    VALIDATION OF INCUBATORS

    1.0 OBJECTIVE

    To lay down a procedure for validation of incubator in

    microbiology laboratory.

    2.0 RESPONSIILTY

    Microbiologist / Executive.

    3.0 ACCOUNTABILITY

    Quality Assurance Control

    4.0 PROCEDURE

    4.1 DETAILS OF THE STANDARD THERMOMETER USED FOR

    VALIDATION OF LAB THERMOMETER.

    4.1.1 Type : Liquid - Glass lab Thermometer.

    4.1.2 Place the standard thermometer dipped in glycerin in incubator at

    various locations as mentioned in annexure-II.

    4.1.3 The incubator is set for desired temperature with the knob on

    control panel of incubator

    ` e.g 32.5 0C, 35 0C & 22.5 0C.

    4.1.4 Wait till the temperature reaches at set point & note down the temp.

    displayed on the small screen of incubator and compare this temp.

    with standard thermometer kept near the RTD probe.Likewise check

    the temperature after every 15 minutes up to one hour and note

    down the readings on the format.

    4.1.5 Record any difference between the displayed temp. & temp. showed

    by standard thermometer and set the incubator accordingly

    4.1.6 Frequency of validation : once in a year.4.1.7 The observations are noted in the format of annexure-I

    4.2 PROCEDURE FOR VALIDATION OF D.H.S (HOT AIR

    OVEN):

    4.2.1 Keep all the apparatus wrapped thrice with aluminum

    foil inside DHS as locations shown

    in the diagram.

  • 7/30/2019 VALIDATION OF INCUBATORS.doc

    2/4

    4.2.2 Keep the Spore loaded strips (having spore of B subilis)

    & Endotoxin indicator indicators

    having 10,000EU/vial in DHS(hot air oven) kept in 30ml vial wrapped

    thrice with aluminum foil at locations shown in the diagram keep one

    vial unbaked as PPC. Set the hot air oven at 250C ,wait till the

    temperature reaches upto the set temperature.

    4.2.3 After reaching the set temperature, note the time & temperature,

    hold for one hour

    4.2.4 After completion of depyrogenation cycle ,switch off the

    power supply and take out

    Indicators for testing.

    4.3 Procedure for testing reduction of endotoxin as under.

    4.3.1 Take out the baked endotoxin loaded vial from the DHS.

    4.3.2 Reconstitute 1ml of LRW in the baked vial and transfer

    100l of sample to the

    depyrogenated reaction tube kept in heating block at 370 C and

    add 100l of LAL in the

    same tube in duplicate.4.3.3 Prepare the dilutions for PPC for confirming

    10,000EU/vial .

    4.3.4 Note the gel clot after incubation of one hour

    4.4. Procedure for testing of reduction of prepared spore

    loaded strip(B subtilis):

    4.4.1 Transfer (spore loaded) strip from backed vials and are

    inoculated in100ml sterile SCDmedia and incubate at 30 350C for 7 days to observe for any

    turbidity, if any, report to Manager QC..

    5.0 REASON FOR REVISION

    Harmonization of format.

    6.0 TRAINING:

    Trainer - Head Quality Control

    Trainees-- Quality Control Chemists & AssistantsPeriod -- One day

  • 7/30/2019 VALIDATION OF INCUBATORS.doc

    3/4

    7.0 DISTRIBUTION:

    Certified Copy No. 1 : Head of Department Quality Control

    Original Copy : Head QUALITY ASSURANCE.

    8.0 ANNEXURES:

    Annexure 1. Formats for Instrument validation.

    9.0 REFERENCES:

    USP 25 page no.-1890 & 2251

    ANNEXURE-1

    INSTRUMENT VALIDATION

    . LOCATION Microbiology

    laboratory

    NAME OF

    INSTRUMENT

    Incubator

    MANUFACTURED

    BY

    MODEL No.

    IDENTIFICATION

    No.

    INSTALLED ON

    SET TEMPERATURE = C

    TIME TEMP. RECORDED DISPLAY

    ED TEMP.

    MINIMUMMAXIMUMDEVIATION

    LOCATIONS INSIDE

    THE CHAMBER

    + -

    1 2 3 4 5 6

    INITI

    AL

    TEMP

    C C C C C C C

    C C

    C C

    AFTE

    R 15

    MIN

    C C C C C C C

    C C

    C C

    AFTE

    R 30

    MIN

    C C C C C C C

    C C

    C C

    AFTE

    R 45

    MIN

    C C C C C C C

    C C

    C C

    AFTE

    R 60 C C C C C C CC C

    C C

  • 7/30/2019 VALIDATION OF INCUBATORS.doc

    4/4

    MIN

    LIMIT:- 2% of Set Temperature

    CONCLUSION: The incubator is maintaining / does not maintaining set

    temperature within

    Specified limits.

    VALIDATED BY:

    CHECKED BY:

    DATE:

    DATE:

    NEXT VALIDATION

    MANAGER /QC

    DUE ON DATE

    Posted by Niazi at11:13 PMLabels: VALIDATION OF INCUBATORS

    0 COMMENTS:

    http://pharmaceuticalsops.blogspot.com/2011/01/validation-of-incubators.htmlhttp://pharmaceuticalsops.blogspot.com/search/label/VALIDATION%20OF%20INCUBATORShttp://www.blogger.com/post-edit.g?blogID=647491574162861344&postID=6928900405023891919http://pharmaceuticalsops.blogspot.com/2011/01/validation-of-incubators.htmlhttp://pharmaceuticalsops.blogspot.com/search/label/VALIDATION%20OF%20INCUBATORS