
50/50 Water/ACN
260.3 - 93 %
291.3 - 95 %
354.4 - 37 %
411.5 - 62 %
472.6 - 71 %485.6 - 84 %
591.7 - 45 %
609.6 - 95 %
260 280 300 320 340 360 380 400 420 440 460 480 500 520 540 560 580 600 620m/z0
100
%
0
100
%
Scan ES+
354.4
291.3260.3
591.7
472.6411.5485.6 609.6
Scan ES+354.4
260.3
291.3
609.6
485.6472.6
411.5591.7
260 280 300 320 340 360 380 400 420 440 460 480260 280 300 320 340 360 380 400 420 440 460 480 500 520 540 560 580 600 620m/z0
100
%
500 520 540 560 580 600 620m/z0
100
%
0
100
%
Scan ES+
354.4
291.3260.3
591.7
472.6411.5485.6 609.6
Scan ES+354.4
260.3
291.3
609.6
485.6472.6
411.5591.7
50/50 Water/ACN + 0.1M NaCl
50/50 Water/ACN
260.3 - 93 %
291.3 - 95 %
354.4 - 37 %
411.5 - 62 %
472.6 - 71 %485.6 - 84 %
591.7 - 45 %
609.6 - 95 %
Salt Adducts
Rat Plasma
-
8/12/2019 Waters LCMSWaters LCMS Troubleshooting
77/82
2004 Waters Corporation2004 Waters Corporation
260 280 300 320 340 360 380 400 420 440 460 480 500 520 540 560 580 600 620m/z0
100
%
0
100
%
Scan ES+
591.7
354.4 518.6
472.6 546.6
609.6
Scan ES+354.4
260.2
291.2
609.6
485.6472.6
411.4 591.6
.
50/50 Water/ACN
260.3 - 98 %
291.3 - 98 %
354.4 - 87 %
411.4 - 94 %
472.6 - 92 %
485.6 - 95 %
591.7 - 42 %
609.6 - 94 %
50/50 Water/ACN + rat plasma supernatant
260 280 300 320 340 360 380 400 420 440 460 480 500 520 540 560 580 600 620m/z0
100
%
0
100
%
Scan ES+
591.7
354.4 518.6
472.6 546.6
609.6
Scan ES+354.4
260.2
291.2
609.6
485.6472.6
411.4 591.6
.
260 280 300 320 340 360 380 400 420 440 460 480260 280 300 320 340 360 380 400 420 440 460 480 500 520 540 560 580 600 620m/z0
100
%
500 520 540 560 580 600 620m/z0
100
%
0
100
%
Scan ES+
591.7
354.4 518.6
472.6 546.6
609.6
Scan ES+354.4
260.2
291.2
609.6
485.6472.6
411.4 591.6
.
50/50 Water/ACN
260.3 - 98 %
291.3 - 98 %
354.4 - 87 %
411.4 - 94 %
472.6 - 92 %
485.6 - 95 %
591.7 - 42 %
609.6 - 94 %
50/50 Water/ACN + rat plasma supernatant
Rat Plasma
Human Plasma
-
8/12/2019 Waters LCMSWaters LCMS Troubleshooting
78/82
2004 Waters Corporation2004 Waters Corporation
260 280 300 320 340 360 380 400 420 440 460 480 500 520 540 560 580 600 620m/z0
100
%
0
100
%
Scan ES+
591.7
354.4 518.5472.6 485.5 609.6
Scan ES+354.4
260.2
291.3
609.6
485.6472.6
411.5 591.6
50/50 Water/ACN + human plasma supernatant
50/50 Water/ACN
260.2 - 97 %
291.2 - 96 %
354.4 - 86 %
411.4 - 93 %
472.6 - 93 %485.6 - 95 %
591.6 - 89 %
609.5 - 93 %
260 280 300 320 340 360 380 400 420 440 460 480 500 520 540 560 580 600 620m/z0
100
%
0
100
%
Scan ES+
591.7
354.4 518.5472.6 485.5 609.6
Scan ES+354.4
260.2
291.3
609.6
485.6472.6
411.5 591.6
50/50 Water/ACN + human plasma supernatant
50/50 Water/ACN
260 280 300 320 340 360 380 400 420 440 460 480260 280 300 320 340 360 380 400 420 440 460 480 500 520 540 560 580 600 620m/z0
100
%
500 520 540 560 580 600 620m/z0
100
%
0
100
%
Scan ES+
591.7
354.4 518.5472.6 485.5 609.6
Scan ES+354.4
260.2
291.3
609.6
485.6472.6
411.5 591.6
50/50 Water/ACN + human plasma supernatant
