Verification of Compendial Procedu. - Microbiology … · Verification of Compendial Procedu. --...

5
Verification of Compendial Procedu. -- Horacio Pappa, David Porter,and Karen Russo Horacio Pappa, PhD,* IS a senlor scientist, D a v i d Porter, PhD, is a dirxtor, and Karen Russo, PhD, IS an associate director, all in the Department of Standards Development at the US Pharmacopeia, 12601 Twinbrook Parkway, Rockville, MD 20852- 1790, tel. 301 -81 6.831 9, hpBusp.org. Ta whom all correspondence should be addressed. C alibration, validation, and traceability are central components of metrology (1-3). The process of validating a new analytical procedure for compendial urage is addressed in US Phamacopeia (USP) General Chapter {I 225) "Validation of Compendial Proceduresm* (4). Even though a USP procedure is fully vali- dated, one inay not have assurance that the pro- cedure is snitable for use with a specific ingre- dient or product, in a specific laboratory, with specific laboratory personnel, equipment, and reagents. The latter falls within the area termed verj;fication, whereby analysts demonstrate that a validated procedure to be used for the first time with a particular product will yield accept- able results using a given laboratory's equip- ment, personnel, and reagents. Therefore, the purposc of the proposed General Chapter (1226) "Verification of Compendial Procedures'" (5) is to provide guidance about the verifica- tion process. USP developed this chapter in response to industry's requests to provide illstructions for verifying compendial procedures. The authors note that verification is not mandatory unless an enforcing body adopts it by reference, it is referenced in a specific USP monograph, or it is adopted by a specific manufacturer in stan- dard operating procedures subject to current good manufacturing practices (CGMPs) regu- lations (see also General Chapters, USP Gen- eral Notices). A compendia1 procedure is considered vali- dated if it is published as official text in USP-National Formulary (NF), in a supple- ment, or as an Interim Revision Announcement in Pharrnacopeial Forum (PFj. Therefore, users of such procedures are not required to perform validation studies. But, they muss demonstrate that the use of the official procedure is suitable given the actual conditions of use. This require- ment is established by the US Food and Drug Administration's CGMPs, which state "The snit- ability of all testing merhods used shall be ver- ified under actual conditions of use" (6). The suitability of a cornpendial procedure may be an issue for an article under test for sev- eral reasons (e.g., different impurity profiIes from different routes of synthesis, composition of formulations, or interference from excipi- ents). General Chapters already exist in the USP on rhe general topic. For example, various Gen- eral Chapters for microbiological testing de- scribe apprnaches that will ensure the article to be tested is suitable for use with the validated procedure. Similarly, many USP monographs have sections devoted TO establishing the suit- ability of chromatographic systems. If multiple laboratories are expected to use the cornpendial procedure with essentially the same conditions of actual usage, then one lab- oratory may perform the verification. With ap- propriate verification of adequate robustness, other locations may use the verified procedure. Some factors that must be considered when as- certaining usage conditions are: article to he tested; laboratories to be used; equipment to be used; personnet - reagents. Verification process It is assumed that personnel involved in the ver- ification process possess appropriate experi- ence, knowledge, and training to execute a ver- ification properly. A critical step in the verification process is the preparation of a doc- ument describing: procedures to be verified; number and identity of lots or batches of ar- ticles that will be used (note that the verifi- cation may need to be repeated for different types of articles); analytical performance characteristics to be 7 h e proposal to replaoe "Methods" in the title with "Procedures" was made in Pharmacopeial Forum 31 (2) and will become official in USP29 (Suppl. 1).

Transcript of Verification of Compendial Procedu. - Microbiology … · Verification of Compendial Procedu. --...

Verification of Compendial Procedu. -- Horacio Pappa, David Porter,and Karen Russo

Horacio Pappa, PhD,* IS a senlor scientist, D a v i d Porter, PhD, is a dirxtor, and Karen Russo, PhD, IS

an associate director, all in the Department of Standards Development at the US Pharmacopeia, 12601 Twinbrook Parkway, Rockville, MD 20852- 1790, tel. 301 -81 6.831 9, hpBusp.org.

Ta whom all correspondence should be addressed.

