Validation of the G-185 incubator Ronny Janssens Romy Souffreau 29/06/2009

Transcript of Validation of the G-185 incubator · 16,0 18,0 1 5 9 13 17 21 25 29 33 37 41 Time(min) O2...

Page 1: Validation of the G-185 incubator · 16,0 18,0 1 5 9 13 17 21 25 29 33 37 41 Time(min) O2 concentratie CO2 concentratie 1 CO 2 en O 2 after opening one chamber ... Closed incubator:

Validation of the G-185 incubator

Ronny Janssens – Romy Souffreau

29/06/2009

Page 2: Validation of the G-185 incubator · 16,0 18,0 1 5 9 13 17 21 25 29 33 37 41 Time(min) O2 concentratie CO2 concentratie 1 CO 2 en O 2 after opening one chamber ... Closed incubator:

2 2-7-2013

Validatie K-systems G185 Direct Heat

Incubator

Introduction

Page 3: Validation of the G-185 incubator · 16,0 18,0 1 5 9 13 17 21 25 29 33 37 41 Time(min) O2 concentratie CO2 concentratie 1 CO 2 en O 2 after opening one chamber ... Closed incubator:

3 2-7-2013

Validation

ISO 15189, §5.3.2

“Equipment shall be shown

(upon installation and in

routine use) to be capable of

achieving the performance

required and shall comply

with specifications relevant

to the examinations

concerned”

ISO 15189 §5.5.2

“The laboratory shall use only

validated procedures for

confirming that the

examination procedures are

suitable for the intended use”

Page 4: Validation of the G-185 incubator · 16,0 18,0 1 5 9 13 17 21 25 29 33 37 41 Time(min) O2 concentratie CO2 concentratie 1 CO 2 en O 2 after opening one chamber ... Closed incubator:

4 2-7-2013

TOC

Technical validation

Clinical validation

3PN zygotes

Sibling oocyte study: G-185 vs Heracell240

Randomised Controled Trial

Page 5: Validation of the G-185 incubator · 16,0 18,0 1 5 9 13 17 21 25 29 33 37 41 Time(min) O2 concentratie CO2 concentratie 1 CO 2 en O 2 after opening one chamber ... Closed incubator:

5 2-7-2013

Validation

User requirements specifications Use as safe incubator for IVF

Space saving alternative for standard incubators

Functional specifications Grow viable embryos

Good embryo quality

Easy to use

Alarm functions

Design specifications 37°C ± 0,5°C

6% CO2 ± 0,5%– 5% O2 ± 1%

Stable culture conditions

No humidification

Air quality (HEPA – VOC)

Page 6: Validation of the G-185 incubator · 16,0 18,0 1 5 9 13 17 21 25 29 33 37 41 Time(min) O2 concentratie CO2 concentratie 1 CO 2 en O 2 after opening one chamber ... Closed incubator:

6 2-7-2013

TOC

Technical validation

Pre production model (2007 - 2008)

Final product (2008 - 2009)

Clinical validation (2009)

3PN zygotes

Sibling oocyte study: G-185 vs Heracell240

Randomised Controled Trial (started 26/05/2009)

Page 7: Validation of the G-185 incubator · 16,0 18,0 1 5 9 13 17 21 25 29 33 37 41 Time(min) O2 concentratie CO2 concentratie 1 CO 2 en O 2 after opening one chamber ... Closed incubator:

7 2-7-2013

Technical validation: pre production model

User interface software: bugs removed – extra functionality added

Minor changes in design: improvements on seals – design of gas mixing chamber

Requirements on safety

Development of monitoring software

permanent monitoring and registration of critical parameters (Temp – CO2 – O2)

Audit trail

E-mail messages

Alarm switch connected to modem

Independent monitoring

Installation of Vaisala CO2 sensor in gas mixing chamber

Temperature monitoring?

Page 8: Validation of the G-185 incubator · 16,0 18,0 1 5 9 13 17 21 25 29 33 37 41 Time(min) O2 concentratie CO2 concentratie 1 CO 2 en O 2 after opening one chamber ... Closed incubator:

8 2-7-2013

Monitoring software

Page 9: Validation of the G-185 incubator · 16,0 18,0 1 5 9 13 17 21 25 29 33 37 41 Time(min) O2 concentratie CO2 concentratie 1 CO 2 en O 2 after opening one chamber ... Closed incubator:

9 2-7-2013

Page 10: Validation of the G-185 incubator · 16,0 18,0 1 5 9 13 17 21 25 29 33 37 41 Time(min) O2 concentratie CO2 concentratie 1 CO 2 en O 2 after opening one chamber ... Closed incubator:

10 2-7-2013

Measurements

•Gas sampling in closed chamber

•Temperature: calibrated thermometer with

thermocouple fixed in culture dish

•pH: ISFET probes – suitable for small volumes

RI pH meter

Page 11: Validation of the G-185 incubator · 16,0 18,0 1 5 9 13 17 21 25 29 33 37 41 Time(min) O2 concentratie CO2 concentratie 1 CO 2 en O 2 after opening one chamber ... Closed incubator:

11 2-7-2013

Temperature – closed chambers

Page 12: Validation of the G-185 incubator · 16,0 18,0 1 5 9 13 17 21 25 29 33 37 41 Time(min) O2 concentratie CO2 concentratie 1 CO 2 en O 2 after opening one chamber ... Closed incubator:

12 2-7-2013

Temperature in Nunc 4well dish in opended

chamber

Time opening a chamber

36

36,2

36,4

36,6

36,8

37

37,2

37,4

0 1 2 3 4 5 6Tijd (min)

Tem

per

atu

re (

°C)

10 sec 20 sec 40 sec 60 sec 120 sec 180 sec240 sec 150 sec

Page 13: Validation of the G-185 incubator · 16,0 18,0 1 5 9 13 17 21 25 29 33 37 41 Time(min) O2 concentratie CO2 concentratie 1 CO 2 en O 2 after opening one chamber ... Closed incubator:

13 2-7-2013

Gas concentration (closed chambers)

Average (n=10) Stdev

[O2] [CO2] [O2] [CO2]

Chamber1 4,9 5,9 0,16 0,09

Chamber2 4,8 5,9 0,11 0,11

Chamber3 4,8 5,8 0,12 0,10

Chamber 4 4,9 5,9 0,14 0,13

Chamber 5 4,7 5,9 0,08 0,16

Chamber 6 4,8 6,0 0,13 0,13

Chamber 7 4,7 5,9 0,04 0,05

Chamber 8 4,8 5,9 0,04 0,09

Chamber 9 4,9 5,9 0,07 0,06

Chamber10 4,9 5,9 0,11 0,15

Average 4,8 5,9 0,10 0,11

t-test

O2 CO2

p-waarde 0,00 0,168

Page 14: Validation of the G-185 incubator · 16,0 18,0 1 5 9 13 17 21 25 29 33 37 41 Time(min) O2 concentratie CO2 concentratie 1 CO 2 en O 2 after opening one chamber ... Closed incubator:

14 2-7-2013

Recovery time

0,0

2,0

4,0

6,0

8,0

10,0

12,0

14,0

16,0

18,0

1 5 9 13 17 21 25 29 33 37 41

Time(min)

Concentr

atio

n O

2 e

n C

O2 (

%)

O2 concentratie CO2 concentratie

1

CO2 en O2 after opening one chamber

Fast return to homeostatic conditions: 3 minutes

Page 15: Validation of the G-185 incubator · 16,0 18,0 1 5 9 13 17 21 25 29 33 37 41 Time(min) O2 concentratie CO2 concentratie 1 CO 2 en O 2 after opening one chamber ... Closed incubator:

15 2-7-2013

Impact of opening one chamber on gas

concentrations

↑ ↓

4,5

5,0

5,5

6,0

6,5

7,0

7,5

8,0

0 1 2 3 4 5Tijd (min)

4,0

4,5

5,0

5,5

6,0

6,5

7,0

kamer 1 kamer 2 kamer 3 kamer 4 kamer 5 kamer 6kamer 7 kamer 9 kamer 10 Serie1 Serie11 Serie12Serie13 Serie14 Serie15 Serie16 Serie17 Serie18

not relevant

O2 en CO2 in closed chambers after opening

of one chamber O2(%) CO2 (%)

Page 16: Validation of the G-185 incubator · 16,0 18,0 1 5 9 13 17 21 25 29 33 37 41 Time(min) O2 concentratie CO2 concentratie 1 CO 2 en O 2 after opening one chamber ... Closed incubator:

16 2-7-2013

Impact of frequent opening

Measurement in 1 room and open 1 room every 5 minutes

7,25

7,26

7,27

7,28

7,29

7,30

0 15 30 45 60 75 90 105 120 135

Tijd (min)

pH

4,9

5,1

5,3

5,5

5,7

5,9

6,1

6,3

CO2 en O2 (%)

pH O2 concentratie (%) CO2 concentratie (%)

Page 17: Validation of the G-185 incubator · 16,0 18,0 1 5 9 13 17 21 25 29 33 37 41 Time(min) O2 concentratie CO2 concentratie 1 CO 2 en O 2 after opening one chamber ... Closed incubator:

17 2-7-2013

Particle count

Maximum permitted number of particles per m3

equal to or greater than the tabulated size

At rest In operation

Grade 0.5 μm 5.0μm 0.5 μm 5.0μm

A 3520 20 3520 20

B 3520 29 352000 2900

C 352000 2900 3520000 29000

D 3520000 29000 Not

defined

Not

defined

Volume 4 - EU Guidelines for Good Manufacturing Practice Medicinal Products for Human and Veterinary Use – annex 1 (feb 2008)

Page 18: Validation of the G-185 incubator · 16,0 18,0 1 5 9 13 17 21 25 29 33 37 41 Time(min) O2 concentratie CO2 concentratie 1 CO 2 en O 2 after opening one chamber ... Closed incubator:

18 2-7-2013

Validation G185 Direct Heat Incubator

Requirements

* [O2] ………………. 4,8% 5,0% ± 1%

* [CO2]……………… 5,9% 6,0% ± 0,5%

* Temperatuur…….. 37,1°C 37,0°C ± 0,5°C

* pH ………………... 7,28 ± 7,20 – 7,30

* Class A (GMP)

Page 19: Validation of the G-185 incubator · 16,0 18,0 1 5 9 13 17 21 25 29 33 37 41 Time(min) O2 concentratie CO2 concentratie 1 CO 2 en O 2 after opening one chamber ... Closed incubator:

19 2-7-2013

Technical validation: conclusions

Closed incubator: homogeneous and stable gas and

temperature distribution

After opening:

1. Fast recovery of gas conditions in opened chamber

2. No significant impact on other chambers

Conclusions

Optimal culture parameters are maintained

Chambers can be considered as individual culture

Page 20: Validation of the G-185 incubator · 16,0 18,0 1 5 9 13 17 21 25 29 33 37 41 Time(min) O2 concentratie CO2 concentratie 1 CO 2 en O 2 after opening one chamber ... Closed incubator:

20 2-7-2013

TOC

Technical validation

Preproduction model

Final product

Clinical validation

3PN zygotes (n=20)

Sibling oocyte study: G-185 vs Heracell240

Randomised Controled Trial (started 26/05/2009)

Page 21: Validation of the G-185 incubator · 16,0 18,0 1 5 9 13 17 21 25 29 33 37 41 Time(min) O2 concentratie CO2 concentratie 1 CO 2 en O 2 after opening one chamber ... Closed incubator:

21 2-7-2013

Sibling oocyte study (n=24)

G-185

N (%)

Heracell 240

N (%)

Injected

(inseminated)

126 133

Deg 11 10

1PN 9 2

2PN 86 (68.3) 90 (67.7) NS

3PN 3 2

Page 22: Validation of the G-185 incubator · 16,0 18,0 1 5 9 13 17 21 25 29 33 37 41 Time(min) O2 concentratie CO2 concentratie 1 CO 2 en O 2 after opening one chamber ... Closed incubator:

22 2-7-2013

Embryo development day 3

Cleavage stage D3 G-185

Uncl

MNB

2-4

5-7

8

>8

Cleavage stage D3 Standard

Uncl

MNB

2-4

5-7

8

>8

Embryo quality D3 G-185

A

B

C

D

Embryo quality D3 Standard

A

B

C

D

Page 23: Validation of the G-185 incubator · 16,0 18,0 1 5 9 13 17 21 25 29 33 37 41 Time(min) O2 concentratie CO2 concentratie 1 CO 2 en O 2 after opening one chamber ... Closed incubator:

23 2-7-2013

Embryo replacements - cyropreservation

G-185

Discard

ET

Cryo

Standard

Discard

ET

Cryo

Cycles 24 +hCG

No replacement 2

Standard 4 1

Mixed 10 5

G-185 8 2

Page 24: Validation of the G-185 incubator · 16,0 18,0 1 5 9 13 17 21 25 29 33 37 41 Time(min) O2 concentratie CO2 concentratie 1 CO 2 en O 2 after opening one chamber ... Closed incubator:

24 2-7-2013

RCT

Started: 27/05/2009

Cycles randomised: 100 (26/06/2009)

+ hCG: ?

Page 25: Validation of the G-185 incubator · 16,0 18,0 1 5 9 13 17 21 25 29 33 37 41 Time(min) O2 concentratie CO2 concentratie 1 CO 2 en O 2 after opening one chamber ... Closed incubator:

25 2-7-2013

Conclusions

Design specifications: 37°C ± 0,5°C

6% CO2 ± 0,5%– 5% O2 ± 1%

Stable culture conditions

Functional specifications: Grow viable embryos

Good embryo quality

Easy to use

Alarm functions

User requirements: Use as safe incubator for IVF

Space saving alternative for standard incubators