VDSP Reference Measurement System: The Role of Accuracy-Based EQA/PT
Schemes
Graham Carter Imperial College Health Care NHS Trust
www.deqas.org
![Page 2: VDSP Reference Measurement System: The Role of Accuracy ... · Mean inter-laboratory imprecision (CV%) of 25OHD results; distribution cycles since 1995. Use of a common standard by](https://reader030.fdocuments.us/reader030/viewer/2022040616/5f1698c5c7f1c444ab266752/html5/thumbnails/2.jpg)
DEQAS and the VDSP
• The ODS has funded NIST to assign values to DEQAS samples.
• DEQAS is an ‘accuracy’ based Proficiency
Testing scheme (from April 2013)
![Page 3: VDSP Reference Measurement System: The Role of Accuracy ... · Mean inter-laboratory imprecision (CV%) of 25OHD results; distribution cycles since 1995. Use of a common standard by](https://reader030.fdocuments.us/reader030/viewer/2022040616/5f1698c5c7f1c444ab266752/html5/thumbnails/3.jpg)
The Role of Accuracy based 25-OHD EQA/PT Schemes
To monitor over time: • The accuracy of individual results • Inter-laboratory imprecision • Trueness of 25-OHD methods EQA/PT samples must be commutable
![Page 4: VDSP Reference Measurement System: The Role of Accuracy ... · Mean inter-laboratory imprecision (CV%) of 25OHD results; distribution cycles since 1995. Use of a common standard by](https://reader030.fdocuments.us/reader030/viewer/2022040616/5f1698c5c7f1c444ab266752/html5/thumbnails/4.jpg)
Accuracy of results: DEQAS sample 440 (July 2013)
Target value 48.4 nmol/L (ALTM 51.3 nmol/L)
![Page 5: VDSP Reference Measurement System: The Role of Accuracy ... · Mean inter-laboratory imprecision (CV%) of 25OHD results; distribution cycles since 1995. Use of a common standard by](https://reader030.fdocuments.us/reader030/viewer/2022040616/5f1698c5c7f1c444ab266752/html5/thumbnails/5.jpg)
DEQAS October 2012; Effect on Mean % Bias of Switching Performance target from ALTM to NIST assigned values
![Page 6: VDSP Reference Measurement System: The Role of Accuracy ... · Mean inter-laboratory imprecision (CV%) of 25OHD results; distribution cycles since 1995. Use of a common standard by](https://reader030.fdocuments.us/reader030/viewer/2022040616/5f1698c5c7f1c444ab266752/html5/thumbnails/6.jpg)
DEQAS results; October 2012 –July 2013
![Page 7: VDSP Reference Measurement System: The Role of Accuracy ... · Mean inter-laboratory imprecision (CV%) of 25OHD results; distribution cycles since 1995. Use of a common standard by](https://reader030.fdocuments.us/reader030/viewer/2022040616/5f1698c5c7f1c444ab266752/html5/thumbnails/7.jpg)
ALTM vs NIST; October 2012 –July 2013
![Page 8: VDSP Reference Measurement System: The Role of Accuracy ... · Mean inter-laboratory imprecision (CV%) of 25OHD results; distribution cycles since 1995. Use of a common standard by](https://reader030.fdocuments.us/reader030/viewer/2022040616/5f1698c5c7f1c444ab266752/html5/thumbnails/8.jpg)
Mean inter-laboratory imprecision (CV%) of 25OHD results; distribution cycles since 1995
![Page 9: VDSP Reference Measurement System: The Role of Accuracy ... · Mean inter-laboratory imprecision (CV%) of 25OHD results; distribution cycles since 1995. Use of a common standard by](https://reader030.fdocuments.us/reader030/viewer/2022040616/5f1698c5c7f1c444ab266752/html5/thumbnails/9.jpg)
Use of a common standard by LC-MS/MS users; effect on inter-laboratory imprecision (CV%)
Total 25-OHD
![Page 10: VDSP Reference Measurement System: The Role of Accuracy ... · Mean inter-laboratory imprecision (CV%) of 25OHD results; distribution cycles since 1995. Use of a common standard by](https://reader030.fdocuments.us/reader030/viewer/2022040616/5f1698c5c7f1c444ab266752/html5/thumbnails/10.jpg)
Use of a common standard in LC-MS/MS assays – effect on inter-laboratory imprecision (CV%)
25-OHD3
0
2
4
6
8
10
12
14
16
18
20
326 327 328 329 330 C1 C2
Inte
r-la
bora
tory
impr
ecis
ion
CV
%
In-house standards Chromsystems standard
![Page 11: VDSP Reference Measurement System: The Role of Accuracy ... · Mean inter-laboratory imprecision (CV%) of 25OHD results; distribution cycles since 1995. Use of a common standard by](https://reader030.fdocuments.us/reader030/viewer/2022040616/5f1698c5c7f1c444ab266752/html5/thumbnails/11.jpg)
Suggested Assay Performance Limits Based on Biological Variation*
*Stöckl D et al. Clin Chim Acta 2009;408:8-13
Measurements CV (%) Bias (%) Reference Labs ≤ 5% ≤ 1.7% “Routine” Labs ≤ 10% ≤ 5%
![Page 12: VDSP Reference Measurement System: The Role of Accuracy ... · Mean inter-laboratory imprecision (CV%) of 25OHD results; distribution cycles since 1995. Use of a common standard by](https://reader030.fdocuments.us/reader030/viewer/2022040616/5f1698c5c7f1c444ab266752/html5/thumbnails/12.jpg)
25-OHD Methods; % of users submitting results within the suggested bias limits of ± 5% (Sample 436, July 2013)
![Page 13: VDSP Reference Measurement System: The Role of Accuracy ... · Mean inter-laboratory imprecision (CV%) of 25OHD results; distribution cycles since 1995. Use of a common standard by](https://reader030.fdocuments.us/reader030/viewer/2022040616/5f1698c5c7f1c444ab266752/html5/thumbnails/13.jpg)
Trueness (%Bias) of automated Competitive Binding Assays for 25-OHD (1)
![Page 14: VDSP Reference Measurement System: The Role of Accuracy ... · Mean inter-laboratory imprecision (CV%) of 25OHD results; distribution cycles since 1995. Use of a common standard by](https://reader030.fdocuments.us/reader030/viewer/2022040616/5f1698c5c7f1c444ab266752/html5/thumbnails/14.jpg)
Trueness (% Bias) of automated Competitive Binding Assays for 25-OHD (2)
![Page 15: VDSP Reference Measurement System: The Role of Accuracy ... · Mean inter-laboratory imprecision (CV%) of 25OHD results; distribution cycles since 1995. Use of a common standard by](https://reader030.fdocuments.us/reader030/viewer/2022040616/5f1698c5c7f1c444ab266752/html5/thumbnails/15.jpg)
Trueness (% Bias) of LC-MS/MS and HPLC/UV
![Page 16: VDSP Reference Measurement System: The Role of Accuracy ... · Mean inter-laboratory imprecision (CV%) of 25OHD results; distribution cycles since 1995. Use of a common standard by](https://reader030.fdocuments.us/reader030/viewer/2022040616/5f1698c5c7f1c444ab266752/html5/thumbnails/16.jpg)
Imprecision of automated Competitive Binding Assays for 25-OHD (1)
![Page 17: VDSP Reference Measurement System: The Role of Accuracy ... · Mean inter-laboratory imprecision (CV%) of 25OHD results; distribution cycles since 1995. Use of a common standard by](https://reader030.fdocuments.us/reader030/viewer/2022040616/5f1698c5c7f1c444ab266752/html5/thumbnails/17.jpg)
Imprecision of automated Competitive Binding Assays for 25-OHD (2)
![Page 18: VDSP Reference Measurement System: The Role of Accuracy ... · Mean inter-laboratory imprecision (CV%) of 25OHD results; distribution cycles since 1995. Use of a common standard by](https://reader030.fdocuments.us/reader030/viewer/2022040616/5f1698c5c7f1c444ab266752/html5/thumbnails/18.jpg)
Imprecision of LC-MS/MS and HPLC/UV
![Page 19: VDSP Reference Measurement System: The Role of Accuracy ... · Mean inter-laboratory imprecision (CV%) of 25OHD results; distribution cycles since 1995. Use of a common standard by](https://reader030.fdocuments.us/reader030/viewer/2022040616/5f1698c5c7f1c444ab266752/html5/thumbnails/19.jpg)
Summary • The VDSP is designed to standardize 25-OHD
results to the true value (RMP) • Accuracy based PT schemes can (should?) be
used to monitor: • Accuracy of individual 25-OHD results • Inter-laboratory imprecision • Trueness of 25-OHD methods
But can we ever guarantee the accuracy of an
individual result?
![Page 20: VDSP Reference Measurement System: The Role of Accuracy ... · Mean inter-laboratory imprecision (CV%) of 25OHD results; distribution cycles since 1995. Use of a common standard by](https://reader030.fdocuments.us/reader030/viewer/2022040616/5f1698c5c7f1c444ab266752/html5/thumbnails/20.jpg)
Thank you!
![Page 21: VDSP Reference Measurement System: The Role of Accuracy ... · Mean inter-laboratory imprecision (CV%) of 25OHD results; distribution cycles since 1995. Use of a common standard by](https://reader030.fdocuments.us/reader030/viewer/2022040616/5f1698c5c7f1c444ab266752/html5/thumbnails/21.jpg)
Acknowledgements
• DEQAS: Jayne Shannon, Julia Jones
• ODS: Christopher Sempos, Paul Coates
• NIST: Johanna Camara, Lane Sander, Karen Phinney, Stephen Wise,
![Page 22: VDSP Reference Measurement System: The Role of Accuracy ... · Mean inter-laboratory imprecision (CV%) of 25OHD results; distribution cycles since 1995. Use of a common standard by](https://reader030.fdocuments.us/reader030/viewer/2022040616/5f1698c5c7f1c444ab266752/html5/thumbnails/22.jpg)
DEQAS: an accuracy based Proficiency Testing Scheme for 25-OHD assays
Aim of DEQAS is to monitor/improve performance of 25-OHD assays through:
• Quarterly distribution of unadulterated human serum samples (from haemochromatosis patients).
• Calculation of %Bias from a Target value (ALTM until January 2013, NIST values from April 2013)
• Educational/investigative exercises