N2-Whisper-0 LC-MS
User ManualThis Operating Manual is applicable for the following models:
N2-Whisper-0 LC-MS
Ver. 1.00.0100 - 1 -
VICI AG InternationalParkstrasse 2CH-6214 SchenkonSwitzerlandPhone Int +41 41 925 62 00Fax Int +41 41 925 62 01E-mail [email protected]
Table of Contents
1. Version history...................................................................................................32. Introduction.......................................................................................................42.1...................................................................................................................................FUNCTIONING PRINCIPLE............................................................................................................................................................. 4
2.2..............................................................................................................................................Symbols Definition............................................................................................................................................................. 4
3. Safety................................................................................................................54. Warning.............................................................................................................55. Specifications.....................................................................................................65.1.....................................................................................................................................Technical Specifications............................................................................................................................................................. 6
5.2................................................................................................................................................................Capacity............................................................................................................................................................. 6
5.3...........................................................................................................................................................Spare Parts............................................................................................................................................................. 7
5.4............................................................................................................................................Separation principle............................................................................................................................................................. 8
5.5.......................................................................................................................................................................Parts............................................................................................................................................................. 9
5.6..................................................................................................................................................Process Diagram........................................................................................................................................................... 11
6. Installation.......................................................................................................126.1...............................................................................................................................................................Transport........................................................................................................................................................... 12
6.2.................................................................................................................................................................Location........................................................................................................................................................... 12
6.3....................................................................................................Connecting the unit to the Air Compressor........................................................................................................................................................... 12
6.4.....................................................................................................................Connecting the unit to the LC/MS........................................................................................................................................................... 13
7. Operation.........................................................................................................147.1...................................................................................................................Starting the N2-WHISPER-0 LC-MS........................................................................................................................................................... 14
7.2.................................................................................................................Stopping the N2-WHISPER-0 LC-MS........................................................................................................................................................... 14
Ver. 1.00.0100 - 2 -
VICI AG InternationalParkstrasse 2CH-6214 SchenkonSwitzerlandPhone Int +41 41 925 62 00Fax Int +41 41 925 62 01E-mail [email protected]
8. Troubleshooting...............................................................................................159. Maintenance....................................................................................................169.1.........................................................................................................................................................Maintenance........................................................................................................................................................... 16
9.2.............................................................................................................................................Changing the filters........................................................................................................................................................... 16
Ver. 1.00.0100 - 3 -
VICI AG InternationalParkstrasse 2CH-6214 SchenkonSwitzerlandPhone Int +41 41 925 62 00Fax Int +41 41 925 62 01E-mail [email protected]
1. VERSION HISTORY
Version Date Comment/Change
1.00.0100 18th January 2017 VICI DBS layout, initial release VICI DBS
1.01.0000 update
Ver. 1.00.0100 - 4 -
VICI AG InternationalParkstrasse 2CH-6214 SchenkonSwitzerlandPhone Int +41 41 925 62 00Fax Int +41 41 925 62 01E-mail [email protected]
2. INTRODUCTION
This manual is an integral part of the product. The manual describes the installation and maintenance operations as well as user modes and troubleshooting.
ContentsRead the manual carefully before starting up the N2-Whisper-0 LC-MS and familiarize yourself with its contents.
ModificationsDo not make any modifications to the N2-Whisper-0 LC-MS unless explicit written permission has been given by the manufacturer. If such modifications are made without the manufacturer’s consent, the manufactures deny any and all responsibility for eventual consequential damage.
InformationAll the information in this manual, including drawings, tables, and technical descriptions, remain the property of the manufacturer and may not be used (with the exception of the purposes herein), copied, reproduced, or published without explicit consent.
2.1 FUNCTIONING PRINCIPLE
The N2-Whisper-0 LC-MS separates compressed air into nitrogen and oxygen-enriched air. The separation system is based on a special membrane.
2.2 SYMBOLS DEFINITION
Symbol Description
or
General warning. Caution is necessary when operating the device or control close to where the symbol is placed, or to indicate that the current situation needs operator awareness or operator action in order to avoid undesirable consequences.The manual must be consulted in all case where this symbol is marked on the device.
or Risk of asphyxia. Caution is necessary when operating the device in a not sufficiently ventilated area.
or Oxidant material. Caution is necessary due to the presence of oxidant material.
WEEE symbol. When discard this product, it must be sent to separate collection facilities for recovery and recycling.
Ver. 1.00.0100 - 5 -
VICI AG InternationalParkstrasse 2CH-6214 SchenkonSwitzerlandPhone Int +41 41 925 62 00Fax Int +41 41 925 62 01E-mail [email protected]
3. SAFETY
The N2-Whisper-0 is a Nitrogen Generator.Do not let it run with the outlet open to the air in a low volume room.