C alibration, validation, and traceability are central components of metrology (1-3). The process of validating a new analytical

procedure for compendial urage is addressed in US Phamacopeia (USP) General Chapter {I 225) "Validation of Compendial Proceduresm* (4).

Even though a USP procedure is fully vali- dated, one inay not have assurance that the pro- cedure is snitable for use with a specific ingre- dient or product, in a specific laboratory, with specific laboratory personnel, equipment, and reagents. The latter falls within the area termed verj;fication, whereby analysts demonstrate that a validated procedure to be used for the first time with a particular product will yield accept- able results using a given laboratory's equip- ment, personnel, and reagents. Therefore, the purposc of the proposed General Chapter (1226) "Verification of Compendial Procedures'" (5) is to provide guidance about the verifica- tion process.

USP developed this chapter in response to industry's requests to provide illstructions for verifying compendial procedures. The authors note that verification is not mandatory unless an enforcing body adopts it by reference, it is referenced in a specific USP monograph, or it is adopted by a specific manufacturer in stan- dard operating procedures subject to current good manufacturing practices (CGMPs) regu- lations (see also General Chapters, USP Gen- eral Notices).

A compendia1 procedure is considered vali- dated if it is published as official text in USP-National Formulary (NF), in a supple- ment, or a s an Interim Revision Announcement in Pharrnacopeial Forum (PFj. Therefore, users of such procedures are not required to perform validation studies. But, they muss demonstrate that the use of the official procedure is suitable given the actual conditions of use. This require- ment is established by the US Food and Drug Administration's CGMPs, which state "The snit- ability of all testing merhods used shall be ver-

ified under actual conditions of use" (6). The suitability of a cornpendial procedure

may be an issue for an article under test for sev- eral reasons (e.g., different impurity profiIes from different routes of synthesis, composition of formulations, or interference from excipi- ents). General Chapters already exist in the USP on rhe general topic. For example, various Gen- eral Chapters for microbiological testing de- scribe apprnaches that will ensure the article to be tested is suitable for use with the validated procedure. Similarly, many USP monographs have sections devoted TO establishing the suit- ability of chromatographic systems.

If multiple laboratories are expected to use the cornpendial procedure with essentially the same conditions of actual usage, then one lab- oratory may perform the verification. With ap- propriate verification of adequate robustness, other locations may use the verified procedure. Some factors that must be considered when as- certaining usage conditions are:

article to he tested; laboratories to be used; equipment to be used; personnet - reagents.

Verification process It is assumed that personnel involved in the ver- ification process possess appropriate experi- ence, knowledge, and training to execute a ver- ification properly. A critical step in the verification process is the preparation of a doc- ument describing:

procedures to be verified; number and identity of lots or batches of ar- ticles that will be used (note that the verifi- cation may need to be repeated for different types of articles); analytical performance characteristics to be

7 h e proposal to replaoe "Methods" in the title with "Procedures" was made in Pharmacopeial Forum 31 (2) and will become official in USP29 (Suppl. 1).

Inside USP

evaluated, along w i t h a spzi i t ic, i t io l~

of the acceptable range of results ex- pressed in terms of acceptable accu- racy, precision, or linearity parame- ters; justification of any deviations from the recommendations in General Chapter (1226). A critical part of the process is to pre-

establish acceptance criteria for the veri- fication attributes being tested. These cri- teria vary with the intended use of the procedure. Upon completion of sample analysis, laboratory personnel should compare the data obtained with the pre- established acceptance criteria. l%e ensu- ing documentation shorsld include a sim- mary of analytical data, a comparison of the data with acceptance criteria, and a conclusion about the suitability of the d- idated pmcedure under actual usage cnn- ditions. If the data meet the acceptance criteria, it may be assumed that the com- pendial procedure will perform as in- tended under actual usage conditions.

If the analytical data do not compare

recision

Spectro- photometrld Precision colorimetric

Titrimetric Precision

TLC

Gas electro- phoresis

Category I!

Quantitative

Precision, speciffcity.

quantltatlon. limit

Precision, quantitation,

Firnit

Specificity, quantitation,

limit

Specificity.