50/50 Water/ACN
260.2 - 97 %
291.2 - 96 %
354.4 - 86 %
411.4 - 93 %
472.6 - 93 %485.6 - 95 %
591.6 - 89 %
609.5 - 93 %
Human Plasma
Reversed Phase SPE
-
8/12/2019 Waters LCMSWaters LCMS Troubleshooting
79/82
2004 Waters Corporation2004 Waters Corporation
Condition/Equilibrate
1.0 mL methanol / 1.0 mL water
Load
1.0 mL plasma
Wash
1.0 mL 5% methanol in water
Elute
0.5 mL MeOH
Dilute with 0.5 ml water
Plasma Sample
* 30 mg HLB 96 plate
Reversed Phase SPE
Reversed Phase SPE Rat Plasma
-
8/12/2019 Waters LCMSWaters LCMS Troubleshooting
80/82
2004 Waters Corporation2004 Waters Corporation
260 280 300 320 340 360 380 400 420 440 460 480 500 520 540 560 580 600 620m/z0
100
%
0
100
%
Scan ES+
354.4
260.1 291.2
591.6
472.5
411.4
485.5
609.5
Scan ES+354.2
260.2
291.2
609.5
485.4472.5
411.4
591.6
50/50 Water/ACN + rat plasma HLB 1D extract
50/50 Water/ACN
260.2 - 41 %
291.2 - 26 %
354.4 - 9 %
411.4 - 32 %
472.6 - 23 %485.6 - 38 %
591.6 + 26 %
609.5 - 49 %
260 280 300 320 340 360 380 400 420 440 460 480 500 520 540 560 580 600 620m/z0
100
%
0
100
%
Scan ES+
354.4
260.1 291.2
591.6
472.5
411.4
485.5
609.5
Scan ES+354.2
260.2
291.2
609.5
485.4472.5
411.4
591.6
50/50 Water/ACN + rat plasma HLB 1D extract
50/50 Water/ACN
260 280 300 320 340 360 380 400 420 440 460 480 500 520 540 560 580 600 620m/z0
100
%
0
100
%
Scan ES+
354.4
260.1 291.2
591.6
472.5
411.4
485.5
609.5
Scan ES+354.2
260.2
291.2
609.5
485.4472.5
411.4
591.6
50/50 Water/ACN + rat plasma HLB 1D extract
50/50 Water/ACN
260.2 - 41 %
291.2 - 26 %
354.4 - 9 %
411.4 - 32 %
472.6 - 23 %485.6 - 38 %
591.6 + 26 %
609.5 - 49 %
Reversed Phase SPE - Rat Plasma
Mixed Mode Cation Exchange SPE
-
8/12/2019 Waters LCMSWaters LCMS Troubleshooting
81/82
2004 Waters Corporation2004 Waters Corporation
Condition/Equilibrate
1.0 mL methanol / 1.0 mL water
Load
1.0 mL plasma
Prepare Sample Solution
Wash 2
1.0 mL MeOH
Elute
0.5 mL MeOH + 2% NH4OH
Dilute with 0.5 ml water
Wash 1
1.0 mL Water + 2 % FALocks basic drug
on ion exchanger
Removes polar
interferences
* 30 mg Oasis MCX 96 well plate
Mixed Mode Cation-Exchange SPE
Mixed Mode SPE Rat Plasma
-
8/12/2019 Waters LCMSWaters LCMS Troubleshooting
82/82
260.2 - 9 %
291.2 - 11%
354.4 - 0.5 %
411.4 - 13 %
472.6 - 9 %
485.6 - 2 %
591.6 - 8 %
609.5 - 8 %
100
%
0
100
%
Scan ES+
354.4
291.3260.3
609.6
472.6
411.5
485.6
591.7
Scan ES+354.4
260.3
291.3
609.6
485.6
472.6
411.5 591.7
50/50 Water/ACN + rat plasma MCX extract
50/50 Water/ACN
260.2 - 9 %
291.2 - 11%
354.4 - 0.5 %
411.4 - 13 %
472.6 - 9 %
485.6 - 2 %
591.6 - 8 %
609.5 - 8 %
100
%
0
100
%
Scan ES+
354.4
291.3260.3
609.6
472.6
411.5
485.6
591.7
Scan ES+354.4
260.3
291.3
609.6
485.6
472.6
411.5 591.7
50/50 Water/ACN + rat plasma MCX extract
50/50 Water/ACN100
%
0
100
%
Scan ES+
354.4
291.3260.3
609.6
472.6
411.5
485.6
591.7
Scan ES+354.4
260.3
291.3
609.6
485.6
472.6
411.5 591.7
50/50 Water/ACN + rat plasma MCX extract
50/50 Water/ACN
Mixed Mode SPE Rat Plasma