The increase of the Nitrogen (and decrease of oxygen) may cause human disease.It is the responsibility of the user to check if the room is appropriate or if an O2
measurement is necessary.
The N2-Whisper-0 reject Oxygen as waste.Oxygen ratio in the air could change around the device in a low volume or not sufficient
ventilated room.If fuel gases are present around, their explosive or flammable limits could be affected.
It is the responsibility of the user to check if the room is appropriate or if an O2 measurement is necessary.
4. WARNING
The quantity of dust in the ambient Air can be very different from one place to another.We recommend to replace the dust filters every 4000H.
In some dusty area, this maintenance need to be perform more frequently.
Clogged dust filter may damage the device or give a bad Nitrogen quality
If the N2-Whisper-0 is used in a way not specified by VICI DBS srl, the protection provided by the equipment could be compromised.
Ver. 1.00.0100 - 6 -
VICI AG InternationalParkstrasse 2CH-6214 SchenkonSwitzerlandPhone Int +41 41 925 62 00Fax Int +41 41 925 62 01E-mail [email protected]
5. SPECIFICATIONS
5.1 TECHNICAL SPECIFICATIONS
Pressure of N2 producedMaximum pressure 8 bar(g) / 116 psig
Environmental conditionsTemperature From 10 to 35 °C / from 50 to 95 °FAir quality Normal atmospheric air, relative humidity < 90%Noisiness dB(A) < 58
Dimensions and connectionsDimensions (H x W x D) [mm] 1300 x 480 x 300Dimensions (H x W x D) [inches] 51 x 19 x 39Net weight 25 kg
Connections N2 Outlet: G ¼ “ OD 6mm
PartsN2-WHISPER-0 1x N2-WHISPER-0
1x Manual1x Tubing for tank connection
A special characteristic of the N2-WHISPER-0 LC-MS is the momentary stop of the compressor compartment if the LC-MS is not actually using nitrogen. This momentary stop is regulated by a pneumatic control valve pre-calibrated by VICIDBS to lengthen the life span of the compressors and guarantee the user savings in terms of electric energy.
5.2 CAPACITY
STP* Production Capacity
Flow (l/min) N2 Purity (%)
5 99.98010 99.95915 99.91018 99.85520 99.824
All the tests were performed with Systech EC92D/IS Oxygen Analyzer.
*STP:Room temperature: 20 °C / 68 °FEnvironmental pressure: 1013 mbar(a).
Ver. 1.00.0100 - 7 -
VICI AG InternationalParkstrasse 2CH-6214 SchenkonSwitzerlandPhone Int +41 41 925 62 00Fax Int +41 41 925 62 01E-mail [email protected]
5.3 SPARE PARTS
Maintenace kit:
Spare part CodeAnnual Maintenance Kit 1x internal filter
Contact your dealer
Ver. 1.00.0100 - 8 -
VICI AG InternationalParkstrasse 2CH-6214 SchenkonSwitzerlandPhone Int +41 41 925 62 00Fax Int +41 41 925 62 01E-mail [email protected]
5.4 SEPARATION PRINCIPLE
A Compressed air intakeB Hollow fibre membraneC Separation layerD Support layer
E Nitrogen outputF Rapid permeationG Slow permeation
Atmospheric air contains nitrogen (78.1%), oxygen (20.9%), argon (1%), carbon dioxide, water vapour, and other traces of inert gas. The compressed air (A) is forced inside the hollow fibre membrane (B). The various components of the air permeate the porous membrane walls. The speed of permeation varies from gas to gas:
Oxygen and water vapour have a high permeation speed and pass quickly through the membrane wall.
Nitrogen has a slow permeation speed and passes slowly through the membrane wall. The resulting nitrogen flow is then forced out of the membrane (E) and used directly.
Ver. 1.00.0100 - 9 -
VICI AG InternationalParkstrasse 2CH-6214 SchenkonSwitzerlandPhone Int +41 41 925 62 00Fax Int +41 41 925 62 01E-mail [email protected]
5.5 PARTS
A. Active carbon filter
B. Air Inlet / Coalescent filter
C. Coalescent filter for fine particles.
D. Pressure gauge.
E. Nitrogen outlet (coupling for 6 mm tube).
Ver. 1.00.0100 - 10 -
F
G
F
VICI AG InternationalParkstrasse 2CH-6214 SchenkonSwitzerlandPhone Int +41 41 925 62 00Fax Int +41 41 925 62 01E-mail [email protected]
F. Switcher
G. Nitrogen Separation Membrane
H. O2/Air Drain
Ver. 1.00.0100 - 11 -
VICI AG InternationalParkstrasse 2CH-6214 SchenkonSwitzerlandPhone Int +41 41 925 62 00Fax Int +41 41 925 62 01E-mail [email protected]
5.6 PROCESS DIAGRAM
Ver. 1.00.0100 - 12 -
LC-MS (or other )Compressed air N2-Whisper-0 LC-MS
VICI AG InternationalParkstrasse 2CH-6214 SchenkonSwitzerlandPhone Int +41 41 925 62 00Fax Int +41 41 925 62 01E-mail [email protected]
6. INSTALLATION
Follow the paragraphs of this chapter to install the N2-Whisper-0 LC-MS.
6.1 TRANSPORT
ATTENTION:Transport the N2-Whisper-0 LC-MS without tipping it.
Insert the N2-Whisper-0 LC-MS in the original packaging to transport it long distances.
6.2 LOCATION
Install the N2-Whisper-0 LC-MS in an area with the following characteristics:
Minimum distance of 10 cm on front and lateral sides to allow sufficient cooling. An enclosed (indoor), dry room Avoid direct sunlight Keep away from sources of heat Room with adequate aeration Connect a tube for waste humidity from filter B
6.3 CONNECTING THE UNIT TO THE AIR COMPRESSOR
There are 2 configuration possibilities:
If you have a big flow compressor (like Scroll Compressor) and no tank after the WHISPER-0, please turn the red tap behind the instrument as showed in photo.
If you have a normal/small flow compressor and a tank (min 100 lt) after the WHISPER-0,
Ver. 1.00.0100 - 13 -
VICI AG InternationalParkstrasse 2CH-6214 SchenkonSwitzerlandPhone Int +41 41 925 62 00Fax Int +41 41 925 62 01E-mail [email protected]
please turn the red tap behind the instrument as showed in photo.
6.4 CONNECTING THE UNIT TO THE LC/MS
ATTENTION:
Check that the tubes connecting the N2-Whisper LC-MS® and the LC/MS are free from dust, grease, drops of liquid, etc. before making the connection.
Connect the N2-Whisper LC-MS® and the LC/MS with stainless steel tubes
Ver. 1.00.0100 - 14 -
VICI AG InternationalParkstrasse 2CH-6214 SchenkonSwitzerlandPhone Int +41 41 925 62 00Fax Int +41 41 925 62 01E-mail [email protected]
7. OPERATION
7.1 STARTING THE N2-WHISPER-0 LC-MS
1. Place the generator in the set you have selected. Connect the Air Inlet (B fig.1) to the Air Line (oil free)
2. Cut a piece of tubing from Nitrogen Inlet to the LC/MS or what needs nitrogen.3. Slowly open the red tap upstream from the coalescent filter B.4. The generator is ready to be used.
7.2 STOPPING THE N2-WHISPER-0 LC-MS
To stop the generator is sufficient to close the red tap upstream from the coalescent filter B.
Ver. 1.00.0100 - 15 -
VICI AG InternationalParkstrasse 2CH-6214 SchenkonSwitzerlandPhone Int +41 41 925 62 00Fax Int +41 41 925 62 01E-mail [email protected]
8. TROUBLESHOOTING
Ver. 1.00.0100 - 16 -
Error Possible cause Possible solutionNitrogen flow insufficient or lacking The internal filters are
contaminatedReplace the filters
The N2-WHISPER-0 LC-MS is off
Start the N2-WHISPER-0 LC-MS
There is a leak in the tube Check for possible leaksThe nitrogen outlet is obstructed
Check and clear the nitrogen outlet
Oxygen residue too high
The room temperature is less than 10°C
Raise the room temperature
The membrane no longer functions correctly
The membrane column must be replaced
VICI AG InternationalParkstrasse 2CH-6214 SchenkonSwitzerlandPhone Int +41 41 925 62 00Fax Int +41 41 925 62 01E-mail [email protected]
9. MAINTENANCE
9.1 MAINTENANCE
Parts Action FrequencyFilters Replace the internal filters Annually
9.2 CHANGING THE FILTERS
1. Turn off the N2-WHISPER LC-MS®2. Close the red tap upstream from the coalescent filter B.3. Let the generator depressurize till the pressure is 0 bar.4. Remove the cap, turning it to unscrew it from the main body (1) and after that, pulling it
down (2);5. Extract the filter cartridge pulling it down (3);6. Replace the filtering element and reassemble the parts in reverse order.
(1) (2) (3)
Ver. 1.00.0100 - 17 -
VICI AG InternationalParkstrasse 2CH-6214 SchenkonSwitzerlandPhone Int +41 41 925 62 00Fax Int +41 41 925 62 01E-mail [email protected]