Limit tests

Specificity. detection,

limit

Specificity, detection,

limit

Specificity, detection,

lirnlt

Specificity. detect~on.

limit

Categow Category 111 IV

N!A I Specificity

NIA Specificity ! "Reproduced w~th permlsslon from the US Pharrnacopela Abbrev~at~ons: HPLC-GC IS high-performance liqu~d chromatography-gas chromatography. NIA 1s not appl~cable TLC IS th~n-layer chromatography

favorably with the prc-established ac- propriate), which may include rectify- ceptance criteria, one should try to de- ing training deficiencies or contacting termine the sources of the deviations USP personnet for craxification of the and initiate corrective action (when ap- procedure. Any corrective actions should

is just what the doctor ordered.

COMPANIES (281)530-8712

~ n m r . mund ycos .corn

I I -- 166 PhamracmtiEal Technolagf M A ~ 2006

be indicated in a revised verification ap- proval document. The final document and its components should contain the initial unacceptable resuIts.

Despite repeated attempts to perform the procedure and obtain acceptable re- sults, the results may demonstrate that the procedure is not suitable for appli- cation under specific usage conditions. In such a case, it may be necessary to modify the official compendial proce- dure (in which casc, suggestions for im- provement should be submitted to USP), develop and validate an alterna- tive procedure, change the matrix to ovcrcome potential interferences, or make other changes.

Tf an alternative procedure Is validated, the new procedure and supporting vali- dation data can be submitted to USP for consideration as part of a flexible mono- graph. Flcxible monographs allow alter- native tests, procedures, and acceptance criteria to account for control of quality attributes that may differ between arti- cles of the samc cornpcndial name. If a compendial procedure is deemed unsuit- able after consultation with USP scien- tists, the11 the final verification document should summarize the steps to dernon- strate suitability of the new procedure and to describe the final action taken in dc- veloping the alternative procedure.

The characteristics chosen for use in the verification prOCESS typically are de- rived from those used in validation stud- ies. %ble I1 in General Chapter (1225) providcs a listing of the performance characteristics that should be assessed based in different analytical categories. Caregory 1 encompasses procedures in- tended to quantify major components of drug substances in finished dosage forms. Category J l covers procedures in- tended for use as either quantitative or limit testing for impurities. Category IIT refers to procedures such as dissolution intended to assess dosage-form per- formance. Category lY includes identi- fication procedures. All thc characteris- tics in Table 11 are not required for procedure verification.

Table I in proposed General Chapter (1226) lists the data elements that should be evaluated for procedure verification

for drug substances and excipients (sec Table I).

Table IT in proposed General Chapter (1 226) lists the data elements that should be evaluated for procedureveritication for do~age forms (see Tahle 11). The number of tests for the verification of procedures for use with drug products is larger for several categories hecause of the increased conlplmity of drug products, in compar- ison with drug substances and excipients.

Additional performance characteris-

tics may be appropriate for verification purposes, For ex;anple, if a pharmaceu- tical manufacturer wants to make delib- erate modifications in a procedure in terms of analyst, equipment, reagent, or environmental condition, then robust- ness studies are appropriate. Definitions for robl~mess are listed in General Chap- ter (1225). In particular cases, the stabil- ity of the sample preparation should be investigated because different matrices can affect the stability of the solution.

," ,o- -t1c, 1 ~ ! ~ ' ~ - r ! ~ ~ l ~ / ~ ~ l r ; ~ ~ l ~ (>xl)(>-t\ 0 thr OPTIMA Grot113 whcn i t corner, to industry srllutions for production, iilling, ~ n t l pnckae~ng cji phnrma- crwtical prnclr~cts. Siat~clhrd inn- c h i n ~ s or innovntivc conccpts for syringes, boltlps, v~nts, or c:u<tr)m dcviccs. Cornplerc, modular l i n r l for thy t1trnor;t in flcui1,ilitv.

In addll~or), s t ~ ~ ~ E ~ z n l ) l ~ f r t -1~7~ cly- rrs, loarl~ng anrl i~nloacl~ng cystems,

automatrt~n technoloyv, IT, and production clnta acqi~~si imn.

The OPTlhl11 Lr r~r rp qtanrl~ fnr rxpcricncr, svncrg~rs, ancl ServIw worlrfcvidp. Int l r ~ d ~ n ~ branches in liic U.S , Hraz~l. Mrxrct,, Kor~a, japan, and Grrat Rrilain.

