Viscosity-based detection system - LAPROPHARM PLUS · 2019. 7. 29. · Diagnostica Stago S.A.S RCS...

12
Viscosity-based detection system www.epicea.com - ©2009 Diagnostica Stago - All rights reserved - Non-contractual photos - 06/2017 - Ref. 27719

Transcript of Viscosity-based detection system - LAPROPHARM PLUS · 2019. 7. 29. · Diagnostica Stago S.A.S RCS...

Page 1: Viscosity-based detection system - LAPROPHARM PLUS · 2019. 7. 29. · Diagnostica Stago S.A.S RCS Nanterre B305 151 409 3 allée Thérésa 92600 Asnières sur Seine France Ph. +33

Viscosity-baseddetection system

www.

epice

a.co

m -

©20

09 D

iagno

stica

Sta

go -

All r

ights

rese

rved

- No

n-co

ntra

ctua

l pho

tos

- 06/

2017

- Re

f. 27

719

Page 2: Viscosity-based detection system - LAPROPHARM PLUS · 2019. 7. 29. · Diagnostica Stago S.A.S RCS Nanterre B305 151 409 3 allée Thérésa 92600 Asnières sur Seine France Ph. +33

The change in viscosity is measured by monitoringthe amplitude of an oscillating steel ball

in a specially designed cuvette.

Clot detection is based on the increasein viscosity of the plasma being tested.

Page 3: Viscosity-based detection system - LAPROPHARM PLUS · 2019. 7. 29. · Diagnostica Stago S.A.S RCS Nanterre B305 151 409 3 allée Thérésa 92600 Asnières sur Seine France Ph. +33

ACTIVATING COILACTIVATING COIL ACTIVATING COILACTIVATING COIL

Movement of the steel ball is mediated by 2 activating coils,working alternatively to induce and maintain

a natural oscillation.

Page 4: Viscosity-based detection system - LAPROPHARM PLUS · 2019. 7. 29. · Diagnostica Stago S.A.S RCS Nanterre B305 151 409 3 allée Thérésa 92600 Asnières sur Seine France Ph. +33

When the start reagent is added,the detection and the chronometer start

immediately and simultaneously.

Page 5: Viscosity-based detection system - LAPROPHARM PLUS · 2019. 7. 29. · Diagnostica Stago S.A.S RCS Nanterre B305 151 409 3 allée Thérésa 92600 Asnières sur Seine France Ph. +33

As the ball oscillates left and right, the amplitude of the motion is measured.

The chronometer times the clotting of the sample.

Page 6: Viscosity-based detection system - LAPROPHARM PLUS · 2019. 7. 29. · Diagnostica Stago S.A.S RCS Nanterre B305 151 409 3 allée Thérésa 92600 Asnières sur Seine France Ph. +33

As the ball oscillates left and right, the amplitude of the motion is measured.

The chronometer times the clotting of the sample.

Page 7: Viscosity-based detection system - LAPROPHARM PLUS · 2019. 7. 29. · Diagnostica Stago S.A.S RCS Nanterre B305 151 409 3 allée Thérésa 92600 Asnières sur Seine France Ph. +33

Amplitude is monitored during the entire clotting process.The amplitude remains constant while no clot is present.

Page 8: Viscosity-based detection system - LAPROPHARM PLUS · 2019. 7. 29. · Diagnostica Stago S.A.S RCS Nanterre B305 151 409 3 allée Thérésa 92600 Asnières sur Seine France Ph. +33

As the clot develops the viscosity increases,and the amplitude decreases.

Page 9: Viscosity-based detection system - LAPROPHARM PLUS · 2019. 7. 29. · Diagnostica Stago S.A.S RCS Nanterre B305 151 409 3 allée Thérésa 92600 Asnières sur Seine France Ph. +33

Based on different algorithms, the chronometer is stoppedeven if the clot is weak, and even if the ball is not stopped.

Page 10: Viscosity-based detection system - LAPROPHARM PLUS · 2019. 7. 29. · Diagnostica Stago S.A.S RCS Nanterre B305 151 409 3 allée Thérésa 92600 Asnières sur Seine France Ph. +33

Maximum sensitivity for:

• Weak clots

• Fibrinogen testing linearity 1.5 to 9 g/L

• All clotting tests

Insensitivity to:

• Coloured plasmas • Haemolysed plasmas • Lipemic plasmas • Bilirubin • Turbid reagents

Standardisation:

• For each clotting process • For the entire range of

Stago analysers

Page 11: Viscosity-based detection system - LAPROPHARM PLUS · 2019. 7. 29. · Diagnostica Stago S.A.S RCS Nanterre B305 151 409 3 allée Thérésa 92600 Asnières sur Seine France Ph. +33

STart Max STA Compact Max

STA Satellite Max STA R Max

Page 12: Viscosity-based detection system - LAPROPHARM PLUS · 2019. 7. 29. · Diagnostica Stago S.A.S RCS Nanterre B305 151 409 3 allée Thérésa 92600 Asnières sur Seine France Ph. +33

Diagnostica Stago S.A.SRCS Nanterre B305 151 4093 allée Thérésa92600 Asnières sur Seine FrancePh. +33 (0)1 46 88 20 20Fax +33 (0)1 47 91 08 [email protected]

This document contains information on products which is targeted to a wide range of audiences and could contain product details or information otherwise not accessible or valid in your country.

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .