The values listed represent typical performance. Formal controlled specificaons for use in equipment design may be obtained from CPI [email protected] www.cpii.com +1-905-877-0161 Model VJB2002 is a refrigerated, re- circulang water condioning heat ex- changer specifically configured for opera- on with Extended Interacon Klystrons (EIKs). This easy to use chiller is opmized for the demanding purity and cooling re- quirements of water cooled EIKs. It fea- tures a robust refrigeraon system de- signed for connuous operaon, providing accurate control of temperature and ensur- ing water purity needs are met. With a focus on reduced maintenance, the chiller comes with a level indicator and built -in funnel to make filling easier. The inte- grated air filter is located behind an easy to remove condenser grill for quick and simple cleaning to opmize chiller performance and maximize operaonal life. When purchased as part of a system, the chiller comes pre-configured to match the EIK cooling needs and interlocked with the power supply unit to ensure safe operaon of the EIK. Fault condions of out-of-range temperature, flow or conducvity will pro- vide an open circuit signal to the power supply that will shut down the operaon of the EIK for fault resoluon. CLOSED LOOP HEAT EXCHANGER Water to Air Laboratory Chiller Features Coolant Temperature +5 °C to +40 °C Temperature Stability ±0.1 °C Cooling Capacity 2.2 kW @ 20 °C Coolant Flow 6.4 l/min @ 415 kPa Temperature Operang +10 to +40 °C Reservoir Volume 7.2 l Power 200 to 230 VAC 50/60 Hz, 10 A Dimensions (LxWxH mm) 673x435x735 Mass 80 kg Model VJB2002 Funcons: A pump in the chiller circulates water throughout a closed-loop system. Heat from various sources is conducted into the water and a heat ex- changer in the chiller transfers the heat into the surrounding air A paral flow filtering system maintains water purity by removing parcu- lates and regulang the conducvity (resisvity) of the water The chiller displays process water temperature, pressure, flow rate and resisvity. Addionally, these parameters are monitored to warning and trip level ranges An EIK power supply interlock funcon enables EIK high voltage operaon when cooling system parameters are within correct operang values Water cooled EIKs operate at high power densies. Careful aenon to ade- quate water flow and purity is required to ensure proper operaon and long life. Water with contaminant levels that exceed the purity requirements can result in severe corrosion and scaling, which includes normal municipal tap water. Unchecked corrosion of the metals in the EIK coolant passages will re- sult in leaks and reduced operang life. Scaling obstructs cooling passages, prevenng efficient heat transfer. Overheang and permanent damage may be the result. Connuous filtering and ion regulaon is necessary for main- taining high water purity. For further details see CPI Bullen #MMAP004. These issues are addressed in the CPI model VJB2002 refrigerated re- circulang heat exchanger.

Transcript of Water to Air Laboratory Chiller - CPI i Datasheet.pdf · This easy to use chiller is optimized for...

Page 1: Water to Air Laboratory Chiller - CPI i Datasheet.pdf · This easy to use chiller is optimized for the demanding purity and cooling re- ... An EIK power supply interlock function

The values listed represent typical performance. Formal controlled specifications for use in equipment design may be obtained from CPI

[email protected] www.cpii.com +1-905-877-0161

Model VJB2002 is a refrigerated, re-

circulating water conditioning heat ex-

changer specifically configured for opera-

tion with Extended Interaction Klystrons

(EIKs). This easy to use chiller is optimized

for the demanding purity and cooling re-

quirements of water cooled EIKs. It fea-

tures a robust refrigeration system de-

signed for continuous operation, providing

accurate control of temperature and ensur-

ing water purity needs are met.

With a focus on reduced maintenance, the

chiller comes with a level indicator and built

-in funnel to make filling easier. The inte-

grated air filter is located behind an easy to

remove condenser grill for quick and simple

cleaning to optimize chiller performance

and maximize operational life.

When purchased as part of a system, the

chiller comes pre-configured to match the

EIK cooling needs and interlocked with the

power supply unit to ensure safe operation

of the EIK. Fault conditions of out-of-range

temperature, flow or conductivity will pro-

vide an open circuit signal to the power

supply that will shut down the operation of

the EIK for fault resolution.

CLOSED LOOP HEAT EXCHANGER Water to Air Laboratory Chiller

Features

Coolant Temperature +5 °C to +40 °C

Temperature Stability ±0.1 °C

Cooling Capacity 2.2 kW @ 20 °C

Coolant Flow 6.4 l/min @ 415 kPa

Temperature Operating +10 to +40 °C

Reservoir Volume 7.2 l

Power 200 to 230 VAC 50/60 Hz, 10 A

Dimensions (LxWxH mm) 673x435x735

Mass 80 kg

Mo

de

l VJB

20

02

Functions:

A pump in the chiller circulates water throughout a closed-loop system.

Heat from various sources is conducted into the water and a heat ex-

changer in the chiller transfers the heat into the surrounding air

A partial flow filtering system maintains water purity by removing particu-

lates and regulating the conductivity (resistivity) of the water

The chiller displays process water temperature, pressure, flow rate and

resistivity. Additionally, these parameters are monitored to warning and

trip level ranges

An EIK power supply interlock function enables EIK high voltage operation

when cooling system parameters are within correct operating values

Water cooled EIKs operate at high power densities. Careful attention to ade-

quate water flow and purity is required to ensure proper operation and long

life. Water with contaminant levels that exceed the purity requirements can

result in severe corrosion and scaling, which includes normal municipal tap

water. Unchecked corrosion of the metals in the EIK coolant passages will re-

sult in leaks and reduced operating life. Scaling obstructs cooling passages,

preventing efficient heat transfer. Overheating and permanent damage may

be the result. Continuous filtering and ion regulation is necessary for main-

taining high water purity. For further details see CPI Bulletin #MMAP004.

These issues are addressed in the CPI model VJB2002 refrigerated re-

circulating heat exchanger.

Page 2: Water to Air Laboratory Chiller - CPI i Datasheet.pdf · This easy to use chiller is optimized for the demanding purity and cooling re- ... An EIK power supply interlock function

CPI Canada 45 River Dr, Georgetown, ON, L7G 2J4, Canada

201501VJB2002 ©CPI 2015

Options

Voltage 100 to 115 VAC, 50/60 Hz

High volume pump 15 l/min @ 415 kPa

Power Supply

Pump

Reservoir

Filtering

Temperature Control

Pressure Control

Flow Monitoring

Ion Control

RFin RFout

Inte

rlo

ck

EIK