INOVA KUGLEK KI-FE IMLD~CON AUTOCON / n"*'kl& packacing group I;rrlI>H , Stt.inhrikwrg 20

1 St-h~$.ihl<rh Hall - Gcrmanv - \t~wiv.c~lit~n~.i p,rr~~rp.<lc.

'44 .12ar~hlr>vrv I-r)rpor;rtlnn - 133(1 Ccrntrac: 1)rlr.e

1 R J L . !3'1 i 1 JO4 $rw\$ o~)ljrlil~-l~rl~.c 13m / P h a r n S ~ t ~ f i ~ I b?Chu/ogy MARCH mos f 67

Inside USP

Status of the new chapter After General Chapter (1226) was pub- lished in PF, USP received numerous comments horn pharmaceutical man- ufacturers and regulators. As specified in the briefing, the deadline was July 15, 2005. USP deferred this deadline in an- ticipation of discussions at the USP An- nual Scientific Meeting in September 2005. A summary of the comments and USP responses follows.

Some respondents indicated that by including elements of validation in the verification process, USP is calling into question one key principle (and advan- tage) of a compendia1 procedure: it is robust and validated. It should be un- derstood that the proposed chapter is not focused on challenging the validity of the procedure per se, but rather the suitability of the analyticid system (i.e., a matrix of factors such as the article to be tested, the laboratories to be used, the equipment to be used, the person- nel involved, and reagents used, etc.).

Other respondents suggested that the

Tflehnique Category l

Ptecrsron, spec~ficity,

-- . quantitatior , I ranqe limit

Spectro- photome 1 I,

colnrimetr~c 1 range l~mrt

cis~on, arity, I

I ge

. - -. . . . .

HPLC-G limit

'recision, Spec detet limit

quantitatior . .. . . . - . . -

Titrimetri c line I P:' 'recision

Specificity, Spec detet limit

Specif icit TLC

Specificit Gas elec phoresis

, cea wlrn perrnlssron rrom rne us rnarrnacopela. Abbrwiations: HPLC-GC is high-performance liquid chromatography-gas chromatography. NIA is not applicable. TLC is thin-layer chromatography

range of results that will be considered for the first time will have no way of acceptable represents knowledge typi- knowing in advance what results would cally avahble only to the group that pa- be acceptable4ther than those defined formed the original validation. Thus, a by the specification range. USP believes laboratory testing a compendia1 article that acceptance criteria verification

Carl Zeis~: Your par tne~ in PAT.

aquasyn

Better process unaerstanalng leads to successrur YAI

~rnplementat~on From lab to plant, Carl Ze~ss Corona

Dryer and Blender and the MCS 600 rnult~chann~l

frber OPTIC ~nnovative spectrometer systems, coverlng - UV-VIS-NIS w~ll help you generate better results for

on-l~ne powder and l~qu~d analys~s And for real tlme

data and trends, use our new ProcessXplarer wh~ch

1s 2 1 CFR compl~ant IP65. GI l ied enclo- sures are ava~lable.

Call I 1T solution tailored to your spec~ftc

ptoo iltion.

11P and Ex r; Aquasyn has developed the ultimate high purrty valve for the biopmcess and pharrnaceut~caf industr~es. The complete product line spec~al~zes in rnnovative and hybrid valve assemblies, setting a new standard In engineering, destgn and manufacturing.

us for a PI

25s applio

When you want it done rfgM and an tlme ... the Iirst time!

Carl Zeiss M~crolma 5pectml 5ensors Thornwoad. NY 105' l Y Y l v Y I C I J , C Y I I I J lkIrL_LI(ll We make r t vlrlble

See Us at INTERPHEX, Booth 2705 i ~ ~ n r w , o h a r m t e c h , c o m

ging Inc. : Aquasyn California wc Tel(8113) 350-0423 Fax (81 8) 350-0424

www.aquasyn corn email sales 0 aouasvn.com

See Us at ~ ~ T E R P H E X , Booth T380 P h ~ e f l ~ I T#hRoIwy MARCH 2006

should be selected according to a risk-based approach according to the in-tended use of the procedure and justi-fied in the verification protocol.

Another respondent stated concernrelated to the issueof "grandfathering"because numerous official procedureshave been in place for years and havebeen used without being subjected toformal verification procedures. Somerespondents asked that a provision beadded indicating that the intent of theproposed chapter is to provide guidancefor verifying the suitability of newlyadopted proceduresand not for retroac-tive application to those being per-formed successfully. USP agrees withthis concern and will provide publicstatement to support itsprospectiveuse.

Some respondents suggested that anorganization performing a verificationexerciseshouldhavethe freedomto doc-ument its process in a defensible man-ner.These pharmaceutical manufactur-ers were concerned that words like"typically" and "may" sometimes arereinterpreted, particularly in regulatorycontexts, as "must." Further, these re-spondents suggestedthat submission ofinformation associatedwith the resolu-tion of verification issues should be atthe discretion of the pharmaceuticalmanufacturer that was performing theverification activities.Asa variation onthese observations, some respondentsnoted that contract laboratoriesperformnumerous procedures for different ma-terials submitted by client companies.In an outsourcingcontext,it maybeverydifficult to verify the suitability of thecompendial procedure for a largenum-ber of articles. USP generally supportsthese observations and notes (as statedabove) that the proposed chapter doesnot specifya requirement exceptunderthe circumstances noted. Manufactur-ers may use all,none, or part of the pro-posed chapter, when finalized, at theirdiscretion.

Finally,some respondents suggestedthat it is inappropriate to combine drugsubstances and excipients in the sametable because the types and require-ments of procedures for these two verydifferent articlescan vary broadly. USP

notes this point and will attempt to clar-ify the distinction as the proposed chap-ter moves to finalization.

ConclusionProcedures published as officialtext inthe USP-NF are considered validated.Therefore, the user of an officialproce-dure does not haveto perform a fullval-idation study. But, CGMPs require thedemonstration that a compendial pro-cedure can be used successfully for agiven article, a processknown as verifi-cation. USP'sproposed new general in-formation chapter (1226)"Verificationof Compendial Procedures"is intendedto fill the gap in the proper usage ofcompendial procedures by outlining aprocess for verifyingtheir suitability.

A newversion of (1226) is scheduledto appear in the May-June issue of PF.USP is amenable to the receipt of com-ments about chapters and monographsthat are in development or are official.USP encourages industry and others toreview the chapter and submit com-ments. Finalization of the proposedchapter is the General Chapters ExpertCommittee's responsibility.

Referencesl. W. May et al., Definitions of Terms and

Modes Usedat NIST for Value-Assignmentof Reference Materials for Chemical Mea-surements (National Institute of Standardsand Technology [NISTJ, Gaithersburg,MD, Jan. 2000).

2. D.L. Duewer et al., An Approachto theMetrologicallySound TraceableAssessmentof the Chemical Purity of OrganicReferenceMaterials (NIST, Gaithersburg, MD, Sept.2004).

3. EURACHEM/CITAC Guide. QuantifyingUncertainty in Analytical Measurement.S.L.R. Elllison, M. Rosslein, and A.Williams, Eds. 2d ed. (NIST, Gaithersburg,MD, April, 2000).

4. Pharm. Forum. 31 (2),549 (2005).

5. General Information Chapter (1226) "Ver-

ification of Compendial Procedures,"Pharm. Forum. 31 (2),555-558 (2005).

6. 21 CFR 211.194(a)(2). PT

-- ---

BULK PRODUCTIONOF API & CONTROLLED

SUBSTANCES

We provide the strengthand know-how behind yourgeneric and branded prod-ucts in the US.

DEA registered facilitieshouse strictly controlled,self contained reactortrains (up to 2,000 gallons;7,500 litres)

Phase II, Phase III throughcommercial bulk API's spe-cializing in Class I-V con-trolled substances andoncology products undercGMP conditions.

Products include: cisplatin,carboplatin, fentanyl citrateand base, hydrocodone,bitartrate, hydromorphoneHCI and oxycodone HC!.

Johnson MattheyPharTTlaceuticalMaterials- USA

2003 Nolte Drive

West Deptford,NJ 08066Tel: 1-800-444-8544

www.jmpharTTla.com